Search Filters

Search Results

Found 12 results

510(k) Data Aggregation

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Profoject™ Low Dead Space Syringe /Profoject™ Tuberculin Syringe is used to inject fluids into, or withdraw fluids from, the body.

    Profoject™ Low Dead Space Syringe with Needle/ Profoject™ Tuberculin Syringe with Needle is used to inject fluids into, or withdraw fluids from, the body.

    Profoject™ Low Dead Space Syringe with Permanent Needle /Profoject™ Tuberculin Syringe with Permanent Needle is intended for use to inject fluid into or withdraw fluids from the parts of body below the surface of the skin.

    Profoject™ Low Dead Space Safety Syringe /Profoject™ Tuberculin Safety Syringe is intended for use to inject fluid into or withdraw fluids from the parts of body below the surface of the skin. The Sharps Injury Prevention Feature (SIPF) is designed to aid in the prevention of needle stick injuries and syringe reuse.

    Device Description

    Profoject™ Low Dead Space Syringe /Profoject™ Tuberculin Syringe: Profoject™ Low Dead Space Syringe /Profoject™ Tuberculin Syringe consists of a barrel with a Luer Slip or Luer Lock connector which comply with ISO 80369-7, a plunger, and a plunger stopper. The device is available in 0.5 mL and 1 mL capacities. It is sterile, single-use, and non-toxic device that operates on the principle of a piston syringe.

    Profoject™ Low Dead Space Syringe with Needle/ Profoject™ Tuberculin Syringe with Needle: Profoject™ Low Dead Space Syringe with Needle/ Profoject™ Tuberculin Syringe with Needle is sterile, single-use device. Each consists of either a Luer slip or Luer lock syringe assembly integrated with a pre-attached needle, and the syringe is available in two capacities (0.5mL and 1mL). For each capacity variant, the pre-attached needle is available in the following gauge specifications: 16G, 18G, 19G, 21G, 22G, 23G, 25G, 27G and 30G.

    Profoject™ Low Dead Space Syringe with Permanent Needle /Profoject™ Tuberculin Syringe with Permanent Needle: Profoject™ Low Dead Space Syringe with Permanent Needle /Profoject™ Tuberculin Syringe with Permanent Needle is sterile, single-use, non-toxic devices that work on the principle of a piston syringe, and the syringes are available in three capacities: 0.3mL, 0.5mL, and 1mL. Each capacity variant is equipped with a permanently attached needle, available in the following gauge specifications: 21G, 23G, 25G, 26G, 27G, 28G, 29G, 30G and 31G.

    Profoject™ Low Dead Space Safety Syringe /Profoject™ Tuberculin Safety Syringe: Profoject™ Low Dead Space Safety Syringe /Profoject™ Tuberculin Safety Syringe is sterile, single-use, non-toxic devices that work on the principle of a piston syringe, and the syringes are available in three capacities: 0.3mL, 0.5mL, and 1mL. Each capacity variant is equipped with a permanently attached needle, available in the following gauge specifications: 21G, 23G, 25G, 26G, 27G, 28G, 29G, 30G and 31G.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K250127

    Validate with FDA (Live)

    Device Name
    TSK Syringe
    Date Cleared
    2025-07-30

    (194 days)

    Product Code
    Regulation Number
    880.5860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TSK Syringe is a low dead volume syringe intended for use in patients that need injection and withdrawal of substances for examination, treatment, diagnosis or prevention.

    Device Description

    The TSK Syringe is a syringe intended for use with a needle to inject or withdraw substances, for general use. The TSK Syringe leaves a low volume of fluid when the plunger is fully depressed.

    There are two model numbers for the TSK Syringe. The technological differences are as follows:

    NumberSyringe Nominal Capacity (mL)
    TSY-0071P0.7
    TSY-0101Y1.0
    AI/ML Overview

    The provided text is a 510(k) Clearance Letter and 510(k) Summary for the TSK Syringe. It details the device's characteristics, indications for use, and a comparison to a predicate device. However, it does not contain the specific information required to answer your request about acceptance criteria and a study proving device performance in the context of an AI/human reader study.

    This document describes a physical medical device (syringe), not a software or AI-powered medical device that would involve AI performance metrics, human reader studies, ground truth establishment, or training/test set sample sizes in the way you've outlined.

    Therefore, I cannot provide the requested information from this document. The details about acceptance criteria and study design for AI-based performance evaluation are not present.

    Ask a Question

    Ask a specific question about this device

    K Number
    K250138

    Validate with FDA (Live)

    Date Cleared
    2025-04-02

    (75 days)

    Product Code
    Regulation Number
    880.5860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Small Volume 0.2mL Syringe is intended to be used for medical purposes to inject fluids into or withdraw fluids from the body.

    Device Description

    The Small Volume 0.2mL Syringe is intended for manual, single use. It is a small volume 0.2 mL syringe. It is comprised of a PP barrel, ABS plunger, TPE plunger stopper, PP plunger shield and PP Luer protective cap. The device has a Luer lock connection type that enables connection to a hypodermic needle for liquid injection or withdrawal. The device has a plunger stopper to prevent any liquid residual from remaining in the syringe barrel, maximizing the efficiency of liquid usage. The feature is called Low Dead Space/Volume. The Luer protective cap serves to safeguard the small Luer connector from potential damage during transportation. Similarly, the plunger shield is utilized to ensure the small plunger remains undamaged during transportation. The Luer protective cap and the plunger shield are optional to be used. The device has been manufactured without any silicone oil. The device has six models, each equipped with a 0.2 mL Luer lock syringe. Among these models, three include an additional feature for enhanced dosing technology (EDT), which comprises a colored reference line on the plunger surface to improve visibility. These include identifying markings to distinguish the EDT models from the regular 0.2 mL syringe.

    AI/ML Overview

    The provided FDA 510(k) clearance letter and summary for the "Small Volume 0.2mL Syringe" do not contain any information related to AI/ML or software-driven medical devices. The submission is for a physical medical device (syringe) and, as such, the acceptance criteria and study descriptions are entirely focused on the physical and material properties, performance, biocompatibility, and sterility of the syringe itself.

