Search Filters

Search Results

Found 65 results

510(k) Data Aggregation

    K Number
    K252313

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2026-01-14

    (173 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. The gloves was tested for use with Chemotherapy Drugs and Fentanyl Citrate as per ASTM D6978, "Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs".

    The following drugs have been tested with these gloves:

    Chemotherapy Drug & ConcentrationMinimum Breakthrough Detection Time in Minutes, 0.01 μg/cm2/min:
    Bendamustine HCl, 5 mg/ml (5,000 ppm)>240 min
    Bleomycin Sulfate, 15 mg/ml (15,000 ppm)>240 min
    Busulfan, 6 mg/ml (6,000 ppm)>240 min
    Carboplatin, 10 mg/ml (10,000 ppm)>240 min
    Carmustine 3.3 mg/ml (3,300 ppm)10.0 min
    Chloroquine, 50 mg/ml (50,000 ppm)>240 min
    Cisplatin, 1mg/ml (1000 ppm)>240 min
    Cyclophosphamide 20mg/ml (20,000 ppm)>240 min
    Cytarabine HCl (Cytosine), 100 mg/ml (100,000 ppm)>240 min
    Dacarbazine (DTIC), 10 mg/ml (10,000 ppm)>240 min
    Doxorubicin HCL, 2 mg/ml (2,000 ppm)>240 min
    Epirubicin HCl, 2 mg/ml (2,000 ppm)>240 min
    Etoposide, 20 mg/ml (20,000 ppm)>240 min
    Fludarabine, 25 mg/ml (25,000 ppm)>240 min
    Fluorouracil (5 Flu), 50mg/ml (50,000ppm)>240 min
    Gemcitabine HCl, 38 mg/ml (38,000 ppm)>240 min
    Idarubicin HCl, 1 mg/ml (1,000 ppm)>240 min
    Ifosfamide (IFEX), 50 mg/ml (50,000 ppm)80.0 min
    Mechlorethamine HCl, 1 mg/ml (1,000 ppm)>240 min
    Melphalan HCl, 5 mg/ml (5,000 ppm)>240 min
    Methotrexate, 25mg/ml (25,000ppm)>240 min
    Mitomycin C, 0.5mg/ml (500 ppm)>240 min
    Mitoxantrone HCl (2.0 mg/mL) (2000 ppm)>240 min
    Oxaliplatin, 5 mg/ml (5,000 ppm)>240 min
    Paclitaxel, 6mg/ml (6,000ppm)>240 min
    Rituximab, 10 mg/ml (10,000 ppm)>240 min
    Thiotepa, l0mg/ml (10,000ppm)20.0 min
    Trisenox (Arsenic Trioxide), 1 mg/ml (1,000 ppm)>240 min
    Vincristine sulfate, 1mg/ml (1000 ppm)>240 min
    Vinorelbine, 10 mg/ml (10,00 ppm)>240 min
    Opioid DrugMinimum Breakthrough Detection Time in Minutes, 0.01 μg/cm2/min:
    Fentanyl Citrate Injection, 100mcg/2mg>240 min

    *Please note that the following drugs have low permeation times:
    Carmustine: 10.0 minutes, Thiotepa: 20.0 minutes
    Warning: Do not use with Carmustine & Thiotepa.
    Caution: Testing showed breakthrough time of 80.0 minutes with Ifosfamide IFEX (50.0 mg/mL).

    Device Description

    Not Found

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K243959

    Validate with FDA (Live)

    Date Cleared
    2025-02-24

    (63 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. Gloves have been tested for use with chemotherapy drugs, Fentanyl Citrate and Other Select Drugs using ASTM D6978-05(2019)

    Device Description

    Powder Free Nitrile Examination Gloves (Violet, Black, Green), Tested for Use with Chemotherapy Drugs, Fentanyl Citrate and Select Other Drugs Are Class I Patient Examination Gloves and Specialty Chemotherapy Gloves. They are ambidextrous and come in different sizes - Extra Small, Medium, Large, Extra Large and XXL, Gloves meet the specification of ASTM D6319 and have been tested for resistance to permeation by chemotherapy drugs, Fentanyl Citrate and Select Other Drugs as per ASTM D6978. The gloves are single use, disposable, and non-sterile.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Powder Free Nitrile Examination Gloves:

    1. Table of Acceptance Criteria and Reported Device Performance:

    MethodologyPurposeAcceptance CriteriaReported Device Performance
    ASTM D6319- 19To determine the length of the glovesXS: ≥ 220 mm; S: ≥ 220 mm; M: ≥ 230 mm; L: ≥ 230 mm; XL: ≥ 230 mm; XXL: ≥ 230 mmPass
    ASTM D6319- 19To determine the Physical Dimensions (Palm Width)XS: 70±10mm; S: 80±10mm; M: 95±10mm; L:110±10mm; XL:120±10mm; XXL:130±10mmPass
    ASTM D6319- 19To determine the Physical Dimensions (Thickness)Finger: 0.05mm (min); Palm: 0.05mm (min)Pass
    ASTM D6319- 19, ASTM D412-16(2021)To determine the Physical PropertiesTensile Strength (Min 14 MPa) and Elongation (Before Aging 500% and after aging 400%) MinPass
    ASTM D6319- 19, ASTM D6124-06 (2022)To determine the Powder ResidueMax 2mg/glovePass
    ISO 10993 Part 10-2021Skin Sensitization TestingUnder the conditions of the study, the device is not a sensitizerUnder the conditions of this protocol, the test article did not elicit a sensitization response.
    ISO 10993 Part 23-2021Skin irritation TestingUnder the conditions of the study, the device is not an irritantThe test result showed that the requirements of the ISO Intracutaneous Reactivity Test have been met by the test article.
    ISO 10993-5:2009Cytotoxicity testingNo Cytotoxicity reactivityThe test article is considered cytotoxic under the conditions of this test. Cytotoxicity concern was addressed by Acute Systemic Injection Test.
    ISO 10993-11:2017Acute systemic toxicity studyNo systemic toxicityThe test result indicates that the requirements of the ISO Acute Systemic Injection Test have been met by the test article.
    ASTM D6978-05 (Reapproved 2019)Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy DrugsMinimum Breakthrough Detection Time for various chemotherapy drugs, Fentanyl Citrate, and other select drugs (specific values listed in the "Indications for Use" and "Chemotherapy Permeation, Fentanyl Citrate, and Select Other Drugs Comparison Claim" tables).See detailed breakthrough times in original document, with most being >240 minutes, except for Carmustine and Thiotepa which have lower breakthrough times (e.g., Carmustine 11.7-15.3 min, Thiotepa 15.4-36.4 min).

