Found 155 results

510(k) Data Aggregation

    K Number
    K252497

    Validate with FDA (Live)

    Device Name
    SANSA HSAT
    Manufacturer
    Date Cleared
    2025-12-10

    (124 days)

    Product Code
    Regulation Number
    868.2375
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K250882

    Validate with FDA (Live)

    Device Name
    SANSA HSAT
    Manufacturer
    Date Cleared
    2025-10-29

    (219 days)

    Product Code
    Regulation Number
    868.2375
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Huxley Home Sleep Apnea Test (SANSA) is a wearable device intended for use in the recording, analysis, and storage of biophysical parameters to aid in the evaluation of sleep-related breathing and cardiac disorders of adults suspected of sleep apnea. The device is intended for the clinical and home use setting under the direction of a Healthcare Professional (HCP). The system is prescription use only.

    The SANSA device records and stores ECG recording for up to 10 hours of wear time which can be displayed in the software portal for manual annotation and analysis. The SANSA does not provide automated analysis of the ECG and is not intended to be used with a 3rd party automated algorithm and is not intended for pacemaker analysis.

    Device Description

    The Huxley Home Sleep Apnea Test (SANSA™) is a wearable device intended for use in the recording, analysis, and storage of biophysical parameters to aid in the evaluation of sleep-related breathing disorders of adults suspected of sleep apnea. The device is intended for clinical and home use setting under the direction of a Healthcare Professional (HCP). The system is prescription use only. The SANSA™ device records and stores ECG recording for up to 10 hours of wear time which can be displayed in the software portal for manual annotation and analysis. The SANSA™ does not provide automated analysis of the ECG and is not intended to be used with a 3rd party automated algorithm and is not intended for pacemaker analysis.

    The SANSA HSAT collects multiple physiological signals using a single wearable patch worn on the chest. The SANSA device contains a reflective PPG sensor, a single-lead ECG sensor, and a 3-axis accelerometer. The signals from these sensors are passed into a cloud-based algorithm which utilizes a combination of signal processing and AI/ML components to compute time-series data for clinician review and summary metrics for report output. The device outputs the following time-series channels: Oximetry, Heart Rate, Chest Movement, Snoring, Body Position, Respiratory Effort, Actigraphy, Sleep staging (Sleep/Wake), and ECG. The following summary metrics are calculated: sansa-Apnea Hypopnea Index (sAHI) and Total Sleep Time (TST).

    Recorded data are uploaded to a software portal where physiological tracings are made available for review and event editing by a qualified healthcare professional. The device is intended to be worn for 10 hours per study.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study information for the SANSA HSAT device, based on the provided FDA 510(k) clearance letter:

    1. Acceptance Criteria and Reported Device Performance

    The acceptance criteria and performance data are primarily found on Page 9 of the document, under the "Performance" section within "Table 1: Device Comparison."

    Performance MetricAcceptance Criteria / Predicate Performance (K202359)Reported Device Performance (SANSA HSAT)Comparison
    Heart Rate (Arms)No Heart Rate Analysis (Predicate)≤ 3 bpm (range 30-250 bpm)N/A
    SpO2 (Arms)Does not collect SpO2 (Predicate)≤ 3% (range 70-100%)N/A
    Aid to Diagnosis of Moderate to Severe OSA (AHI≥15) - SensitivityNot applicable (Predicate does not diagnose OSA)88.2%N/A
    Aid to Diagnosis of Moderate to Severe OSA (AHI≥15) - SpecificityNot applicable (Predicate does not diagnose OSA)87.3%N/A
    ECG Recording Accuracy: Frequency Response0.67 Hz to 40 Hz0.67 Hz to 40 HzIdentical
    ECG Recording Accuracy: Input Impedance>10 MΩ>10 MΩIdentical
    ECG Recording Accuracy: Gain AccuracyMaximum amplitude error ±10%Maximum amplitude error ±10%Identical
    ECG Recording Accuracy: Gain Stability<3% over a 24-hour period<3% over a 24-hour periodIdentical
    ECG Recording Accuracy: Timing Accuracy<30 seconds<30 secondsIdentical

    Note: For Heart Rate, SpO2, and OSA diagnosis metrics, the predicate device (iRhythm Zio Monitor) does not provide these analyses or collect SpO2. Therefore, the "acceptance criteria" for these aspects are implicitly met by the absence of safety/effectiveness concerns with the SANSA HSAT's reported performance, which aligns with its intended use in sleep apnea evaluation. The ECG recording accuracy metrics are directly comparable and are identical between the subject device and the predicate.

    2. Sample Size and Data Provenance for the Test Set

    The document does not explicitly state the sample size used for the clinical performance validation of the SANSA HSAT's ECG, nor does it specify the country of origin or whether the data was retrospective or prospective. It only mentions:

    • Sample Size: Not specified.
    • Data Provenance: Not specified (country/retrospective/prospective).

    For the Aid to Diagnosis of OSA, a sensitivity of 88.2% and specificity of 87.3% are reported. The document does not provide the sample size or provenance for this particular study, but it was leveraged from a previous clearance (K244027).

    3. Number of Experts and Qualifications for Ground Truth - Test Set

    The document does not provide details on the number or qualifications of experts used to establish ground truth for the test set specifically for the ECG clinical performance validation.

    For the "Aid to Diagnosis of Moderate to Severe OSA", the percentages suggest a comparison against a diagnostic standard, which would typically involve expert interpretation, but the details are not provided in this document.

    4. Adjudication Method for the Test Set

    The document does not specify any adjudication method (e.g., 2+1, 3+1) used for establishing the ground truth for the test set.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No information is provided about a multi-reader multi-case (MRMC) comparative effectiveness study, or any effect size related to human reader improvement with or without AI assistance. The SANSA device explicitly states it "does not provide automated analysis of the ECG and is not intended to be used with a 3rd party automated algorithm." (Pages 4 and 5).

    6. Standalone Performance Study

    Yes, a standalone performance study was done for the SANSA HSAT's ECG:

    • Study Type: "Sansa ECG clinical performance was validated through comparison to a simultaneously collected reference standard Holter monitor." (Page 11)
    • Performance: "Clinically acceptable performance was demonstrated through qualitative and quantitative analysis of the ECG signal." (Page 11)

    The reported sensitivity and specificity for OSA diagnosis (Page 9) also represent standalone algorithm performance.

    7. Type of Ground Truth Used

    • For ECG Clinical Performance: "simultaneously collected reference standard Holter monitor." (Page 11)
    • For Aid to Diagnosis of OSA (reported sensitivity/specificity): While not explicitly stated, the context of "Aid to Diagnosis" for Sleep Apnea typically implies comparison to a polysomnography (PSG) study, which is the gold standard, interpreted by sleep specialists. This data was "leveraged from previous clearance."

    8. Sample Size for the Training Set

    The document does not provide any information about the sample size used for a training set. This is consistent with the statement that the device does not provide automated analysis of the ECG and explicitly states that it is not intended for use with a 3rd party automated algorithm. While the device uses "AI/ML components" for other signals (Snoring, Body Position, Respiratory Effort, Actigraphy, Sleep staging), the training set size for these components is not detailed in this document.

    9. How Ground Truth for the Training Set Was Established

    The document does not provide information on how the ground truth for any potential training set was established. Given the focus on manual annotation and analysis for ECG, and the lack of detail on the AI/ML components for other signals, this information is not present in the clearance letter.

