Search Filters

Search Results

Found 18 results

510(k) Data Aggregation

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    HemoSphere Alta™ Advanced Monitor with Swan-Ganz Technology
    The HemoSphere Alta monitor when used with the HemoSphere Alta Swan-Ganz patient cable and Edwards Swan-Ganz catheters is indicated for use in adult and petical care patients requiring monitoring of cardiac output (continuous [CO] and intermittent [iCO]) and derived hemodynamic parameters in a hospital environment. Pulmonary artery blood temperature monitoring is used to compute continuous and intermittent CO with thermodilution technologies. It may be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed in a hospital environment. Refer to the Edwards Swan-Ganz Ir catheter indications for use statement for information on target patient population specific to the catheter being used.

    The Global Hypoperfusion Index (GHI) algorithm provides the clinician with physiological insight into a patient's likelihood of future hemodynamic instability. The GHI algorithm is intended for use in surgical or non-surgical patients receiving advanced hemodynamic monitoring with the Swan-Ganz catheter. The GHI algorithm is considered to provide additional information regarding the patient's predicted future risk for clinical deterioration, as well as identifying patients at low risk for deterioration. The product predictions are for reference only and no therapeutic decisions should be made based solely on the GHI algorithm predictions.

    When used in combination with a Swan-Ganz catheter connected to a pressure transducer, the Edwards Lifesciences Smart Wedge algorithm measures and provides pulmonary artery occlusion pressure and assesses the quality of the pulmonary artery occlusion pressurement. The Smart Wedge algorithm is indicated for use in critical care patients over 18 years of age receiving advanced hemodynamic monitoring. The Smart Wedge algorithm is considered to be additional quantitative information regarding the patient's physiological condition for reference only and no therapeutic decisions should be made based solely on the Smart Wedge algorithm parameters.

    HemoSphere Alta Advanced Monitor with HemoSphere Oximetry Cable
    The HemoSphere Alta monitor when used with the HemoSphere oximetry cable and Edwards oximetry catheters is indicated for use in adult and pediatric critical care patients requiring of venous oxygen saturation (SvO2 and ScvO2) and derived hemodynamic parameters in a hospital environment. Refer to the Edwards oximetry catheter indications for use statement for information on target patient population specific to the catheter being used.

    HemoSphere Alta Advanced Monitor with HemoSphere Pressure Cable or HemoSphere Alta Monitor Pressure Cable
    The HemoSphere Alta monitor when used with the HemoSphere Pressure Cable or HemoSphere Alta monitor Pressure cable is indicated for use in adult and pediatric critical care patients in which the balance between cardiac fluid status, vascular resistance and pressure needs continuous assessment. It may be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. Refer to the Edwards FloTrac, FloTrac Jr, Acumen IQ, and TruWave DPT sensor indications for use statement for information on target patient population specific to the sensor being used.

    The Edwards Lifesciences Acumen Hypotension Prediction Index software feature provides the clinician with physiological insight into a patient's likelihood of future hypotensive events and the associated hemodynamics. The Acumen HPI feature is intended for use in surgical patients receiving advanced hemodynamic monitoring. The Acumen HPI feature is considered to be additional quantitative information regarding the patient's physiological condition for reference only and no therapeutic decisions should be made based solely on the Hypotension Prediction Index (HPI) parameter.

    When used in combination with the Swan-Ganz technology connected to a compatible Swan-Ganz catheter, the Edward Lifesciences Right Ventricular Pressure (RVP) algorithm provides the clinician with physiological insight into the hemodynamic status of the right ventricle of the heart. The RVP algorithm is indicated for critically ill patients over 18 years of age receiving advanced hemodynamic monitoring in the operating room (OR) and intensive care unit (ICU). The RVP algorithm is considered to be additional quantitative information regarding the patient's physiological condition for reference only and no therapeutic decisions should be made based solely on the Right Ventricular Pressure (RVP) parameters.

    When used in combination with the HemoSphere Pressure Cable connected to a compatible Swan-Ganz catheter, the Right Ventricular Cardiac Output (RVCO) feature provides the clinician with physiological insight into the hemodynamic status of the right ventricle of the heart. The RVCO algorithm is intended for use in surgical patients over 18 years of age that require advanced hemodynamic monitoring. The Right Ventricular Cardiac a continuous cardiac output and derived parameters.

    The Cerebral Adaptive Index (CAI) Algorithm is an informational index to help assess the level of coherence or lack thereof between Mean Arterial Pressure (MAP) and the Absolute Levels of Blood Oxygenation Saturation (StO2) in patient's cerebral tissue. MAP is acquired by the HemoSphere pressure cable or HemoSphere Alta Pressure Cable and StO2 is acquired by the ForeSight oximeter cable. CAI is intended for use in patients over 18 years of age receiving advanced hemodynamic monitoring. CAI is not indicated to be used for treatment of any disease or condition and no therapeutic decisions should be made based solely on the Cerebral Adaptive Index (CAI) Algorithm.

    HemoSphere Alta Advanced Monitor with Acumen Assisted Fluid Management Feature and Acumen IQ Sensor
    The Acumen assisted fluid management (AFM) software feature provides the clinician with physiological insight into a patient's estimated response to fluid therapy and the associated hemodynamics. The Acumen AFM software feature is intended for use in surgical patients ≥18 years of age, that require advanced hemodynamic monitoring. The Acumen AFM software feature offers suggestions regarding the patient's physiological condition and estimated response to fluid therapy.

    Acumen AFM fluid administration suggestions are offered to the clinician; the decision to administer a fluid bolus is made by the clinician, based upon review of the patient's hemodynamics. No therapeutic decisions should be made based solely on the assisted fluid management suggestions.

    Acumen IQ Fluid Meter
    The Acumen IQ fluid meter is a sterile single use device that is intended to be used with the HemoSphere Alta AFM cable and AFM software feature to inform the user of the rate of flow. The device is intended to be used by qualified personnel or clinicians in a clinical setting for up to 24 hours.

