K Number
K231856

Validate with FDA (Live)

Manufacturer
Date Cleared
2024-03-15

(266 days)

Product Code
Regulation Number
880.5860
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

TopFine® LDS (Low Dead Space) Syringe is intended for use to inject fluid into or withdraw fluids from the parts of body below the surface of the skin.

Device Description

The TopFine® LDS (Low Dead Space) Syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a hypodermic single lumen needle which is a permanently attached. The device is used to inject fluids into, or withdraw fluids from, the body. It is made of plastic and silicone materials that allows for smooth plunger movement, and is manually operated. This is a single-use device. This product is packed by sterile paper and sterilized by Ethylene Oxide gas.

The proposed device has 2 models depending on the diameter of the needle.

AI/ML Overview

The provided document is a 510(k) premarket notification for a medical device, specifically a syringe. It does not contain information about an Artificial Intelligence/Machine Learning (AI/ML) device. Therefore, it is not possible to describe acceptance criteria or a study related to AI/ML device performance.

The document focuses on demonstrating substantial equivalence of the "TopFine® LDS (Low Dead Space) Syringe" to a predicate syringe through various performance tests and material comparisons.

Here's an analysis based on the information provided, specifically addressing the technical performance of the syringe, not AI/ML:

Acceptance Criteria and Reported Device Performance (Non-AI/ML Syringe)

The document details various performance criteria for the syringe. There isn't a single table explicitly labeled "Acceptance Criteria and Reported Device Performance" with detailed numerical results for each criterion. However, the "Comparison of technological characteristics with the predicate device" section (page 5) and "Performance data" section (page 7) implicitly define the acceptance criteria by stating compliance with recognized standards and providing general results.

Here's an attempt to construct a table based on the provided information, focusing on criteria where reported performance is available, even if only generally stated as "Same" or "Complies."

Acceptance Criteria CategorySpecific Criteria / StandardAcceptance Criteria (Expected/Reference)Reported Device Performance (TopFine® LDS Syringe)
General CharacteristicsRegulation Number21 CFR 880.5860Same (21 CFR 880.5860)
Product CodeQNQSame (QNQ)
Indications for UseIntended for use to inject fluid into or withdraw fluids from the parts of body below the surface of the skin.Same
Type of usePrescription UseSame
Mechanism of actionPlunger can be pulled and pushed along inside the barrel, allowing the syringe to take in and expel fluids through the connector to the patient.Same
Number of usesSingle Use OnlySame
Material (Needle)SUS 304Same (SUS 304)
Material (Cap)PolypropyleneSame (Polypropylene)
Material (Barrel)PolypropyleneSame (Polypropylene)
Material (Plunger)PolypropyleneSame (Polypropylene)
Material (Gasket)PolyisopreneSame (Polyisoprene)
Needle Gauge23G, 25GSame (23G, 25G)
Needle OD (23G)0.600 ~ 0.673 mmSame (0.600 ~ 0.673 mm)
Needle OD (25G)0.500 ~ 0.530 mmSame (0.500 ~ 0.530 mm)
Syringe nozzle typePermanently attachedSame
Syringe capacity1 mlSame (1 ml)
Performance - BiocompatibilityISO 10993-1 ClassificationExternally Communicating Device, Blood Path Indirect, Limited Contact (<24hours)Device classified the same.
CytotoxicityComplies with ISO 10993-1Testing conducted and met criteria (implied by "following testing was conducted").
SensitizationComplies with ISO 10993-1Testing conducted and met criteria.
IrritationComplies with ISO 10993-1Testing conducted and met criteria.
Acute Systemic ToxicityComplies with ISO 10993-1Testing conducted and met criteria.
Material Mediated PyrogenicityComplies with ISO 10993-1Testing conducted and met criteria.
HemolysisComplies with ISO 10993-1Testing conducted and met criteria.
Particulate MatterUSP <788> acceptance criteriaMet USP acceptance criteria.
Performance - PhysicalNeedle Length (23G, 25G)Predicate: 25.0 mm, 25.4 mm, 38.1 mmProposed: 26.5 mm (+1.5 / -2.5 mm) - Different, but performance verified.
Hub/needle bond strength (23G)≥ 34 N (ISO 7864)Same (≥ 34 N)
Hub/needle bond strength (25G)≥ 22 N (ISO 7864)Same (≥ 22 N)
Dead space≤ 0.03 mlSame (≤ 0.03 ml)
Performance Needle StandardISO 7864:2016Same (ISO 7864:2016)
Needle Tubing StandardISO 9626:2016Same (ISO 9626:2016)
Syringe StandardISO 7886-1:2017Same (ISO 7886-1:2017)
Performance - SterilitySterilization MethodEO Gas, SAL=10-6Same (EO Gas, SAL=10-6)
EO ResidueNot detected (LOD: 0.001)EO not detected.
ECH ResidueNot detected (LOD: 0.02)ECH not detected.
Pyrogen TestBacterial endotoxin USP <161>, USP <85>Testing conducted and met criteria.
Package Integrity (ASTM D4169-22)Acceptable for protection and sterility maintenance after environmental conditioning and simulated transportation.All packaging is acceptable.
Seal Peel TestComplies with ASTM F88/F88 -15Performed.
Dye MigrationComplies with ASTM F1929-15Performed.
Shelf Life3 years (validated using ASTM F1980-21)3 years validated.
LabelingComplianceComplies with 21 CFR part 801Same (Complies with 21 CFR part 801)

