K Number
K092973

Validate with FDA (Live)

Manufacturer
Date Cleared
2010-10-07

(377 days)

Product Code
Regulation Number
890.1385
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BIONEN Disposable Concentric Needle electrodes are sterile electrodes indicated for recording muscle activity for Electro-myography (EMG) applications. The electrodes are for single patient use only.

The BIONEN Disposable Monopolar/Subdermal Needle electrodes are sterile electrodes indicated for recording muscle activity for Electro - myography (EMG) and/or Electroencephalography (EEG) applications. The electrodes are for single patient use only.

The BIONEN Disposable Monopolar Needle electrodes are sterile electrodes indicated for injection of Botulinum Toxin while recording muscle activity with Electroymyography (EMG) applications. The electrodes are for single patient use only.

Device Description

The BIONEN Disposable Concentric Needle Electrode is used for electromyography (EMG) recording for examination of the peripheral neuromuscular system, by registration of the electrical activity from the muscles.

BIONEN Disposable Concentric Needle Electrode consists of a stainless steel cannula and an inner conductor of stainless steel or platinum. Between these two conductors there is an insulation layer. The inner conductor is the active measure point and the outer conductor of stainless steel is the reference point. The Stainless steel cannula is coated with a low friction lubricant.

The BIONEN Disposable Monopolar/Subdermal Needle Electrode is intended to be inserted in the subdermal, muscle or nerve tissue for use with recording equipment for the recording of biopotentials signals, EEG or EMG, and proximally connected to electromyography/Electroencephalogram recording equipment. The electrodes consist of a formed stainless steel needle with a lead wire attached. The wire can be directly connected or removable and terminates in a safety connector that cannot be connected to an AC power outlet or and cannot get in touch with possible hazardous voltage.

The BIONEN Disposable Monopolar Needle Electrode is intended to be inserted in the muscle while recording electromyography activity, and proximally connected to electromyography recording equipment. The electrodes consist of an hypodermic stainless steel needle with an open lumen and a lead wire attached. The wire can be directly connected or removable and terminates in a safety connector that cannot be connected to an AC power outlet or and cannot get in touch with possible hazardous voltage.

AI/ML Overview

The provided text describes BIONEN Needle Electrodes (Disposable Concentric, Disposable Monopolar/Subdermal, and Disposable Monopolar Needle Electrodes) and summarizes the non-clinical tests performed to demonstrate their functionality and safety, asserting substantial equivalence to predicate devices. However, it does not include specific acceptance criteria with quantifiable metrics for device performance (e.g., sensitivity, specificity, accuracy, or precision in signal recording) nor a study that directly proves the device meets such criteria through a clinical validation or comparative effectiveness study.

Instead, the document focuses on compliance with established standards for medical devices and material safety, and performance characteristics related to the physical integrity and electrical properties of the electrodes.

Therefore, many of the requested sections about acceptance criteria and study details cannot be fully extracted as they are not present in the provided text.

Here's a breakdown of what can be inferred and what is explicitly stated:


1. Table of Acceptance Criteria and Reported Device Performance

As specific quantitative acceptance criteria for device performance (e.g., signal-to-noise ratio, diagnostic accuracy) are not explicitly stated, nor are comparative performance values against the predicate devices for these aspects. The document focuses on performance related to physical attributes and safety.

Acceptance Criteria CategorySpecific Criteria (Implied/Explicit)Reported Device Performance
Mechanical PerformancePenetration and friction force (according to DIN 13097) (Concentric Needle)"The result of testing indicates that the electrodes are as safe, as effective, and perform at least as safely and effectively as the legally marketed device." (Concentric Needle)
Sharpening, Bevel and burrs (visual examination with microscope camera)"Needs meet criteria" (implied by "Quality control tests" and "Tests performed")
Electrical PerformanceElectrical continuity and insulation of poles (automated 100% control for Concentric; audible ohmmeter for Monopolar/Subdermal on 10%)"Needs meet criteria" (implied by "Quality control tests" and "Tests performed")
BiocompatibilityNon-cytotoxicity (ISO 10993-5), Non-irritancy (ISO 10993-10), Non-sensitization (ISO 10993-10)Cytotoxicity: Test article treated extract cells were 97% viable (NON-CYTOTOXIC)
Intracutaneous Reactivity: No erythema (Grade 0) or oedema (Grade 0) effects (NON-IRRITANT)
Sensitization: No hypersensitivity effects (Grade 0 erythema) (NON-SENSITIZING)
SterilityBioburden < 100UFCBioburden results consistently "extensively under this value"
Shelf-life / AgingFunctionality maintained during shelf life"Ageing tests was performed to verify and ensure the functionality during the shelf life of the product." (Conclusion of successful maintenance implied)
Electrical Safety"Touch-proof" safety connector prevents connection to AC power outlet and contact with hazardous voltageConnector is "specifically designed" to meet this (implied successful design and function)
Mechanical SafetyNeedle covered with a needle cover"Packaged needle covered with a needle cover." (Implied compliance)
Compliance with StandardsIEC 60601-1, ISO 10993-1, ISO 10993-10, ISO 10993-5, ISO 11137, UNI EN 11607-1, UNI EN 868-5, DIN 13097, UNI EN 556-1, ASTM F1980-07, ASTM F2054-00, DIN 58953-6, ASTM F1929-98, ASTM F1886-98, ISO 11607-1"The following standards were applied during testing:" (Implied compliance)

