K Number
K250603
Manufacturer
Date Cleared
2025-03-24

(24 days)

Product Code
Regulation Number
888.3080
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Innovasis AxTiHA® Stand-Alone ALIF System is an intervertebral body fusion device for use in patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbar spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. In addition, these patients may have up to a Grade 1 spondylolisthesis or retrolisthesis at the involved levels(s). These implants are used to facilitate fusion in the lumbar spine and are placed via an anterior (ALIF) approach. Hyperlordotic implants (those with a lordotic angle greater than or equal to 20°) are indicated for use with a supplemental spinal fixation system such as the Innovasis® Excella® Spinal System. The AxTiHA Stand-Alone interbody implants with a lordotic angle less than 20°, when used with all three internal fixation screws, do not require use of supplemental fixation. The interior of the AxTiHA implant is intended to be packed with autograft or allogenic bone graft composed of cancellous and/or corticocancellous bone graff.

Device Description

The AxTiHA system is for Anterior Lumbar Interbody Fusion (ALIF). The IBF implants are an additive manufactured device comprised of Ti-6Al-4V (ELI) per ASTM F3001 and Hydroxyapatite (HA) and are available in multiple size options to facilitate a more precise anatomical fit. The IBF implants have a tapered leading edge which aids in imited anatomical space, feature a bi-convex profile to match the anatomy, and include anti-migration features to ensure implant stability during the fusion process. The large graft cavity and open geometric Tetracell® Technology structure provide increased volume for autograft loading and bone through-growth. The IBF devices include integrated fixation by way of three converging bone screws and optional screw anti-backout locking clips manufactured from Ti-6Al-4V (ELI) per ASTM F136. Reusable instruments to support the ALIF surgery are provided with the implants in sterilization travs.

AI/ML Overview

The provided FDA 510(k) clearance letter for the AxTiHA® Stand-Alone ALIF System does not contain information related to a study proving the device meets specific acceptance criteria in the context of AI/software performance.

The document details the clearance of a medical device implant (an intervertebral body fusion device) based on substantial equivalence to predicate devices, primarily considering its mechanical and material properties, and intended use.

Therefore, many of the requested points regarding acceptance criteria, study design, and ground truth for AI/software evaluation cannot be extracted from this document.

The document states: "Engineering analysis of the subject device indicates that additional verification testing is not required to demonstrate substantial equivalence. Previous evaluations and performance testing (e.g., design verification and validation) performed on the predicate devices apply to the subject devices." This implies the submission focused on the physical characteristics and prior testing of similar devices, rather than a novel AI/software component requiring performance studies with human readers or standalone algorithm evaluations.

Based on the provided text, here's what can be inferred or explicitly stated, and what cannot:


Inferences from the Provided Document:

  • Device Type: The AxTiHA® Stand-Alone ALIF System is an intervertebral body fusion device (an implant), not an AI/software medical device.
  • Clearance Basis: Substantial equivalence to predicate devices (K212967 and K160605) based on indications for use, intended use, design, technology, and performance, primarily through engineering analysis and prior performance testing of predicates.
  • No mention of AI/Software Performance: There is no indication or mention of any AI component, software algorithm, or diagnostic capabilities that would necessitate the types of performance studies (e.g., MRMC, standalone algorithm performance, expert ground truth establishment for image interpretation) you've inquired about.

Responding to Your Specific Questions Based on the Provided Text:

Since the document describes an implantable device and not an AI/software component, most of your questions are not applicable or cannot be answered.

  1. A table of acceptance criteria and the reported device performance

    • Not Applicable. The document does not describe specific performance metrics for an AI/software component. For the physical implant, the "acceptance criteria" are generally met through demonstration of substantial equivalence via engineering analysis and previous testing on predicate devices, rather than a specific performance table in this document.
  2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not Applicable. No test set for AI/software performance is mentioned. The clearance relies on substantial equivalence and engineering analysis.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not Applicable. No ground truth establishment for AI/software performance is mentioned.
  4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable. No adjudication for AI/software performance is mentioned.
  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. No MRMC study mentioned, as there is no AI assistance component.
  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. No standalone algorithm performance study mentioned.
  7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Not Applicable. No ground truth for AI/software performance is mentioned. The ground truth for the physical device's safety and effectiveness would be based on established engineering principles, biocompatibility testing (likely referenced from predicate devices), and clinical performance of predicate devices.
  8. The sample size for the training set

    • Not Applicable. No training set for an AI model is mentioned.
  9. How the ground truth for the training set was established

    • Not Applicable. No training set for an AI model is mentioned.

In summary, the provided FDA 510(k) clearance letter pertains to a an implantable intervertebral body fusion device and provides no information related to the performance of an AI/software component. Therefore, the questions regarding acceptance criteria and study details for AI/software performance cannot be answered from this specific document.

