K Number
K250882

Validate with FDA (Live)

Device Name
SANSA HSAT
Manufacturer
Date Cleared
2025-10-29

(219 days)

Product Code
Regulation Number
868.2375
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Huxley Home Sleep Apnea Test (SANSA) is a wearable device intended for use in the recording, analysis, and storage of biophysical parameters to aid in the evaluation of sleep-related breathing and cardiac disorders of adults suspected of sleep apnea. The device is intended for the clinical and home use setting under the direction of a Healthcare Professional (HCP). The system is prescription use only.

The SANSA device records and stores ECG recording for up to 10 hours of wear time which can be displayed in the software portal for manual annotation and analysis. The SANSA does not provide automated analysis of the ECG and is not intended to be used with a 3rd party automated algorithm and is not intended for pacemaker analysis.

Device Description

The Huxley Home Sleep Apnea Test (SANSA™) is a wearable device intended for use in the recording, analysis, and storage of biophysical parameters to aid in the evaluation of sleep-related breathing disorders of adults suspected of sleep apnea. The device is intended for clinical and home use setting under the direction of a Healthcare Professional (HCP). The system is prescription use only. The SANSA™ device records and stores ECG recording for up to 10 hours of wear time which can be displayed in the software portal for manual annotation and analysis. The SANSA™ does not provide automated analysis of the ECG and is not intended to be used with a 3rd party automated algorithm and is not intended for pacemaker analysis.

The SANSA HSAT collects multiple physiological signals using a single wearable patch worn on the chest. The SANSA device contains a reflective PPG sensor, a single-lead ECG sensor, and a 3-axis accelerometer. The signals from these sensors are passed into a cloud-based algorithm which utilizes a combination of signal processing and AI/ML components to compute time-series data for clinician review and summary metrics for report output. The device outputs the following time-series channels: Oximetry, Heart Rate, Chest Movement, Snoring, Body Position, Respiratory Effort, Actigraphy, Sleep staging (Sleep/Wake), and ECG. The following summary metrics are calculated: sansa-Apnea Hypopnea Index (sAHI) and Total Sleep Time (TST).

Recorded data are uploaded to a software portal where physiological tracings are made available for review and event editing by a qualified healthcare professional. The device is intended to be worn for 10 hours per study.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study information for the SANSA HSAT device, based on the provided FDA 510(k) clearance letter:

1. Acceptance Criteria and Reported Device Performance

The acceptance criteria and performance data are primarily found on Page 9 of the document, under the "Performance" section within "Table 1: Device Comparison."

Performance MetricAcceptance Criteria / Predicate Performance (K202359)Reported Device Performance (SANSA HSAT)Comparison
Heart Rate (Arms)No Heart Rate Analysis (Predicate)≤ 3 bpm (range 30-250 bpm)N/A
SpO2 (Arms)Does not collect SpO2 (Predicate)≤ 3% (range 70-100%)N/A
Aid to Diagnosis of Moderate to Severe OSA (AHI≥15) - SensitivityNot applicable (Predicate does not diagnose OSA)88.2%N/A
Aid to Diagnosis of Moderate to Severe OSA (AHI≥15) - SpecificityNot applicable (Predicate does not diagnose OSA)87.3%N/A
ECG Recording Accuracy: Frequency Response0.67 Hz to 40 Hz0.67 Hz to 40 HzIdentical
ECG Recording Accuracy: Input Impedance>10 MΩ>10 MΩIdentical
ECG Recording Accuracy: Gain AccuracyMaximum amplitude error ±10%Maximum amplitude error ±10%Identical
ECG Recording Accuracy: Gain Stability<3% over a 24-hour period<3% over a 24-hour periodIdentical
ECG Recording Accuracy: Timing Accuracy<30 seconds<30 secondsIdentical

Note: For Heart Rate, SpO2, and OSA diagnosis metrics, the predicate device (iRhythm Zio Monitor) does not provide these analyses or collect SpO2. Therefore, the "acceptance criteria" for these aspects are implicitly met by the absence of safety/effectiveness concerns with the SANSA HSAT's reported performance, which aligns with its intended use in sleep apnea evaluation. The ECG recording accuracy metrics are directly comparable and are identical between the subject device and the predicate.

2. Sample Size and Data Provenance for the Test Set

The document does not explicitly state the sample size used for the clinical performance validation of the SANSA HSAT's ECG, nor does it specify the country of origin or whether the data was retrospective or prospective. It only mentions:

  • Sample Size: Not specified.
  • Data Provenance: Not specified (country/retrospective/prospective).

