K Number
K242937

Validate with FDA (Live)

Date Cleared
2025-05-05

(222 days)

Product Code
Regulation Number
878.4040
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Dream Medi Sterile Surgical Gowns are intended to be worn by operating room personnel during surgical procedures to protect the surgical patient and operating room personnel from the transfer of microorganisms, body fluids and particulate matter.

In addition, this surgical gown meets the requirements of AAMI Level 3 barrier protection for a surgical gown per ANSI/AAMI PB70 Liquid barrier performance and classification of protective apparel and drapes intended for use in health care facilities (AAMI PB70). This is a single use, disposable device, provided sterile.

Device Description

Dream Medi Surgical Gowns are made from SMS material. They are full length, constructed with raglan sleeves, hook and loop neck closures, and tie waist closures. They provide AAMI Level-3 protection.

The Level 3 Gowns have five different sizes: Small (S), Medium (M), Large (L), Extra Large (XL), Extra Extra Large (XXL).

SMS is a multi-ply material consisting of layers of spunbond and meltblown polypropylene.

The body is made from 44g Blue SMS and the sleeve is made from 54g Blue SMS. Sleeve opening is made from pure polyester. The collar closure is made from dacron. All gowns are sterilized with ethylene oxide.

AI/ML Overview

This document, a 510(k) Clearance Letter, describes the regulatory approval of a surgical gown (Dream Medi Sterile Surgical Gown), not a medical device that relies on complex algorithms, image analysis, or AI. Therefore, most of the requested information regarding "acceptance criteria and the study that proves the device meets the acceptance criteria" in the context of an AI/algorithm-driven device (e.g., sample size for test/training sets, expert ground truth, MRMC studies, standalone performance) is not applicable to this document.

The "acceptance criteria" and "study that proves the device meets the acceptance criteria" in this context refer to the performance of the surgical gown against established standards for protective apparel, as detailed in Section 8.0 "Summary of Non-Clinical Testing."

Here's a breakdown of the relevant information from the provided document, addressing the prompt as best as possible given the nature of the device:

1. Table of Acceptance Criteria and Reported Device Performance

PurposeTestAcceptance Criteria (from AAMI Level 3)Reported Device Performance (Results)
Impact PenetrationAATCC 42Level 3, ≤1.0 gPass
Hydrostatic ResistanceAATCC 127Level 3, ≥50 cmPass
Tensile strengthASTM D5034≥30 N (≥ 7 lbf)Pass
Tear resistanceASTM D5587≥10 N (≥ 2.3 lbf)Pass
Seam strengthASTM D1683≥30 N (≥ 7 lbf)Pass
Lint and Other particles generation in the dry stateISO 9073-10Log 10 < 4Pass
Flammability testing16 CFR Part 1610Class 1Pass
Biocompatibility cytotoxicityISO 10993-5Non-cytotoxicPass
Biocompatibility irritationISO 10993-23Non-irritantPass
Biocompatibility SensitizationISO 10993-10Non-sensitizingPass

2. Sample size used for the test set and the data provenance

  • Sample Size for Test Set: The document does not specify the exact sample size for each non-clinical test. However, these tests are typically performed on a statistically representative number of units according to the respective ASTM, AAMI, and ISO standards (e.g., a certain number of specimens cut from gowns for tensile strength, tear resistance, etc.).
  • Data Provenance: The tests were performed to confirm the device met design specifications and acceptance criteria. The document states the manufacturer is Dae Myung Chemical Co., Ltd. in Vietnam. The testing itself is non-clinical, meaning it's laboratory-based performance testing rather than patient data. It is inherently "prospective" in the sense that the gowns were produced and then tested to confirm properties.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Not Applicable. For a surgical gown, "ground truth" is established by the standardized test methods themselves (e.g., AATCC 42 for impact penetration, ISO 10993-5 for cytotoxicity). These are objective measurements following established protocols, not subjective interpretations by human experts in the way that ground truth for medical image analysis would be.

4. Adjudication method for the test set

  • Not Applicable. Adjudication methods (like 2+1 or 3+1) are used for resolving discrepancies in human expert interpretations (e.g., in radiology studies). For physical product testing against objective criteria, there is no such adjudication method. Results are either within the specified limits ("Pass") or not.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done

  • No. This type of study involves human readers (e.g., radiologists) interpreting cases with and without AI assistance to measure improvement in diagnostic accuracy or efficiency. It is entirely irrelevant for a sterile surgical gown.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

  • Not Applicable. This concept applies to AI algorithms. A surgical gown has no algorithm.

7. The type of ground truth used

  • Standardized Test Methods and Biocompatibility Assessment: The "ground truth" here is adherence to the performance requirements specified in recognized industry standards (e.g., AAMI PB70, ASTM, ISO, 16 CFR Part 1610) for fluid barrier protection, physical strength, flammability, and biocompatibility. These are objective, measurable properties.