    Ask a Question

    Ask a specific question about this device

    K Number
    K250460

    Validate with FDA (Live)

    Date Cleared
    2025-09-05

    (199 days)

    Product Code
    Regulation Number
    868.2375
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The WatchPAT400 (WP400) is a non-invasive home care device for use with patients suspected to have sleep related breathing disorders. The device is a diagnostic aid for the detection of sleep related breathing disorders, sleep staging (REM Sleep, Light Sleep, Deep Sleep and Wake), snoring level and body position. The WP400 generates a peripheral arterial tonometry ("PAT"), Respiratory Disturbance Index ("pRDI"), Apnea-Hypopnea index ("pAHI"), Central Apnea-Hypopnea index ("pAHIc"), sleep staging identification (pSTAGES) and optional snoring level and body position discrete states from an external integrated snoring and body position sensor. The device's pSTAGES and snoring level and body position provide supplemental information to its pRDI/pAHI/pAHIc. The device's pSTAGES and snoring level and body position are not intended to be used as the sole or primary basis for diagnosing any sleep related breathing disorder, prescribing treatment, or determining whether additional diagnostic assessment is warranted.

    pAHIc is indicated for use in patients 17 years and older. All other parameters are indicated for 12 years and older.

    Device Description

    The WP400 is a ventilatory effort recorder that utilizes PAT technology. The controller part of the device is worn on the wrist and records the PAT signal and arterial blood oxygen saturation levels by a finger-mounted probe based on an optical plethysmographic method. An actigraph, embedded in the wrist worn controller unit, records wrist motion that is used to determine periods of sleep vs wake. A chest sensor is attached to the patient's chest right under the sternal notch for measuring snoring level, body position states and chest movements. The device is battery powered and connects via Bluetooth to a mobile application.

    The WP400 is a re-usable device, with sensors common with the WP300 and wireless connectivity identical to the WP1. With the same core technology, signal acquisition, and signal processing identical across all three WatchPAT devices, the WP400 is a combination of its predecessors: device housing leveraged from the WP300 for its re-usable nature, combined with the ability for wireless communication leveraged from the WP1. Signals recording, processing, and analysis has not changed from the predicate devices.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K252383

    Validate with FDA (Live)

    Device Name
    Somfit D
    Date Cleared
    2025-08-28

    (28 days)

    Product Code
    Regulation Number
    868.2375
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Somfit D is a single-use, non-invasive prescription device for home use with patients suspected to have sleep-related breathing disorders. The Somfit D is a diagnostic aid for the detection of sleep-related breathing disorders, sleep staging (REM, N1, N2, N3, Wake), and snoring level. The Somfit D system acquires electrical data from three frontal electrodes, tri-axial accelerometer data, acoustical and plethysmographic data. The Somfit D calculates and reports to clinicians EEG/EOG channels, Sleep Stages, SpO2, Peripheral Arterial Tonometry signal, pulse rate, and snoring level. The Somfit D calculates and reports to clinicians derived parameters such as Peripheral Arterial Tonometry-derived Apnea Hypopnea Index, Obstructive Desaturation Index; and hypnogram-derived indices such as time in each sleep stage. Somfit D data is not intended to be used as the sole or primary basis for diagnosing any sleep-related breathing disorder, prescribing treatment, or determining whether additional diagnostic assessment is warranted. The Somfit D is not intended for use as life support equipment, for example vital signs monitoring in intensive care unit. The Somfit D is a prescription device indicated for adult patients aged 21 years and over.

    Device Description

    The Somfit D is a home-based sleep monitoring device which records signals from the patient's forehead. Somfit D is a wearable, low voltage, battery operated device which is attached to subject forehead via a self-adhesive and disposable skin electrode patch. The electrodes are placed on the anterior Prefrontal cortex (PFC) at the Fp1 and Fp2 positions according to the 10/20 EEG system. The device allows for recording of two frontal EEG signals, pulse rate, SPO2, PAT, PPG, motion, and snore. Somfit D uses a mobile phone application to acquire data wirelessly via Bluetooth BLE technology, then transfer into a secure cloud, for management, storage and post-processing. The software reports measured parameters in a format compatible with the American Academy of Sleep Medicine guidelines, including sleep time, ODI, pAHI and conventional graphical displays such as a hypnogram.

    AI/ML Overview

    The provided FDA 510(k) clearance letter and Somfit D 510(k) Summary describe the acceptance criteria and the study that proves the device meets those criteria. However, it explicitly states that the Somfit D is substantially equivalent to the predicate device, Somfit (K231546), and therefore, the performance data for the Somfit D is derived from the studies conducted on the Somfit. The summary then refers to already submitted and approved studies for the Somfit device.

    Here's a breakdown of the requested information based on the provided document:

    Acceptance Criteria and Device Performance (Derived from Predicate Device, Somfit)

    Table 1: Acceptance Criteria and Reported Device Performance

    ParameterAcceptance Criteria (Implicit from Predicate Studies)Reported Device Performance (Somfit D, by Equivalence to Somfit)
    Oximeter PerformanceISO 80601-2-61 complianceAchieved (Formal controlled desaturation study conducted as per standard, already submitted and approved for Somfit)
    PAT-derived Apnea-Hypopnea Index (pAHI)Meaningful validation as an HSAT device (implied by previous clearance of Predicate)Meaningful validation as an HSAT device (study conducted on Somfit, already submitted and approved)
    Oxygen Desaturation Index (ODI)Meaningful validation as an HSAT device (implied by previous clearance of Predicate)Meaningful validation as an HSAT device (study conducted on Somfit, already submitted and approved)
    Sleep Staging Concordance (REM, N1, N2, N3, Wake)Meaningful validation as an HSAT device (implied by previous clearance of Predicate)Meaningful validation as an HSAT device (study conducted on Somfit, already submitted and approved)
    Electrical SafetyIEC 60601-1:2005 (Third Edition) + COR1:2006 + COR2:2007 + A1:2012 complianceAchieved (Testing activities on Somfit, identical to Somfit D in this aspect)
    Electromagnetic Compatibility (EMC)IEC 60601-1-2:2014, EN 60601-1-2:2015 complianceAchieved (Testing activities on Somfit, identical to Somfit D in this aspect)
    Home Healthcare EnvironmentalIEC 60601-1-11:2015 complianceAchieved (Testing activities on Somfit)
    Electroencephalograph safety and performanceIEC 60601-2-26:2012 complianceAchieved (Testing activities on Somfit, identical to Somfit D in this aspect)
    Pulse oximeter safety and performanceISO 80601-2-61:2011 compliance (including functional simulator)Achieved (Testing activities on Somfit, identical to Somfit D in this aspect)
    Lithium Battery SafetyIEC 60086-4 (single use lithium batteries) complianceAchieved (Specific test report for Somfit D's CR2032 battery)
    Hardware Bench Testing / Electrical Parameters/Design SpecificationsVerification of electrical parameters and design specificationsAchieved (For Somfit)
    Software Functional Requirements / System IntegrationVerification of functional requirements and system integrationAchieved (For Somfit)
    BiocompatibilityOvernight use on intact skin (implied by materials and intended use)Met (For Somfit)

    Study Details (Pertaining to the Predicate Device, Somfit, as explicitly stated for Somfit D equivalence)

