Search Filters

Search Results

Found 25 results

510(k) Data Aggregation

    K Number
    K231612

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-05-07

    (340 days)

    Product Code
    Regulation Number
    892.5770
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Leo Cancer Care Patient Positioning Subsystem is indicated for precise positioning of human patients to facilitate delivery of external beam radiation when integrated with a separate beam therapy treatment delivery device.

    Device Description

    The Leo Cancer Care (LCC) Patient Positioning Subsystem (EVE) is designed to deliver precise positioning of a patient when combined with an external Beam Therapy device, to provide the intended treatment therapy prescribed for patients requiring radiation treatments.

    The Patient Positioning System has 6 degrees of freedom to provide full alignment capability for a patient treatment session, with the patient being located in an upright position. The Eve Patient Positioning Subsystem is held in a static position for the duration of the external Beam Therapy System radiation exposure. After the beam delivery has been completed, the patient positiong subsystem is then aligned to the next patient position required by the treatment system and the process repeated for the required series of different beam placements to achieve the full prescribed treatment delivery for the patient.

    AI/ML Overview

    The provided text is an FDA 510(k) summary for the "Leo Cancer Care Eve Patient Positioner System." It describes a medical device used for patient positioning in radiation therapy.

    However, this document DOES NOT describe an AI/ML powered device, nor does it contain information about acceptance criteria or a study that proves a device meets acceptance criteria related to AI/ML performance.

    Instead, the document outlines the substantial equivalence of the Eve Patient Positioner System to a predicate device (LEONI Orion System K160518) based on non-clinical performance testing for device functionality and safety. The document states:

    • "No animal or clinical tests were performed to establish substantial equivalence with the predicate device." Therefore, there's no clinical study involving human readers or expert consensus for ground truth as would be relevant for an AI/ML device.
    • The performance testing focuses on mechanical and electrical performance (e.g., accuracy of positioning, speed, patient load, motion axes, and compliance with general medical device and software life-cycle standards like ISO 14971, IEC 60601-1, IEC 62304).

    Therefore, it is impossible to extract the requested information regarding AI/ML acceptance criteria, study details, human reader performance, or ground truth establishment relevant to an AI/ML algorithm from this document.

    If you have a document describing an AI/ML powered medical device, please provide that.

    Ask a Question

    Ask a specific question about this device

    K Number
    K182683

    Validate with FDA (Live)

    Date Cleared
    2019-06-27

    (274 days)

    Product Code
    Regulation Number
    892.5770
    Age Range
    18 - 120
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Z-Robot™ Patient Positioning System is an electro-mechanical robotic arm for patient positioning in radiotherapy and medical imaging. It is designed for positioning a patient with a high degree of accuracy and repeatability.

    Device Description

    The Z-Robot™ Patient Positioning System (hereinafter referred to as Z-Robot™) is an electro-mechanical robotic arm with the motion capability in six degrees of freedom. These axes allow the couch to be precisely positioned translationally and rotationally within the specified movement limits. Therefore, Z-Robot™ could be used to accurately position patient in radiotherapy and medical imaging.

    AI/ML Overview

    The requested information is contained within the provided text. The device in question is the "Z-Robot™ Patient Positioning System", which is an electro-mechanical robotic arm for patient positioning in radiotherapy and medical imaging. The acceptance criteria and the study that proves the device meets these criteria are primarily based on non-clinical testing.

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly provide a "table of acceptance criteria and reported device performance" for all tests. However, it lists the performance parameters tested and indicates that "All test results were acceptable per the acceptance criteria detailed in the corresponding protocol and test report." It also provides a direct comparison to the predicate device for several key performance metrics, implying these are the relevant acceptance criteria for substantial equivalence.