For the Aid to Diagnosis of OSA, a sensitivity of 88.2% and specificity of 87.3% are reported. The document does not provide the sample size or provenance for this particular study, but it was leveraged from a previous clearance (K244027).

3. Number of Experts and Qualifications for Ground Truth - Test Set

The document does not provide details on the number or qualifications of experts used to establish ground truth for the test set specifically for the ECG clinical performance validation.

For the "Aid to Diagnosis of Moderate to Severe OSA", the percentages suggest a comparison against a diagnostic standard, which would typically involve expert interpretation, but the details are not provided in this document.

4. Adjudication Method for the Test Set

The document does not specify any adjudication method (e.g., 2+1, 3+1) used for establishing the ground truth for the test set.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No information is provided about a multi-reader multi-case (MRMC) comparative effectiveness study, or any effect size related to human reader improvement with or without AI assistance. The SANSA device explicitly states it "does not provide automated analysis of the ECG and is not intended to be used with a 3rd party automated algorithm." (Pages 4 and 5).

6. Standalone Performance Study

Yes, a standalone performance study was done for the SANSA HSAT's ECG:

  • Study Type: "Sansa ECG clinical performance was validated through comparison to a simultaneously collected reference standard Holter monitor." (Page 11)
  • Performance: "Clinically acceptable performance was demonstrated through qualitative and quantitative analysis of the ECG signal." (Page 11)

The reported sensitivity and specificity for OSA diagnosis (Page 9) also represent standalone algorithm performance.

7. Type of Ground Truth Used

  • For ECG Clinical Performance: "simultaneously collected reference standard Holter monitor." (Page 11)
  • For Aid to Diagnosis of OSA (reported sensitivity/specificity): While not explicitly stated, the context of "Aid to Diagnosis" for Sleep Apnea typically implies comparison to a polysomnography (PSG) study, which is the gold standard, interpreted by sleep specialists. This data was "leveraged from previous clearance."

8. Sample Size for the Training Set

The document does not provide any information about the sample size used for a training set. This is consistent with the statement that the device does not provide automated analysis of the ECG and explicitly states that it is not intended for use with a 3rd party automated algorithm. While the device uses "AI/ML components" for other signals (Snoring, Body Position, Respiratory Effort, Actigraphy, Sleep staging), the training set size for these components is not detailed in this document.

9. How Ground Truth for the Training Set Was Established

The document does not provide information on how the ground truth for any potential training set was established. Given the focus on manual annotation and analysis for ECG, and the lack of detail on the AI/ML components for other signals, this information is not present in the clearance letter.

FDA 510(k) Clearance Letter - SANSA HSAT

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.08.00

October 29, 2025

Huxley Medical
℅ Grace Powers
Founder / Principal Consultant
Powers Regulatory Consulting
2451 Cumberland Blvd SE
Suite 3740
Atlanta, Georgia 30339

Re: K250882
Trade/Device Name: SANSA HSAT
Regulation Number: 21 CFR 868.2375
Regulation Name: Breathing Frequency Monitor
Regulatory Class: Class II
Product Code: MNR, BZQ, MWJ
Dated: September 26, 2025
Received: September 30, 2025

Dear Grace Powers:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K250882 - Grace Powers Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K250882 - Grace Powers Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

JENNIFER W. SHIH -S

Jennifer Kozen
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

Indications for Use

Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions.
K250882

Please provide the device trade name(s).
SANSA HSAT

Please provide your Indications for Use below.

The Huxley Home Sleep Apnea Test (SANSA) is a wearable device intended for use in the recording, analysis, and storage of biophysical parameters to aid in the evaluation of sleep-related breathing and cardiac disorders of adults suspected of sleep apnea. The device is intended for the clinical and home use setting under the direction of a Healthcare Professional (HCP). The system is prescription use only.

The SANSA device records and stores ECG recording for up to 10 hours of wear time which can be displayed in the software portal for manual annotation and analysis. The SANSA does not provide automated analysis of the ECG and is not intended to be used with a 3rd party automated algorithm and is not intended for pacemaker analysis.

Please select the types of uses (select one or both, as applicable).
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

SANSA HSAT Page 8 of 61

Page 5

510(k) Summary Traditional 510(k) – Huxley Medical

K250882 SANSA HSAT

510(k) Summary

In accordance with 21 CFR §807.92 and the Safe Medical Devices Act of 1990, the following information is provided for the Huxley Medical Traditional 510(k) premarket notification.