    HemoSphere Alta Advanced Monitor with ForeSight Oximeter Cable
    The non-invasive ForeSight oximeter cable is intended for use as an adjunct monitor of absolute regional hemoglobin oxygen saturation of blood under the sensors in individuals at risk for reduced flow or no-flow ischemic states. The ForeSight oximeter cable is also intended to monitor relative changes of total hemoglobin of blood under the sensors. The ForeSight oximeter cable is intended to allow for the display of StO2 and relative change in total hemoglobin on the HemoSphere Alta monitor.

    • When used with large sensors, the ForeSight Oximeter Cable is indicated for use on adults and transitional adolescents ≥40 kg.
    · When used with Medium Sensors, the ForeSight Oximeter Cable is indicated for use on pediatric subjects ≥3 kg.
    · When used with Small Sensors, the ForeSight Oximeter Cable is indicated for cerebral use on pediatric subjects < 8 kg and non-cerebral use on pediatric subjects <5kg.

    The Edwards Algorithm for Measurement of Blood Hemoglobin is indicated for continuously monitoring changes to hemoglobin concentration in the circulating blood of adults and transitional adolescents ≥ 40 kg receiving advanced hemodynamic monitoring using HemoSphere ForeSight Oximeter Cable and ForeSight IQ Sensors in cerebral locations.

    HemoSphere Alta Advanced Monitor with ClearSight Technology
    The HemoSphere Alta Monitor when used with the HemoSphere ClearSight technology, pressure controller and a compatible Edwards finger cuff are indicated for Adult and Pediatric patients of age in which the balance between cardiac function, fluid status and vascular resistance needs continuous assessment. It may be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. In addition, the noninvasive system is indicated for use in patients with co-morbidities for which hemodynamic optimization is desired and invasive measurements are difficult. The HemoSphere Alta monitor and compatible Edwards' finger cuffs noninvasively measures blood pressure and associated hemodynamic parameters.

    The Edwards Lifesciences Acumen Hypotension Prediction Index feature provides the clinician with physiological insight into a patient's likelihood of future hypotensive events and the associated hemodynamics. The Acure is intended for use in surgical patients or non-surgical patients receiving advanced hemodynamic monitoring. The Acumen HPI feature is considered to be additional quantitative information regarding the patient's physiological condition for reference only and no therapeutic decisions should be made based solely on the Hypotension Prediction Index (HPI) parameter.

    Device Description

    The HemoSphere Alta Advanced Monitoring Platform is Edwards' next-generation platform that provides a means to interact with and visualize hemodynamic and volumetric data on a screen. It incorporates a comprehensive view of patient hemodynamic parameters with an intuitive and easy user interface. The HemoSphere Alta Advanced Monitoring Platform is designed to provide monitoring of cardiac flow with various core technologies coupled with other technologies-based features such as Algorithms and Interactions. It integrates Edwards existing Critical Care technologies into a unified platform.

    The Right Ventricular Cardiac Output (RVCO) feature is a machinelearning algorithm that calculates and displays continuous cardiac output (CORV) from the right ventricle using as inputs the right ventricular pressure waveform and derived right ventricular pressure parameters such as SYSRVF, DIARVP, MRVP, RVEDP, PRRV and Max RV dP/dt from the existing Right Ventricular Pressure (RVP) algorithm and if available, intermittent cardiac output (iCO).

    AI/ML Overview

    The provided text describes the HemoSphere Alta Advanced Monitoring Platform and its various features, as well as the testing conducted to support its 510(k) clearance. However, it does not contain specific acceptance criteria and detailed device performance data in the format of a table, nor does it provide a detailed study that proves the device meets specific acceptance criteria for any of its algorithms.

    The document makes general statements about testing, such as:

    • "Completion of all verification and validation activities demonstrated that the subject devices meet their predetermined design and performance specifications."
    • "Measured and derived parameters were tested using a bench simulation."
    • "All tests passed."
    • "Software verification testing were conducted, and documentation was provided per FDA's Guidance..." "All tests passed."
    • "Usability study was conducted per FDA's guidance document... The usability study demonstrated that the intended users can perform primary operating functions and critical tasks of the system without any usability issues that may lead to patient or user harm."

    While it mentions the Right Ventricular Cardiac Output (RVCO) algorithm as a new algorithm and states that "clinical data (waveforms) were collected in support of the design and validation of the RVCO algorithm," it does not present the detailed results of this validation study, nor does it define specific acceptance criteria for the RVCO algorithm and its performance against those criteria.

    Therefore,Based on the provided text, I cannot provide the requested information in the form of a table of acceptance criteria and reported device performance for any specific algorithm, nor can I describe a detailed study that proves the device meets these criteria. The document contains general statements about testing and compliance but lacks the specific quantitative data and study design details needed to answer all aspects of your request.

    To provide a complete answer, specific study reports and performance data would be required, which are not present in the provided FDA 510(k) summary.

    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Swan-Ganz pacing-TD catheters are indicated for use for temporary, atrial, ventricular, or A-V sequential cardiac pacing of adult surgical and/or critically ill patients. This includes, but is not limited to, patients with arrythmias, acute myocardial infarction, cardiac surgery, and coronary angiography. In addition, these catheters are indicated for use in hemodynamic monitoring of patients experiencing medical conditions such as, but not limited to, post-major surgical recovery, trauma, sepsis, burns, pulmonary disease, pulmonary failure, stroke, and cardiac disease including heart failure.

    The Swan-Ganz bipolar pacing catheters are indicated for use in surgical and/or critically ill cardiac patients who may benefit from temporary, transvenous right ventricular pacing.

    The Chandler transluminal V-pacing probe is a transluminal bipolar pacing probe, which must be used with a Swan-Ganz Paceport catheter is indicated for use in adult surgical and/or critically ill cardiac patients. This includes, but is not limited to, patients with arrhythmias, acute myocardial infarction, cardiac surgery and coronary angiography.

