Search Filters

Search Results

Found 20 results

510(k) Data Aggregation

    K Number
    K150819

    Validate with FDA (Live)

    Device Name
    Triglycerides
    Date Cleared
    2015-08-06

    (132 days)

    Product Code
    Regulation Number
    862.1705
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For the quantitative in vitro determination of Triglycerides in serum. Triglyceride measurements are used in the diagnosis and treatment of diseases involving lipid metabolism and various endocrine disorders e.g Diabetes mellitus, nephrosis and liver obstruction

    This in vitro diagnostic device is intended for prescription use only.

    Device Description

    The Randox Triglycerides kit assay consists of ready to use reagent solutions.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study information for the Triglycerides (TRIGS) device, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't present a formal table of "acceptance criteria" for the entire device's performance followed by a direct comparison in a single table, but rather describes various performance characteristics and their findings. I will construct a table based on the individual performance criteria and the results presented.

    Performance CharacteristicAcceptance Criteria (Implicit/Explicit)Reported Device Performance
    Precision/ReproducibilityNot explicitly stated as a single numerical criterion. Evaluated consistent with C.L.S.I documents EP5-A2. Implied goal is acceptable variability across runs, days, and total.Lot 1 (RX Daytona Plus): - Sensitivity Pool (13.3 mg/dl): Total CV 13.4% - Serum Pool 1 (96.4 mg/dl): Total CV 2.1% - QC 1 (96.5 mg/dl): Total CV 2.3% - Patient Pool 2 (104 mg/dl): Total CV 2.5% - Patient Pool 1 (117 mg/dl): Total CV 2.5% - Serum Pool 2 (237 mg/dl): Total CV 2.1% - CAL (240 mg/dl): Total CV 2.0% - QC 2 (259 mg/dl): Total CV 1.5% - Serum Pool 3 (326 mg/dl): Total CV 1.6% Lot 2 (RX Daytona Plus): - Sensitivity Pool (17.7 mg/dl): Total CV 11.6% - 801UN QC 2 (97.3 mg/dl): Total CV 3.2% - Serum Pool 1 (111 mg/dl): Total CV 3.6% - 832UE CAL (235 mg/dl): Total CV 3.0% - Serum Pool 2 (252 mg/dl): Total CV 2.6% - 587UE QC 3 (265 mg/dl): Total CV 2.5% - Serum Pool 3 (326 mg/dl): Total CV 3.7% - Serum Pool 4 (514 mg/dl): Total CV 2.8%
    Linearity/Assay Reportable RangeDeviation from linearity less than 5%.Linearity: Slope 0.96, Intercept 3.30, r = 1.000. Reportable Range: 12.4 – 1000 mg/dl. (The linearity study covered up to approximately 1000 mg/dl, and the device has an auto-dilution feature for samples >1000 mg/dL).
    Detection LimitNot explicitly stated as acceptance criteria, but rather defined properties.LoD: 3.96 mg/dl (based on 240 determinations, 4 low-level samples) LoB: 2.65 mg/dl LoQ: 12.4 mg/dl (lowest concentration at which precision is met)
    Analytical Specificity (Interference)Recovery within ±10% of the initial value of Triglycerides concentration of 150mg/dL and 496mg/dL.Hemoglobin: No significant interference up to 750mg/dL Total Bilirubin: No significant interference up to 60mg/dL Conjugate Bilirubin: No significant interference up to 60mg/dL Ascorbic Acid: No significant interference up to 3.0mg/dL
    Method Comparison with Predicate DeviceNot explicitly stated as a single numeric criterion for the regression, but the goal is "substantially equivalent" to the predicate. Implied acceptable correlation (r) and agreement (slope, intercept).Comparison: Y = 0.97x + 1.22 Correlation coefficient: r = 0.999 (This indicates a very strong positive correlation)

