K Number
K160518

Validate with FDA (Live)

Date Cleared
2016-07-14

(141 days)

Product Code
Regulation Number
892.5770
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The LEONI Orion System in an electro-mechanical robotic arm for patient positioning in radiotherapy and medical imaging. It is designed for positioning a patient with a high degree of accuracy and repeatability.

Device Description

The LEONI ORION System is an electro-mechanical robotic arm capable of motion in six degrees of freedom. The purpose of the device is to position a patient during radiotherapy, radiology and other medical applications with a high degree of accuracy and repeatability.
The LEONI ORION System consists of the electro-mechanical unit that is a 6 axes robot which supports a standard radiotherapy table couch or other approved patient support device, and a Control Unit that includes computers and application software. The robot is linked to the Control Unit by cables.

AI/ML Overview

The provided text is a 510(k) summary for the LEONI ORION System, an electro-mechanical robotic arm for patient positioning in radiotherapy and medical imaging. It focuses on demonstrating substantial equivalence to a predicate device, rather than a clinical study evaluating diagnostic performance. Therefore, many of the requested items related to AI device evaluation (like MRMC studies, ground truth establishment for training data, and expert qualifications) are not applicable or cannot be extracted from this document.

Here's an analysis based on the information available:

1. Table of Acceptance Criteria and Reported Device Performance

Feature/MetricAcceptance CriteriaReported Device Performance
Payload CapacityNot explicitly stated as "acceptance criteria" but tested.375 kg / 826 lbs
Accurate Treatment VolumeNot explicitly stated as "acceptance criteria" but tested.400 mm x 1000 mm x 500 mm
AccuracyNot explicitly stated as "acceptance criteria" but tested.± 0.5 mm and ± 0.2°
SpeedNot explicitly stated as "acceptance criteria" but tested.0.1 m/s and 6 °/s
Safety (Collision Detection)Not explicitly stated as "acceptance criteria" but tested (detection of a 150N force).Detection of a 150N force
Safety ("Overtravel" in emergency stop)Not explicitly stated as "acceptance criteria" but tested (< 5 mm).< 5 mm
EMC (Electromagnetic Compatibility)(Implied standard compliance)Tests performed by third party
Electrical and Mechanical Safety(Implied standard compliance)Tests performed by third party

2. Sample size used for the test set and the data provenance

The document describes bench tests rather than a "test set" in the context of diagnostic data. The tests were performed by LEONI, and some by a third-party test organization. The "data provenance" would refer to these test environments rather than patient data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. This device is an electro-mechanical robotic arm, and the "ground truth" for its performance is determined by physical measurements and test procedures, not expert interpretation of medical images or conditions.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. This concept applies to human reader studies, not to the performance testing of a robotic patient positioning system.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No MRMC study was done. This document does not describe an AI diagnostic device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This document describes the performance of the device itself (standalone in terms of its mechanical and electrical function). The "algorithm only" concept is typically for diagnostic AI, which is not what this device is.

7. The type of ground truth used

The ground truth for the performance metrics (payload, accuracy, speed, safety) are based on physical measurements and engineering standards/specifications inherent to the design and function of a robotic arm, not medical outcomes, pathology, or expert consensus on medical conditions.

8. The sample size for the training set

Not applicable. This device does not use a "training set" in the machine learning sense. Its performance is based on engineering design, calibration, and physical testing.

9. How the ground truth for the training set was established

Not applicable, for the same reason as above.

Summary of the Study that Proves Acceptance Criteria:

The study proving the device meets its acceptance criteria involved a series of non-clinical bench tests. These tests were conducted by LEONI and third-party test organizations. The tests evaluated the physical and functional characteristics of the LEONI ORION System, including:

  • Payload Capacity: Verified the maximum weight the system can support.
  • Accurate Treatment Volume: Demonstrated the spatial envelope within which the specified accuracy can be maintained.
  • Accuracy: Measured the precision of positioning (± 0.5 mm and ± 0.2°).
  • Speed: Measured the maximum linear and rotational movement speeds (0.1 m/s and 6 °/s).
  • Safety (Collision Detection): Tested the system's ability to detect a 150N force, indicating its safety mechanism.
  • Safety ("Overtravel" in Emergency Stop): Verified that the system stops within a very short distance (< 5 mm) after an emergency stop, ensuring patient safety.
  • Electromagnetic Compatibility (EMC): Confirmed compliance with relevant standards to prevent interference with other medical equipment.
  • Electrical and Mechanical Safety: Ensured the device meets safety standards for electrical operation and mechanical integrity.