2. Sample Size Used for the Test Set and the Data Provenance

  • Test Set Sample Size:
    • Concentric Needle Electrode: Not explicitly stated for each test, but "Quality control tests are performed on every lot" with "100% control" for needle continuity and pole isolation. Penetration and friction force testing, and aging tests are mentioned but sample sizes are not provided.
    • Monopolar/Subdermal Needle Electrode: Tested on "10% of every lot" for sharpening, bevel/burrs, and needle continuity.
    • Biocompatibility:
      • Cytotoxicity: "Test article treated extract cells" - specific number not given.
      • Intracutaneous Reactivity: "Test article treated extract" - specific number not given.
      • Sensitization: Conducted "on guinea pig" - specific number not given, but refers to reports 408/10, 281/10, 1653 A-10 respectively.
    • Bioburden: "Bioburden tests and Dose Audit are performed every 9 months by a third laboratory on a sample of the specific batch manufactured." - specific sample size not given.
  • Data Provenance: The studies are non-clinical laboratory tests, likely conducted in Italy where BIONEN s.a.s. is based. The FDA 510(k) summary is a submission for marketing in the USA. The biocompatibility reports have numbers (e.g., 408/10) but no country of origin is explicitly stated for the lab that performed them. These are retrospective tests conducted on samples of the manufactured devices.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

  • Not applicable / Not provided. The "ground truth" for these tests is based on objective measurements against established international and national standards (e.g., ISO, DIN, ASTM) and physical evaluations (visual, automated electrical checks). There is no indication of expert consensus for these technical performance parameters.

4. Adjudication Method for the Test Set

  • Not applicable / Not provided. Since the studies are non-clinical technical and safety assessments against defined standards, an adjudication method for human interpretation is not relevant. Results are based on objective measurements or visual inspection against specified criteria.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

  • Not applicable. This document describes medical electrodes, not an AI or imaging device that would typically undergo an MRMC study. There is no mention of AI assistance or human reader performance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

  • Not applicable. This document describes a physical medical device (electrodes), not an algorithm.

7. The Type of Ground Truth Used

  • Objective Technical Standards and Laboratory Measurements: The ground truth for the device's acceptable performance is established by compliance with international and national standards (e.g., ISO, IEC, DIN, ASTM) for electrical properties, mechanical properties, sterility, and biocompatibility. This includes laboratory test results like cell viability rates, skin irritation grades, and sensitization responses.

8. The Sample Size for the Training Set

  • Not applicable / Not provided. This section discusses the performance characteristics of electrodes and does not involve a "training set" in the context of an algorithm or AI.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable / Not provided. As there is no training set mentioned, the method for establishing its ground truth is also not relevant here.

{0}------------------------------------------------

092973

Image /page/0/Picture/1 description: The image shows the logo for Bionen Medical Devices. The logo features a stylized letter "B" in a bold, sans-serif font, with a smaller, outlined "B" nested within the larger one. Below the "B" symbol, the word "BIONEN" is printed in a simple, sans-serif typeface, followed by the words "medical devices" in a smaller font size.