    2. Sample Size Used for the Test Set and Data Provenance

    • Precision/Reproducibility:
      • Test Set: Serum-based control material and unaltered human serum samples (some spiked or diluted). Specific numbers of individual patient samples beyond "Pool 1, Pool 2, Pool 3, Pool 4" are not given. For each sample type, 2 replicates per run were performed, twice daily for 20 non-consecutive days, using 2 reagent lots and 2 systems.
      • Data Provenance: Not explicitly stated, but implies laboratory testing with control materials and human serum samples. Given the manufacturer is based in the UK, it's likely the testing was done there, but this is not confirmed. It is a prospective study design for precision.
    • Linearity:
      • Test Set: 11 levels prepared from low and high serum pools, each run in replicates of five.
      • Data Provenance: Not explicitly stated, but implies laboratory testing with serum pools. Prospective study design.
    • Detection Limit:
      • Test Set: 240 determinations were made, including 4 low-level samples, to determine LoD, LoB, and LoQ.
      • Data Provenance: Not explicitly stated, implies laboratory testing. Prospective study design.
    • Analytical Specificity (Interference):
      • Test Set: Spiked interferent samples with corresponding control solutions. Specific number of samples not provided. Triglycerides concentrations of 150 mg/dL and 496 mg/dL were examined.
      • Data Provenance: Not explicitly stated, implies laboratory testing. Prospective study design.
    • Method Comparison:
      • Test Set: 109 serum patient samples spanning the range 14.2 to 986 mg/dl. Each tested in singlicate.
      • Data Provenance: Not explicitly stated (e.g., country of origin), but states "serum patient samples." Implies retrospective collection of samples or prospective collection for this study.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This device (Triglycerides assay) is an in-vitro diagnostic device for quantitative chemical analysis. The "ground truth" for these types of devices is established through analytical reference methods or highly characterized reference materials, not typically by expert consensus of human readers.

    • No mention of human experts or their qualifications for establishing ground truth for the test set.

    4. Adjudication Method for the Test Set

    Not applicable for this type of quantitative analytical assay. Adjudication is typically used in image-based diagnostic studies involving human interpretation.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. This type of study involves assessing the performance of human readers, sometimes with and without AI assistance, and is not relevant for a quantitative chemical assay.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Yes, the studies described (precision, linearity, detection limit, analytical specificity, method comparison) represent the standalone performance of the device/system (RX Daytona plus analyzer with Randox Triglycerides reagent) without human intervention in the analytical measurement process beyond sample loading and general operation.

    7. The Type of Ground Truth Used

    • Precision/Reproducibility, Linearity, Detection Limit, Analytical Specificity: The "ground truth" or reference for these studies refers to the true concentration of triglycerides in the samples (control materials, spiked samples, diluted samples) as determined by a highly accurate method or known values of reference materials.
    • Method Comparison: The "ground truth" is the results obtained by a legally marketed predicate device (Randox Triglyceride Assay, K923508). For calibrators within the system, Randox Calibration Serum Level 3 is stated to be traceable to the Triglycerides reference method ID-GC/MS. This indicates a high-accuracy chemical method is the ultimate ground truth for calibration.

    8. The Sample Size for the Training Set

    This document describes a medical device (an in-vitro diagnostic reagent/system) for which "training sets" are not typically applicable in the same way as for AI/machine learning algorithms. The device's performance characteristics are inherent to its chemical formulation and the analytical instrument. There is no mention of a "training set" for an algorithm.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no specific "training set" for an algorithm mentioned in the context of this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K133067

    Validate with FDA (Live)

    Date Cleared
    2013-11-22

    (56 days)

    Product Code
    Regulation Number
    862.1705
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ADVIA® Chemistry Triglycerides 2 assay (TRIG 2) is intended for in vitro diagnostic use in the quantitative measurement triglycerides in human serum and plasma on the ADVIA® Chemistry systems. Measurements of triglycerides are used in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver obstruction, other diseases involving lipid metabolism, or various endocrine disorders.

    Device Description

    The Triglyceride 2 reagent is a ready-to-use liquid reagent packaged for use on the automated ADVIA 1800 Chemistry system. It is supplied as a 358 tests/wedge. 4 wedges/Kit with a 38 mL fill in a 40 mL wedge size.

    Existing Siemens Set-Point Chemistry calibrator (classified under CFR 862.1150 - calibrator, multi-analyte mixture, product code JIX), cleared under 510k K030169, is used to calibrate the assay on the ADVIA Chemistry systems.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the ADVIA® Chemistry Triglycerides 2 Reagents (TRIG 2) device, based on the provided text:

    Acceptance Criteria and Device Performance

    The acceptance criteria for the ADVIA® Chemistry Triglycerides 2 assay appear to be established by comparison with a predicate device and adherence to CLSI guidelines for various performance characteristics. Specific quantifiable acceptance criteria are often implied by "acceptable results" or meeting certain thresholds, but the document primarily reports the performance observed.