    Device Description

    The Swan-Ganz Pacing Catheters and Pacing Probes are a series of specialized catheters and pacing probes, which are temporarily inserted into the central circulatory system vasculature to facilitate temporary atrial, ventricular, or A-V sequential pacing and to serve as diagnostic and/or therapeutic tools. The catheters may be used to overdrive suppression of arrhythmias and diagnosis of complex arrhythmias. The pacing catheters and probes may be used for ECG detection during placement but are not intended to be used for ECG monitoring.

    AI/ML Overview

    I am sorry, but based on the provided text, there is no information about the acceptance criteria and study proving the device meets those criteria in the context of an AI/ML powered medical device. The document is a 510(k) summary for a set of physical medical devices (catheters and pacing probes) and does not mention any AI or machine learning components.

    Therefore, I cannot provide the requested information, including:

    1. A table of acceptance criteria and the reported device performance
    2. Sample size used for the test set and data provenance
    3. Number of experts used to establish ground truth and their qualifications
    4. Adjudication method for the test set
    5. MRMC comparative effectiveness study results or effect size
    6. Standalone (algorithm only) performance
    7. Type of ground truth used
    8. Sample size for the training set
    9. How the ground truth for the training set was established

    The document focuses on the substantial equivalence of the Swan-Ganz Pacing Probe and Catheters to predicate devices based on design, material, chemical composition, principle of operation, and compliance with biocompatibility and bench testing standards.

    Ask a Question

    Ask a specific question about this device

    K Number
    K233824

    Validate with FDA (Live)

    Date Cleared
    2024-06-06

    (188 days)

    Product Code
    Regulation Number
    870.1240
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Swan-Ganz catheters are diagnostic and monitoring tools used for hemodynamic monitoring of adult critically ill patients including but not limited to post major surgical recovery, trauma, sepsis, burns, pulmonary disease, pulmonary failure, cardiac disease including heart failure.

    Device Description

    The Swan-Ganz catheters are flow-directed pulmonary artery catheters used to monitor hemodynamic pressures. The Swan-Ganz thermodilution catheters provide diagnostic information to rapidly determine hemodynamic pressures and cardiac output when used with a compatible cardiac output computer.

    AI/ML Overview

    The provided text is an FDA 510(k) clearance letter and summary for the Swan-Ganz catheter. It primarily details the regulatory process, device description, and indications for use.

    Crucially, it does not contain information about acceptance criteria, device performance studies, sample sizes, ground truth establishment, or expert qualifications in the context of proving device performance against specific metrics. The document explicitly states:

    "The subject Swan-Ganz Base and Advanced catheters are identical to the predicate devices cleared in K160084 and K222117 in terms of design, performance specifications, and technological characteristics with the exception of the indications for use statement and other portions of the labeling. The are no changes to the design, technology, performance, materials, or specifications of the devices in this 510(k). The modifications to the subject devices are limited to labeling changes."

    This indicates that the current submission (K233824) is for a labeling change only, and therefore, the performance of the device itself (its ability to accurately monitor hemodynamic pressures) would have been established during the clearance of the predicate devices (K160084 and K222117), not in this particular submission.

    Therefore, I cannot provide the requested information from the given text. The text does not describe an acceptance criteria or a study proving the device meets acceptance criteria related to its performance.

    Ask a Question

    Ask a specific question about this device

    K Number
    K231248

    Validate with FDA (Live)

    Date Cleared
    2023-09-21

    (143 days)

    Product Code
    Regulation Number
    870.1240
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Swan-Ganz catheters:
    The Swan-Ganz catheters are diagnostic and monitoring tools used for hemodynamic monitoring of adult critically ill patients including but limited to post major surgical recovery, trauma, sepsis, burns, pulmonary disease, pulmonary failure, cardiac disease including heart failure.
    Models 096F6, 096F6P, TS105F5, 132F5, 131F7, 131F7P 831F75, 831F75P, 834F75, and 834F75P are intended for adult and pediatric patients:
    The Swan-Ganz catheters are diagnostic and monitoring tools used for hemodynamic monitoring of critically ill adult and pediatric patients ≥ 12 years of age including but not limited to post major surgical recovery, trauma, sepsis, burns, pulmonary disease, pulmonary failure, cardiac disease including heart failure.

    FloTrac sensors:
    The FloTrac sensor is indicated for use in intravascular pressure monitoring. It is also indicated for use with the Edwards arterial pressure based cardiac output monitoring devices or hardware to measure cardiac output. They are intended to be used in adult patients.
    Models MHD6, MHD8, MHD65, and MHD85 are intended for adult and pediatric patients:
    The FloTrac sensor is indicated for use in intravascular pressure monitoring. It is also indicated for use with the Edwards arterial pressure based cardiac output monitoring devices or hardware to measure cardiac output. The FloTrac sensor is indicated for use in adult and pediatric patients ≥ 12 years of age.

    ClearSight finger cuffs:
    The Acumen IQ are indicated for patients over 18 years of age to non-invasively measure blood pressure and associated hemodynamic parameters when used with EV1000 clinical platform or HemoSphere Advanced Monitoring Platform.
    The ClearSight finger cuffs are indicated for adult and pediatric patients ≥ 12 years of age to noninvasively measure blood pressure and associated hemodynamic parameters when used with EV1000 clinical platform or HemoSphere Advanced Monitoring Platform.

    HemoSphere Advanced Monitor with HemoSphere Swan-Ganz Module:
    The HemoSphere advanced monitor when used with the HemoSphere Swan-Ganz module and Edwards Swan-Ganz catheters is indicated for use in adult and pediatric critical care patients requiring monitoring of cardiac output (continuous [CO] and intermittent [iCO]) and derived hemodynamic parameters in a hospital environment. It may be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. Refer to the Edwards Swan-Ganz catheter indications for use statement for information on target patient population specific to the catheter being used.
    Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.

