Search Results
Found 30 results
510(k) Data Aggregation
(118 days)
Ask a specific question about this device
(196 days)
The Allergan Botox Needle Electrode are for SINGLE USE ONLY in electromyography (EMG) in situations wherein it is desired to insert an electrode, in the form of a probe, into a patient to locate a particular muscle and then inject a medication into that muscle. The hypodermic needle with an open lumen is designed for muscle stimulation, motor unit action potential recording and drug delivery. Once the physician is satisfied with the location, he/she injects a drug therein via the lumen of the needle.
The Allergan Botox Needle Electrode is used in EMG applications to record EMG activity while allowing injection of medication. The Allergan Botox Needle Electrode is a sterile, disposable, monopolar needle electrode with attached lead wire. The needle shaft is insulated so the needle records signals received only from the tip. The needle hub is designed to allow for precise manipulation of the electrode.
The provided text describes a 510(k) premarket notification for the "Allergan Botox Needle Electrode." This document focuses on demonstrating substantial equivalence to a predicate device (K973444 Teca Myoject Disposable Needle Electrode) through non-clinical performance data. Therefore, the "acceptance criteria" and "study that proves the device meets the acceptance criteria" refer to the non-clinical performance tests conducted to show the device's functional equivalence and safety/effectiveness compared to its predicate.
Here's a breakdown of the requested information based on the provided document:
1. A table of acceptance criteria and the reported device performance
| Test | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Dimensional measurement of Needle Electrode – Needle Length | Needle length should be 37.0 ± 1.5mm. | Pass |
| Dimensional measurement of Needle Electrode - Needle Diameter with and without coating | The diameter of the Needle (without PTFE coat) Electrode should be of 27 Gauge (0.40± 0.1mm). The diameter of the Needle with PTFE coating should be between 0.420 to 0.456 mm. | Pass |
| Device Continuity measurement | The resistance between needle tip to the standard 1.5mm connector end should be less than 1.5Ω to allow the biopotential signals. | Pass |
| Biocompatibility Evaluation | Biocompatibility testing per ISO 10993-1, classifying it as an Invasive Electrode, external communicating device in limited (≤ 24 hours) contact with tissue. Tests considered: cytotoxicity (ISO 10993-5), sensitization (ISO 10993-10), irritation (ISO 10993-10), and Blood Compatibility Hemolysis Test (ISO 10993-4). | Pass |
| Tensile testing | Acceptance Criteria: $>/=22N$ (Needle to Wire Crimp Pull Test). | Pass |
| Penetration testing | Comparative data (Verify that the needle consistently produces the comparative sharpness and frictional force). | Pass |
| Leak Testing | The resistance of the needle should be greater than $10MΩ$ (measured across the needle with probe on the PTFE coating). | Pass |
| Tip geometry measurements | The tip geometry of the Product should be compliant with BS EN ISO 7864:2016 standard (Verified by certificate). | Pass |
| Luer Lock Design | The Luer lock and Luer slip fit syringes should be conform to ISO 594-1 & 594-2 standard (Compliance of certificates for Luer lock and Luer slip fit syringes). | Pass |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document does not specify the exact sample sizes (number of units tested) for each non-clinical performance test. It only states "Pass" for each test.
The data provenance is from Natus Manufacturing Limited, Ireland, the manufacturer submitting the 510(k). The tests are non-clinical performance tests, typically conducted in laboratory settings.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This is a non-clinical performance study of a physical device (needle electrode). The "ground truth" for these tests is established by engineering and quality control standards (e.g., precise measurements, material specifications, mechanical strength tests, and biocompatibility standards like ISO 10993). Therefore, human expert review in the sense of clinical image interpretation or diagnosis is not applicable. The expertise involved would be in manufacturing, quality assurance, materials science, and biomedical engineering.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. This is not a clinical study involving human interpretation or subjective assessment that would require adjudication. The tests are objective measurements and compliance checks against established engineering and safety standards.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This 510(k) submission is for a medical device (a needle electrode), not an AI-powered diagnostic or assistive tool that would involve human readers. The study performed is a non-clinical performance comparison to a predicate device, focusing on physical and functional characteristics.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is for a physical medical device, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The "ground truth" for this device's performance relies on established engineering standards, material specifications, and regulatory requirements (e.g., ISO standards) for medical devices. For instance, needle length must fall within a specified range, resistance must be below a certain threshold, and materials must demonstrate biocompatibility through standardized tests. It's objective performance against predefined technical criteria rather than subjective clinical "ground truth."
8. The sample size for the training set
Not applicable. This is a 510(k) submission for a physical device, not an AI model requiring a training set.
9. How the ground truth for the training set was established
Not applicable, as there is no training set for a physical medical device.
Ask a specific question about this device
(283 days)
The disposable hypodermic monopolar needles are FOR SINGLE USE ONLY in electromyography (EMG) in situations wherein it is desired to insert an electrode, in the form of a probe, into a patient to locate a particular muscle and then inject a medication into that muscle. The hypodermic needle with an open lumen is designed for muscle stimulation, motor unit action potential recording and drug delivery. Once the physician is satisfied with the location, he/she injects a drug therein via the lumen of the needle.
The Myoject™ Luer Lock Needle Electrode is a Disposable Hypodermic Needle Electrode
Here's a breakdown of the acceptance criteria and the study information for the Natus Manufacturing Limited Myoject™ Luer Lock Needle Electrode, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance:
The document primarily focuses on demonstrating substantial equivalence to a predicate device (Teca Myoject Disposable Needle Electrode K973444) through a comparison of characteristics and performance testing. There isn't a direct "acceptance criteria" table with specific pass/fail metrics explicitly stated alongside the new device's performance in a quantitative manner for most characteristics. Instead, the "Discussion of Differences" column in Table 5A (starting on page 5) serves as the primary mechanism for demonstrating equivalence and addressing performance.
