Search Results
Found 167 results
510(k) Data Aggregation
(197 days)
The Intellidrop system is indicated for use to provide external drainage of cerebrospinal fluid (CSF) and/or monitoring of CSF drainage and intracranial pressure (ICP) for ventricular or lumbar use.
The Intellidrop system continuously monitors intracranial or spinal pressure and manages external drainage of cerebrospinal fluid (CSF) and other body fluids to a user-selected pressure target or volume target.
Key features of the Intellidrop include:
- Wearable sensor array facilitating position-agnostic pressure monitoring
- Closed-loop, gravity-based drainage to a pressure or volume target
- Real-time display of intracranial pressure (ICP) waveform
- Historical trend data of daily, hourly, and minute-by-minute ICP and drainage volume
- User selectable audio-visual alarm thresholds for both pressure and volume
- Patented Continuous Drift Detection Technology
- Continuous operation on AC power and up to 4 hours of continuous operation on rechargeable battery power
- Real-time data streaming of ICP to a patient monitor via cable
- Proximal sampling port and optional distal sampling port
N/A
Ask a specific question about this device
(270 days)
The Sophy Mini Monopressure Valve is designed for the treatment of hydrocephalus by shunting cerebrospinal fluid (CSF) to the abdominal cavity or right atrium of the heart.
The Sophy® Mini Monopressure Valve is intended for the treatment of hydrocephalus by shunting the cerebrospinal fluid (CSF) to the abdominal cavity or right atrium of the heart.
The complete CSF shunting system provides shunting of CSF from the ventricular cavities in the brain to another part of the body where the CSF can be absorbed in blood.
At the minimum, a complete CSF shunt consists of:
- A proximal catheter to drain CSF from either lateral brain ventricles;
- A distal catheter to drain CSF to a body cavity where it can be absorbed in blood (most often, the peritoneal cavity or the right atrium of the heart so respectively called peritoneal or atrial catheter also);
- Between the two, a valve which regulates the flow of CSF.
N/A
Ask a specific question about this device
(86 days)
Phasor EVAC-MRI is indicated when access to and evacuation of a cranial subacute or chronic hematoma or hygroma is necessary. The Phasor EVAC-MRI system is intended for drainage of subdural fluid accumulations such as hygromas and chronic or subacute hematomas to an external suction reservoir. The Phasor EVAC–MRI system is also intended for draining air and fluids from the subdural space immediately following craniotomy procedures performed to remove a chronic or subacute subdural hematoma.
Phasor EVAC-MRI™ subdural evacuation system is indicated when access to the subdural space and evacuation of a cranial subacute or chronic hematoma or hygroma is necessary. The Phasor EVAC-MRI™ system consists of surgical instruments and accessories used for draining subdural fluid accumulations such as hygromas and liquid-state subdural hematomas to an external suction reservoir without touching the brain. Utilizing a minimally invasive technique, the Phasor EVAC-MRI™ system components are designed to promote gradual brain re-expansion by creating a low homogeneous negative pressure throughout the subdural space as fluid is drained to an external suction reservoir. The completeness of the Phasor EVAC-MRI™ system saves time and eliminates the need to gather supplies; this convenience can be critical in an emergency situation.
N/A
Ask a specific question about this device
(267 days)
For use in the treatment of hydrocephalus as a component of a shunt system when draining or shunting of cerebrospinal fluid (CSF) is indicated.
The Libertís™ Shunt Catheter includes a ventricular and distal (peritoneal) drainage catheter that are used as part of a cerebrospinal fluid (CSF) shunting system to treat hydrocephalus. Both catheters are attached to the valve portion of a shunting system. The ventricular catheter diverts excess CSF from the ventricles of the brain through the valve. After passing through the valve, the excess CSF is drained through the distal catheter into another part of the body, such as the peritoneal cavity, where it is reabsorbed into the bloodstream. The catheters have Endexo® polymer, a surface modifying macromolecule, blended into their base silicone. The catheters are subjected to a treatment process by which the silicone is impregnated with two antimicrobials, rifampicin and clindamycin hydrochloride. The Libertís™ Shunt Catheter has been shown in laboratory studies to reduce the colonization of gram-positive bacteria on the tubing surface. The catheters contain barium sulfate for radiopacity and have ink markings on the silicone tubing to aid in positioning of the catheter.
The catheters are packaged with two accessories, a stylet and a right-angle adapter. The stylet is used to help introduce the ventricular catheter into the brain's ventricles. The right-angle adapter is used to fixate the ventricular catheter to the cranium.
This document is a 510(k) clearance letter for a medical device, specifically a shunt catheter. It does not describe an AI medical device or software. Therefore, most of the requested information regarding AI acceptance criteria, training/test sets, expert adjudication, MRMC studies, or standalone algorithm performance cannot be extracted from the provided text.
The document focuses on demonstrating substantial equivalence to a predicate device through non-clinical bench testing and biocompatibility testing, rather than performance studies of an AI algorithm.
However, I can provide the acceptance criteria and performance information that is present in the document, which pertains to the physical and biological characteristics of the shunt catheter.
Acceptance Criteria and Device Performance (for the physical device, not an AI algorithm):
Since the provided document is for a medical device (shunt catheter) and not an AI/software device, the questions related to AI-specific criteria (e.g., sample size for test/training sets of AI, expert adjudication, MRMC studies, standalone algorithm performance, type of ground truth for AI) are not applicable.