    Performance CharacteristicAcceptance Criteria (Implied by Predicate & "Same")Reported Device Performance (Z-Robot™ Patient Positioning System)
    Accuracy± 0.5 mm and ± 0.2°± 0.5 mm and ± 0.2° (Same as predicate)
    Linear XYZ maximal medical speed10 cm/s10 cm/s (Same as predicate)
    Linear XYZ maximum maintenance speed20 cm/s20 cm/s (Same as predicate)
    Angular Yaw Pitch Roll maximum medical speed5 °/s5 °/s (Same as predicate)
    Angular Yaw Pitch Roll maximum maintenance speed12 °/s12 °/s (Same as predicate)
    Collision detectionDetection of a 150 N forceDetection of a 150 N force (Same as predicate)
    "Overtravel" in case of emergency stop<5 mm<5 mm (Same as predicate)
    Speed Accuracy(Not explicitly detailed, but stated "acceptable")"All test results were acceptable"
    Translational Accuracy(Not explicitly detailed, but stated "acceptable")"All test results were acceptable"
    Rotational Accuracy(Not explicitly detailed, but stated "acceptable")"All test results were acceptable"
    Repeatability(Not explicitly detailed, but stated "acceptable")"All test results were acceptable"

    2. Sample Size Used for the Test Set and the Data Provenance

    The document states that "A series of safety and performance tests were conducted on the subject device, Z-Robot™ Patient Positioning System." It does not specify the sample size for these tests (e.g., number of devices tested or number of times each test was performed). The data provenance is also not specified in terms of country of origin or whether it was retrospective or prospective. It is implied to be prospective testing of the manufactured device.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This information is not applicable and not provided. The testing described is non-clinical performance testing of a physical device, not a diagnostic or AI-based system requiring expert interpretation or ground truth establishment in the typical sense. The "ground truth" for these tests would be the physical measurements taken during the performance evaluation.

    4. Adjudication Method for the Test Set

    This information is not applicable and not provided. As the testing is non-clinical performance testing, there's no "adjudication method" in the context of expert review or consensus as there would be for an image classification or diagnostic device.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC study was done. The device is a patient positioning system, not an AI-assisted diagnostic or imaging interpretation tool that would typically involve human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    The device itself is a mechanical system, not an algorithm in the sense of AI. The performance tests ("Reliability test," "Software Validation test," "Electromagnetic compatibility and electrical safety test," and "Performance test") evaluate the standalone performance of the Z-Robot™ Patient Positioning System, as it is a machine.

    7. The Type of Ground Truth Used

    The ground truth for the non-clinical performance tests would be precise measurements obtained through calibrated instruments and established engineering standards. For example, for "Accuracy (± 0.5 mm and ± 0.2°)", the ground truth would be the actual position and orientation measured by high-precision sensors, compared against the commanded movements. This is a form of empirical measurement against a known standard.

    8. The Sample Size for the Training Set

    This information is not applicable. The Z-Robot™ Patient Positioning System is a mechanical device, not an AI or machine learning model that requires a "training set."

    9. How the Ground Truth for the Training Set was Established

    This information is not applicable, as there is no "training set" for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K170502

    Validate with FDA (Live)

    Date Cleared
    2017-04-13

    (51 days)

    Product Code
    Regulation Number
    892.5770
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The patient positioning system is a SCARA designed to position a patient for medical procedures prescribed by oncologists and others that require a high degree of accuracy and repeatability.

    Device Description

    Not Found

    AI/ML Overview

    This document is a 510(k) premarket notification decision letter from the FDA for a device called "Patient Positioning System" by Forte Automation Systems, Inc. It does not contain information about acceptance criteria, device performance, study details, or ground truth establishment.

    Therefore, I cannot provide a response to your request, as the necessary information is not present in the provided text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K160518

    Validate with FDA (Live)

    Date Cleared
    2016-07-14

    (141 days)

    Product Code
    Regulation Number
    892.5770
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LEONI Orion System in an electro-mechanical robotic arm for patient positioning in radiotherapy and medical imaging. It is designed for positioning a patient with a high degree of accuracy and repeatability.

    Device Description

    The LEONI ORION System is an electro-mechanical robotic arm capable of motion in six degrees of freedom. The purpose of the device is to position a patient during radiotherapy, radiology and other medical applications with a high degree of accuracy and repeatability.
    The LEONI ORION System consists of the electro-mechanical unit that is a 6 axes robot which supports a standard radiotherapy table couch or other approved patient support device, and a Control Unit that includes computers and application software. The robot is linked to the Control Unit by cables.

