Search Filters

Search Results

Found 197 results

510(k) Data Aggregation

    K Number
    K251860

    Validate with FDA (Live)

    Date Cleared
    2026-02-26

    (254 days)

    Product Code
    Regulation Number
    878.4040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Copioumed AAMI 3 Surgical Gown is intended to be worn by operating room personnel during surgical procedures to protect both surgical patient and the operating room personnel from the transfer of microorganisms, body fluids, and particulate materials. The gown is single-use, disposable, and provided sterile, and is classified as AAMI Level 3.

    Device Description

    Copioumed AAMI 3 Surgical Gown provides protection with the materials of SMS. Copioumed AAMI 3 Surgical Gown has comfort and breathability and for a better fit on the user. The neck design of Copioumed Surgical Gown is adjustable and closes with a Velcro. Copioumed Surgical Gown is available in several sizes and lengths with two models (standard performance and high standard).

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K252301

    Validate with FDA (Live)

    Date Cleared
    2026-02-04

    (195 days)

    Product Code
    Regulation Number
    878.4040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Henan Yadu Level 4 Surgical Gown (Reinforced) are sterile, single-use surgical apparel intended to be used by operating room personnel during surgical operations and other invasive examinations to provide a liquid and microbial barrier for patients and medical staff.

    These products meet the Level 4 liquid barrier performance requirements of ANSI/AAMI PB70:2022 "Liquid barrier performance and classification of protective apparel and drapes intended for use in health care facilities."

    Device Description

    Henan Yadu Level 4 Surgical Gowns (Reinforced) are sterile, single-use surgical apparel intended to be worn by healthcare professionals during surgical procedures to provide barrier protection against the transfer of microorganisms, body fluids, and particulate matter in critical surgical zones for both patients and healthcare personnel.

    These products meet the Level 4 liquid barrier performance requirements of ANSI/AAMI PB70:2022 "Liquid barrier performance and classification of protective apparel and drapes intended for use in health care facilities.

    Gowns have six different sizes: Small (S), Medium (M), Large (L), Extra Large (XL), Extra Extra Large (XXL), Extra extra extra Large (XXXL).

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K250564

    Validate with FDA (Live)

    Date Cleared
    2025-12-03

    (280 days)

    Product Code
    Regulation Number
    878.4040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Level 3 Surgical Gown (Standard & Reinforced) are sterile, single-use surgical apparel intended to be worn by healthcare professionals during surgical procedures to provide barrier protection against the transfer of microorganisms, body fluids, and particulate matter in critical surgical zones for both patients and healthcare personnel.

    The products meet the respective liquid barrier performance requirements for Level 3 as specified in ANSI/AAMI PB70:2022 "Liquid barrier performance and classification of protective apparel and drapes intended for use in health care facilities".

    Device Description

    Surgical Gowns are class II medical devices under the FDA product code FYA, General & Plastic Surgery Panel, and Regulation 21 CFR 878.4040. The subject surgical gowns are single use, disposable and provided sterile. They are available as Standard or Reinforced at Critical areas of the gown to increase its resistance to fluid penetration. The gowns are designed for short to medium length procedures and procedures where there is moderate risk (AAMI PB70 level 3).

    The subject gowns are single-use and provided sterile (Ethylene Oxide sterilization). The subject gowns provide barrier protection against the transfer of microorganisms, body fluids, and particulate matter in critical zones of the gowns for both patients and healthcare personnel.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K243522

    Validate with FDA (Live)

    Date Cleared
    2025-05-30

    (197 days)

    Product Code
    Regulation Number
    878.4040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ViVi® Toga Premium is a component of the ViVi® System and is intended to be worn by surgical personnel to provide a barrier between the operating environment and the surgical personnel in order to help protect against contamination and/or exposure of infectious body fluids and harmful microorganisms.

    Device Description

    The ViVi® Toga Premium is a one-piece head and body cover that is worn by healthcare professionals. The ViVi® Toga Premium is a component of the ViVi® Surgical Helmet System and worn over the air-exchanging surgical helmet. The ViVi® Toga Premium is used with the ViVi® Helmet or the ViVi® Helmet HPL (High Power LED) with Light Emitting Diode (LED) light to provide a barrier between the operating environment and the surgical personnel to protect against contamination and/or exposure of infectious body fluids and harmful microorganisms.

    ViVi® Toga Premium is designed and has been tested to meet the applicable AAMI PB70 standards for level 4 compliance. The AAMI standard does not cover apparel for the head, face, and eyes. Therefore, the hood area and the lens are exempt from classification under this standard.

    AI/ML Overview

    The provided FDA 510(k) clearance letter for the ViVI® Toga Premium is for a Class II surgical apparel device. This document primarily details the substantial equivalence of the ViVi® Toga Premium to a predicate device based on non-clinical performance data. It explicitly states that "No clinical evaluation is necessary for this device."

    Therefore, this document does not describe a study involving an AI/Machine Learning (ML) algorithm with a test set, ground truth, expert adjudication, or MRMC studies. The device is a physical product (surgical toga/gown), and its "performance" is related to its physical properties and barrier capabilities, not diagnostic or predictive accuracy.