    HemoSphere Advanced Monitor with HemoSphere Pressure Cable (compatible with FloTrac sensors):
    The HemoSphere advanced monitor when used with the HemoSphere pressure cable is indicated for use in adult and pediatric critical care patients in which the balance between cardiac function, fluid status, vascular resistance and pressure needs continuous assessment. It may be used for monitoring of hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. Refer to the Edwards FloTrac sensor, Acumen IQ sensor, and TruWave DPT indications for use statements for information on target patient populations specific to the sensor/transducer being used.
    The Edwards Acumen Hypotension Prediction Index feature provides the clinician with physiological insight into a patient's likelihood of future hypotensive events (defined as mean arterial pressure < 65 mmHg for at least one minute in duration) and the associated hemodynamics. The Acumen HPI™ feature is intended for use in surgical or non-surgical patients receiving advanced hemodynamic monitoring. The Acumen HPI™ feature is considered to be additional quantitative information regarding the patient's physiological condition for reference only and no therapeutic decisions should be made based solely on the Acumen Hypotension Prediction Index (HPI) parameter.
    Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.

    HemoSphere Advanced Monitor with ClearSight Module:
    The HemoSphere advanced monitor when used with the HemoSphere ClearSight module, pressure controller and a compatible Edwards finger cuff are indicated for adult and pediatric patients in which the balance between cardiac function, fluid status and vascular resistance needs continuous assessment. It may be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. In addition, the noninvasive system is indicated for use in patients with co-morbidities for which hemodynamic optimization is desired and invasive measurements are difficult. The HemoSphere advanced monitor and compatible Edwards finger cuffs noninvasively measures blood pressure and associated hemodynamic parameters.
    Refer to the ClearSight finger cuff and Acumen IQ finger cuff indications for use statements for information on target patient population specific to the finger cuff being used.
    The Edwards Acumen Hypotension Prediction Index (HPI) feature provides the clinician with physiological insight into patient's likelihood of future hypotensive events (defined as mean arterial pressure < 65 mmHg for at least one minute in duration) and the associated hemodynamics. The Acumen HPI feature is intended for use in surgical patients receiving advanced hemodynamic monitoring. The Acumen HPI feature is considered to be additional quantitative information regarding the patient's physiological condition for reference only and no therapeutic decisions should be made based solely on the Acumen Hypotension Prediction Index (HPI) parameter.
    Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.

    Device Description

    Swan-Ganz Catheters:
    The Swan-Ganz catheters are flow-directed pulmonary artery catheters used to monitor hemodynamic pressures. The Swan-Ganz thermodilution catheters provide diagnostic information to rapidly determine hemodynamic pressures and cardiac output when used with a compatible cardiac output computer.

    FloTrac Sensors:
    The FloTrac sensor is a sterile, single use kit that monitors pressures when attached to pressure monitoring catheters. When connected to a compatible monitor, the FloTrac sensor minimally-invasively measures cardiac output and key hemodynamic parameters, which assist the clinician in assessing the patient's physiologic status and support clinical decisions related to hemodynamic optimization. The disposable sterile cable, with a red-connector interfaces, exclusively with an Edwards cable that is specifically wired for the pressure monitor being used. The disposable sterile cable, with a green-connector interfaces, exclusively with the Edwards cables for use with the Edwards arterial pressure based cardiac output monitoring devices or hardware.
    The FloTrac sensor has a straight, flow-through design across the pressure sensors with an integral flush device.

    ClearSight finger cuffs:
    The ClearSight finger cuffs, when used with an appropriate Edwards monitoring system, provide continuous, noninvasive hemodynamic monitoring. The ClearSight finger cuffs utilize the volume-clamp method to measure blood pressure with an inflatable bladder wrapped around the middle phalanx of the finger.

    HemoSphere Advanced Monitor Platform:
    The HemoSphere Advanced Monitoring Platform was designed to simplify the customer experience by providing one platform with modular solutions for their hemodynamic monitoring needs. The user can choose from the available optional sub-system modules or use multiple sub-system modules at the same time. This modular approach provides the customer with the choice of purchasing and/or using specific monitoring applications based on their needs. Users are not required to have all of the modules installed at the same time for the platform to function.

    AI/ML Overview

    This document is a 510(k) summary for Edwards Lifesciences' Swan-Ganz catheters, FloTrac sensors, ClearSight finger cuffs, and HemoSphere Advanced Monitoring Platform. The core of this submission is to expand the indications for use of these devices to include pediatric patients (≥12 years of age), in addition to their existing adult indications.

    The information regarding acceptance criteria and the study that proves the device meets those criteria, as detailed in your request, is primarily found in the "Device Verification and Validation" and "Conclusion" sections. However, the FDA summary document only provides a high-level overview. It does not contain the detailed acceptance criteria table, precise performance metrics, sample sizes, ground truth establishment methods, or specific details about expert adjudication or MRMC studies, which are typically found in the full 510(k) submission or supporting clinical trial reports.

    Based on the provided text, here's what can be extracted and what information is not present:

    Key Takeaway from the FDA 510(k) Summary:

    The primary purpose of this 510(k) submission is to expand the indicated patient population for the listed devices from adult-only to include pediatric patients aged 12 years and older. The submission claims that the devices perform similarly in this expanded pediatric population compared to adults, meeting predefined acceptance criteria.


    Table of Acceptance Criteria and Reported Device Performance

    Not Provided in this Document. The document states "within predefined acceptance criteria" but does not list these criteria or the specific reported device performance metrics against them.

    Acceptance Criteria CategorySpecific Metric (e.g., Accuracy, Precision, AUC Threshold)Acceptance ThresholdReported Device Performance
    Hemodynamic Parameter MeasurementNot specifiedNot specifiedNot specified
    SafetyNot specifiedNot specified"no issues of safety"
    EffectivenessNot specifiedNot specified"effective in measuring cardiac output"
    ConsistencyNot specifiedNot specified"consistent and within predefined acceptance criteria." "similar performance to an adult population."