However, for Biocompatibility and Bench Testing, the performance is reported against established ISO and ASTM standards.
| Characteristic / Test | Acceptance Criteria (Implicit from Predicate/Standards) | Myoject™ Luer Lock Needle Electrode Performance (Reported) |
|---|---|---|
| Biocompatibility | ||
| Cytotoxicity | Non-cytotoxic (ISO 10993-5:2009) | Non-cytotoxic to L929 cells |
| Irritation (Intracutaneous) | Average reaction not greater than control (ISO 10993-10:2010) | Meets requirements; average reaction not greater than control for polar and non-polar extracts |
| Sensitization (Local Lymph Node Assay) | Non-sensitizer (ISO 10993-10:2010) | Polar and non-polar extracts considered non-sensitizers. Positive controls gave SI > 3 (as expected for positive control function). |
| Systemic Toxicity | PASS (ISO 10993-11:2006) | 0.9% sodium chloride solution extract: PASS; Cottonseed oil extract: PASS |
| Bench Testing | ||
| Tensile testing | Implied to meet design specifications | Conducted; meets performance and design specifications |
| Coating thickness | Implied to meet design specifications | Conducted; meets performance and design specifications |
| Leak Testing | Implied to meet design specifications | Conducted; meets performance and design specifications |
| Tip geometry grind measurements | Implied to meet design specifications | Conducted; meets performance and design specifications |
| Penetration testing | Reduced initial penetration force compared to predicate (discussed as an improvement) | Conducted; meets performance and design specifications. Implied improved penetration. |
| Luer Lock Design Verification | Secure Luer Lock connection (reduced seepage/leakage) | Conducted; meets performance and design specifications. Luer lock is described as adding an additional safety mechanism. |
| Other Comparisons (from Table 5A) | ||
| Labelling | Equivalent, with minor changes not affecting safety/effectiveness | Minor changes (manufacturer name, address, branding, warnings, harmonized symbols); do not raise questions of safety/effectiveness. |
| Intended Use | Same as predicate | Same as predicate; minor labeling nuance regarding drug supply does not affect safety/effectiveness. |
| Physical Characteristics | Monopolar Needle | Same |
| Connector | Single Contact Touch Proof DIN42 802 | Same |
| Material of Needle | Stainless 304 | Same |
| Needle Diameter (mm) | 0.30, 0.41, 0.46, 0.51, 0.71 | Same |
| Needle Coating Material | PTFE (different formulation) | PTFE (Xylan 8820 G4075); improved to reduce initial penetration force, similar safety/effectiveness. |
| Hub | Luer Slip (predicate) | Luer Lock/Slip; improved to reduce risk of medication leakage. Doesn't affect overall function/safety/effectiveness. |
| Hub Material | Polycarbonate (predicate) | Polypropylene (Exxonmobil) Escorene PP9074 MED; equivalent material performance, no new safety/effectiveness concerns. |
| Needle grind angle | 10-20 degrees conical (predicate) | 10-16 degrees conical; slight change in upper range, not significant, no new safety/effectiveness concerns, no altered diagnostic effects. |
| Recording area (sqmm) | 0.25 (predicate) | 0.19 (fine) to 0.89 (depending on size); differences vary, no effect on safety/effectiveness or diagnostic effects. |
| Length (mm) | 25, 37, 50, 75, 100 (predicate) | 25, 37, 50, 75; 100mm version obsolete. No new safety/effectiveness concerns. |
| Connecting lead | 14 or 26 inches, 2mm or touchproof plug (predicate) | 770±10mm, 'ST' moulded 1.5mm touch proof. Cable length "30" (76cm) to 1.5mm female touchproof connector; increased length, minor changes, no new safety/effectiveness concerns. |
| Sterilization | Supplied sterilized gamma irradiated | Same |
| Protective Sheath | Polyethylene (predicate) | Polypropylene Escorene PP9074 MED (Exxon Mobil); equivalent material performance, no new safety/effectiveness concerns. |
| Protective Pouch | Tyvek/Mylar | Same |
| Sterility Assurance Level (SAL) | 10 minus 6 (predicate) | 10 minus 6; improved needle tip sharpness. No new safety/effectiveness concerns. |
| Tip geometry | Lancet Point (predicate) | Trocar Point, 15 degree bevel; preferred for range of needles, improves tip sharpness, reduces patient discomfort. |
2. Sample Size Used for the Test Set and Data Provenance:
- Sample Size for Test Set: The document does not explicitly state specific sample sizes for each of the performance tests (tensile, coating thickness, leak, tip geometry, penetration, Luer Lock verification). It refers to "Performance Testing" and "Bench Testing" being performed. For biocompatibility tests, details are provided on the methods (e.g., L929 cells for cytotoxicity, rabbit for intracutaneous reactivity, mouse for local lymph node assay), but the exact number of test items or animals used is not specified in this summary.
- Data Provenance: The studies were conducted by Natus Manufacturing Limited, which is based in Ireland. The nature of these tests (biocompatibility, bench testing) usually implies controlled laboratory settings rather than patient data. Therefore, the data Provenance is the manufacturer's internal testing. Not retrospective or prospective patient data.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:
- This is not applicable as the studies described are non-clinical performance tests (biocompatibility and bench testing) and technical comparisons to a predicate device. The "ground truth" is established by adherence to recognized international standards (ISO, ASTM) for materials, design, and manufacturing quality, not expert consensus on clinical cases.
4. Adjudication Method for the Test Set:
- Not applicable, as this refers to non-clinical performance and design verification tests, not clinical studies with human assessors.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:
- No, an MRMC comparative effectiveness study was not done. This device is a diagnostic electromyograph needle electrode, not an image-based diagnostic AI device that would typically involve an MRMC study. The comparison is primarily against a predicate device's design specifications and performance standards through bench testing.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:
- Not applicable. This device is a medical instrument (needle electrode), not an algorithm or AI system.
7. The Type of Ground Truth Used:
- The "ground truth" for demonstrating the device's acceptable performance and substantial equivalence is based on:
- Adherence to International Standards: ISO 10993 series (Biocompatibility), ISO 2859-1 (Sampling), ASTM F 1980 (Sterile Barrier).
- Predicate Device Specifications: The Teca Myoject Disposable Needle Electrode (K973444) serves as the primary benchmark for comparison of design and technological characteristics.
- Design and Performance Specifications: The device's own predetermined specifications for physical characteristics and functional performance (e.g., tensile strength, leak integrity, penetration force).
8. The Sample Size for the Training Set:
- Not applicable. There is no training set mentioned, as this is a physical medical device, not an AI or machine learning model.
9. How the Ground Truth for the Training Set Was Established:
- Not applicable, as there is no training set.
Ask a specific question about this device
(238 days)
Disposable Concentric Needle electrodes are sterile electrodes indicated for recording muscle activity for Electromyography (EMG) applications. The electrodes are for single patient use only.
Disposable Monopolar Needle electrodes / EP Needle electrodes indicated for recording muscle activity for Electromyography (EMG) and/or Electroencephalography (EEG) applications. The electrodes are for single patient use only.
Disposable Hypodermic Needle electrodes indicated for injection of Botulium Toxin while recording muscle activity with Electromyography (EMG) applications. The electrodes are for single patient use only.
