Search Results
Found 198 results
510(k) Data Aggregation
(100 days)
The Ventana® A Anterior Lumbar Interbody System is indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). The Ventana® A Anterior Lumbar System is intended for patients who have had at least six (6) months of non-operative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone.
The Ventana® A Anterior Lumbar Interbody System may be used as a standalone device when used with three integrated fixation screws and has a lordotic angle of 20° or less. When an interbody device with fewer than three screws is used, supplemental spinal fixation systems cleared by the FDA for use in the lumbar spine in addition to the integrated fixation must be used. Hyperlordotic sizes (greater than 20°) are intended for use with supplemental spinal fixation systems cleared by the FDA for use in the lumbar spine in addition to the integrated fixation.
The Ventana® A Anterior Lumbar Interbody System is an integrated intervertebral body fusion device for use in lumbar spinal surgery. Ventana® A Anterior Lumbar Interbody System devices are provided in multiple footprints with varying lengths, widths, heights, and angles of lordosis to accommodate individual patient anatomy. All interbody spacers feature an internal graft aperture for placement of graft material to promote fusion through the cage. The devices include a lid which further facilitates the installation of bone graft.
The interbody devices are additively manufactured from titanium alloy (Ti-6Al-4V ELI) conforming to ASTM F3001 and machined titanium alloy (Ti-6Al-4V ELI) components conforming to ASTM F136. The interbody device includes three screw holes to accept bone screws that are made of titanium alloy (Ti-6Al-4V ELI) per ASTM F136 in varying lengths and diameters.
All implants are intended for single patient use only and should not be reused (used in additional patients) under any circumstances. Reuse may result in serious injury or death. Components from this system should not be used in conjunction with components from other systems.
N/A
Ask a specific question about this device
(178 days)
The Solar-S™ PSI (Standalone) and Solar-A™ PSI (Non-Standalone) are lumbar interbody fusion devices intended for use at one or more levels of the lumbosacral spine (L1- S1) in patients with degenerative disc disease (DDD), disc herniation (myelopathy and/or radiculopathy), spondylolisthesis or retrolisthesis, deformity (degenerative scoliosis or kyphosis), spinal stenosis, and failed previous fusion (pseudarthrosis). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. Solar-S™ PSI and Solar-A™ PSI Spacers are to be filled with autograft bone and/or allogenic bone graft composed of cancellous and/or corticocancellous bone.
When used with three (3) screws, interbody devices with a lordotic angle ≤20° can be used as standalone interbody fusion devices at 1 or 2 contiguous levels.
When used with three (3) tusks, interbody devices must always be used with supplemental fixation and may be used at one or more levels.
Hyperlordotic interbody devices (≥20° lordosis) used with screws or tusks, must always be used with supplemental fixation, and may be used at one or more levels.
When used with supplemental fixation, all Solar-S™ PSI Spacers used with screws or tusks may be used at one or more levels.
These devices are intended to be used at one or more levels with supplemental fixation systems that have been cleared for use in the lumbosacral spine (e.g., posterior pedicle screw and rod systems, anterior plate systems, anterior screw and rod systems). Hyperlordotic interbody devices (≥20° lordosis) must be used with at least anterior supplemental fixation.
Solar-M™ PSI Spacers are lumbar interbody fusion devices indicated at one or more levels of the thoracic spine (T1-T12), thoracolumbar junction (T12-L1), or lumbosacral spine (L1-S1) as an adjunct to fusion in patients with the following indications: degenerative disc disease (DDD), disc herniation (with myelopathy and/or radiculopathy), spondylolisthesis or retrolisthesis, deformity (degenerative scoliosis or kyphosis), spinal stenosis, and failed previous fusion (pseudarthrosis). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. Solar-M™ PSI Spacers are to be filled with autograft bone and/or allogenic bone graft composed of cancellous and/or corticocancellous bone. These devices are intended to be used with supplemental fixation systems that have been cleared for use in the thoracolumbosacral spine (e.g., posterior pedicle screw and rod systems, anterior plate systems, anterior screw and rod systems). Hyperlordotic interbody devices (≥20° lordosis) must be used with at least anterior supplemental fixation.
The Impulse™ PSI is an intervertebral body fusion device intended for use in skeletally mature patients with Degenerative Disk Disease (DDD) of the lumbar spine with up to Grade 1 Spondylolisthesis at one or more levels in the lumbosacral spine (L1-S1) as an adjunct to fusion. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. The Impulse™ PSI is indicated to be used with autograft and/or allograft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion and are intended to be used with supplemental fixation systems cleared for use in the lumbar spine. The device is to be used in patients who have had six months of nonoperative treatment.
The DeGen Medical Patient-Specific Implant (PSI) System is a series of patient specific lumbar interbody spacers intended for use in anterior (Solar™ PSI) or posterior / transforaminal (Impulse™ PSI) lumbar fusion procedures. The DeGen Medical PSI devices are comprised of a single component that is additively manufactured. The superior and inferior endplates feature a porous surface to mitigate subsidence and expulsion. Superior and inferior faces feature a central aperture to constrain bone graft. The Impulse™ PSI spacers are available in two configurations: Impulse™ PSI and Impulse™ PSI Hyperlordotic. The Solar™ PSI patient specific spacers are available in three configurations; Solar-S™ (integrated fixation), Solar-A™ (anterolateral, with integrated fixation), and Solar-M™ (monolithic, without integrated fixation).
The DeGen Medical PSI spacers are additively manufactured from Puri-Ti™ unalloyed titanium. The DeGen Medical Solar-S™ PSI and Solar-A™ PSI incorporate integrated fixation in the form of screws manufactured from Ti-6AL-4V ELI Alloy per ASTM F136 or tusks additively manufactured from Puri-Ti™ unalloyed titanium. The Solar-S™ PSI Lumbar Interbody Spacer must be used with three (3) integrated screws and the spacer must have ≤20° of lordosis to be considered for standalone use. All implants may include patient matched endplates when provided with an adequate CT scan.