The acceptance criteria here refer to the performance of the physical device as demonstrated through various bench and biocompatibility tests. The document states that all tests were "Pass," implying that the device met the pre-defined acceptance criteria for each test.
1. Table of Acceptance Criteria and Reported Device Performance
| Test | Acceptance Criteria (Implied by "Pass") | Reported Device Performance |
|---|---|---|
| Evaluation of Safety and Performance per ISO 7197 and ASTM F647-94 | Complies with the safety and performance requirements of these standards. | Pass |
| Functional/Mechanical Performance per ASTM F647-94 and ASTM D412-16 | Complies with the functional and mechanical performance requirements of these standards. | Pass |
| MRI Compatibility per ASTM F2053-20 | Meets MRI compatibility standards (specifically, "MR-safe" designation). | Pass |
| Antimicrobial Efficacy – Zone of Inhibition (ZOI) per internal methods | Demonstrates adequate zone of inhibition as per internal methods. | Pass |
| Antibiotic Content Analysis using High-performance Liquid Chromatography (HPLC), per internal methods | Meets specified antibiotic content levels as per internal methods. | Pass |
| Residual Solvents Testing via Headspace Gas Chromatography – Flame Ionization Detection (HS GC-FID), per FDA guidance "Q3C – Tables and List Guidance for Industry" | Meets specified limits for residual solvents. | Pass |
| Usability Evaluation and Design Validation per IEC 62366-1 | Complies with usability and design validation requirements of IEC 62366-1. | Pass |
| In vitro flow loop thrombogenicity test | Observed significantly less thrombus deposition compared to the predicate device. | Observed significantly less thrombus deposition on Codman Libertis Distal Shunt Catheter with Bactiseal and Endexo Technology compared to predicate device |
| Cytotoxicity per ISO 10993-5:2009 | No cytotoxic effects. | Pass |
| Sensitization per ISO 10993-10:2021 | No sensitization reactions. | Pass |
| Irritation per ISO 10993-10:2021 and ISO 10993-23:2021 | No irritation. | Pass |
| Acute Systemic Toxicity per ISO 10993-11:2017 | No acute systemic toxicity. | Pass |
| Pyrogenicity per ISO 10993-11:2017 | No pyrogenic effects. | Pass |
| Subacute/Subchronic Toxicity per ISO 10993-6:2016 and ISO 10993-11:2017 | No subacute/subchronic toxicity. | Pass |
| Genotoxicity per ISO 10993-3:2014 | No genotoxic effects. | Pass |
| Implantation per ISO 10993-6:2016 | Acceptable tissue response upon implantation. | Pass |
| Chronic Toxicity per ISO 10993-11:2017 | No chronic toxicity. | Pass |
| Hemocompatibility per ISO 10993-4:2017 | Hemocompatible. | Pass |
| Carcinogenicity per ISO 10993-3:2014 | No carcinogenic effects. | Pass |
| Sterilization Validation (E-beam) per ISO 11137-1, ISO 11137-2, ISO 11737-1, and ISO 11737-2 | Achieves a sterility assurance level (SAL) of 10-6. | Achieved SAL of 10-6 |
| Shelf Life | At least 1 year. | 1 year |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: The document does not specify general "sample sizes" in terms of patient data or typical "test sets" as one would for an AI model. Instead, it refers to "production equivalent devices" used for the various bench and biocompatibility tests. The specific number of devices or replicates for each test is not detailed.
- Data Provenance: The data is generated from laboratory studies and bench testing performed on the physical device. This is not clinical patient data. The provenance for this type of testing is typically the manufacturer's internal testing facilities or contract labs. No country of origin for data or retrospective/prospective nature is applicable as this is not patient data.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
- Not applicable. This pertains to an AI/software device. The "ground truth" for a physical device is established by objective measurements and standardized test methods, not expert consensus on data interpretation.
4. Adjudication Method (e.g. 2+1, 3+1, none) for the Test Set
- Not applicable. This pertains to an AI/software device.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
- No, not applicable. This pertains to an AI/software device evaluating human reader performance.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done
- No, not applicable. This pertains to an AI algorithm; the device is a physical shunt catheter.
7. The Type of Ground Truth Used
- For the performance bench tests and biocompatibility tests, the "ground truth" is defined by established international (ISO, ASTM) and FDA-recognized consensus standards, as well as pre-defined internal test methods and specifications. For instance, a "Pass" for biocompatibility means the device met the criteria specified in the relevant ISO 10993 standards. For thrombogenicity, the ground truth was a statistically significant reduction in thrombus deposition compared to the predicate device.
8. The Sample Size for the Training Set
- Not applicable. This pertains to an AI/software device. No training set is involved for validating a physical medical device.
9. How the Ground Truth for the Training Set was Established
- Not applicable. This pertains to an AI/software device.
Ask a specific question about this device
(264 days)
The Codman Libertís™ EVD Catheter with Bactiseal® and Endexo® Technology is indicated for gaining access to the ventricles of the brain and can be used with dimensionally compatible devices for draining cerebrospinal fluid (CSF) and other fluids of similar physical characteristics as a means of reducing intracranial pressure and CSF volume.