    1. Sample sizes used for the test set and the data provenance:

      • Oximeter Validation: Controlled desaturation study in accordance with ISO 80601-2-61. Specific sample size not specified in this document, but implied to be sufficient for standard compliance. Data provenance is implied to be from a "Hypoxia Lab." The document doesn't specify if it was retrospective or prospective, or country of origin, but clinical studies for FDA clearance are typically prospective to ensure controlled conditions.
      • Home Sleep Apnea Test Validation (pAHI, ODI, Sleep Staging): A "Multi-Site Clinical Study" was conducted. Specific sample sizes for each of these validations are not provided in this summary. The document states "clinical data for the purpose of the predicate device Somfit (K231546), already submitted and approved," indicating these details would be found in the original Somfit 510(k) submission. The provenance is from "Multi-Site Clinical Study," implying multiple locations, likely within the regulatory jurisdiction where the clearance was sought (e.g., US, Australia). It's implied to be prospective for validation purposes.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This information is not explicitly stated in the provided 510(k) summary for any of the clinical validations (Oximeter, pAHI, ODI, Sleep Staging). It only mentions that the studies were "meaningful validations" and "concordance," implying comparison to a gold standard. For sleep staging, ground truth is typically established by certified polysomnography technologists and/or sleep physicians.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • This information is not explicitly stated in the provided 510(k) summary for any of the clinical validations.
    4. If a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No MRMC study is mentioned. The Somfit D (and its predicate Somfit) is described as a diagnostic aid that calculates and reports parameters to clinicians. It's not presented as an AI-assissted reading tool for human interpretation, but rather a device that quantifies specific physiological signals and derives standard indices. The human role is in interpretation of the reported data, not in an AI-assisted reading workflow.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Yes, the device (Somfit D, through its predicate Somfit) performs standalone algorithmic analysis to calculate:
        • Sleep Stages (REM, N1, N2, N3, Wake)
        • SpO2
        • Peripheral Arterial Tonometry (PAT) signal
        • Pulse rate
        • Snoring level
        • PAT-derived Apnea Hypopnea Index (pAHI)
        • Obstructive Desaturation Index (ODI)
        • Hypnogram-derived indices (e.g., time in each sleep stage)
      • The "meaningful validation" studies for these parameters (pAHI, ODI, Sleep Staging) suggest a comparison of the device's algorithmic outputs against established ground truth.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • Oximeter Validation: Performed in a "Hypoxia Lab," implying comparison to a highly accurate laboratory reference oximeter or blood gas analysis (gold standard for SpO2 measurements).
      • Home Sleep Apnea Test Validation (pAHI, ODI, Sleep Staging): While not explicitly stated, for HSAT devices, "ground truth" for pAHI and ODI is typically derived from comparison to full in-lab Polysomnography (PSG) data scored by a certified sleep technologist and/or interpreted by a board-certified sleep physician, which is considered the clinical gold standard. For sleep staging, the ground truth would be expert-scored PSG recordings.
    7. The sample size for the training set:

      • This information is not provided in the summary. The summary focuses on the validation studies, which imply the device's algorithms were already developed and trained.
    8. How the ground truth for the training set was established:

      • This information is not provided in the summary, as it pertains to the development phase rather than the validation phase described.
    Ask a Question

    Ask a specific question about this device

    K Number
    K243220

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-07-03

    (269 days)

    Product Code
    Regulation Number
    868.2375
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Onera STS 2 measures and records multiple physiological parameters from a patient during a sleep study, which are used by clinicians to decide on the diagnosis of sleep disorders.

    Onera STS 2 is intended to be used on a patient who has been prescribed a polysomnography study by a healthcare professional. The device is designed to be used under the direction of a physician or trained technician but applied by a layperson.

    The recorded data will be made available to the healthcare professional to assist in the diagnosis of sleep disorders.

    The device is intended to be used for adults.

    The device is not a life supporting physiological monitor.

    Device Description

    Onera Sleep Test System 2 (Onera STS 2) is a hardware, wearable system for measuring physiological signals during a sleep study. The device can be used in the home (Home Healthcare Environment) as well as in Professional Healthcare Facilities.

    The device measures EEG, EOG, EMG, ECG, respiratory signals, cannula based respiratory flow, oxygen saturation, activity, position, ambient light level and sound pressure level.

    The ECG signal is not intended for diagnosis of cardiac disorders, except for the manual determination of arrhythmia during polysomnography studies.

    The Onera STS 2 does not provide any automated output like heart rate, assessments of arrhythmia, heart rate variability, or other related heart rate measurement functions.

    Onera STS 2 consists of five Sensors applied on the forehead, upper chest area, abdomen and lower leg area (both left and right leg).

    The Sensors should be placed on intact skin. During the night measurement it is not needed to inspect the application sites. The Sensors encrypt the recorded signals and upload the measurement data during the night to the Patient App installed on the patient's phone via Bluetooth. The Patient App securely transfers the data to the cloud interface for further processing once all data has been uploaded. The output of the system is represented by a file in EDF format which contains the recorded (physiological) signals. EDF files can be read by any application software that accepts such files as input.

    It is not possible to relocate the Sensors during a sleep study since the Sensor is single-use and for one sleep study only.

    AI/ML Overview

    Here's a summary of the acceptance criteria and study information for the Onera STS 2, based on the provided FDA 510(k) clearance letter:


    Onera STS 2: Acceptance Criteria and Performance Study Summary

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document primarily focuses on the SpO2 measurement accuracy as a key performance metric with specific acceptance criteria.

    ParameterAcceptance CriteriaReported Device Performance
    SpO2 Accuracy (70-100% SpO2 range)$\pm$3% (ISO 80601-2-61:2019 Clause 201.12.1.101.1)$\pm$2.5%

    Note: While other parameters are listed as "Identical" to the predicate, specific numerical acceptance criteria for those parameters are not explicitly stated in this document beyond their qualitative equivalence.

    2. Sample Size and Data Provenance for the Test Set

    • Sample Size: 12 healthy subjects
      • 9 male, 3 female
      • Aged between 23 and 46 years old
    • Data Provenance: The study was conducted in an "independent research laboratory." The country of origin is not explicitly stated in the provided text. The study involved "induced hypoxia," indicating a prospective, controlled experimental design.

    3. Number and Qualifications of Experts for Ground Truth

    • The document does not mention the use of experts to establish ground truth for the SpO2 accuracy test.
    • The ground truth for SpO2 was established by "laboratory co-oximeter" measurements of arterial hemoglobin oxygen (SaO2) values from blood samples.

    4. Adjudication Method for the Test Set

    • The document does not describe any adjudication method. The SpO2 accuracy was determined by direct comparison of the device's SpO2 measurements to SaO2 values from a laboratory co-oximeter.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not conducted or described in this document. The device primarily measures physiological parameters, and the study focused on the accuracy of these measurements rather than human reader interpretation with or without AI assistance.

    6. Standalone (Algorithm Only) Performance Study

    • Yes, a standalone performance study was conducted for SpO2 measurement accuracy. The device's SpO2 readings were directly compared to reference SaO2 values without human intervention in the SpO2 measurement process itself. The Onera STS 2 is described as measuring and recording parameters, with the output as an EDF file to be read by other software. The SpO2 accuracy assessment is specifically for the device's measurement capability.

    7. Type of Ground Truth Used

    • Objective Measurement (Laboratory Co-oximeter): For SpO2 accuracy, the ground truth was established by arterial hemoglobin oxygen (SaO2) values determined from blood samples using a laboratory co-oximeter, which is considered a gold standard for blood oxygen saturation.

    8. Sample Size for the Training Set

    • The document does not provide information regarding the sample size for a training set. This is likely because the performance study described (SpO2 accuracy) is a validation of the device's sensor capabilities, not an evaluation of a machine learning algorithm that would typically require a training set. The device outputs raw physiological signals in EDF format for clinicians to interpret, rather than providing automated diagnoses based on an internal algorithm.