    Performance CharacteristicAcceptance Criteria (Implied/Reference)Reported Device Performance (ADVIA TRIG 2)
    PrecisionCLSI document EP5-A2, "Evaluation of Precision Performance of Quantitative Measurement Methods: Approved Guideline." (Specific CV/SD thresholds are not explicitly stated but are implied by the predicate device's expected performance and established laboratory practice for triglyceride assays.)Serum Control (183 mg/dL): Total SD: 1.77 mg/dL, Total CV: 1.0% Serum Control (92 mg/dL): Total SD: 0.82 mg/dL, Total CV: 0.9% Serum Pool (254 mg/dL): Total SD: 2.04 mg/dL, Total CV: 0.8% Serum Pool (503 mg/dL): Total SD: 2.00 mg/dL, Total CV: 0.4%
    Linearity/Assay RangeDeviation from linearity of ≤ 5%. Range should be comparable to the predicate device or clinically appropriate.Linear/measuring range: 10 - 550 mg/dL with a deviation from linearity of ≤ 5%. R-squared value of 0.9998. (Predicate range: 15-1000 mg/dL)
    Limit of Blank (LoB)Determined according to CLSI guideline EP17-A2. Proportions of false positives (a) less than 5%.LoB: 5 mg/dL
    Limit of Detection (LoD)Determined according to CLSI guideline EP17-A2. Proportions of false negatives (β) less than 5%. Smallest amount reliably detected.LoD: 8 mg/dL
    Limit of Quantitation (LoQ)Determined according to CLSI guideline EP17-A2.LoQ: 10 mg/dL
    Method ComparisonDemonstrated substantial equivalence through correlation and agreement with the predicate device. (Specific acceptable regression parameters like slope and intercept range are typical for such studies, but not explicitly stated as "acceptance criteria" here, rather implied by "acceptable results" and demonstrating substantial equivalence).Serum Samples (n=101): ADVIA TRIG 2 = 0.94 (predicate) + 4.4 mg/dL. Slope 95% CI: 0.93 to 0.95. Intercept 95% CI: 2.88 to 5.85. Sample range: 20 - 540 mg/dL.Plasma (Lithium Heparin) (n=60): ADVIA TRIG 2 = 1.01 (predicate) - 2.6 mg/dL. Slope 95% CI: 0.99 to 1.3. Intercept 95% CI: -5.95 to 0.73. Sample range: 34 - 509 mg/dL.Plasma (Potassium EDTA) (n=60): ADVIA TRIG 2 = 1.02 (predicate) - 7.5 g/dL. Slope 95% CI: 1.00 to 1.03. Intercept 95% CI: -10.18 to -4.76. Sample range: 34 - 509 mg/dL.
    Analytical SpecificityInterference criteria: Bias exceeding 10% is considered significant interference. (CLSI EP7-A2 referenced)Bilirubin (conjugated): Interference at 22.5 mg/dL (-12.0%) with 148.0 mg/dL Trig concentration. NSI at other levels. Bilirubin (unconjugated): Interference at 15.0 mg/dL (+12.4%) with 148.0 mg/dL Trig concentration. NSI at other levels. Hemolysis: Interference at 750.0 mg/dL (+11.9%) with 138.0 mg/dL Trig concentration. NSI at other levels. Ascorbic Acid: Interference at 6.0 mg/dL (-13.1%) with 144.0 mg/dL Trig concentration. NSI at other levels.
    StabilityReagent stability for opened and unopened products should meet manufacturer's claims. (No explicit criteria given, but the purpose of the study is to establish these values).Opened reagent stable for 60 days on system. Shelf life of unopened product is 12 months at 2-8°C.
    TraceabilityTraceable to an accepted reference material.Traceable to SRM909c from NIST.

    Study Details

    1. Sample size used for the test set and the data provenance:

      • Precision: 80 replicates for each of the two serum controls and two serum pools. The data provenance is not explicitly stated (e.g., country of origin); however, the context of in vitro diagnostic testing in a 510(k) submission generally implies laboratory testing under controlled conditions, likely in the US by the manufacturer (Siemens Healthcare Diagnostics, Inc., Tarrytown, NY). The data is prospective as it was generated specifically for this submission.
      • Linearity: Not explicitly stated how many samples/replicates were used, but it involved mixing low and high serum pools to create 12 levels (9 intermediate, 3 low-end).
      • LoB, LoD: 320 determinations, comprising 160 blank and 160 low-level sample replicates.
      • Method Comparison (Serum): 101 serum samples.
      • Method Comparison (Plasma Matrices): 60 plasma samples (Lithium Heparin) and 60 plasma samples (Potassium EDTA).
      • Analytical Specificity: Not explicitly stated, but involved testing samples with varying concentrations of interferents and triglycerides.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