The Codman LibertísTM EVD Catheter with Bactiseal® and Endexo® Technology (LibertísTM EVD Catheters) include a ventricular catheter that is supplied with component accessories that facilitate placement and use of the catheter for reducing and controlling intracranial pressure due to excess cerebrospinal fluid (CSF). The ventricular catheter is subjected to a treatment process by which the silicone is impregnated with two antimicrobials, rifampicin and clindamycin hydrochloride. Laboratory studies show Bactiseal treated catheters reduce the colonization of gram-positive bacteria on the tubing surface. Additionally, LibertísTM EVD Catheters contain an Endexo® polymer additive; a surface modifying macromolecule (SMM) polymer blended into the catheter's base silicone. The ventricular catheter is placed in the ventricles of the brain and CSF enters the fluid conduit through the inlet holes near the tip of the catheter and drains into the external drainage system connected to the catheter. The catheter contains a barium sulfate stripe for radiopacity and includes numerical depth markings and circumferential bands, made of ink, from the proximal tip.
The provided document is a 510(k) clearance letter from the FDA for a medical device (Codman Libertís™ EVD Catheter). It describes the device, its intended use, a comparison to a predicate device, and summaries of nonclinical testing.
However, the question asks for details related to acceptance criteria and studies that prove a device meets acceptance criteria, specifically those typically associated with AI/software medical devices. This document does not describe an AI/software medical device; it describes a physical medical device (a catheter for draining CSF). Therefore, much of the information requested in the prompt, such as "number of experts used to establish ground truth," "adjudication method," "multi-reader multi-case (MRMC) study," "standalone performance," "training set size," and "how ground truth for training set was established," are not applicable to this type of device and are not present in the provided text.
The document focuses on demonstrating substantial equivalence to a predicate device through bench testing, biocompatibility, and sterilization validation, which are standard for physical medical devices.
Here's an attempt to answer the applicable parts of your request based on the provided text, and explicitly state where information is not available or not applicable.
Description of Acceptance Criteria and Proving Device Meets Criteria
The Codman Libertís™ EVD Catheter is a physical medical device. Its acceptance criteria are primarily demonstrated through a battery of nonclinical performance bench tests, biocompatibility testing, and sterilization validation, to show that it is substantially equivalent to a legally marketed predicate device. The performance criteria are generally considered a "Pass" or "Fail" based on established standards and internal methods.
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly list quantitative acceptance criteria values but rather states a "Pass" or "Fail" for each test. The acceptance criterion for each test is implicitly that the device performs as expected according to the specified standard or internal method.
| Test | Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|---|
| Evaluation of Codman Libertís EVD Catheters per ASTM F647 and ISO 7197 | Meets specifications of ASTM F647 and ISO 7197 for EVD catheters. | Pass |
| Evaluation of Aged Codman Libertís EVD Catheters per ASTM F647 and ISO 7197 (T = 12 months) | Meets specifications of ASTM F647 and ISO 7197 for EVD catheters after 12 months simulated aging. | Pass |
| Functional/Mechanical Performance of Codman Libertís EVD Catheters | Meets predefined functional and mechanical performance specifications. | Pass |
| Functional/Mechanical Performance of Aged Codman Libertís EVD Catheters (T = 12 months) | Meets predefined functional and mechanical performance specifications after 12 months simulated aging. | Pass |
| MRI Compatibility Assessment of Codman Libertís EVD Catheters per ASTM F2503 | Meets specifications of ASTM F2503 for MRI compatibility (Note: Catheter is MR Conditional, Trocar and Stylet are MR-Unsafe, as noted in labeling). | Pass |
| Antimicrobial Efficacy of Codman Libertís EVD Catheters – Zone of Inhibition (ZOI) after 28 Days of Simulated Use | Demonstrates sufficient Zone of Inhibition against specified bacteria after 28 days of simulated use to show antimicrobial efficacy. | Pass |
| Antimicrobial Efficacy of Aged Codman Libertís EVD Catheters – Zone of Inhibition (ZOI) after 28 Days of Simulated Use (T = 12 months) | Demonstrates sufficient Zone of Inhibition after 12 months simulated aging and 28 days of simulated use. | Pass |
| Antibiotic Content Testing of Codman Libertís EVD Catheters | Contains the specified levels of antibiotics (rifampicin and clindamycin hydrochloride). | Pass |
| Residual Solvents Testing of Codman Libertís EVD Catheters | Levels of residual solvents are within acceptable limits. | Pass |
| Libertís EVD Catheters with Bactiseal® and Endexo® Technology – Summative Usability Study | Meets user-related safety and effectiveness requirements, indicating the device can be used safely and effectively by its intended users. | Pass |
| Product Shelf Life Testing per ASTM F1980 and ISO 11607-1 | Maintains integrity and performance over its stated shelf life (12 months) as per ASTM F1980 and ISO 11607-1. | Pass |
| Biocompatibility Testing per ISO 10993-1 | Demonstrates biocompatibility in accordance with ISO 10993-1 for medical devices in contact with the body. | Pass |
| Sterilization Validation (Electron Beam) | Achieves a sterility assurance level (SAL) of 10-6. | Pass |
2. Sample size used for the test set and the data provenance
The document states, "All testing was performed on production equivalent devices." However, it does not specify the sample size for any of the individual bench tests.
Data provenance (e.g., country of origin of the data, retrospective or prospective) is not applicable in the traditional sense for these bench tests, which are conducted in a laboratory setting to evaluate physical properties and performance characteristics.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This concept is not applicable to this type of medical device (physical catheter). "Ground truth" in this context would refer to established engineering standards, material specifications, and validated test methods, rather than expert interpretation of medical images or clinical outcomes.
4. Adjudication method for the test set
This concept is not applicable to this type of medical device. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies involving multiple readers (e.g., radiologists) to resolve disagreements in interpretation, which is not relevant for bench testing of a physical device.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No MRMC comparative effectiveness study was done, as this is a physical medical device (catheter), not an AI/software device. Therefore, this question is not applicable.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This concept is not applicable since the device is a physical catheter, not a software algorithm.
