K Number
K161853

Validate with FDA (Live)

Manufacturer
Date Cleared
2017-02-27

(236 days)

Product Code
Regulation Number
882.5550
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Miethke proGAV Programmable Shunt System is intended to shunt cerebrospinal fluid (CSF) from the lateral ventricles of the brain into the peritoneum.

The Miethke proGAV 2.0 Adjustable Shunt System is intended to shunt cerebrospinal fluid (CSF) from the lateral ventricles of the brain into the peritoneum.

The Miethke proSA Adjustable Shunt System is intended to shunt cerebrospinal fluid (CSF) from the lateral ventricles of the brain into the peritoneum.

Device Description

The proGAV is a "programmable" shunt that can be set for a range of pressures. The valve in the proGAV is a leaf spring and ball mechanism that is mechanically controlled by internal magnets. The outer case for the device is made of titanium. The shunt comes with a manual device to verify the pressure setting and another to set or re-set the pressure. These manual accessories are both for external use by the physician. Various Miethke shunt system accessories such as shunt assistants, catheters, connectors, deflectors and reservoirs are also offered with the proGAV.

proGAV 2.0 is an adjustable differential pressure valve that can be set for a range of pressures. The proGAV 2.0 valve is comprised of a titanium housing that contains a leaf spring and ball mechanism that is mechanically controlled by internal magnets. Manual devices are available to locate, verify the pressure setting and to set or re-set the pressure pre and postoperatively. These manual accessories are for external use by the physician. The device will be distributed by itself or in combination with the ShuntAssistant valve or proSA valve. The proGAV 2.0 adjustable differential pressure valve includes the same legally marketed accessories that are available with the Miethke Shunt Systems.

proSA is an adjustable gravitational valve that can be set for a range of pressures. The proSA valve is comprised of a titanium housing that contains a tantalum weight, leaf spring and ball mechanism that is mechanically controlled by internal magnets. Several manual devices are available to verify the pressure setting and to set or re-set the pressure pre and postoperatively. These manual accessories are for external use by the physician. The device will be distributed by itself or in combination with the miniNAV valve or proGAV valve. The proSA adjustable gravitational valve includes the same legally marketed accessories that are available with the Miethke Shunt Systems.

AI/ML Overview

The provided text describes the acceptance criteria and performance data for three medical devices: the Miethke proGAV Programmable Shunt System, the Miethke proGAV 2.0 Adjustable Shunt System, and the Miethke proSA Adjustable Shunt System.

The primary purpose of the studies was to demonstrate that each of these devices, when used with specific non-invasive verification tools (proGAV Verification Tool, proGAV Verification Compass, or proGAV 2.0 Compass), could accurately verify the shunt setting, thereby removing the need for radiographic (X-ray) confirmation. The acceptance criterion for each device was to demonstrate "substantial equivalence" between the required radiographic verification method and the non-invasive method.

Here's a breakdown of the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria for all three devices:

  • Demonstrate substantial equivalence between the radiographic verification method and the non-invasive verification method using the specified tools.
  • The non-invasive method using the specified tools must be able to accurately measure the shunt setting, making X-ray confirmation no longer required.
DeviceAcceptance CriteriaReported Device Performance
Miethke proGAV Programmable Shunt SystemThe test results met the acceptance criteria by demonstrating substantial equivalence between the required radiographic verification method and the non-invasive method using the proGAV Verification Tool and proGAV Verification Compass. This implies that the measurement agreement between these non-invasive tools and X-ray confirmation was within acceptable limits, allowing for the removal of the radiographic verification requirement in the labeling.Verification testing assessing the measurement agreement between X-ray confirmation and the proGAV Verification Tool when used with the proGAV Programmable Shunt System. Verification testing assessing the measurement between X-ray confirmation and the proGAV Verification Compass when used with the proGAV Programmable Shunt System. Result: Test results met acceptance criteria, demonstrating substantial equivalence between radiographic and non-invasive methods.
Miethke proGAV 2.0 Adjustable Shunt SystemThe test results met the acceptance criteria by demonstrating substantial equivalence between the required radiographic verification method and the non-invasive method using the proGAV 2.0 Compass, proGAV Verification Compass, and proGAV Verification Tool. This implies that the measurement agreement between these non-invasive tools and X-ray confirmation was within acceptable limits, allowing for the removal of the radiographic verification requirement in the labeling.Verification testing assessing the measurement agreement between X-ray confirmation and the proGAV 2.0 Compass when used with the proGAV 2.0 Adjustable Shunt System. Verification testing assessing the measurement between X-ray confirmation and the proGAV Verification Compass when used with the proGAV 2.0 Adjustable Shunt System. Verification testing assessing the measurement agreement between X-ray confirmation and the proGAV Verification Tool when used with the proGAV 2.0 Adjustable Shunt System. Result: Test results met acceptance criteria, demonstrating substantial equivalence between radiographic and non-invasive methods.
Miethke proSA Adjustable Shunt SystemThe test results met the acceptance criteria by demonstrating substantial equivalence between the required radiographic verification method and the non-invasive method using the proSA Verification Tool and proSA Verification Compass. This implies that the measurement agreement between these non-invasive tools and X-ray confirmation was within acceptable limits, allowing for the removal of the radiographic verification requirement in the labeling.Verification testing assessing the measurement between X-ray confirmation and the proSA Verification Tool when used with the proSA Adjustable Shunt System. Verification testing assessing the measurement between X-ray confirmation and the proSA Verification Compass when used with the proSA Adjustable Shunt System. Result: Test results met acceptance criteria, demonstrating substantial equivalence between radiographic and non-invasive methods.

2. Sample size used for the test set and the data provenance

The document specifies "Bench testing" for the performance data. It does not provide specific sample sizes (e.g., number of shunts tested, number of measurements taken) for the test set or the data provenance (e.g., country of origin, retrospective or prospective).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

The ground truth for comparison in these studies was "X-ray confirmation." The text does not mention the use of human experts to establish this ground truth or their qualifications. The X-ray is presented as the objective, established method of verification.

4. Adjudication method for the test set

Not applicable. The study involved comparing a device-tool measurement to an X-ray confirmation, not expert adjudication of subjective assessments.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not an AI-based diagnostic or assistive device. The study focused on the performance of mechanical devices and their verification tools.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable, as this is not an algorithm or AI device. The "standalone" performance here would refer to the performance of the non-invasive tools in indicating the shunt setting, which was the core of the bench testing.

7. The type of ground truth used

The ground truth used was radiographic (X-ray) confirmation of the shunt setting.

8. The sample size for the training set

The document does not mention a training set, as these are mechanical devices undergoing bench testing for performance and equivalence, not machine learning algorithms requiring training data.

9. How the ground truth for the training set was established

Not applicable, as there is no training set for these types of device studies based on the provided document.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol of three human profiles facing right, arranged in a cascading manner, representing the department's focus on people and health.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 27, 2017

Aesculap, Inc. Kathy A. Racosky Senior Regulatory Affairs Specialist 3773 Corporate Parkway Center Valley, Pennsylvania 18034

Re: K161853

Trade/Device Name: Miethke proGAV Programmable Shunt System, Miethke proGAV 2.0 Adjustable Shunt System, Miethke proSA Adjustable Shunt System Regulation Number: 21 CFR 882.5550 Regulation Name: Central Nervous System Fluid Shunt and Components Regulatory Class: Class II Product Code: JXG Dated: January 20, 2017 Received: January 23, 2017

Dear Ms. Racosky:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Carlos L. Pena-S

Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K161853

Device Name

Miethke proGAV Programmable Shunt System Miethke proGAV 2.0 Adjustable Shunt System Miethke proSA Adjustable Shunt System

Indications for Use (Describe)

The Miethke proGAV Programmable Shunt System is intended to shunt cerebrospinal fluid (CSF) from the lateral ventricles of the brain into the peritoneum.

