K Number
K122201

Validate with FDA (Live)

Date Cleared
2012-08-30

(36 days)

Product Code
Regulation Number
892.5770
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Protura Couch Software is interface between a record and verify system, linear accelerator (Linac) software, Linac safeguard systems, and/or image guidance systems and the Protura Couch. The Protura Couch Software is also capable of operating the Protura Couch (6 Degree Axis Couch).

Device Description

The proposed device, the Protura Couch Software (MT6XSMEU), is patient positioning software used in radiation treatments in conjunction with an Image Guided Radiation Therapy (IGRT) system. The Protura Couch Software interfaces with the Protura Couch Hardware (K031866) and is intended to position the patient after diagnostic decisions have been made based on results from an IGRT system. The Protura Couch Software allows the user to control patient positioning with six degrees of freedom from outside of the treatment room. The Protura Couch Software also includes the ability to interface with a record and verify system, Linac software, Linac safeguard systems, and/or image guidance systems.

AI/ML Overview

The provided text describes a 510(k) summary for the Protura Couch Software (device name: MT6XSMEU) by MED-TEC, Inc. d/b/a CIVCO Medical Solutions. This software is designed for patient positioning in radiation therapy.

Here's an analysis of the acceptance criteria and study information, based solely on the provided document:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly state quantitative acceptance criteria in a table format. Instead, it describes performance testing conducted. The reported device performance is that "All testing confirmed that the Protura Couch Software is safe and effective for its intended use" and that it is "substantially equivalent to the predicate device in regards to safety, effectiveness, and performance."

The characteristics tested include:

  • Movement of the Protura Couch
  • Interfacing with External Systems
  • Couch Pedestal and Isocenter Alignment
  • Patient Record Handling

Without specific numerical targets for these characteristics, a precise table of acceptance criteria and reported performance cannot be generated. The document indicates these tests were performed and passed.

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

The document does not specify a sample size for the test set. It mentions "non-clinical performance testing" but does not detail the number of test cases, simulated patients, or data points used. Data provenance is also not mentioned, as no clinical data or data from human subjects was used for testing.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

This information is not applicable and hence not provided. As the testing was "non-clinical" and involved evaluating software functionality and interface, there was no need for experts to establish ground truth in the context of medical image interpretation or patient outcomes.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

This information is not applicable and hence not provided. The non-clinical performance data described does not involve expert adjudication as it pertains to software functionality and system interfaces, not diagnostic or prognostic outcomes requiring consensus.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No MRMC study was performed or is mentioned. The device is software for controlling a radiation therapy couch, not an AI for image interpretation or diagnosis that would typically be evaluated with human readers.

6. If a standalone (i.e., algorithm only without human-in-the loop performance) was done

Yes, based on the description, the performance testing appears to be a standalone evaluation of the software's functionality and its ability to interface with other systems and control the couch. The "non-clinical performance data" section describes testing conducted on the software itself and its interactions, without human-in-the-loop performance being a primary evaluation point for its safety and effectiveness.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The concept of "ground truth" as typically applied in AI/clinical studies (e.g., expert consensus for diagnosis, pathological findings) is not relevant here. For this device (software controlling a medical couch), the "ground truth" would be the expected functional behavior and precise physical movements of the couch, as defined by engineering specifications and safety standards. The testing verified that the software met these operational specifications.

8. The sample size for the training set

This information is not applicable and not provided. The Protura Couch Software is described as an interface and control system, programmed in C#. It is not an AI/machine learning model that typically requires a distinct "training set" in the sense of labeled data for learning.

9. How the ground truth for the training set was established

This information is not applicable and not provided, for the same reasons as in point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for CIVCO Medical Solutions. The logo features a stylized globe on the left, followed by the text "CIVCO" in bold, sans-serif font. Below the company name, the words "MEDICAL SOLUTIONS" are printed in a smaller, sans-serif font, with spacing between the letters. The overall design is clean and professional.

WWW.CIVCO.COM

AUG 30 2012

Section 5 - 510(k) Summary

A. Submitter InformationSubmitter Name & Address: MED-TEC, Inc. d/b/a CIVCO Medical Solutions1401 8th Street SEOrange City, lowa 51041
Contact Person:Amanda Stahle, Regulatory Affairs SpecialistTelephone: 319-248-6628, Fax: 877-218-0324amanda.stahle@civco.com
Date Summary Prepared:July 10, 2012
Trade Name:Common Name:Classification Name:Classification Number:Product Code:Protura Couch SoftwareTreatment Couch SoftwarePowered radiation therapy patient support assembly892.5770JAI

B. Predicate Device

MED-TEC, Inc. d/b/a CIVCO Medical Solutions claims the proposed device to be substantially equivalent to the following device:

510(k) NumberDevice NameProductClassification/CodeSubmitterName
K034054PRO SERIES COUCHSOFTWARE, MODEL MT-PRO100892.5770/JAIMED-TEC, Inc.

Both the proposed device and the predicate device are patient positioning software systems that interface with the Couch Hardware (K031866). Six degrees of freedom patient positioning corrections are sent from the software to the Couch Hardware for implementation. The predicate device also interfaces with a treatment planning system whereas the proposed device interfaces with a record and verify system, linear accelerator (Linac) software, Linac safeguard systems, and/or image guidance systems. Testing has demonstrated that the proposed device is substantially equivalent to the predicate device in regards to safety, effectiveness, and performance, and the additional interfaces provided by the proposed device did not diminish the safety or effectiveness of the device.