7. The type of ground truth used
The "ground truth" for the performance tests of this physical device consists of:
- Established industry standards (e.g., ASTM F647, ISO 7197, ASTM F2503, ISO 10993-1, ASTM F1980, ISO 11607-1).
- Internal predefined functional and mechanical specifications.
- Validated laboratory methods for chemical content, antimicrobial efficacy (Zone of Inhibition), and sterility assurance (SAL 10-6).
- Usability requirements for the Summative Usability Study.
8. The sample size for the training set
This concept is not applicable as this is a physical medical device, not an AI/software device that would involve a "training set."
9. How the ground truth for the training set was established
This concept is not applicable for the same reason as above.
Ask a specific question about this device
(267 days)
The Sophysa external CSF drainage catheters are intended to temporarily drain the Cerebrospinal Fluid (CSF), for less than 30 days (up to 29 days maximum).
The Sophysa external CSF drainage catheters are intended to temporarily drain the Cerebrospinal Fluid (CSF), for less than 30 days.
There are two types of external CSF drainage catheters:
- Ventricular catheters
- Lumbar catheters
The catheter is provided sterile and with accessories to facilitate the surgical procedures and connectors. The external CSF drainage catheters need to be connected to collection systems.
The provided FDA 510(k) clearance letter and summary discuss the "External CSF Drainage" device. However, this document does not contain information related to a study proving the device meets specific performance acceptance criteria in the context of medical imaging or AI/algorithm performance.
The document primarily focuses on demonstrating substantial equivalence to predicate devices through technical, clinical, and biological safety comparisons, along with non-clinical bench testing. It does not describe a study involving human readers, AI assistance, ground truth establishment, or specific diagnostic performance metrics (like sensitivity, specificity, AUC) that would typically be associated with performance acceptance criteria for an AI or imaging device.
Therefore, I cannot fulfill the request to provide:
- A table of acceptance criteria and reported device performance specific to diagnostic accuracy or AI performance.
- Sample size used for the test set and data provenance.
- Number of experts used to establish ground truth and their qualifications.
- Adjudication method.
- Multi-reader multi-case (MRMC) comparative effectiveness study results.
- Standalone algorithm performance.
- Type of ground truth used (expert consensus, pathology, outcomes data).
- Sample size for the training set.
- How ground truth for the training set was established.
The document details bench testing performed, which validates physical and mechanical properties of the device, but not its performance in a clinical diagnostic or AI-assisted setting.
The information available regarding acceptance criteria and testing from the provided document is as follows:
Acceptance Criteria (Implicit from Bench Testing):
| Performance Standard/Characteristic | Acceptance Criteria (Implicit from Tests) | Reported Device Performance (Summary) |
|---|---|---|
| Visual Inspection | (Not explicitly stated, but implies meeting design specifications) | Performed (Summary of Non-clinical testing) |
| Dimensional Analysis | (Not explicitly stated, but implies meeting design specifications) | Performed (Summary of Non-clinical testing) |
| Radiopacity | (Not explicitly stated, but implies visible under imaging) | Performed (Summary of Non-clinical testing) |
| Pressure / Flow Characteristics | (Not explicitly stated, but implies adequate CSF drainage) | Performed (Summary of Non-clinical testing) |
| Blockage / Hemorrhagic CSF Exposure | (Not explicitly stated, but implies resistance to blockage/degradation) | Performed (Summary of Non-clinical testing) |
| Air and Water Tightness | (Not explicitly stated, but implies no leakage) | Performed (Summary of Non-clinical testing) |
| Dynamic Breaking Strength | (Not explicitly stated, but implies resistance to fracture under use) | Performed (Summary of Non-clinical testing) |
| Catheter Tensile Strength | (Not explicitly stated, but implies resistance to pulling forces) | Performed (Summary of Non-clinical testing) |
| Kink Resistance | (Not explicitly stated, but implies resistance to kinking) | Performed (Summary of Non-clinical testing) |
| Catheter Fixation | (Not explicitly stated, but implies secure attachment) | Performed (Summary of Non-clinical testing) |
| Corrosion Resistance | (Not explicitly stated, but implies material integrity) | Performed (Summary of Non-clinical testing) |
| Guidewire Tensile Strength | (Not explicitly stated) | Performed (Summary of Non-clinical testing) |
| Guidewire Bending Strength | (Not explicitly stated) | Performed (Summary of Non-clinical testing) |
| Guidewire Breakage | (Not explicitly stated) | Performed (Summary of Non-clinical testing) |
| Tuohy Needle/Hub Connection Strength | (Not explicitly stated) | Performed (Summary of Non-clinical testing) |
| Sterility | Supplied sterile | Supplied sterile (Predicate Comparison Table) |
| Biocompatibility | Biocompatible materials | Biocompatible materials established by testing and rationale for Cytotoxicity, Sensitization, Irritation, Acute Systemic Toxicity, Material-Mediated Pyrogenicity, Subacute/Subchronic Toxicity, Genotoxicity, Implantation (brain, 4 weeks, muscle 4 weeks), Indirect Hemolysis, Neurotoxicity (Predicate Comparison Table) |
| MRI Compatibility (Ventricular) | MR Safe | MR Safe (Predicate Comparison Table) |
| MRI Compatibility (Lumbar) | MR Conditional | MR Conditional due to metallic parts in Compression Hub (Predicate Comparison Table) |
| Shelf life | 5 years | Target shelf life (Predicate Comparison Table) |
| Performance Standards | ISO 7197, ISO 20697, ISO 20698, ISO 11070 | Substantially Equivalent (Predicate Comparison Table) |
Study Details from the document (focused on Substantial Equivalence and Bench Testing):
- Sample size used for the test set and the data provenance: Not applicable for diagnostic performance metrics. Bench testing was performed on "representative samples" of the product line. No information on data provenance (e.g., country of origin, retrospective/prospective) is associated with these bench tests.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for bench tests typically involves direct measurement against engineering specifications.