The Impulse™ and Solar™ PSIs are designed individually for each patient using patient imaging data (X-Ray, MRI, CT). Variable dimensions include anterior and posterior heights, bone graft window, lordosis, coronal angle, length, and width. When provided with an adequate CT scan, implants may also include patient-matched endplates. All implants are provided in their final geometry and are not intended to be altered or reshaped at the time of surgery.
N/A
Ask a specific question about this device
(104 days)
IdentiTi II ALIF Standalone Interbody System
The IdentiTi II ALIF Standalone Interbody System is indicated for spinal fusion procedures from L2 to S1 in skeletally mature patients for the treatment of symptomatic degenerative disc disease (DDD), degenerative spondylolisthesis and/or spinal stenosis at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. When used with screws or blades, these patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved levels.
Additionally, the IdentiTi II ALIF Standalone Interbody System can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity. However, when used in these patients at multiple levels, and for patients with degenerative spondylolisthesis (>Grade 1) and spinal stenosis at one or two adjacent levels, the IdentiTi II ALIF Standalone Interbody System must be used with supplemental spinal fixation systems cleared by the FDA for use in the lumbar spine. The IdentiTi II ALIF Standalone Interbody System is intended for use on patients who have had at least six months of nonoperative treatment. It is intended to be used with autograft and/or allogenic bone graft comprised of cortical, cancellous, and/or corticocancellous bone, and/or demineralized allograft bone with bone marrow aspirate, or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion.
When used with three (3) screws, interbody implants of <=20° can be used as standalone interbody fusion devices at 1 or 2 contiguous levels without the need for supplemental fixation.
When used with three (3) blades, a combination of three (3) screws and blades, or less than three (3) screws and/or blades, interbody implants must always be used with supplemental fixation.
When used with or without screws and/or blades, interbody implants of >20° must always be used with supplemental fixation.
Transend ALIF Standalone Interbody System
The Transcend ALIF Standalone Interbody System is indicated for spinal fusion procedures from L2 to S1 in skeletally mature patients for the treatment of symptomatic degenerative disc disease (DDD), degenerative spondylolisthesis and/or spinal stenosis at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. When used with screws or blades, these patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved levels.
Additionally, the Transcend ALIF Standalone Interbody System can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity. However, when used in these patients at multiple levels, and for patients with degenerative spondylolisthesis (>Grade 1) and spinal stenosis at one or two adjacent levels, the Transcend ALIF Standalone Interbody System must be used with supplemental spinal fixation systems cleared by the FDA for use in the lumbar spine. The Transcend ALIF Standalone Interbody System is intended for use on patients who have had at least six months of nonoperative treatment. It is intended to be used with autograft and/or allogenic bone graft comprised of cortical, cancellous, and/or corticocancellous bone, and/or demineralized allograft bone with bone marrow aspirate, or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion.
When used with three (3) screws, interbody implants of <=20° can be used as standalone interbody fusion devices at 1 or 2 contiguous levels without the need for supplemental fixation.
When used with three (3) blades, a combination of three (3) screws and blades, or less than three (3) screws and/or blades, interbody implants must always be used with supplemental fixation.
When used with or without screws and/or blades, interbody implants of >20° must always be used with supplemental fixation.
The IdentiTi II and Transcend ALIF Standalone Interbody Systems are integrated intervertebral body fusion systems with integrated screw and/or blade fixation for use in anterior and anterolateral procedures. The subject IdentiTi II interbody spacers are additively manufactured via full-melt powder bed fusion (laser melting) from titanium alloy powder (Ti-6Al-4V ELI) conforming to ASTM F3001. The subject Transcend interbody spacers are an assembly of additively manufactured titanium alloy (Ti-6Al-4V ELI) conforming to ASTM F3001, PEEK (polyetheretherketone) Optima LT1 per ASTM F2026, and titanium alloy (Ti-6Al-4V ELI) conforming to ASTM F136. The IdentiTi II and Transcend ALIF Standalone interbody spacers are offered in various widths, heights, lengths, lordotic, and anti-migration options to accommodate individual patient anatomy and surgeon preference. The IdentiTi II and Transcend interbody implants are not patient matched. All interbody implants feature an internal graft aperture for placement of graft material to promote fusion through the cage. The interbody spacers accept up to three integrated bone screws and/or blades. The screws and blades are made of titanium alloy (Ti-6Al-4V ELI) per ASTM F136 in varying lengths and diameters, and they mitigate the risk of expulsion of the interbody spacers. To mitigate risk of expulsion, the interbody endplates consist of roughed anti-migration surfaces and teeth. Additionally, the additively manufactured portions of the IdentiTi II (full spacer) and Transcend (anterior faceplate) implants are offered with a microporous/macroporous lattice structure that spans the entirety of the implant and extends to the superior and inferior surfaces of the device for biological fixation. The interbody implants are provided terminally sterile via gamma irradiation while the bone screws, blades, and reusable instruments are provided non-sterile to be steam sterilized by the end user.
N/A
Ask a specific question about this device
(181 days)
The CONDUIT™ SYNFIX™ Evolution Secured Spacer System is a stand-alone anterior interbody fusion device with a microscope roughened surface and micro and nano-scale features indicated for use in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade I spondylolisthesis at the involved level(s). The interior of the spacer component of the CONDUIT SYNFIX Evolution can be packed with autograft. If used with less than the four integrated bone screws, or for hyperlordotic implants (>20Deg), implants must be used with supplemental fixation systems cleared by the FDA for use in the lumbosacral spine.
DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment.
The CONDUIT SYNFIX Evolution Secured Spacers are intervertebral body fusion devices intended for lumbar interbody fusion (ALIF). Four Screws are inserted through the anteriorly-located Plate into the adjacent vertebral bodies. The Screws lock securely to the Plate using a tapered-thread locking mechanism.
The CONDUIT SYNFIX Evolution Secured Spacer System is available as non-assembled Cage and Plate components in various heights and geometries to suit individual pathology and anatomical conditions. The Cage and Plate components are intended to be assembled at the point of use prior to implantation.