1 510(k) Summary or 510(k) Statement

Bionen s.a.s Via P. Petrocchi, 42/1 Florence, Italy 50127 P: 39 055 4265945 F: 39 055 412188

Contact: Allison Scott, RAC

Date: September 21, 2010

Trade Name: Disposable Concentric Needle Electrode Disposable Monopolar/Subdermal Needle Electrode Disposable Monopolar Needle Electrode

Common Name: Needle Electrodes

Classification Name: EMG Diagnostic Needle (21 CFR 890.1385, Product Code IKT) Needle Electrode (21 CFR 882.1350, Product Code GXZ)

Predicate Devices:

510(k) NumberTrade or Proprietary or Model NameManufacturer
Concentric Needle Electrode
K071186Neuroline, Disposable Concentric needleelectrodeAmbu A/S
K931966Needle electrode EMGAlpine Biomed(ex Dantec)
Monopolar/Subdermal Needle Electrode
K050194Subdermal Needle Electrodes, Twisted Pair NeedleElectrodes, Corkscrew (spiral) Needle ElectrodeAxon Systems, Inc.
Disposable Monopolar Needle Electrode
K091410Myobot NeedlesSpes Medica
K062437Disposable Hypodermic Needle ElectrodeTechnomed Europe

{1}------------------------------------------------

Device Description:

The BIONEN Disposable Concentric Needle Electrode is used for electromyography (EMG) recording for examination of the peripheral neuromuscular system, by registration of the electrical activity from the muscles.

BIONEN Disposable Concentric Needle Electrode consists of a stainless steel cannula and an inner conductor of stainless steel or platinum. Between these two conductors there is an insulation layer. The inner conductor is the active measure point and the outer conductor of stainless steel is the reference point. The Stainless steel cannula is coated with a low friction lubricant.

The BIONEN Disposable Monopolar/Subdermal Needle Electrode is intended to be inserted in the subdermal, muscle or nerve tissue for use with recording equipment for the recording of biopotentials signals, EEG or EMG, and proximally connected to electromyography/Electroencephalogram recording equipment. The electrodes consist of a formed stainless steel needle with a lead wire attached. The wire can be directly connected or removable and terminates in a safety connector that cannot be connected to an AC power outlet or and cannot get in touch with possible hazardous voltage.

The BIONEN Disposable Monopolar Needle Electrode is intended to be inserted in the muscle while recording electromyography activity, and proximally connected to electromyography recording equipment. The electrodes consist of an hypodermic stainless steel needle with an open lumen and a lead wire attached. The wire can be directly connected or removable and terminates in a safety connector that cannot be connected to an AC power outlet or and cannot get in touch with possible hazardous voltage.

Indications for Use:

The BIONEN Disposable Concentric Needle electrodes are sterile electrodes indicated for recording muscle activity for Electroymyography (EMG) applications. The electrodes are for single patient use only.

The BIONEN Disposable Monopolar/Subdermal Needle electrodes are sterile electrodes indicated for recording muscle activity for Electroymyography (EMG) and/or Electroencephalography (EEG) applications. The electrodes are for single patient use only.

The BIONEN Disposable Monopolar Needle electrodes are sterile electrodes indicated for injection of Botulinum Toxin while recording muscle activity with Electroymyography (EMG) applications. The electrodes are for single patient use only.

Technological Characteristics:

BIONEN Disposable Concentric Needle Electrode consists of a stainless steel cannula and an inner conductor that can be of stainless steel or platinum. Between these two conductors there is an insulation layer. The inner conductor is the active measure point and the outer conductor of stainless steel is the reference point. The Stainless steel cannula is coated with a low friction lubricant. The Electrode has a plastic (NYLON) ergonomic hub.

BIONEN Disposable Monopolar/Subdermal Needle Electrode consists of a formed stainless steel needle with a lead wire attached. The lead wire terminates in a safety connector that cannot be connected to an AC power outlet.

{2}------------------------------------------------

BIONEN Disposable Monopolar Needle Electrode consists of an hypodermic stainless steel needle with an open lumen and a lead wire attached. The lead wire terminates in a safety connector that cannot be connected in a AC power outlet.