8. The sample size for the training set

  • Not Applicable. There is no "training set" for a surgical gown. The device is manufactured, and select samples are tested for quality control and regulatory submission.

9. How the ground truth for the training set was established

  • Not Applicable. See point 8.

U.S. Food & Drug Administration 510(k) Clearance Letter

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

May 5, 2025

Dae Myung Chemical Co., Ltd. (vietnam)
Tri Do Minh
QA Manager
Lot CI.II-7a and Lot CI.II-7a', Street No. 5,
Long Thanh Industrial Zone, Tam An Ward
Long Thanh, Dong Nai 810000
Vietnam

Re: K242937
Trade/Device Name: Dream Medi Sterile Surgical Gown (SurgicalGownUL)
Regulation Number: 21 CFR 878.4040
Regulation Name: Surgical Apparel
Regulatory Class: Class II
Product Code: FYA
Dated: June 14, 2024
Received: March 31, 2025

Dear Tri Do Minh:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K242937 - Tri Do Minh
Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K242937 - Tri Do Minh
Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Bifeng Qian -S

Bifeng Qian, M.D., Ph.D.
Assistant Director
DHT4C: Division of Infection Control Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known)
K242937

Device Name
Dream Medi Sterile Surgical Gown (SurgicalGownUL)

Indications for Use (Describe)
The Dream Medi Sterile Surgical Gowns are intended to be worn by operating room personnel during surgical procedures to protect the surgical patient and operating room personnel from the transfer of microorganisms, body fluids and particulate matter.

In addition, this surgical gown meets the requirements of AAMI Level 3 barrier protection for a surgical gown per ANSI/AAMI PB70 Liquid barrier performance and classification of protective apparel and drapes intended for use in health care facilities (AAMI PB70). This is a single use, disposable device, provided sterile.

Type of Use (Select one or both, as applicable)
☐ Prescription Use (Part 21 CFR 801 Subpart D)
☒ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

510(k) Summary K242937

(As required by 21 CFR 807.92)

1.0 Submitter's information

Name: DAE MYUNG CHEMICAL CO., LTD. (VIETNAM)
Address: Lot CI.II-7a and Lot CI.II-7a', Street No. 5, Long Thanh Industrial Zone, Tam An Ward, Long Thanh Dong Nai 810000 Vietnam
Contact: Mr. Tri Do Minh
Phone Number: +84 251-3514 017
Email: qc.tri@dmpoly.com
Date of Preparation: April 04, 2025

2.0 Device information

Trade name: Dream Medi Sterile Surgical Gown (SurgicalGownUL)
Common name: Surgical Gown
Classification name: Gown, Surgical
Model: SurgicalGownUL
Size: S, M, L, XL, XXL

3.0 Classification

Production code: FYA - Gown, Surgical
Regulation number: 21CFR 878.4040 - Surgical apparel
Classification: Class II
Panel: Surgical apparel

4.0 Predicate device information

Name: Hubei Woozon Healthcare Co.,Ltd.
Address: Nongfeng Road, Nonwoven Fabric Industrial Park, Wangzhou Avenue, Pengchang Town, Xiantao City, Hubei, China
K number: K222999

5.0 Intended Use/Indication for Use Statement

The Dream Medi Sterile Surgical Gowns are intended to be worn by operating room personnel during surgical procedures to protect the surgical patient and operating room personnel from the transfer of microorganisms, body fluids and particulate matter.

In addition, this surgical gown meets the requirements of AAMI Level 3 barrier protection for a surgical gown per ANSI/AAMI PB70 Liquid barrier performance and classification of protective apparel and drapes intended for use in health care facilities (AAMI PB70). This is a single use, disposable device, provided sterile.

6.0 Device description

Dream Medi Surgical Gowns are made from SMS material. They are full length, constructed with raglan sleeves, hook and loop neck closures, and tie waist closures. They provide AAMI Level-3 protection.

The Level 3 Gowns have five different sizes: Small (S), Medium (M), Large (L), Extra Large (XL), Extra Extra Large (XXL).

SMS is a multi-ply material consisting of layers of spunbond and meltblown polypropylene.

Page 6

The body is made from 44g Blue SMS and the sleeve is made from 54g Blue SMS. Sleeve opening is made from pure polyester. The collar closure is made from dacron. All gowns are sterilized with ethylene oxide.

