K Number
K243531

Validate with FDA (Live)

Date Cleared
2025-08-08

(267 days)

Product Code
Regulation Number
882.5550
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

For use in the treatment of hydrocephalus as a component of a shunt system when draining or shunting of cerebrospinal fluid (CSF) is indicated.

Device Description

The Libertís™ Shunt Catheter includes a ventricular and distal (peritoneal) drainage catheter that are used as part of a cerebrospinal fluid (CSF) shunting system to treat hydrocephalus. Both catheters are attached to the valve portion of a shunting system. The ventricular catheter diverts excess CSF from the ventricles of the brain through the valve. After passing through the valve, the excess CSF is drained through the distal catheter into another part of the body, such as the peritoneal cavity, where it is reabsorbed into the bloodstream. The catheters have Endexo® polymer, a surface modifying macromolecule, blended into their base silicone. The catheters are subjected to a treatment process by which the silicone is impregnated with two antimicrobials, rifampicin and clindamycin hydrochloride. The Libertís™ Shunt Catheter has been shown in laboratory studies to reduce the colonization of gram-positive bacteria on the tubing surface. The catheters contain barium sulfate for radiopacity and have ink markings on the silicone tubing to aid in positioning of the catheter.

The catheters are packaged with two accessories, a stylet and a right-angle adapter. The stylet is used to help introduce the ventricular catheter into the brain's ventricles. The right-angle adapter is used to fixate the ventricular catheter to the cranium.

AI/ML Overview

This document is a 510(k) clearance letter for a medical device, specifically a shunt catheter. It does not describe an AI medical device or software. Therefore, most of the requested information regarding AI acceptance criteria, training/test sets, expert adjudication, MRMC studies, or standalone algorithm performance cannot be extracted from the provided text.

The document focuses on demonstrating substantial equivalence to a predicate device through non-clinical bench testing and biocompatibility testing, rather than performance studies of an AI algorithm.

However, I can provide the acceptance criteria and performance information that is present in the document, which pertains to the physical and biological characteristics of the shunt catheter.


Acceptance Criteria and Device Performance (for the physical device, not an AI algorithm):

Since the provided document is for a medical device (shunt catheter) and not an AI/software device, the questions related to AI-specific criteria (e.g., sample size for test/training sets of AI, expert adjudication, MRMC studies, standalone algorithm performance, type of ground truth for AI) are not applicable.

The acceptance criteria here refer to the performance of the physical device as demonstrated through various bench and biocompatibility tests. The document states that all tests were "Pass," implying that the device met the pre-defined acceptance criteria for each test.

1. Table of Acceptance Criteria and Reported Device Performance

TestAcceptance Criteria (Implied by "Pass")Reported Device Performance
Evaluation of Safety and Performance per ISO 7197 and ASTM F647-94Complies with the safety and performance requirements of these standards.Pass
Functional/Mechanical Performance per ASTM F647-94 and ASTM D412-16Complies with the functional and mechanical performance requirements of these standards.Pass
MRI Compatibility per ASTM F2053-20Meets MRI compatibility standards (specifically, "MR-safe" designation).Pass
Antimicrobial Efficacy – Zone of Inhibition (ZOI) per internal methodsDemonstrates adequate zone of inhibition as per internal methods.Pass
Antibiotic Content Analysis using High-performance Liquid Chromatography (HPLC), per internal methodsMeets specified antibiotic content levels as per internal methods.Pass
Residual Solvents Testing via Headspace Gas Chromatography – Flame Ionization Detection (HS GC-FID), per FDA guidance "Q3C – Tables and List Guidance for Industry"Meets specified limits for residual solvents.Pass
Usability Evaluation and Design Validation per IEC 62366-1Complies with usability and design validation requirements of IEC 62366-1.Pass
In vitro flow loop thrombogenicity testObserved significantly less thrombus deposition compared to the predicate device.Observed significantly less thrombus deposition on Codman Libertis Distal Shunt Catheter with Bactiseal and Endexo Technology compared to predicate device
Cytotoxicity per ISO 10993-5:2009No cytotoxic effects.Pass
Sensitization per ISO 10993-10:2021No sensitization reactions.Pass
Irritation per ISO 10993-10:2021 and ISO 10993-23:2021No irritation.Pass
Acute Systemic Toxicity per ISO 10993-11:2017No acute systemic toxicity.Pass
Pyrogenicity per ISO 10993-11:2017No pyrogenic effects.Pass
Subacute/Subchronic Toxicity per ISO 10993-6:2016 and ISO 10993-11:2017No subacute/subchronic toxicity.Pass
Genotoxicity per ISO 10993-3:2014No genotoxic effects.Pass
Implantation per ISO 10993-6:2016Acceptable tissue response upon implantation.Pass
Chronic Toxicity per ISO 10993-11:2017No chronic toxicity.Pass
Hemocompatibility per ISO 10993-4:2017Hemocompatible.Pass
Carcinogenicity per ISO 10993-3:2014No carcinogenic effects.Pass
Sterilization Validation (E-beam) per ISO 11137-1, ISO 11137-2, ISO 11737-1, and ISO 11737-2Achieves a sterility assurance level (SAL) of 10-6.Achieved SAL of 10-6
Shelf LifeAt least 1 year.1 year