Since this is concerning a non-AI/ML medical device, the following points are not applicable or cannot be answered from the provided text:

  1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • The document does not specify exact sample sizes for each physical or biological test. It only mentions that tests were "conducted."
    • Data provenance (country of origin, retrospective/prospective) is not mentioned for the testing data.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable as this is not an AI/ML device requiring expert ground truth for image/data interpretation. This refers to standard performance testing.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. Adjudication methods are typically for clinical studies involving human interpretation or AI/ML output.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This is not an AI/ML device. "Clinical studies were not required to support substantial equivalence."
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is not an AI/ML device.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • For the syringe, "ground truth" would be established by direct physical measurements, chemical analyses, and biological assays according to the specified international and national standards (e.g., ISO, USP, ASTM). For example, the "dead space" measurement itself would be the ground truth for that specific characteristic, and the acceptance criterion is a predefined maximum value (≤ 0.03 ml).
  7. The sample size for the training set

    • Not applicable. This is not an AI/ML device with a training set.
  8. How the ground truth for the training set was established

    • Not applicable. This is not an AI/ML device.

In summary, the document addresses a traditional medical device (syringe) and demonstrates its substantial equivalence to a predicate device through adherence to manufacturing standards, material comparisons, and performance testing against established physical, chemical, and biological criteria, rather than through AI/ML performance metrics.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

March 15, 2024

MedExel Co., Ltd. % Peter Chung, President PlusGlobal 300 Atwood Street Pittsburgh, Pennsylvania 15213

Re: K231856

Trade/Device Name: TopFine® LDS (Low Dead Space) Syringe (2 models (TS-LDS2334, TS-LDS2534)) Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: QNQ Dated: February 14, 2024 Received: February 14, 2024

Dear Peter Chung:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

2

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.70) and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely, Shruti N. Mistry -S

Shruti Mistry Assistant Director, Injection Team Division of Drug Delivery and General Hospital Devices, and Human Factors Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality

Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K231856

Device Name

TopFine® LDS (Low Dead Space) Syringe (2 models (TS-LDS2334, TS-LDS2534))

Indications for Use (Describe)

TopFine® LDS (Low Dead Space) Syringe is intended for use to inject fluid into or withdraw fluids from the parts of body below the surface of the skin.