    Given the information provided, it's not possible to fulfill the request for acceptance criteria and study details related to an AI/ML device. The "study" mentioned in the document refers to non-clinical laboratory testing of material properties.

    However, I can extract the acceptance criteria and the "reported device performance" based on the provided document as it relates to this specific physical device.


    Acceptance Criteria and Reported Device Performance for ViVi® Toga Premium (Surgical Apparel)

    Based on the provided FDA 510(k) clearance letter for the ViVi® Toga Premium, the acceptance criteria and reported device performance relate to its physical properties and barrier capabilities, as it is a surgical apparel device and not an AI/ML device.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implied by the standards and tests performed, with "PASS" indicating successful compliance.

    Property or CharacteristicTest Method / StandardAcceptance Criteria (Implied)Reported Device Performance (ViVi® Toga Premium)
    Flammability of clothing textiles16 CFR 1610Class 1 compliantClass 1 compliant PASS
    Biocompatibility: CytotoxicityISO 10993-5CompliantCompliant PASS
    Biocompatibility: SensitizationISO 10993-10CompliantCompliant PASS
    Biocompatibility: IrritationISO 10993-23CompliantCompliant PASS
    Tear ResistanceASTM D5587CompliantCompliant PASS
    Tensile StrengthASTM D5034CompliantCompliant PASS
    Seam StrengthASTM F88/F88MCompliantCompliant PASS
    Laser ResistanceISO 11810:2015CompliantCompliant PASS
    Linting (Lint and other particles generation in the dry state)ISO 9073-10:2005CompliantCompliant PASS
    Water Vapor Transmission rateASTM E96/E96M-24ACompliantCompliant PASS
    Barrier Performance (Liquid)ANSI/AAMI PB70:2012; ASTM F1671/F1671MLevel 4 compliantCompliant Level 4 PASS
    Sterilization Assurance Level (SAL)Not explicitly detailed by test method but stated as requirement10⁻⁶10⁻⁶

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document does not specify the exact sample sizes (number of garments or material samples) used for each individual non-clinical test. The testing is reported as "tests were conducted and completed," indicating standard laboratory evaluations.
    • Data Provenance: Not applicable in the context of clinical data. The tests are material and performance tests conducted in a laboratory setting. No geographical origin of "data" in the sense of patient data is mentioned, as this is a physical product. The manufacturer is THI Total Healthcare Innovation GmbH, based in Austria. The predicate device manufacturer is Zimmer Surgical, Inc.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    • Not applicable. This device is surgical apparel, and its "performance" is assessed through standardized physical and biological material testing, not through expert human interpretation of data like images or clinical outcomes. There is no "ground truth" in the diagnostic AI sense.

    4. Adjudication Method for the Test Set

    • Not applicable. As described above, there is no expert adjudication process for this type of device. Performance is determined by meeting pre-defined regulatory standards and test method specifications.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    • No. An MRMC study is a type of clinical study used to evaluate the performance of diagnostic devices (often imaging-based AI) by comparing the accuracy of multiple human readers with and without AI assistance across multiple cases. This does not apply to a surgical apparel device.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study was Done

    • Not applicable. This is not an AI algorithm.

    7. The Type of Ground Truth Used

    • Not applicable in the context of AI/ML. The "ground truth" for this device's performance is derived from established international and national standards (e.g., ISO, ASTM, ANSI/AAMI, CFR), which define objective physical and chemical properties and test methodologies. Meeting these standards serves as the "truth" for device functionality.

    8. The Sample Size for the Training Set

    • Not applicable. This is not an AI/ML device requiring a training set.

    9. How the Ground Truth for the Training Set was Established

    • Not applicable. This is not an AI/ML device and does not have a training set or associated ground truth.
    Ask a Question

    Ask a specific question about this device

    K Number
    K242937

    Validate with FDA (Live)

    Date Cleared
    2025-05-05

    (222 days)

    Product Code
    Regulation Number
    878.4040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Dream Medi Sterile Surgical Gowns are intended to be worn by operating room personnel during surgical procedures to protect the surgical patient and operating room personnel from the transfer of microorganisms, body fluids and particulate matter.

    In addition, this surgical gown meets the requirements of AAMI Level 3 barrier protection for a surgical gown per ANSI/AAMI PB70 Liquid barrier performance and classification of protective apparel and drapes intended for use in health care facilities (AAMI PB70). This is a single use, disposable device, provided sterile.

    Device Description

    Dream Medi Surgical Gowns are made from SMS material. They are full length, constructed with raglan sleeves, hook and loop neck closures, and tie waist closures. They provide AAMI Level-3 protection.

    The Level 3 Gowns have five different sizes: Small (S), Medium (M), Large (L), Extra Large (XL), Extra Extra Large (XXL).

    SMS is a multi-ply material consisting of layers of spunbond and meltblown polypropylene.

    The body is made from 44g Blue SMS and the sleeve is made from 54g Blue SMS. Sleeve opening is made from pure polyester. The collar closure is made from dacron. All gowns are sterilized with ethylene oxide.