FDA 510(k) Clearance Letter - AxTiHA® Stand-Alone ALIF System

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

March 24, 2025

Innovasis
Mr. Mike Thomas
Director Regulatory Affairs
614 E 3900 S
Salt Lake City, Utah 84107

Re: K250603
Trade/Device Name: AxTiHA® Stand-Alone ALIF System
Regulation Number: 21 CFR 888.3080
Regulation Name: Intervertebral Body Fusion Device
Regulatory Class: Class II
Product Code: OVD
Dated: February 27, 2025
Received: February 28, 2025

Dear Mr. Mike Thomas:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

Page 2

K250603 - Mr. Mike Thomas Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Page 3

K250603 - Mr. Mike Thomas Page 3

Sincerely,

Katherine D. Kavlock -S

for
Brent Showalter, Ph.D.
Assistant Director
DHT6B: Division of Spinal Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

FORM FDA 3881 (7/17) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2020
See PRA Statement below.

510(k) Number (if known): K250603

Device Name: AxTiHA® Stand-Alone ALIF System

Indications for Use (Describe)

The Innovasis AxTiHA Stand-Alone ALIF System is an intervertebral body fusion device for use in patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbar spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. In addition, these patients may have up to a Grade 1 spondylolisthesis or retrolisthesis at the involved levels(s). These implants are used to facilitate fusion in the lumbar spine and are placed via an anterior (ALIF) approach. Hyperlordotic implants (those with a lordotic angle greater than or equal to 20°) are indicated for use with a supplemental spinal fixation system such as the Innovasis® Excella® Spinal System. The AxTiHA Stand-Alone interbody implants with a lordotic angle less than 20°, when used with all three internal fixation screws, do not require use of supplemental fixation. The interior of the AxTiHA implant is intended to be packed with autograft or allogenic bone graft composed of cancellous and/or corticocancellous bone graft.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.


This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

K250603
Page 1 of 2

510(k) Summary

510(k) #: K250603
Prepared on: 2025-02-27

Contact Details

21 CFR 807.92(a)(1)

Applicant NameInnovasis
Applicant Address614 E 3900 S Salt Lake City UT 84107 United States
Applicant Contact Telephone8012612236
Applicant ContactMr. Mike Thomas
Applicant Contact Emailmthomas@innovasis.com

Device Name

21 CFR 807.92(a)(2)

Device Trade NameAxTiHA® Stand-Alone ALIF System
Common NameIntervertebral Fusion Device With Integrated Fixation, Lumbar
Classification NameIntervertebral body fusion device
Regulation Number888.3080
Product Code(s)OVD

Legally Marketed Predicate Devices

21 CFR 807.92(a)(3)

Predicate #Predicate Trade Name (Primary Predicate is listed first)Product Code
K212967AxTiHA® Stand-Alone ALIF SystemOVD
K160605MectaLIF Anterior Stand-AloneOVD

Device Description Summary

21 CFR 807.92(a)(4)

The AxTiHA system is for Anterior Lumbar Interbody Fusion (ALIF). The IBF implants are an additive manufactured device comprised of Ti-6Al-4V (ELI) per ASTM F3001 and Hydroxyapatite (HA) and are available in multiple size options to facilitate a more precise anatomical fit. The IBF implants have a tapered leading edge, which aids in implant insertion due to the limited anatomical space, feature a bi-convex profile to match the anatomy, and include anti-migration features to ensure implant stability during the fusion process. The large graft cavity and open geometric Tetracell® Technology structure provide increased volume for autograft loading and bone through-growth. The IBF devices include integrated fixation consisting of three converging bone screws and optional screw anti-backout locking clips manufactured from Ti-6Al-4V (ELI) per ASTM F136. Reusable instruments to support the ALIF surgery are provided with the implants in sterilization trays.

Intended Use/Indications for Use

21 CFR 807.92(a)(5)

The Innovasis AxTiHA Stand-Alone ALIF System is an intervertebral body fusion device for use in patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbar spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. In addition, these patients may have up to a Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These implants are used to facilitate fusion in the lumbar spine and are placed via an anterior (ALIF) approach. Hyperlordotic implants (those with a lordotic angle greater than or equal to 20°) are indicated for use with a supplemental spinal fixation system such as the Innovasis® Excella® Spinal System. The AxTiHA Stand-Alone interbody implants with a lordotic angle less than 20°, when used with all three internal fixation screws, do not require use of supplemental fixation. The interior of the AxTiHA implant is intended to be packed with autograft or allogenic bone graft composed of cancellous and/or corticocancellous bone graft.

Page 1 of 2

Page 6

K250603
Page 2 of 2

Indications for Use Comparison

21 CFR 807.92(a)(5)

The indications for use statement of the subject device is substantially equivalent to the predicate devices (K212967 and K160605).

Technological Comparison

21 CFR 807.92(a)(6)

The subject device has substantially equivalent technological characteristics (i.e., design, material, chemical composition, principle of operation, energy source, etc.) as the predicate devices (K212967 and K160605).

Non-Clinical and/or Clinical Tests Summary & Conclusions

21 CFR 807.92(b)

Engineering analysis of the subject device indicates that additional verification testing is not required to demonstrate substantial equivalence. Previous evaluations and performance testing (e.g., design verification and validation) performed on the predicate devices apply to the subject devices. Based on the information presented above and within the submission, the subject device is substantially equivalent to the predicate devices regarding indications for use, intended use, design, technology, and performance.


K250603

Page 2 of 2

N/A