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: The document does not specify general "sample sizes" in terms of patient data or typical "test sets" as one would for an AI model. Instead, it refers to "production equivalent devices" used for the various bench and biocompatibility tests. The specific number of devices or replicates for each test is not detailed.
  • Data Provenance: The data is generated from laboratory studies and bench testing performed on the physical device. This is not clinical patient data. The provenance for this type of testing is typically the manufacturer's internal testing facilities or contract labs. No country of origin for data or retrospective/prospective nature is applicable as this is not patient data.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • Not applicable. This pertains to an AI/software device. The "ground truth" for a physical device is established by objective measurements and standardized test methods, not expert consensus on data interpretation.

4. Adjudication Method (e.g. 2+1, 3+1, none) for the Test Set

  • Not applicable. This pertains to an AI/software device.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

  • No, not applicable. This pertains to an AI/software device evaluating human reader performance.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • No, not applicable. This pertains to an AI algorithm; the device is a physical shunt catheter.

7. The Type of Ground Truth Used

  • For the performance bench tests and biocompatibility tests, the "ground truth" is defined by established international (ISO, ASTM) and FDA-recognized consensus standards, as well as pre-defined internal test methods and specifications. For instance, a "Pass" for biocompatibility means the device met the criteria specified in the relevant ISO 10993 standards. For thrombogenicity, the ground truth was a statistically significant reduction in thrombus deposition compared to the predicate device.

8. The Sample Size for the Training Set

  • Not applicable. This pertains to an AI/software device. No training set is involved for validating a physical medical device.

9. How the Ground Truth for the Training Set was Established

  • Not applicable. This pertains to an AI/software device.

FDA 510(k) Clearance Letter - Codman Libertís Shunt Catheter

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.08.00

August 8, 2025

Integra LifeSciences Production Corporation
David Arroyo
Sr. Regulatory Affairs Specialist
11 Cabot Boulevard
Mansfield, Massachusetts 02048

Re: K243531
Trade/Device Name: Codman Libertís Shunt Catheter with Bactiseal and Endexo Technology
Regulation Number: 21 CFR 882.5550
Regulation Name: Central Nervous System Fluid Shunt And Components
Regulatory Class: Class II
Product Code: JXG
Dated: November 14, 2024
Received: July 11, 2025

Dear David Arroyo:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

Page 2

K243531 - David Arroyo Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Page 3

K243531 - David Arroyo Page 3

Sincerely,

Yen-chih Lin -S Digitally signed by Yen-chih Lin -S Date: 2025.08.08 17:44:25 -04'00'

For
Adam Pierce, Ph.D.
Assistant Director
DHT5A: Division of Neurosurgical, Neurointerventional, and Neurodiagnostic Devices
OHT5: Office of Neurological and Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

Indications for Use

Submission Number (if known)
K243531

Device Name
Codman Libertís Shunt Catheter with Bactiseal and Endexo Technology

Indications for Use (Describe)
For use in the treatment of hydrocephalus as a component of a shunt system when draining or shunting of cerebrospinal fluid (CSF) is indicated.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

Integra® LifeSciences Production Corporation
Codman® Libertís™ Shunt Catheter with Bactiseal® and Endexo® Technology
Page 1 of 4

510(k) Summary

A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92.