The Miethke proGAV 2.0 Adjustable Shunt System is intended to shunt cerebrospinal fluid (CSF) from the lateral ventricles of the brain into the peritoneum.

The Miethke proSA Adjustable Shunt System is intended to shunt cerebrospinal fluid (CSF) from the lateral ventricles of the brain into the peritoneum.

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY

Miethke proGAV Programmable Shunt SystemFebruary 15, 2017
COMPANY:Aesculap®, Inc.3773 Corporate ParkwayCenter Valley, PA 18034Establishment Registration Number: 2916714
CONTACT:Kathy A. Racosky610-984-9291 (phone)610-791-6882 (fax)kathy.racosky@aesculap.com
TRADE NAME:Aesculap Miethke proGAV Programmable Shunt System
COMMON NAME:Hydrocephalus Shunt System
CLASSIFICATION:Class II
CLASSIFICATION NAME: Shunt, Central Nervous System and Components
REGULATION NUMBER:882.5550
PRODUCT CODE:JXG

DEVICE DESCRIPTION

The proGAV is a "programmable" shunt that can be set for a range of pressures. The valve in the proGAV is a leaf spring and ball mechanism that is mechanically controlled by internal magnets. The outer case for the device is made of titanium. The shunt comes with a manual device to verify the pressure setting and another to set or re-set the pressure. These manual accessories are both for external use by the physician. Various Miethke shunt system accessories such as shunt assistants, catheters, connectors, deflectors and reservoirs are also offered with the proGAV .

INDICATIONS FOR USE

The Miethke proGAV Programmable Shunt System is intended to shunt cerebrospinal fluid (CSF) from the lateral ventricles of the brain into the peritoneum.

SUBSTANTIAL EQUIVALENCE

Comparative information presented in the 510(k) supports the substantial equivalence of the subject proGAV Programmable Shunt System to the predicate proGAV Programmable Shunt System (K062009 / K103003). The proGAV Programmable Shunt System is substantially equivalent to the identified predicate with respect to performance characteristics, design, principles of operation and materials. The difference between the subject device and the

{4}------------------------------------------------

predicate is the removal of the required radiographic verification stated in the indications for use and labeling.

TECHNOLOGICAL CHARACTERISTICS(compared to Predicate(s))

The Aesculap Miethke proGAV Programmable Shunt System is substantially equivalent to the predicate Miethke proGAV Programmable Shunt System. The subject device is shown to be substantially equivalent and has the same performance characteristic to its predicate device through comparison in design, principles of operation, and materials. The proGAV device characteristics are summarized below.

Miethke proGAVProgrammable Shunt SystemK161853Miethke proGAVProgrammable Shunt SystemK103003/K062009
Indication:The Miethke proGAVProgrammable Shunt System isintended to shunt cerebrospinalfluid (CSF) from the lateralventricles of the brain into theperitoneum.The Miethke proGAVProgrammable Shunt System isintended to shunt cerebrospinalfluid (CSF) from the lateralventricles of the brain into theperitoneum. Adjustments of theproGAV shunt can be verified byusing the verification instrumentbut must be confirmed byradiograph (X-ray).
AdjustableSameYes
ValveTypeSameAdjustable differential pressure
Material:
HousingSameTitanium Alloy Ti4Al6V
MagnetsSameNeodym Ferrite Boron
BallSameAlpha Sapphire
SpringSameTitanium Alloy Ti4Al6V
DesignSameCircular
Magnet:
# of magnetsSame2
DimensionsSame1.5 x 1.5 mm
VolumeSame5.3 mm3
Pressure levels:SameAdjustable0 - 20 cmH2O
Sterile/ Single UseSameYes
SterilizationSameSteam
Packaging:SameDouble Peel Pouch