    AI/ML Overview

    This document, a 510(k) Clearance Letter, describes the regulatory approval of a surgical gown (Dream Medi Sterile Surgical Gown), not a medical device that relies on complex algorithms, image analysis, or AI. Therefore, most of the requested information regarding "acceptance criteria and the study that proves the device meets the acceptance criteria" in the context of an AI/algorithm-driven device (e.g., sample size for test/training sets, expert ground truth, MRMC studies, standalone performance) is not applicable to this document.

    The "acceptance criteria" and "study that proves the device meets the acceptance criteria" in this context refer to the performance of the surgical gown against established standards for protective apparel, as detailed in Section 8.0 "Summary of Non-Clinical Testing."

    Here's a breakdown of the relevant information from the provided document, addressing the prompt as best as possible given the nature of the device:

    1. Table of Acceptance Criteria and Reported Device Performance

    PurposeTestAcceptance Criteria (from AAMI Level 3)Reported Device Performance (Results)
    Impact PenetrationAATCC 42Level 3, ≤1.0 gPass
    Hydrostatic ResistanceAATCC 127Level 3, ≥50 cmPass
    Tensile strengthASTM D5034≥30 N (≥ 7 lbf)Pass
    Tear resistanceASTM D5587≥10 N (≥ 2.3 lbf)Pass
    Seam strengthASTM D1683≥30 N (≥ 7 lbf)Pass
    Lint and Other particles generation in the dry stateISO 9073-10Log 10 < 4Pass
    Flammability testing16 CFR Part 1610Class 1Pass
    Biocompatibility cytotoxicityISO 10993-5Non-cytotoxicPass
    Biocompatibility irritationISO 10993-23Non-irritantPass
    Biocompatibility SensitizationISO 10993-10Non-sensitizingPass

    2. Sample size used for the test set and the data provenance

    • Sample Size for Test Set: The document does not specify the exact sample size for each non-clinical test. However, these tests are typically performed on a statistically representative number of units according to the respective ASTM, AAMI, and ISO standards (e.g., a certain number of specimens cut from gowns for tensile strength, tear resistance, etc.).
    • Data Provenance: The tests were performed to confirm the device met design specifications and acceptance criteria. The document states the manufacturer is Dae Myung Chemical Co., Ltd. in Vietnam. The testing itself is non-clinical, meaning it's laboratory-based performance testing rather than patient data. It is inherently "prospective" in the sense that the gowns were produced and then tested to confirm properties.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not Applicable. For a surgical gown, "ground truth" is established by the standardized test methods themselves (e.g., AATCC 42 for impact penetration, ISO 10993-5 for cytotoxicity). These are objective measurements following established protocols, not subjective interpretations by human experts in the way that ground truth for medical image analysis would be.

    4. Adjudication method for the test set

    • Not Applicable. Adjudication methods (like 2+1 or 3+1) are used for resolving discrepancies in human expert interpretations (e.g., in radiology studies). For physical product testing against objective criteria, there is no such adjudication method. Results are either within the specified limits ("Pass") or not.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done

    • No. This type of study involves human readers (e.g., radiologists) interpreting cases with and without AI assistance to measure improvement in diagnostic accuracy or efficiency. It is entirely irrelevant for a sterile surgical gown.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Not Applicable. This concept applies to AI algorithms. A surgical gown has no algorithm.

    7. The type of ground truth used

    • Standardized Test Methods and Biocompatibility Assessment: The "ground truth" here is adherence to the performance requirements specified in recognized industry standards (e.g., AAMI PB70, ASTM, ISO, 16 CFR Part 1610) for fluid barrier protection, physical strength, flammability, and biocompatibility. These are objective, measurable properties.

    8. The sample size for the training set

    • Not Applicable. There is no "training set" for a surgical gown. The device is manufactured, and select samples are tested for quality control and regulatory submission.

    9. How the ground truth for the training set was established

    • Not Applicable. See point 8.
    Ask a Question

    Ask a specific question about this device

    K Number
    K242844

    Validate with FDA (Live)

    Date Cleared
    2025-02-26

    (159 days)

    Product Code
    Regulation Number
    878.4040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Medline Level 4 Surgical Gown with Breathable Sleeves is a sterile, single use surgical apparel intended to be worn by healthcare professionals to help protect both the patient and the healthcare worker from the transfer of microorganisms, body fluids, and particulate matter. The Medline Level 4 Surgical Gown with Breathable Sleeves meets the Level 4 requirements of ANSI/AAMIPB70:2022 Liquid barrier performance and classification of protective apparel and drapes intended for use in healthcare facilities.

    The Medline Level 4 Surgical Gown with Breathable Sleeves is also sold in a non-sterile version to repackagers/kitting entities, to be sterilized using the validated ethylene oxide sterilization method according to ISO 11135-1 in its final finished form, as a kit component.