    Therefore, I cannot provide details on:

    • A table of acceptance criteria for AI/ML performance
    • Sample sizes for a test set or data provenance for AI/ML
    • Number of experts and their qualifications for establishing AI/ML ground truth
    • Adjudication methods for AI/ML test sets
    • MRMC comparative effectiveness study for AI with human readers
    • Standalone AI algorithm performance
    • Type of ground truth for AI/ML
    • Training set sample size for AI/ML
    • How ground truth for AI/ML training was established

    The document explicitly states "9. Clinical testing: Not applicable for this submission." This further confirms that no studies involving human interaction or clinical performance for a software/AI component were conducted or required for this purely mechanical device.

    If the request was intended to be about the acceptance criteria and studies for the physical device as described in the 510(k) summary, here is the relevant information extracted:


    Acceptance Criteria and Study for the "Small Volume 0.2mL Syringe" (Physical Device)

    The device, a "Small Volume 0.2mL Syringe," is a physical medical device. The acceptance criteria and the studies proving it meets these criteria are related to its mechanical performance, material safety, and sterility, in accordance with recognized medical device standards. There is no AI/ML component described in this 510(k) submission.

    1. Table of Acceptance Criteria and Reported Device Performance

    The 510(k) summary references compliance with specific standards, rather than providing explicit numerical acceptance criteria directly in a table. The device performance is generally stated as "met the standards' requirements."

    CategoryAcceptance Criteria (Standard Reference)Reported Device Performance
    Material SafetyComplies with ISO 10993-1: - Cytotoxicity - Irritation - Sensitization - Acute systemic toxicity - Hemocompatibility - Pyrogenicity - Particulate Matter per USP <788> - Bacterial Endotoxins per USP <85>All tested endpoints met the acceptance criteria (Implied: complied with ISO 10993-1, USP <788>, and USP <85>).
    Syringe PerformanceComplies with ISO 7886-1:2017 (e.g., accuracy, force, freedom from leakage, residual volume)Met all necessary requirements of the standard.
    Luer Connector PerformanceComplies with ISO 80369-7:2016 and ISO 80369-20:2015 (e.g., connection integrity, leakage)Met the standards' requirements.
    Dead Space Specification≤ 0.0058 mL (95% confidence/ 95% reliability)Meets requirements (stated as "a lower dead space volume is achieved" compared to predicate).
    SterilizationComplies with ISO 11135 (EO sterilization process validation)Validation results confirm compliance.
    Ethylene Oxide (EO) & Ethylene Chlorohydrin (ECH) ResidualsDo not exceed limits according to ISO 10993-7Residues do not exceed the limits.
    Shelf-LifeDevice functionality maintained through 3 years (accelerated aging per ASTM F1980-16)Demonstrated device functionality through the claimed shelf-life of 3 years. (Real-time testing is ongoing to confirm).
    Packaging IntegrityComplies with ISO 11607-1, ISO 11607-2, ASTM D4169-22, ASTM F88/F88-15, ASTM F1929-15All packaging was deemed acceptable for protection of product and sterility maintenance.
    Sterile Barrier TestingComplies with USP <71>Confirmed compliance.

    2. Sample Size Used for the Test Set and Data Provenance

    The exact sample sizes used for each specific test (e.g., biocompatibility endpoints, performance testing per ISO 7886-1) are not detailed in the 510(k) summary. The summary generally states that testing was conducted on "samples" or "unconditioned samples as well as samples aged... and samples transit conditioned."

    • Data Provenance: The studies were conducted by the manufacturer, Prosum Medical Limited, and their testing partners, likely in the United Kingdom or through contract laboratories that adhere to international standards. The data is prospective as it involves new testing specifically for this device submission.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    Not Applicable. For a physical device like a syringe, "ground truth" as it relates to expert consensus for image interpretation or clinical outcomes is not relevant. The ground truth here is derived from objective measurements and compliance with established engineering and biological standards. Testing is performed by qualified laboratory personnel following validated protocols.

    4. Adjudication Method for the Test Set

    Not Applicable. Adjudication typically applies to subjective assessments (e.g., expert reads). For objective measurements on a physical device, statistical analysis of measured values against predefined acceptance limits determines compliance.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and its effect size

    Not Applicable. MRMC studies are used for assessing the impact of AI/software on human performance in diagnostic tasks and do not apply to a physical syringe device.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not Applicable. This refers to AI algorithm performance, which is not part of this submission.

    7. The type of ground truth used

    The "ground truth" for this device's performance and safety is based on objective, quantifiable measurements and compliance with internationally recognized consensus standards (e.g., ISO 7886-1 for syringe performance, ISO 10993-1 for biocompatibility). These standards define acceptable ranges for various parameters (e.g., dead space, force to push plunger, leakage, material toxicity).

    8. The sample size for the training set

    Not Applicable. There is no "training set" as this is not an AI/ML device.

    9. How the ground truth for the training set was established

    Not Applicable. There is no "training set" as this is not an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Luer Lock/Slip Tip Syringe (Low Dead Space) without needle is to inject fluids into, or withdraw fluids from, the body.
    The Luer Lock/Slip Tip Syringe (Low Dead Space) with exchangeable is used to inject fluids into, or withdraw fluids from, the body.
    The Luer Lock Syringe (Low Dead Space) with Safety needle is used to inject fluids into, or withdraw fluids from, the body. The safety mechanism covers the needle after use. In the activated position, the protective arm guards against accidental needle stick during normal handling and disposal of the used needle/syringe combination.

    Device Description

    The Luer Lock and Slip Tip syringe (Low Dead Space) without Needle is a sterile, single-use, standard 3-piece piston hypodermic syringe.
    The Luer lock and Slip Tip syringe (Low Dead Space) with exchangeable needle is a sterile, single-use, standard 3-piece piston syringe with a detachable pre-attached needle.
    The Luer Lock Syringe (Low Dead Space) with Safety Needle is a Low Dead Space syringe in combination with a safety needle. The safety needle contains a mechanism that covers the needle point after use. In the activated position, the needle cover guards against accidental needle stick during normal handling and disposal of the used needle/syringe combination.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification from the FDA for a medical device (syringes). It details the device description, indications for use, comparison to a predicate device, and summaries of non-clinical testing.