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K231856- 510(K) Summary

I. Submitter
510(k) Submitter: MedExel Co., Ltd.252, Geumgwangosan-ro, Geumgwang-myeonAnseong Korea, SouthPhone: +82-10-9992-9912
Primary Contact: Peter ChungUS Agentcontact@plusglobal.net
Secondary Contact: Juntaek SeoRA managerjtseo@medexel.co.kr
Date prepared: 15 March, 2024

{4}------------------------------------------------

II. Device Name

Device Trade NameTopFine® LDS (Low Dead Space)Syringe(2 models (TS-LDS2334, TS-LDS2534))
Common NamePiston syringe
Classification NameLow Dead Space Piston Syringe
Regulation Number21 CFR 880.5860
Product CodeQNQ
  • III. Legally Marketed Predicate Device
Predicate #K213010
Predicate Trade NameFEELject LDV (Low dead volume) syringe

Device Description IV.

The TopFine® LDS (Low Dead Space) Syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a hypodermic single lumen needle which is a permanently attached. The device is used to inject fluids into, or withdraw fluids from, the body. It is made of plastic and silicone materials that allows for smooth plunger movement, and is manually operated. This is a single-use device. This product is packed by sterile paper and sterilized by Ethylene Oxide gas.

The proposed device has 2 models depending on the diameter of the needle.

No.Model nameDiameter of needleLength of needle
1TS-LDS233423G (0.600 ~ 0.673 mm)26.5 mm (+1.5 / -2.5 mm)
2TS-LDS253425G (0.500~ 0.530 mm)26.5 mm (+1.5 / -2.5 mm)

The difference between the above 2 models is only the needle diameter.

V. Indications for Use

TopFine® LDS (Low Dead Space) Syringe is intended for use to inject fluid into or withdraw fluids from the parts of body below the surface of the skin.

{5}------------------------------------------------

ItemProposed DevicePredicate DeviceRemark
510(k) No.K231856K213010-
Regulation No.21 CFR 880.586021 CFR 880.5860Same
Product CodeQNQQNQ, FMISame
Indications for UseTopFine® LDS (Low Dead Space) Syringe is intended for use to inject fluid into or withdraw fluids from the parts of body below the surface of the skin.FEELject LDV (Low dead volume) Syringe is intended for use to inject fluid into or withdraw fluids from the parts of body below the surface of the skin.Same
Type of usePrescription UsePrescription UseSame
Mechanism of actionThe plunger of syringe can be pulled and pushed along inside the barrel, allowing the syringe to take in and expel the fluids through the connector to the patient.The plunger of syringe can be pulled and pushed along inside the barrel, allowing the syringe to take in and expel the fluids through the connector to the patient.Same
Number of usesSingle Use OnlySingle Use OnlySame
Configuration and materialsNeedle- Needle: SUS 304- Cap: Polypropylene- Barrel: Polypropylene- Plunger: Polypropylene- Gasket: PolyisopreneNeedle- Needle: SUS 304- Cap: Polypropylene- Barrel: Polypropylene- Plunger: Polypropylene- Gasket: PolyisopreneSame
Needle Gauge23G, 25G23G, 25GSame
Needle dimension – OD23G: 0.600 ~ 0.673 mm25G: 0.500 ~ 0.530 mm23G: 0.600 ~ 0.673 mm25G: 0.500 ~ 0.530 mmSame
Needle dimension – needle length26.5 mm(+1.5 / -2.5 mm)25.0 mm25.4 mm38.1 mmDifferentSee note 1
Syringe nozzle typePermanently attachedPermanently attachedSame
Syringe capacity1 ml1 mlSame
Dead spaceLow dead volume ≤ 0.03 mlLow dead volume ≤ 0.03 mlSame
BiocompatibilityAccording to ISO 10993-1According to ISO 10993-1Same
Performance NeedleISO 7864:2016 Sterile hypodermic needles for single use – Requirements and test methodsISO 7864:2016 Sterile hypodermic needles for single use – Requirements and test methodsSame
ISO 9626:2016 Stainlesssteel needle tubing for themanufacture of medicaldevices - Requirementsand test methodsISO 9626:2016 Stainlesssteel needle tubing for themanufacture of medicaldevices – Requirementsand test methods
SyringeISO 7886-1:2017 Sterilehypodermic syringes forsingle use - Part 1:Syringes for manual useISO 7886-1:2017 Sterilehypodermic syringes forsingle use - Part 1:Syringes for manual useSame
SterilitySterilized by EO GasSAL=10-6Sterilized by EO GasSAL=10-6Same
Endotoxin limit20 EU/Device20 EU/DeviceSame
LabelingComplies with 21 CFRpart 801Complies with 21 CFRpart 801Same
PhysicalspecificationTip type: Slip TipImage: Syringe tipTip type: Slip TipImage: Syringe tipSame
MechanicalspecificationHub/needle bond strength(ISO 7864)1) 23 Gauge: ≥ 34 N2) 25 Gauge: ≥ 22 NHub/needle bond strength(ISO 7864)1) 23 Gauge: ≥ 34 N2) 25 Gauge: ≥ 22 NSame