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not explicitly state the sample sizes used for each specific test (e.g., number of gloves tested for length, width, thickness, etc.). However, the tests were conducted according to established ASTM and ISO standards, which typically specify appropriate sample sizes for such evaluations.

    • Provenance: This information is not directly stated. The tests were conducted according to international standards (ASTM and ISO). The manufacturing company is Lingshi Hongruida Health Protection Technology Co., Ltd. located in Jinzhong, Shanxi, China. It is highly probable that the testing was performed in a laboratory either in China or an accredited facility capable of performing these international standard test methods. The study appears to be retrospective as it involves the testing of a finished device against established performance criteria.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

    This information is not applicable. The device is a physical product (nitrile gloves), and its performance is evaluated against objective, measurable criteria defined by established engineering and biological safety standards (ASTM and ISO), not subjective expert interpretation of medical images or patient data. The "ground truth" is derived from direct measurements and laboratory analyses, not expert consensus.

    4. Adjudication Method for the Test Set:

    This information is not applicable. As the testing involved objective physical and chemical performance standards, there was no need for expert adjudication. The results were quantifiable measurements against pre-defined acceptance criteria.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

    No, an MRMC comparative effectiveness study was not done. This type of study is typically used for diagnostic or screening devices involving human interpretation of medical data (e.g., radiologists reading images). This product is a medical glove, and its effectiveness is determined through standardized physical, chemical, and biological testing.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Yes, the studies conducted are standalone performance evaluations of the physical glove itself, without any human-in-the-loop component in the evaluation of its core performance characteristics (e.g., tensile strength, resistance to chemical permeation, cytotoxicity). The testing directly measured the attributes of the glove.

    7. The Type of Ground Truth Used:

    The ground truth for this device's performance evaluation is based on established objective technical standards and laboratory measurements. Specifically:

    • Physical properties (length, width, thickness, tensile strength, elongation) are measured directly according to ASTM D6319-19 and ASTM D412-16(2021).
    • Chemical resistance (chemotherapy drug permeation) is measured according to ASTM D6978-05 (Reapproved 2019).
    • Biocompatibility (skin sensitization, irritation, cytotoxicity, systemic toxicity) is evaluated through laboratory tests following ISO 10993 series standards.
    • Freedom from holes is evaluated by ASTM D5151-19.
    • Powder residue is evaluated by ASTM D6124-06 (Reapproved 2022).

    8. The Sample Size for the Training Set:

    This information is not applicable. This device is a physical product, not an AI/ML algorithm. Therefore, there is no "training set" in the context of machine learning. The manufacturing process and quality control would involve statistical sampling typical for medical device production.

    9. How the Ground Truth for the Training Set was Established:

    This information is not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K221157

    Validate with FDA (Live)

    Device Name
    JR Medic
    Date Cleared
    2022-08-30

    (131 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    JR Medic Blue Latex Examination Powder Free Gloves Tested for use with Chemotherapy drugs are disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. Additionally, the gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 (2019) Standard Practice for Assessment of Medical Glove to Permeation by Chemotherapy Drugs.

    The tested chemotherapy drugs and their breakthrough detection times are as follows:

    Tested Chemotherapy Drug Name & ConcentrationMinimum Breakthrough Detection Time
    Carmustine (BCNU) (3.3 mg/ml)23.8 Minutes
    Carboplatin (10 mg/ml)>240 Minutes
    Cisplatin (1 mg/ml)>240 Minutes
    Cyclophosphamide (Cytoxan) (20 mg/ml)>240 Minutes
    Dacarbazine (10.0 mg/ml)>240 Minutes
    Doxorubicin HCl (2 mg/ml)>240 Minutes
    Etoposide (20 mg/ml)>240 Minutes
    Fluorouracil (50 mg/ml)>240 Minutes
    Ifosfamide (50 mg/ml)>240 Minutes
    Methotrexate (25 mg/ml)>240 Minutes
    Mitomycin C (0.5 mg/ml)>240 Minutes
    Mitoxantrone (2 mg/ml)>240 Minutes
    Paclitaxel (6 mg/ml)>240 Minutes
    Thiotepa (10 mg/ml)24.1 Minutes
    Vincristine Sulfate (1 mg/ml)>240 Minutes

    Please note that the following drugs have low permeation times: Carmustine (BCNU) (3.3 mg/ml) 23.8 Minutes Thiotepa (10 mg/ml) 24.1 Minutes

    Warning: Do not use with Carmustine or Thiotepa.

    Device Description

    The subject device in 510(K) notification K221157 is JR Medic Blue Latex Examination Powder Free Gloves tested for use with Chemotherapy drugs. The subject device is a patient examination glove made from Natural Rubber Latex, blue color, powder free and non-sterile (as per 21CFR 880.6250, class I).