    However, it does not contain information related to an AI/ML powered device, acceptance criteria or a study proving that an AI/ML meets these criteria.

    Therefore, I cannot fulfill your request to describe the acceptance criteria and the study that proves the device (AI/ML based) meets the acceptance criteria using this document.

    The document discusses performance testing against established ISO standards for medical devices like sterile hypodermic syringes and needles, biocompatibility testing, sterility, and shelf life for physical devices, not AI/ML model performance.

    Ask a Question

    Ask a specific question about this device

    K Number
    K232943

    Validate with FDA (Live)

    Date Cleared
    2024-05-31

    (254 days)

    Product Code
    Regulation Number
    880.5860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This device is intended for use to inject fluids into or withdraw fluids from the body. The needle protection device covers the needle after use to help prevent needle sticks.

    Device Description

    The Hypodermic Needle-Pro® EDGE™ Safety Device with Syringe is intended for injection or aspiration of fluids using a Luer lock (LL) or Luer slip syringe. The device is provided sterile and is for single use only. The Hypodermic Needle-Pro® EDGE™ Safety Device with Syringe is an extension of Smiths Medical's Hypodermic Needle-Pro® EDGE™ product portfolio. The Low Dead-Space (LD) Syringe new component - is added to form new device combo with the previously cleared Hypodermic Needle-Pro® EDGE™ Safety Device. The device combo is intended to offer an optimization of the volume of fluids injected to patients, and thus reduce the amount of medical product remaining in the needle and the syringe's hub after injection. The Needle-Protection (Needle-Pro) EDGE™ safety device covers the needle after use to help prevent needle sticks. The device features a one-viece design of polypropylene needle hub and protective sheath with a living hinge. The stainless-steel needle cannula is permanently affixed into the hub. After a procedure is completed, the needle is pressed into the sheath using a one-handed technique and the needle is contained within the sheath. The Low Dead-Space (LD) Syringe is a new component to be used along with the Hypodermic Needle-Pro® EDGE™ Safety Device. The LD Syringe is designed with luer lock (LL) connection to provide a secure and leak-free connection between the syringe and the needle and maintain a continuous lumen for fluid flow. The LD Syringe plunger is designed for less product waste without changing the use of the device. The Syringe plunger tip extends into the syringe neck reducing fluid waste. The fluid waste is about 0.010 mL (10 µL), max = 0.015 mL (15 µL) for the LD syringe.

    AI/ML Overview

    This document is a 510(k) Premarket Notification from the FDA regarding a medical device, specifically a hypodermic needle and syringe. It does not contain information about acceptance criteria or a study that proves the device meets those criteria in the context of an Artificial Intelligence (AI) or software-based medical device.

    The document primarily focuses on demonstrating "substantial equivalence" of a new physical device (Hypodermic Needle-Pro® EDGE™ Safety Device with Low Dead Space Syringe) to a previously cleared predicate device. The performance data mentioned (biocompatibility, sterilization, shelf-life, non-clinical bench testing) refers to the physical properties and safety of the needle and syringe, not the performance of an AI algorithm.

    Therefore, I cannot extract the requested information regarding AI device acceptance criteria and study details from this document. The provided text does not describe an AI medical device or its performance studies.

    Ask a Question

    Ask a specific question about this device

    K Number
    K241190

    Validate with FDA (Live)

    Date Cleared
    2024-05-29

    (30 days)

    Product Code
    Regulation Number
    880.5860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    PLPT LDV (Low Dead Volume) LC(Luer-Cone) Sterile Syringe / PLPT LDV (Low Dead Volume) LL(Luer-Lock) Sterile Syringe is intended for use by health care professionals for general purpose aspiration of fluid from vials, ampoules and liquid injection below the surface of the skin.

    Device Description

    PLPT LDV LC Sterile Syringe / PLPT LDV LL Sterile Syringe is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the skin. The device consists of a calibrated barrel (cylinder) with a plunger, and metal tube that is sharpened at one end and at the other end joined to a female connector (hub). This device is intended for various medical applications and is not dedication administration. The distal end of the barrel has a Luer-Cone or Luer-lock connector for the attachment to a hypodermic needle or an administration set. The attached needle is covered by the protective cap which is intended to provide physical protection to the needle tube. The attached needle may have a safety guard that protects the user after use. This product is packed by sterile paper and sterilized by EO gas, and is a single-use device.

    AI/ML Overview

    This document is a 510(k) summary for the PLPT LDV (Low Dead Volume) LC (Luer-Cone) Sterile Syringe and PLPT LDV (Low Dead Volume) LL (Luer-Lock) Sterile Syringe. It outlines the performance data and equivalence comparison to a predicate device.

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document primarily relies on compliance with established ISO standards and USP requirements rather than defining specific acceptance criteria values with corresponding measured performance values in a single table. Instead, it states that tests were "passed" or "met requirements." This is typical for medical device submissions demonstrating substantial equivalence to a predicate device by adhering to recognized consensus standards.

    Here's a summary of the acceptance criteria (standards/test criteria) and reported performance:

    Test ItemTest Method / Test CriteriaReported Device Performance
    BiocompatibilityAll evaluation acceptance criteria were met.
    Cytotoxicity TestWhen tested accordingly to ISO 10993-5, tests for in vitro cytotoxicity - Test on extracts method, it should satisfy the requirements.Pass
    Hemolysis TestWhen tested accordingly to ISO 10993-4, Selection of tests for interactions with blood - evaluation of hemolytic properties of medical devices and medical device materials, it should satisfy the requirements.Pass
    Intracutaneous Reactivity TestWhen tested accordingly to ISO 10993-10, Tests for irritation and skin sensitization - Animal intracutaneous (Intradermal) reactivity test, it should satisfy the requirements.Pass
    Skin Sensitization TestWhen tested accordingly to ISO 10993-10, Tests for irritation and skin sensitization - Guinea pig maximization test (GPMT), it should satisfy the requirements.Pass
    Acute Systemic Toxicity TestWhen tested accordingly to ISO 10993-11, Tests for systemic toxicity - Acute systemic toxicity, it should satisfy the requirements.Pass
    Pyrogen TestWhen tested accordingly to ISO 10993-11, Tests for systemic toxicity - Information on material-mediated pyrogens, it should satisfy the requirements.Pass
    LAL TestUSP39 <85>, Bacterial Endotoxins TestPass
    Particulate Matter InjectionUSP <788>, Particulate Matter for Injections (Method 1 Light Obscuration Particle Count Test). Test result should satisfy the requirements described in the USP <788>.Pass
    Sterility and LAL Test
    LAL testUSP39 <85>, Bacterial Endotoxins Test (Unit : EU/Device)Pass
    E.O sterilization validationAccording to ISO 11135:2014 E.O 30%, CO2 70% Temperature: 50 $\pm$ 7°C Exposure time: 5 hoursPass
    Sterility testAccording to ISO 11737-2Pass
    E.O Residual testUnder the conditions of ISO 10993-7:2008, Ethylene oxide sterilization residuals, the test articles should meet the test requirements.Pass
    Needle Injury Test
    Needle penetration forceISO 23908: 2011 Sharps injury protectionPass
    Pull-out forceISO 23908: 2011 Sharps injury protectionPass
    Needle cap removal forceISO 23908: 2011 Sharps injury protectionPass
    Activation (locking) forceISO 23908: 2011 Sharps injury protectionPass
    Unlocking forceISO 23908: 2011 Sharps injury protectionPass
    Mechanical PerformanceComplies with ISO 7886-1:2017 Sterile hypodermic syringes for single use - Part 1: Syringes for manual use.Complies
    Needle Performance RequirementsISO 7864:2016 Sterile hypodermic needles for single use - Requirements and test methods. ISO 9626:2016 Stainless steel needle tubing for the manufacture of medical devices - Requirements and test methods.Complies
    Safety Feature Performance SpecsISO 23908:2011 Sharps Injury Protection - Requirements And Test Methods - Sharps Protection Features For Single-Use Hypodermic Needles, Introducers For Catheters And Needles Used For Blood Sampling.Complies
    Dead Space Specification (Syringe)≤ 0.023mL with 95% confidence/95% reliability≤ 0.023mL
    Dead Space Specification (Needle)≤ 0.0054ml (No specific confidence/reliability mentioned for needle, but implied to meet this target)≤ 0.0054ml
    Shelf LifeISO 11607-1:2006 and ISO 11607-2:2006 compliance for 5 years.Validated for 5 years

    2. Sample Size Used for the Test Set and Data Provenance:

    The document mentions "Bench tests for the device's performance were conducted." and "Performance testing was conducted for this change". However, it does not specify the sample sizes used for the test sets for any of the performance or biocompatibility evaluations.

    The data provenance is not explicitly stated in terms of country of origin, but the manufacturer is "Poonglim Pharmatech Inc." located in "Gunsan, 54001, Republic of Korea." The nature of the tests (bench, biocompatibility, sterility) implies prospective testing on manufactured devices, not retrospective data analysis.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    This information is not applicable and not provided in the document. The tests performed are objective, laboratory-based physical, chemical, and biological tests governed by international standards (ISO, USP). They do not require interpretation by a panel of human experts in the way clinical diagnostic or AI-based image analysis studies might.

    4. Adjudication Method for the Test Set:

    This information is not applicable as the tests are objective, laboratory-based evaluations against predetermined criteria. There is no human subjective assessment that requires adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This information is not applicable and not provided. This document pertains to the market clearance of a sterile syringe, which is a physical medical device. It does not involve any artificial intelligence (AI) component or human diagnostic interpretation that would warrant an MRMC study.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    This information is not applicable and not provided. This device is a physical syringe and does not have an algorithm or AI component.

    7. The Type of Ground Truth Used:

    The "ground truth" for the performance criteria and biocompatibility assessments is based on established international consensus standards (ISO) and pharmacopeial monographs (USP). For instance, cytotoxicity is evaluated against requirements of ISO 10993-5, sterility against ISO 11737-2, and mechanical properties against ISO 7886-1 and ISO 23908. These standards define the acceptable limits and methodologies, which serve as the "ground truth" for compliance.

    8. The Sample Size for the Training Set:

    This information is not applicable and not provided. This device is a physical syringe and does not involve machine learning or a training set.

    9. How the Ground Truth for the Training Set was Established:

    This information is not applicable and not provided as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K231856

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-03-15

    (266 days)

    Product Code
    Regulation Number
    880.5860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    TopFine® LDS (Low Dead Space) Syringe is intended for use to inject fluid into or withdraw fluids from the parts of body below the surface of the skin.

    Device Description

    The TopFine® LDS (Low Dead Space) Syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a hypodermic single lumen needle which is a permanently attached. The device is used to inject fluids into, or withdraw fluids from, the body. It is made of plastic and silicone materials that allows for smooth plunger movement, and is manually operated. This is a single-use device. This product is packed by sterile paper and sterilized by Ethylene Oxide gas.

    The proposed device has 2 models depending on the diameter of the needle.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a medical device, specifically a syringe. It does not contain information about an Artificial Intelligence/Machine Learning (AI/ML) device. Therefore, it is not possible to describe acceptance criteria or a study related to AI/ML device performance.

    The document focuses on demonstrating substantial equivalence of the "TopFine® LDS (Low Dead Space) Syringe" to a predicate syringe through various performance tests and material comparisons.

    Here's an analysis based on the information provided, specifically addressing the technical performance of the syringe, not AI/ML:

    Acceptance Criteria and Reported Device Performance (Non-AI/ML Syringe)

    The document details various performance criteria for the syringe. There isn't a single table explicitly labeled "Acceptance Criteria and Reported Device Performance" with detailed numerical results for each criterion. However, the "Comparison of technological characteristics with the predicate device" section (page 5) and "Performance data" section (page 7) implicitly define the acceptance criteria by stating compliance with recognized standards and providing general results.

    Here's an attempt to construct a table based on the provided information, focusing on criteria where reported performance is available, even if only generally stated as "Same" or "Complies."