    Study Details (Based on available information):

    1. Sample sizes used for the test set and the data provenance:

      • Sample Size: Not explicitly stated. The document mentions "the target pediatric population in a clinical study" but does not provide the number of patients.
      • Data Provenance:
        • Country of Origin: Not specified in this summary. Clinical studies are often multi-center, but no specific locations are mentioned.
        • Retrospective or Prospective: Not explicitly stated, but the phrase "clinical study" typically implies prospective data collection for regulatory submissions, especially for expanding indications to a new population.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Number of Experts: Not specified.
      • Qualifications of Experts: Not specified. Given the nature of these hemodynamic devices, ground truth would likely be established by clinical measurements and potentially other validated reference methods, rather than expert interpretation of images or signals in the same way a radiologist might interpret an X-ray. Clinical experts (e.g., intensivists, anesthesiologists, cardiologists) would be involved in patient selection, data collection, and overseeing the study.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • Adjudication Method: Not specified. This type of adjudication (e.g., multi-reader consensus) is more common in image-based diagnostic AI studies. For device performance measurement, agreement with a reference standard is the typical method.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • MRMC Study: Not applicable/not performed for this type of device and study. MRMC studies are primarily for evaluating AI's impact on human interpretation tasks (e.g., radiologists reading images). This submission is for devices that directly measure physiological parameters and provide data, with one feature (Acumen HPI) providing "physiological insight" for reference. The study focuses on the device's performance in a new patient population, not on how it assists human readers in interpreting complex cases, except perhaps by providing accessible data. The Acumen HPI feature specifically states "no therapeutic decisions should be made based solely on the Acumen Hypotension Prediction Index (HPI) parameter," indicating it's an informational tool rather than a definitive diagnostic AI assist requiring a MRMC study.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Standalone Performance: The study evaluated the devices' performance in the pediatric population. The devices themselves are "standalone" in that they acquire and process physiological signals to output parameters. The statement "The data demonstrate the performance in the expanded pediatric population ≥ 12 years of age and is substantially equivalent to their respective predicate devices for adult patient population ≥ 18 years of age" suggests an evaluation of the device's measurement accuracy and consistency.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • Type of Ground Truth: Not explicitly stated, but for hemodynamic monitoring devices, ground truth is typically established by:
        • Reference Standards/Comparative Measurements: Comparing the device's measurements (e.g., cardiac output, blood pressure) against established, precise, and often more invasive or gold-standard methods (e.g., thermodilution, invasive arterial line measurements from validated transducers).
        • Clinical Outcomes/Safety Data: Demonstrating safety and effectiveness in the target population through clinical monitoring and observation.
      • The document states "clinical data presented in this 510(k) demonstrates the performance of hemodynamic parameters expanded to pediatric patients ≥ 12 years of age... are consistent and within predefined acceptance criteria," implying comparisons to a reference or expected range.
    7. The sample size for the training set:

      • Training Set Size: Not applicable/not specified. These devices measure physiological parameters; they are not "AI algorithms" in the sense of needing a large training dataset to learn patterns for classification or prediction (except for the Acumen HPI feature, but no details on its development or training are provided here). The focus of this 510(k) is the application of existing technology to a new age group, not the development of a new pattern-recognition algorithm.
    8. How the ground truth for the training set was established:

      • Ground Truth for Training Set: Not applicable, as there's no mention of a traditional AI training set. The devices likely use established physiological models and algorithms for calculating parameters. If the Acumen HPI feature involved machine learning, its training ground truth would likely be based on recorded physiological data and subsequent hypotensive events. However, the document does not elaborate on the development of this specific feature in the context of this 510(k).
    Ask a Question

    Ask a specific question about this device

    K Number
    K222117

    Validate with FDA (Live)

    Date Cleared
    2022-12-09

    (144 days)

    Product Code
    Regulation Number
    870.1240
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The primary indications for the Swan-Ganz IQ Pulmonary Arterial catheter includes:

    • Acute heart failure
    • Severe hypovolemia
    • Complex circulatory situations
    • Medical emergencies
    • Adult respiratory distress syndrome
    • Gram negative sepsis
    • Drug intoxication
    • Acute renal failure
    • Hemorrhagic pancreatitis
    • Intra and post-operative management of high risk patients
    • History of pulmonary or cardiac disease
    • Fluid shifts (e.g., extensive intra-abdominal operations)
    • Management of high-risk obstetrical patients
    • Diagnosed cardiac disease
    • Toxemia
    • Premature separation of placenta
    • Cardiac output determinations
    • Differential diagnosis of mitral regurgitation and ventricular septal rupture
    • Diagnosis of cardiac tamponade

    Secondary indications include the following:

    • Blood Sampling
    • Infusion of saline and dextrose solutions

    Intended Use- Swan-Ganz IQ catheter:
    The Swan-Ganz IQ catheter (model AIQSGF8) is for use in patients who require hemodynamic monitoring. It is intended to be used in combination with clinical pressure monitoring, equipment to measure right heart and pulmonary artery pressures, and with a compatible cardiac output computer to measure intermittent cardiac output. Model AIOSGF8 also measures mixed venous oxygen saturation.

    Device Description

    The Edwards Swan-Ganz catheters are used to monitor the hemodynamic status of critically ill and injured patients. The existing catheters give clinicians the ability to measure right heart pressures, pulmonary artery occlusion pressure ("wedge"), sample mixed venous blood from the pulmonary artery, as well as measure cardiac output through thermodilution when used with a bedside physiologic monitor and pressure transducers.

    The Edwards Swan-Ganz product line is comprised of Standard (Base) and Advanced Technology monitoring catheters. The standard Swan-Ganz catheter measures: right heart pressures, thermodilution cardiac output (room temperature and chilled) and provides a mechanism for pulmonary artery blood sampling for laboratory analysis.