The CONDUIT SYNFIX Evolution Cages are made from Ti-6Al-4V ELI conforming to ASTM F3001 with an additive manufacturing process (Selective Laser Melting). The design contains solid structures and porous structures. The hollow geometry of the implants allows them to be packed with autogenous bone graft.
The 3D Printed Conduit Cellular Titanium Cages have a microscopic roughened surface with micro and nano-scale features. The micro and nano features are on all surfaces of the Cage, including the superior, inferior, and peripheral surfaces, as well as each member of the internal cell structure.
The provided text is a 510(k) clearance letter for a medical device called the "CONDUIT™ SYNFIX™ Evolution Secured Spacer System," which is an intervertebral body fusion device.
Crucially, this document is for a traditional medical device (an implantable hardware system), not an Artificial Intelligence (AI) or software-as-a-medical-device (SaMD) product.
Therefore, the information requested in the prompt, such as acceptance criteria for AI performance (e.g., sensitivity, specificity, AUC), sample sizes for test/training sets for AI, expert adjudication of AI ground truth, MRMC studies for AI, or standalone AI performance, is not applicable to this submission.
The acceptance criteria and study proving the device meets them, as described in this 510(k), relate to the mechanical performance, material properties, and biocompatibility of the physical interbody fusion device, not an AI algorithm.
Here's an analysis of the provided information relevant to the device's acceptance criteria and studies:
Acceptance Criteria and Device Performance (for a physical medical device):
The document mentions that mechanical testing was performed to "allow comparison with established acceptance criteria." While the specific numerical acceptance criteria (e.g., minimum load to failure, maximum displacement) are not explicitly detailed in the publicly available summary (which is common for 510(k) summaries), the types of tests conducted and the general conclusion indicate that the device met these criteria.
1. Table of Acceptance Criteria and Reported Device Performance (as inferred for a physical device):
| Acceptance Criteria Category | Reported Device Performance (Summary) |
|---|---|
| Mechanical Integrity | Performed per ASTM F2077-24 (Static and Dynamic Axial Compression, Compression Shear), ASTM F2267-24 (Subsidence), and Expulsion |
| MR Compatibility | Evaluated per ASTM F2052-21 (Magnetically Induced Displacement Force), ASTM F2213-17 (Magnetically Induced Torque), ASTM F2119-24 (MR Image Artifact), and ASTM F2182-19e2 (RF-induced Heating) to support MR Conditional labeling. |
| Biocompatibility | Implied to be equivalent to predicate devices, as materials are medical grade titanium (Ti-6Al-4V ELI, Ti-6Al-4V, Ti-6Al-7Nb) which are well-established for this application. Specific tests not detailed but generally required for implants. |
| Sterility | Implied to be equivalent to predicate devices. Specific tests not detailed but generally required. |
2. Sample size used for the test set and the data provenance:
- Sample Size: Not specified in the summary document. For mechanical testing of medical devices, "sample size" refers to the number of physical device units tested for each configuration or stress condition. These studies are typically conducted in a laboratory setting.
- Data Provenance: Laboratory testing (mechanical, MR compatibility). The country of origin of the data is not specified but is typically internal lab data or contracted third-party lab data.
- Retrospective or Prospective: Not applicable in the traditional sense for physical device testing. The tests are designed to assess the device's properties under simulated conditions.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not Applicable. This pertains to AI/software performance evaluation. For physical device testing, "ground truth" is established by calibrated measuring equipment and standardized test methods (e.g., ASTM standards). The "experts" are typically engineers and technicians responsible for conducting the tests and interpreting the results against pre-defined engineering acceptance criteria.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not Applicable. This relates to human expert consensus for AI ground truth labeling. For mechanical testing, the results are objective measurements from testing equipment validated against industry standards.
5. If a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not Applicable. This is specific to AI/software for diagnostic or image-interpretation tasks. No such study was conducted or required for this physical device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not Applicable. This is specific to AI/software.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
- For mechanical testing, the "ground truth" is derived from standardized test methods (e.g., ASTM F2077, F2267) that define specific performance thresholds based on mechanical properties (e.g., ultimate strength, displacement within limits) and material science. The goal is to demonstrate that the device performs equivalently to predicate devices under defined mechanical loads and environmental conditions.
8. The sample size for the training set:
- Not Applicable. This pertains to AI model development.
9. How the ground truth for the training set was established:
- Not Applicable. This pertains to AI model development.
Conclusion from the 510(k) Summary:
The manufacturer "demonstrated substantial equivalence" by showing that their device performs mechanically and in terms of MR compatibility similarly to legally marketed predicate devices, and that the materials and design are appropriate for the intended use. The reliance on established ASTM standards and comparison to predicate devices are the primary methods for demonstrating safety and effectiveness for this type of medical implant.
The summary explicitly states: "No clinical data was necessary to demonstrate substantial equivalence, nor safety and effectiveness of this system." This further confirms that the evaluation was based on non-clinical (mechanical, material, and MR compatibility) testing and comparison to predicates.
Ask a specific question about this device
(26 days)
The ProAM ALIF System is an intervertebral body fusion device intended for spinal fusion procedures at one or two contiguous levels of the lumbosacral (L2-S1) spine in skeletally mature patients with degenerative disc disease (DDD), degenerative spondylolisthesis, and/or spinal stenosis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and imaging studies (radiographs, CT, MRI). DDD patients may also have up to grade 1 spondylolisthesis or retrolisthesis at the involved levels. These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s).
The ProAM ALIF System can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity. However, when used in these patients at multiple levels, and for patients with degenerative spondylolisthesis and spinal stenosis at one or two adjacent levels, the system must be used with supplemental spinal fixation systems cleared by the FDA for fusion of the lumbar spine in addition to the integrated fixation.
Interbody devices used with three fully threaded bone screws and which have a lordotic angle of 20° or less can be used as standalone interbody fusion devices at one or two contiguous levels without the need for supplemental fixation. Interbody devices with a lordotic angle greater than 20° must be used with supplemental spinal fixation systems cleared by the FDA for fusion of the lumbar spine in addition to the integrated fixation. Interbody devices used with fewer than three fixation components (OLIF devices and 2-Hole ALIF devices) or used with anchors or impacted screws must be used with supplemental spinal fixation systems cleared by the FDA for fusion of the lumbar spine in addition to the integrated fixation.