    Acceptance Criteria CategorySpecific Criteria / StandardAcceptance Criteria (Expected/Reference)Reported Device Performance (TopFine® LDS Syringe)
    General CharacteristicsRegulation Number21 CFR 880.5860Same (21 CFR 880.5860)
    Product CodeQNQSame (QNQ)
    Indications for UseIntended for use to inject fluid into or withdraw fluids from the parts of body below the surface of the skin.Same
    Type of usePrescription UseSame
    Mechanism of actionPlunger can be pulled and pushed along inside the barrel, allowing the syringe to take in and expel fluids through the connector to the patient.Same
    Number of usesSingle Use OnlySame
    Material (Needle)SUS 304Same (SUS 304)
    Material (Cap)PolypropyleneSame (Polypropylene)
    Material (Barrel)PolypropyleneSame (Polypropylene)
    Material (Plunger)PolypropyleneSame (Polypropylene)
    Material (Gasket)PolyisopreneSame (Polyisoprene)
    Needle Gauge23G, 25GSame (23G, 25G)
    Needle OD (23G)0.600 ~ 0.673 mmSame (0.600 ~ 0.673 mm)
    Needle OD (25G)0.500 ~ 0.530 mmSame (0.500 ~ 0.530 mm)
    Syringe nozzle typePermanently attachedSame
    Syringe capacity1 mlSame (1 ml)
    Performance - BiocompatibilityISO 10993-1 ClassificationExternally Communicating Device, Blood Path Indirect, Limited Contact (<24hours)Device classified the same.
    CytotoxicityComplies with ISO 10993-1Testing conducted and met criteria (implied by "following testing was conducted").
    SensitizationComplies with ISO 10993-1Testing conducted and met criteria.
    IrritationComplies with ISO 10993-1Testing conducted and met criteria.
    Acute Systemic ToxicityComplies with ISO 10993-1Testing conducted and met criteria.
    Material Mediated PyrogenicityComplies with ISO 10993-1Testing conducted and met criteria.
    HemolysisComplies with ISO 10993-1Testing conducted and met criteria.
    Particulate MatterUSP <788> acceptance criteriaMet USP acceptance criteria.
    Performance - PhysicalNeedle Length (23G, 25G)Predicate: 25.0 mm, 25.4 mm, 38.1 mmProposed: 26.5 mm (+1.5 / -2.5 mm) - Different, but performance verified.
    Hub/needle bond strength (23G)≥ 34 N (ISO 7864)Same (≥ 34 N)
    Hub/needle bond strength (25G)≥ 22 N (ISO 7864)Same (≥ 22 N)
    Dead space≤ 0.03 mlSame (≤ 0.03 ml)
    Performance Needle StandardISO 7864:2016Same (ISO 7864:2016)
    Needle Tubing StandardISO 9626:2016Same (ISO 9626:2016)
    Syringe StandardISO 7886-1:2017Same (ISO 7886-1:2017)
    Performance - SterilitySterilization MethodEO Gas, SAL=10-6Same (EO Gas, SAL=10-6)
    EO ResidueNot detected (LOD: 0.001)EO not detected.
    ECH ResidueNot detected (LOD: 0.02)ECH not detected.
    Pyrogen TestBacterial endotoxin USP <161>, USP <85>Testing conducted and met criteria.
    Package Integrity (ASTM D4169-22)Acceptable for protection and sterility maintenance after environmental conditioning and simulated transportation.All packaging is acceptable.
    Seal Peel TestComplies with ASTM F88/F88 -15Performed.
    Dye MigrationComplies with ASTM F1929-15Performed.
    Shelf Life3 years (validated using ASTM F1980-21)3 years validated.
    LabelingComplianceComplies with 21 CFR part 801Same (Complies with 21 CFR part 801)

    Since this is concerning a non-AI/ML medical device, the following points are not applicable or cannot be answered from the provided text:

    1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

      • The document does not specify exact sample sizes for each physical or biological test. It only mentions that tests were "conducted."
      • Data provenance (country of origin, retrospective/prospective) is not mentioned for the testing data.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

      • Not applicable as this is not an AI/ML device requiring expert ground truth for image/data interpretation. This refers to standard performance testing.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set

      • Not applicable. Adjudication methods are typically for clinical studies involving human interpretation or AI/ML output.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

      • Not applicable. This is not an AI/ML device. "Clinical studies were not required to support substantial equivalence."
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

      • Not applicable. This is not an AI/ML device.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

      • For the syringe, "ground truth" would be established by direct physical measurements, chemical analyses, and biological assays according to the specified international and national standards (e.g., ISO, USP, ASTM). For example, the "dead space" measurement itself would be the ground truth for that specific characteristic, and the acceptance criterion is a predefined maximum value (≤ 0.03 ml).
    7. The sample size for the training set

      • Not applicable. This is not an AI/ML device with a training set.
    8. How the ground truth for the training set was established

      • Not applicable. This is not an AI/ML device.

    In summary, the document addresses a traditional medical device (syringe) and demonstrates its substantial equivalence to a predicate device through adherence to manufacturing standards, material comparisons, and performance testing against established physical, chemical, and biological criteria, rather than through AI/ML performance metrics.

    Ask a Question

    Ask a specific question about this device

    K Number
    K221860

    Validate with FDA (Live)

    Date Cleared
    2023-09-15

    (445 days)

    Product Code
    Regulation Number
    880.5860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    PLPT LDV (Low Dead Volume) LC(Luer-Cone) Sterile Syringe / PLPT LDV (Low Dead Volume) LL(Luer-Lock) Sterile Syringe is intended for use by health care professionals for general purpose aspiration of fluid from vials, ampoules and liquid injection below the surface of the skin.

    Device Description

    PLPT LDV LC Sterile Syringe / PLPT LDV LL Sterile Syringe is a device intended to inject fluids into, or withdraw fluids from, parts of the body the surface of the skin. The device consists of a calibrated barrel (cylinder) with plunger, and metal tube that is sharpened at one end and at the other end joined to a female connector (hub). This device is intended for various medical applications and is not dedication administration. At the distal end of the barrel has a Luer-Cone or Luer-lock connector for the attachment to a hypodermic needle or an administration set. Attached needle has covered by the protective cap which is intended to provide physical protection to the needle tube. Attached needle may have safety guard that protect the user after use. This product is packed by sterile paper and sterilized by EO gas, and single-use device.