    Device Description

    The Medline Level 4 Surgical Gown with Breathable Sleeves is a single-use, sterile, disposable medical device that is provided in a variety of sizes (Large, X-Large, 3X-Large, 3X-Large and X-Long (only Sirus)) and styles (Sirus, Aurora and Eclipse). The Medline Level 4 Surgical Gown with Breathable Sleeves is constructed from nonwoven SMS (spunbond, meltblown, spunbond), a poly-fabric reinforcement on the chest front panel, and a breathable film (Polyester ether and polyester) for the sleeves. Each gown is designed with a hook and loop closure at the neck, ties at the waist, knitted cuffs at the end of the sleeves as well as a transfer tab for ease in donning. The Medline Level 4 Surgical Gown with Breathable Sleeves has been tested according to ANSI/AAMI PB70:2022 and meets AAMI Level 4 barrier level protection for a surgical gown.

    AI/ML Overview

    This document describes the performance testing for the Medline Level 4 Surgical Gown with Breathable Sleeves (K242844), a surgical apparel device. The information provided is based on non-clinical testing only, as no clinical testing was performed for this 510(k) submission.

    Here's an analysis of the acceptance criteria and the study proving the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance:

    StandardStandard TitleAcceptance CriteriaReported Device Performance (Results)
    ISO 10993-5 CytotoxicityISO MEM Elution Using L-929 Mouse Fibroblast CellsNon-cytotoxicPass - Non-cytotoxic
    ISO 10993-23 IrritationISO Intracutaneous Irritation TestNon-irritatingPass - Non-irritating
    ISO 10993-10 SensitizationISO Guinea Pig Maximization Sensitization TestNon-sensitizingPass - Non-sensitizing
    ASTM F1671Resistance of Materials Used in Protective Clothing to Penetration by Blood-Borne Pathogens4.0 AQLPass
    AATCC 42Water Resistance: Impact Penetration Test≤4.5 gPass
    AATCC 127Water Resistance: Hydrostatic Pressure Test≥50 cmH20Pass
    ASTM D5034-09 (2013)Breaking Strength and Elongation of Textile Fabrics (Grab Test)≥20 NPass
    ASTM D5587-15Tearing Strength of Fabrics by Trapezoid Procedure≥20 NPass
    ASTM D1683Standard Method for Failure in Sewn Seams of Woven Apparel Fabrics≥20 NPass
    16 CFR 1610Flammability of Clothing TextilesMeets Class 1 RequirementsPass
    ASTM E96Water Vapor Transmission of MaterialsReinforced Outside Material >800 g/m²/24 hrsPass
    Sleeve Material >1200 g/m²/24 hrsPass
    ANSI/AAMI PB70:2022Liquid Barrier performance and classification of protective apparel and drapes intended for use in healthcare facilities.Meets ANSI/AAMI PB 70:2012 Level 3 and Level 4 Liquid Barrier requirementsPass

    2. Sample Size Used for the Test Set and Data Provenance:

    The document describes non-clinical tests conducted to verify design specifications. For each test listed in the table, the "Results" column indicates a "Pass," meaning the acceptance criteria were met.

    • Sample Size: The specific sample sizes for each non-clinical test (e.g., number of fabric specimens tested for ASTM F1671) are not detailed in this summary. The 510(k) summary typically provides a high-level overview, and the detailed test reports (which FDA reviews) would contain this information.
    • Data Provenance: The tests are generally performed by accredited testing laboratories or the manufacturer's own qualified labs. The document does not specify the country of origin of the data. Given it's an FDA submission, the tests would typically adhere to recognized international standards like ISO and ASTM. These are prospective tests performed specifically to support the regulatory submission of this new device.

    3. Number of Experts Used to Establish Ground Truth and Qualifications:

    • Not Applicable. For this type of medical device (surgical gowns), "ground truth" is established through adherence to standardized, objective performance tests (e.g., measuring water resistance, breaking strength, biocompatibility). These are not studies that rely on expert interpretation of images or clinical outcomes to establish ground truth in the way an AI diagnostic device would.

    4. Adjudication Method for the Test Set:

    • None. Adjudication methods (like 2+1, 3+1) are relevant for studies involving human interpretation or clinical endpoints, often in the context of diagnostic accuracy or clinical trials. As this is a non-clinical performance evaluation against objective standards, no adjudication method is used. The results are quantitative measurements interpreted against pre-defined thresholds.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done:

    • No, an MRMC comparative effectiveness study was not done. This type of study is typically performed for diagnostic devices (especially those involving image interpretation by human readers) to assess how an AI algorithm impacts reader performance. The Medline Level 4 Surgical Gown is a protective apparel, not a diagnostic device, and its performance is evaluated through physical and material property tests, not human reader studies.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Not Applicable. This question typically pertains to AI/ML software as a medical device (SaMD). The Medline Level 4 Surgical Gown is a physical medical device. Its performance is inherent in its material properties and manufacturing, not an algorithm.

    7. The Type of Ground Truth Used:

    The "ground truth" for this device's performance is established by objective, quantitative measurements against published national and international standards. This includes:

    • Biocompatibility Standards: ISO 10993 series (Cytotoxicity, Irritation, Sensitization).
    • Barrier Performance Standards: ASTM F1671 (blood-borne pathogens), AATCC 42 (impact penetration), AATCC 127 (hydrostatic pressure), and ANSI/AAMI PB70:2022 (overall liquid barrier performance classification).
    • Material Strength Standards: ASTM D5034 (breaking strength), ASTM D5587 (tearing strength), ASTM D1683 (sewn seam failure).
    • Other Performance Standards: 16 CFR 1610 (flammability), ASTM E96 (water vapor transmission/breathability).