Substantial Equivalence Discussion:

Compared to the predicate, the subject devices have intended use, similar physical and performance characteristics and are manufactured using similar processes.

productcharacteristicsNeedle electrodesNeuroline, DisposableConcentric needleelectrodeDCN™ DisposableConcentric NeedleElectrodes
manufacturerBIONEN s.a.s.Ambu A/SAlpine Biomed (ex Dantec)
510K numberK092973K071186K931966
Device classClass IIClass IIClass II
Product codeIKTIKTIKT
Device typeNeedle electrodeDisposable ConcentricneedleelectrodeElectrode, needle,diagnostic electromyograph
Regulationnumber882.1350890.1385890.1385890.1385
indications foruseThe BIONEN DisposableConcentric Needleelectrodes are sterileelectrodes indicated forrecording muscle activityfor Electroymyography(EMG) applications. Theelectrodes are for singlepatient use only.The Neuroline, DisposableConcentric needleelectrodes are made formuscle activity recordingfor Electromyography(EMG) applications. Theelectrodes are for singlepatient use only.Alpine Biomed DisposableNeedle Electrodes cannotbe re-sterilized and are forsingle patient use only. Donot reuse but discard usedneedle electrodes in aproperly marked medicalbiohazard sharps container.
anatomical sitessubdermal, muscle or nervetissuemusclessubdermal, muscle or nervetissue
where used(hospital, home,ambulance, etc)Electrode preparation andapplication should besupervised by a qualifiedhealthcare professionalsamesame
productcharacteristicsNeedle electrodesNeuroline, DisposableConcentric needleelectrodeDCN™ DisposableConcentric NeedleElectrodes
manufacturerBIONEN s.a.s.Ambu A/SAlpine Biomed (ex Dantec)
designErgonomic connector andgeometric sharper tip.Color-coded hubErgonomic and Color-codedhubPre-sterilized DCN™Electrodes have trueergonomic design and arealways sharp.Each needle has a matchingcolour-coded hub andpackage for easyidentification andreordering.The hub has a raised bevelindicator, so you alwaysknow the direction of therecording surface. Theneedle features a coaxialhub that allows you toeasily connect the needlesinto the HUSH™ electrodecables without orientation.A wiping contact is usedinside the hub to optimizethe connection throughoutthe life of the cable andreduce oxide build-up. Thecable can withstand a 50 lbspull test, so it is made tolast.
performanceTested for penetrationand friction force andelectrical properties(according to DIN 13097).Ageing tests are performedto verify and ensure thefunctionality during theshelf life of the product.Tested for penetrationand friction force andelectrical properties(according to DIN 13097).Ageing tests are performedto verify and ensure thefunctionality during theshelf life of the product.Unknown
standards metIEC 60601-1ISO 10993-1ISO 10993-10ISO 10993-5ISO 11137UNI EN 11607-1UNI EN 868-5unknownUnknown
productcharacteristicsNeedle electrodesNeuroline, DisposableConcentric needleelectrodeDCN™ DisposableConcentric NeedleElectrodes
manufacturerBIONEN s.a.s.Ambu A/SAlpine Biomed (ex Dantec)
materialsStainless Steel cannula,Platinum / Stainless Steelsensor,Polyethylene (PE) Hub,Epoxy Insulator,PELD Plastic protector,Stainless steel/gold platedconnectionStainless steel cannula,Silver sensor,ABS Hub,Epoxy Insulator,Polyethylene (PE) Plasticprotector,brass/gold platedconnectionStainless Steel cannula,Platinum sensor
DimensionsDiameter = 0.45/0.35 mmL=25-30-35-40-45-50-65mmDiameter = 0.45/0.35 mmL=25-30-38-50-75mmDiameter = 0.30/0.46/0.64mmL=25-37-50-75mm
Recording area$0,02 - 0,07mm^2$$0,02 - 0,07mm^2$$0.019 - 0.07mm^2$
Connector cableDin 5 polesDin 5 polesDin 5 poles
biocompatibilityselection of materials,which demonstrateappropriatelevels of biocompatibility.Tests on the basis of ISO10993-1selection of materials,which demonstrateappropriatelevels of biocompatibility.Tests on the basis of ISO10993-1Unknown
compatibilitywith theenvironmentand otherdevicesCompatibility is achievedthrough theconnecting cable toEMG/EEG machines orsimilar physiologicalrecording devices.unknownUnknown
sterilityGamma irradiationE-beamE-Beam.
Shelf life60 months36 months36 months
electrical safetyThe "touch-proof' safetyconnector is specificallydesigned so that it cannotbe plugged into AC poweroutlet and cannot get intouch with possiblehazardous voltageunknownUnknown
mechanicalsafetyPackaged needle coveredwith a needle cover.Packaged needle coveredwith a needle cover.Packaged needle coveredwith a needle cover.
chemical safetyNot applicableNot applicableNot applicable
thermal safetyNot applicableNot applicableNot applicable
radiation safetyNot applicableNot applicableNot applicable

Concentric Needle Electrodes

{3}------------------------------------------------

:

:

.