    The subject device meets all the current specifications listed under the ASTM Specification D3578-2019, Standard Specification for Rubber Examination Gloves for Medical Application. This device also complies with requirements for standard practice for assessment of medical gloves to permeation by chemotherapy drugs as per ASTM D6978- 05(2019).

    AI/ML Overview

    The provided document describes the safety and performance testing of "JR Medic Blue Latex Examination Powder Free Gloves Tested for use with Chemotherapy drugs" to demonstrate substantial equivalence to a predicate device, rather than a study proving the device meets general acceptance criteria for a new medical device technology like an AI algorithm.

    Therefore, the requested information elements related to AI/algorithm performance, such as sample size for test set with provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, training set details, and how ground truth was established for the training set are not applicable to this document.

    The document focuses on the physical, chemical, and biological properties of examining gloves and their resistance to chemotherapy drugs.

    However, I can extract the acceptance criteria and reported device performance from the provided "G. SUMMARY OF NON-CLINICAL PERFORMANCE DATA" section, focusing on the glove's performance criteria.


    1. Table of Acceptance Criteria and the Reported Device Performance

    Test MethodPurposeAcceptance CriteriaReported Device Performance
    ASTM D3578-2019 Standard Specification for Rubber Examination GlovesTo determine the length of the glovesMin 230 mm for all sizesSmall: 304 mm, Medium: 304 mm, Large: 305 mm, X-Large: 305 mm
    ASTM D3578-2019 Standard Specification for Rubber Examination GlovesTo determine the width of the glovesSmall: 80+/-10mm, Medium: 95+/-10 mm, Large: 111+/-10 mm, X-Large: 115+/-10 mmSmall: 84 mm, Medium: 94 mm, Large: 105 mm, X-Large: 114 mm
    ASTM D3578-2019 Standard Specification for Rubber Examination GlovesTo determine the thickness of the glovesPalm: 0.08 mm min, Finger: 0.08 mm min for all sizesPalm: 0.31mm (all sizes), Finger: 0.38mm (all sizes)
    ASTM D3578-2019 Standard Specification for Rubber Examination GlovesTo Determine the physical properties - Tensile strength (Before Ageing)18Mpa Min for all sizesSmall: 33.16 Mpa, Medium: 32.93 Mpa, Large: 32.43 Mpa, X-Large: 32.36 Mpa
    ASTM D3578-2019 Standard Specification for Rubber Examination GlovesTo Determine the physical properties - Tensile strength (After Ageing)14Mpa Min for all sizesSmall: 30.78 Mpa, Medium: 30.73 Mpa, Large: 30.23 Mpa, X-Large: 30.16 Mpa
    ASTM D3578-2019 Standard Specification for Rubber Examination GlovesTo Determine the physical properties - Ultimate Elongation (Before Ageing)650% Min for all sizesSmall: 1320%, Medium: 1288%, Large: 1376%, X-Large: 1200%
    ASTM D3578-2019 Standard Specification for Rubber Examination GlovesTo Determine the physical properties - Ultimate Elongation (After Ageing)500% Min for all sizesSmall: 1066%, Medium: 1105%, Large: 1225%, X-Large: 1041%
    ASTM D3578-2019 Standard Specification for Rubber Examination GlovesTo Determine the physical properties - Stress at 500% Elongation (Before Ageing)5.5 Mpa Max for all sizesSmall: 5.1 Mpa, Medium: 5.2 Mpa, Large: 5.2 Mpa, X-Large: 5.2 Mpa
    ASTM D5151-2019 Standard Test Method for Detection of Holes in Medical GlovesTo determine the holes in the glovesAQL 2.5Gloves Passes AQL 1.5
    ASTM D6124-06 (Reapproved 2017) Standard Test Method for Residual PowderTo determine the residual powder in the gloves2 Mg/Glove MaxSmall: 0.21 mg/glove, Medium: 0.22 mg/glove, Large: 0.22 mg/glove, X-Large: 0.22 mg/glove
    ASTM D 5712-95 (Reapproved 2010) Standard Test Method for the Analysis of ProteinTo determine the extractable protein in the gloves.200 µg/ dm² Max for all sizesSmall: 43.12 µg/ dm², Medium: 41.16 µg/ dm², Large: 42.06 µg/ dm², X-Large: 46.24 µg/ dm²
    ISO 10993-10 Biological Evaluation of Medical Devices: Irritation TestTo determine the potential of the material under test to produce dermal irritation.Under the condition of study not an irritantUnder the condition of study not an irritant
    ISO 10993-10 Biological Evaluation of Medical Devices: Skin Sensitization TestTo determine the skin sensitization potential of the material.Under the conditions of the study not a sensitizerUnder the conditions of the study not a sensitizer
    ISO 10993-5:2009 Biological Evaluation of Medical Devices: Cytotoxicity TestTo evaluate the in vitro cytotoxic potential of the test item.Under the conditions of study non cytotoxicUnder the conditions of the study cytotoxic.
    Acute systemic toxicityTo determine the acute systemic toxicity potential of the test item.No specific acceptance criteria stated in table, but implication is to show no systemic toxicityUnder the conditions of the study, no systemic toxicity
    Material Mediated Pyrogenicity ISO 10993-11:2017(E) / USP 41<151>To determine the pyrogenic potential of the test item extract.Under the conditions of the study non pyrogenicUnder the conditions of the study non pyrogenic
    ASTM D6978-05 (2019) Standard Practice for Assessment of Medical Glove to Permeation by Chemotherapy DrugsTo assess breakthrough time for various chemotherapy drugs.Varies per drug, generally aiming for longer breakthrough times. Specific minimums were not explicitly stated as "acceptance criteria" but rather as "minimum breakthrough detection time" in the Indications for Use.Carmustine (BCNU) (3.3 mg/ml): 23.8 MinutesCarboplatin (10 mg/ml): >240 MinutesCisplatin (1 mg/ml): >240 MinutesCyclophosphamide (Cytoxan) (20 mg/ml): >240 MinutesDacarbazine (10.0 mg/ml): >240 MinutesDoxorubicin HCl (2 mg/ml): >240 MinutesEtoposide (20 mg/ml): >240 MinutesFluorouracil (50 mg/ml): >240 MinutesIfosfamide (50 mg/ml): >240 MinutesMethotrexate (25 mg/ml): >240 MinutesMitomycin C (0.5 mg/ml): >240 MinutesMitoxantrone (2 mg/ml): >240 MinutesPaclitaxel (6 mg/ml): >240 MinutesThiotepa (10 mg/ml): 24.1 MinutesVincristine Sulfate (1 mg/ml): >240 Minutes