The ProAM ALIF System is intended for use on patients who have had at least six months of nonoperative treatment. It is intended to be used with autograft and/or allogenic bone graft comprised of cortical, cancellous, and/or corticocancellous bone, and/or demineralized allograft bone with bone marrow aspirate.
The ProAM ALIF System is an intervertebral body fusion system that consists of an intervertebral body fusion device manufactured from titanium alloy (Ti-6Al-4V conforming to ASTM F3001) and fixation devices manufactured from titanium alloy (Ti-6Al-4V conforming to ASTM F136).
The intervertebral body fusion device has a cross-sectional shape that is somewhat oval, and the device is otherwise box-shaped in stature with cavities for containment of bone graft material. The device is available in sizes that vary in footprint size, height, and lordotic angle. It is available in configurations meant for direct anterior insertion (ALIF) or oblique-anterior (OLIF) insertion. Devices for implantation via a direct anterior approach are available with 2 or 3 holes through which screws are placed to affix to bone. Devices for implantation via an oblique-anterior approach have 2 holes through which screws are placed to affix to bone. All devices have an open, porous architecture throughout and a pyramidal textured porous surface at the endplate contacting surfaces that helps prevent device migration.
The fixation devices consist of Bone Screws, Impacted Screws, and Curved Anchors. Bone Screws are available in various diameters and lengths and engage the vertebral bodies by passing through the intervertebral body fusion device and into bone. They are turned to fixate to the bone. Impacted Screws are available in various lengths. They are first impacted through the intervertebral body fusion device into bone and then turned to further fixate to the bone. Curved Anchors are available in various lengths and are impacted into the bone through the intervertebral body fusion device.
The purpose of this premarket notification is to add the Impacted Screws and Curved Anchors as fixation devices for the ProAM ALIF System
The provided FDA 510(k) clearance letter for the ProAM ALIF System does not contain any information regarding acceptance criteria or a study proving the device meets those criteria, as typically seen for AI/ML-based medical devices.
This document is a standard 510(k) clearance for an intervertebral body fusion device, which is a physical implant. The information provided focuses on material composition, mechanical testing (static and dynamic cantilever, expulsion), and substantial equivalence to predicate devices based on design, indications for use, and performance of the physical implant.
Therefore, I cannot answer the questions posed regarding acceptance criteria, study details, sample sizes, expert ground truth, MRMC studies, standalone performance, training sets, or adjudication methods, as these are all relevant to the evaluation of AI/ML or software-driven medical devices, not a physical orthopedic implant like the ProAM ALIF System.
The questions are highly appropriate for an AI/ML medical device, but the provided text is a clearance letter for a mechanical implant.
Ask a specific question about this device
(90 days)
The Curiteva Porous PEEK Standalone ALIF devices are intended for use in patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbosacral spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. In addition, these patients may have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). The system spacers are intended to be used with autograft and/or allograft bone (comprised of cancellous and/or corticocancellous bone graft) or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion.
The Curiteva Porous PEEK Standalone ALIF spacer is an interbody fusion device to be used with three titanium alloy screws or anchors. When used with screws, the system is a standalone interbody fusion device. When used with anchors, the system is intended for use with supplemental fixation (e.g. facet screws or posterior fixation). Hyperlordotic implants (>20° lordosis) are intended for use with supplemental fixation (e.g. facet screws or posterior fixation).
The Curiteva Porous PEEK ALIF devices are intended for use at one or more levels of the thoracic spine (T1-T12), thoracolumbar junction (T12-L1), or lumbosacral spine (L1-S1) as an adjunct to fusion in patients with the following indications: degenerative disc disease (DDD), disc herniation (with myelopathy and/or radiculopathy), spondylolisthesis, deformity (degenerative scoliosis or kyphosis), spinal stenosis, and failed previous fusion (pseudarthrosis). DDD is defined as discogenic back pain with degeneration of the disc as confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. The system spacers are intended to be used with autograft and/or allograft bone (comprised of cancellous and/or corticocancellous bone graft) or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion. These devices are intended to be used with supplemental fixation systems that have been cleared for use in the thoracolumbosacral spine (e.g., posterior pedicle screw and rod systems, anterior plate systems, and anterior screw and rod systems). Hyperlordotic interbody devices (>20° lordosis) must be used with at least anterior supplemental fixation.
The Curiteva Porous PEEK Standalone ALIF System implants are provided sterile or non-sterile, are single-use devices, and available in a variety of different footprints, styles and sizes to accommodate the individual pathology and anatomical conditions of the patient. The subject system consists of interbody spacer implants that are generally box-shaped with an open central corridor to permit packing with autograft and/or allograft bone (comprised of cancellous and/or corticocancellous bone graft) or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion. The system also provides screws and anchors that can be inserted through the anterior portion of the spacers into adjacent vertebral bodies for bony fixation.
The Curiteva Porous PEEK Standalone ALIF System implants are manufactured from implant-grade PEEK (per ASTM F2026), Titanium alloy (per ASTM F136), and Tantalum (per ASTM F560). All Porous PEEK interbody spacers are additionally surface treated with a hydroxyapatite (HA) coating that is approximately 20nm thick.
The provided FDA 510(k) Clearance Letter for the Curiteva Porous PEEK Standalone ALIF System does not contain the information requested regarding acceptance criteria and the study that proves the device meets those criteria, as typically found for devices involving AI/ML and diagnostic performance.
This document describes a traditional medical device (an intervertebral body fusion device) primarily evaluated through non-clinical performance testing (mechanical testing) for substantial equivalence to predicate devices. Therefore, many of the requested categories are not applicable to the information provided in this regulatory submission.
Here's an assessment based on the provided text, indicating where information is not present or not applicable:
Description of Acceptance Criteria and the Study that Proves the Device Meets the Acceptance Criteria:
The Curiteva Porous PEEK Standalone ALIF System is an intervertebral body fusion device. Its clearance is based on demonstrating substantial equivalence to legally marketed predicate devices through comparison of design, intended use, material composition, function, range of sizes, and non-clinical performance testing.