.

.

{4}------------------------------------------------

:

{5}------------------------------------------------

Monopolar/Subdermal Needle Electrodes

productNeedle electrodesSubdermial Needle Electrodes,
characteristicsTwisted Pair Needle Electrodes,
Corkscrew (spiral) Needle Electrode
manufacturerBIONEN s.a.s.Axon Systems, Inc.
510K numberK092973K050194
Device classClass IIClass II
Product codeGXZGXZ
Device typeMonopolar/Subdermal Needle electrodeSubdermal Needle electrode
Regulation number882.1350882.1350
indications for useThe BIONEN DisposableAxon Systems' Subdermal Needle
Monopolar/Subdermal Needle electrodesElectrodes are intended for use with
are sterile electrodes indicated forrecording, monitoring and
recording muscle activity forstimulation/recording equipment for the
Electroymyography (EMG) and/orrecording of biopotential signals including
Electroencephalography (EEG)electroencephalograph (EEG),
applications. The electrodes are forelectromyograph (EMG) and nerve
single patient use only.potential signals and for stimulation
during the intraoperative diagnosis of
acute dysfunction in corticospinal axonal
conduction.
anatomical sitessubdermal, muscle or nerve tissuesubdermal
where usedElectrode preparation and applicationUse by a licensed physician or
(hospital, home,should be supervised by a qualifiedtechnologist under the supervision of a
ambulance, etc)healthcare professional.physician.
designErgonomic connector and geometricUnknown
sharper tip.
Color-coded hub
performanceSharpening; Camera visualUnknown
examination with special
attention to bevel and burrs; Electrical
continuity and isolation of all
poles;
standards metIEC 60601-1unknown
ISO 10993-1
ISO 10993-10
ISO 10993-5ISO 11137
UNI EN 11607-1
UNI EN 868-5
materialsStainless Steel / platinum needleStainless steel needle
DimensionsDiameter=0,35mmDiameter=0,4mm
L=15-20mm
biocompatibilityselection of materials, whichunknown
demonstrate appropriate
levels of biocompatibility. Tests on the
basis of ISO 10993-1
compatibility withCompatibility is achieved through theunknown
the environment andconnecting cable to EMG/EEG machines
other devicesor similar physiological recording devices.
productcharacteristicsNeedle electrodesSubdermial Needle Electrodes,Twisted Pair Needle Electrodes,Corkscrew (spiral) Needle Electrode
manufacturerBIONEN s.a.s.Axon Systems, Inc.
Shelf life60 monthsunknown
electrical safetyThe "touch-proof' safety connector isspecifically designed so that it cannot beplugged into AC power outlet and cannotget in touch with possible hazardousvoltageThe DIN 42802 "touch-proof' safetyconnector is specifically designed so thatit cannot be plugged into AC power outlet
mechanical safetyPackaged needle covered with aneedle cover.Packaged needle covered with aneedle cover.
chemical safetyNot applicableNot applicable
thermal safetyNot applicableNot applicable
radiation safetyNot applicableNot applicable

.