- Adjudication method: Not applicable for bench testing.
- If a multi-reader multi-case (MRMC) comparative effectiveness study was done: No, not mentioned.
- If a standalone (i.e. algorithm only without human-in-the loop performance) was done: No, this is not an AI/algorithm-driven device for diagnosis.
- The type of ground truth used: For the bench tests, the "ground truth" would be the engineering specifications and established test methodologies (e.g., ISO standards) for physical and material properties. For biocompatibility, established ISO standards and toxicological assessments.
- The sample size for the training set: Not applicable (not an AI/ML device).
- How the ground truth for the training set was established: Not applicable.
In summary, the provided document from the FDA is for a physical medical device (External CSF drainage catheters) and demonstrates its safety and effectiveness through substantial equivalence to existing devices and extensive bench testing of its physical and material properties. It does not describe a study involving AI, image analysis, or human reader performance, and therefore, cannot provide the information requested for such types of studies.
Ask a specific question about this device
(245 days)
Phasor EVAC™ is indicated when access to and evacuation of a cranial subacute or chronic hematoma or hygroma is necessary. The EVAC system is intended for drainage of subdural fluid accumulations such as hygromas and chronic or subacute hematomas to an external suction reservoir. The EVAC system is also intended for draining air and fluids from the subdural space immediately following craniotomy procedures performed to remove a chronic or subacute subdural hematoma.
Phasor EVAC™ subdural evacuation system is indicated when access to the subdural space and evacuation of a cranial subacute or chronic hematoma or hygroma is necessary. The EVAC system consists of surgical instruments and accessories used for draining subdural fluid accumulations such as hygromas and chronic or subacute hematomas to an external suction reservoir without touching the brain. Utilizing a minimally invasive technique, the EVAC system components are designed to promote gradual brain re-expansion by creating a low homogeneous negative pressure throughout the subdural space as fluid is drained to an external suction reservoir.
This 510(k) clearance letter for the EVAC device (K243205) indicates that it is a central nervous system fluid shunt and components and it's regulated under 21 CFR 882.5550. While the letter confirms the device's substantial equivalence to predicate devices and states that various tests were conducted, it does not provide detailed information about specific acceptance criteria, study methodologies (like sample size, ground truth establishment, expert qualifications, or MRMC studies), or actual performance data in the format requested.
The document primarily focuses on explaining why the device is considered substantially equivalent to existing devices based on its intended use, indications for use, and technological characteristics. It mentions that "Various tests including for biocompatibility, packaging, sterility, safety, and performance were conducted and passed successfully," and that "testing for a closed system with firm purchase into bone was established and verified by testing." However, it does not disclose the specific data or criteria for these tests. It explicitly states, "No clinical testing was needed or performed otherwise."
Therefore, based on the provided FDA 510(k) clearance letter, I cannot populate the detailed table and answer all the specific questions about acceptance criteria and study proving device performance as requested, because this information is typically contained in the actual 510(k) submission, not the clearance letter itself. The clearance letter summarizes the FDA's decision, but not the detailed technical data from the submission.
Here's what can be inferred or explicitly stated from the provided document, with notes where information is not present:
Acceptance Criteria and Device Performance (Based on Inferred Safety and Effectiveness)
The acceptance criteria for a 510(k) cleared device are implicitly that it demonstrates substantial equivalence to a predicate device, meaning it is as safe and effective while having the same intended use. For this device, the "performance" described is its ability to meet the design intent and demonstrate equivalence through bench testing, rather than clinical efficacy.
Table 1: Inferred Acceptance Criteria and Reported Device Performance
| Acceptance Criterion (Inferred from 510(k) Basis) | Reported Device Performance (from document) |
|---|---|
| Biocompatibility | "Various tests including for biocompatibility... were conducted and passed successfully." (Specific criteria and results not provided) |
| Packaging Integrity | "Various tests including for... packaging... were conducted and passed successfully." (Specific criteria and results not provided) |
| Sterility | "Various tests including for... sterility... were conducted and passed successfully." (Specific criteria and results not provided) |
| Safety | "Various tests including for... safety... were conducted and passed successfully." (Specific criteria and results not provided) |
| Performance (Functional Equivalence) | "Various tests including for... performance were conducted and passed successfully." "Specifically, testing for a closed system with firm purchase into bone was established and verified by testing." "The performance and design validation testing conducted on the EVAC™ device on the bench demonstrated that it performs equivalent to the stated predicates..." (Specific criteria and results not provided) |
| Manufacturing Quality Systems | Compliance with Quality System (QS) regulation (21 CFR Part 820) and other general controls is required for marketing. (No specific test results provided in letter) |
Study Details (Based on Available Information)
-
Sample size used for the test set and the data provenance:
- The document states "No clinical testing was needed or performed otherwise." This indicates that the "test set" for demonstrating substantial equivalence was primarily based on bench testing and engineering evaluations of the device components and system.