    Non-Applicable Information for this Document:

    The following requested information is not applicable to the provided document as it describes the FDA 510(k) clearance for examination gloves and their resistance to chemotherapy drugs, which is a physical device, not an AI/algorithm-based diagnostic or therapeutic tool.

    1. Sample size used for the test set and the data provenance: Not applicable - tests are on gloves, not a data set for an algorithm.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable - "ground truth" concept does not apply in this context; instead, it's about objective physical/chemical measurements and biological responses.
    7. The sample size for the training set: Not applicable.
    8. How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K222449

    Validate with FDA (Live)

    Date Cleared
    2022-08-27

    (12 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. Gloves have been tested for use with chemotherapy drugs and Fentanyl Citrate using ASTM D6978-05(2019)

    Device Description

    Nitrile Powder Free Examination Gloves, Tested For Use With Chemotherapy Drugs And Fentanyl Citrate (Dark Blue) Are Class I Patient Examination Gloves and Specialty Chemotherapy Gloves. They are ambidextrous and come in different sizes - Extra Small, Medium, Large, Extra Large and XXL. Gloves meet the specification of ASTM D6319-19 and have been tested for resistance to permeation by chemotherapy drugs and Fentanyl Citrate as per ASTM D6978-05(2019). The gloves are single use, disposable, and provided non-sterile.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Nitrile Powder Free Examination Gloves, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    MethodologyTest PerformedAcceptance CriteriaReported Device Performance
    ASTM D6319-19Physical Dimensions LengthMinimum 230mm for all sizesPass
    ASTM D6319-19Physical Dimensions Palm Width (XS, S, M, L, XL, XXL)XS: 70±10mm, S: 80±10mm, M: 95±10mm, L:110±10mm, XL: 120±10mm, XXL: 130±10mmPass
    ASTM D6319-19Physical Dimensions Thickness (Finger, Palm)Finger: 0.05mm (min), Palm: 0.05mm (min)Pass
    ASTM D6319-19, ASTM D412-16(2021)Physical Properties (Tensile Strength, Elongation)Tensile Strength (Min 14 MPa) and Elongation (Before Aging 500% and after aging 400%) Min AQL 2.5 (ISO 2859-1)Pass
    ASTM D6319-19, ASTM D5151-19Water leak testAQL 2.5 (ISO 2859-1)Pass
    ASTM D6319-19, ASTM D6124-06 (2017)Powder ResidueMax 2mg/glovePass
    ASTM D6978-05 (2019)Permeation by Chemotherapy DrugsRefer to the table below for specific drug BDTsPass (Based on individual BDTs)
    ISO 10993-10:2010Irritation and Skin SensitizationNo Skin sensitization and Skin irritationIs non-sensitization and Non-irritation
    ISO 10993-5:2009CytotoxicityNo Cytotoxicity reactivityshowed potential toxicity to L929 cells. (Note: Cytotoxicity concern was addressed by acute systematic toxicity testing to still meet acceptance)
    ISO 10993-11:2017Acute systemic toxicity studySubject showed no adverse biological reactionno evidence of systemic toxicity

    Chemotherapy Drug and Fentanyl Citrate Permeation (Breakthrough Detection Time - BDT)

    Chemotherapy DrugMinimum Breakthrough Detection Time (BDT) in Minutes (Acceptance Criteria)Reported Device Performance (BDT in Minutes)
    Carmustine 3.3 mg/ml (3,300 ppm)Explicit criteria not listed, but compared to predicate and results are 11.2 min, with a caution against use.11.2
    Cisplatin 1mg/ml (1,000 ppm)>240 (Implicitly, as predicate is >240 and proposed is same)>240
    Cyclophosphamide 20mg/ml (20,000 ppm)>240 (Implicitly)>240
    Dacarbazine 10 mg/ml (10,000 ppm)>240 (Implicitly)>240
    Doxorubicin HCL, 2 mg/ml (2,000 ppm)>240 (Implicitly)>240
    Etoposide, 20 mg/ml (20,000 ppm)>240 (Implicitly)>240
    Fluorouracil, 50mg/ml (50,000ppm)>240 (Implicitly)>240
    Methotrexate, 25mg/ml (25,000ppm)>240 (Implicitly)>240
    Paclitaxel, 6mg/ml (6,000ppm)>240 (Implicitly)>240
    Thiotepa, 10mg/ml (10,000ppm)Explicit criteria not listed, but compared to predicate and results are 29.4 min, with a caution against use.29.4
    Fentanyl Citrate Injection (100 mcg/2ml)>240 (Implicitly)>240

    Note on Permeation Acceptance: For chemotherapy drugs where the predicate device showed ">240 minutes", the proposed device also achieving ">240 minutes" indicates it meets or exceeds the previous performance. For Carmustine and Thiotepa, while the BDT is lower, the acceptance is ultimately related to the explicit warning not to use them with these drugs, indicating that the observed breakthrough times are treated as insufficient for safe handling.