The Bio Protech Disposable Concentric Needle Electrode is used for electromyography (EMG) recording for examination of the peripheral neuromuscular system, by registration of the electrical activity from the muscles. Bio Protech Disposable Concentric Needle Electrode consists of a stainless steel cannula and an inner conductor of silver. Between these two conductors there is an insulation layer. The inner conductor is the active measure point and the outer conductor of stainless steel is the reference point. The Stainless steel cannula is coated with a low friction lubricant.
The Bio Protech Disposable Monopolar / EP Needle Electrode is intended to be inserted in the subdermal, muscle or nerve tissue for use with recording equipment for the recording of biopotentials signals, EEG or EMG, and proximally connected to electromyography / Electroencephalogram recording equipment. The electrodes consist of a formed stainless steel needle with a lead wire attached. The wire can be directly connected or removable and terminates in a safety connector that cannot be connected to an AC power outlet or and cannot get in touch with possible hazardous voltage.
The Bio Protech Disposable Hypodermic Needle Electrode is intended to be inserted in the muscle while recording electromyography activity, and proximally connected to electromyography recording equipment. The electrodes consist of an hypodermic stainless steel needle with an open lumen and a lead wire attached. The wire can be directly connected or removable and terminates in a safety connector that cannot be connected to an AC power outlet or and cannot get in touch with possible hazardous voltage.
This document describes the premarket notification for Bio Protech, Inc.'s Disposable Concentric Needle Electrodes, Disposable Monopolar Needle Electrodes, Disposable EP Needle Electrodes, and Disposable Hypodermic Needle Electrodes. The submission seeks to demonstrate substantial equivalence to legally marketed predicate devices.
Acceptance Criteria and Device Performance:
The document does not explicitly present a table of acceptance criteria with corresponding performance metrics in a pass/fail format. Instead, it relies on demonstrating similar performance characteristics to predicate devices and adherence to relevant standards. The "Substantial Equivalence Chart" (Pages 5-7) directly compares the new device with predicate devices across various characteristics.
For "Performance", the report states:
- Acceptance Criteria (Implied): Equivalent penetration and friction force to predicate devices, and electrical properties conforming to DIN 13097. Assurance of functionality during the shelf life through aging tests. Optimal bevel and burr for sharpening, and electrical continuity and isolation of all poles.
- Reported Device Performance:
- Disposable Concentric Needle electrodes: "Tested for penetration and friction force and electrical properties (according to DIN 13097). Ageing tests are performed to verify and ensure the functionality during the shelf life of the product."
- Disposable Hypodermic Needle electrodes (under the "Performance" row for Concentric, but appears to relate to Hypodermic based on context within the table): "Sharpening; Camera visual examination with special attention to bevel and burrs; Electrical continuity and isolation of all poles;"
- Disposable Monopolar Needle electrodes / EP Needle electrodes: "Sharpening; Camera visual examination with special attention to bevel and burrs; Electrical continuity and isolation of all poles;"
The overall conclusion states: "The result of testing shows that the electrodes are as safe, effective and perform at least as the legally marketed device."
Summary of Studies and Information:
- Table of Acceptance Criteria and Reported Device Performance:
| Feature/Test | Acceptance Criteria (Implied/Predicate Match) | Reported Device Performance (Bio Protech Needle Electrodes) |
|---|---|---|
| Penetration and Friction Force | Equivalent to predicate devices (according to DIN 13097 for Concentric/Hypodermic, unknown for Monopolar/EP as predicate performance is "Unknown") | Tested for penetration and friction force according to DIN 13097. Results show the electrodes are "as safe, effective and perform at least as the legally marketed device." |
| Electrical Properties | Conforms to DIN 13097 (for Concentric). Electrical continuity and isolation of all poles (for Hypodermic/Monopolar/EP). | Tested for electrical properties (according to DIN 13097) for Concentric. Sharpening; Camera visual examination with special attention to bevel and burrs; Electrical continuity and isolation of all poles for Hypodermic/Monopolar/EP. |
| Shelf-Life | Functionality during shelf life (implied by predicate's aging tests). Specific duration for predicates: 60 months (BIONEN Concentric), Unknown (Technomed, Axon). | Accelerated aging testing conducted according to ASTM F 1980. Test results indicate a shelf-life of 3 years. |
| Sterility | EO sterilization or Gamma irradiation (predicate dependent). | EO sterilization. Passed EO residual testing per ISO 10993-7: 2008. |
| Biocompatibility | Appropriate levels of biocompatibility (based on ISO 10993-1). | Evaluations performed for cytotoxicity (ISO MEM elution), intracutaneous reactivity, Guinea pig maximization sensitization, acute systemic toxicity, material-mediated pyrogenicity (rabbit), and hemolysis (based on ISO 10993-1). |
| Electrical Safety | "Touch-proof" safety connector, not pluggable into AC power outlet, cannot contact hazardous voltage. | The "touch-proof" safety connector is specifically designed so that it cannot be plugged into an AC power outlet and cannot get in touch with possible hazardous voltage. |
| Mechanical Safety | Packaged needle covered with a needle cover. | Packaged needle covered with a needle cover. |
| Materials | (Matches predicate materials: Stainless Steel, Silver/Platinum/Stainless Steel sensor, ABS/PE Hub, Epoxy Insulator, PELD Plastic protector, etc.) | Stainless Steel cannula, silver, ABS Hub, Epoxy Insulator, PELD Plastic protector, Stainless steel / gold plated connection (Concentric). Stainless steel cannula, ABS Hub, PELD Plastic protector, stainless steel/gold plated connection, lead wire (Hypodermic). Stainless Steel (Monopolar/EP). |
| Dimensions | (Similar to predicate dimensions) | Concentric: Diameter 0.30/0.45/0.36/0.65mm, L=25-30-38-50-75mm. Hypodermic: Diameter 0.30/0.40/0.45/0.65mm, L=25-37-40-50-75mm. Monopolar/EP: Diameter 0.36/0.45mm, L=25-37-50-75mm; Diameter 0.40mm, L=12mm. |
| Indications for Use / Anatomical Sites / Where Used | Functionally equivalent to predicates. | Demonstrated to be equivalent to BIONEN s.a.s. (K092973), Technomed Europe (K130136), and Axon System, Inc. (K050194) for their respective types of needle electrodes. |
-
Sample size used for the test set and the data provenance:
- The document does not specify the sample sizes used for the bench testing (penetration, friction force, electrical properties, aging, EO residual, biocompatibility).
- Data Provenance: The accelerated aging testing was conducted by KTR (Korea Testing and Research Institute). Biocompatibility evaluations were performed for manufactured and sterilized Bio Protech electrodes. The provenance of the specific test samples (e.g., country of manufacture) is not detailed beyond the company's location in Republic of Korea. All testing appears to be prospective testing of the Bio Protech device.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- This question is not applicable as the studies are bench performance and material characterization tests, not studies requiring expert interpretation of medical images or patient data to establish ground truth.