    The "ground truth" is therefore the measured physical and chemical properties of the gown, compared against the predefined acceptance criteria of these standard test methods.

    8. The Sample Size for the Training Set:

    • Not Applicable. The concept of a "training set" is relevant for machine learning algorithms. This device is a manufactured product, not an AI algorithm. Its design and manufacturing process would involve material selection and engineering, not machine learning training.

    9. How the Ground Truth for the Training Set Was Established:

    • Not Applicable. As there is no training set for an AI algorithm, this question is not relevant.
    Ask a Question

    Ask a specific question about this device

    K Number
    K241272

    Validate with FDA (Live)

    Date Cleared
    2024-10-16

    (163 days)

    Product Code
    Regulation Number
    878.4040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Stryker Steri-Shield 8 disposables, including all hood and toga models, are components of the Stryker Steri-Shield 8 Personal Protection System and are intended to protect the patient, healthcare personnel, and operating room personnel against contamination, exposure of infectious bodily fluids, and the transfer of microorganism and particulate material. The Stryker Steri-Shield 8 disposables are sterile, disposable, single-use only devices.

    Device Description

    The Stryker Steri-Shield 8 Surgical Protective Equipment consist of Stryker Steri-Shield 8 Surgical Hoods and Steri-Shield 8 Surgical Togas.

    The Stryker Steri-Shield 8 Surgical Hood is a single piece device that covers the user's head, neck, and shoulder region and is intended to be worn with commercially available sterile surgical gowns. The Steri-Shield 8 Hoods are available with Standard (NPA), Peel Away (PA) and Anti-Reflective (AR) face shield options.

    The Stryker Steri-Shield 8 Surgical Togas are a single piece device that covers the user's head, neck, arms, torso, and upper legs region. The Stryker Steri-Shield 8 Surgical Togas is offered in three configurations: Steri-Shield 8 Pullover Togas, Steri-Shield 8 Zippered Togas, and Steri-Shield 8 Tie Back Togas.

    AI/ML Overview

    The Stryker Steri-Shield 8 Surgical Hoods and Togas is a medical device intended to protect the patient, healthcare personnel, and operating room personnel against contamination, exposure to infectious bodily fluids, and the transfer of microorganisms and particulate material. The device is sterile, disposable, and for single-use only.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. A table of acceptance criteria and the reported device performance:

    Acceptance Criteria (Standard/Test)Reported Device Performance
    Barrier Performance (Gown - Critical Zone Level 4 Protection)
    ASTM F1671 testing - BVB & Sleeve SeamPasses requirements for PB70 Level 4 protection
    Barrier Performance (Gown - Non-Critical Zone)
    AATCC Test Method 42 Per AAMI/ANSI PB70Passes AATCC Test Method 42 (< 4.5g)
    Barrier Performance (Hood - Exempt)
    All materials and seamsNon-protective (meets acceptance criteria as intended)
    Physical Performance
    Flammability per 16 CFR Part 1610 (Class 1)Class 1 requirements met
    Water Spray Impact per AATCC 42 (Water Penetration ≤ 4.5 g)Water Penetration ≤ 4.5 g
    Synthetic Blood per ASTM F1862 (No visible penetration)No visible penetration
    Fabric Viral Penetration per ASTM F1671 (no detectable (< 1 PFU/mL))No detectable (< 1 PFU/mL)
    Particulate Filtration Efficiency per ASTM F2299 (≥ 95%)≥ 95%
    Bacterial Filtration Efficiency per ASTM F2101 (≥ 95%)≥ 95%
    Particle Release per ISO 9073-10 (Lint Count ≤ 4.0 (Log10, Upper Quartile))Lint Count ≤ 4.0 (Log10, Upper Quartile)
    Air Permeability per ASTM D737Better or equivalent to predicate product
    Tear Strength per ASTM D5587 (≥ 2.3 lbf)≥ 2.3 lbf
    Tensile Strength per ASTM D5034 (≥ 7.0 lbf)≥ 7.0 lbf
    Seam Strength per ASTM D1683 (≥ 7.0 lbf)≥ 7.0 lbf (Predicate was ≥ 6.0 lbf, so this is an improvement or meets threshold)
    Evaporative Resistance per ASTM F1868Better or equivalent to predicate product
    Light Transmittance and Haze per ASTM D1003Better or equivalent to predicate product
    Perforation Tear Strength Testing (Stryker Internal Testing) (0.15 lbf < Tear Force ≤ 1.5 lbf)0.15 lbf < Tear Force ≤ 1.5 lbf
    Biocompatibility
    Human Factors and UsabilityProduct is found to be safe and effective for the intended users, uses and use environment
    Noise and Speech Interference LevelsBetter or equivalent to predicate product
    Carbon Dioxide (CO2) Testing (Inhale CO2 ≤ 5000ppm per OSHA Exposure Limits)Inhale CO2 ≤ 5000ppm
    ISO 10993-1, -5, -10, -23, -11 (Cytotoxicity, Sensitization, Irritation, Acute Systemic Toxicity)Cytotoxicity – MEM Grade ≤ 2; Sensitization – Sensitization response ≤ 0; Irritation – Intracutaneous Reactivity ≤ 1.0; Acute Systemic Toxicity – No signs of toxicity, no weight loss in excess of 10% (Note: Cytotoxic response in one test article was mitigated by in vivo testing and chemical characterization)
    ISO 10993-17 (Toxicological risk assessment)Assessment performed
    ISO 10993-18 (Chemical characterization of medical device materials)Characterization performed
    Sterilization
    Finished (Terminal) Product Sterilization Method (SAL 10-6 Terminally sterilized via Ethylene Oxide (EO))SAL 10-6 Terminally sterilized via Ethylene Oxide (EO) in accordance with ISO 11135, Half cycle overkill method