    9. How the Ground Truth for the Training Set Was Established

    • As no training set is mentioned or implied for the device's core functionality (measuring and recording parameters for clinician interpretation), this information is not applicable and not provided in the document.
    Ask a Question

    Ask a specific question about this device

    K Number
    K242224

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-06-18

    (324 days)

    Product Code
    Regulation Number
    868.2375
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Happy Health Home Sleep Test is a Software as a Medical Device that uses data from wearable devices to record, analyze, display, export, and store biophysical parameters to aid in evaluating sleep‐related breathing disorders of adult patients suspected of sleep apnea. The device is intended for use on individuals who are 22 years of age or older in clinical and home settings under the direction of a trained healthcare provider.

    Device Description

    The Happy Health Home Sleep Test is a Software as a Medical Device that uses data from wearable devices to record, analyze, display, export, and store biophysical parameters to aid in evaluating sleep-related breathing disorders of adult patients suspected of sleep apnea.

    The device is intended for use on individuals who are 22 years of age or older in clinical and home settings under the direction of a trained healthcare provider. The device is intended to process input data streams received from an external hardware device (i.e., a smart ring, K240236) and uses these signals to determine various sleep parameters that may be used and interpreted by a clinician in diagnosing sleep disorders such as sleep apnea.

    The input physiologic signals from the external device are:

    • Acceleration / Movement
    • Photoplethysmography (PPG)

    The external hardware device (K240236) includes a PPG sensor and accelerometer embedded within a housing to capture the above physiological signals. The K240236 device is worn on the finger and is indicated for continuous data collection of the above signals. Data from the external hardware device is transmitted over a secure API to the subject device for analysis.

    The device then uses a set of algorithms to compute the following outputs:

    • Happy Health Apnea Hypopnea Index (hAHI)
    • Total Sleep Time

    The outputs are available for a clinician to review as a report, accessible through a web-based viewer application.

    AI/ML Overview

    The provided FDA 510(k) clearance letter and summary for the Happy Health Home Sleep Test give a good overview of the device's performance testing. Here's a structured breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't explicitly list "acceptance criteria" as a separate section with specific thresholds that were agreed upon before the study. Instead, it presents performance metrics of the Happy Health Home Sleep Test and compares them to the predicate and reference devices, implying these metrics are the basis for demonstrating substantial equivalence. For clarity, I'm inferring the acceptance criteria from the "Equivalent" column in the comparison tables and the detailed performance results.

    Metric (Inferred Acceptance Criteria)Happy Health Home Sleep Test Reported PerformanceJustification for Acceptance (from document)
    hAHI Regression Slope (Regression: PSG_AHI = Slope * hAHI + Intercept, Slope between 0.9 and 1.1)0.98 [0.91, 1.06]"Equivalent - both subject and predicate devices demonstrate strong correlation with manually scored AHI, each with a regression slope between 0.9 and 1.1 and intercept between -5 and 5."
    hAHI Regression Intercept (Intercept between -5 and 5)0.81 [-0.35, 1.91]"Equivalent - both subject and predicate devices demonstrate strong correlation with manually scored AHI, each with a regression slope between 0.9 and 1.1 and intercept between -5 and 5."
    hAHI Bland-Altman Mean Bias (Not explicitly quantified as a criterion, but a low bias is desired)0.5 [-0.1, 1.1] events/hrDemonstrates low systematic difference from PSG AHI.
    hAHI Bland-Altman Limits of Agreement (LOA) (Comparable to predicate/reference, generally aiming for tighter LOA)Lower LOA: -9.8 [-10.6, -9] events/hrUpper LOA: 10.7 [-9.9, 11.5] events/hr"Equivalent - both subject and predicate devices demonstrate strong correlation with manually scored AHI..." (Implied that these LOA are acceptable/comparable to predicate when predicate's full data is considered).
    Total Sleep Time (TST) Mean Absolute Difference (Comparable to predicate/reference, around 30 minutes or less)24.9 minutes (SD 32.6 minutes)"Equivalent - both subject and reference devices demonstrate strong correlation with manually scored AHI, each with a mean absolute difference of around 30 minutes or less."

    2. Sample Size and Data Provenance

    • Test Set Sample Size: 90 subjects.
    • Data Provenance:
      • Country of Origin: Not explicitly stated, but the study was conducted at "two sleep labs", implying a clinical setting within the country of submission (likely USA, given FDA submission).
      • Retrospective or Prospective: The wording "Data from a total of 90 subjects referred to the sleep clinic by a physician was manually scored" suggests the data was collected prospectively for the purpose of the study. The phrasing "A clinical study was performed to evaluate the performance..." also indicates a planned, prospective study.

    3. Number of Experts and Qualifications for Ground Truth

    • Number of Experts: Not explicitly stated how many experts were involved in the manual scoring. The text only mentions "manually scored in accordance with the American Academy of Sleep Medicine (AASM) guidelines."
    • Qualifications of Experts: Not explicitly stated, but implied to be qualified sleep technicians/physicians capable of AASM-compliant scoring.

    4. Adjudication Method for the Test Set

    The adjudication method for reconciling discrepant manual scores (if multiple scorers were used) is not specified in the provided text. It simply states "manually scored." If only one scorer per patient, no adjudication would be needed.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    There is no indication that a multi-reader multi-case (MRMC) comparative effectiveness study was done to evaluate how human readers improve with AI vs. without AI assistance. The study focuses solely on the device's standalone performance compared to manual PSG scoring.

    6. Standalone (Algorithm Only) Performance

    Yes, a standalone performance study was done. The entire clinical testing section details the performance of the "Happy Health Home Sleep Test" algorithm (hAHI and TST) compared to manually scored Polysomnography (PSG) data, without human-in-the-loop assistance.

    7. Type of Ground Truth Used

    The primary ground truth used was expert consensus / manual scoring of Polysomnography (PSG) data in accordance with American Academy of Sleep Medicine (AASM) guidelines. This is the gold standard for sleep studies.

    8. Sample Size for the Training Set

    The sample size for the training set is not provided in this document. The clinical study details describe the test set used for validation.

    9. How the Ground Truth for the Training Set Was Established

    The document does not provide information on how the ground truth for the training set was established. It only discusses the ground truth for the clinical validation (test) set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K243092

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-03-19

    (170 days)

    Product Code
    Regulation Number
    868.2375
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    AcuPebble Ox200 is a wearable device intended for use in the recording, analysis, displaying, exporting, and storage of biophysical parameters to aid in the evaluation of adult patients with, or with suspected, obstructive sleep apnea (OSA). The device is primarily intended for home setting use (although can also be used in healthcare settings) under the direction of a Healthcare Professional (HCP).

    Device Description

    AcuPebble Ox200 is prescribed by a Health Care Professional for the patient to use in the home as a 'home sleep apnea test' (HSAT). AcuPebble Ox200 comprises: 1- A small microphones/accelerometer sensor, identical to the one in the previously cleared devices AcuPebble SA100 (K210480) and AcuPebble Ox100 (K222950) that is worn on the front of the neck around the suprasternal notch area (the "AcuPebble SA100 sensor"), and attaches with a single use biocompatible adhesive tape (same one as for "AcuPebble SA100" and "AcuPebble Ox100)); 2- Another PPG/accelerometer sensor ("AcuPebble_oximetry sensor") worn either on the finger or on the forehead; 2- A mobile device app (the "AcuPebble Ox200 app") that guides the patient through the steps of the test, collects the data from the sensors, and uploads them in the cloud ; 3- A cloud-based software that analyses the collected signals as in the previously cleared devices; 4- A web app user interface for healthcare professionals where they can set up a study and see the results, including different physiological channel traces and OSA diagnostic parameters, as in the previously cleared devices.