The "acceptance criteria" in this context are related to the mechanical performance and safety of the implant, demonstrating that it is as safe and effective as existing cleared devices. The "study" proving this involves a series of in-vitro biomechanical tests.
1. Table of Acceptance Criteria and the Reported Device Performance
| Acceptance Criteria Category | Specific Criteria (Implicit from Testing) | Reported Device Performance (Summary from Document) |
|---|---|---|
| Mechanical Strength | Static Axial Compression (ASTM F2077) | Sufficient for intended use; substantially equivalent to predicates |
| Dynamic Axial Compression (ASTM F2077) | Sufficient for intended use; substantially equivalent to predicates | |
| Static Compression-Shear (ASTM F2077) | Sufficient for intended use; substantially equivalent to predicates | |
| Dynamic Compression-Shear (ASTM F2077) | Sufficient for intended use; substantially equivalent to predicates | |
| Stability | Subsidence (ASTM F2267) | Sufficient for intended use; substantially equivalent to predicates |
| Expulsion Resistance | Expulsion Testing | Sufficient for intended use; substantially equivalent to predicates |
| Material/Wear | Particle Characterization (ASTM F1877) | Sufficient for intended use; substantially equivalent to predicates |
Note: The document states that the results "demonstrate that the strength and performance... is sufficient for its intended use and is substantially equivalent to legally marketed predicate devices." Specific numerical acceptance values or detailed performance data are not provided in this summary, as is typical for 510(k) summary documents which focus on the conclusion of substantial equivalence rather than raw data.
2. Sample Size Used for the Test Set and the Data Provenance
- Sample Size for Test Set: This information is not provided in the document. For biomechanical testing, sample sizes are typically specified per test (e.g., n=6, n=10 per group).
- Data Provenance: The testing is described as "non-clinical testing" and performed on the "subject Curiteva Porous PEEK Standalone ALIF System." This indicates in-vitro (benchtop) testing of the device itself, not human or patient data. Therefore, country of origin or retrospective/prospective classification is not applicable.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of those Experts
- Not Applicable. This type of device (a physical implant) does not involve AI/ML or diagnostic image interpretation where expert ground truth establishment for a test set would be relevant. The "ground truth" for mechanical testing is established by the specified ASTM standards and the physical properties of the material and design.
4. Adjudication Method for the Test Set
- Not Applicable. As there is no ground truth established by experts, an adjudication method is not relevant.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done
- No. An MRMC study is relevant for diagnostic devices (e.g., AI for image interpretation) to compare human performance with and without AI assistance. This device is an implantable surgical device, so an MRMC study was not performed, nor is it applicable.
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was Done
- Not Applicable. This concept applies to AI/ML algorithms, not to a physical implantable device.
7. The Type of Ground Truth Used
- For the non-clinical performance testing, the "ground truth" is defined by the established engineering and biomechanical principles and standards (e.g., ASTM F2077, ASTM F2267, ASTM F1877). These standards define acceptable methodologies and performance metrics for intervertebral fusion devices.
8. The Sample Size for the Training Set
- Not Applicable. This device is not an AI/ML algorithm that requires a "training set" of data.
9. How the Ground Truth for the Training Set Was Established
- Not Applicable. As there is no training set, there is no ground truth to be established for it.
Ask a specific question about this device
(31 days)
The OneLIF™ Interbody Fusion System is an interbody fusion device system indicated for use in skeletally mature patients at one or more levels of the lumbosacral spine (L2-S1), in patients with the following indications: degenerative disc disease (DDD) defined as back pain with degeneration of the disc confirmed by patient history and radiographic studies, spinal deformity (degenerative scoliosis or kyphosis), spondylolisthesis or retrolisthesis, and failed previous fusion (pseudoarthrosis). Patients should have received 6 months of nonoperative treatment prior to treatment with the devices.
The OneLIF™ Interbody Fusion System is intended to be used with or without the screws which accompany the implants. These devices are intended for stand-alone use in patients with DDD or degenerative spondylolisthesis at one or two contiguous levels only when used with at least three screws per implant (including at least one screw in each endplate) and when ≤ 20° lordotic implants are used. When used at more than 2 contiguous levels, or for treatment of conditions other than DDD or degenerative spondylolisthesis, or with fewer than 3 accompanying screws, or when using implants greater than a 20° lordotic angle, the system must be supplemented by posterior fixation (e.g., pedicle screw system) cleared for use in the lumbar spine.
The implants can be placed via a variety of open or minimally invasive approaches. These include anterior and oblique approaches. The implant is designed for use with autograft bone and/or allogenic bone graft comprised of cancellous or corticocancellous bone graft.
The OneLIF™ Interbody Fusion System includes interbody cages and retention plates designed for anterior or oblique insertion techniques. The purpose of this submission is to add additional implant sizes to the previously cleared system.
The provided text is a 510(k) clearance letter for a medical device called the OneLIF™ Interbody Fusion System. It does not contain information about acceptance criteria and a study proving device performance in the way typically observed for AI/ML-based devices.
Instead, this document focuses on demonstrating substantial equivalence to existing legally marketed predicate devices, which is a common pathway for clearance of conventional medical devices like interbody fusion systems. The "performance data" mentioned refers to mechanical testing of the physical hardware, not a clinical performance study with human subjects, AI accuracy, or human-in-the-loop metrics.
Therefore, many of the requested categories cannot be populated from the provided text.