    2. Sample size used for the test set and the data provenance:

    The document does not explicitly state the sample sizes for each test mentioned (e.g., number of hoods or togas tested for flammability, water resistance, etc.). The tests are described as "Performance Data (Non-Clinical Tests)," indicating that they were conducted in a laboratory setting on product samples. The data provenance is not specified beyond being generated by Stryker Instruments as part of their 510(k) submission. Given the nature of performance testing for surgical apparel standards, these tests are typically conducted on representative samples of the manufactured devices.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not applicable. The acceptance criteria for the Stryker Steri-Shield 8 Surgical Hoods and Togas are based on recognized industry standards (e.g., ASTM, AATCC, ISO, 16 CFR) for material properties and device performance, not on expert interpretations of outcomes.

    4. Adjudication method for the test set:

    This information is not applicable as the acceptance criteria are based on objective, quantifiable measurements from recognized standards, not subjective assessments requiring adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This information is not applicable. The device is a surgical hood and toga, not an AI-assisted diagnostic or imaging device that would involve human readers or MRMC studies.

    6. If a standalone (i.e., algorithm only without human-in-the loop performance) was done:

    This information is not applicable. The device is a physical product (surgical apparel), not an algorithm or software.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The "ground truth" for this device is based on established, quantifiable metrics and pass/fail criteria defined by recognized national and international standards for medical textiles, barrier materials, and biological safety. Examples include:

    • Physical properties (e.g., tear strength, tensile strength, air permeability).
    • Barrier effectiveness against fluids, synthetic blood, viruses, and particulates.
    • Biocompatibility (cytotoxicity, sensitization, irritation, systemic toxicity).
    • Flammability.
    • CO2 levels within the hood.

    These "ground truths" are objective measurements against defined thresholds in the standards.

    8. The sample size for the training set:

    This is not applicable as the device is a physical product and not an AI/ML algorithm that requires a training set.

    9. How the ground truth for the training set was established:

    This is not applicable as the device is a physical product and not an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K241214

    Validate with FDA (Live)

    Date Cleared
    2024-09-10

    (133 days)

    Product Code
    Regulation Number
    878.4040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ProPel SG3TM Surgical Gown is intended to protect both surgical patients and operating room personnel from the transfer of microorganisms, body fluids, and particulate material. They are classified as Level 3 surgical gowns in accordance with ANSI/AAMI PB70.

    The reusable Propel Surgical Gown is provided non-sterile and must be sterilized before use. Stertlization parameters are as follows:

    Prevacuum Steam Sterilization: 132°C/270°F (temperature)/ 4 minutes (exposure time)/15 minute (dry time)

    Device Description

    The ProPel SG3TM Surgical Gown is a reusable, woven surgical gown that provides an ANSI /AAMI PB70:2022 Level 3 Liquid Barrier Performance Barrier using 100% polyester with silicone coating in the critical zone. The critical zone inner ply and back of the surgical gown is made of polyester with electrostatic dissipative yarn (graphene) and silicone coating. The gown has polyester mesh and vented cape on the back panels of the gown for ventilation. The ProPel SG3TM Surgical Gown may undergo up to 60 reprocessing cycles.

    The ProPel SG3™ Surgical Gown is available in small, medium, large, extra large (XL), 2XL and 3XL. The device will be sold as a non-sterile surgical gown that is to be laundered, sterilized and processed by the healthcare facility prior to first use and after each subsequent use.

    AI/ML Overview

    The provided text describes the non-clinical testing performed on the ProPel SG3™ Surgical Gown to demonstrate its equivalence to a predicate device. It is important to note that this is a 510(k) submission, which aims to show substantial equivalence, not necessarily a study proving clinical effectiveness in the same way a new drug or high-risk device might.