-
Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- This question is not applicable as the studies are bench performance and material characterization tests, not studies requiring adjudication of medical interpretations.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No MRMC or comparative effectiveness study involving human readers or AI assistance was conducted or described in this document. This device is a disposable needle electrode, not an imaging or diagnostic AI device.
-
If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- No standalone algorithm performance study was conducted. This device is a physical medical device (needle electrodes), not an algorithm or AI.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- For the performance testing, the "ground truth" or reference was based on established industry standards (e.g., DIN 13097 for mechanical and electrical properties, ASTM F 1980 for aging, ISO 10993 for biocompatibility, ISO 10993-7 for EO residuals) and direct comparison to the performance characteristics of the legally marketed predicate devices.
-
The sample size for the training set:
- This question is not applicable as there is no mention of a training set or an algorithm in this submission.
-
How the ground truth for the training set was established:
- This question is not applicable as there is no mention of a training set or an algorithm in this submission.
Ask a specific question about this device
(270 days)
Dantec™ DCN Disposable Concentric Needle Electrodes are intended for use with recording, monitoring and stimulation/recording equipment for the stimulation/recording of biopotential signals including electromyograph (EMG) and nerve potential signals.
The Dantec™ DCN Disposable Concentric Needle Electrodes are designed as an electrical bipolar recording device for use in electromyography. The Dantec™ DCN Disposable Concentric Needle Electrode consists of an insulated core wire located inside a stainless steel cannula. The hub has a raised bevel indicator, allowing the user to always know the direction of the recording surface. A coaxial hub allows easy connection between needle and the electrode cable.
Acceptance Criteria and Device Performance Study for Dantec™ DCN Disposable Concentric Needle Electrodes
This document describes the acceptance criteria and the study proving that the Dantec™ DCN Disposable Concentric Needle Electrodes meet these criteria, based on the provided 510(k) submission.
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for the Dantec™ DCN Disposable Concentric Needle Electrodes are primarily demonstrated through equivalence to the predicate device (K112034 Teca™ Elite Disposable Concentric Needle Electrodes) and successful completion of various performance tests.
| Acceptance Criterion | Predicate Device Specification (K112034 Teca™ Elite) | Dantec™ DCN Performance | Discussion of Equivalence/Compliance |
|---|---|---|---|
| Material | |||
| Core Material | Tungsten | Tungsten | Same |
| Low Friction Lubricant | Silicone | Silicone | Same |
| Cannula | Stainless Steel | Stainless Steel | Same |
| Internal Insulating Polymer Coating (Core to Lumen) | Polyesterimide | Polyesterimide/Araldite | Dantec™ DCN uses Polyesterimide with Araldite, still considered equivalent. |
| Physical Dimensions | |||
| Length Diameter (mm) | 25mm 0.30 (30g), 25mm 0.46 (26g), 37mm 0.46 (26g), 50mm 0.46 (26g), 75mm 0.64 (23g) | 25mm 0.30 (30g), 25mm 0.46 (26g), 37mm 0.46 (26g), 50mm 0.46 (26g), 75mm 0.64 (23g) | Same |
| Tip geometry | Trocar Point | Trocar Point | Same |
| Recording Area | Fine gauge = 0.03 mm², Medium gauge = 0.07mm² | Fine gauge = 0.02 mm², Medium gauge = 0.07mm² | Fine gauge is slightly smaller but meets functional criteria, demonstrating no significant difference. |
| Mechanical Properties | |||
| Hub/Cover Tensile | Specification: >1.2 kgf, Mean: 3.9 kgf | Specification: >1.2 kgf, Mean: 3.9 kgf | Same |
| Needle/Hub Tensile | Specification: >6 lbf, Mean: 19.7 lbf | Specification: >6 lbf, Mean: 22.7 lbf | Exceeds predicate results, demonstrating no significant difference. |
| Penetration Testing | (Implicit: Demonstrates acceptable sliding force) | (Result not explicitly quantified in table, but stated as performed) | Stated that testing was performed, implying satisfactory results to deem equivalence. |
| Electrical Properties | |||
| Impedance | Specification: <200 K Ohms ($KΩ$), 100% Pass | Specification: <200 K Ohms ($KΩ$), 100% Pass | Same |
| Tip Geometry - Grind Measurements | |||
| Pencil Point Height (mm) | LSL = 0.345, USL = 0.377, Mean: 0.362 | LSL = 0.345, USL = 0.377, Mean: 0.354 | No significant difference |
| Facet Angle (degrees) | LSL = 40.0, USL = 46.0, Mean: 44.9 | LSL = 40.0, USL = 46.0, Mean: 44.9 | No significant difference |
| Biocompatibility | |||
| Cytotoxicity (ISO 10993-5) | No cytotoxic effect | Non-cytotoxic to L929 cells | Complies with standard |
| Sensitization (ISO 10993-10) | Non-sensitizer effect | Non-sensitizer (Polar and Non-Polar extracts) | Complies with standard |
| Irritation (ISO 10993-10) | Meets requirements | Meets requirements (polar and non-polar extracts) | Complies with standard |
| Endotoxin Levels (USP <161>) | <20 EU/device | Below 20 EU/device (3 lots tested) | Complies with standard |
| Sterilization & Packaging | |||
| Sterilization Method | Gamma Irradiation | Gamma Irradiation | Same |
| Sterility Assurance Level (SAL) | $10^{-6}$ | $10^{-6}$ | Same |
| Shelf Life | Three years | Three years | Same |
2. Sample Size Used for the Test Set and Data Provenance
The document describes several non-clinical tests to demonstrate device performance and substantial equivalence.
- Mechanical Testing (Hub/Cover tensile, Needle/Hub tensile, Impedance):
- Sample Size: For Needle/Hub tensile, performance results show a "Mean (lbf)" suggesting multiple samples were tested, but the exact number is not explicitly stated. For Impedance, "All samples pass" indicates that multiple samples were tested, and "Attribute data - 100% Pass" implies a robust assessment.
- Data Provenance: The tests were conducted by Natus Manufacturing Limited as part of the 510(k) submission for a new device. This represents prospective testing of the manufactured device. The country of origin for the data is Ireland, where Natus Manufacturing Limited is located.
- Tip Geometry-Grind Measurements:
- Sample Size: Similar to tensile testing, "Mean" values for Pencil Point Height and Facet Angle suggest multiple units were measured, but the specific number is not provided.