Here's a breakdown based on the given document:
1. A table of acceptance criteria and the reported device performance
| Acceptance Criteria (from text) | Reported Device Performance (from text) |
|---|---|
| Mechanical performance per ASTM F2077 | "Previously tested per ASTM F2077 under K211769" |
| No new worst-cases with modified implant designs | "Dynamic testing per ASTM F2077 and engineering analysis was performed on the modified subject implant designs to confirm no new worst-cases" |
| Substantial equivalence to predicate device (K211769) | "Overall technology characteristics and mechanical performance data lead to the conclusion that the OneLIF™ Interbody Fusion System is substantially equivalent to the predicate device." |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Not applicable / Not mentioned. The "test set" here refers to physical implant designs and their mechanical properties, not a clinical dataset of patients or images.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not applicable / Not mentioned. Ground truth in this context would be related to the mechanical properties and compliance with ASTM standards, typically assessed by engineers or testing laboratories, not clinical experts.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not applicable / Not mentioned. Adjudication methods are typically used for clinical endpoints or image interpretation studies.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No. This is a hardware device (interbody fusion system), not an AI-assisted diagnostic or treatment system.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- No. This is a hardware device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- Not applicable / Not mentioned for clinical ground truth. For the mechanical testing, the "ground truth" would be established by the ASTM F2077 standard's specifications for mechanical behavior (e.g., strength, fatigue life) under specific loading conditions.
8. The sample size for the training set
- Not applicable / Not mentioned. This device does not use a training set in the context of AI/ML.
9. How the ground truth for the training set was established
- Not applicable / Not mentioned. This device does not use a training set.
Ask a specific question about this device
(131 days)
The Stable-L Lumbar Interbody System is indicated for spinal fusion procedures in skeletally mature patients. Stable-L is designed for use with autogenous and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft. The devices are to be used in patients who have had at least six months of non-operative treatment. Stable-L is intended for use in interbody fusions in the lumbar spine from L2 to S1 at one or two adjacent levels in the treatment of symptomatic degenerative disc disease (DDD). The DDD patients may also have up to Grade 1 spondylolisthesis at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies.
When used with the provided blades, supplemental fixation is required.
When used with the provided screws, each interbody fusion device having a lordotic angle 20° or less is intended to be used with the bone screws provided and requires no additional fixation. However, for hyperlordotic devices (> 20° lordosis), due to the increased risk of anterior migration with hyperlordotic implants, the devices should be used with the bone screws provided and supplemental fixation such as posterior fixation.
The existing Stable-L Interbody System is a lumbar interbody system including an interbody cage additively manufactured from titanium alloy (Ti-6Al-4V ELI) per ASTM F3001 and associated bone screws manufactured from titanium alloy per ASTM F136. The device is offered in a variety of sizes to accommodate patient anatomy. Stable-L Interbody devices are provided either non-sterile to be sterilized by the end-user or sterile via gamma irradiation.
The subject submission seeks to add interbody cages with integrated blade fixation as well as expand the size offerings available. Additionally, this submissions seeks to gain clearance for a minor modification to the Stable L Cover Plate as well as extend the shelf life of devices provided sterile.
The provided FDA 510(k) clearance letter and summary for the "Stable-L Lumbar Interbody System" describe a spinal implant, not an AI/software medical device.
Therefore, the requested information regarding AI/software device acceptance criteria, performance studies (e.g., MRMC studies, ground truth establishment, training/test set details), and expert qualifications is not applicable to this document.
The document focuses on the physical characteristics and mechanical performance of the intervertebral body fusion device. The "Performance Testing" section mentions:
- Expulsion
- Blade compression/collapse
- Confirmatory Dynamic compression and dynamic shear per ASTM F2077
- Confirmatory insertion and screw pull out testing
These tests are standard for implantable mechanical devices and aim to demonstrate that the device performs equivalently to previously cleared predicate devices and does not raise new questions of safety or effectiveness. The criteria for these tests would typically involve specific force, displacement, or fatigue cycle thresholds, and the reported performance would be the successful completion of these predefined values.
In summary, this document does not contain the information needed to answer your request about AI/software device acceptance criteria and performance studies.
Ask a specific question about this device
(173 days)
Ventris anterior cages are indicated for intervertebral body fusion of the spine in skeletally mature patients who have had at least six months of non-operative treatment. The device systems are designed for use with allogenic bone graft comprised of cancellous and/or corticocancellous bone graft and/or autograft to facilitate fusion. One device is used per intervertebral body space. Ventris anterior cages are intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies.
Ventris 2-screw anterior cages may be used as a stand alone device only when two (2) vertebral body bone screws are used. Ventris 4-screw anterior cages may be used as a stand alone device only when at least two (2) vertebral body bone screws are inserted in the two medial fixation holes with one inferior and one superior screw trajectory. If the physician chooses to use Ventris anterior cages with fewer than two (2) screws in the two medial fixation holes with one inferior and one superior screw trajectory, then an additional supplemental spinal fixation system cleared for use in the lumbosacral spine must be used. Ventris anterior cages are intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, in patients having an ODI >40 and diagnosed with severe symptomatic adult spinal deformity (ASD) conditions. When used to treat ASD, or when cages with more than 20° of lordosis are used, Ventris anterior cages must be used with a supplemental spinal fixation system cleared for use in the lumbosacral spine and cannot be used as a stand alone.
Ventris anterolateral cages are indicated for intervertebral body fusion of the spine in skeletally mature patients who have had at least six months of non-operative treatment. The device systems are designed for use with allogenic bone graft comprised of cancellous and/or corticocancellous bone graft and/or autograft to facilitate fusion. One device is used per intervertebral body space. Ventris anterolateral cages are intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, in patients having an ODI >40 and diagnosed with severe symptomatic adult spinal deformity (ASD) conditions. Ventris anterolateral cages are intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. Ventris anterolateral cages are intended to be used with a supplemental spinal fixation system cleared for use in the lumbosacral spine.
Ventris lateral cages are indicated for intervertebral body fusion of the spine in skeletally mature patients who have had at least six months of non-operative treatment. The device systems are designed for use with allogenic bone graft comprised of cancellous and/or corticocancellous bone graft and/or autograft to facilitate fusion. One device is used per intervertebral body space. Ventris lateral cages are intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, in patients having an ODI >40 and diagnosed with severe symptomatic adult spinal deformity (ASD) conditions. Ventris lateral cages are intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. Ventris lateral cages are intended to be used with a supplemental spinal fixation system cleared for use in the lumbosacral spine.