    2. Sample size used for the test set and the data provenance

    The document does not specify the exact sample sizes used for each individual test against the acceptance criteria (e.g., number of gloves tested for each physical dimension, water leak, or permeation). However, it implies standardized testing according to relevant ASTM and ISO standards. The data provenance is industrial testing, most likely performed in a lab setting, to comply with the specified ASTM and ISO standards.

    • Data Provenance: The device manufacturer, Better Care Plastic Technology Co., Ltd. (located in Shenze County, Hebei, China), would have overseen these tests, either internally or through a qualified testing laboratory. The data is thus of industrial origin, specifically from the manufacturer's testing or a third-party lab they used.
    • Retrospective/Prospective: These are likely prospective tests conducted specifically for this 510(k) submission to demonstrate compliance with the standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable to this type of device and study. The "ground truth" for examination gloves is established by objective, standardized laboratory tests (e.g., measuring physical dimensions, tensile strength, breakthrough detection time for chemicals, biological reactivity as per ISO standards). It does not involve human expert consensus on interpretations of images or clinical outcomes.

    4. Adjudication method for the test set

    This is not applicable. Adjudication methods like "2+1" or "3+1" are typically used in clinical studies or image-based diagnostic device evaluations where expert opinions need to be reconciled. For the physical and chemical testing of gloves, the results are quantitative measurements or direct observations (e.g., pass/fail for a water leak test) and do not require expert adjudication in the same sense.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. An MRMC study assesses the performance of human readers (e.g., radiologists, pathologists) in interpreting medical data, often with and without AI assistance, across a range of cases. This device (examination gloves) is a physical product and does not involve human readers or AI assistance in its primary function, nor in the testing performed for its clearance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable. This question pertains to AI/algorithm performance. The device is a physical product, not an algorithm.

    7. The type of ground truth used

    The "ground truth" for the performance of these gloves is defined by:

    • Standardized Test Methods and Specifications: Specific values and criteria (e.g., minimum tensile strength, maximum powder residue, minimum breakthrough detection time) defined by national and international standards like ASTM D6319-19, ASTM D6978-05, ISO 10993 series.
    • Direct Measurement and Observation: The device's physical properties and chemical resistance are directly measured and observed in laboratory settings according to these established protocols.
    • Biological Reactivity Endpoints: For biocompatibility tests (irritation, sensitization, cytotoxicity, systemic toxicity), the "ground truth" is determined by established biological endpoints and criteria within the ISO 10993 standards.

    8. The sample size for the training set

    This is not applicable. This question refers to the training data for an AI algorithm. This device is a physical medical product, and no AI algorithm development is described or relevant to its clearance.

    9. How the ground truth for the training set was established

    This is not applicable for the reasons stated in point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K220545

    Validate with FDA (Live)

    Date Cleared
    2022-05-26

    (90 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
    In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05, Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs:
    Carmustine (BCNU) (3.3 mg/ml) has a minimum breakthrough time of 23.4 minute and Thiotepa (10.0 mg/ml) has a minimum breakthrough time of 77.5 minute. Warning statements has been included.
    WARNING: DO NOT USE WITH CARMUSTINE WARNING: DO NOT USE WITH THIOTEPA

    Device Description

    Powder-Free Nitrile Examination Gloves, Blue, Tested for Use With Chemotherapy Drugs

    AI/ML Overview

    The provided text is a 510(k) premarket notification letter from the FDA regarding "Powder Free Nitrile Examination Gloves, Blue, Tested for use with Chemotherapy Drugs." This document outlines the FDA's substantial equivalence determination for this medical device.

    It does not contain information about acceptance criteria and a study proving a device meets those criteria in the context of an AI-powered diagnostic or similar system.

    Instead, the document details the "Indications for Use" for the gloves, which includes their general purpose as examination gloves and their tested performance against specific chemotherapy drugs according to ASTM D6978-05.

    Therefore, I cannot extract the requested information (acceptance criteria table, sample sizes, expert qualifications, adjudication, MRMC study, standalone performance, ground truth types, or training set details) from the provided text, as these are not relevant to this type of medical device submission.

    The table in the document you provided describes the "Minimum Breakthrough Detection Time (Minutes)" for different chemotherapy drugs when tested with the gloves. This is a performance specification for the gloves regarding chemical permeation, not acceptance criteria for an AI product.

    Ask a Question

    Ask a specific question about this device

    K Number
    K220491

    Validate with FDA (Live)

    Date Cleared
    2022-05-14

    (81 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This device is intended for medical purpose that is worn on the examination between patient and examiner. In addition, these examination gloves were tested for use with the following chemotherapy drug concentrations per ASTN D6978-05(2019):

    The following drugs had NO breakthrough detected up to 240 minutes:
    Cisplatin, 1.0 mg/ml (1,000 ppm)
    Cyclophosphamide (Cytoxan),20.0 mg/ml (20,000 ppm)
    Dacarbazine, 10.0 mg/ml (10,000 ppm)
    Doxorubicin HCI, 2.0 mg/ml (2,000 ppm)
    Etoposide, 20.0 mg/ml (20,000 ppm)
    Fluorouracil, 50.0 mg/ml (50,000 ppm)
    Paclitaxel, 6.0 mg/ml (6,000 ppm)

    Please note that the following drugs have low permeation times of less than 240 minutes: Carmustine (BCNU) 3.3 mg/ml (3,300 ppm), 34.4minutes ThioTepa10.0 mg/ml (10,000 ppm),88.5 minutes

    WARNING: Do not use with Carmustine and Thiotepa

    Device Description

    Disposable Medical Nitrile Examination Gloves -Tested for Use with Chemotherapy Drugs

    AI/ML Overview

    This document describes the FDA's clearance of "Disposable Medical Nitrile Examination Gloves - Tested for Use with Chemotherapy Drugs" (K220491). This is a medical device, specifically a non-powdered patient examination glove, and the information provided details its intended use and testing against chemotherapy drugs.

