Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K060677

    Validate with FDA (Live)

    Device Name
    N LATEX CDT
    Manufacturer
    Date Cleared
    2006-08-25

    (164 days)

    Product Code
    Regulation Number
    862.1360
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    In vitro diagnostic for the quantitative determination of carbohydrate-deficient transferrin (CDT) in human serum by means of particle-enhanced immunonephelometry using the BN™ II and BN ProSpec® System. The N Latex CDT assay must be run concurrently with the N Antisera to Human Transferrin assay so that the result can be expressed as a relative ratio, i.e., %CDT of the total transferrin. The calculation of %CDT can be used as a tool to identify possible chronic heavy alcohol consumption.

    Device Description

    The CDT in the sample competes with CDT-coated polystyrene particles for the bond to specific monoclonal antibodies against human CDT, which are likewise bound to polystyrene particles. In the presence of CDT in the sample, there is no or little aggregation of the polystyrene particles. In the absence of CDT in the sample, the polystyrene particles aggregate. The higher the CDT content in the assay, the lower the scattered light signal. The evaluation is performed by comparison with a standard of known concentration.

    AI/ML Overview

    The provided text describes the 510(k) summary for the N Latex CDT Kit, which is an in vitro diagnostic device. The study presented focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than defining and meeting new acceptance criteria for standalone performance or comparative effectiveness.

    Here's an analysis based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document does not explicitly state "acceptance criteria" in the traditional sense of pre-defined thresholds for performance metrics. Instead, it demonstrates performance by comparing the new device (N Latex CDT assay) against a predicate device (Axis-Shield %CDT assay). The criteria for "acceptance" appear to be based on a high correlation and agreement with the predicate.

    Performance MetricAcceptance Criteria (Implicit)Reported Device Performance (N Latex CDT vs. Predicate)
    CorrelationClose to 1.0Correlation coefficient: 0.99
    Agreement (Linear Regression)Slope close to 1, Intercept close to 0y = 0.720x + 0.75 %CDT

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size for Test Set: 116 serum samples
    • Data Provenance: The document does not specify the country of origin of the data or whether the study was retrospective or prospective. It only states that 116 serum samples were "evaluated."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    Not applicable. The study does not establish "ground truth" through expert consensus. Instead, it uses a legally marketed predicate device's results as the reference for comparison.

    4. Adjudication Method for the Test Set:

    Not applicable. There was no expert adjudication process described, as the comparison was against a predicate device's results.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This device is an in vitro diagnostic assay, not an AI-assisted diagnostic tool involving human readers. Therefore, an MRMC study or evaluation of human reader improvement with AI assistance is not relevant to this submission.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    The device's performance, as described by the regression analysis (y = 0.720x + 0.75 %CDT, correlation coefficient 0.99), is essentially its "standalone" performance when compared against the predicate device. It's a chemical assay, so its performance is inherent to the assay itself.

    7. The Type of Ground Truth Used:

    The "ground truth" in this context is the results obtained from the predicate device (Axis-Shield %CDT Assay) for the same serum samples. This is a form of reference standard comparison rather than ground truth established by pathology or clinical outcomes.

    8. The Sample Size for the Training Set:

    Not applicable. This is an in vitro diagnostic assay, not a machine learning or AI-based device that typically requires a distinct training set in the same manner. The "training" of such a device generally refers to its chemical formulation and calibration, which is not detailed in terms of a "training set sample size" in this document.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as there is no mention or implication of a "training set" in the context of machine learning or AI. The development of an in vitro diagnostic largely relies on chemical and biological principles, calibration, and validation, rather than ground truth established for a training dataset in the AI sense.

    Ask a Question

    Ask a specific question about this device

    K Number
    K001943

    Validate with FDA (Live)

    Date Cleared
    2000-08-23

    (58 days)

    Product Code
    Regulation Number
    862.1360
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K992502

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1999-12-21

    (148 days)

    Product Code
    Regulation Number
    862.1360
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Axis ® %CDT Turbidimetric Immunoassay (TIA) is intended for the quantitative measurement of carbohydrate deficient transferrin (CDT) in human serum, as a tool to identify possible chronic heavy alcohol consumption.

    The Axis ® %CDT Turbidimetric Immunoassay is designed for the quantitative determination of carbohydrate deficient transferrin (CDT) in human serum.

    Device Description

    The Axis® %CDT is a heterogeneous immunoassay with column separation followed by turbidimetric measurement. Serum transferrin in the sample is saturated with Fe ". The mixture is applied to an ion-exchange column. Due to the different amounts of sialic residues on transferrin, the isoforms carry different charges and are separated in the column. The CDT isoforms are eluted. The CDT content of the collected eluate is determined by turbidimetric measurement.