    The differences between AcuPebble Ox200 and the previously cleared device (K222950) are:

    • A new sensor for sensing the PPG signal and extraction of the SpO2 and pulse rate channel from. This new sensor can be placed either on the finger or the forehead.
    • The addition of new parameters extracted from the acoustic neck sensor including position, respiratory efforts and respiratory rate channels.

    The OSA diagnostic indexes are still extracted from the signals measured with the identical sensor as in the predicate (neck sensor only), and with the same algorithms, remain unchanged with respect to the predicate device (K222950) and hence have not been separately tested in this device.

    AI/ML Overview

    Here's a summary of the acceptance criteria and the studies that prove the device meets those criteria, based on the provided text:

    AcuPebble Ox200 Acceptance Criteria and Performance Study Summary

    1. Table of Acceptance Criteria and Reported Device Performance

    ParameterAcceptance Criteria (Target or Reference Performance)Reported Device Performance (AcuPebble Ox200)
    OSA Diagnostic IndexesAHI (3% desaturation criteria): Diagnostic Sensitivity > 81.3% (from reference device)92.73%
    AHI (4% desaturation criteria): Diagnostic Sensitivity > 81.3% (from reference device)95.92%
    AHI (3% desaturation criteria): Diagnostic Specificity > 82.1% (from reference device)96.84%
    AHI (4% desaturation criteria): Diagnostic Specificity > 82.1% (from reference device)97.03%
    Pulse Rate (Acoustics)50-120bpm, RMS error < 3.62bpm (from predicate device)50-120bpm, 3.62bpm validation accuracy (from acoustics)
    Pulse Rate (PPG)50-120bpm, RMS error < 2 bpm (from predicate device)Arms (validation accuracy) < 1.23bpm (for both forehead and finger, average RMSE of 1.23 bpm across varying skin tones), worst case RMSE 1.66 bpm (using functional simulator)
    Respiratory Rate4-35 bpm, accuracy ±2.7 bpm (from reference device Airmod K2427984)Overall RMSD 2.52 bpm (4-30bpm, lab study with capnography); RMSD 2.46 bpm (during natural sleep, compared to Embletta MPR chest/abdomen bands). Claimed range: 4-30 bpm, resolution 1 bpm, accuracy ±3 bpm.
    Respiratory EffortNot inferior to reference device (Embletta MPR RIP bands) for differentiating no effort vs. effortAgreement of 96.8% with manual scoring by two clinicians when distinguishing apnoea with no effort vs. with effort. Outperformed reference RIP bands of Embletta MPR (K122516) with 96.3% accuracy vs. 75.9% accuracy when using Embla RemLogic (K162140) automatic scoring as reference.
    Body Position DetectionNot inferior to reference device (implied standard from reference device for accelerometer-based position sensing)Lab study: 93.93% overall accuracy (supine, prone, left, right, upright). Natural sleep study: 79.00% accuracy overall; 83.59% accuracy in detecting supine vs. non-supine.
    SpO2 Accuracy (Arms)Comparable to predicate (implied standard, likely ISO 80601-2-61:2017 standards for accuracy with varying skin tones for pulse oximeters)1.86 (forehead), 2.25 (finger) in the range of 70% to 100%. Lightest skin tone (Fitzpatrick 1, Monk 1): 1.63 (forehead), 1.44 (finger). Darkest skin tone (Fitzpatrick 6, Monk 10): 0.66 (forehead), 1.36 (finger). (All values for SpO2 accuracy (Arms) are within typical acceptance for pulse oximetry).

    2. Sample Sizes and Data Provenance

    • SpO2 and Pulse Rate: 12 subjects with varying skin tone. The study was performed by an independent testing lab. (Country of origin not specified, but the device manufacturer is UK-based.) Data appears to be prospective clinical validation.
    • Respiratory Rate:
      • Study 1: 31 healthy subjects. Conducted under laboratory conditions. (Country of origin not specified, but the device manufacturer is UK-based.) Data appears to be prospective clinical validation.
      • Study 2: 150 patients referred for OSA diagnosis. Conducted during natural sleep. Royal Free London Hospital NHS Foundation Trust (UK). Data appears to be prospective clinical validation, with a trial registration number (NCT03544086).
    • Respiratory Effort: 150 patients referred for OSA diagnosis. Conducted during natural sleep. Royal Free London Hospital NHS Foundation Trust (UK). Data appears to be prospective clinical validation, with a trial registration number (NCT03544086).
    • Body Position Detection:
      • Study 1: 17 volunteers. Conducted under laboratory conditions. (Country of origin not specified, but the device manufacturer is UK-based.) Data appears to be prospective clinical validation.
      • Study 2: 60 patients referred for OSA diagnosis. Conducted during natural sleep. Virgen Macarena Hospital, Spain. Data appears to be prospective clinical validation, with a trial registration number (NCT04028011).

    3. Number of Experts and Qualifications for Ground Truth (Test Set)

    • Respiratory Effort: "manual scoring by two clinicians". Qualifications of the clinicians are not explicitly stated (e.g., specific specialties or years of experience), but they are described as "clinicians." For the automatic scoring comparison, an FDA-cleared device (Embla RemLogic, K162140) served as a reference.
    • Other parameters (SpO2, Pulse Rate, Respiratory Rate, Body Position): The ground truth relied on established reference devices and standards (e.g., functional simulator, capnography, RIP bands, ISO 80601-2-61:2017), rather than expert consensus on individual cases for the test set. For OSA diagnostic indexes, the performance is inherited from the predicate device, so the ground truth setup for that specific parameter is not detailed for AcuPebble Ox200 in this document.

    4. Adjudication Method (Test Set)

    • Respiratory Effort: Manual scoring was done by "two clinicians," and the agreement between them or a further adjudication process (e.g., 2+1) is not specified beyond "the agreement in distinguishing." However, a primary comparison for superiority was made against an FDA-cleared device's automatic scoring (Embla RemLogic).
    • For other parameters, adjudication is not applicable as ground truth was established by reference devices/standards.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study is explicitly mentioned as having been performed to assess how much human readers improve with or without AI assistance for this device. The evaluations focus on the standalone performance of the device's outputs.

    6. Standalone Performance (Algorithm Only)

    Yes, standalone performance was done for several new parameters:

    • SpO2 and Pulse Rate Output: Validated as per ISO 80601-2-61:2017 using an independent testing lab.
    • Respiratory Rate: Validated in two studies against capnography and respiratory inductance plethysmography (RIP) bands.
    • Respiratory Effort Detection: Validated against RIP bands and compared to automatic scoring of an FDA-cleared device.
    • Body Position Detection: Validated in laboratory conditions and during natural sleep.
    • The OSA diagnostic indexes (AHI) are explicitly stated to be "extracted from the signals measured with the identical sensor as in the predicate (neck sensor only), and with the same algorithms, remain unchanged with respect to the predicate device (K222950) and hence have not been separately tested in this device." Their performance is inherited from the predicate.