    The subject Swan-Ganz IO catheter combines from the existing Advanced catheter design (model 777F8) and the Paceport catheter design (model 780F75M). Using the 777F8 model as the base design, the new catheter modifies the design to remove the thermal filament functionality and replaces it with an additional port which will allow for pressure monitoring at the right ventricular level. The RV port is utilized just like the Swan-Ganz Paceport Oximetry catheter model 780F75M.

    Just like the existing advanced catheters, the Swan-Ganz IQ catheter provides the same functionality as the standard and Advanced Technology Swan-Ganz catheters by providing the ability to continuously monitor the patient's balance between oxygen delivery and consumption as well as the ability to help investigate the root cause of an imbalance through analysis of the components of stroke volume (preload, afterload, and contractility).

    AI/ML Overview

    The provided text is a 510(k) summary for the Edwards Lifesciences Swan-Ganz IQ Pulmonary Arterial Catheter. This document is a premarket notification to the FDA to demonstrate that the device is substantially equivalent to a legally marketed predicate device.

    Crucially, this document is NOT for an AI/ML device. It describes a physical medical device (a catheter) and its modifications. Therefore, the concept of "acceptance criteria" and "study proving the device meets the acceptance criteria" in the context of AI/ML performance metrics (like sensitivity, specificity, or human-in-the-loop improvement) is not applicable here.

    The document discusses "functional and performance testing" which are typical for physical medical devices, such as:

    • Packaging, shelf life, sterilization, biocompatibility
    • Bench testing:
      • Frequency response testing
      • Pressure tubing pull testing
      • Lumen burst pressure testing per ISO 10555-1:2013 and IEC 60601:2015 Amd.1:2020.

    The conclusion states that "Completion of all performance verification and validation activities demonstrated that the subject device meets its predetermined design and performance specifications." This refers to engineering and safety specifications of a physical device, not the performance of an AI algorithm on a dataset.

    Therefore, I cannot provide the requested information (table of acceptance criteria, sample sizes for AI test/training sets, expert ground truth establishment, MRMC studies, standalone performance, etc.) because the device described in this 510(k) summary is a physical catheter, not an AI/ML diagnostic tool.

    Ask a Question

    Ask a specific question about this device

    K Number
    K193466

    Validate with FDA (Live)

    Date Cleared
    2020-02-16

    (62 days)

    Product Code
    Regulation Number
    870.1240
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Swan-Ganz flow-directed monitoring catheters are indicated for the assessment of a patient's hemodynamic condition through direct intracardiac and pulmonary artery pressure monitoring. Secondary indications are for sampling blood and infusing solutions.

    Swan-Ganz thermodilution catheters are indicated for the assessment of a patient's hemodynamic condition through direct intracardiac and pulmonary artery pressure monitoring, cardiac output determination, and for infusing solutions. The distal (pulmonary artery) port also allows sampling of mixed venous blood for the assessment of oxygen transport balance and the calculation of derived parameters such as oxygen consumption, oxygen utilization coefficient, and intrapulmonary shunt fraction.

    ControlCath thermodilution catheters and Torque Support Thermodilution catheters are indicated for the assessment of a patient's hemodynamic condition through direct intracardiac and pulmonary artery pressure monitoring, cardiac output determination, and for infusing solutions. The distal (pulmonary artery) port also allows sampling of mixed venous blood for the assessment of oxygen transport balance and the calculation of derived parameters such as oxygen consumption, oxygen utilization coefficient, and intrapulmonary shunt fraction.

    Swan-Ganz oximetry TD catheters are indicated for the assessment of a patient's hemodynamic condition through direct intracardiac and pulmonary artery pressure monitoring, cardiac output determination, continuous mixed venous oxygen saturation monitoring, and for infusing solutions. For 631NX55, the distal (pulmonary artery) port also allows sampling of mixed venous blood for the assessment of oxygen transport balance and the calculation of derived parameters such as oxygen consumption, oxygen utilization coefficient, and intrapulmonary shunt fraction.

    Device Description

    The Swan-Ganz catheters are cardiovascular diagnostic catheters intended for use on critical care patients. Swan-Ganz catheters are used to monitor the hemodynamic status of critically ill and injured patients. The catheters give clinicians the ability to measure right heart pressures, pulmonary artery occlusion pressure ("wedge"), sample mixed venous blood from the pulmonary artery, as well as measure cardiac output through thermodilution when used with a bedside physiologic monitor.

    The subject Edwards Swan-Ganz product line is comprised of Base and Advanced Technology monitoring catheters. The Base Swan-Ganz catheter measures: right heart pressures, thermodilution cardiac output (room temperature and chilled) and provides a mechanism for pulmonary artery blood sampling for laboratory analysis. In addition to providing most of the same functionality as the Base models, the Advanced Swan-Ganz catheters provide the ability to continuously monitor the patient's mixed venous oxygen saturation, the balance between oxygen delivery and consumption, when used with an Edwards Lifesciences oximetry monitor or compatible bedside module system. It also monitors cardiac output with a compatible cardiac output computer.

    AI/ML Overview

    This is a 510(k) premarket notification for a Class II medical device, specifically Swan-Ganz Flow-Directed Monitoring Catheters. The document is about demonstrating substantial equivalence to a legally marketed predicate device, not about proving the performance of a novel AI/ML device against specific acceptance criteria in a clinical study. As such, most of the requested information (e.g., sample size for test set, number of experts for ground truth, MRMC study, standalone performance) is not applicable or present in this document.

    The "acceptance criteria" discussed here are related to demonstrating that the modified device (with SEBS balloons) performs equivalently to the predicate device (with latex balloons) in terms of its functional and safety aspects.

