K Number
K240236

Validate with FDA (Live)

Manufacturer
Date Cleared
2024-09-24

(239 days)

Product Code
Regulation Number
870.2300
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Happy Ring Health Monitoring System is a wearable device system to remotely monitor physiologic parameters of patients in professional healthcare facilities, such as hospitals or skilled nursing facilities, or their own home. The device is intended for use on individuals who are 22 years of age or older.

The device supports continuous data collection for monitoring of the following physiological parameters:

  • Acceleration / Movement
  • Electrodermal Activity (EDA)
  • Blood Oxygen Saturation
  • Pulse Rate
  • Peripheral Skin Temperature

The Happy Ring Health Monitoring System is intended for peripheral skin temperature monitoring, where monitoring temperature at the finger is clinically indicated.

The Happy Ring Health Monitoring System is not intended for SpO2, pulse rate, respiration rate monitoring in conditions of motion or low perfusion.

Device Description

The Happy Ring Health Monitoring System is a wearable device and software platform comprising:

  • A wearable medical device smart ring,
  • A mobile app-based bluetooth-to-internet gateway,
  • A cloud-based API,
  • A set of data processing algorithms, and
  • A Physician data viewer.The Ring is worn on the user's finger and continuously collects raw data via specific sensors. These raw data are transmitted via Bluetooth Low Energy to a paired mobile device. The data received are transmitted by the mobile app gateway, via the cloud-based AP, to the data processing algorithms where various physiological parameters are computed. The raw and processed data are stored, further analyzed, and accessible by healthcare providers or researchers via the Physician data viewer.

The Happy Ring Health Monitoring System is intended for retrospective remote monitoring of physiological parameters in ambulatory adults in home-healthcare environments. It is designed to continuously collect data to support intermittent monitoring of the following physiological parameters by trained healthcare professions or researchers: Acceleration / movement, electrodermal activity (EDA), blood oxygen saturation, pulse rate, and peripheral skin temperature.

AI/ML Overview

The provided text describes information about the Happy Ring Health Monitoring System, including its intended use, technological comparison to a predicate device, and a summary of non-clinical and clinical tests performed. However, it does not contain specific acceptance criteria for the device performance or detailed results of a study proving the device meets those criteria, beyond a general statement about SpO2 accuracy.

Therefore, I cannot fully complete the requested table and answer all questions with the provided text. I will extract all available information and explicitly state what is not present.

Here's a breakdown of the available information and what's missing:

1. Table of Acceptance Criteria and Reported Device Performance

ParameterAcceptance Criteria (Not explicitly stated as such, but inferred from testing standards)Reported Device Performance
Blood Oxygen Saturation (SpO2)Per ISO 80601-2-61 (Accuracy for pulse oximeters, typically ARMS within a certain percentage)Was within 3.5% ARMS for the range of oxygen saturation measured by the device.
Pulse RatePer ISO 80601-2-61 (Accuracy for pulse oximeters)Tested in accordance with ISO 80601-2-61, but specific quantitative performance not reported.
Peripheral Skin TemperaturePer ISO 80601-2-56 (Accuracy for clinical thermometers)Tested in accordance with relevant sections of ISO 80601-2-56, but specific quantitative performance not reported.
Electrodermal Activity (EDA)Bench testing to verify performanceBench testing to verify performance, but specific quantitative performance not reported.
Acceleration / MovementBench testing to verify performanceBench testing to verify performance, but specific quantitative performance not reported.
Electrical, Mechanical & Thermal SafetyIEC 60601-1 and IEC 60601-1-11 complianceTesting in accordance with standards.
Electromagnetic CompatibilityIEC 60601-1-2 complianceTesting in accordance with standards.
Wireless CoexistenceFDA's guidance: Radio Frequency Wireless Technology in Medical Devices complianceTesting in accordance with guidance.
UsabilityIEC 62366 and FDA's guidance: Applying Human Factors and Usability Engineering to Medical Devices complianceTesting in accordance with standards/guidance.
Software V&V & CybersecurityFDA's guidance for Software Contained in Medical Devices and Cybersecurity in Medical Devices complianceDocumentation provided as recommended by guidance.

2. Sample size used for the test set and the data provenance:

  • Sample Size (SpO2 Clinical Study): 12 subjects
  • Data Provenance: Not explicitly stated regarding country of origin or whether it was retrospective or prospective. The study was described as "clinical tests" and involved "subjects," which implies a prospective clinical trial.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not specified. The text mentions "oxyhemoglobin saturation using a radial arterial line" as the comparison method for SpO2, which is an objective measurement and doesn't typically involve expert consensus for ground truth establishment in the same way image interpretation might. For other physiological parameters, the ground truth establishment method is not detailed beyond "bench testing to verify performance."