    The eluted CDT isoforms form immune complexes with anti-transferrin antibodies. Total transferrin content of the sample is determined separately, using the same anti- transferrin antibodies. The measurements are evaluated using a calibration curve, and the Axis® %CDT value is calculated.

    AI/ML Overview

    The Axis® %CDT Turbidimetric Immunoassay (TIA) is designed for the quantitative determination of carbohydrate deficient transferrin (CDT) in serum to identify possible chronic heavy alcohol consumption.

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria CategorySpecific CriteriaReported Device PerformanceComments
    Clinical Performance (vs. GGT)Sensitivity at 5% CDT cut-offMales: 0.70 Females: 0.55Compared to Gamma-glutamyl transferase (GGT)
    Specificity at 5% CDT cut-offMales: 0.93 Females: 0.91Compared to GGT
    Correlation with GGTLow and not significantConsistent with literature
    Area Under the Curve (ROC)Identical to GGT for all subjects (combined genders)Suggests similar overall diagnostic accuracy
    AUC for MalesSlightly better for CDT than GGT
    AUC for FemalesSlightly better for GGT than CDT
    PrecisionWithin-run CV (low control)4.8%NCCLS document EP5-T2 followed
    Within-run CV (high control)2.7%
    Total Precision (low control)6.4%
    Total Precision (high control)3.4%
    Analytical SensitivityLimit of Detection (LOD)1.0 mg/L transferrin
    Limit of Quantification (LOQ)1.0 mg/L CDT and 1.5 mg/L total transferrin
    LinearityMeasuring Range0 - 50 mg/mL
    AccuracyCorrelation with HPLC %CDT (r)0.96Comparison to a recognized gold standard
    Slope vs. HPLC %CDT1.12
    Y-Intercept vs. HPLC %CDT-0.15
    InterferenceHemoglobin interference< 10%
    Bilirubin interference< 10%
    Lipid interference> 10% for concentrations above 10g/LNote: Interference observed at high lipid levels
    Plasma vs. SerumSignificantly higher %CDT in plasmaSample type impact
    Drug interferenceNo effect on CDT level
    Hook effect (Total Transferrin)Critical at > 100mg/L; no clinical consequenceHuman TT values are far below this level
    Temperature SensitivityLow control %CDT (18°C, 25°C, 30.5°C)2.5%, 4.9%, 8.3% respectivelyIndicates temperature dependence
    High control %CDT (18°C, 25°C, 30.5°C)8.3%, 9.5%, 11.2% respectivelyIndicates temperature dependence

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not explicitly state the exact sample size for the test set used in study CT-C8001. It describes the study group as "social drinkers and in chronic heavy drinkers," differentiating between males and females in the performance analysis.

    The data provenance is not explicitly stated in terms of country of origin. The study appears to be retrospective as it analyzes associations between CDT, GGT, and drinking habits.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    The document does not specify the number or qualifications of experts used to establish the ground truth for the test set.

    4. Adjudication Method for the Test Set:

    The document does not describe an adjudication method for the test set. The ground truth appears to be based on "social drinkers" vs. "chronic heavy drinkers" which would typically be established through self-reporting or clinical assessment, but the method for confirming these classifications is not detailed.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance:

    This device is an in vitro diagnostic (IVD) immunoassay, not an AI-powered diagnostic imaging or interpretation tool designed to assist human readers. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is not applicable and was not performed.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    Yes, the study describes the standalone performance of the Axis® %CDT TIA. The reported sensitivities, specificities, and AUC values are for the device itself, not in conjunction with human interpretation or intervention for the actual measurement. The calculation of %CDT from measured concentrations is an intrinsic function of the assay.

    7. The Type of Ground Truth Used:

    The ground truth appears to be a clinical classification based on behavioral and lifestyle factors: "social drinkers" and "chronic heavy drinkers." While not specified, these classifications are typically established through patient history, questionnaires, or other clinical assessments rather than pathology or direct outcomes data provided within this summary.

    8. The Sample Size for the Training Set:

    The document does not explicitly mention a separate "training set" or its sample size. The description pertains to the overall clinical study (CT-C8001) which evaluates the device's performance in different drinking groups. It's possible the entire dataset was used for evaluation, or aspects of the assay development were done with internal data not detailed here.

    9. How the Ground Truth for the Training Set Was Established:

    As no explicit training set is mentioned, the method for establishing ground truth for a training set is not provided. If the clinical study data was also used for "training" or refinement, the ground truth would have been established by classifying individuals as "social drinkers" or "chronic heavy drinkers" as described above (likely based on self-reported alcohol intake or clinical assessment for drinking patterns).

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1