- Data Provenance: Prospective testing conducted by Natus Manufacturing Limited (Ireland).
- Biocompatibility Testing (Cytotoxicity, Sensitization, Irritation, Endotoxin):
- Sample Size: For Endotoxin levels, "three tested lots" were evaluated. For other biocompatibility tests, the sample size (number of devices or extract samples) is not explicitly stated but implied to be sufficient for standard ISO 10993 testing.
- Data Provenance: Prospective testing conducted according to ISO 10993 series standards, likely by a certified lab. The provenance is associated with the test sponsor, Natus Manufacturing Limited (Ireland).
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
This type of device (disposable concentric needle electrodes) does not typically involve human expert interpretation of results to establish "ground truth" in the way image analysis or diagnostic algorithms do. The performance is assessed against engineering specifications, material properties, and biological safety standards. Therefore, the concept of "experts establishing ground truth" in a clinical/interpretive sense is not applicable here.
Instead, the "ground truth" is derived from:
- Established international standards (e.g., ISO 10993 for biocompatibility, ASTM F 1980 for aging).
- Engineering specifications for mechanical and electrical properties.
- Material science principles for component selection.
- Predicate device characteristics for demonstrating substantial equivalence.
The experts involved would be the engineers, quality assurance personnel, and regulatory affairs specialists at Natus Manufacturing Limited, along with personnel at independent testing laboratories, who are qualified in their respective fields to design, conduct, and interpret these types of tests against the established standards and specifications.
4. Adjudication Method for the Test Set
Adjudication methods (e.g., 2+1, 3+1) are typically used in clinical studies or performance evaluations where there can be inter-reader variability in interpreting diagnostic results, often involving human readers and a gold standard. Since the tests described are primarily bench testing and biocompatibility evaluations against predefined acceptance criteria and standards, a formal adjudication method for the "test set" is not relevant. The results are quantitative measurements or qualitative compliance assessments against specified thresholds or documented standards.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done
No, an MRMC comparative effectiveness study was not done. This type of study is relevant for evaluating the impact of an AI algorithm on human reader performance, typically in diagnostic imaging or similar interpretive tasks. The Dantec™ DCN Disposable Concentric Needle Electrode is a medical device for recording biopotential signals (EMG, nerve potential) and its performance is evaluated through physical, electrical, and biological safety testing, not through human interpretation of cases or AI assistance.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
No, a standalone (algorithm only) performance study was not done. This also relates to AI/software as a medical device. The Dantec™ DCN is a physical needle electrode, not an algorithm. Its function is to acquire signals, not to interpret them or provide diagnostic output independently.
7. The Type of Ground Truth Used
The "ground truth" for evaluating the Dantec™ DCN Disposable Concentric Needle Electrodes consists of:
- Engineering Specifications and Design Requirements: These are the predefined performance targets for mechanical strength, electrical impedance, and physical dimensions established during the device design phase.
- Applicable International and National Standards: Including ISO 10993 series for biocompatibility, USP <161> for endotoxin, ISO 2859-1 for sampling, and ASTM F 1980 for accelerated aging. Compliance with these standards serves as a form of ground truth for safety and performance.
- Predicate Device Characteristics: For many aspects (materials, dimensions, sterilization, intended use), the characteristics and established performance of the legally marketed predicate device (K112034 Teca™ Elite Disposable Concentric Needle Electrodes) serve as the "ground truth" for demonstrating substantial equivalence.
8. The Sample Size for the Training Set
The concept of a "training set" is usually applicable to machine learning algorithms. Since the Dantec™ DCN Disposable Concentric Needle Electrodes are a physical medical device and not an AI/ML algorithm, there is no training set in this context. The manufacturing process and device design are optimized through engineering principles, quality control, and adherence to established standards, rather than through iterative training with a dataset.
9. How the Ground Truth for the Training Set was Established
As there is no training set for this device, the question of how its ground truth was established is not applicable.
Ask a specific question about this device
(140 days)
Ask a specific question about this device
(28 days)
TECA™ ELITE Disposable Concentric Needles are intended for use with recording, monitoring and stimulation/recording equipment for the stimulation/recording of biopotential signals including electromyograph (EMG) and nerve potential signals.
CareFusion 209, Inc. TECATM ELITE Disposable Concentric Needles (TECATM ELITE Needles) are disposable, single use, sterile products intended to sense subcutaneous bioelectrical activity, or to stimulate nerve or muscular response by insertion into the patients muscle(s) through the skin where the recording area can differentiate between individual motor units. TECATM ELITE Needles are connected to an electromyographic (EMG) device through a cable. The EMG device detects and displays the biopotential information to the user.
TECATM ELITE Needles consist of a stainless steel tube with a conductive core. A polyesterimide isolative coating resides between the core and the stainless steel tube. The exterior surface of the needle is dipped in a low-friction lubricant. TECATM ELITE Needles are passive devices, and do not contain electrical hardware components or software.
The following device modifications were made to the TECATM ELITE Needles: Modification of the core material; Addition of a low friction lubricant to the exterior surface of the needle; and Modification of the packaging materials.
The provided document is a Special 510(k) Premarket Notification for TECA™ ELITE Disposable Concentric Needles. This document details device modifications and performance testing for demonstrating substantial equivalence to a predicate device, rather than a study proving new acceptance criteria for a novel device or AI. As such, information regarding AI-specific criteria, human reader studies, or training set details will not be available.
Here's an analysis based on the provided text:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly state "acceptance criteria" in a tabular format with corresponding "reported device performance" as one might expect for a new device establishing clinical efficacy. Instead, it details verification and validation activities conducted to demonstrate that the modified device continues to meet its predetermined specifications and applicable standards. The acceptance criteria are implicitly derived from these standards and internal requirements.
