Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K182274

    Validate with FDA (Live)

    Date Cleared
    2018-11-20

    (90 days)

    Product Code
    Regulation Number
    880.5860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K010305

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Orbit Subretinal Delivery System is indicated for microinjection into the subretinal space.

    Device Description

    The Orbit Subretinal Delivery System is designed for microinjection into the subretinal space. The Orbit Subretinal Delivery System is comprised of 2 kits: the Subretinal Access Kit and the Third Arm Kit. All Subretinal Access Kit components are supplied sterile and intended for single-use only and cannot be reused or resterilized. The Subretinal Access Kit contains the following device components: Subretinal Injection Cannula, Ophthalmic Marker, Syringes, and Tubing Set. The assembled Third Arm Kit functions as a "third arm" to stabilize the subretinal injection cannula and free the surgeon's hands during the procedure. The Third Arm Kit (supplied nonsterile) is reusable and must be cleaned and sterilized prior to each use. Together, the Subretinal Access Kit and the Third Arm Kit are to deliver the infusate (e.g., BSS. BSS® PLUS) to the subretinal space as described in the Instructions for Use.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study information based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance
    BiocompatibilityIn accordance with ISO 10993-01 standards (Cytotoxicity, Sensitization, Irritation, Systemic Toxicity)Product performance met all prior established acceptance criteria.
    Sterilization (Injection Cannula Set & Tubing Set)Sterility Assurance Level (SAL) of 10-6 in accordance with ANSI/AAMI/ISO 11137-2:2013Testing demonstrated product performance met all prior established acceptance criteria.
    Sterilization (Syringe Set)In accordance with ISO 11135:2014; EO and ECH residuals within acceptable limits.Testing demonstrated product performance met all prior established acceptance criteria.
    Cleaning and Disinfection (Third Arm Kit)Meets established cleaning and disinfection requirements per FDA Guidance "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling"Confirmed that the Third Arm Kit meets the established cleaning and disinfection requirements.
    Packaging Qualification (Subretinal Access Kit)Whole package physical integrity requirements and seal integrity requirements met.Met.
    Real-time Shelf Life (Subretinal Access Kit)Establishes expiration date.29 month expiration date established.
    Post-Distribution Performance: Syringe SetVisual inspection of device and package integrity testing.All tests passed pre-established test criteria.
    Post-Distribution Performance: Tubing SetLeak pressure, tensile strength, dimensional integrity.All tests passed pre-established test criteria.
    Post-Distribution Performance: Subretinal Injection Cannula SetFlexural stiffness, flow rate, dimensional integrity, assembly tensile strength, and leak pressure.All tests passed pre-established test criteria.
    Simulated Use (Subretinal Delivery Device)Confirmation of saline injection forming a subretinal bleb.Successfully performed; injections formed subretinal blebs.
    Functionality (Third Arm Kit)Verification of stability to the injection cannula during use, dose clip delivery accuracy, and device durability.All tests passed.
    Overall Product Design Requirements & Applicable StandardsMeets all product design requirements and applicable standards.Met.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size:
      • Simulated use evaluation: Eight (8) procedures were conducted in a porcine model.
      • Specific sample sizes for other tests (e.g., biocompatibility, sterilization, packaging) are not detailed in the provided document. The document states "Testing demonstrated product performance met all prior established acceptance criteria," implying that adequate samples were used for these regulatory tests, but the numbers themselves are not specified.
    • Data Provenance: The simulated use test was performed in a porcine model, indicating animal testing. The document does not specify the country of origin for the data, nor does it explicitly state whether it was retrospective or prospective. Given the nature of medical device validation testing, it would generally be prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • The document describes laboratory and animal model testing, not human clinical studies involving expert observers to establish ground truth in the way one might for diagnostic AI.
    • For the simulated use evaluation, the procedures were performed in a porcine model. It is implied that qualified personnel (e.g., surgeons, researchers) carried out these procedures and observed the results (formation of subretinal blebs), but the number and specific qualifications of these individuals are not stated.

    4. Adjudication Method for the Test Set

    • The document describes engineering and simulated use testing, not a scenario where human observers would require an adjudication method for ground truth. Observations from the simulated use evaluation (formation of subretinal blebs) are presumably directly observable outcomes rather than requiring expert consensus. Therefore, an adjudication method in the context of expert consensus is not applicable/described.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    • No, an MRMC comparative effectiveness study was not done. The Orbit Subretinal Delivery System is a physical medical device (a microinjection system), not an AI algorithm or a diagnostic tool that assists human readers.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    • Not applicable. This device is a manual ophthalmic surgical instrument designed for human operators; it is not an algorithm, and therefore, standalone algorithmic performance is not relevant.

    7. The Type of Ground Truth Used

    • For the simulated use evaluation, the "ground truth" was the direct observation of the physical outcome: the successful formation of a subretinal bleb after injection into a porcine model.
    • For other tests like biocompatibility, sterilization, and performance testing, the ground truth refers to established scientific and regulatory standards (e.g., ISO standards, FDA guidance, pre-established test criteria) that define acceptable performance.

    8. The Sample Size for the Training Set

    • Not applicable. The Orbit Subretinal Delivery System is a physical medical device, not an AI model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. As stated above, this is not an AI model requiring a training set.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1