    AI/ML Overview

    This document describes the regulatory approval for a medical device, specifically a Low Dead Volume (LDV) Sterile Syringe (PLPT LDV LC/LL Sterile Syringe). The provided text is an FDA 510(k) Premarket Notification, which focuses on demonstrating substantial equivalence to a predicate device, rather than proving novel effectiveness or clinical superiority through extensive comparative clinical trials.

    Therefore, the information about acceptance criteria and study proving the device meets those criteria will be focused on bench testing and biocompatibility assessments demonstrating equivalence to established standards and predicate devices, rather than complex multi-reader, multi-case studies typically seen for AI/ML devices.

    Here's the breakdown based on the provided text:

    Device: PLPT LDV (Low Dead Volume) LC (Luer-Cone) Sterile Syringe and PLPT LDV (Low Dead Volume) LL (Luer-Lock) Sterile Syringe.

    Indications for Use: Intended for use by healthcare professionals for general purpose aspiration of fluid from vials, ampoules, and liquid injection below the surface of the skin.


    Acceptance Criteria and Reported Device Performance

    The acceptance criteria are primarily based on established international standards for medical devices and comparison to predicate devices. The device's performance is demonstrated through adherence to these standards and the fulfillment of specific regulatory requirements.

    Acceptance Criteria CategorySpecific Acceptance Criteria (Standard / Test Method / Limit)Reported Device Performance (Compliance / Result)
    Mechanical PerformanceISO 7886-1:2017 Sterile hypodermic syringes for single use (Syringe performance)Complies with ISO 7886-1:2017
    ISO 7864:2016 Sterile hypodermic needles for single use (Needle performance)Complies with ISO 7864:2016
    ISO 9626:2016 Stainless steel needle tubing for the manufacture of medical devicesComplies with ISO 9626:2016
    ISO 80369-7:2021 Small-bore connectors for liquids and gases in healthcare applications (Connector performance)Complies with ISO 80369-7:2021
    BiocompatibilityISO 10993-1:2018 (Body contact: External communication device – Blood path indirect; Duration: Limited (< 24 hours))All evaluation acceptance criteria were met.
    Cytotoxicity (ISO 10993-5)Satisfied requirements
    Hemolysis (ISO 10993-4)Satisfied requirements
    Intracutaneous Reactivity (ISO 10993-10)Satisfied requirements
    Skin Sensitization (ISO 10993-10, GPMT)Satisfied requirements
    Acute Systemic Toxicity (ISO 10993-11)Satisfied requirements
    Pyrogenicity (ISO 10993-11)Satisfied requirements
    LAL Test (USP39 <85>, Bacterial Endotoxins Test)Satisfied requirements
    Particulate Matter Injection (USP <788>)Satisfied requirements
    SterilityISO 11135:2014 Ethylene Oxide (EO) Sterilization Validation (Parameters: EO 30%, CO2 70%; Temp: 50 ±7°C; Exposure: 5 hours)Validated and deemed acceptable.
    Sterility Test (ISO 11737-2)Complied
    EO Residual Test (ISO 10993-7:2008)Met test requirements
    Sterility Assurance Level (SAL)10⁻⁶
    Endotoxin Limit20 EU per device
    Low Dead VolumeSyringe Dead Space: ≤ 0.023 mL≤ 0.023 mL (Same as predicate)
    Needle Dead Space: < 0.0054 mL< 0.0054 mL (Same as predicate)
    Sharps Injury ProtectionISO 23908:2011 Sharps injury protectionComplies with ISO 23908:2011
    Shelf-LifeISO 11607-1:2006 and ISO 11607-2:2006 (for 3 years)Validated for 3 years
    Labeling21 CFR part 801Complied

    2. Sample size used for the test set and the data provenance:

    The document does not explicitly state the numerical sample sizes for each specific test (e.g., how many syringes were tested for each mechanical property, or how many biological samples were used for biocompatibility tests). However, for medical devices cleared via 510(k), it is expected that sufficient samples were used to meet the statistical requirements of the referenced ISO standards for design verification and validation.

    • Data Provenance: The studies were bench tests and biocompatibility evaluations. The data would be prospective, generated specifically for this 510(k) submission, likely conducted at the manufacturer's facilities or accredited testing labs. The country of origin for the data generation is not explicitly stated, but the manufacturer is based in Republic of Korea.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This type of information is generally not applicable for a 510(k) submission for a physical medical device like a syringe. The "ground truth" for these tests comes from adherence to established international engineering and biological standards (ISO, USP). There isn't a subjective "ground truth" established by human experts in the way there would be for an AI-based diagnostic tool. The "experts" are the engineers and scientists conducting the tests according to the specified protocols.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    Not applicable. Adjudication methods like 2+1 or 3+1 are used for establishing consensus "ground truth" in human-in-the-loop diagnostic studies, typically involving image interpretation. For physical device performance, the results are quantitative measurements against predefined pass/fail criteria established by engineering standards.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This device is a physical syringe, not an AI-assisted diagnostic tool. No MRMC study was conducted.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    Not applicable. This device is a physical syringe, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The "ground truth" for this device's performance is based on objective measurements and results from standardized bench tests and validated biological assays as defined by:

    • International Organization for Standardization (ISO) standards (e.g., ISO 7886-1, ISO 80369-7, ISO 10993 series, ISO 11135, ISO 11607, ISO 23908).
    • United States Pharmacopeia (USP) guidelines (e.g., USP <85>, USP <788>).

    These standards define the acceptable performance characteristics for syringes and needles.

    8. The sample size for the training set:

    Not applicable. This device is a physical product and does not involve AI/ML requiring a "training set" in the computational sense.

    9. How the ground truth for the training set was established:

    Not applicable for the same reason as above.

    Ask a Question

    Ask a specific question about this device

    K Number
    K213010

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-05-31

    (618 days)

    Product Code
    Regulation Number
    880.5860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    FEELject LDV(Low dead volume) Syringe is intended for use to inject fluid into or withdraw fluids from the parts of body below the surface of the skin.