{5}------------------------------------------------

Miethke proGAVProgrammable Shunt SystemK161853Miethke proGAVProgrammable Shunt SystemK103003/K062009
Tools:
Verification toolMasterdiscVerification CompassAdjustment toolAdjustment discSameManually wipe with an alcoholbase cleaner (>75% alcohol). Maynot be re-sterilized.
Check-mateSameFor use in a sterile surgical field.Sterile, reusable, may be re-sterilized
Tool settings andreadings:Same0 - 20 cmH2O

PERFORMANCE DATA

Bench testing on the proposed device, proGAV Programmable Shunt System, included the following:

  • Verification testing assessing the measurement agreement between X-ray confirmation . and the proGAV Verification Tool when used with the proGAV Programmable Shunt System
  • . Verification testing assessing the measurement between X-ray confirmation and the proGAV Verification Compass when used with the proGAV Programmable Shunt System

Test results demonstrated the substantial equivalence between the required radiographic verification method and the non-invasive method with the proGAV Verification Tool and proGAV Verification Compass. The test results met the acceptance criteria. Results of verification testing have demonstrated that the proposed proGAV Programmable Shunt System is substantially equivalent to the predicate proGAV Programmable Shunt System.

{6}------------------------------------------------

Page 1 of 3

510(k) SUMMARY

Miethke proGAV 2.0 Adjustable Shunt System
February 15, 2017
COMPANY:Aesculap®, Inc.3773 Corporate ParkwayCenter Valley, PA 18034Establishment Registration Number: 2916714
CONTACT:Kathy A. Racosky610-984-9291 (phone)610-791-6882 (fax)kathy.racosky@aesculap.com
TRADE NAME:Aesculap Meithke proGAV 2.0 Programmable Shunt System
COMMON NAME:Hydrocephalus Shunt System
CLASSIFICATION:Class II
CLASSIFICATION NAME:Shunt, Central Nervous System and Components
REGULATION NUMBER:882.5550
PRODUCT CODE:JXG

DEVICE DESCRIPTION

proGAV 2.0 is an adjustable differential pressure valve that can be set for a range of pressures. The proGAV 2.0 valve is comprised of a titanium housing that contains a leaf spring and ball mechanism that is mechanically controlled by internal magnets. Manual devices are available to locate, verify the pressure setting and to set or re-set the pressure pre and postoperatively. These manual accessories are for external use by the physician. The device will be distributed by itself or in combination with the ShuntAssistant valve or proSA valve. The proGAV 2.0 adjustable differential pressure valve includes the same legally marketed accessories that are available with the Miethke Shunt Systems.

INDICATIONS FOR USE

The Miethke proGAV 2.0 Adjustable Shunt System is intended to shunt cerebrospinal fluid (CSF) from the lateral ventricles of the brain into the peritoneum.

SUBSTANTIAL EQUIVALENCE

Comparative information presented in the 510(k) supports the substantial equivalence of the subject proGAV 2.0 Adjustable Shunt System to the predicate proGAV 2.0 Adjustable Shunt

{7}------------------------------------------------

System (K141687). The proGAV 2.0 Adjustable Shunt System is substantially equivalent to the identified predicate with respect to performance characteristics, design, principles of operation and materials. The difference between the subject device and the predicate is the removal of the required radiographic verification stated in the indications for use and labeling.

TECHNOLOGICAL CHARACTERISTICS(compared to Predicate(s))

The Aesculap Miethke proGAV 2.0 Adjustable Shunt System is substantially equivalent to the predicate Miethke proGAV 2.0 Adjustable Shunt System. The subject device is shown to be substantially equivalent and has the same performance characteristic to its predicate device through comparison in design, principles of operation, and materials. The proGAV 2.0 device characteristics are summarized below.