    This section does not contain information about an AI/ML device or its performance criteria, study design, or ground truth establishment. The document is a clearance letter for a physical medical device (gloves) and focuses on their chemical resistance properties.

    Therefore, I cannot provide the requested information about acceptance criteria, study details, sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, or ground truth for an AI/ML device based on this input.

    Ask a Question

    Ask a specific question about this device

    K Number
    K220353

    Validate with FDA (Live)

    Date Cleared
    2022-03-31

    (51 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This device is intended for medical purpose that is worn on the examination between patient and examiner. In addition, these gloves were tested for use with the following chemotherapy drug concentrations per ASTM D6978-05(2019): The following drugs had NO breakthrough detected up to 240 minutes:
    Cisplatin, 1.0 mg/ml (1,000 ppm) >240 min.
    Cyclophosphamide (Cytoxan),20.0 mg/ml (20,000 ppm) >240 min.
    Dacarbazine,10.0 mg/ml (10,000 ppm) >240 min.
    Doxorubicin HCI, 2.0 mg/ml (2,000 ppm) >240 min.
    Etoposide, 20.0 mg/ml (20,000 ppm) >240 min.
    Fluorouracil,50.0 mg/ml (50,000 ppm) >240 min.
    Paclitaxel,6.0 mg/ml (6,000 ppm) >240 min.
    Carmustine (BCNU) 3.3 mg/ml (3,300 ppm), 38.7 min.
    ThioTepa10.0 mg/ml (10,000 ppm), 56.9 min
    Please note that the following drug has low permeation times of less than 60 minutes:
    Carmustine (BCNU) 3.3 mg/ml (3,300 ppm), 38.7minutes
    ThioTepa10.0 mg/ml (10.000 ppm),56.9 minutes
    WARNING: Do not use Carmustine and Thiotepa

    Device Description

    Not Found

    AI/ML Overview

    This is an FDA 510(k) clearance letter for Disposable Medical Nitrile Examination Gloves -Tested for Use with Chemotherapy Drugs. It's a Class I reserved device and, as such, the provided information does not describe an AI/ML device but rather a physical device (gloves). Therefore, many of the requested criteria regarding AI/ML study design are not applicable.

    Here's the relevant information that can be extracted from the provided text, focusing on the device's performance regarding chemotherapy drug resistance:

    Acceptance Criteria and Device Performance for Chemotherapy Drug Permeation

    The acceptance criteria are implicitly defined by the test results indicating "NO breakthrough detected up to 240 minutes" for most listed chemotherapy drugs, in accordance with ASTM D6978-05(2019). For drugs with breakthrough times below 240 minutes, those times are reported, along with a warning for those under 60 minutes.

    Acceptance Criteria (Target Breakthrough Time per ASTM D6978-05(2019))Reported Device Performance (Minimum Breakthrough Detection Time)
    No breakthrough for at least 240 minutesCisplatin, 1.0 mg/ml (1,000 ppm): >240 min.
    No breakthrough for at least 240 minutesCyclophosphamide (Cytoxan), 20.0 mg/ml (20,000 ppm): >240 min.
    No breakthrough for at least 240 minutesDacarbazine, 10.0 mg/ml (10,000 ppm): >240 min.
    No breakthrough for at least 240 minutesDoxorubicin HCI, 2.0 mg/ml (2,000 ppm): >240 min.
    No breakthrough for at least 240 minutesEtoposide, 20.0 mg/ml (20,000 ppm): >240 min.
    No breakthrough for at least 240 minutesFluorouracil, 50.0 mg/ml (50,000 ppm): >240 min.
    No breakthrough for at least 240 minutesPaclitaxel, 6.0 mg/ml (6,000 ppm): >240 min.
    Breakthrough time reportedCarmustine (BCNU) 3.3 mg/ml (3,300 ppm): 38.7 min.
    Breakthrough time reportedThioTepa 10.0 mg/ml (10,000 ppm): 56.9 min.

    Important Note: The document explicitly warns against using the gloves with Carmustine and ThioTepa due to their low permeation times (less than 60 minutes).


    The following points are not applicable to this device as it is a physical medical device (examination gloves) and not an AI/ML powered diagnostic or therapeutic device. The requested information pertains to studies typically conducted for AI/ML performance evaluation.

    1. Sample size used for the test set and the data provenance: Not applicable. This refers to physical material testing, not data analysis.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Material testing to a standard like ASTM D6978-05(2019) involves laboratory procedures, not expert consensus on medical images or diagnoses.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): The "ground truth" used here is the physical barrier performance of the glove against specific chemical permeation, determined according to the standardized laboratory methods described in ASTM D6978-05(2019). This is a scientific, objective measurement, not based on expert consensus or clinical outcomes for diagnosis.
    7. The sample size for the training set: Not applicable. This concept is for AI/ML models.
    8. How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K213910

    Validate with FDA (Live)

    Date Cleared
    2022-03-10

    (85 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K213325

    Validate with FDA (Live)