    AI/ML Overview

    The provided text is a 510(k) summary for the LEONI ORION System, an electro-mechanical robotic arm for patient positioning in radiotherapy and medical imaging. It focuses on demonstrating substantial equivalence to a predicate device, rather than a clinical study evaluating diagnostic performance. Therefore, many of the requested items related to AI device evaluation (like MRMC studies, ground truth establishment for training data, and expert qualifications) are not applicable or cannot be extracted from this document.

    Here's an analysis based on the information available:

    1. Table of Acceptance Criteria and Reported Device Performance

    Feature/MetricAcceptance CriteriaReported Device Performance
    Payload CapacityNot explicitly stated as "acceptance criteria" but tested.375 kg / 826 lbs
    Accurate Treatment VolumeNot explicitly stated as "acceptance criteria" but tested.400 mm x 1000 mm x 500 mm
    AccuracyNot explicitly stated as "acceptance criteria" but tested.± 0.5 mm and ± 0.2°
    SpeedNot explicitly stated as "acceptance criteria" but tested.0.1 m/s and 6 °/s
    Safety (Collision Detection)Not explicitly stated as "acceptance criteria" but tested (detection of a 150N force).Detection of a 150N force
    Safety ("Overtravel" in emergency stop)Not explicitly stated as "acceptance criteria" but tested (< 5 mm).< 5 mm
    EMC (Electromagnetic Compatibility)(Implied standard compliance)Tests performed by third party
    Electrical and Mechanical Safety(Implied standard compliance)Tests performed by third party

    2. Sample size used for the test set and the data provenance

    The document describes bench tests rather than a "test set" in the context of diagnostic data. The tests were performed by LEONI, and some by a third-party test organization. The "data provenance" would refer to these test environments rather than patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This device is an electro-mechanical robotic arm, and the "ground truth" for its performance is determined by physical measurements and test procedures, not expert interpretation of medical images or conditions.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This concept applies to human reader studies, not to the performance testing of a robotic patient positioning system.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC study was done. This document does not describe an AI diagnostic device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This document describes the performance of the device itself (standalone in terms of its mechanical and electrical function). The "algorithm only" concept is typically for diagnostic AI, which is not what this device is.

    7. The type of ground truth used

    The ground truth for the performance metrics (payload, accuracy, speed, safety) are based on physical measurements and engineering standards/specifications inherent to the design and function of a robotic arm, not medical outcomes, pathology, or expert consensus on medical conditions.

    8. The sample size for the training set

    Not applicable. This device does not use a "training set" in the machine learning sense. Its performance is based on engineering design, calibration, and physical testing.

    9. How the ground truth for the training set was established

    Not applicable, for the same reason as above.

    Summary of the Study that Proves Acceptance Criteria:

    The study proving the device meets its acceptance criteria involved a series of non-clinical bench tests. These tests were conducted by LEONI and third-party test organizations. The tests evaluated the physical and functional characteristics of the LEONI ORION System, including:

    • Payload Capacity: Verified the maximum weight the system can support.
    • Accurate Treatment Volume: Demonstrated the spatial envelope within which the specified accuracy can be maintained.
    • Accuracy: Measured the precision of positioning (± 0.5 mm and ± 0.2°).
    • Speed: Measured the maximum linear and rotational movement speeds (0.1 m/s and 6 °/s).
    • Safety (Collision Detection): Tested the system's ability to detect a 150N force, indicating its safety mechanism.
    • Safety ("Overtravel" in Emergency Stop): Verified that the system stops within a very short distance (< 5 mm) after an emergency stop, ensuring patient safety.
    • Electromagnetic Compatibility (EMC): Confirmed compliance with relevant standards to prevent interference with other medical equipment.
    • Electrical and Mechanical Safety: Ensured the device meets safety standards for electrical operation and mechanical integrity.