- Specific sample sizes for these bench tests are not provided in the clearance letter.
- Data provenance (country of origin, retrospective/prospective): Not applicable or specified for bench testing.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
- Given that "No clinical testing was needed or performed," there's no mention of expert panels establishing ground truth in a clinical context. The "ground truth" for the engineering tests would be derived from accepted engineering standards and specifications.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not applicable, as no clinical study or human reader evaluation requiring adjudication was performed.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, an MRMC comparative effectiveness study was not done. The device is a surgical system (shunt and components), not an AI-assisted diagnostic tool.
-
If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- This question is more pertinent to software/AI devices. The EVAC system is a mechanical device. Its "performance" refers to its physical functioning (e.g., maintaining a closed system, firm purchase into bone), not an algorithmic output. The "Phasor Drill" component is explicitly stated to have been previously cleared (K161704), indicating its standalone performance was established previously.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc):
- For this device, the "ground truth" for its functional performance was based on engineering specifications, material science standards, and mechanical testing requirements (e.g., ability to form a "closed system with firm purchase into bone"). There's no mention of clinical ground truth (like pathology or outcomes data) because no clinical study was performed.
-
The sample size for the training set:
- Not applicable. This is a hardware device, not an AI/machine learning algorithm requiring a "training set."
-
How the ground truth for the training set was established:
- Not applicable for the same reason as above.
Ask a specific question about this device
(103 days)
The Duet EDMS is indicated for draining of CSF flow from the lateral ventricles or lumbar subarachnoid space in selected patients to:
- · Reduce intracranial pressure (ICP), e.g., pre-, intra- or postoperative.
- · Monitor CSF chemistry, cytology, and physiology.
- · Provide temporary CSF drainage in patients with infected cerebrospinal fluid shunts.
Monitoring of intracranial pressure (ICP) is indicated in selected patients with:
- · Severe head injury
- · Subarachnoid hemorrhage graded III, IV, or V preoperatively
- · Reyes syndrome or similar encephalopathies
- · Hydrocephalus
- · Intracranial hemorrhage
- · Miscellaneous problems when drainage is to be used as a therapeutic maneuver.
Monitoring can also be used to evaluate the status pre- and postoperatively for space-occupying lesions.
The Medtronic Duet™ External Drainage and Monitoring System (Duet™ EDMS) is a complete draining and monitoring cerebrospinal system for externally fluid (CSF) and monitoring intracranial pressure (ICP). It can be used for both external ventricular and lumbar drainage.
The Duet™ EDMS is an external drainage and monitoring system that uses gravity to drain cerebrospinal fluid (CSF) from the patient's ventricles or lumbar space to an external drainage receptacle. The drainage flow of CSF into the Duet™ EDMS is uni-directional and gravity-driven; there is no recirculation of the CSF. The Duet's catheter is surgically attached to the patient before it is attached to the drainage system. An opening is made in the patient's skull lumbar region and a catheter is inserted into patient's ventricle or the lumbar or subarachnoid space. The catheter is then attached to the drainage system. The CSF or blood is drained and monitored until the patient is stabilized, the infection successfully treated, or a long-term drainage method is implemented.
The provided document is a 510(k) Premarket Notification from the FDA for a medical device called the "Duet External Drainage and Monitoring System (EDMS)". It describes the device, its intended use, and its comparison to a predicate device. However, it does not contain information about an AI/ML-based device or a study involving human readers or ground truth established by experts/pathology.
The performance testing summarized in the document focuses on the physical and functional characteristics of the Duet EDMS (e.g., dimensional accuracy, torque strength, leak integrity, flow initiation pressure, and attachment strength) rather than diagnostic or prognostic accuracy of an AI model.
Therefore, many of the requested points, such as sample size for test/training sets, data provenance, number/qualifications of experts, adjudication methods, MRMC studies, standalone algorithm performance, and ground truth establishment for AI models, are not applicable and cannot be extracted from this document.
If you are looking for an example of an AI/ML-based medical device submission, this document is not suitable.
Despite the irrelevance of the document to AI/ML, I will attempt to frame the acceptance criteria and study as presented for this non-AI device:
Acceptance Criteria and Device Performance for Duet External Drainage and Monitoring System (EDMS)
The Duet External Drainage and Monitoring System (EDMS) is a non-AI/ML medical device. Its performance testing focused on demonstrating its physical and functional suitability for its intended use, rather than diagnostic accuracy or AI model performance.
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are implicitly defined by the "Pass" results for each test. The study demonstrated that the device "passed" all these checks, indicating it met the pre-defined performance requirements for each test.