    7. Type of Ground Truth Used (Test Set)

    The ground truth varied by parameter:

    • SpO2 and Pulse Rate: Reference pulse oximetry and functional simulator, following ISO 80601-2-61:2017 standards for accuracy assessment.
    • Respiratory Rate: Capnography (K150272) in a lab setting, and chest and abdomen respiratory effort bands of Embletta MPR Sleep Data Recording System (K122516) during natural sleep.
    • Respiratory Effort: Manual scoring by two clinicians based on RIP bands (Respiratory Inductance Plethysmography) and automatic scoring from the FDA-cleared device Embla RemLogic (K162140).
    • Body Position Detection: Presumed to be observational ground truth established by study personnel in lab, and potentially manually annotated based on video/observation in natural sleep studies, or potentially using other reference accelerometers. The document does not specify the exact method for ground truth for body posture.

    8. Sample Size for the Training Set

    The document does not explicitly state the sample size used for the training set for any of the algorithms. It focuses on the validation studies.

    9. How the Ground Truth for the Training Set Was Established

    As the training set information is not provided, the method for establishing its ground truth is also not detailed in this document. The document primarily describes the validation of the device's performance against established clinical standards and reference devices.

    Ask a Question

    Ask a specific question about this device

    K Number
    K243268

    Validate with FDA (Live)

    Device Name
    TipTraQ (TTQ001)
    Manufacturer
    Date Cleared
    2025-02-03

    (111 days)

    Product Code
    Regulation Number
    868.2375
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TipTraQ is a wearable device intended for aiding in sleep apnea evaluation/diagnosis for adult patients suspected of sleep apnea in both home-based and clinical-use environments.

    Device Description

    TipTraQ is a prescription-only medical device aiding in sleep apnea evaluation/diagnosis comprising a fingertip wearable device, a companion mobile app, a cloud-based AI analysis and an information display system. It collects essential physiological waveform information, including Photoplethysmogram (PPG) and accelerometer data. The TipTraQ Algorithm System determines Total Sleep Time (TST), Total REM Time (TREMT), Oxygen Desaturation Index (ODI), and Apnea-Hypopnea Index (AHI) from the recorded waveform signals.

    TipTraQ Sensor includes a PPG sensor, which uses light sources (green, red, and infrared) and two photodetectors to measure the blood volume changes in the microvascular bed of tissue. Additionally, the TipTraQ Sensor utilizes an Inertial Measurement Unit (IMU) with three-axis accelerometers (device-function) and three-axis gyroscopes (non-device function), monitoring movement and actigraphy to assess sleep stages. The TipTraQ Expert Panel can display recorded physiological waveforms, while the TipTraQ API System generates output for essential physiological parameters analyzed by the TipTraQ Algorithm System.

    The TipTraQ Sensor is rechargeable and designed to provide at least 12 hours of usage after fully charged. The TipTraQ Charging Case, which is also rechargeable, provides a battery capacity of eight full charges of the TipTraQ Sensor for user convenience. The TipTraQ Charging Case can be powered and charged using an external USB-C.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the TipTraQ (TTQ001) device, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    Note: The document presents the performance results directly without explicitly stating the "acceptance criteria" as a separate, quantified threshold. However, for the SpO2 validation, it states "met the predefined acceptance criteria <3.5". For the sleep validation study, it mentions "met all the predetermined acceptance criteria (set by prior studies and reference literatures)," implying these are the results shown in the table.

    MeasureMetricAcceptance Criteria (Implied / Stated)Reported Device Performance
    SpO2 AccuracyArms (Artificial Root Mean Square)< 3.5% (stated)1.41% (Overall)
    SpO2 Range 70-100%N/A (implied by meeting overall Arms)1.41
    SpO2 Range 60-70%N/A1.55
    SpO2 Range 70-80%N/A1.7
    SpO2 Range 80-90%N/A1.35
    SpO2 Range 90-100%N/A1.11
    Pulse Rate (PR) AccuracyArms (Artificial Root Mean Square)N/A1.64
    PR RangeN/A56-118 bpm
    Total Sleep TimePearson Correlation Coefficient (PCC)N/A0.786
    Epoch-wise Sleep Stage (Wake)SensitivityN/A0.655
    SpecificityN/A0.901
    AccuracyN/A0.837
    Epoch-wise Sleep Stage (REM)SensitivityN/A0.713
    SpecificityN/A0.930
    AccuracyN/A0.905
    Epoch-wise Sleep Stage (NREM)SensitivityN/A0.824
    SpecificityN/A0.738
    AHI based on 1a ruleOverall AccuracyN/A0.791
    AHI cutoff 5 sensitivityN/A0.868
    AHI cutoff 5 specificityN/A0.741
    AHI cutoff 15 sensitivityN/A0.876
    AHI cutoff 15 specificityN/A0.755
    AHI cutoff 30 sensitivityN/A0.806
    AHI cutoff 30 specificityN/A0.905
    AHI based on 1b ruleAHI cutoff 5 sensitivityN/A0.924
    AHI cutoff 5 specificityN/A0.801
    AHI cutoff 15 sensitivityN/A0.909
    AHI cutoff 15 specificityN/A0.908
    AHI cutoff 30 sensitivityN/A1.000
    AHI cutoff 30 specificityN/A0.933

    2. Sample Size and Data Provenance for the Test Set

    • SpO2 Validation Study:
      • Sample Size: 12 subjects
      • Data Provenance: University of California San Francisco (UCSF) - implies prospective study data from the US.
    • Sleep Validation Study:
      • Sample Size: 147 qualified samples
      • Data Provenance: Duke University Hospital - implies prospective study data from the US.

    3. Number of Experts and Qualifications for Ground Truth

    The document does not explicitly state the number or specific qualifications of experts used to establish the ground truth for either the SpO2 validation study or the sleep validation study.

    • SpO2 Validation: The ground truth was established by "reference arterial blood gas sampling," which is a clinical standard, not an expert panel.
    • Sleep Validation: The ground truth was established by "the gold standard in-lab polysomnography (PSG)." PSG scoring typically involves trained sleep technologists and often review by a board-certified sleep physician, but the number and qualifications of these individuals are not specified in the document.

    4. Adjudication Method for the Test Set

    The document does not specify any adjudication method (e.g., 2+1, 3+1) for either the SpO2 validation study or the sleep validation study.


    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No mention of a Multi-Reader Multi-Case (MRMC) comparative effectiveness study or any effect size for human readers improving with AI assistance is present in the document. The studies described are standalone performance validations against a gold standard.


    6. Standalone Performance Study (Algorithm Only)

    Yes, standalone performance studies were done. Both the SpO2 validation study and the sleep validation study describe the TipTraQ device's performance (which includes its algorithm, "TipTraQ Algorithm System") compared to a "gold standard" or reference method, without human-in-the-loop assessment as part of the primary performance metrics. The stated "TipTraQ Algorithm System determines Total Sleep Time (TST), Total REM Time (TREMT), Oxygen Desaturation Index (ODI), and Apnea-Hypopnea Index (AHI) from the recorded waveform signals," clearly indicates an algorithm-only evaluation for these parameters.


    7. Type of Ground Truth Used

    • SpO2 Validation Study: "Reference arterial blood gas sampling." (Clinical standard measurement)
    • Sleep Validation Study: "Gold standard in-lab polysomnography (PSG)." (Clinical standard physiological monitoring and expert scoring)

    8. Sample Size for the Training Set

    The document does not provide any information regarding the sample size for the training set used for the TipTraQ Algorithm System.


    9. How the Ground Truth for the Training Set Was Established

    The document does not provide any information regarding how the ground truth for the training set was established. It only describes the validation studies.