    Date Cleared
    2022-01-28

    (115 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This device is intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. In addition, these gloves were tested for use with the following chemotherapy drug concentrations per ASTM D6978-05(2019): The following drugs had NO breakthrough detected up to 240 minutes: Cisplatin, 1.0 mg/ml (1,000 ppm); Cyclophosphamide (Cytoxan), 20.0 mg/ml (20,000 ppm); Dacarbazine, 10.0 mg/ml (10,000 ppm); Doxorubicin HCI, 2.0 mg/ml (2,000 ppm); Etoposide, 20.0 mg/ml (20,000 ppm); Fluorouracil, 50.0 mg/ml (50,000 ppm); Paclitaxel, 6.0 mg/ml (6,000 ppm). Please note that the following drug has low permeation times of less than 60 minutes: Carmustine (BCNU) 3.3 mg/ml (3.300 ppm), 23.9minutes; ThioTepa10.0 mg/ml (10,000 ppm),25.3 minutes. WARNING: Do not use Carmustine and Thiotepa

    Device Description

    Disposable Medical Nitrile Examination Gloves -Tested for Use with Chemotherapy Drugs

    AI/ML Overview

    This document is a 510(k) premarket notification decision letter from the FDA for medical examination gloves. It does not contain any information about an AI/ML device or a study proving its performance. Therefore, I cannot provide the requested information about acceptance criteria and a study proving device performance in the context of an AI/ML medical device.

    The document discusses:

    • The clearance of "Disposable Medical Nitrile Examination Gloves - Tested for Use With Chemotherapy Drugs"
    • The regulation number and product codes
    • General controls provisions and other regulatory requirements for medical devices
    • Indications for Use for the gloves, including specific chemotherapy drugs they were tested against and their breakthrough times.

    There is no mention of acceptance criteria, study designs, sample sizes, expert involvement, ground truth, or any other elements related to the evaluation of an AI/ML-driven medical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K213076

    Validate with FDA (Live)

    Device Name
    AspenMed+
    Date Cleared
    2021-12-20

    (88 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Blue Nitrile Examination Gloves Powder Free tested for use with Chemotherapy drugs is a disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. Additionally, the gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 (2019) Standard Practice for Assessment of Medical Glove to Permeation by Chemotherapy Drugs.

    Device Description

    The subject device in 510(K) notification is a blue nitrile examination gloves powder free tested for use with Chemotherapy drugs. The subject device is a patient examination glove made from acrylonitrile-butadiene copolymer dispersion, blue color, powder free and non sterile (as per 21CFR 880.6250, class I). The subject device meets all the current specifications listed under the ASTM Specification D 6319 -2019, Standard Specification for Nitrile Examination Gloves for Medical Application. This device also complies with requirements for standard practice for assessment of resistance of medical gloves to permeation by chemotherapy drugs as per ASTM D6978- 05(2019)

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification for "Blue Nitrile Examination Gloves Powder Free Tested for Use with Chemotherapy Drugs." This document clears a medical device (examination gloves) for market and is not an AI/ML device study. Therefore, most of the requested information about acceptance criteria and study design for AI/ML performance is not applicable within this context.

    However, I can extract information related to the performance of these physical medical gloves as detailed in the provided document, specifically related to their physical properties and chemical resistance.

    Here's the relevant information based on the document:

    Study Purpose: To demonstrate that the "Blue Nitrile Examination Gloves Powder Free tested for use with Chemotherapy Drugs" are as safe, effective, and perform as well as the legally marketed predicate device (Harbour Health Powder Free Nitrile Examination Glove, Blue (Tested for use with Chemotherapy Drugs) K210944).

    Lack of AI/ML Study: This document does not describe a study involving an AI/ML device. It pertains to a physical medical device (examination gloves). Therefore, specific criteria and study details related to AI/ML device performance (such as sample size for test/training sets for AI, data provenance for AI, number of experts for AI ground truth, human reader improvement with AI assistance, standalone AI performance) are not present in this document and cannot be provided.

    Acceptance Criteria and Device Performance (for Physical Device - Gloves):

    The performance of the gloves was evaluated against ASTM standards for various physical properties and for resistance to chemotherapy drugs. The acceptance criteria and reported device performance are presented in the tables below.

    Table 1: Physical Properties Acceptance Criteria and Device Performance

    Test Method (Standard)PurposeAcceptance CriteriaReported Device Performance
    ASTM D6319-2019To determine the length of the glovesMin 220 mm for Size Small & Min 230 mm for all other sizesSmall: 245 mm; Medium: 246 mm; Large: 248 mm; X-Large: 248 mm
    ASTM D6319-2019To determine the width of the glovesSmall: 80+/-10 mm; Medium: 95+/-10 mm; Large: 110+/-10 mm; X-Large: 120+/-10 mmSmall: 80 mm; Medium: 92 mm; Large: 105 mm; X-Large: 115 mm
    ASTM D6319-2019To determine the thickness of the glovesPalm 0.05 mm min; Finger 0.05 mm min for all sizesPalm: 0.09 mm (all sizes); Finger: 0.15mm (all sizes)
    ASTM D6319-2019To Determine the physical properties - Tensile strengthBefore Ageing: Tensile Strength 14Mpa Min for all sizes; After Ageing: Tensile Strength 14Mpa Min for all sizesBefore Ageing: >18.52 Mpa (all sizes); After Ageing: >17.54 Mpa (all sizes)
    ASTM D6319-2019To Determine the physical properties - Ultimate ElongationBefore Ageing: Ultimate Elongation 500% Min for all sizes; After Ageing: Ultimate Elongation 400% Min for all sizesBefore Ageing: >690% (all sizes); After Ageing: >665% (all sizes)
    ASTM D5151-2019To determine the holes in the glovesAQL 2.5Gloves Passes AQL 1.5
    ASTM D6124-06 (Reapproved 2017)To determine the residual powder in the gloves2 Mg/Glove Max<0.22 mg/glove (all sizes)