VI. Comparison of technological characteristics with the predicate device

{6}------------------------------------------------

Note 1: Needle length

The needle length of proposed device is different from the needle of predicate device. In order to confirm the performance between the different needle specifications, the proposed device was verified by measuring the needle length at each time point of accelerated aging test. The proposed device was found to be substantially equivalent to the predicate device.

{7}------------------------------------------------

VII. Performance data

1. Biocompatibility Testing

In accordance with ISO 10993-1 the subject device is classified as: Externally Communicating Device, Blood Path Indirect, Limited Contact (<24hours). The following testing was conducted.

    1. Cytotoxicity
    1. Sensitization
    1. Irritation
    1. Acute Systemic Toxicity
    1. Material Mediated Pyrogenicity
    1. Hemolysis

Particulate matter testing was conducted in accordance with USP <788> Particulate Matter in Injection and met the USP acceptance criteria.

2. Physical Performance Testing

  • (1) ISO 7864:2016-Sterile hypodermic needles for single use requirements and test methods
  • (2) ISO 7886-1:2017-Sterile hypodermic syringes for single use Part 1: Syringes for manual use
  • (3) ISO 9626:2016-Stainless steel needle tubing for the manufacture of medical devices

3. Sterility, Shipping and Shelf-Life

The sterilization method is Ethylene Oxide. EO not detected (Limit of Detection: 0001), ECH- Not detected. (Limit of Detection: 0.02), validation method: Overkill Approach (e.g., Half-Cycle method), pyrogen test: Bacterial endotoxin USP <161>Medical devices -Bacterial Endotoxin and pyrogen test, USP <85> Bacterial endotoxins test.

  • Package integrity testing, after environmental conditioning and simulated transportation using ASTM D4169-22, was conducted on final packaged, and sterile devices. All packaging is acceptable for protection of final finished product and sterility maintenance.
  • Sterile Barrier Packaging Testing performed on the proposed device:
    • Seal peel test ASTM F88/F88 -15 O
    • Dye migration- ASTM F1929-15 o
  • Shelf life of 3 years is validated using the FDA recognized standard ASTM F1980-21 Standard Guide for Accelerated Aging of Sterile Barrier System for Medical Devices.

4. Clinical Studies

Clinical testing was not required to support substantial equivalence.

{8}------------------------------------------------

VIII. Conclusion

The differences between the predicate device and the subject device do not raise any new or different questions of safety or effectiveness. The TopFine® LDS (Low Dead Space) Syringe (2 models (TS-LDS2334, TS-LDS2534) is substantially equivalent to the FEELject LDV (low dead volume) syringe with respect to the indications for use, target populations, treatment method, and technological characteristics.

§ 880.5860 Piston syringe.

(a)
Identification. A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.(b)
Classification. Class II (performance standards).