{6}------------------------------------------------

Disposable Monopolar Needle electrode

productDisposable MonopolarMyobot NeedleDisposable Hypodermic
characteristicsNeedle electrodeNeedle Electrode
manufacturerBIONEN s.a.s.Spes MedicaTechnomed Europe
510K numberK092973K091410K062437
Device classClass IIClass IIClass II
Product codeIKTIKTIKT
Device typeNeedle electrodeNeedle electrodeNeedle electrode
Regulation number890.1385890.1385890.1385
indications for useThe BIONEN DisposableMonopolar Needleelectrodes are sterileelectrodes indicated forinjection of BotulinumToxin while recordingmuscle activity withElectromyography (EMG)applications. Theelectrodes are for singlepatient use only.Recording muscle activityfor Electromyography(EMG) applications. Forsingle patient use only.The disposablehypodermic needle isintended to be used forinjection of the BotulinumToxin into a muscle, whilerecordingelectromyography activity.The electrode has an openlumen and is designed formuscle stimulation, motorunit action, potentialrecording and BotulinumToxin injection.The disposablehypodermic EMG needleelectrode is used formuscle stimulation, motorunit action potentialrecording and drugdelivery. The motornerves are monitored bydetecting EMG activity inthe muscles theyinnervate. The drug usedshould be Botox BotulinumToxin type A.Note: Technomed Europedoes not supply any drugswith the Disposablehypodermic EMG needleelectrodes nor doesTechnomed Europe offerfor sale any form of drugs.
anatomical sitesmusclemusclemuscle
productcharacteristicsDisposable MonopolarNeedle electrodeMyobot NeedleDisposable HypodermicNeedle Electrode
manufacturerBIONEN s.a.s.Spes MedicaTechnomed Europe
where used(hospital, home,ambulance, etc)Electrode preparation andapplication should besupervised by a qualifiedhealthcare professional.The specific Botox® type tobe injected must bechosen by the physician.unknownunknown
designErgonomic connector andgeometric sharper tip.The disposable MonopolarNeedle electrode consistsof a stainless steel cannulaelectrically insulated witha PTFE coating, except forthe lancet point and theinner surface of the tube.The coating is to ensureeasy skin penetration andto ensure electricalinsulation on the entirecannula, except for thepoint. A husk fittingtogether with a wire withconnection to anextension cable has beenattached to the cannula.This cable will enable theelectrical signal to betransferred to astimulating or recordingdeviceunknownThe disposablehypodermic EMG needleelectrode consists of astainless steel cannulaelectrically insulated witha PTFE coating, except forthe lancet point and theinner surface of the tube.The coating is to ensureeasy skin penetration andto ensure electricalinsulation on the entirecannula, except for thepoint. A husk fittingtogether with a wire withconnection to anextension cable has beenattached to the cannula.This cable will enable theelectrical signal to betransferred to astimulating or recordingdevice.
performanceSharpening;Camera visualexamination with specialattention to bevel andburrs;Electrical continuityunknownunknown
standards metIEC 60601-1ISO 10993-1ISO 10993-10ISO 10993-5ISO 11137UNI EN 11607-1UNI EN 868-5unknownunknown
materialsStainless Steel cannula,Polyethylene (PE) Hub,PELD Plastic protector,Stainless steel/gold platedStainless Steel cannulaStainless steel cannula,PTFE coating, lead wire
productcharacteristicsDisposable MonopolarNeedle electrodeMyobot NeedleDisposable HypodermicNeedle Electrode
manufacturerBIONEN s.a.s.Spes MedicaTechnomed Europe
DimensionsDiameter=0,50mmL=20-30-40-50-60mmDiameter=0,30-0,40-0,45-0,50-0,55-0,70mmL=25-37-50-75mmDiameter=0,30-0,40-0,45-0,50mmL=25-37-50-75mm
biocompatibilityTests on the basis of ISO10993-1 table 1 made onthe Concentric needleelectrode as the worsecase are considered validand extensible to this kindof device.unknownunknown
compatibility withthe environment andother devicesCompatibility is achievedthrough the connectingcable to EMG machines orsimilar physiologicalrecording devices.unknownunknown
sterilityGamma irradiationunknownunknown
Shelf life60 monthsunknownunknown
electrical safetyThe "touch-proof' safetyconnector is specificallydesigned so that it cannotbe plugged into AC poweroutlet and cannot get intouch with possiblehazardous voltageunknownunknown
mechanical safetyPackaged needle coveredwith a needle cover.Packaged needle coveredwith a needle cover.Packaged needle coveredwith a needle cover.
chemical safetyNot applicableNot applicableNot applicable
thermal safetyNot applicableNot applicableNot applicable
radiation safetyNot applicableNot applicableNot applicable

:

{7}------------------------------------------------

.

.

:

1

{8}------------------------------------------------

:

·

…………

{9}------------------------------------------------

Discussion of Nonclinical Tests Submitted:

The non-clinical tests performed are laboratory tests to verify the functionality of the BIONEN Needle Electrodes. The following standards were applied during testing:

IEC 60601-1:2005UNI EN 868-5:2002
UNI EN ISO 10993-1UNI EN 556-1:200
UNI EN ISO 10993-10ASTM F1980-07
UNI EN ISO 10993-5ASTM F2054-00
UNI EN ISO 11137-1DIN 58953-6
UNI EN ISO 11137-2ASTM F1929-98
UNI EN ISO 11137-3ASTM F1886-98
DIN 13097:2009ISO 11607-1

The Concentric Needle Electrode is tested for penetration and friction force according to DIN 13097. The result of testing indicates that the electrodes are as safe, as effective, and perform at least as safely and effectively as the legally marketed device.