| Test Name | Acceptance Criteria (Implicit from "Pass" Result) | Reported Device Performance |
|---|---|---|
| Dimensional | Vertical/horizontal alignment of scale label, pressure scale lengths, patient line tubing ID/length, cord length, drainage path ID, and stopcock flow path diameter must meet specified dimensions. | Pass |
| Drip Chamber Graduations | Drip chamber graduations must show correct readings. | Pass |
| Main System Stopcock (MSS) Assembly Torque Applied to Arm of Stopcock | MSS assembly must withstand a specified peak torque without failure or detachment. | Pass |
| MSS Assembly Load Applied to Core of Stopcock Arm | MSS assembly must withstand a specified peak load without failure. | Pass |
| Clamp to I.V. Pole Attachment Strength | Clamp must securely attach to the I.V. pole. | Pass |
| Cord to I.V. Pole Attachment Strength | Cord and cord lock must maintain secure hanging of the system. | Pass |
| Drip Chamber to Back Panel Attachment Strength | Drip chamber/bag subassembly must securely attach to the panel. | Pass |
| Strength of Attached Junctions (Tubing to Luer) | Junctions of tubes to luer must be securely attached. | Pass |
| Bottom Cap to Stopcock Junction Torque | Stopcock/bottom cap bond must be secure. | Pass |
| Drip Assembly and Drainage Bag Vent Integrity | Drip assembly and drainage bag vent must withstand appropriate fluid pressures. | Pass |
| Tensile Strength of Drainage Bag Inlet Port | Drainage bag inlet port must exhibit sufficient tensile strength to failure. | Pass |
| Drainage Bag Seal Weld | Drainage bag must have no leaks. | Pass |
| Flow Initiation Pressure | Flow must initiate at a specified pressure for each drainage bag. | Pass |
| Drip Assembly Vent Test (Exposure of Vent to Blood Solution) | Drip assembly vent must allow drainage of blood and CSF flow with minimal resistance. | Pass |
| Drip Assembly Vent Integrity | Drip assembly vent must allow fluid withdrawal without compromising mechanical integrity. | Pass |
| Leakage of UV-Cure Bonds | No leakage from UV-cure bonds between patient line and drip chamber subassembly. | Pass |
| Leakage of Drainage Bag | Drainage bag must withstand inversion without leaking. | Pass |
| Drip Chamber Volume | Drip chamber fluid weight must be verifiable. | Pass |
| Attachment of I.V. Pole and Position of Adjustable Drip Chamber | Clamping thumbscrews and cord locks must not slip from initial positions. | Pass |
| Leakage of UV Cure Bonds (Air Pressure) | UV-cure bonds must withstand air pressure without leaks. | Pass |
| Attachment of Junctions (Axial Load) | Junctions must withstand a minimum 5-pound axial load. | Pass |
| Bottom Cap to Stopcock Junction Torque | Stopcock/bottom cap bond must withstand specified torque. | Pass |
| Hydrophobic Microbial Barrier Vent on the Drainage Bag | Material must demonstrate 99.9% Bacterial Filtration Efficiency (BFE). | Pass |
2. Sample Size and Data Provenance
- Sample Size for Test Set: The document does not specify the quantitative sample size for each individual performance test (e.g., number of devices tested for each parameter). It only states that "Testing demonstrated the performance of the device."
- Data Provenance: The document does not provide details about the country of origin of the data or whether the testing was retrospective or prospective in the context of patient data. The tests described are laboratory-based engineering performance tests conducted on the physical device.
3. Number of Experts and Qualifications for Ground Truth
- Number of Experts: Not applicable. The "ground truth" for these tests relates to engineering specifications and physical measurements, rather than clinical interpretation requiring expert consensus.
- Qualifications of Experts: Not applicable. Testing was likely conducted by engineers or technicians involved in product development and quality assurance, following established protocols.
4. Adjudication Method for the Test Set
- Adjudication Method: Not applicable. The tests involve objective measurements and physical properties, which typically have clear pass/fail criteria based on engineering specifications, not subjective interpretation requiring adjudication.
5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study
- MRMC Study: No, an MRMC comparative effectiveness study was not done. This type of study is relevant for evaluating the clinical performance of diagnostic or AI-assisted interpretation systems, which is not the function of the Duet EDMS.
6. Standalone (Algorithm Only) Performance
- Standalone Performance: No, a standalone performance study was not done. The Duet EDMS is a physical medical device, not a software algorithm.
7. Type of Ground Truth Used
- Type of Ground Truth: The "ground truth" for the performance tests was based on engineering specifications, design requirements, and established physical measurement standards. For example, a dimension must be within a certain tolerance, a torque value must exceed a minimum, or a seal must not leak.
8. Sample Size for the Training Set
- Sample Size for Training Set: Not applicable. This device does not involve machine learning; therefore, there is no "training set."
9. How Ground Truth for the Training Set Was Established
- Ground Truth Establishment for Training Set: Not applicable, as there is no training set for this non-AI device.
Ask a specific question about this device
(160 days)
The XABO Catheters are used for cerebrospinal fluid (CSF) shunting.
The XABO Catheters are manufactured using barium sulfate fillcone elastomer and are impregnated with clindamycin hydrochloride and rifampicin designed to be released over time from the exterior and inner lumen surface once implanted.
The XABO Ventricular Catheters will be offered in 18 cm in length with an inner diameter of 1.2 mm and an outer diameter of 2.5 mm. Lengths are marked in 1 cm intervals starting from the catheter tip, thus enabling the surgeon to qauge the depth of penetration of the catheter into the lateral ventricle. The proximal end of the catheter has 16 flow holes around the catheter circumference.
Components supplied with the XABO Ventricular Catheter include a pre-loaded stainless steel stylet and depending on the confiquration may contain a deflector.
The XABO Peritoneal Catheters measure 60 cm or 120 cm in length, 1.2 mm in inner diameter, and 2.5 mm in outer diameter. There are no length markers or wall slits on the tip is open ended. The catheter may be trimmed to the proper length.
The XABO Catheters are designed to articulate with existing Miethke Shunt Systems, such as the M.blue Adjustable Shunt System. Miethke Shunt System GAV 2.0 and SA 2.0 Valves, proGAV 2.0 Adjustable Shunt System miniNAV valve, and the Miethke Shunt System (DSV, connectors, and reservoirs) cleared by FDA (K192266/K190174/K161853/K141687/K110206/K030698/K011030).