807.92(a) (1) Submitter Information

Name and AddressIntegra LifeSciences Production Corporation11 Cabot BoulevardMansfield, MA 02048
Telephone number(609) 627-9053
Primary ContactDavid A. Arroyo
Date Summary PreparedAugust 8, 2025

807.92(a) (2) Name of Device

Trade or Proprietary NameCodman® Libertís™ Shunt Catheter with Bactiseal® and Endexo® Technology
Common NameHydrocephalus Shunt SystemHydrocephalus Catheters
Classification NameShunt, Central Nervous System and Components(21 CFR 882.5550)
Device ClassII
Product CodeJXG

807.92(a) (3) Predicate Information

Predicate DeviceCodman® Bactiseal® Shunt Catheter: K233445

807.92(a) (4) Device Description

The Libertís™ Shunt Catheter includes a ventricular and distal (peritoneal) drainage catheter that are used as part of a cerebrospinal fluid (CSF) shunting system to treat hydrocephalus. Both catheters are attached to the valve portion of a shunting system. The ventricular catheter diverts excess CSF from the ventricles of the brain through the valve. After passing through the valve, the excess CSF is drained through the distal catheter into another part of the body, such as the peritoneal cavity, where it is reabsorbed into the bloodstream. The catheters have Endexo® polymer, a surface modifying macromolecule, blended into their base silicone. The catheters are subjected to a treatment process by which the silicone is impregnated with two antimicrobials, rifampicin and clindamycin hydrochloride. The Libertís™ Shunt Catheter has been shown in laboratory studies to reduce the colonization of gram-positive bacteria on the tubing surface. The catheters contain barium sulfate for radiopacity and have ink markings on the silicone tubing to aid in positioning of the catheter.

The catheters are packaged with two accessories, a stylet and a right-angle adapter. The stylet is used to help introduce the ventricular catheter into the brain's ventricles. The right-angle adapter is used to fixate the ventricular catheter to the cranium.

807.92(a) (5) Indications for Use

For use in the treatment of hydrocephalus as a component of a shunt system when draining or shunting of cerebrospinal fluid (CSF) is indicated.

807.92(a) (6) Technological Characteristics Compared to Predicate

The Libertís™ Shunt Catheter has the same intended use, same functionality, same antibiotics, same radiopaque material, and similar design specifications as the predicate device. Compared to the predicate, the proposed device adds a new material, Endexo®, is made of an equivalent silicone, has markings made of ink instead of tantalum, offers a new length of ventricular catheter and stylet, and uses different packaging materials. The differences in technological characteristics do not raise different questions of safety and effectiveness, compared to the predicate.

Page 6

Integra® LifeSciences Production Corporation
Codman® Libertís™ Shunt Catheter with Bactiseal® and Endexo® Technology

510(k) Summary

A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92.

807.92(a) (1) Submitter Information

Name and AddressIntegra LifeSciences Production Corporation11 Cabot BoulevardMansfield, MA 02048
Telephone number(609) 627-9053
Primary ContactDavid A. Arroyo
Date Summary PreparedAugust 8, 2025

807.92(a) (2) Name of Device

Trade or Proprietary NameCodman® Libertís™ Shunt Catheter with Bactiseal® and Endexo® Technology
Common NameHydrocephalus Shunt SystemHydrocephalus Catheters
Classification NameShunt, Central Nervous System and Components(21 CFR 882.5550)
Device ClassII
Product CodeJXG

807.92(a) (3) Predicate Information

Predicate DeviceCodman® Bactiseal® Shunt Catheter: K233445

807.92(a) (4) Device Description

The Libertís™ Shunt Catheter includes a ventricular and distal (peritoneal) drainage catheter that are used as part of a cerebrospinal fluid (CSF) shunting system to treat hydrocephalus. Both catheters are attached to the valve portion of a shunting system. The ventricular catheter diverts excess CSF from the ventricles of the brain through the valve. After passing through the valve, the excess CSF is drained through the distal catheter into another part of the body, such as the peritoneal cavity, where it is reabsorbed into the bloodstream. The catheters have Endexo® polymer, a surface modifying macromolecule, blended into their base silicone. The catheters are subjected to a treatment process by which the silicone is impregnated with two antimicrobials, rifampicin and clindamycin hydrochloride. The Libertís™ Shunt Catheter has been shown in laboratory studies to reduce the colonization of gram-positive bacteria on the tubing surface. The catheters contain barium sulfate for radiopacity and have ink markings on the silicone tubing to aid in positioning of the catheter.

The catheters are packaged with two accessories, a stylet and a right-angle adapter. The stylet is used to help introduce the ventricular catheter into the brain's ventricles. The right-angle adapter is used to fixate the ventricular catheter to the cranium.

807.92(a) (5) Indications for Use

For use in the treatment of hydrocephalus as a component of a shunt system when draining or shunting of cerebrospinal fluid (CSF) is indicated.