Miethke proGAV 2.0 AdjustableShunt SystemK161853Miethke proGAV 2.0 AdjustableShunt SystemK141687
Indication:The Miethke proGAV 2.0Adjustable Shunt System isintended to shunt cerebrospinalfluid (CSF) from the lateralventricles of the brain into theperitoneum.The Miethke proGAV 2.0Adjustable Shunt System isintended to shunt cerebrospinalfluid (CSF) from the lateralventricles of the brain into theperitoneum. Adjustments of theproGAV 2.0 shunt can be verifiedby using the verificationinstrument but must be confirmedby radiograph (X-ray).
AdjustableSameYes
ValveType:SameAdjustable differential pressure
Material:
HousingSameTitanium Alloy Ti4Al6V
MagnetsSameNeodym Ferrite Boron
BallSameAlpha Sapphire
SpringSameTitanium Alloy Ti4Al6V
DesignSameCircular
Magnet:
# of magnetsSame4
DimensionsSame1.5 x 1.5 mm
VolumeSame10.6 mm3
Pressure levels:SameAdjustable
Same0 - 20 cmH2O
Sterile/ Single UseSameYes
SterilizationSameSteam
Packaging:SameDouble Peel Pouch

{8}------------------------------------------------

Miethke proGAV 2.0 AdjustableShunt SystemMiethke proGAV 2.0 AdjustableShunt System
K161853K141687
proGAV 2.0 Tools:
CompassSameManually wipe with an alcohol
Adjustment toolSamebase cleaner (>75% alcohol).May not be re-sterilized.
Compatible w/proGAV Tools:
SameVerification tool, Verificationcompass, Master disk, Adjustmenttool, Adjustment disk, Check-mate
Tool settings andreadings:Same0 – 20 cmH2O

PERFORMANCE DATA

Bench testing on the proposed device, proGAV 2.0 Adjustable Shunt System, included the following:

  • Verification testing assessing the measurement agreement between X-ray confirmation . and the proGAV 2.0 Compass when used with the proGAV 2.0 Adjustable Shunt System
  • Verification testing assessing the measurement between X-ray confirmation and the proGAV Verification Compass when used with the proGAV 2.0 Adjustable Shunt System
  • Verification testing assessing the measurement agreement between X-ray confirmation and the proGAV Verification Tool when used with the proGAV 2.0 Adjustable Shunt System

Test results demonstrated the substantial equivalence between the required radiographic verification method and the non-invasive method with the proGAV 2.0 Compass, proGAV Verification Compass and proGAV Verification Tool. The test results met the acceptance criteria. Results of verification testing have demonstrated that the proposed proGAV 2.0 Adjustable Shunt System is substantially equivalent to the predicate proGAV 2.0 Adjustable Shunt System.

{9}------------------------------------------------

510(k) SUMMARY

Miethke proSA Adjustable Shunt System
February 15, 2017
COMPANY:Aesculap®, Inc.3773 Corporate ParkwayCenter Valley, PA 18034Establishment Registration Number: 2916714
CONTACT:Kathy A. Racosky610-984-9291 (phone)610-791-6882 (fax)kathy.racosky@aesculap.com
TRADE NAME:Aesculap Miethke proSA Adjustable Shunt System
COMMON NAME:Hydrocephalus Shunt System
CLASSIFICATION:Class II
CLASSIFICATION NAME:Shunt, Central Nervous System and Components
REGULATION NUMBER:882.5550
PRODUCT CODE:JXG

DEVICE DESCRIPTION

proSA is an adjustable gravitational valve that can be set for a range of pressures. The proSA valve is comprised of a titanium housing that contains a tantalum weight, leaf spring and ball mechanism that is mechanically controlled by internal magnets. Several manual devices are available to verify the pressure setting and to set or re-set the pressure pre and postoperatively. These manual accessories are for external use by the physician. The device will be distributed by itself or in combination with the miniNAV valve or proGAV valve. The proSA adjustable gravitational valve includes the same legally marketed accessories that are available with the Miethke Shunt Systems.