    Ask a Question

    Ask a specific question about this device

    K Number
    K244027

    Validate with FDA (Live)

    Device Name
    SANSA HSAT
    Manufacturer
    Date Cleared
    2025-01-28

    (29 days)

    Product Code
    Regulation Number
    868.2375
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Huxley Home Sleep Apnea Test (SANSA) is a wearable device intended for use in the recording, and storage of biophysical parameters to aid in the evaluation of sleep-related breathing disorders of adults suspected of sleep apnea. The device is intended for the clinical and home use setting under the direction of a Healthcare Professional (HCP).

    Device Description

    The Huxley Home Sleep Apnea Test (SANSA™) is a wearable device intended for use in the recording, analysis, and storage of biophysical parameters to aid in the evaluation of sleep-related breathing disorders of adults suspected of sleep apnea. The device is intended for clinical and home use setting under the direction of a Healthcare Professional (HCP). The system is prescription use only.

    The SANSA HSAT collects multiple physiological signals using a single wearable patch worn on the chest. The SANSA device contains a reflective PPG sensor, a single-lead ECG sensor, and a 3-axis accelerometer. The signals from these sensors are passed into a cloud-based algorithm which utilizes a combination of signal processing and AI/ML components to compute time-series data for clinician review and summary metrics for report output. The device outputs the following time-series channels: Oximetry, Heart Rate, Chest Movement, Snoring, Body Position, Respiratory Effort, Actigraphy, Sleep staging (Sleep/Wake), and ECG (reference channel only). The following summary metrics are calculated: sansa-Apnea Hypopnea Index (sAHI) and Total Sleep Time (TST).

    Recorded data are uploaded to a software portal where physiological tracings are made available for review and event editing by a qualified healthcare professional. The device is intended to be worn for 10 hours per study. The ECG channel is used for diagnostic purposes and is not intended by any automated ECG analysis system or algorithm.

    AI/ML Overview

    This document (K244027) is a Special 510(k) submission for the SANSA HSAT device, focusing on adding cellular capability. This type of submission is used when a modification to an already cleared device does not significantly change its safety or effectiveness or its indications for use. Therefore, the information provided primarily concerns the equivalence of the modified device to its predicate (which is its own previous version, K240285).

    The document states that there are no changes to the Indications for Use, physical hardware, principles of operation, device offerings, manufacturing process, or device packaging as part of the cellular update. This means that the core performance characteristics related to sleep apnea detection and physiological measurements have already been established and cleared with the predicate device.

    Given this context, the document does not contain a new, comprehensive clinical study proving the device meets new acceptance criteria for its core function (sleep apnea detection). Instead, it focuses on non-clinical testing to demonstrate that the addition of cellular capability does not negatively impact the device's original performance and introduces no new safety or effectiveness concerns.

    Therefore, many of the requested points regarding acceptance criteria and clinical study details for the sleep apnea detection function are not included in this specific Special 510(k) document, as they were addressed in the original clearance (K240285). The performance claims for the device (e.g., Sensitivity, Specificity for OSA) are listed as "Identical" to the predicate, meaning these were established in the predicate's clearance.

    However, I can extract the information related to the changes made and the testing performed for this specific submission (K244027):

    Here's an analysis based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    For the specific change (cellular capability), the document does not present a table of new clinical acceptance criteria. Instead, it asserts that the performance remains "Identical" to the predicate, and non-clinical testing verifies that the cellular addition does not degrade this performance.

    The "Performance" criteria listed under the "Device Comparison" table are for the main function of the device (sleep apnea detection and physiological measurements), which are stated to be identical to the predicate. These are not new acceptance criteria for the cellular update, but rather the previously established performance of the base device.

    Performance Metric (from Predicate)Acceptance Criteria (from Predicate)Reported Device Performance (for Subject Device)
    Heart RateRms ≤ 3 bpm (range 30-250 bpm)Rms ≤ 3 bpm (range 30-250 bpm) (Identical)
    SpO2Rms ≤ 3% (range 70-100%)Rms ≤ 3% (range 70-100%) (Identical)
    Aid to Diagnosis of Moderate to Severe OSA (AHI ≥ 15) Sensitivity88.2%88.2% (Identical)
    Aid to Diagnosis of Moderate to Severe OSA (AHI ≥ 15) Specificity87.3%87.3% (Identical)

    For the cellular capability itself, the acceptance is implicitly based on:

    • Successful wireless data transfer via Cellular (LTE-M).
    • No regression in existing functionality.
    • Maintenance of data integrity during upload.
    • Meeting cybersecurity requirements.
    • Compliance with wireless testing standards (FCC 47 CFR Part 15, Part 24, Part 27, Part 2).

    2. Sample Size Used for the Test Set and Data Provenance

    This Special 510(k) primarily relies on non-clinical testing for the cellular capability. Therefore, there is no "test set" in the sense of a patient cohort for a clinical study.

    • Test Set (for the cellular update): Not applicable in terms of patient data. The testing involved software testing, cybersecurity testing, and wireless testing. The sample size for these types of tests would be in terms of test cases, configurations, and test environments, not patient data.
    • Data Provenance: Not applicable for the cellular update. The statement "Identical" for performance refers to the original data from the predicate device's clearance.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    Not applicable for this Special 510(k) submission, as no new clinical study requiring ground truth establishment by experts for sleep apnea diagnosis was performed for the cellular update. The "Identical" performance refers to the ground truth established for the predicate device.

    4. Adjudication Method for the Test Set

    Not applicable for this Special 510(k) submission.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No. This submission is for a modification (cellular capability) to an existing device, not a new AI-assisted diagnostic tool requiring a human-in-the-loop MRMC study. The "device performance" listed is for the standalone device, not in assistance with human readers in a comparative setting.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) was Done

    Yes, the performance metrics (Sensitivity, Specificity for OSA, Heart Rate, SpO2 accuracy) listed in the "Device Comparison" table represent the standalone performance of the SANSA HSAT device's algorithms, without human intervention in the interpretation of these core metrics. These metrics are stated to be "Identical" to the predicate, implying this standalone performance was established during the predicate's clearance.

    7. The Type of Ground Truth Used

    For the core function of sleep apnea evaluation (sAHI, etc.), the document does not explicitly state the ground truth used. However, for sleep apnea diagnostics, the common ground truth is Polysomnography (PSG) scored by board-certified sleep technologists and interpreted by board-certified sleep physicians. Since the device aids in the evaluation of sleep-related breathing disorders, it's highly likely that full PSG served as the ground truth for the predicate device's performance validation. This document does not provide details of the ground truth for the predicate.

    8. The Sample Size for the Training Set

    The document does not provide information on the training set sample size. This would have been part of the original K240285 submission, as this document focuses only on changes related to cellular connectivity.

    9. How the Ground Truth for the Training Set was Established

    The document does not provide information on how the ground truth for the training set was established. This would have been part of the original K240285 submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K242290

    Validate with FDA (Live)

    Device Name
    DormoTech NLab
    Date Cleared
    2025-01-08

    (159 days)

    Product Code
    Regulation Number
    868.2375
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DormoTech Nlab is a physiological data recorder intended to collect and record data from multiple physiological channels for use by clinical software used in polysomnography and sleep disorder studies. It is intended for use by or on the order of a physician. It is intended for use on patients greater than 6 years of age in a supervised (hospital) or unsupervised (home) environments.