    The document states: "The results from these performance assessments demonstrated that the LEONI ORION System met the acceptance criteria defined in the product specifications. Moreover these results proved that the subject device is comparable to the predicate device in terms of safety and effectiveness performance."

    Therefore, the "study" was a series of engineering and safety tests designed to validate the mechanical and electrical specifications of the robotic positioning system, rather than a clinical study or an AI performance evaluation.

    Ask a Question

    Ask a specific question about this device

    K Number
    K132084

    Validate with FDA (Live)

    Date Cleared
    2013-09-26

    (83 days)

    Product Code
    Regulation Number
    892.5770
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Protura Couch Software is intended to interface between record and verify systems, linear accelerator (Linac) software systems, Linac safeguard systems, and/or image guidance systems and the Protura Couch. The Protura Couch Software is also capable of operating the Protura Couch (6 Degree Axis Couch).

    Device Description

    The proposed device, the Protura Couch Software 1.3.0 (Part No. MT6XSM1.3.0), is patient positioning software used in radiation treatments in conjunction with an Image Guided Radiation Therapy (IGRT) system. The Protura Couch Software interfaces with the Couch Hardware (K031866) and is intended to position the patient after diagnostic decisions have been made based on results from an IGRT system. The Protura Couch Software allows the user to control patient positioning with six degrees of freedom from outside of the treatment room. The Protura Couch Software also includes the ability to interface with record and verify systems, Linac software systems, Linac safeguard systems, and/or image guidance systems.

    AI/ML Overview

    The provided document is a 510(k) summary for the Protura Couch Software 1.3.0, a patient positioning software used in radiation treatments. It states that non-clinical performance testing was conducted, but no clinical testing was performed. Therefore, it does not contain the specific information required to complete a table of acceptance criteria and reported device performance based on a study, nor does it provide details about patient data, expert reviews, or comparative effectiveness studies.

    However, based on the information provided, here's what can be extracted and inferred:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document lists "Non-Clinical Performance Data" areas tested and states "All testing confirmed that the Protura Couch Software 1.3.0 is safe and effective for its intended use." However, it does not provide specific acceptance criteria (e.g., numerical thresholds, error limits) or quantitative results of the device's performance against such criteria.

    Acceptance Criterion (Inferred from testing areas)Reported Device Performance (General Statement)
    Movement of the Protura CouchConfirmed safe and effective
    Interfacing with External SystemsConfirmed safe and effective
    Couch Pedestal and Isocenter AlignmentConfirmed safe and effective
    Patient Record HandlingConfirmed safe and effective

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: Not applicable. No clinical testing with patient data was performed. The testing described is non-clinical performance testing of the software and hardware interfaces.
    • Data Provenance: Not applicable, as no patient data was used.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    • Number of Experts: Not applicable. No clinical testing or expert review of patient data was conducted.
    • Qualifications of Experts: Not applicable.

    4. Adjudication Method for the Test Set:

    • Adjudication Method: Not applicable. No clinical testing or expert review requiring adjudication was conducted.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

    • MRMC Study: No. The document explicitly states: "No clinical testing was performed in the evaluation of this medical device." Therefore, no MRMC study, human reader improvement analysis, or AI assistance effect size was reported.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Standalone Performance: The non-clinical performance testing focused on the device's functionality, interfacing, and mechanical aspects. While this could be considered a form of "standalone" testing for the software's operational characteristics, it's not in the context of diagnostic or interpretive performance where "standalone" typically refers to an AI algorithm's accuracy in analysis without human input. The document describes testing of the software's technical operation (e.g., couch movement, interfacing), not its interpretation of data.

    7. The Type of Ground Truth Used:

    • Ground Truth: For the non-clinical performance tests, the "ground truth" would have been established by engineering specifications, calibration standards, and expected operational parameters for the couch movement, interface communication, and alignment accuracy. For example, for "Movement of the Protura Couch," the ground truth would be the programmed movement patterns and expected physical responses of the couch hardware. For "Couch Pedestal and Isocenter Alignment," it would be the precise physical alignment specifications.