INDICATIONS FOR USE

The Miethke proSA Adjustable Shunt System is intended to shunt cerebrospinal fluid (CSF) from the lateral ventricles of the brain into the peritoneum.

SUBSTANTIAL EQUIVALENCE

Comparative information presented in the 510(k) supports the substantial equivalence of the subject proSA Adjustable Shunt System to the predicate proSA Adjustable Shunt System

{10}------------------------------------------------

(K120559). The proSA Adjustable Shunt System is substantially equivalent to the identified predicate with respect to performance characteristics, design, principles of operation and materials. The difference between the subject device and the predicate is the removal of the required radiographic verification stated in the indications for use and labeling.

TECHNOLOGICAL CHARACTERISTICS(compared to Predicate(s))

The Aesculap Miethke proSA Adjustable Shunt System is substantially equivalent to the predicate Miethke proSA Adjustable Shunt System. The subject device is shown to be substantially equivalent and has the same performance characteristic to its predicate device through comparison in design, principles of operation, and materials. The proSA device characteristics are summarized below.

Miethke proSA AdjustableShunt SystemK161853Miethke proSA AdjustableShunt SystemK120559
Indication:The Miethke proSA AdjustableShunt System is intended to shuntcerebrospinal fluid (CSF) fromthe lateral ventricles of the braininto the peritoneum.The Miethke proSA AdjustableShunt System is intended to shuntcerebrospinal fluid (CSF) fromthe lateral ventricles of the braininto the peritoneum. Adjustmentsof the proSA shunt can be verifiedby using the verificationinstrument but must be confirmedby radiograph (X-ray).
AdjustableSameYes
ValveTypeSameAdjustable Gravitational valve
Material:
HousingSameTitanium Alloy Ti4Al6V
MagnetsSameNeodym Ferrite Boron
BallSameAlpha Sapphire
SpringSameTitanium Alloy Ti4Al6V
WeightSameTantalum
DesignSameCircular
Magnet:
# of magnetsSame2
DimensionsSame1.5 x 1.5 mm
VolumeSame5.3 mm3
Pressure levels:SameAdjustable
Same0 - 40 cmH2O
Sterile/ Single UseSameYes
SterilizationSameSteam
Packaging:SameDouble Peel Pouch

{11}------------------------------------------------

Miethke proSA AdjustableShunt SystemMiethke proSA AdjustableShunt System
K161853K120559
Tools:
Verification toolMasterdiscVerification CompassAdjustment toolAdjustment discSameManually wipe with an alcoholbase cleaner (>75% alcohol). Maynot be re-sterilized.
Check-mateSameFor use in a sterile surgical field.Sterile, reusable, may be re-sterilized
Tool settings andreadings:Same0 - 40 cmH2O

PERFORMANCE DATA

Bench testing on the proposed device, proSA Adjustable Shunt System, included the following:

  • Verification testing assessing the measurement between X-ray confirmation ● and the proSA Verification Tool when used with the proSA Adjustable Shunt System
  • Verification testing assessing the measurement between X-ray confirmation and the proSA Verification Compass when used with the proSA Adjustable Shunt System

Test results demonstrated the substantial equivalence between the required radiographic verification method and the non-invasive method with the proSA Verification Tool and proSA Verification Compass. The test results met the acceptance criteria. Results of verification testing have demonstrated that the proposed proSA Adjustable Shunt System is substantially equivalent to the predicate proSA Adjustable Shunt System.

§ 882.5550 Central nervous system fluid shunt and components.

(a)
Identification. A central nervous system fluid shunt is a device or combination of devices used to divert fluid from the brain or other part of the central nervous system to an internal delivery site or an external receptacle for the purpose of relieving elevated intracranial pressure or fluid volume (e.g., due to hydrocephalus). Components of a central nervous system shunt include catheters, valved catheters, valves, connectors, and other accessory components intended to facilitate use of the shunt or evaluation of a patient with a shunt.(b)
Classification. Class II (performance standards).