| Acceptance Criteria (Implicit from Standards/Requirements) | Reported Device Performance |
|---|---|
| Biocompatibility: | |
| - Compliance with ISO 10993-1: 2009 for patient-contact materials. | - Verification results indicated compliance. |
| - Comparison to predicate devices (K991522, K033173) for core material. | - Comparison made; deemed compliant. |
| - Comparison to predicate device (K111131) for low friction lubricant. | - Comparison made; deemed compliant. |
| Packaging Integrity: | |
| - Seal Strength (ASTM F88-07). | - Verification results indicated compliance with predetermined specifications. |
| - Seal Leaks (ASTM F1929-04). | - Verification results indicated compliance with predetermined specifications. |
| Sterile Hypodermic Needles: | |
| - Compliance with ISO 7864:1993 (for single use). | - Verification results indicated compliance with predetermined specifications. |
| Functional Performance (Internal Requirements): | |
| - Lubricant performance. | - Verification results indicated compliance. |
| - Insertion force. | - Validation results indicated compliance. |
| - Recording characteristics (noise and impedance). | - Validation results indicated compliance. |
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
The document does not specify sample sizes for any of the performance tests. The data provenance is not explicitly stated as "country of origin" or "retrospective/prospective." However, the tests are described as "Performance testing was performed on device characteristics... in accordance with internal requirements and the applicable portions of the following Standards." This implies the testing was conducted internally, likely prospective, to evaluate the specific modifications.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)
This type of information is not relevant or included in the provided document. The device is a medical needle for electromyography, not an AI or diagnostic imaging device that requires expert interpretation for "ground truth" establishment in a clinical study context. The performance assessments are based on engineering standards and physical/electrical measurements.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
Not applicable. This device's performance evaluation does not involve clinical interpretation or a "ground truth" to be adjudicated by multiple experts.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. The device is a passive, single-use needle electrode, not an AI system or an imaging modality that would involve a multi-reader multi-case study.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. The device is a physical medical instrument, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The "ground truth" for this device's performance is established by engineering and biocompatibility standards (e.g., ISO 10993-1, ASTM F88, ASTM F1929, ISO 7864) and the predetermined specifications of the device itself and its predicate. There is no clinical or pathological "ground truth" in the traditional sense for these types of manufacturing and design verification tests.
8. The sample size for the training set
Not applicable. This document describes a medical device, not a machine learning model, so there is no concept of a "training set."
9. How the ground truth for the training set was established
Not applicable. As there is no training set for this device, a method for establishing its "ground truth" is irrelevant.
Ask a specific question about this device
(57 days)
TECA™ elite Disposable Monopolar Needles are intended for use with recording, monitoring and stimulation/recording equipment for the stimulation/recording of biopotential signals including electromyograph (EMG) and nerve potential signals.
CareFusion TECAT™ elite Disposable Monopolar Needles (TECATM elite Needles) are disposable, single use, sterile products intended to sense subcutaneous bioelectrical activity, or to stimulate nerve or muscular response by insertion into the patients muscle(s) through the skin where the recording area can differentiate between individual motor units. TECATM elite Needles are connected to an electromyographic (EMG) device through a cable. The EMG device detects and displays the biopotential information to the user. TECAT™ elite Needles are stainless steel, platinum alloy, and/or gold plated stainless steel, and coated with a polytetrafluoroethylene (PTFE) insulator. A low-friction lubricant is applied to the exterior surface of the needle. TECAT™ elite Needles are passive devices, and do not contain electrical hardware components or software. The following device modifications were made to the TECAT™ elite Needles: Reformulation of the PTFE exterior needle coating; Addition of a low friction lubricant to the exterior surface of the needle; and Modification of the needle tip geometry.
The provided text describes the 510(k) Premarket Notification for CareFusion's TECA™ elite Disposable Monopolar Needles. This document focuses on demonstrating substantial equivalence to a predicate device, rather than a clinical study establishing the performance of a novel AI/ML device against specific acceptance criteria. Therefore, most of the requested information regarding AI/ML device performance, sample sizes, expert ground truthing, MRMC studies, or training sets is not applicable here.
However, I can extract information related to the device's characteristics and the testing performed to show its compliance.
Here's the relevant information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The document primarily focuses on verifying that the modified TECA™ elite Needles comply with predetermined specifications and applicable standards, specifically referencing ISO 10993-1 for biocompatibility and ISO 7864 for performance. The "acceptance criteria" here are implied by these standards and the internal requirements, and the "reported device performance" is a general statement of compliance.
| Acceptance Criterion (Implied) | Reported Device Performance |
|---|---|
| Biocompatibility: Meet biological evaluation requirements. | "Biocompatibility verification was performed on direct patient contact materials... in accordance with... ISO 10993-1: 2009... Verification results and device material comparisons indicated that the TECAT™ elite Needles comply with their predetermined specifications, and with the applicable portions of the Standard." |
| Performance - Coating: Coating performance | "Verification of coating performance... indicated that the TECAT™ elite Needles comply with their predetermined specification and with the applicable Standard." |
| Performance - Tip Geometry: Tip geometry | "Verification of tip geometry... indicated that the TECAT™ elite Needles comply with their predetermined specification and with the applicable Standard." |
| Performance - Insertion Force: Validated insertion force | "Validation of insertion force... indicated that the TECAT™ elite Needles comply with their predetermined specification and with the applicable Standard." |
| Performance - Recording Characteristics: Noise and impedance testing | "Validation of recording characteristics including noise and impedance testing... indicated that the TECAT™ elite Needles comply with their predetermined specification and with the applicable Standard." |
| Overall Safety and Effectiveness | "Verification and validation activities were conducted to establish the performance and safety characteristics of the device modifications... The results... demonstrate that the TECA™ elite Needles are safe and effective when used in accordance with the intended use and labeling." |
2. Sample Size Used for the Test Set and Data Provenance
This is not applicable as the study described is a device modification verification, not a clinical study with a specific test set of patient data. The testing involved bench performance and biocompatibility assessments of the physical device and its materials.
3. Number of Experts Used to Establish the Ground Truth and Qualifications of Those Experts
This information is not provided and is not applicable to the type of device verification described. Ground truth for clinical outcomes or diagnostic accuracy by experts is not relevant here.
4. Adjudication Method for the Test Set
Not applicable. The testing described involves bench tests and biocompatibility assessments, not clinical data requiring adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
No, an MRMC comparative effectiveness study was not done. This device is a passive, disposable electromyograph needle, not an AI-powered diagnostic tool that would typically undergo such a study to evaluate human reader improvement with AI assistance.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
No, this is not an AI/ML algorithm. It's a physical medical device.
7. The Type of Ground Truth Used
For biocompatibility, the "ground truth" was established by compliance with the ISO 10993-1 standard and comparison of direct patient contact materials. For performance testing, the "ground truth" was established by compliance with internal requirements and the ISO 7864 standard for sterile hypodermic needles.
8. The Sample Size for the Training Set
Not applicable. There is no training set mentioned as this is not an AI/ML device.
9. How the Ground Truth for the Training Set Was Established
Not applicable. There is no training set.
Ask a specific question about this device
(377 days)
The BIONEN Disposable Concentric Needle electrodes are sterile electrodes indicated for recording muscle activity for Electro-myography (EMG) applications. The electrodes are for single patient use only.
The BIONEN Disposable Monopolar/Subdermal Needle electrodes are sterile electrodes indicated for recording muscle activity for Electro - myography (EMG) and/or Electroencephalography (EEG) applications. The electrodes are for single patient use only.