The Ventris system consists of intervertebral body fusion devices intended for lumbar interbody fusion using an anterior lumbar interbody fusion surgical approach (ALIF), anterolateral (i.e., oblique) lumbar interbody fusion surgical approach (AOLIF), or a lateral lumbar interbody fusion surgical approach (LLIF). The devices are intended to improve stability of the spine while supporting fusion. The Ventris constructs are intended for use at one or two contiguous levels in the lumbar spine (L2-S1). The components are offered in different shapes and sizes to meet the requirements of the individual patient's anatomy and are provided sterile.
Ventris cages are available in six configurations: Ventris Ti Interbody anterior four-hole constructs for ALIF approach, Ventris Ti FRA Interbody anterior fully round ALIF (FRA) constructs for ALIF approach, Ventris Ti Open Interbody anterior two-hole constructs for ALIF approach, Ventris Ti AL Interbody anterolateral (i.e., oblique) constructs for OLIF approach, Ventris Ti Lateral FX two-screw lateral constructs for LLIF approach, and Ventris Ti Lateral Interbody lateral constructs for LLIF approach. All cages are also available with increased surface area options that provide additional endplate surface area. For the increased surface area cages, the outer footprint remains the same, but the volume of the internal graft window is reduced, creating more endplate surface area.
Ventris cages are secured on the vertebral bodies using bone screws. A cover plate assembly prevents the screws from backing out after insertion. The cages and cover plates are made of titanium alloy (Ti-6Al-4V ELI) per ASTM F3001 Standard Specification for Additive Manufacturing Titanium-6 Aluminum-4 Vanadium ELI (Extra Low Interstitial) with Powder Bed Fusion. The bone screw and cover plate screws are made from titanium alloy (Ti-6Al-4V ELI) per ASTM F136 Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications. All anterior and anterolateral constructs are zero profile, reducing potential for vessel interference with the anterior column.
All Ventris cages and cover plates are manufactured using the L-PBF (laser powder bed fusion) additive manufacturing method. L-PBF allows for the formation of solid, non-porous cages with a layered porous lattice structure on the surfaces of the components, including the surfaces of the interior graft window. This intricate structure facilitates bone in-growth by providing a larger surface of implant/bone contact than a buffed surface. Interbody cages and bone screws may also be titanium anodized to allow for identification of various heights/sizes by color.
Ventris Ti Interbody cages, Ventris Ti FRA Interbody cages, Ventris Ti Open Interbody cages, Ventris Ti AL Interbody cages, Ventris Ti Lateral FX Interbody cages, Ventris Ti Lateral Interbody cages, and Ventris bone screws are also available with a Proximal HA Surface (hydroxyapatite) coating. Cages and bone screws are coated with 20mm HA layer composed of crystalline hydroxyapatite particles that mimic human bone tissue through shape, composition, and structure. This surface treatment increases implant anchoring by facilitating osseointegration and enhancing early bone growth. All Ventris devices are only available sterile packaged.
Non-sterile, reusable surgical instruments to support implantation of the system are provided for use with Ventris devices are provided in steam sterilization trays.
The provided FDA 510(k) clearance letter and summary for the Ventris IBFD focuses primarily on demonstrating substantial equivalence to predicate devices through non-clinical performance testing (mechanical, biocompatibility, sterilization, and packaging validation), rather than outlining specific clinical acceptance criteria for a diagnostic AI device or a direct human observational study.
Therefore, the requested information regarding human reader studies, training data, and ground truth establishment, which are typical for AI/diagnostic device clearances, is not present in the provided document. The Ventris IBFD is a physical intervertebral body fusion device, not a diagnostic AI or imaging analysis tool.
However, I can extract the acceptance criteria and performance related to the mechanical and physical properties of the device as described in the non-clinical testing.
Here's the information based on the provided text, with the understanding that this is for a physical medical implant and not a diagnostic AI system:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria Category | Specific Test Standard / Requirement | Device Performance / Conclusion |
|---|---|---|
| Mechanical Strength | Static Compression (ASTM F2077) | Met required mechanical strength criteria, demonstrated substantial equivalence to predicate devices. |
| Static Compression Shear (ASTM F2077) | Met required mechanical strength criteria, demonstrated substantial equivalence to predicate devices. | |
| Dynamic Compression (ASTM F2077) | Met required mechanical strength criteria, demonstrated substantial equivalence to predicate devices. | |
| Dynamic Compression Shear (ASTM F2077) | Met required mechanical strength criteria, demonstrated substantial equivalence to predicate devices. | |
| Subsidence Resistance | Subsidence Testing (ASTM F2267) | Met required mechanical strength criteria, demonstrated substantial equivalence to predicate devices. |
| Material Properties | Tensile Testing (ASTM F3001) | Performance demonstrated substantial equivalence to predicate devices. |
| Microstructure Assessment (ASTM F3001) | Performance demonstrated substantial equivalence to predicate devices. | |
| Chemical Composition Assessment (ASTM F3001) | Performance demonstrated substantial equivalence to predicate devices. | |
| Biocompatibility | (Testing conducted) | Successfully validated. |
| Sterilization | (Validation conducted) | Successfully validated. |
| Packaging | (Validation conducted) | Successfully validated. |
| Risk Assessment | (Worst-case analysis) | No new or increased risks identified. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: Not explicitly stated as a number of devices or test articles for each mechanical test. The text mentions "worst-case subject device" and "all process steps, including HIP and Non-HIP test articles" were taken into account. This implies a selection of devices representing various configurations and manufacturing processes were tested.
- Data Provenance: The data comes from non-clinical laboratory testing performed by the manufacturer, Acuity Surgical Devices LLC, or a contract lab. The provenance is internal testing to establish product performance against industry standards. There is no mention of country of origin for test data, as it's not a clinical study. It's not retrospective or prospective in the clinical sense, but rather pre-market engineering validation.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts
- Not applicable. For mechanical testing of a medical device, "ground truth" is established by adherence to recognized ASTM standards and validated testing methodologies. The "experts" would be the engineers and technicians conducting the tests and interpreting the results against the standard's specifications. The document does not specify their number or qualifications but implies standard engineering and regulatory expertise would be involved in such validation.