Ageing tests was performed to verify and ensure the functionality during the shelf life of the product.

Quality control tests are performed on every lot:

  • Needle sharpening: visual examination with a microscope camera �
  • . Bevel and burrs: visual examination with a microscope camera
  • Needle continuity: automated process (100% control) .
  • Needle pole isolation: automated process (100% control) �

The Monopolar/Subdermal needle electrode is tested on 10% of every lot:

  • � Needle sharpening: visual examination under microscope
  • Bevel and burrs: visual examination under microscope .
  • Needle continuity: audible ohmmeter .

Bioburden tests and Dose Audit are performed every 9 months by a third laboratory to verify the microbic load on a sample of the specific batch manufactured.

BIONEN has determined its reference value at 100UFC and Bioburden results are always extensively under this value.

The biological safety of the BIONEN Concentric Needle Electrode has been assured through the selection of materials, which demonstrate appropriate levels of biocompatibility.

{10}------------------------------------------------

Some tests were selected for the Biological evaluation of Medical Device on the basis of ISO 10993-10 and ISO 10993-5. The following tests were performed:

TestReportnumberStandardResultConclusion
Cytotoxicity -Neutral Red Uptake408/10 &AddendumMay 3,2010ISO 10993-5:2009Test article treatedextract cells were foundto be 97% viableTest sample extractis consideredNON-CYTOTOXIC
IntracutaneousReactivity281/10ISO 10993-10: 2009Test article treatedextract were found tohave no erythema(Grade 0) or oedema(Grade 0) effectsTest sample extractis consideredNON-IRRITANT
Sensitization teston guinea pig -Delayed contacthypersensitivity -Maximizationmethod1653 A-10 &AddendumSeptember30, 2010ISO 10993-10: 2009Test article treatedextract were found tohave nohypersensitivity effects(Grade 0 erythema)Test sample extractis consideredNON- SENSITIZING

From the tests' results, no adverse effects or sign of sensitization have been detected. The requirements of the test protocol were met.

From the results of the non clinical verification test and biocompatibility test, it has been concluded that BIONEN Needle Electrodes fulfill the products' specifications set for the design and are safe and effective needle electrodes comparable to the predicate devices.

{11}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/11/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines forming its body and wings. The eagle is positioned within a circle that contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Bionen SAS c/o Ms. Allison Scott Anson Group, LLC 11460 N. Meridian St., Ste 150 Carmel, IN 46032

DCT - 7 2010

Re: K092973

Trade/Device Name: Disposable Concentric Needle Electrode

Disposable Monopolar/Subdermal Needle Electrode Disposable Monopolar Needle Electrode

Regulation Number: 21 CFR 890.1385 Regulation Name: EMG Diagnostic Needle Regulatory Class: Class II Product Code: IKT, GXZ Dated: September 22, 2010 Received: September 24, 2010

Dear Ms. Scott:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{12}------------------------------------------------

Page 2 - Ms. Allison Scott

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You nut . comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm1118001/pp.flast the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Malvina B. Evdelman M.D Director Division of Ophthalmic, Neurological. and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{13}------------------------------------------------

Indications for Use

510(k) Number (if known): H 09297 3

Device Name: BIONEN Needle Electrodes

Indications for Use:

The BIONEN Disposable Concentric Needle electrodes are sterile electrodes indicated for recording muscle activity for Electro-myography (EMG) applications. The electrodes are for single patient use only.

The BIONEN Disposable Monopolar/Subdermal Needle electrodes are sterile electrodes indicated for recording muscle activity for Electro - myography (EMG) and/or Electroencephalography (EEG) applications. The electrodes are for single patient use only.

The BIONEN Disposable Monopolar Needle electrodes are sterile electrodes indicated for injection of Botulinum Toxin while recording muscle activity with Electroymyography (EMG) applications. The electrodes are for single patient use only.

Prescription Use _ X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use_ (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Noll

(Division Sign-Off) (Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices

Page / of /

510(k) Number K092973

§ 890.1385 Diagnostic electromyograph needle electrode.

(a)
Identification. A diagnostic electromyograph needle electrode is a monopolar or bipolar needle intended to be inserted into muscle or nerve tissue to sense bioelectrical signals. The device is intended for medical purposes for use in connection with electromyography (recording the intrinsic electrical properties of skeletal muscle).(b)
Classification. Class II (performance standards).