807.92(a) (6) Technological Characteristics Compared to Predicate

The Libertís™ Shunt Catheter has the same intended use, same functionality, same antibiotics, same radiopaque material, and similar design specifications as the predicate device. Compared to the predicate, the proposed device adds a new material, Endexo®, is made of an equivalent silicone, has markings made of ink instead of tantalum, offers a new length of ventricular catheter and stylet, and uses different packaging materials. The differences in technological characteristics do not raise different questions of safety and effectiveness, compared to the predicate.

Page 1 of 4

Component AffectedProposed ModificationRationale
Catheters MaterialsThe proposed device and the predicate device are made of base silicones from different manufacturers.Addition of Endexo® additive to the base silicone in the proposed device.The proposed device is made of implant-grade platinum-cured silicone, equivalent to that of the predicate.Endexo® is a surface modifying macromolecule (SMM) polymer blended into the catheter's base silicone.
Ventricular Catheter MarkingsThe proposed device's markings consist of numbers with circumferential bands, which are made of implant-grade ink and silicone adhesive. The predicate device's markings consist of tantalum dots.The proposed device also has additional markings compared to the predicate device.The addition of numerical markings and circumferential bands is part of continuous improvement in catheter usability.
Ventricular Catheter LengthThe proposed device's ventricular catheter comes in 14 cm and 23 cm lengths. In comparison, the predicate device's ventricular catheter only has a 14 cm option.Adding a 23 cm ventricular catheter supports the variety of approaches and valve placements used in hydrocephalus shunting.
Ventricular Catheter Inlet HolesThe proposed device has a smaller number of inlet holes, compared to the predicate devices. The inlet holes in the proposed device are of different size and shape, and have a different spatial arrangement, compared to the predicate.The change in the number, size, shape, and spatial arrangement of the inlet holes is part of continuous improvement in catheter usability.
StyletThe stylets of the proposed device have a square shaped cross section, compared to a circular shape in the proposed device's stylet.A longer stylet is paired with the longer ventricular catheter (23 cm).The square shaped cross section of the stylet is part of continuous improvement in catheter usability.The longer stylet is required for aiding the insertion of the new 23 cm ventricular catheter.
Ventricular and Distal Catheters RadiopacityBoth the proposed and predicate devices use barium sulfate for radiopacity. The proposed ventricular and distal catheters are translucent with a radiopaque stripe along the entire length of the catheterThe use of a barium sulfate stripe in a translucent catheter is part of continuous improvement in catheter usability.

Page 7

Integra® LifeSciences Production Corporation
Codman® Libertís™ Shunt Catheter with Bactiseal® and Endexo® Technology

Page 2 of 4

Component AffectedProposed ModificationRationale
Catheters MaterialsThe proposed device and the predicate device are made of base silicones from different manufacturers.Addition of Endexo® additive to the base silicone in the proposed device.The proposed device is made of implant-grade platinum-cured silicone, equivalent to that of the predicate.Endexo® is a surface modifying macromolecule (SMM) polymer blended into the catheter's base silicone.
Ventricular Catheter MarkingsThe proposed device's markings consist of numbers with circumferential bands, which are made of implant-grade ink and silicone adhesive. The predicate device's markings consist of tantalum dots.The proposed device also has additional markings compared to the predicate device.The addition of numerical markings and circumferential bands is part of continuous improvement in catheter usability.
Ventricular Catheter LengthThe proposed device's ventricular catheter comes in 14 cm and 23 cm lengths. In comparison, the predicate device's ventricular catheter only has a 14 cm option.Adding a 23 cm ventricular catheter supports the variety of approaches and valve placements used in hydrocephalus shunting.
Ventricular Catheter Inlet HolesThe proposed device has a smaller number of inlet holes, compared to the predicate devices. The inlet holes in the proposed device are of different size and shape, and have a different spatial arrangement, compared to the predicate.The change in the number, size, shape, and spatial arrangement of the inlet holes is part of continuous improvement in catheter usability.
StyletThe stylets of the proposed device have a square shaped cross section, compared to a circular shape in the proposed device's stylet.A longer stylet is paired with the longer ventricular catheter (23 cm).The square shaped cross section of the stylet is part of continuous improvement in catheter usability.The longer stylet is required for aiding the insertion of the new 23 cm ventricular catheter.
Ventricular and Distal Catheters RadiopacityA longer stylet is paired with the longer ventricular catheter (23 cm).Both the proposed and predicate devices use barium sulfate for radiopacity. The proposed ventricular and distal catheters are translucent with a radiopaque stripe along the entire length of the catheterThe longer stylet is required for aiding the insertion of the new 23 cm ventricular catheter.The use of a barium sulfate stripe in a translucent catheter is part of continuous improvement in catheter usability.