      • For this type of in vitro diagnostic device (chemical assay), the "ground truth" is typically established by the reference method (the predicate device) or by known concentrations of analytes in calibrators and control materials, rather than by human expert consensus or pathology. No human experts are mentioned for establishing ground truth in this context.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • Adjudication methods like "2+1" or "3+1" are characteristic of studies involving human interpretation (e.g., imaging studies). This information is not applicable to a chemical assay where results are quantifiable values based on instrument measurements.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No MRMC comparative effectiveness study was done. This type of study is not relevant for an automated chemical assay.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Yes, this is a standalone device in the sense that the assay itself generates the quantitative results without direct human intervention in the result determination. The performance characteristics described are "algorithm only" or "device only" performance. Humans operate the instrument and interpret results, but the measurement itself is automated.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The ground truth for performance evaluation (e.g., method comparison) is based on the predicate device (Dimension Clinical Chemistry Triglycerides FLEX reagent cartridge, K010871) and known concentrations for controls, calibrators, and linearity pools. For traceability, the assay is referenced to NIST SRM909c, a certified reference material.
    7. The sample size for the training set:

      • This is a traditional IVD assay, not a machine learning or AI-based device. Therefore, there is no "training set" in the context of data used to train an algorithm. The development of the reagent and its parameters would have involved R&D studies, but these are not analogous to training data for AI.
    8. How the ground truth for the training set was established:

      • Not applicable, as there is no training set for an AI/ML algorithm.
    Ask a Question

    Ask a specific question about this device

    K Number
    K130332

    Validate with FDA (Live)

    Date Cleared
    2013-05-31

    (109 days)

    Product Code
    Regulation Number
    862.1705
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For in vitro diagnostic use only. VITROS Chemistry Products TRIG Slides quantitatively measure triglyceride (TRIG) concentration in serum and plasma using VITROS® Systems. Triglyceride measurements are used in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver obstruction, other diseases involving lipid metabolism, or various endocrine disorders.

    Device Description

    Not Found

    AI/ML Overview

    The provided document describes a 510(k) summary for a modified medical device, the "VITROS Chemistry Products TRIG Slides," and references a study conducted to demonstrate substantial equivalence to a predicate device. This is a Special 510(k) submission, indicating that the modifications did not alter the fundamental scientific technology or intended use, but rather involved changes to the product that necessitated new verification and validation. Therefore, the study described focuses on demonstrating that the modified device continues to meet established performance criteria.

    Here's a breakdown of the requested information based on the provided text:


    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of acceptance criteria with corresponding performance metrics in a pass/fail format. Instead, it states that various performance characteristics were considered for potential hazards and that validation and verification testing were conducted and the modified device met the pre-determined acceptance criteria for all the performance testing.

    The characteristics considered, implying where acceptance criteria would have been applied, are:

    Performance CharacteristicReported Device Performance
    AccuracyMet pre-determined acceptance criteria. (Implies performance within acceptable limits for comparison to a reference method or the predicate device.)
    PrecisionMet pre-determined acceptance criteria. (Implies acceptable variability in results.)
    LinearityMet pre-determined acceptance criteria. (Implies accurate measurement across the specified assay range.)
    Potential InterferentsMet pre-determined acceptance criteria. (Implies results are not significantly affected by common interferents.)
    Long term and On-analyzer StabilityMet pre-determined acceptance criteria. (Implies acceptable performance over time and during use on the analyzer.)
    Limit of DetectionMet pre-determined acceptance criteria. (Implies ability to detect triglycerides at the lower end of the assay range.)
    Specimen TypeMet pre-determined acceptance criteria. (Implies acceptable performance with both serum and plasma.)

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample size used for the test set or the data provenance (country of origin of the data, retrospective or prospective). This information is typically detailed in the validation and verification reports, which are summarized in the 510(k) but not fully reproduced here.


    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided. For this type of in vitro diagnostic device (IVD), ground truth is established through reference methods or comparative studies with an established predicate device, rather than expert interpretation of images or other subjective data. Therefore, "experts" in the context of establishing ground truth for a clinical diagnostic test would typically refer to laboratory professionals performing the reference measurements, for which specific qualifications might not be explicitly stated in the 510(k) summary itself.


    4. Adjudication Method for the Test Set

    This information is not applicable and therefore not provided. Adjudication methods (like 2+1, 3+1) are typically used in studies involving subjective interpretation, such as image analysis, where multiple readers provide opinions that might need to be resolved. For an IVD device measuring a quantitative analyte like triglycerides, the "ground truth" is established by a reference method or predicate device, and differences are analyzed statistically, not through direct adjudication of expert opinions on the test results.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    An MRMC comparative effectiveness study was not done. This is an In Vitro Diagnostic (IVD) device for measuring triglyceride concentration, not an AI-assisted diagnostic imaging or interpretation device that would involve human readers and AI assistance.