Sponsor: Huxley Medical, Inc.
1465 Northside Dr NW Ste 217
Atlanta, GA, 30318
info@huxleymed.com

Submission Contact: Grace Powers, MS, MBA, RAC
Founder/Principal Consultant
Powers Regulatory Consulting
grace@powersregulatory.com
404-931-8730

Submission Date: March 24, 2025

Subject Device: Trade Name: SANSA™ HSAT
Common Name: Ventilatory Effort Recorder
Classification Name: Breathing Frequency Monitor
Regulation: 21 CFR §868.2375
Regulatory Classification: Class 2
Product Code: MNR, BZQ, MWJ

Predicate Device: iRhythm Zio Monitor (K202359)
Reference Device: Huxley Medical SANSA HSAT (K244027)

Device Description

The Huxley Home Sleep Apnea Test (SANSA™) is a wearable device intended for use in the recording, analysis, and storage of biophysical parameters to aid in the evaluation of sleep-related breathing disorders of adults suspected of sleep apnea. The device is intended for clinical and home use setting under the direction of a Healthcare Professional (HCP). The system is prescription use only. The SANSA™ device records and stores ECG recording for up to 10 hours of wear time which can be displayed in the software portal for manual annotation and analysis. The SANSA™ does not provide automated analysis of the ECG and is not intended to be used with a 3rd party automated algorithm and is not intended for pacemaker analysis.

The SANSA HSAT collects multiple physiological signals using a single wearable patch worn on the chest. The SANSA device contains a reflective PPG sensor, a single-lead ECG sensor, and a 3-axis accelerometer. The signals from these sensors are passed into a cloud-based algorithm which utilizes a combination of signal processing and AI/ML components to compute time-series data for clinician review and summary metrics for report output. The device outputs the following time-series channels: Oximetry, Heart Rate,

Page 6

510(k) Summary Traditional 510(k) – Huxley Medical
K250882 SANSA HSAT

Chest Movement, Snoring, Body Position, Respiratory Effort, Actigraphy, Sleep staging (Sleep/Wake), and ECG. The following summary metrics are calculated: sansa-Apnea Hypopnea Index (sAHI) and Total Sleep Time (TST).

Recorded data are uploaded to a software portal where physiological tracings are made available for review and event editing by a qualified healthcare professional. The device is intended to be worn for 10 hours per study.

Intended Use/Indications for Use

The Huxley Home Sleep Apnea Test (SANSA™) is a wearable device intended for use in the recording, analysis, and storage of biophysical parameters to aid in the evaluation of sleep-related breathing and cardiac disorders of adults suspected of sleep apnea. The device is intended for the clinical and home use setting under the direction of a Healthcare Professional (HCP). The system is prescription use only.

The SANSA device records and stores ECG recording for up to 10 hours of wear time which can be displayed in the software portal for manual annotation and analysis. The SANSA does not provide automated analysis of the ECG and is not intended to be used with a 3rd party automated algorithm and is not intended for pacemaker analysis.

An overview comparison of the SANSA HSAT (subject device) to the predicate device and reference device are presented in the table below.

Table 1: Device Comparison

Device ComparisonSubject Device: SANSA HSATPredicate Device: Zio Monitor (K202359)Reference Device: SANSA HSAT (K244027)Comparison
ManufacturerHuxley Medical, Inc.iRhythm Technologies, IncHuxley Medical, Inc.Identical to reference device.
FDA Product CodesMNR, BZQ, MWJDSH, MWJMNR, BZQIdentical to reference device with the addition of MWJ from the predicate device.
Primary Regulation868.2375870.2800868.2375Identical to reference device.
ClassificationIIIIIIIdentical to reference device.
Classification NameVentilatory Effort RecorderMedical magnetic tape recorderVentilatory Effort RecorderIdentical to reference device.
Advisory CommitteeAnesthesiologyCardiovascularAnesthesiologyIdentical to reference device.
Prescription Use?YesYesYesIdentical
Indications for UseThe Huxley Home Sleep Apnea Test (SANSA™) is aThe Zio Monitor is a prescription-only, single-patient-use,The Huxley Home Sleep Apnea Test (SANSA™) is aEquivalent - The addition of ECG monitoring