    8. The Sample Size for the Training Set:

    • Sample Size: Not applicable. This device is software that controls a medical device (a radiation therapy couch). It is not an AI/ML algorithm that requires a training set in the typical sense (i.e., for learning from data to perform a diagnostic or predictive task). The software is programmed based on engineering principles and specifications, not trained on a dataset of patient cases.

    9. How the Ground Truth for the Training Set was Established:

    • Ground Truth for Training Set: Not applicable, as there is no "training set" in the context of an AI/ML algorithm for this device. The software functions based on its coded logic and interface protocols, not on learned patterns from a dataset.
    Ask a Question

    Ask a specific question about this device

    K Number
    K122413

    Validate with FDA (Live)

    Date Cleared
    2012-09-13

    (36 days)

    Product Code
    Regulation Number
    892.5770
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The patient positioning system is a SCARA designed robotic arm designed to position a patient for medical procedures prescribed by oncologists and others that require a high degree of accuracy and repeatability.

    Device Description

    The patient positioning system is a SCARA designed robotic arm allowing six degrees of freedom.

    AI/ML Overview

    The provided 510(k) summary for the Forte Automation Systems Patient Positioning System (K122413) does not contain information about acceptance criteria and a study proving the device meets those criteria in the context of clinical performance or diagnostic accuracy.

    This 510(k) pertains to a patient positioning system, which is a robotic arm for positioning patients during medical procedures. The submission focuses on non-clinical testing to demonstrate substantial equivalence to a predicate device. Clinical efficacy or diagnostic performance, which would typically involve acceptance criteria related to accuracy, sensitivity, specificity, and a clinical study to prove them, are not applicable to this type of device submission.

    Here's a breakdown of the available information based on your requested points, highlighting why some are not present for this specific device:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Non-Clinical)Reported Device Performance
    Electromagnetic Compatibility and Susceptibility TestsPassed (by a third party)
    Surge and Static TestsPassed (by a third party)
    Vibration TestsPassed (by a third party)
    Speed TestsPassed (by Forte Automation)
    Accuracy TestsPassed (by Forte Automation)
    Collision Detection TestsPassed (by Forte Automation)

    It's crucial to understand that these are engineering/safety performance criteria, not clinical performance criteria in the context of diagnostic devices. The document states, "The submitted device does not impart energies into a patient. Therefore no clinical testing was needed." This confirms that the 510(k) submission relies on non-clinical engineering and safety performance to demonstrate substantial equivalence.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    Not applicable. The tests conducted were non-clinical engineering tests (e.g., electromagnetic compatibility, vibration, speed, accuracy, collision detection) on the device itself, not on a patient population. There is no "test set" of patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. Ground truth, in the context of clinical or diagnostic performance, is established by experts (e.g., radiologists, pathologists). Since no clinical testing was performed and no patient data was analyzed, this information is not relevant to this 510(k) submission.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. Adjudication methods are used to resolve discrepancies in expert interpretations of clinical or diagnostic data. This was not part of the non-clinical testing performed.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. MRMC studies are used to assess the clinical effectiveness of diagnostic devices and often involve human readers interpreting medical images. This device is a patient positioning system, not an imaging or diagnostic device with AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. Standalone performance refers to the diagnostic accuracy of an algorithm without human intervention. This is not a diagnostic algorithm. The "performance" discussed for this device relates to its mechanical and electrical functions.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the non-clinical tests, the "ground truth" was established by engineering specifications, industry standards, and the physical properties of the device. For example, the accuracy tests would compare the device's measured positioning to its programmed target position, with "ground truth" being the theoretically correct position based on the design.

    8. The sample size for the training set

    Not applicable. This device is not an AI/ML algorithm that requires a "training set" for model development.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K122201

    Validate with FDA (Live)

    Date Cleared
    2012-08-30

    (36 days)

    Product Code
    Regulation Number
    892.5770
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Protura Couch Software is interface between a record and verify system, linear accelerator (Linac) software, Linac safeguard systems, and/or image guidance systems and the Protura Couch. The Protura Couch Software is also capable of operating the Protura Couch (6 Degree Axis Couch).