The document states: "The results from these performance assessments demonstrated that the LEONI ORION System met the acceptance criteria defined in the product specifications. Moreover these results proved that the subject device is comparable to the predicate device in terms of safety and effectiveness performance."

Therefore, the "study" was a series of engineering and safety tests designed to validate the mechanical and electrical specifications of the robotic positioning system, rather than a clinical study or an AI performance evaluation.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo features a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 14, 2016

LEONI CIA Cable System SAS % Ms. Aurelie Genho Quality Engineer 5 Avenue Victor Hugo 28000 Chartres FRANCE

Re: K160518

Trade/Device Name: LEONI Orion System Regulation Number: 21 CFR 892.5770 Regulation Name: Powered radiation therapy patient support assembly Regulatory Class: II Product Code: JAI Dated: July 5, 2016 Received: July 7, 2016

Dear Ms. Genho:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Michael D'Hara

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

For

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K160518

Device Name LEONI Orion System

Indications for Use (Describe)

The LEONI Orion System in an electro-mechanical robotic arm for patient positioning in radiotherapy and medical imaging. It is designed for positioning a patient with a high degree of accuracy and repeatability.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

005 - 510(K) Summary

LEONI ORION System

LEONI

Traditional 510(k) Notification

510 (K) SUMMARY

This summary of 510 (k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

ApplicantLEONI CIA Cable Systems SAS5, avenue Victor HugoZAC du Jardin d'Entreprises28000 CHARTRESFrance
Contact personMrs. Aurélie GENHOPh : +33 2 37 91 21 82Fax : +33 2 37 91 21 81
Preparation DateJanuary 29, 2016
Device NameCommon Name: LEONI ORION System
Trade Name: LEONI ORION System
Classification NamePowered radiation therapy patient support assembly (21 CFR 892.5770)
Product Code: JAI
Substantial EquivalencePatient Positioning System by Forte Automation Systems (K122413)
Device DescriptionThe LEONI ORION System is an electro-mechanical robotic arm capable ofmotion in six degrees of freedom. The purpose of the device is to position apatient during radiotherapy, radiology and other medical applications with ahigh degree of accuracy and repeatability.The LEONI ORION System consists of the electro-mechanical unit that is a 6axes robot which supports a standard radiotherapy table couch or otherapproved patient support device, and a Control Unit that includes computersand application software. The robot is linked to the Control Unit by cables.
Intended UseThe LEONI ORION System is an electro-mechanical robotic arm for patientpositioning in radiotherapy, radiology and other medical applications. It isdesigned for positioning a patient with a high degree of accuracy andrepeatability.

{4}------------------------------------------------

LEONI

005 - 510(K) Summary

LEONI ORION System

Traditional 510(k) Notification

Comparison of Technological Characteristics with the Predicate DeviceThe LEONI ORION System is identical or similar to the predicate device in regard to:Intended Use Shape, dimensions and materials Behavior and technology of movement Safety features
The LEONI ORION System has the following technological differences compared with the predicate device:Rotational axis linked to the floor Greater range of motion thanks to its hybrid kinematic and the length of the axes A higher payload capacity A UPS battery that allows the slow release of the brakes in order to lower the couch into a safe position to remove the patient in case of power failure.
Non clinical TestsThe bench tests below were carried out by LEONI in support of the substantial equivalence determination:Payload: 375 kg / 826 lbs. Accurate treatment volume : 400 mm x 1000 mm x 500 mm Accuracy: ± 0.5 mm and ± 0.2° Speed: 0.1 m/s and 6 °/s Safety: collision detection: detection of a 150N force; "overtravel" in case of emergency stop: < 5 mm. Moreover, electromagnetic compatibility (EMC) and electrical and mechanical safety tests were performed by third party test organization.
Clinical TestsNot Applicable
Testing ConclusionsThe results from these performance assessments demonstrated that the LEONI ORION System met the acceptance criteria defined in the product specifications. Moreover these results proved that the subject device is comparable to the predicate device in terms of safety and effectiveness performance.

§ 892.5770 Powered radiation therapy patient support assembly.

(a)
Identification. A powered radiation therapy patient support assembly is an electrically powered adjustable couch intended to support a patient during radiation therapy.(b)
Classification. Class II.