Page 7

510(k) Summary Traditional 510(k) – Huxley Medical
K250882 SANSA HSAT

Device ComparisonSubject Device: SANSA HSATPredicate Device: Zio Monitor (K202359)Reference Device: SANSA HSAT (K244027)Comparison
wearable device intended for use in the recording, analysis, and storage of biophysical parameters to aid in the evaluation of sleep-related breathing disorders of adults suspected of sleep apnea. The device is intended for the clinical and home use setting under the direction of a Healthcare Professional (HCP). The system is prescription use only. The SANSA™ device records and stores ECG recording for up to 10 hours of wear time which can be displayed in the software portal for manual annotation and analysis. The SANSA™ does not provide automated analysis of the ECG and is not intended to be used with a 3rd party automated algorithm and is not intended for pacemaker analysis.ECG monitor that continuously records data for up to 14 days. It is indicated for use on patients who may be asymptomatic or who may suffer from transient symptoms such as palpitations, shortness of breath, dizziness, lightheadedness, pre-syncope, syncope, fatigue or anxiety.wearable device intended for use in the recording, analysis, and storage of biophysical parameters to aid in the evaluation of sleep-related breathing disorders of adults suspected of sleep apnea. The device is intended for the clinical and home use setting under the direction of a Healthcare Professional (HCP).capability for manual annotation and analysis of the ECG signal with no automated ECG analysis is supported by the indication for use of the predicate device.
Target PopulationAdults (22 years age and older)18 years and olderAdults (22 years age and older)Identical to reference device.
Intended UseClinics and HomeClinics and HomeClinics and HomeIdentical

Page 8

510(k) Summary Traditional 510(k) – Huxley Medical
K250882 SANSA HSAT

Device ComparisonSubject Device: SANSA HSATPredicate Device: Zio Monitor (K202359)Reference Device: SANSA HSAT (K244027)Comparison
EnvironmentUseUseUse
Device Placement for Data CollectionChest sensorChest SensorChest sensorIdentical
Device SensorsChest Sensor Accelerometer ECG Reflectance PhotoplethysmographyECGChest Sensor Accelerometer ECG Reflectance PhotoplethysmographyEquivalent - Subject and Predicate contain ECG.
ChannelsOximetry Heart rate Chest movement Snoring Body position Respiratory effort Actigraphy Sleep stage (Sleep/Wake) ECGECGOximetry Heart rate Chest movement Snoring Body position Respiratory effort Actigraphy Sleep stage (Sleep/Wake) ECG (Reference channel only)Equivalent – Official ECG channel (not reference only) is equivalent to the ECG channel of the reference device.
ECG Recording CharacteristicsNumber of Leads: Single Lead ECG Lead Orientation: Modified Lead II ECG A/D Sampling Rate: 250 Hz ECG Resolution:Number of Leads: Single Lead ECG Lead Orientation: Modified Lead II ECG A/D Sampling Rate: 200 Hz ECG Resolution:Number of Leads: Single Lead ECG Lead Orientation: Modified Lead II ECG A/D Sampling Rate: 250 Hz ECG Resolution:Identical to reference device. Equivalent to predicate device in performance in recording capability and lead orientation. Higher sample rate in subject device does not raise questions of safety or effectiveness with predicate device as the output ECG is

Page 9

510(k) Summary Traditional 510(k) – Huxley Medical
K250882 SANSA HSAT

Device ComparisonSubject Device: SANSA HSATPredicate Device: Zio Monitor (K202359)Reference Device: SANSA HSAT (K244027)Comparison
15.5 bits15.5 bits15.5 bitsequivalent.
Analysis OutputssAHI Body Position Discrete States Heart Rate Total Sleep Time SpO2No analysis provided by device.sAHI Body Position Discrete States Heart Rate Total Sleep Time SpO2Identical to reference device. Subject device and predicate device lack ECG analysis output.
PerformanceHeart Rate: Arms ≤ 3 bpm (range 30-250 bpm) SpO2 Arms ≤ 3% (range 70-100%) Aid to Diagnosis of Moderate to Severe OSA (AHI≥15): Sensitivity 88.2%, Specificity 87.3% ECG Recording Accuracy: • Frequency Response: 0.67 Hz to 40 Hz • Input Impedance: >10 MΩ • Gain Accuracy: Maximum amplitude error ±10%No Heart Rate Analysis Does not collect SpO2 Does not diagnose OSA ECG Recording Accuracy: • Frequency Response: 0.67 Hz to 40 Hz • Input Impedance: >10 MΩ • Gain Accuracy: Maximum amplitude error ±10% • Gain Stability: <3% over a 24-hour periodHeart Rate: Arms ≤ 3 bpm (range 30-250 bpm) SpO2 Arms ≤ 3% (range 70-100%) Aid to Diagnosis of Moderate to Severe OSA (AHI≥15): Sensitivity 88.2%, Specificity 87.3% ECG Recording Accuracy: • Frequency Response: 0.67 Hz to 40 Hz • Input Impedance: >10 MΩ • Gain Accuracy: Maximum amplitude error ±10%Identical to reference device. Identical ECG recording accuracy to predicate device.