    Device Description

    The proposed device, the Protura Couch Software (MT6XSMEU), is patient positioning software used in radiation treatments in conjunction with an Image Guided Radiation Therapy (IGRT) system. The Protura Couch Software interfaces with the Protura Couch Hardware (K031866) and is intended to position the patient after diagnostic decisions have been made based on results from an IGRT system. The Protura Couch Software allows the user to control patient positioning with six degrees of freedom from outside of the treatment room. The Protura Couch Software also includes the ability to interface with a record and verify system, Linac software, Linac safeguard systems, and/or image guidance systems.

    AI/ML Overview

    The provided text describes a 510(k) summary for the Protura Couch Software (device name: MT6XSMEU) by MED-TEC, Inc. d/b/a CIVCO Medical Solutions. This software is designed for patient positioning in radiation therapy.

    Here's an analysis of the acceptance criteria and study information, based solely on the provided document:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state quantitative acceptance criteria in a table format. Instead, it describes performance testing conducted. The reported device performance is that "All testing confirmed that the Protura Couch Software is safe and effective for its intended use" and that it is "substantially equivalent to the predicate device in regards to safety, effectiveness, and performance."

    The characteristics tested include:

    • Movement of the Protura Couch
    • Interfacing with External Systems
    • Couch Pedestal and Isocenter Alignment
    • Patient Record Handling

    Without specific numerical targets for these characteristics, a precise table of acceptance criteria and reported performance cannot be generated. The document indicates these tests were performed and passed.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The document does not specify a sample size for the test set. It mentions "non-clinical performance testing" but does not detail the number of test cases, simulated patients, or data points used. Data provenance is also not mentioned, as no clinical data or data from human subjects was used for testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This information is not applicable and hence not provided. As the testing was "non-clinical" and involved evaluating software functionality and interface, there was no need for experts to establish ground truth in the context of medical image interpretation or patient outcomes.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    This information is not applicable and hence not provided. The non-clinical performance data described does not involve expert adjudication as it pertains to software functionality and system interfaces, not diagnostic or prognostic outcomes requiring consensus.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC study was performed or is mentioned. The device is software for controlling a radiation therapy couch, not an AI for image interpretation or diagnosis that would typically be evaluated with human readers.

    6. If a standalone (i.e., algorithm only without human-in-the loop performance) was done

    Yes, based on the description, the performance testing appears to be a standalone evaluation of the software's functionality and its ability to interface with other systems and control the couch. The "non-clinical performance data" section describes testing conducted on the software itself and its interactions, without human-in-the-loop performance being a primary evaluation point for its safety and effectiveness.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The concept of "ground truth" as typically applied in AI/clinical studies (e.g., expert consensus for diagnosis, pathological findings) is not relevant here. For this device (software controlling a medical couch), the "ground truth" would be the expected functional behavior and precise physical movements of the couch, as defined by engineering specifications and safety standards. The testing verified that the software met these operational specifications.

    8. The sample size for the training set

    This information is not applicable and not provided. The Protura Couch Software is described as an interface and control system, programmed in C#. It is not an AI/machine learning model that typically requires a distinct "training set" in the sense of labeled data for learning.

    9. How the ground truth for the training set was established

    This information is not applicable and not provided, for the same reasons as in point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K120520

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-06-29

    (129 days)

    Product Code
    Regulation Number
    892.5770
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K120916

    Validate with FDA (Live)

    Date Cleared
    2012-06-20

    (85 days)

    Product Code
    Regulation Number
    892.5770
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Qfix Radiotherapy Chair is designed as a support and fixation device for patients undergoing radiotherapy treatment(s). Its design includes multiple adjustments on the device to accommodate patients of varying sizes.

    It is designed to interface with a robotic multi-axis arm by way of standard (off the shelf) robotic tooling changers. It incorporates the use of a kVue IGRT couchtop insert (K060671 The device includes a Ofix One Touch™ mechanism for kVue Couchtop Insert insertion and for the ability to use other inserts for various treatment and fixation styles.