    Table 2: Biocompatibility Testing Acceptance Criteria and Device Performance

    Test Method (Standard)PurposeAcceptance CriteriaReported Device Performance
    ISO 10993-10 (Irritation)To determine the potential of the material to produce dermal irritation in RabbitsUnder the condition of study not an irritantUnder the condition of study not an irritant
    ISO 10993-10 (Skin Sensitization)To determine the skin sensitization potential of the material in Guinea Pig.Under the conditions of the study not a sensitizerUnder the conditions of the study not a sensitizer
    ISO 10993-5:2009 (In vitro cytotoxicity)To evaluate the in vitro cytotoxic potential of the test item extracts in L-929 mouse fibroblasts cells.Under the conditions of study non cytotoxicUnder the conditions of the study cytotoxic. (Note: This indicates it did not meet the "non cytotoxic" criteria, but was still cleared due to other compensating factors or the nature of the cytotoxicity test that might be acceptable for the device's use)
    ISO 10993-11:2017 (Acute systemic toxicity)To determine the acute systemic toxicity potential of the test item extracts in Swiss Albino mice.Under the conditions of study the device extracts do not pose a systemic toxicity concernUnder the conditions of study the device extracts do not pose a systemic toxicity concern
    ISO 10993-11:2017 / USP 41<151> (Material Mediated Pyrogenicity)To determine the pyrogenic potential of the test item extract following intravenous injection in New Zealand white Rabbits.Under the conditions of the study, the device did not demonstrate a material mediated pyrogenicity response.Under the conditions of the study, the device did not demonstrate a material mediated pyrogenicity response.

    Table 3: Chemotherapy Drug Permeation Acceptance Criteria and Device Performance

    (Note: The "acceptance criteria" here is implied by the predicate device's performance for comparison, and the requirement to report the minimum breakthrough detection time. A longer time generally indicates better performance.)

    Tested Chemotherapy Drug Name & ConcentrationPredicate Device Minimum Breakthrough Detection TimeProposed Device Minimum Breakthrough Detection Time
    Carmustine (BCNU) (3.3 mg/ml)14.5 Minutes34.3 Minutes
    Carboplatin (10 mg/ml)>240 Minutes>240 Minutes
    Cisplatin (1 mg/ml)>240 Minutes>240 Minutes
    Cyclophosphamide (Cytoxan) (20 mg/ml)>240 Minutes>240 Minutes
    Dacarbazine (10.0 mg/ml)>240 Minutes>240 Minutes
    Doxorubicin HCl (2 mg/ml)>240 Minutes>240 Minutes
    Etoposide (20 mg/ml)>240 Minutes>240 Minutes
    Fluorouracil (50 mg/ml)>240 Minutes>240 Minutes
    Ifosfamide (50 mg/ml)>240 Minutes>240 Minutes
    Methotrexate (25 mg/ml)>240 Minutes>240 Minutes
    Mitomycin C (0.5 mg/ml)>240 Minutes>240 Minutes
    Mitoxantrone (2 mg/ml)>240 Minutes>240 Minutes
    Paclitaxel (6 mg/ml)>240 Minutes>240 Minutes
    Thiotepa (10 mg/ml)47.4 Minutes87.3 Minutes
    Vincristine Sulfate (1 mg/ml)>240 Minutes>240 Minutes
    Drugs tested by Predicate but NOT Proposed:Busulfan (6mg/ml) (>240 min), Cytarabine (100 mg/ ml) (>240 min), Docetaxel (10 mg/ml) (>240 min), Mechlorethamine HCL (1 mg/ml) (>240 min)Not Tested
    Note on warnings:Warning: Do not use with Carmustine (BCNU) & Thiotepa

    2. Sample Size Used for the Test Set and Data Provenance:

    • The document primarily refers to testing done in accordance with ASTM standards. These standards specify sample sizes for testing (e.g., for AQL testing for pinholes, or for physical properties). The exact number of samples for each specific test (e.g., how many gloves for tensile strength or how many for chemical permeation) is not explicitly stated in the summary, but it would be prescribed by the referenced ASTM standards (e.g., ASTM D6319-2019, ASTM D5151-2019, ASTM D6124-06, ASTM D6978-05).
    • Data Provenance: The applicant for the device is Aspen Glove Sdn. Bhd., located in Malaysia. The testing would presumably have been conducted there or by a certified lab, but specific country of origin for the data (i.e., where the tests were physically performed) is not explicitly stated beyond the applicant's location. The testing would be prospective in nature as it evaluates the manufactured product.

    3. Number of Experts Used to Establish Ground Truth and Qualifications:

    • This is not applicable as this is a physical medical device and not an AI/ML device requiring establishment of "ground truth" by clinical experts for image interpretation or similar tasks. The "ground truth" for the physical properties and chemical resistance is established by validated laboratory testing methods conforming to recognized international standards (ASTM, ISO).

    4. Adjudication Method for the Test Set:

    • Not applicable. This is not a study requiring adjudication of expert interpretations but rather laboratory testing against established physical and chemical standards.

    5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study:

    • Not applicable. This is a physical medical device, not an AI/ML device that assists human readers.

    6. Standalone (Algorithm Only) Performance:

    • Not applicable. This is a physical medical device, not an algorithm.

    7. Type of Ground Truth Used:

    • For physical and biocompatibility properties: Ground truth is established by validated laboratory testing methods against recognized industry standards (ASTM, ISO) with numerical or pass/fail criteria.
    • For chemotherapy drug permeation: Ground truth is established by laboratory testing per ASTM D6978-05 (2019), measuring the minimum breakthrough detection time.

    8. Sample Size for the Training Set:

    • Not applicable. This is not an AI/ML device requiring a training set.

    9. How the Ground Truth for the Training Set Was Established:

    • Not applicable. This is not an AI/ML device.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 7