    Here's the information requested, based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    Test MethodologyTest Method/Applicable Standard(s)Acceptance CriteriaReported Device Performance
    Tensile StrengthASTM D5034/ASTM F2407-20≥ 30 N (≥ 7 lbf)PASS
    Tear StrengthASTM D5587/ASTM F2407-20≥ 10 N (≥ 2.3 lbf)PASS
    Seam StrengthASTM D1683/ASTM F2407-20≥ 30 N (≥ 7 lbf)PASS
    Hydrostatic ResistanceAATCC Method 127/ANSI/AAMI PB70≥ 50 cm per AAMI PB70 for Level 3PASS
    Impact PenetrationATCC Method 42/ANSI/AAMI PB70≤ 1.0 g per AAMI PB70 for Level 3PASS
    Lint GenerationISO 9073-10--- (No explicit quantitative acceptance criteria listed; qualitative comparison)Coefficient for linting is 3 for both predicate and subject device
    Flammability16 CFR 1610Pass Class I FlammabilityPASS (Class I)
    Snap StrengthASTM D4846-96Fasteners provide peel strength of ≤ 1.5 lbf (easy doff) and shear force ≥ 4.5 lbf (secure)PASS
    Evaporative ResistanceASTM F1868-17Evaporative resistance of critical zone (average of 3 specimens) ≤ 350 Pa m²/W; evaporative resistance of the back mesh (2-ply average of 3 specimens) ≤ 5 Pa m²/WPASS
    Critical Zone Fabric Weight TestingASTM D3776/D3776M-20--- (No explicit acceptance criteria; result is a specific measurement)Gown critical zone fabric construction weighs less than 7.83 OSY
    Use Life Tracking---QCM label is legible and RFID Chip can be read at the beginning and end of use life (after 60 reuses)PASS
    CytotoxicityISO 10993-5Less than/equal to grade 2 (mild reactivity)PASS
    SensitizationISO 10993-10No evidence of causing delayed dermal contact sensitizationPASS
    IrritationISO 10993-23No erythema, no edemaPASS
    PyrogenicityUSP <151>/ISO 10993-11Temperature rise within acceptable limitsPASS
    Acute Systemic ToxicityISO 10993-11No mortality or evidence of systemic toxicityPASS

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document refers to non-clinical testing of the physical and biological properties of the surgical gown material and design. For each test, "specimens" or "gowns" would be used as samples. However, the exact sample size (N) for each specific test (e.g., number of gowns for tensile strength, number of fabric pieces for hydrostatic resistance, number of biological samples for cytotoxicity) is not explicitly stated in this summary. For Evaporative Resistance, it mentions "average of 3 specimens".

    The data provenance (country of origin, retrospective/prospective) is not provided in this document. Given that it's non-clinical testing of a manufactured product, it is likely that the testing was performed in a controlled laboratory environment, but the location is not specified.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is a non-clinical, laboratory-based study for a surgical gown. It does not involve medical imaging, diagnosis, or clinical decision-making by human experts where "ground truth" would be established by referring radiologists or other clinicians. The "ground truth" or "acceptance criteria" for these tests are based on established national and international standards (like ANSI/AAMI, ASTM, ISO, AATCC, 16 CFR, USP), which reflect a consensus of scientific and technical experts in materials science, biology, and medical device regulations. Therefore, this question is not applicable in the context of this device and study type.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Since this is non-clinical laboratory testing against predefined, objective standards, an adjudication method like 2+1 or 3+1 (typically used for resolving discrepancies in expert interpretations) is not applicable. The 'adjudication' is inherent in the test method itself, where results are measured objectively and compared against numerical acceptance criteria in a pass/fail manner.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    A Multi-Reader, Multi-Case (MRMC) comparative effectiveness study is performed for diagnostic devices, particularly those involving human interpretation (e.g., radiologists reading images). This device is a surgical gown, which is a physical barrier protection device. Therefore, an MRMC study and effects related to human reader improvement with AI assistance are not applicable. The document explicitly states "SUMMARY OF CLINICAL TESTING Not Applicable", further confirming no clinical effectiveness study (which would include MRMC) was performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This refers to the performance of an AI algorithm without human involvement. Since the device is a surgical gown and not an AI-powered diagnostic or assistive tool, this question is not applicable.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the performance claims of the surgical gown is based on established national and international standards and test methods. These standards define measurable physical and biological properties (e.g., tensile strength, tear strength, barrier performance, biocompatibility) and their respective acceptance criteria.

    8. The sample size for the training set

    This question typically applies to machine learning or AI models. Since this is non-clinical testing of a physical medical device (surgical gown), there is no "training set" in the context of an AI algorithm. The gown samples tested were used to demonstrate compliance with the specified performance standards.

    9. How the ground truth for the training set was established

    As there is no "training set" for an AI algorithm in this context, this question is not applicable. The "ground truth" for evaluating the gown's performance was established by recognized industry standards and test methodologies.

    Ask a Question

    Ask a specific question about this device

    K Number
    K233571

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-08-02

    (270 days)

    Product Code
    Regulation Number
    878.4040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    GCI surgical gown is intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Per ANSI/AAMI PB70:2012 Liquid barrier performance and classification of protective apparel and drapes intended for use in health care facilities, the surgical gowns met the requirements for Level 2 barrier classification.

    Device Description

    GCI Surgical Gowns are intended to be used during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate materials. The gown is available in multiple sizes. They are provided non- sterile and intended to be sterilized prior to use. They are reusable up to 75 uses. Per ANSI/AAMI PB70 Liquid barrier performance and classification of protective apparel intended for use in health care facilities, the proposed devices meet the requirements for 2, as indicated by their labeling.