    The Qfix Radiotherapy Chair will also be able to accommodate the use of the Qfix Accubite TM hite block stabilization system along with an array of WFR/Aquaplast masks (K935067) to assist in the fixation of the pationt during said treatment(s).

    Device Description

    The Qfix Radiotherapy Chair is designed as a support and fixation device for patients undergoing radiotherapy treatment(s). Its design includes multiple adjustments on the device to accommodate patients of varying sizes.

    It is designed to interface with a robotic multi-axis arm by way of standard (off the shelf) robotic tooling changers. It incorporates the use of a kVue IGRT couchtop insert (K060671 The device includes a Ofix One Touch™ mechanism for kVue Couchtop Insert insertion and for the ability to use other inserts for various treatment and fixation styles.

    The Qfix Radiotherapy Chair will also be able to accommodate the use of the Qfix Accubite TM hite block stabilization system along with an array of WFR/Aquaplast masks (K935067) to assist in the fixation of the pationt during said treatment(s).

    AI/ML Overview

    The provided document is an FDA 510(k) clearance letter for the "Qfix Radiotherapy Chair". It does not contain information about the acceptance criteria and study proving device performance as typically expected for software or AI/ML devices. This document is a regulatory approval for a physical medical device, not a software algorithm.

    Therefore, I cannot answer your specific questions about acceptance criteria, sample sizes, ground truth establishment, or human-in-the-loop performance, as this information is not present in the provided text. The document primarily focuses on FDA's determination of substantial equivalence for a physical radiotherapy support and fixation device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K072898

    Validate with FDA (Live)

    Date Cleared
    2007-10-25

    (14 days)

    Product Code
    Regulation Number
    892.5770
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The intended use of the device is to support and aid in positioning a patient during radiation therapy.

    Device Description

    The HexaPOD evo consists of two platforms, which are connected by six length adjustable elements which are powered. By appropriate coordinative adjustment of these elements, the system is able to move the upper platform relative to the lower one. The movement can occur in all three dimensions in space. Additionally the upper platform can rotate around these three axes which results in a tilt or a rotation of the upper platform relative to the lower one. Finally an accurate positioning within all six degrees of freedom (6DOF) can be provided. The HexaPOD consists of a controller unit which is directed by a cable connected hand control. Additionally it can be directed via an external graphics user interface (GUI) which is installed on a PC.

    AI/ML Overview

    The provided text is a 510(k) summary for the HexaPOD™ evo RT Couch Top, which is a powered radiation therapy patient support assembly. It describes the device, its intended use, and its substantial equivalence to a predicate device.

    However, the document explicitly states: "No performance data is required for this Class II device nor requested by the Food and Drug Administration (Office of Device Evaluation)."

    Therefore, I cannot provide a table of acceptance criteria and reported device performance or information about a study proving the device meets acceptance criteria, as such data was not required or provided in this 510(k) submission.

    Here's a breakdown of why the specific questions cannot be answered based on the input:

    • 1. A table of acceptance criteria and the reported device performance: Not applicable. The document states no performance data was required or requested.
    • 2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective): Not applicable. No performance data or testing on a test set was mentioned.
    • 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience): Not applicable. No performance data or ground truth establishment was mentioned.
    • 4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable. No performance data or adjudication was mentioned.
    • 5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. The device is a physical patient support assembly, not an AI or imaging interpretation device. MRMC studies are not relevant.
    • 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. The device is a physical patient support assembly, not an algorithm.
    • 7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable. No performance data or ground truth establishment was mentioned.
    • 8. The sample size for the training set: Not applicable. As no performance data from a study was required, there's no mention of a training set.
    • 9. How the ground truth for the training set was established: Not applicable. No training set or ground truth establishment was mentioned.

    In summary, the 510(k) submission for the HexaPOD™ evo RT Couch Top focuses on its substantial equivalence to a predicate device due to minor hardware changes, and explicitly states that no performance data was required or requested.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 3