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • Not applicable/Not specified. Given the nature of objective physiological measurements (like radial arterial line for SpO2), a multi-reader adjudication method as seen in image interpretation studies is not typically used.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No. An MRMC study was not described. The study focused on the device's accuracy in measuring physiological parameters, not on how it assists human readers or clinicians in interpreting data or making decisions. The "Physician data viewer" is mentioned, suggesting human review, but no comparative effectiveness study with human readers is detailed.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Yes, implicitly. The SpO2 and other bench tests (EDA, accelerometer, temperature) assess the device's ability to measure physiological parameters independently of human interpretation. The "data processing algorithms" compute the parameters, and their accuracy is evaluated.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • Reference Devices/Objective Measurements:
    • For SpO2: Oxyhemoglobin saturation using a radial arterial line (a gold standard, objective measurement).
    • For other parameters: Implied reference instruments for "bench testing to verify the performance." This typically means comparing the device's output against a highly accurate, calibrated reference measurement.

8. The sample size for the training set:

  • Not specified. The document discusses the test set but provides no information about the training set size for the device's algorithms.

9. How the ground truth for the training set was established:

  • Not specified. Since information about the training set size or its existence is not provided, the method for establishing its ground truth is also not elaborated upon in this document.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 24, 2024

Happy Health Inc. % Rakesh Lal Consultant Rx Device Consulting 7 Courtyard Pl Lexington, Massachusetts 02420

Re: K240236

Trade/Device Name: Happy Ring Health Monitoring System Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac monitor (including cardiotachometer and rate alarm) Regulatory Class: Class II Product Code: MWI, DQA, FLL, DRG Dated: January 10, 2024 Received: January 29, 2024

Dear Rakesh Lal:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

Jennifer W. Shih -S

Jennifer Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)

K240236

Device Name

Happy Ring Health Monitoring System

Indications for Use (Describe)

The Happy Ring Health Monitoring System is a wearable device system to remotely monitor physiologic parameters of patients in professional healthcare facilities, such as hospitals or skilled nursing facilities, or their own home. The device is intended for use on individuals who are 22 vears of age or older.

The device supports continuous data collection for monitoring of the following physiological parameters:

  • · Acceleration / Movement
  • · Electrodermal Activity (EDA)
  • · Blood Oxygen Saturation
  • · Pulse Rate
  • · Peripheral Skin Temperature

The Happy Ring Health Monitoring System is intended for peripheral skin temperature monitoring, where monitoring temperature at the finger is clinically indicated.

The Happy Ring Health Monitoring System is not intended for SpO2, pulse rate, respiration rate monitoring in conditions of motion or low perfusion.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) #: K240236

510(k) Summary

Prepared on: 2024-08-21

tails

21 CFR 807.92(a)(1)

Applicant NameHappy Health Inc.
Applicant Address3200 Gracie Kiltz Ln, #301 Austin TX 78758 United States
Applicant Contact Telephone(512) 686-8572
Applicant ContactDr. Dustin Freckleton
Applicant Contact Emaildustin@happy.ai.com
Correspondent NameRx Device Consulting
Correspondent Address7 Courtyard Pl Lexington MA 02420 United States
Correspondent Contact Telephone8177348303
Correspondent ContactMr. Rakesh Lal
Correspondent Contact Emailrakesh@rxdeviceconsulting.com

21 CFR 807.92(a)(1)

Device Trade NameHappy Ring Health Monitoring System
Common NameCardiac monitor (including cardiotachometer and rate alarm)
Classification NameMonitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms)
Regulation Number870.2300
Product CodeMWI

21 CFR 807.92(a)(2)
Legally Marketed Predicate Devices
21 CFR 807.92(a)(3)

Predicate #Predicate Trade Name (Primary Predicate is listed first)Product Code
K230457Empatica Health Monitoring Platform; EmbracePlus; EmpaticaMWI

Device Description Summary
21 CFR 807.92(a)(4)
The Happy Ring Health Monitoring System is a wearable device and software platform comprising:

  • A wearable medical device smart ring,
  • A mobile app-based bluetooth-to-internet gateway,
  • A cloud-based API,
  • A set of data processing algorithms, and
  • A Physician data viewer.The Ring is worn on the user's finger and continuously collects raw data via specific sensors. These raw data are transmitted via Bluetooth Low Energy to a paired mobile device. The data received are transmitted by the mobile app gateway, via the cloud-based AP, to the data processing algorithms where various physiological parameters are computed. The raw and processed data are stored, further analyzed, and accessible by healthcare providers or researchers via the Physician data viewer.

{5}------------------------------------------------

The Happy Ring Health Monitoring System is intended for retrospective remote monitoring of physiological parameters in ambulatory adults in home-healthcare environments. It is designed to continuously collect data to support intermittent monitoring of the following physiological parameters by trained healthcare professions or researchers: Acceleration / movement, electrodermal activity (EDA), blood oxygen saturation, pulse rate, and peripheral skin temperature.