    AI/ML Overview

    This document describes the non-clinical performance testing of the GCI Surgical Gown to demonstrate its substantial equivalence to a legally marketed predicate device (K211422).

    1. Table of Acceptance Criteria and Reported Device Performance

    AttributeAcceptance CriteriaReported Device Performance
    FlammabilityMeets Class 1 RequirementsMeets Class 1 Requirements
    Water Resistance: Hydrostatic PressureWater Resistant ≥20 cmWater Resistant ≥20 cm
    Water Resistance: Impact Penetration≤1.0 g water≤1.0 g water
    Breaking / Bursting Strength≥20 N>20 N
    Tearing Strength≥20 N>20 N
    Seam Strength≥20 N>20 N
    LintingLog10 < 4Log10 < 4
    BiocompatibilityBiocompatible in alignment with ISO 10993-1Biocompatible in alignment with ISO 10993-1
    - CytotoxicityNon-cytotoxicUnder the conditions of the study, the device is non-cytotoxic.
    - IrritationNon-irritatingUnder the conditions of the study, the device is not an irritant.
    - SensitizationNon-sensitizingUnder the conditions of the study, the device is not a sensitizer.

    2. Sample size used for the test set and the data provenance

    The document does not specify the exact sample sizes for each non-clinical test. However, the tests are laboratory-based performance tests of the device itself (GCI Surgical Gown). The data provenance is from non-clinical laboratory testing performed to specific standards (e.g., AATCC, ASTM, ISO) to verify the device met design specifications. The country of origin of the data is not explicitly stated, but the submission is to the US FDA by a Canadian company (George Courey, Inc.), suggesting the testing was performed to international standards. This is retrospective testing performed on the final device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This section is not applicable as the ground truth for this type of device (surgical gown) is established through objective, standardized physical and biological performance tests, not through expert human assessment of images or clinical data. The "ground truth" is defined by the acceptance criteria of the specified international standards.

    4. Adjudication method for the test set

    This section is not applicable. The outcome of non-clinical, objective tests is not subject to adjudication in the same manner as expert consensus for clinical image interpretation. The tests yield quantitative results that are compared directly against predefined numerical or qualitative acceptance criteria specified in the relevant standards.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This section is not applicable. The device is a surgical gown, not an AI-powered diagnostic or assistive tool. Therefore, MRMC studies and assessment of human reader improvement with AI are irrelevant to this submission.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This section is not applicable. The device is a physical surgical gown, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth used for this device is based on objective, standardized measurements and biological assessments as defined by international and national standards. For example:

    • Physical properties (e.g., water resistance, strength) are evaluated against quantitative thresholds.
    • Biocompatibility is evaluated against qualitative (e.g., "non-cytotoxic," "not an irritant") outcomes determined by specified in-vitro or in-vivo biological tests.

    8. The sample size for the training set

    This section is not applicable. The device is a physical product (surgical gown) that does not utilize machine learning or AI, and therefore does not have a "training set" in the computational sense.

    9. How the ground truth for the training set was established

    This section is not applicable for the same reasons as #8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K233068

    Validate with FDA (Live)

    Date Cleared
    2024-03-29

    (185 days)

    Product Code
    Regulation Number
    878.4040
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Disposable Surgical Gowns are intended to be worn by operating room personnel during surgical procedures to protect the surgical patient and operating room personnel from the transfer of microorganisms, body fluids and particulate material. This surgical gown meets the requirements of AAMI Level 3/Level 4 barrier protection for a surgical gown per ANSU AAMI PB70:2012 Liquid barrier performance and classification of protective apparel and drapes intended for use in health care facilities (AAMI PB70). The Surgical Gowns are single use, disposable medical devices; provided sterile.

    Device Description

    Disposable Surgical Gown

    AI/ML Overview

    This document is a 510(k) clearance letter for a Disposable Surgical Gown. It is a regulatory document addressing the substantial equivalence of the device to a predicate device, not a study report for a medical AI/ML device.

    Therefore, the requested information about acceptance criteria, study details, expert qualifications, adjudication methods, MRMC studies, standalone performance, ground truth, and training set details for an AI/ML device is not applicable to this document.

    The document discusses:

    • The device name: Disposable Surgical Gown
    • Regulation Number: 21 CFR 878.4040
    • Regulatory Class: Class II
    • Product Code: FYA
    • Indications for Use: "Disposable Surgical Gowns are intended to be worn by operating room personnel during surgical procedures to protect the surgical patient and operating room personnel from the transfer of microorganisms, body fluids and particulate material. This surgical gown meets the requirements of AAMI Level 3/Level 4 barrier protection for a surgical gown per ANSU AAMI PB70:2012 Liquid barrier performance and classification of protective apparel and drapes intended for use in health care facilities (AAMI PB70). The Surgical Gowns are single use, disposable medical devices; provided sterile."

    The acceptance criteria mentioned for the surgical gown are related to its physical performance as a barrier: "meets the requirements of AAMI Level 3/Level 4 barrier protection for a surgical gown per ANSU AAMI PB70:2012 Liquid barrier performance and classification of protective apparel and drapes intended for use in health care facilities (AAMI PB70)." However, this document does not provide a study detailing how these criteria were met or the specific results.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 20