    Device Description

    A Disposable Syringe with Needle is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a hypodermic single lumen needle which is permanently attached. The device is used to inject fluids into, or withdraw fluids from, the body. It is made of plastic and silicone materials that allows for smooth plunger movement, and manually operated. This is a single-use device. This product is packed by sterile paper and sterilized by E.O gas.

    AI/ML Overview

    The provided text describes the 510(k) summary for the FEELject LDV (Low dead volume) Syringe, a medical device. This document focuses on demonstrating substantial equivalence to a predicate device, rather than proving a new medical device meets specific acceptance criteria through a clinical study or AI performance evaluation.

    Therefore, the information required to answer your specific questions about acceptance criteria, AI performance, ground truth establishment, sample sizes for training/test sets, expert adjudication, or MRMC studies for an AI/device is not available in this document.

    The document discusses bench testing and biocompatibility testing to show that the device performs in a substantially equivalent manner to the predicate device. These tests demonstrate the physical and biological properties of the syringe, not the performance of an AI algorithm or a diagnostic device.

    Here's a breakdown of what can be extracted from the document, and why the other requested information is absent:


    Information Available from the Document:

    This document is a 510(k) summary for a medical device (syringe), not an AI algorithm. Therefore, the requested information about AI-related acceptance criteria, study design (MRMC, standalone), ground truth establishment for AI, and sample sizes for AI training/test sets is not applicable and not found in this text.

    The closest equivalent to "acceptance criteria" here is compliance with ISO standards and passing various bench tests and biocompatibility tests to show substantial equivalence to the predicate device.

    1. A table of acceptance criteria and the reported device performance

    While not "AI performance" criteria, the document lists various requirements/tests for the syringe and their results (Pass). These serve as the "acceptance criteria" for the syringe's physical and biological properties.

    Requirement - Test (ISO 7886-1)Reported Device Performance (Result)
    Limits for extractable metalsPass
    GeneralPass
    Limits for acidity or alkalinityPass
    Conical fittingPass
    Position of nozzle on end of barrelPass
    Tolerance on graduated capacityPass
    ScalePass
    Numbering of scalesPass
    Overall length of scale to nominal capacity linePass
    Position of scalePass
    Barrel flangesPass
    Plunger stopper/plunger assemblyPass
    Dead spacePass
    Freedom from air and liquid leakage past plunger stopperPass
    Force to operate the pistonPass
    Fit of plunger stopper/plunger in barrelPass
    Unit packaging and self-contained syringe unitsPass
    Quantity of silicone oilPass
    Requirement - Test (ISO 7864)Reported Device Performance (Result)
    CleanlinessPass
    Tolerances on lengthPass
    Needle Point, needles with sharp tip onlyPass
    Bond between hub and needle tubePass
    Patency lumenPass
    Requirement – Test (ISO 9626)Reported Device Performance (Result)
    StiffnessPass
    Resistance to breakagePass
    Resistance to corrosionPass
    Requirement - Test (USP 788)Reported Device Performance (Result)
    Particulate matter injectionsPass
    Biocompatibility Test ItemTest Method / Test CriteriaReported Device Performance (Result)
    CytotoxicityISO 10993-5 Tests for in vitro cytotoxicityPass
    Skin Sensitization TestISO 10993-10 irritation and skin sensitizationPass
    Intracutaneous Reactivity TestISO 10993-10 Test for irritation and skin sensitization, maximization test for delayed hypersensitivityPass
    Acute Systemic Toxicity TestISO 10993-11 Test for systemic toxicity - Acute Systemic ToxicityPass
    Pyrogen TestISO 10993-11 Tests for systemic toxicity, Annex(F) Information on material-mediated pyrogens.Pass
    Hemolysis TestISO 10993-4 Selection of tests for interactions with bloodPass
    Sterility and LAL Test ItemTest StandardReported Device Performance (Result)
    LAL test (Bacterial Endotoxins)USP39 <85> (Unit : EU/Device)Pass
    Sterility testAccording to ISO 11737-2Pass
    E.O ResidualUnder the conditions of ISO 10993-7:2008, Ethylene oxide sterilization residualsPass

    Information Not Applicable or Not Present:

    The following information is specifically requested for AI/Software as a Medical Device (SaMD) performance studies, which are not described in this 510(k) summary for a physical medical device (syringe). Therefore, the answers are "Not Applicable" or "Not Provided" in this context.

    • 2. Sample sized used for the test set and the data provenance: Not applicable. The tests are bench tests of physical characteristics, not data-driven AI test sets with specific sample sizes from a clinical population.
    • 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for an AI algorithm is not established for a physical syringe.
    • 4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable. This is for AI performance evaluation.
    • 5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is for AI performance evaluation.
    • 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is for AI performance evaluation. The device is a physical syringe.
    • 7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable. This is for AI performance evaluation. The "ground truth" for the syringe is its compliance with specified physical and biological standards.
    • 8. The sample size for the training set: Not applicable. There is no AI training set for a syringe.
    • 9. How the ground truth for the training set was established: Not applicable. There is no AI training set for a syringe.
    Ask a Question

    Ask a specific question about this device

    K Number
    K213013

    Validate with FDA (Live)

    Date Cleared
    2022-09-19

    (364 days)

    Product Code
    Regulation Number
    880.5860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    STERILE SINGLE USE LDV SYRINGE/CHOICARE STERILE SINGLE USE LDV SYRINGE is intended for use by health care professionals for general purpose aspiration of fluid from vials, ampoules and liquid injection below the surface of the skin.

    Device Description

    A Disposable Syringe with Needle is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a hypodermic single lumen needle which is permanently attached. The device is used to inject fluids into, or withdraw fluids from, the body. This is a single-use device. This product is packed by sterile paper and sterilized by E.O gas.

    AI/ML Overview

    This document is an FDA 510(k) Premarket Notification for a medical device (Sterile Single-use LDV Syringe, CHOICARE Sterile Single-use LDV Syringe). The provided text describes the device, its intended use, performance data, and a comparison to a predicate device to demonstrate substantial equivalence.