    Device Description

    The DormoTech Nlab is a physiological data recorder intended to collect and record data from multiple physiological channels for use by clinical software used in polysomnography and sleep disorder studies. It is intended for use by or on the order of a physician and is intended for use on patients greater than 6 years of age in a supervised (hospital) or unsupervised (home) environment.

    It consists of:

    The Head Unit
    The head unit acquires electric signals indicative of EEG and eye movement located in the upper part of the unit (on the patient's forehead). Plethysmograph for Heart rate and SpO2 measurements, and head relative position to body position are also measured using accelerometer sensors located in the upper part of the unit. The middle part of the unit is located below the mouth, it contains 2 nasal (one in each nostril) and 1 oral airflow sensors (thermistors), 1 EMG sensor, along with a snoring sensor.

    The Body Unit
    The body unit is made of 2 belts, the upper belt sits on the lower belt sits on the stomach of the patient. Both belts contain respiratory effort and accelerometer sensors, in addition, the upper belt contains an accelerometer to measure body position.

    The ExG Unit
    The ExG unit is put on using an adhesive sticker on the leg/arm/chest (either leg/arm is ok). It consists of 3 electrodes that acquire leg/arm/chest ExG signal.

    The Central Unit
    The head, chest and ExG units communicate with the central unit via Bluetooth, the wearable units send the measured data to the central unit receives the data, stores it within an internal flash drive and then transmits the data via Wi-Fi to online servers for further diagnosis. The central unit is located in the test room (up to 10 meters from the patient). The central unit has no contact with the patient.

    To ensure device reusability between sessions or patients, sections of the head unit have been designed to be detachable or with barriers between the sensor and patient's skin to stop direct contact. Specifically, the head unit incorporates (1) a detachable nasal and oral airflow section to allow for replacement of the section between each use of the device, (2) a removable barrier over the plethysmograph recorder to create a separation between the patient's skin and the recording device which is replaced between each use of the device and (3) a detachable textile on the inner side of the head unit (i.e., the side in direct contact with the patient's skin) to allow for replacement between each use of the device.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the studies that prove the DormoTech NLab device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are effectively demonstrated by the "good agreement" between the DormoTech NLab and the gold standard PSG, and the acceptable RMSE for the SpO2 sensor.

    Acceptance CriteriaReported Device PerformanceComments
    Polysomnography (PSG) Parameters (Compared to Gold Standard PSG)"Good agreement" between devices, mean difference close to zero, and narrow limits of agreement, indicating interchangeability. Performance considered "substantially equivalent."
    AHI (events/h)Mean Difference: 0.2875Accepted
    ODI (events/h)Mean Difference: -0.1042Accepted
    Snore (%)Mean Difference: 1.995Accepted
    Sleep Latency (Minutes)Mean Difference: 0.9727Accepted
    REM Latency (Minutes)Mean Difference: -0.2864Accepted
    Wake after Sleep Onset (Minutes)Mean Difference: -2.091Accepted
    REM (%)Mean Difference: 0.6045Accepted
    N1 (%)Mean Difference: -1.659Accepted
    N2 (%)Mean Difference: -1.095Accepted
    N3 (%)Mean Difference: 2.173Accepted
    Wake (%)Mean Difference: -1.127Accepted
    Total Sleep Time (Minutes)Mean Difference: 4.00Accepted
    Sleep Efficiency (%)Mean Difference: -0.1773Accepted
    Position (Up) (%)Mean Difference: -0.2792Accepted
    Position (Supine) (%)Mean Difference: 1.892Accepted
    Position (Left) (%)Mean Difference: 0.725Accepted
    Position (Right) (%)Mean Difference: -0.3042Accepted
    SpO2 Sensor Accuracy (Compared to Arterial HbO2 Saturations)Root Mean Square Error (RMSE) of 2.53% for 70-100% SpO2, which is explicitly stated as being "within the expected range of accuracy" and allowable per ISO 80601-2-61 (up to 4%).
    SpO2 Accuracy (70-100%)RMSE: 2.53%Accepted
    Pulse rate (20-250 bpm)±3 bpmAccepted (similar to reference)

    2. Sample Size Used for the Test Set and Data Provenance

    • Polysomnography (PSG) Study:
      • Sample Size: 24 subjects (out of 26 recruited) completed the study.
      • Data Provenance: Prospective clinical study conducted in two sleep labs in Israel (Shamir Medical Center Be'er Ya'akov, and Millenium Sleep Clinic Be'er Sheva).
    • SpO2 Sensor Study:
      • Sample Size: 12 patients. 259 data points were included in the analysis.
      • Data Provenance: Prospective clinical study. No specific country of origin is mentioned, but it's likely part of the overall clinical trials for the device, potentially in Israel as well.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    The document does not explicitly state the number of experts or their specific qualifications (e.g., "radiologist with 10 years of experience") used to establish the ground truth for human-scored parameters in the PSG study. However, the ground truth for the PSG study is based on a "gold standard polysomnogram (PSG) study," which inherently implies scoring by trained sleep specialists.

    For the SpO2 sensor study, the ground truth was established by "simultaneous monitoring of arterial HbO2 saturations at six different levels of oxyhemoglobin saturation between 70-100%," which is a direct physiological measurement, not an expert-based ground truth.

    4. Adjudication Method for the Test Set

    The document does not specify an adjudication method (e.g., 2+1, 3+1) for the human-scored PSG parameters. The mention of "gold standard PSG" suggests that standard clinical scoring practices were followed, which typically involve certified polysomnography technologists scoring sleep studies.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No multi-reader multi-case (MRMC) comparative effectiveness study was done. The study compares the device's readings directly to a gold-standard PSG for the sleep parameters and directly to arterial blood gas measurements for SpO2, not comparing human readers with and without AI assistance.

    6. Standalone (Algorithm Only) Performance

    The device is described as a "physiological data recorder intended to collect and record data from multiple physiological channels for use by clinical software used in polysomnography and sleep disorder studies." This implies that the device itself is a data acquisition unit, and its performance is evaluated in terms of its ability to accurately record these physiological signals compared to established methods.

    • For PSG parameters: The "DormoTech NLab" is compared to a "gold standard polysomnogram (PSG)." This is a standalone performance evaluation of the device as a data recorder, with the implicit understanding that the data collected by both the NLab and the predicate PSG would then be analyzed by clinical software (or human scorers). The "mean difference" and "limits of agreement" directly assess the NLab's standalone measurement accuracy against the gold standard.
    • For SpO2 sensor: The "DormoTech NLab SpO2 sensor" is compared to "arterial HbO2 saturations." This is a standalone performance evaluation of the sensor's accuracy.

    7. Type of Ground Truth Used

    • Polysomnography (PSG) Parameters: "Gold standard polysomnogram (PSG) study." This refers to a comprehensive sleep study recorded and scored according to established clinical guidelines, serving as the benchmark for various sleep parameters. It's essentially expert-interpreted physiological data.
    • SpO2 Sensor: "Arterial HbO2 saturations." This is direct physiological outcomes data obtained through arterial blood gas measurements, which is considered the most accurate measure of blood oxygen saturation.

    8. Sample Size for the Training Set

    The document does not explicitly state the sample size for any training set. It focuses on the validation studies, which are test sets. This device primarily functions as a physiological data recorder, not necessarily an AI/machine learning algorithm requiring a separate training set in the typical sense. Any internal calibration or algorithm development would likely have used internal datasets, but these are not detailed in this premarket notification.

    9. How Ground Truth for the Training Set Was Established

    Since no training set details are provided, the method for establishing its ground truth is also not mentioned.

    Ask a Question

    Ask a specific question about this device