The BIONEN Disposable Monopolar Needle electrodes are sterile electrodes indicated for injection of Botulinum Toxin while recording muscle activity with Electroymyography (EMG) applications. The electrodes are for single patient use only.
The BIONEN Disposable Concentric Needle Electrode is used for electromyography (EMG) recording for examination of the peripheral neuromuscular system, by registration of the electrical activity from the muscles.
BIONEN Disposable Concentric Needle Electrode consists of a stainless steel cannula and an inner conductor of stainless steel or platinum. Between these two conductors there is an insulation layer. The inner conductor is the active measure point and the outer conductor of stainless steel is the reference point. The Stainless steel cannula is coated with a low friction lubricant.
The BIONEN Disposable Monopolar/Subdermal Needle Electrode is intended to be inserted in the subdermal, muscle or nerve tissue for use with recording equipment for the recording of biopotentials signals, EEG or EMG, and proximally connected to electromyography/Electroencephalogram recording equipment. The electrodes consist of a formed stainless steel needle with a lead wire attached. The wire can be directly connected or removable and terminates in a safety connector that cannot be connected to an AC power outlet or and cannot get in touch with possible hazardous voltage.
The BIONEN Disposable Monopolar Needle Electrode is intended to be inserted in the muscle while recording electromyography activity, and proximally connected to electromyography recording equipment. The electrodes consist of an hypodermic stainless steel needle with an open lumen and a lead wire attached. The wire can be directly connected or removable and terminates in a safety connector that cannot be connected to an AC power outlet or and cannot get in touch with possible hazardous voltage.
The provided text describes BIONEN Needle Electrodes (Disposable Concentric, Disposable Monopolar/Subdermal, and Disposable Monopolar Needle Electrodes) and summarizes the non-clinical tests performed to demonstrate their functionality and safety, asserting substantial equivalence to predicate devices. However, it does not include specific acceptance criteria with quantifiable metrics for device performance (e.g., sensitivity, specificity, accuracy, or precision in signal recording) nor a study that directly proves the device meets such criteria through a clinical validation or comparative effectiveness study.
Instead, the document focuses on compliance with established standards for medical devices and material safety, and performance characteristics related to the physical integrity and electrical properties of the electrodes.
Therefore, many of the requested sections about acceptance criteria and study details cannot be fully extracted as they are not present in the provided text.
Here's a breakdown of what can be inferred and what is explicitly stated:
1. Table of Acceptance Criteria and Reported Device Performance
As specific quantitative acceptance criteria for device performance (e.g., signal-to-noise ratio, diagnostic accuracy) are not explicitly stated, nor are comparative performance values against the predicate devices for these aspects. The document focuses on performance related to physical attributes and safety.
| Acceptance Criteria Category | Specific Criteria (Implied/Explicit) | Reported Device Performance |
|---|---|---|
| Mechanical Performance | Penetration and friction force (according to DIN 13097) (Concentric Needle) | "The result of testing indicates that the electrodes are as safe, as effective, and perform at least as safely and effectively as the legally marketed device." (Concentric Needle) |
| Sharpening, Bevel and burrs (visual examination with microscope camera) | "Needs meet criteria" (implied by "Quality control tests" and "Tests performed") | |
| Electrical Performance | Electrical continuity and insulation of poles (automated 100% control for Concentric; audible ohmmeter for Monopolar/Subdermal on 10%) | "Needs meet criteria" (implied by "Quality control tests" and "Tests performed") |
| Biocompatibility | Non-cytotoxicity (ISO 10993-5), Non-irritancy (ISO 10993-10), Non-sensitization (ISO 10993-10) | Cytotoxicity: Test article treated extract cells were 97% viable (NON-CYTOTOXIC) |
| Intracutaneous Reactivity: No erythema (Grade 0) or oedema (Grade 0) effects (NON-IRRITANT) | ||
| Sensitization: No hypersensitivity effects (Grade 0 erythema) (NON-SENSITIZING) | ||
| Sterility | Bioburden < 100UFC | Bioburden results consistently "extensively under this value" |
| Shelf-life / Aging | Functionality maintained during shelf life | "Ageing tests was performed to verify and ensure the functionality during the shelf life of the product." (Conclusion of successful maintenance implied) |
| Electrical Safety | "Touch-proof" safety connector prevents connection to AC power outlet and contact with hazardous voltage | Connector is "specifically designed" to meet this (implied successful design and function) |
| Mechanical Safety | Needle covered with a needle cover | "Packaged needle covered with a needle cover." (Implied compliance) |
| Compliance with Standards | IEC 60601-1, ISO 10993-1, ISO 10993-10, ISO 10993-5, ISO 11137, UNI EN 11607-1, UNI EN 868-5, DIN 13097, UNI EN 556-1, ASTM F1980-07, ASTM F2054-00, DIN 58953-6, ASTM F1929-98, ASTM F1886-98, ISO 11607-1 | "The following standards were applied during testing:" (Implied compliance) |
2. Sample Size Used for the Test Set and the Data Provenance
- Test Set Sample Size:
- Concentric Needle Electrode: Not explicitly stated for each test, but "Quality control tests are performed on every lot" with "100% control" for needle continuity and pole isolation. Penetration and friction force testing, and aging tests are mentioned but sample sizes are not provided.
- Monopolar/Subdermal Needle Electrode: Tested on "10% of every lot" for sharpening, bevel/burrs, and needle continuity.
- Biocompatibility:
- Cytotoxicity: "Test article treated extract cells" - specific number not given.
- Intracutaneous Reactivity: "Test article treated extract" - specific number not given.
- Sensitization: Conducted "on guinea pig" - specific number not given, but refers to reports 408/10, 281/10, 1653 A-10 respectively.
- Bioburden: "Bioburden tests and Dose Audit are performed every 9 months by a third laboratory on a sample of the specific batch manufactured." - specific sample size not given.
- Data Provenance: The studies are non-clinical laboratory tests, likely conducted in Italy where BIONEN s.a.s. is based. The FDA 510(k) summary is a submission for marketing in the USA. The biocompatibility reports have numbers (e.g., 408/10) but no country of origin is explicitly stated for the lab that performed them. These are retrospective tests conducted on samples of the manufactured devices.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts
- Not applicable / Not provided. The "ground truth" for these tests is based on objective measurements against established international and national standards (e.g., ISO, DIN, ASTM) and physical evaluations (visual, automated electrical checks). There is no indication of expert consensus for these technical performance parameters.
