(28 days)
TECA™ ELITE Disposable Concentric Needles are intended for use with recording, monitoring and stimulation/recording equipment for the stimulation/recording of biopotential signals including electromyograph (EMG) and nerve potential signals.
CareFusion 209, Inc. TECATM ELITE Disposable Concentric Needles (TECATM ELITE Needles) are disposable, single use, sterile products intended to sense subcutaneous bioelectrical activity, or to stimulate nerve or muscular response by insertion into the patients muscle(s) through the skin where the recording area can differentiate between individual motor units. TECATM ELITE Needles are connected to an electromyographic (EMG) device through a cable. The EMG device detects and displays the biopotential information to the user.
TECATM ELITE Needles consist of a stainless steel tube with a conductive core. A polyesterimide isolative coating resides between the core and the stainless steel tube. The exterior surface of the needle is dipped in a low-friction lubricant. TECATM ELITE Needles are passive devices, and do not contain electrical hardware components or software.
The following device modifications were made to the TECATM ELITE Needles: Modification of the core material; Addition of a low friction lubricant to the exterior surface of the needle; and Modification of the packaging materials.
The provided document is a Special 510(k) Premarket Notification for TECA™ ELITE Disposable Concentric Needles. This document details device modifications and performance testing for demonstrating substantial equivalence to a predicate device, rather than a study proving new acceptance criteria for a novel device or AI. As such, information regarding AI-specific criteria, human reader studies, or training set details will not be available.
Here's an analysis based on the provided text:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly state "acceptance criteria" in a tabular format with corresponding "reported device performance" as one might expect for a new device establishing clinical efficacy. Instead, it details verification and validation activities conducted to demonstrate that the modified device continues to meet its predetermined specifications and applicable standards. The acceptance criteria are implicitly derived from these standards and internal requirements.
| Acceptance Criteria (Implicit from Standards/Requirements) | Reported Device Performance |
|---|---|
| Biocompatibility: | |
| - Compliance with ISO 10993-1: 2009 for patient-contact materials. | - Verification results indicated compliance. |
| - Comparison to predicate devices (K991522, K033173) for core material. | - Comparison made; deemed compliant. |
| - Comparison to predicate device (K111131) for low friction lubricant. | - Comparison made; deemed compliant. |
| Packaging Integrity: | |
| - Seal Strength (ASTM F88-07). | - Verification results indicated compliance with predetermined specifications. |
| - Seal Leaks (ASTM F1929-04). | - Verification results indicated compliance with predetermined specifications. |
| Sterile Hypodermic Needles: | |
| - Compliance with ISO 7864:1993 (for single use). | - Verification results indicated compliance with predetermined specifications. |
| Functional Performance (Internal Requirements): | |
| - Lubricant performance. | - Verification results indicated compliance. |
| - Insertion force. | - Validation results indicated compliance. |
| - Recording characteristics (noise and impedance). | - Validation results indicated compliance. |
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
The document does not specify sample sizes for any of the performance tests. The data provenance is not explicitly stated as "country of origin" or "retrospective/prospective." However, the tests are described as "Performance testing was performed on device characteristics... in accordance with internal requirements and the applicable portions of the following Standards." This implies the testing was conducted internally, likely prospective, to evaluate the specific modifications.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)
This type of information is not relevant or included in the provided document. The device is a medical needle for electromyography, not an AI or diagnostic imaging device that requires expert interpretation for "ground truth" establishment in a clinical study context. The performance assessments are based on engineering standards and physical/electrical measurements.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
Not applicable. This device's performance evaluation does not involve clinical interpretation or a "ground truth" to be adjudicated by multiple experts.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. The device is a passive, single-use needle electrode, not an AI system or an imaging modality that would involve a multi-reader multi-case study.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. The device is a physical medical instrument, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The "ground truth" for this device's performance is established by engineering and biocompatibility standards (e.g., ISO 10993-1, ASTM F88, ASTM F1929, ISO 7864) and the predetermined specifications of the device itself and its predicate. There is no clinical or pathological "ground truth" in the traditional sense for these types of manufacturing and design verification tests.
8. The sample size for the training set
Not applicable. This document describes a medical device, not a machine learning model, so there is no concept of a "training set."
9. How the ground truth for the training set was established
Not applicable. As there is no training set for this device, a method for establishing its "ground truth" is irrelevant.