Page 2 of 4

Page 8

Integra® LifeSciences Production Corporation
Codman® Libertís™ Shunt Catheter with Bactiseal® and Endexo® Technology

Page 3 of 4

Component AffectedProposed ModificationRationale
Packagingtubing. In comparison, the predicate device is fully radiopaque.The predicate and the proposed devices use a double sterile barrier system composed of a sealed tray/lid system inside a sealed pouch. Some of the components of the packaging systems are made with different materials.The differences in packaging materials are intended to support sterilization and to reduce degradation of the antibiotics in the catheters.

In addition to the differences in the materials and design of the proposed and the predicate devices, the devices are sterilized using different methods and have different MRI compatibility. The proposed device is sterilized using E-beam radiation while the predicate is sterilized using steam. Both sterilization methods achieve the same Sterility Assurance Level (SAL = 10-6). The proposed device has an MR-safe designation while the predicate has an MR-conditional designation. The difference in MRI compatibility is due to the absence of tantalum, a metallic alloy, in the markings of proposed device.

807.92(b) 1-2: Summary of Nonclinical and Clinical Testing Performed

The following performance testing has been conducted in support of the substantial equivalence determination. The testing was performed according to FDA-recognized consensus standards and internal test methods. All testing was performed on production equivalent devices.

Performance Bench Test Results

TestConclusion
Evaluation of Safety and Performance per ISO 7197 and ASTM F647-94Pass
Functional/Mechanical Performance per ASTM F647-94 and ASTM D412-16Pass
MRI Compatibility per ASTM F2053-20Pass
Antimicrobial Efficacy – Zone of Inhibition (ZOI) per internal methodsPass
Antibiotic Content Analysis using High-performance Liquid Chromatography (HPLC), per internal methodsPass
Residual Solvents Testing via Headspace Gas Chromatography – Flame Ionization Detection (HS GC-FID), per FDA guidance "Q3C – Tables and List Guidance for IndustryPass
Usability Evaluation and Design Validation per IEC 62366-1Pass
In vitro flow loop thrombogenicity testObserved significantly less thrombus deposition on Codman Libertis Distal Shunt Catheter with Bactiseal and Endexo Technology compared to predicate device

Biocompatibility Testing Results

TestConclusion
Cytotoxicity per ISO 10993-5:2009Pass
Sensitization per ISO 10993-10:2021Pass
Irritation per ISO 10993-10:2021 and ISO 10993-23:2021Pass
Acute Systemic Toxicity per ISO 10993-11:2017Pass
Pyrogenicity per ISO 10993-11:2017Pass
Subacute/Subchronic Toxicity per ISO 10993-6:2016 and ISO 10993-11:2017Pass
Genotoxicity per ISO 10993-3:2014Pass
Implantation per ISO 10993-6:2016Pass
Chronic Toxicity per ISO 10993-11:2017Pass
Hemocompatibility per ISO 10993-4:2017Pass
Carcinogenicity per ISO 10993-3:2014Pass

Sterilization/Cleaning

The proposed device is sterilized via electron beam (E-beam) radiation. A sterilization validation was performed according to FDA recognized standards ISO 11137-1, ISO 11137-2, ISO 11737-1, and ISO 11737-2. The sterilization validation demonstrates that the proposed sterilization method achieves a sterility assurance level (SAL) of 10-6.

Shelf Life

The shelf-life established for the predicate and the proposed devices is the same, 1-year.

Clinical Studies

No clinical studies were required as appropriate verification of the subject devices was achieved based on the comparison to the predicate devices and from the results of the bench testing.

807.92(b) (3) Conclusion

Based upon the intended use, design, comparison to the predicate device, and testing performed, Integra LifeSciences believes that the proposed Libertís™ Shunt Catheter does not raise any different questions of safety and effectiveness, and is therefore, substantially equivalent to the predicate Bactiseal® Shunt Catheter.

Page 4 of 4

§ 882.5550 Central nervous system fluid shunt and components.

(a)
Identification. A central nervous system fluid shunt is a device or combination of devices used to divert fluid from the brain or other part of the central nervous system to an internal delivery site or an external receptacle for the purpose of relieving elevated intracranial pressure or fluid volume (e.g., due to hydrocephalus). Components of a central nervous system shunt include catheters, valved catheters, valves, connectors, and other accessory components intended to facilitate use of the shunt or evaluation of a patient with a shunt.(b)
Classification. Class II (performance standards).