4. Adjudication Method for the Test Set
- Not applicable / Not provided. Since the studies are non-clinical technical and safety assessments against defined standards, an adjudication method for human interpretation is not relevant. Results are based on objective measurements or visual inspection against specified criteria.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance
- Not applicable. This document describes medical electrodes, not an AI or imaging device that would typically undergo an MRMC study. There is no mention of AI assistance or human reader performance.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
- Not applicable. This document describes a physical medical device (electrodes), not an algorithm.
7. The Type of Ground Truth Used
- Objective Technical Standards and Laboratory Measurements: The ground truth for the device's acceptable performance is established by compliance with international and national standards (e.g., ISO, IEC, DIN, ASTM) for electrical properties, mechanical properties, sterility, and biocompatibility. This includes laboratory test results like cell viability rates, skin irritation grades, and sensitization responses.
8. The Sample Size for the Training Set
- Not applicable / Not provided. This section discusses the performance characteristics of electrodes and does not involve a "training set" in the context of an algorithm or AI.
9. How the Ground Truth for the Training Set Was Established
- Not applicable / Not provided. As there is no training set mentioned, the method for establishing its ground truth is also not relevant here.
Ask a specific question about this device
(30 days)
Electromyography (EMG) Needle Electrode designed for Botulinum Toxin therapy and nerve block procedures. For single patient use only.
The Ambu Neuroline Disposable Inoject needle electrode is a Single Patient needle electrode for Electromyography (EMG) guided injections. The needle electrode is designed for Botulinum Toxin therapy and nerve block procedures. The Ambu Neuroline Disposable Inoject needle is manufactured in different lengths and diameters. The Ambu Neuroline Disposable Inoject needle is connected to the EMG equipment through a touch proof connector with a pre-attached cable. The Ambu Neuroline Disposable Inoject needle electrode is a sterile product. Ambu Neuroline Disposable Inoject needle electrodes consist of a coated needle made from stainless steel. The Ambu Neuroline Disposable Inoject needle electrode is used with a reference electrode and a ground electrode.
Here's an analysis of the provided text regarding the Ambu Neuroline Disposable Inoject Needle Electrode (K093825), focusing on acceptance criteria and study information:
1. Table of Acceptance Criteria and Reported Device Performance
The provided 510(k) summary does not explicitly state quantitative acceptance criteria or a direct measurement of "device performance" in terms of clinical outcomes or specific numerical metrics. Instead, the justification for clearance relies on substantial equivalence to predicate devices and verification of functionality through non-clinical testing.
Given the information, the acceptance criteria are implicitly tied to meeting product specifications and demonstrating safety and effectiveness comparable to predicate devices. The reported "performance" is that it meets these implicit criteria.
| Acceptance Criterion (Implicit) | Reported Device Performance |
|---|---|
| Functional Verification (Non-clinical) | The device was tested for: - Penetration force - Friction force - Electrical properties - Mechanical properties These tests reportedly verified the functionality and fulfilled product specifications. (Conclusion from non-clinical tests stated that the device fulfills product specifications set for the design.) |
| Biocompatibility | The device passed the following tests based on ISO 10993-1: - Cytotoxicity assay in vitro - Contact hypersensitivity in the guinea pig Maximization study - Intracutaneous test in the rabbit - Systemic Injection test in the mice This ensures biological safety and appropriate levels of biocompatibility. |
| Safety and Effectiveness (Relative to Predicate Devices) | Concluded to be a safe and effective Inoject needle electrode and comparable to the predicate devices based on non-clinical verification and biocompatibility tests. (This is the overarching conclusion required for 510(k) clearance via substantial equivalence). |
| Maintenance of Functionality over Shelf Life (Aging) | Aging tests were performed to verify and ensure the functionality during the shelf life of the product. |
| Intended Use and Indications for Use Match Predicate | The device has the same intended use and similar characteristics as the currently commercially available Ambu Neuroline Disposable Inoject needle electrode (K001869, assuming this is the primary predicate) and other predicate devices (K071185, K973444, K002992). The intended use is "Electromyography (EMG) guided injections" and indications for use are "EMG Needle Electrode designed for Botulinum Toxin therapy and nerve block procedures. For single patient use only." |
Regarding the study proving the device meets acceptance criteria:
The provided document describes a submission for a 510(k) premarket notification, which primarily aims to demonstrate substantial equivalence to legally marketed predicate devices, rather than proving de novo clinical efficacy through a standalone study.
2. Sample size used for the test set and the data provenance
- Sample Size for Test Set: Not applicable in the traditional sense of a clinical test set. The submission relies on non-clinical laboratory tests for functional verification and biocompatibility tests. The sample sizes for these specific lab tests (e.g., number of needles tested for penetration force, number of animals in biocompatibility studies) are not provided in this summary.
- Data Provenance: The data provenance for the non-clinical and biocompatibility tests would be internal laboratory studies conducted by Ambu A/S. The document does not specify a country of origin for the non-clinical data, but Ambu A/S is based in Denmark. These were retrospective studies in the sense that they were performed to support the 510(k) submission.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Not applicable. For this type of submission (510(k) for an EMG needle electrode), ground truth is not established by a panel of clinical experts in the context of diagnostic agreement. The "ground truth" for the non-clinical tests would be the pre-defined product specifications and international standards (e.g., ISO 10993-1 for biocompatibility). Compliance with these objective criteria is assessed by engineers and laboratory personnel, not clinical experts for ground truth establishment.
4. Adjudication method for the test set
- Not applicable. There was no clinical adjudication process described, as no clinical test set requiring expert interpretation was conducted. The "adjudication" for non-clinical tests is against pre-defined performance specifications.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No. An MRMC comparative effectiveness study was not performed. This device is a passive medical instrument (a needle electrode), not an AI-powered diagnostic or assistive tool. Therefore, the concept of "human readers improve with AI vs without AI assistance" is not relevant here.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- No. This device is a physical medical instrument, not an algorithm or AI system. Therefore, a standalone algorithm performance study is not applicable.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- For the non-clinical tests (penetration force, friction force, electrical, mechanical properties): The ground truth is the pre-defined engineering specifications and performance requirements for the device.
- For the biocompatibility tests: The ground truth is established by international standards (ISO 10993-1) and the specific biological endpoints measured in the in vitro and in vivo assays (e.g., cell viability for cytotoxicity, skin reaction for hypersensitivity).
8. The sample size for the training set
- Not applicable. As this is not an AI/machine learning device, there is no "training set" in the computational sense. The device's design is based on established engineering principles and prior iterations of similar devices.
9. How the ground truth for the training set was established
- Not applicable. As there is no training set. The design and manufacturing process are validated against the product specifications and regulatory standards.
Ask a specific question about this device
Page 1 of 3