    Here's a breakdown of the requested information based on the provided text:

    Important Note: The provided document is a 510(k) submission for a syringe, which is a relatively low-risk medical device. The types of acceptance criteria and studies are primarily focused on bench testing and adherence to existing international standards, rather than complex clinical studies or AI/software performance evaluations. Therefore, many of the requested points related to AI, MRMC studies, and expert ground truth establishment for complex image analysis are not applicable or detailed in this document.


    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are primarily defined by adherence to various ISO standards and specific quantitative thresholds for certain tests. The reported device performance is consistently "Pass" for all tested criteria, indicating the device met the acceptance criteria.

    Requirement – Test (Standard)Acceptance Criteria (Implicit for "Pass")Reported Device Performance
    ISO 7886-1:2017 (Syringe Performance)Conformance to standard requirements for manual use syringes (e.g., proper dimensions, tightness, piston/plunger assembly)Pass
    Visual checkN/A (Visual inspection, implied to be defect-free)Pass
    Dimension (out diameter of the needle tube)N/A (Within specified tolerance per standard)Pass
    Length of the needle tubeN/A (Within specified tolerance per standard)Pass
    NozzleN/A (Conforms to design and standard)Pass
    TightnessN/A (No leakage at specified pressures)Pass
    Piston/Plunger AssemblyN/A (Smooth movement, no binding)Pass
    ISO 7864:2016 (Hypodermic Needle Performance)Conformance to standard requirements (e.g., elasticity, flexural strength, pullout)Pass
    ElasticityN/A (Meets specified flexibility)Pass
    Flexural strengthN/A (Withstands specified bending force)Pass
    PulloutN/A (Withstands specified pullout force)Pass
    ISO 9626:2016 (Stainless Steel Needle Tubing)Conformance to standard requirements (e.g., stiffness, resistance to breakage/corrosion)Pass
    StiffnessN/A (Meets specified rigidity)Pass
    Resistance to breakageN/A (Withstands specified forces without breaking)Pass
    Resistance to corrosionN/A (Shows no significant corrosion after specified tests)Pass
    Dead Volume< 0.0284mlConforms (Implied: < 0.0284ml)
    Biocompatibility (ISO 10993 Series)Conformance to the requirements for each specific test (e.g., no cytotoxicity, no hemolysis, acceptable reactivity)Pass (for all listed tests)
    Cytotoxicity test (ISO 10993-5)Satisfy the requirements for in vitro cytotoxicityPass
    Hemolysis test (ISO 10993-4)Satisfy the requirements for hemolytic propertiesPass
    Intracutaneous reactivity test (ISO 10993-10)Satisfy the requirements for irritation and skin sensitizationPass
    Skin sensitization test (ISO 10993-10)Satisfy the requirements for irritation and skin sensitizationPass
    Acute systemic toxicity test (ISO 10993-11)Satisfy the requirements for acute systemic toxicityPass
    Pyrogen Test (ISO 10993-11)Satisfy the requirements for material-mediated pyrogensPass
    LAL Test (USP39 <85>)Satisfy the requirements for bacterial endotoxinsPass
    Particulate Matter Injection (USP <788>)Satisfy the requirements for particulate matterPass
    Sterility & EO ResidualsConformance to standard requirements (e.g., SAL of 10^-6, low EO residuals)Pass (for all listed tests)
    LAL test (USP39 <85>)Satisfy the requirements for bacterial endotoxinsPass
    E.O sterilization validation (ISO 11135:2014)Process validated to achieve sterilityPass
    Sterility test (ISO 11737-2)No microbial growthPass
    E.O Residual test (ISO 10993-7:2008)Meets test requirements for ethylene oxide sterilization residualsPass
    Needle Injury Test (ISO 23908:2011 Sharps Injury Protection)Conformance to standard requirements for sharps injury protection (e.g., acceptable penetration, pull-out, cap removal, activation/locking, unlocking forces)Pass (for all listed tests)
    Needle penetration forceN/A (Within specified range)Pass
    Pull-out forceN/A (Within specified range)Pass
    Needle cap removal forceN/A (Within specified range)Pass
    Activation (locking) force (for safety feature)N/A (Within specified range)Pass
    Unlocking force (for safety feature)N/A (Within specified range)Pass

    2. Sample Size for the Test Set and Data Provenance

    • Sample Size for Test Set: The document does not explicitly state the numerical sample sizes for each bench test conducted. However, for a 510(k) submission, testing is typically performed on a statistically relevant number of units as required by the referenced ISO standards.
    • Data Provenance: The document indicates that the tests were performed by Jeil Tech Co., Ltd. (Republic of Korea). The studies are "bench tests," which are laboratory tests, not clinical studies involving human patients. The nature is prospective in the sense that the tests were performed specifically for this submission to verify the device's adherence to standards and safety.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable to this type of device and study. Bench testing for a syringe relies on objective measurements against engineering specifications and international standards (e.g., ISO, USP). There isn't "ground truth" to be established by human experts in the context of diagnostic interpretation or clinical outcomes. The "experts" involved would be qualified lab technicians and engineers performing and verifying the tests.


    4. Adjudication Method for the Test Set

    This is not applicable. Since the tests are objective bench measurements against quantitative or qualitative (Pass/Fail based on standard adherence) criteria, there is no need for an adjudication method. The results are either within specification or not.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, What was the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance

    This is not applicable. This submission is for a physical medical device (syringe), not an AI-powered diagnostic or assistive tool. Therefore, no MRMC comparative effectiveness study was performed or required.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    This is not applicable. As stated above, this is not an AI/algorithm device.


    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance is the adherence to established international standards (ISO 7886-1, ISO 7864, ISO 9626, ISO 23908, ISO 10993 series, ISO 11135, ISO 11737-2) and specific quantitative requirements (e.g., dead volume, endotoxin limits). This is an engineering specification and regulatory standard conformance type of ground truth.


    8. The Sample Size for the Training Set

    This is not applicable. There is no "training set" as this is not an AI or machine learning device. The device is manufactured based on design specifications, and tested to ensure it meets those specifications and relevant standards.


    9. How the Ground Truth for the Training Set was Established

    This is not applicable as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 2