Intended Use/Indications for Use

21 CFR 807.92

The Happy Ring Health Monitoring System is a wearable device system to remotely monitor physiologic parameters of patients in professional healthcare facilities, such as hospitals or skilled nursing facilities, or their own home. The device is intended for use on individuals who are 22 years of age or older.

The device supports continuous data collection for monitoring of the following physiological parameters:

  • Acceleration / Movement
  • · Electrodermal Activity (EDA)
  • Blood Oxygen Saturation
  • · Pulse Rate
  • · Peripheral Skin Temperature

The Happy Ring Health Monitoring System is intended for perioneral skin temperature monitoring temperature at the finger is clinically indicated.

The Happy Ring Health Monitoring System is not intended for SpO2 or pulse rate monitoring in conditions of motion or low perfusion.

Indications for Use Comparison

21 CFR 807.92(a)(3)

The subject device and predicate device have very similar indications for use, and therefore have the same intended use. Both devices are intended to measure various physiological parameters in adults (18 years of age or older) in home or healthcare facilities for the purpose of retrospective review by a clinician or researcher. In particular, both devices measure blood oxygen saturation, pulse rate, perioheral skin temperature, and electrodermal activity. The subject device measures acceleration while the measures activity associated with movement during sleep. This minor difference in indications for use does not constitute a new intended use.

Technological Comparison

The subject device and predicate device both have very similar technological characteristics. Both devices include a wearable sensor that acquires various raw signals, and transmits these raw signals to a cloud application via a mobile application. The communication between wearable device and mobile application occurs via Bluetooth Low Energy and the communication between mobile application and cloud application occurs via Wi-Fi or cellular for both devices, raw signal data are processed to determine and display various physiological measures (pulse rate, blood oxygen saturation, peripheral skin temperature, and electrodermal activity), and for these measures, the technical method to determine the measures are similar. Both devices power the a rechargeable lithium battery. Both devices include software, and include user interfaces for patient and clinician.

Some minor differences in technological characteristics exist between the two devices. The specific implementation of the software within each device is different but this does not raise different questions of safety and effectiveness as the general function of the software within each device is similar. The sensors used by each device may be different, but this also does not raise different questions of safety and effectiveness as the sensors (PG, Temperature, EDA) acquire the same general raw signal data.

The subject device's wearable sensor is a smart ring worn on the patient's finger whereas the predicate device is a wrist worn smart watch. Both devices measure their respective physiological signals with appropriate accuracy and as a result, this difference in anatomical site for acquiring raw signal data does not raise different questions of safety and effectiveness. The subject device also does not include a screen on the wearable sensor whereas the predicate device does. All relevant information is available on other screens accessible to patient (mobile application) and clinician (Physician data viewer), and as a result, this difference does not raise different questions of safety and effectiveness.

On this basis, the subject and predicate device can be considered substantially equivalent.

Non-Clinical and/or Clinical Tests Summary & Conclusions

21 CFR 807.92(b)

The following non-clinical tests were performed on the Health Monitoring System.

{6}------------------------------------------------

  • Electrical, mechanical and thermal safety testing in accordance with IEC 60601-1 and IEC 60601-1-11.

  • Electromagnetic Compatibility testing in accordance with IEC 60601-1-2.

  • Wireless coexistence testing in accordance with FDA's quidance: Radio Frequency Wireless Technology in Medical Devices.

  • Usability testing in accordance with IEC 62366 and FDA's guidance: Applying Human Factors and Usability Engineering to Medical Devices.

  • SpO2 and Pulse Rate bench testing in accordance with ISO 80601-2-61.

  • Temperature measurement accuracy testing in accordance with the relevant sections of ISO 80601-2-56.

  • Bench testing to verify the performance of EDA and accelerometer sensors.

  • Software V&V testing and Cybersecurity. Documentation was provided as recommended by FDA's guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, and Content of Premarket Submissions for Management of Cybersecurity in Medical Devices.

The following clinical tests were performed on the Happy Ring Health Monitoring System.

  • Pulse Oximetry: blood oxygen saturation data from 12 subjects of varying BMI, skin tone, and ring size, all wearing the subject device were compared against oxyhemoqlobin saturation using a radial arterial line, in accordance with ISO 80601-2-61. The data demonstrates that the subject device's blood oxygen saturation measurements was within 3.5% Arms for the range of oxygen saturation measured by the device. No adverse events were reported during the trial.

Based on the information presented in this 510(k) premarket notification, device performance and safety evaluated in clinical testing, and comparison with the legally marketed predicate device, the Happy Ring Health Monitoring Platform has been shown to be substantially equivalent to the legally marketed predicate device.

§ 870.2300 Cardiac monitor (including cardiotachometer and rate alarm).

(a)
Identification. A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.(b)
Classification. Class II (performance standards).