Corporate Headquarters102 First Street SouthKalona, IA 52247USA0 319.248.67573 319.248.6660
Coralville Office2301 Jones Blvd.Coralville, IA 52241USA0 319.248.67573 319.248.6660
Europe OfficePasteurstraat 62811 DX ReeuwijkThe Netherlands0 +31(0) 182.3944951 +31(0) 182.395014
Orange City Office1401 8th Street SEOrange City, IA 51041USA1 712.737.86889 712.737.8654

{1}------------------------------------------------

WWW.CIVCO.COM

C. Device Description

The proposed device, the Protura Couch Software (MT6XSMEU), is patient positioning software used in radiation treatments in conjunction with an Image Guided Radiation Therapy (IGRT) system. The Protura Couch Software interfaces with the Protura Couch Hardware (K031866) and is intended to position the patient after diagnostic decisions have been made based on results from an IGRT system. The Protura Couch Software allows the user to control patient positioning with six degrees of freedom from outside of the treatment room. The Protura Couch Software also includes the ability to interface with a record and verify system, Linac software, Linac safeguard systems, and/or image guidance systems.

D. Indications for Use/Intended Use

The Protura Couch Software is intended to interface between a record and verify system. linear accelerator (Linac) software, Linac safeguard systems, and/or image guidance systems and the Protura Couch. The Protura Couch Software is also capable of operating the Protura Couch (6 Degree Axis Couch).

E. Technological Characteristics

Both the Protura Couch Software and the predicate device are interface software systems that are intended to operate the Couch Hardware (K031866). The predicate device interfaces with a treatment planning system whereas the Protura Couch Software interfaces with a record and verify system, Linac software, Linac safeguard systems, and/or image guidance systems. The Protura Couch Software is programmed in C#; the predicate device is programmed in C, C#, C++, and Java. The Protura Couch Software runs on Windows XP SP3 (32 bit), Windows 7 (32 bit), and Windows 7 (64 bit); the predicate runs on Windows 2000.

F. Non-Clinical Performance Data

Non-clinical performance testing was conducted for the following characteristics:

  • Movement of the Protura Couch .
  • Interfacing with External Systems .
  • . Couch Pedestal and Isocenter Alignment
  • . Patient Record Handling

All testing confirmed that the Protura Couch Software is safe and effective for its intended use.

G. Clinical Performance Data

No clinical testing was performed in the evaluation of this medical device.

H. Non-Clinical and Clinical Performance Data Conclusions

The conclusions drawn from the tests are that the Protura Couch Software is substantially equivalent to the predicate device in regards to safety, effectiveness, and performance, and the additional interfaces have not diminished the safety or effectiveness of the device.

Corporate Headquarters102 First Street SouthKalona, IA 52247USAⓟ 319.248.6757Ⓕ 319.248.6660
Coralville Office2301 Jones Blvd.Coralville, IA 52241USAⓟ 319.248.6757Ⓕ 319.248.6660
Europe OfficePasteurstraat 62811 DX ReeuwijkThe Netherlandsⓟ +31(0) 182.394495Ⓕ +31(0) 182.395014
Orange City Office1401 8th Street SEOrange City, IA 51041USAⓟ 712.737.8688Ⓕ 712.737.8654

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized image of an eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" arranged in a circular fashion around the eagle. The eagle is depicted in a simple, black and white design.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002

Ms.` Amanda Stahle Regulatory Affairs Specialist MED-TEC, Inc. d/b/a CIVCO Medical Solutions 1401 8th Street SE ORANGE CITY IA 51041

AUG 30 2012

Re: K122201

Trade/Device Name: Protura Couch Software Regulation Number: 21 CFR 892.5770 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: JAI Dated: July 17, 2012 Received: July 25, 2012

Dear Ms. Stahle:

We have reviewed your Section 510(k) premarket notification of intent to market the indication We nave reviewed your Section 910(t) premains is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the Electosule for tegarry manation that of the Medical Device Ameral Food Truc commerce prior to May 26, 1977, the excordance with the provisions of the Federal Food. Drug, devices that have been reciassified in accordines while premarket approval application (PMA). and Cosmetic Act (Act) that to not require approval controls provisions of the Act. The Act. The Act. The Act. The Act. The Act. The Act. The Act. The You may, therefore, market the device, saojeet to decidents for annual registration, listing of
general controls provisions of the Act include requirements for annual regist general controls provisions of the receives and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such If your device is classified (Scc above) into eason affecting your device can be found in Title 21, 2011 additional controls. Existing major regulations a 195. In addition, FDA may publish further
Code of Federal Regulations (CFR), Parts 800 to 1995. In addition, FDA may publ Code of Poderal regarning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that FDA s issuality of a succession with other requirements of the Act
that FDA has made a determination that your devices associes. You must that FDA has made a decemmation that your corred by other Federal agencies. You must or any Federal statutes and regulations and including, but not to: registing (21)
comply with all the Act's requirements, including, but not includes areasting (reporting of comply with all the Act s requirements, therauling, our not many of the one of

{3}------------------------------------------------

medical device-related adverse events) (21 CFR 803); and good manufacturing practice mediem as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter requirences as bet form inrketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed nonicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 11 you desire possible ad res to: Jon Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office

of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Four may of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely Yours,

Janine M. Morris

Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known):

Device Name: Protura Couch Software

Indications for Use: The Protura Couch Software is interface between a record and verify system, linear accelerator (Linac) software, Linac safeguard systems, and/or image guidance systems and the Protura Couch. The Protura Couch Software is also capable of operating the Protura Couch (6 Degree Axis Couch).

Prescription Use × (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Arland D. Johns

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) KLaaaøl

Page 1 of

§ 892.5770 Powered radiation therapy patient support assembly.

(a)
Identification. A powered radiation therapy patient support assembly is an electrically powered adjustable couch intended to support a patient during radiation therapy.(b)
Classification. Class II.