7.0 Technological Characteristic Comparison Table

Table - General Comparison

FeatureDream Medi Sterile Surgical GownPredicate (K222999)Comparison
Regulatory ClassificationClass IIClass IISame
Regulation Number21 CFR 878.404021 CFR 878.4040Same
Product CodeFYAFYASame
Device NameSurgical GownSurgical GownSame
ModelsS, M, L, XL, XXLS, M, L, XL, XXLSame
Intended UseDream Medi Sterile Surgical Gown is intended to be worn by operating room personnel during surgical procedures to protect the surgical patient and operating room personnel from the transfer of microorganisms, body fluids and particulate material. This surgical gown meets the requirements of AAMI Level 3 barrier protection for a surgical gown per ANSI/AAMI PB70:2012 Liquid barrier performance and classification of protective apparel and drapes intended for use in health care facilities (AAMI PB70). Dream Medi Sterile Surgical Gown are single use, disposable medical devices; the Surgical Gowns were sterilized by EO sterilization based on ISO 11135:2014Surgical Gowns are intended to be worn by operating room personnel during surgical procedures to protect the surgical patient and operating room personnel from the transfer of microorganisms, body fluids and particulate material. This surgical gown meets the requirements of AAMI Level 3 barrier protection for a surgical gown per ANSI/AAMI PB70:2012 Liquid barrier performance and classification of protective apparel and drapes intended for use in health care facilities (AAMI PB70). The Surgical Gowns are single use, disposable medical devices; provided non-sterile. Before the use, the Surgical Gowns must be sterilized by EO sterilization based on ISO 11135:2014.Same
MaterialSMS Nonwoven FabricSMS Nonwoven FabricSame
ColorBlueBlueSame
SterilitySterile (by EO sterilization)Sterile (by EO sterilization)Same
Single UseYesYesSame
AAMI Barrier LevelLevel 3Level 3Same
Impact Penetration (AATCC 42)≤1.0 g≤1.0 gSame
Hydrostatic Resistance (AATCC 127)≥50 cm≥50 cmSimilar

Page 7

| Tensile Strength (ASTM D5034) | Machine Direction Wet: 115.44 N (25.9 lbf)Machine Direction Dry: 114.62 N (25.7 lbf)Cross Direction Wet: 88.5 N (19.89 lbf)Cross Direction Dry: 94.66 N (21.28 lbf) | Machine Direction Mean: 32.9 lbfCross Direction Mean: 26.3 lbf | Similar |
| Tear Resistance (ASTM D5587) | Machine Direction Mean: 22.59 NCross Direction Mean: 15.14 N | Machine Direction Mean: 9.3 lbfCross Direction Mean: 18.2 lbf | Similar |
| Seam Strength (ASTM D1683) | Mean 67.17 N | Shoulder: 48.7NArmhole: 33.9NSleeve: 49.6N | Similar |
| Flammability (16 CFR Part 1610) | Ignited but self-extinguished (Class 1) | Class 1, Non Flammable | Same |
| Lint Generation (ISO 9073-10) | Log10 < 4 | Log10 < 4 | Same |
| Cytotoxicity | No potential cytotoxicity | No potential cytotoxicity | Same |
| Irritation | No irritation | No irritation | Same |
| Sensitization | No sensitization | No sensitization | Same |
| Acute Systemic Toxicity | Not performed | No acute systemic toxicity in vivo | Different (Acute Systemic toxicity not required if Cytotoxicity passed.) |
| Sterilization Method | Ethylene Oxide (EO) | Ethylene Oxide (EO) | Similar |
| Shelf life | 03 years | 03 years | Same |

8.0 Summary of Non-Clinical Testing

Non-clinical testing was performed to confirm that the proposed device met all design specifications. The performance of the subject surgical gown was evaluated using established standards and test methods in accordance with FDA-recognized consensus standards and applicable guidance documents to ensure compliance with safety and performance requirements.

The test results demonstrated that the proposed device met its acceptance criteria or testing endpoints

PurposeTestAcceptance CriteriaResults
Impact PenetrationAATCC 42Level 3, ≤1.0 gPass

Page 8

| Hydrostatic Resistance | AATCC 127 | Level 3, ≥50 cm | Pass |
| Tensile strength | ASTM D5034 | ≥30 N (≥ 7 lbf) | Pass |
| Tear resistance | ASTM D5587 | ≥10 N (≥ 2.3 lbf) | Pass |
| Seam strength | ASTM D1683 | ≥30 N (≥ 7 lbf) | Pass |
| Lint and Other particles generation in the dry state | ISO 9073-10 | Log 10 < 4 | Pass |
| Flammability testing | 16 CFR Part 1610 | Class 1 | Pass |
| Biocompatibility cytotoxicity | ISO 10993-5 | Non-cytotoxic | Pass |
| Biocompatibility irritation | ISO 10993-23 | Non-irritant | Pass |
| Biocompatibility Sensitization | ISO 10993-10 | Non-sensitizing | Pass |

9.0 Summary of Clinical Testing

No clinical study was conducted.

10.0 Conclusion

The conclusion drawn from the non-clinical tests demonstrates that the subject device, the Dream Medi Sterile Surgical Gown, is as safe, as effective, and performs as well as or better than the legally marketed predicate device K222999.

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.