    Here's the information that can be extracted or deduced from the provided text:

    1. A table of acceptance criteria and the reported device performance:

    Acceptance Criteria (Evaluation Areas)Reported Device Performance
    Functional PerformancePassed (performance not affected)
    BiocompatibilityPassed (in accordance with ISO 10993-1:2018 and FDA guidance)
    SterilizationPassed (in accordance with ISO 11135:2014 and FDA Quality System Regulation)
    Usability and Human FactorsPassed (in accordance with IEC 62366:2015 and FDA guidance)
    Overall Substantial EquivalenceDemonstrated to be substantially equivalent to the predicate device for intended use.

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not specified for any of the tests. The document refers to "functional testing of the catheter balloon," "biocompatibility testing," "sterilization testing," and "usability and human factors testing" but does not provide quantities for the test articles or subjects involved in these tests.
    • Data Provenance: Not explicitly stated. The tests are bench studies and laboratory tests related to device manufacturing and biological effects, not clinical data from patients.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. This information is typically relevant for studies involving human interpretation or subjective assessment of medical images or conditions where expert consensus is needed to establish ground truth. The tests described (functional, biocompatibility, sterilization, usability) are objective engineering and biological assessments.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not applicable. Adjudication methods are used in clinical studies, particularly for diagnostic devices, to resolve disagreements among expert readers when establishing ground truth. The tests mentioned are objective evaluations and do not involve human adjudication in this context.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC comparative effectiveness study was not done. This type of study is relevant for AI/ML diagnostic devices. The device described is a medical catheter and the submission focuses on demonstrating substantial equivalence of a modified material.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This submission is not for an AI/ML algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Objective Test Standards and Regulations: The "ground truth" for the performance claims comes from objective engineering and biological standards and regulations.
      • Functional testing: Likely based on pre-defined engineering specifications for balloon performance.
      • Biocompatibility: ISO 10993-1:2018 and FDA guidance.
      • Sterilization: ISO 11135:2014 and FDA Quality System Regulation (21 CFR 820.30).
      • Usability and Human Factors: IEC 62366:2015 and FDA guidance.

    8. The sample size for the training set:

    • Not applicable. This submission is not for an AI/ML device that requires a training set.

    9. How the ground truth for the training set was established:

    • Not applicable. As above, no training set is relevant for this device type.
    Ask a Question

    Ask a specific question about this device

    K Number
    K160084

    Validate with FDA (Live)

    Date Cleared
    2016-05-03

    (109 days)

    Product Code
    Regulation Number
    870.1240
    Age Range
    All
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Swan-Ganz flow-directed monitoring catheters are indicated for the assessment of a patient's hemodynamic condition through direct intracardiac and pulmonary artery pressure monitoring. Secondary indications are for sampling blood and infusing solutions.

    Swan-Ganz thermodilution catheters are indicated for the assessment of a patient's hemodynamic condition through direct intracardiac and pulmonary artery pressure monitoring, cardiac output determination, and for infusing solutions. The distal (pulmonary artery) port also allows sampling of mixed venous blood for the assessment of oxygen transport balance and the calculation of derived parameters such as oxygen consumption, oxygen utilization coefficient, and intrapulmonary shunt fraction.

    The Swan-Ganz thermodilution Paceport catheters (Model 931F75) are used for assessment of a patient's hemodynamic condition through direct intracardiac and pulmonary artery pressure monitoring, cardiac output determination, and for infusing solutions. The Paceport catheter (Model 931F75) may also be used for standby temporary ventricular pacing. The distal (pulmonary artery) port also allows sampling of mixed venous blood for the assessment of oxygen transport balance and the calculation of derived parameters such as oxygen consumption, oxygen utilization coefficient, and intrapulmonary shunt fraction.

    The Swan-Ganz thermodilution A-V Paceport catheter (Model 991F8) is indicated for the assessment of a patient's hemodynamic condition through simultaneous right atrial, right ventricular, and pulmonary artery or wedge pressure monitoring, cardiac output determination, and for infusing solutions. The A-V Paceport catheter (Model 991F8) is also indicated for standby temporary ventricular, atrial, or A-V sequential pacing using the Model D98100 Chandler Transluminal V-pacing probe and/or Model D98500 A-pacing probe. The distal (pulmonary artery) port also allows sampling of mixed venous blood for the assessment of oxygen transport balance and the calculation of derived parameters such as oxygen consumption, oxygen utilization coefficient, and intrapulmonary shunt fraction.

    ControlCath thermodilution catheters are indicated for the assessment of a patient's hemodynamic condition through direct intracardiac and pulmonary artery pressure monitoring, cardiac output determination, and for infusing solutions. The distal (pulmonary artery) port also allows sampling of mixed venous blood for the assessment of oxygen transport balance and the calculation of derived parameters such as oxygen consumption, oxygen utilization coefficient, and intrapulmonary shunt fraction.

    Torque Support Thermodilution catheters are indicated for the assessment of a patient's hemodynamic condition through direct intracardiac and pulmonary artery pressure monitoring, cardiac output determination, and for infusing solutions. The distal (pulmonary artery) port also allows sampling of mixed venous blood for the assessment of oxygen transport balance and the calculation of derived parameters such as oxygen consumption, oxygen utilization coefficient, and intrapulmonary shunt fraction.

    The primary indications for the Swan-Ganz CCO thermodilution catheters include: Acute heart failure, Severe hypovolemia, Complex circulatory situations, Medical emergencies, Adult respiratory distress syndrome, Gram negative sepsis, Drug intoxication, Acute renal failure, Hemorrhagic pancreatitis, Intra and post-operative management of high risk patients, History of pulmonary or cardiac disease, Fluid shifts (e.g., extensive intra-abdominal operations), Management of high-risk obstetrical patients, Diagnosed cardiac disease, Toxemia, Premature separation of placenta, Cardiac output determinations, Differential diagnosis of mitral regurgitation and ventricular septal rupture, Diagnosis of cardiac tamponade. Models with CEDV capabilities are also indicated for volumetric determinations. Secondary indications include the following: Blood Sampling, Infusion of saline and dextrose solutions.