{0}------------------------------------------------
Special 510(k) Premarket Notification CareFusion 209, Inc. TECA™ ELITE Disposable Concentric Needles 510(k) Summary
AUG 1 2 2011
| Submission Date: | 10 July 2011 | ||
|---|---|---|---|
| Submitter: | CareFusion 209, Inc.1850 Deming WayMiddleton, WI 53562 | ||
| Submitter Contact: | Mr. Glen HermansonRegulatory Affairs Manager - ManufacturingCareFusion 209, Inc.1850 Deming WayMiddleton, WI 53562+1 (608) 829 8608+1 (608) 829 8737 (fax)Glen.Hermanson@CareFusion.com | ||
| Manufacturing Site: | CareFusion Manufacturing Ireland 241, Ltd.IDA Business ParkGort, Co. GalwayIreland | ||
| Trade Name: | TECA™ ELITE Disposable Concentric Needles | ||
| ClassificationName: | Diagnostic electromyograph needle electrode | ||
| ClassificationRegulation: | 21 CFR §890.1385 | ||
| Product Code: | IKT | ||
| SubstantiallyEquivalent Devices: | New Care FusionModel | 510(k)Number | PredicateManufacturer / Model |
| TECA™ ELITEDisposable ConcentricNeedles | K961013 | Medelec, Limited (now ownedby CareFusion) / DisposableNeedle Electrode |
i
{1}------------------------------------------------
Special 510(k) Premarket Notification CareFusion 209, Inc. TECA™ ELITE Disposable Concentric Needles 510(k) Summary
| Device Description: | CareFusion 209, Inc. TECATM ELITE Disposable Concentric Needles(TECATM ELITE Needles) are disposable, single use, sterile productsintended to sense subcutaneous bioelectrical activity, or to stimulatenerve or muscular response by insertion into the patients muscle(s)through the skin where the recording area can differentiate betweenindividual motor units. TECATM ELITE Needles are connected to anelectromyographic (EMG) device through a cable. The EMG devicedetects and displays the biopotential information to the user.TECATM ELITE Needles consist of a stainless steel tube with aconductive core. A polyesterimide isolative coating resides between thecore and the stainless steel tube. The exterior surface of the needle isdipped in a low-friction lubricant. TECATM ELITE Needles are passivedevices, and do not contain electrical hardware components or software.The following device modifications were made to the TECATM ELITENeedles:Modification of the core material; Addition of a low friction lubricant to the exterior surface of the needle; and Modification of the packaging materials. |
|---|---|
| Intended Use: | TECATM ELITE Disposable Concentric Needles are intended for usewith recording, monitoring and stimulation/recording equipment for thestimulation/recording of biopotential signals including electromyograph(EMG) and nerve potential signals. |
| TechnologyComparison: | TECATM ELITE Needles employ the same technological characteristicsas the predicate device. |
{2}------------------------------------------------
Special 510(k) Premarket Notification CareFusion 209, Inc. TECA™ ELITE Disposable Concentric Needles 510(k) Summary
Summary of Performance Testing:
ﺔ ﺗﻢ
A comparison of the TECA™ ELITE Needles core material was made Biocompatibility with the core material from the following devices: K991522, FHC Micro Targeting Electrode; and . K033173, FHC Micro Targeting Electrode. . Biocompatibility verification was performed on direct patient contact materials comprising the TECA™ ELITE Needles that were affected by the device modifications in accordance with and the applicable portions of the following Standards:
-
ISO 10993-1: 2009, Biological evaluation of medical devices -. Part 1: Evaluation and testing within a risk management process.
Additionally, a comparison of the low friction lubricant characteristics used on the exterior surface of the TECA™ ELITE Needles was made to the low friction lubricant characteristics used on the following device: -
KII1131, TECA™ elite Disposable Monopolar Needles. .
Verification results and device material comparisons indicated that the TECA™ ELITE Needles comply with their predetermined specifications, and with the applicable portions of the Standard.
{3}------------------------------------------------
Special 510(k) Premarket Notification CareFusion 209, Inc. TECA™ ELITE Disposable Concentric Needles 510(k) Summarv
Performance testing was performed on device characteristics of the Performance Testing TECA™ ELITE Needles that were affected by the device modifications - Bench in accordance with internal requirements and the applicable portions of the following Standards:
- ASTM F88 07, Standard Test Method for Seal Strength of Flexible . Barrier Materials
- ASTM F 1929 04, Standard Test Method for Detecting Seal Leaks . in Porous Medical Packaging by Dye Penetration
- ISO 7864: 1993, Sterile hypodermic needles for single use .
Performance testing related to internal requirements consisted of:
- . Verification of lubricant performance;
- . Validation of insertion force; and
- Validation of recording characteristics including noise and . impedance testing.
Verification and validation results indicated that the TECA™ ELITE Needles comply with their predetermined specification and with the applicable Standard.
Conclusion
Verification and validation activities were conducted to establish the performance and safety characteristics of the device modifications made to the TECA™ ELITE Needles. The results of these activities demonstrate that the TECA™ ELITE Needles are safe and effective when used in accordance with the intended use and labeling.
Therefore, the TECA™ ELITE Needles are considered substantially equivalent to the predicate device.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles an abstract representation of a human figure or a bird in flight, composed of three curved lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
CareFusion 209 Inc. % Mr. Glen Hermanson Regulatory Affairs Manager - Manufacturing 1850 Deming Way Middleton, WI 53562
AUG 1 2 2011
Re: K112034
Trade/Device Name: TECA™ ELITE Disposable Concentric Needles Regulation Number: 21 CFR 890.1385 Regulation Name: Diagnostic Electromyograph Needle Electrode Regulatory Class: Class II Product Code: IKT Dated: 10 July 2011 Received: 15 July 2011
Dear Mr. Hermanson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{5}------------------------------------------------
Page 2 - Mr. Glen Hermanson
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Melvina B. Eydelman, M.D.
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Indications for Use
510(k) Number (if known):
Device Name:
Indications for Use:
K | 12034
TECATM ELITE Disposable Concentric Needles
TECA™ ELITE Disposable Concentric Needles are intended for use with recording, monitoring and stimulation/recording equipment for the stimulation/recording of biopotential signals including electromyograph (EMG) and nerve potential signals.
Prescription Use X (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
John Grimes
(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
510(k) Number K112034
N/A