4. Adjudication Method for the Test Set
- Not applicable. Adjudication methods like 2+1 or 3+1 are used in clinical studies involving multiple human readers for diagnostic interpretation. For mechanical testing, the "adjudication" is inherently built into the pass/fail criteria of the ASTM standards and the rigorous interpretation of raw data against these predefined limits.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- No. An MRMC study is not relevant for this type of device (an intervertebral body fusion device) which is a physical implant, not a diagnostic or AI-assisted system. No human readers or AI assistance are mentioned in the context of device performance improvement for an MRMC study.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
- Not applicable. This device is a physical implant. There is no "algorithm only" performance to evaluate. Its "performance" is its mechanical integrity and biological compatibility when implanted.
7. Type of Ground Truth Used
- The "ground truth" for the non-clinical tests is based on established industry standards (ASTM F2077, ASTM F2267, ASTM F3001) for the mechanical, material, and performance characteristics of intervertebral body fusion devices. These standards define the acceptable range of performance for such devices.
8. Sample Size for the Training Set
- Not applicable. This device is a physical implant, not an AI model. There is no concept of a "training set" in the context of its validation for FDA clearance. The design and manufacturing processes are developed based on engineering principles and existing knowledge, not through machine learning training data.
9. How the Ground Truth for the Training Set Was Established
- Not applicable. As there is no training set for an AI model, there is no ground truth, in that sense, to establish.
Ask a specific question about this device
(24 days)
The Innovasis AxTiHA® Stand-Alone ALIF System is an intervertebral body fusion device for use in patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbar spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. In addition, these patients may have up to a Grade 1 spondylolisthesis or retrolisthesis at the involved levels(s). These implants are used to facilitate fusion in the lumbar spine and are placed via an anterior (ALIF) approach. Hyperlordotic implants (those with a lordotic angle greater than or equal to 20°) are indicated for use with a supplemental spinal fixation system such as the Innovasis® Excella® Spinal System. The AxTiHA Stand-Alone interbody implants with a lordotic angle less than 20°, when used with all three internal fixation screws, do not require use of supplemental fixation. The interior of the AxTiHA implant is intended to be packed with autograft or allogenic bone graft composed of cancellous and/or corticocancellous bone graff.
The AxTiHA system is for Anterior Lumbar Interbody Fusion (ALIF). The IBF implants are an additive manufactured device comprised of Ti-6Al-4V (ELI) per ASTM F3001 and Hydroxyapatite (HA) and are available in multiple size options to facilitate a more precise anatomical fit. The IBF implants have a tapered leading edge which aids in imited anatomical space, feature a bi-convex profile to match the anatomy, and include anti-migration features to ensure implant stability during the fusion process. The large graft cavity and open geometric Tetracell® Technology structure provide increased volume for autograft loading and bone through-growth. The IBF devices include integrated fixation by way of three converging bone screws and optional screw anti-backout locking clips manufactured from Ti-6Al-4V (ELI) per ASTM F136. Reusable instruments to support the ALIF surgery are provided with the implants in sterilization travs.
The provided FDA 510(k) clearance letter for the AxTiHA® Stand-Alone ALIF System does not contain information related to a study proving the device meets specific acceptance criteria in the context of AI/software performance.
The document details the clearance of a medical device implant (an intervertebral body fusion device) based on substantial equivalence to predicate devices, primarily considering its mechanical and material properties, and intended use.
Therefore, many of the requested points regarding acceptance criteria, study design, and ground truth for AI/software evaluation cannot be extracted from this document.
The document states: "Engineering analysis of the subject device indicates that additional verification testing is not required to demonstrate substantial equivalence. Previous evaluations and performance testing (e.g., design verification and validation) performed on the predicate devices apply to the subject devices." This implies the submission focused on the physical characteristics and prior testing of similar devices, rather than a novel AI/software component requiring performance studies with human readers or standalone algorithm evaluations.
Based on the provided text, here's what can be inferred or explicitly stated, and what cannot:
Inferences from the Provided Document:
- Device Type: The AxTiHA® Stand-Alone ALIF System is an intervertebral body fusion device (an implant), not an AI/software medical device.
- Clearance Basis: Substantial equivalence to predicate devices (K212967 and K160605) based on indications for use, intended use, design, technology, and performance, primarily through engineering analysis and prior performance testing of predicates.
- No mention of AI/Software Performance: There is no indication or mention of any AI component, software algorithm, or diagnostic capabilities that would necessitate the types of performance studies (e.g., MRMC, standalone algorithm performance, expert ground truth establishment for image interpretation) you've inquired about.
Responding to Your Specific Questions Based on the Provided Text:
Since the document describes an implantable device and not an AI/software component, most of your questions are not applicable or cannot be answered.
-
A table of acceptance criteria and the reported device performance
- Not Applicable. The document does not describe specific performance metrics for an AI/software component. For the physical implant, the "acceptance criteria" are generally met through demonstration of substantial equivalence via engineering analysis and previous testing on predicate devices, rather than a specific performance table in this document.
-
Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Not Applicable. No test set for AI/software performance is mentioned. The clearance relies on substantial equivalence and engineering analysis.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not Applicable. No ground truth establishment for AI/software performance is mentioned.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not Applicable. No adjudication for AI/software performance is mentioned.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not Applicable. No MRMC study mentioned, as there is no AI assistance component.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not Applicable. No standalone algorithm performance study mentioned.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- Not Applicable. No ground truth for AI/software performance is mentioned. The ground truth for the physical device's safety and effectiveness would be based on established engineering principles, biocompatibility testing (likely referenced from predicate devices), and clinical performance of predicate devices.
-
The sample size for the training set
- Not Applicable. No training set for an AI model is mentioned.
-
How the ground truth for the training set was established
- Not Applicable. No training set for an AI model is mentioned.
In summary, the provided FDA 510(k) clearance letter pertains to a an implantable intervertebral body fusion device and provides no information related to the performance of an AI/software component. Therefore, the questions regarding acceptance criteria and study details for AI/software performance cannot be answered from this specific document.
Ask a specific question about this device
Page 1 of 20