    The primary indications for the CCOmbo EDV TD catheters include: Acute heart failure, Severe hypovolemia, Complex circulatory situations, Medical emergencies, Adult respiratory distress syndrome, Gram negative sepsis, Multi system organ failure, Drug intoxication, Acute renal failure, Hemorrhagic pancreatitis, Intra and post-operative management of high risk patients, History of pulmonary or cardiac disease, Fluid shifts (e.g., extensive intra-abdominal operations), Management of high-risk obstetrical patients, Diagnosed cardiac disease, Toxemia, Premature separation of placenta, Cardiac output determinations, Volumetric determinations, Differential diagnosis of mitral regurgitation and ventricular septal rupture, Diagnosis of cardiac tamponade. Secondary indications include the following: Blood Sampling, Infusion of saline and dextrose solutions.

    Swan-Ganz oximetry TD catheters are indicated for the assessment of a patient's hemodynamic condition through direct intracardiac and pulmonary artery pressure monitoring, cardiac output determination, continuous mixed venous oxygen saturation monitoring, and for infusing solutions. The oximetry Paceport catheters (model 780F75M) are also indicated for standby temporary ventricular pacing using the model D98100 Chandler transluminal V-pacing probe. For all models, the distal (pulmonary artery) port also allows sampling of mixed venous blood for the assessment of oxygen transport balance and the calculation of derived parameters such as oxygen consumption, oxygen utilization coefficient, and intrapulmonary shunt fraction.

    The REF/Ox catheters (Models D754F75 and 759F75) are indicated for continuous mixed venous oxygen saturation in addition to the prior listed indications. For Models D754F75 and 759F75, the distal (pulmonary artery) port also allows sampling of mixed venous blood for the assessment of oxygen transport balance and the calculation of derived parameters such as oxygen consumption, oxygen utilization coefficient, and intrapulmonary shunt fraction.

    Device Description

    The Swan-Ganz catheters are well known pulmonary artery catheters intended for use on critical care patients. A Swan-Ganz Catheter includes an inflatable balloon at the tip, which facilitates its placement into the pulmonary artery through the flow of blood. The catheters can be grouped together based on functionality. The Swan-Ganz catheters can be used with compatible cardiac output patient monitors and/or with oximetry modules (depending on the model number) to transmit signals for hemodynamic pressure monitoring, cardiac output measurements, and/or oximetry measurements. The Swan-Ganz catheters are to be used with the Edwards and/or Edwards' compatible patient monitors, such as the Vigilance II.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for Edwards Lifesciences, LLC Swan Ganz Catheters. This document outlines the device's indications for use and includes a 510(k) summary (Section 5).

    However, it does not contain information regarding detailed acceptance criteria, a study proving the device meets these criteria, sample sizes for test or training sets, expert qualifications, or adjudication methods for AI/software-as-a-medical-device (SaMD) performance evaluation. The document primarily focuses on demonstrating substantial equivalence to predicate devices for its intended physiological monitoring functions and includes an update regarding MRI compatibility.

    Therefore, I cannot extract the specific information requested in your bullet points related to acceptance criteria and study details for an AI/SaMD device from this document. The "Key Performance Specifications" section explicitly states: "The proposed contraindication change and the MR Safe claim do not have any effect on the design, materials, technology and operating principles of current legally marketed device. Therefore no additional performance testing was required."

    The "Functional/Safety Testing" section mentions: "Swan-Ganz Flow Directed Monitoring catheter was evaluated for MRI compatibility and was confirmed to be MR Safe". This refers to the physical safety of the catheter in an MRI environment, not a performance evaluation in the context of diagnostic accuracy or clinical effectiveness that would typically involve acceptance criteria and comparative studies for a software device.

    Without the requested details, I am unable to generate the bulleted list of acceptance criteria and study information.

    Ask a Question

    Ask a specific question about this device

    K Number
    K094048

    Validate with FDA (Live)

    Date Cleared
    2011-07-14

    (560 days)

    Product Code
    Regulation Number
    870.1240
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Percu-Pro™ Messenger Balloon Catheter is intended to facilitate the placement of interventional devices during diagnostic and interventional cardiovascular procedures. The device is not intended for coronary or neurovascular use.

    Device Description

    The Cardiosolutions Percu-Pro™ Messenger Balloon Catheter is a 5Fr catheter with a bifurcated proximal hub. The catheter shaft and connection ends are made from Pebax and the distal tip balloon is made from latex.

    AI/ML Overview

    This document does not contain information about a study proving that a device meets acceptance criteria, an AI/ML device, or a standalone algorithm.

    The document is a 510(k) summary for the Percu-Pro™ Messenger Balloon Catheter, which is a physical medical device (a catheter). The summary details its design, intended use, and comparison to predicate devices to establish substantial equivalence, rather than providing performance metrics like accuracy, sensitivity, or specificity that would be typical for an AI system.

    Therefore, I cannot provide the requested information.

    Ask a Question

    Ask a specific question about this device

    K Number
    K081052

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2008-06-18

    (65 days)

    Product Code
    Regulation Number
    870.1240
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Occlusion Balloon Catheter, model Venos™, is intended for a temporary occlusion of coronary sinus during a venogram, or infusion of contrast media or drug, or for possible introduction of devices into the coronary venous system.

    Device Description

    Not Found

    AI/ML Overview

    This document is a 510(k) clearance letter from the FDA for a medical device (Occlusion Balloon Catheter, Model Venos™). It does not contain information about acceptance criteria and a study proving the device meets those criteria.

    Therefore, I cannot provide the requested information from this document. The letter only states that the device is substantially equivalent to a legally marketed predicate device, allowing it to be marketed. It does not detail the specific performance metrics or studies used to demonstrate that equivalence.

    Ask a Question

    Ask a specific question about this device

    K Number
    K010253

    Validate with FDA (Live)

    Date Cleared
    2002-03-19

    (417 days)

    Product Code
    Regulation Number
    870.1240
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    Page 1 of 2