The provided text focuses on the 510(k) premarket notification for the XABO Ventricular Catheter, XABO Peritoneal Catheter, and XABO Catheter Set. It primarily addresses the substantial equivalence of these devices to existing predicate devices. While it mentions performance bench testing and specific tests performed, it does not explicitly state specific acceptance criteria values or detailed study results that would allow for a complete fill of the requested table.
The document indicates that the devices were tested against "performance specifications" and that the results "confirm that the XABO Catheters meet performance specifications." However, the exact numerical or qualitative specifications themselves are not provided.
Therefore, the following information is extracted directly from the provided text, and where information is not explicitly stated, it is noted as "Not explicitly stated in the provided text."
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Performance Bench Testing (ISO 7197) | "Testing results confirm that the XABO Catheters meet performance specifications." (Specific specifications and results not provided). |
| Zone of Inhibition | "Comparison testing of the XABO Catheters and the predicate device show that they have the same characteristics." (Specific characteristics and results not provided). |
| Drug Content | "Comparison testing of the XABO Catheters and the predicate device show that they have the same characteristics." (Specific characteristics and results not provided). |
| Drug Release Kinetic | "Comparison testing of the XABO Catheters and the predicate device show that they have the same characteristics." (Specific characteristics and results not provided). |
| Crush Resistance | "Comparison testing of the XABO Catheters and the predicate device show that they have the same characteristics." (Specific characteristics and results not provided). |
| MRI Safety (ASTM F2213, F2182, F2119) | "the deflector is MR Conditional in 3-Tesla Magnetic Resonance Imaging systems per ASTM F2503." |
| Biocompatibility (ISO 1099-1, FDA Guidance "Use of International Standard ISO 10993-1") | Met requirements for Cytotoxicity, Sensitization/Reactivity, Genotoxicity, Implantation, Acute Systemic Toxicity, Sub-Chronic, Hemolysis, Pyrogenicity. Chronic and developmental/reproductive toxicity addressed by chemical characterization. |
2. Sample size used for the test set and the data provenance
- Sample Size: Not explicitly stated in the provided text. The phrase "All testing was performed on the worst-case final finished device" could imply a limited number of samples, but a specific number is not given.
- Data Provenance: The studies are described as "Performance bench testing" and "Biocompatibility testing," suggesting laboratory-based testing, not human-patient data. The country of origin of the data is not specified. It is laboratory research, not retrospective or prospective patient data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Not applicable as the testing described is bench testing and biocompatibility, not studies requiring expert interpretation of clinical data for ground truth.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
- Not applicable for the types of tests described (bench testing, biocompatibility).
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No, an MRMC comparative effectiveness study was not done. The submission focuses on device equivalence through bench and biocompatibility testing, not AI-assisted human reading.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not applicable. This device is a physical medical device (catheter), not an algorithm or AI system.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- For performance bench testing, the "ground truth" would be the pre-defined engineering specifications and standards (e.g., ISO 7197) that the device must meet.
- For biocompatibility testing, the "ground truth" is compliance with international standards (ISO 1099-1) and FDA guidance, ensuring that the device's biological interactions are within acceptable limits.
8. The sample size for the training set
- Not applicable. This is not an AI/machine learning device that requires a training set.
9. How the ground truth for the training set was established
- Not applicable. This is not an AI/machine learning device that requires a training set and ground truth establishment.
Ask a specific question about this device
(89 days)
The EVAC is indicated to facilitate the evacuation of a chronic or subacute hematoma or hygroma in the cranium. The EVAC is intended for draining of fluid accumulated in the subdural space, including chronic or subacute hematomas and hygromas. The EVAC is also intended for removing air and fluid from the subdural space following open surgical procedures to remove subdural collections.
The Borvo EVAC System is intended for the drainage of subdural fluid accumulations. The EVAC System is designed to gradually drain fluid to an external reservoir by creating a low negative pressure in the subdural space. It consists of 3 components:
-
- EVAC Port (Ergo or Classic)
-
- Tubing
-
- Vacuum bulb
An optional Cranial Access Kit can be purchased with the EVAC System, or a commercially available Cranial Access Kit with a 5.31mm drill bit can be used.
- Vacuum bulb
The provided text is a 510(k) premarket notification letter and summary for the Borvo EVAC System. It describes the device, its intended use, and a comparison to a predicate device. However, it does not include information about acceptance criteria or a study proving that the device meets those criteria, as typically found in clinical performance or software validation reports for AI/ML-driven devices.
The document focuses on demonstrating substantial equivalence to a predicate device based on:
- Same intended use.
- Similar technological characteristics (with detailed comparisons of materials, dimensions, etc.).
- Performance data from non-clinical testing.
Therefore, I cannot provide the requested information about acceptance criteria, device performance against those criteria, sample sizes for test sets, data provenance, expert ground truth establishment, adjudication methods, MRMC studies, standalone performance, or training set details because this information is not present in the provided text.
The "Performance Data" section (Section 1.7 on page 3 of 4) lists the types of non-clinical testing performed, such as:
- Dimensional
- Tensile
- Pull-Out Force
- Simulated Use
- Sterile Barrier/Packaging Testing
- Shelf Life
- Biocompatibility (ISO 10993-1)
These tests are typically used to demonstrate the physical and biological safety and performance of a medical device, but they are not related to an AI/ML algorithm's analytical or clinical performance requiring ground truth, expert review, or statistical performance metrics.
Ask a specific question about this device
Page 1 of 17