Page 10

510(k) Summary Traditional 510(k) – Huxley Medical
K250882 SANSA HSAT

Device ComparisonSubject Device: SANSA HSATPredicate Device: Zio Monitor (K202359)Reference Device: SANSA HSAT (K244027)Comparison
• Gain Stability: <3% over a 24-hour period • Timing Accuracy: <30 seconds• Timing Accuracy: <30 seconds• Gain Stability: <3% over a 24-hour period • Timing Accuracy: <30 seconds
Data Collection and TransferPatient data is wirelessly transferred via Cellular (LTE-M) at the conclusion of the study. In the event of no connectivity or failed transfer, patient data is physically transferred via USB after study conclusion.Patient data is transferred after study conclusionPatient data is wirelessly transferred via Cellular (LTE-M) at the conclusion of the study. In the event of no connectivity or failed transfer, patient data is physically transferred via USB after study conclusion.Identical to reference device.
Recording CapacityApprox. 10 hours per study. 2 nights of study maximum.Continuous recording for 14 days.Approx. 10 hours per study. 2 nights of study maximum.Identical to reference device. Predicate device is focused on 14 days of data and subject and reference are focused on sleep.
Energy SourceRechargeable Lithium Polymer BatteryLithium Manganese Dioxide Coin CellRechargeable Lithium Polymer BatteryIdentical to reference device.
Analysis SoftwareAnalysis performed off the recording device, on a compatible cloud-based software platform. No automated ECG analysis conducted.No ECG analysis provided by the device.Analysis performed off the recording device, on a compatible cloud-based software platform. No automated ECG analysis conducted.Identical to reference device. Equivalent to predicate.

Page 11

510(k) Summary Traditional 510(k) – Huxley Medical
K250882 SANSA HSAT

Non-Clinical Performance Data

The following non-clinical testing was conducted to support device changes:

• Software Unit & Integration Testing to test portal functionality. This testing verified the additional features, and that the data integrity was not altered in the EDF or PDF export process.

• The following performance testing verifies the accuracy of the SANSA device's ECG recording:

  • Testing to FDA recognized consensus standard IEC 60601-2-47 Ed. 2.0 (2012-02) Medical electrical equipment - Part 2-47: Particular requirements for the basic safety and essential performance of ambulatory electrocardiographic systems
  • ECG recorder performance testing (Retested with latest device firmware)
  • Database Accuracy testing for beat detection and heart rate (testing leveraged from previous clearance due to lack of changes)
  • Testing to FDA recognized consensus standard ANSI AAMI EC12:2000/(R)2015 Disposable ECG electrodes (testing leveraged from previous clearance due to lack of changes)
  • ECG electrode performance testing
  • On-body adhesive performance testing

• Human Factors Usability Testing conducted on patients and clinicians in accordance with FDA Guidance Document "Applying Human Factors and Usability Engineering to Medical Devices" Issued February 3, 2016. (Testing leveraged from previous clearance due to lack of changes)

No further system or performance testing was required, as all updates made were isolated in cloud software functionality.

Clinical Performance Data

Sansa ECG clinical performance was validated through comparison to a simultaneously collected reference standard Holter monitor. Clinically acceptable performance was demonstrated through qualitative and quantitative analysis of the ECG signal.

Conclusion

The SANSA System is substantially equivalent to the legally marketed predicate device as demonstrated by the similar intended use and similar technologies, with the reference device leveraged to address the sleep-disordered breathing indications for use. The addition to the indications for use to allow usage of the existing ECG waveform for diagnostic purposes is supported by the predicate device. The updates made to the subject device in this submission do not raise questions of safety and effectiveness compared to the predicate and reference devices.

§ 868.2375 Breathing frequency monitor.

(a)
Identification. A breathing (ventilatory) frequency monitor is a device intended to measure or monitor a patient's respiratory rate. The device may provide an audible or visible alarm when the respiratory rate, averaged over time, is outside operator settable alarm limits. This device does not include the apnea monitor classified in § 868.2377.(b)
Classification. Class II (performance standards).