    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This concept is not directly applicable to this device in the way it is typically understood for AI-based algorithms. The device itself performs the measurement (algorithmically via chemical reactions and spectrophotometry). The study implicitly evaluates the "standalone" performance of the modified VITROS Chemistry Products TRIG Slides by comparing its results to a predicate device and ensuring it meets established specifications (accuracy, precision, etc.). There is no "human-in-the-loop" component in the direct measurement process of this specific IVD device.


    7. The Type of Ground Truth Used

    For this IVD device, the "ground truth" would have been established by:

    • Reference Methods: Highly accurate and precise laboratory methods for triglyceride measurement.
    • Predicate Device Comparison: The performance of the modified device was compared to the legally marketed predicate device (VITROS Chemistry Products TRIG Slides cleared on August 3, 1981 (K812029)). This comparison ensures "substantial equivalence" to a device already deemed safe and effective.

    The testing considered accuracy, precision, linearity, potential interferents, stability, limit of detection, and specimen type, all of which would be assessed against established scientific and clinical standards, and in comparison to the predicate.


    8. The Sample Size for the Training Set

    This information is not provided. For a traditional IVD device based on chemical reactions and spectrophotometry, there isn't a "training set" in the machine learning sense. The device's operational parameters (like calibration curves, reaction kinetics) are established through rigorous experimental design and optimization, not by "training" an algorithm on a 'training set' of patient data in the way a modern AI model would be.


    9. How the Ground Truth for the Training Set Was Established

    As explained above, the concept of a "training set" and establishing ground truth for it, as typically applied to AI/ML devices, is not directly applicable here. The device's function is based on established biochemical principles and analytical chemistry, with its design and performance specifications developed through scientific experimentation and engineering.

    Ask a Question

    Ask a specific question about this device

    K Number
    K113830

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-08-30

    (247 days)

    Product Code
    Regulation Number
    862.1705
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Vantera® Clinical Analyzer is an automated laboratory test analyzer which measures the 400 MHz proton nuclear magnetic resonance (NMR) spectrum of clinical samples to produce signal amplitudes, converting these signal amplitudes to analyte concentration. The device includes a 400 MHz NMR spectrometer and software to analyze digitized spectral data. This instrumentation is intended to be used with NMR based assays to detect multiple analytes from clinical samples.

    The NMR LipoProfile® test, when used with the Vantera® Clinical Analyzer, an automated NMR spectrometer, measures lipoprotein particles to quantify LDL particle number (LDL-P), HDL cholesterol (HDL-C), and triglycerides in human serum and plasma using nuclear magnetic resonance (NMR) spectroscopy. LDL-P and these NMR-derived concentrations of HDL-C and triglycerides are used in conjunction with other lipid measurements and clinical evaluation to aid in the management of lipoprotein disorders associated with cardiovascular disease.

    Device Description

    The Vantera Clinical Analyzer is a clinical laboratory analyzer that employs nuclear magnetic resonance spectroscopic detection to quantify multiple analytes in biological fluid specimens, specifically blood plasma and serum. The Vantera Clinical Analyzer system design is divided into 3 major subassemblies: a sample handling assembly, an NMR subassembly, and an enclosure. The Vantera Clinical Analyzer control system is distributed across three separate computers: The Host (1U) controls user interface, data handling, results calculation, system startup and shutdown. The Process Control (4U) schedules and manages all activities required to process a sample, controls all hardware in the sample handling subsystem, and manages remote access to the system. The NMR Control Computer controls all magnet operations. Two of these computers are contained within the Sample Handling Subassembly (1U and 4U) and one in the NMR Subassembly (NMR Console).

    The NMR LipoProfile test involves measurement of the 400 MHz proton NMR spectrum of a plasma/serum sample, deconvolution of the composite signal at approximately 0.8 ppm to produce signal amplitudes of the lipoprotein subclasses that contribute to the composite plasma/serum signal, and conversion of these subclass signal amplitudes to lipoprotein subclass concentrations.

    AI/ML Overview

    The provided 510(k) summary focuses on the analytical performance of the Vantera® Clinical Analyzer and the NMR LipoProfile® test compared to predicate devices, establishing substantial equivalence rather than providing explicit acceptance criteria as would be typical for a novel device. The study described primarily demonstrates that the proposed device performs comparably to its predicate devices in terms of analytical accuracy and precision.

    Here's an analysis of the acceptance criteria and study information provided:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state acceptance criteria in a pass/fail format. Instead, it demonstrates performance by comparing the analytical results of the Vantera® Clinical Analyzer with the NMR LipoProfile® test to its predicate device (NMR Profiler for the assay) across various metrics. The unstated acceptance criteria for each analytical performance metric would be that the proposed device's performance must be comparable to or better than the predicate device's performance.

    MetricAcceptance Criteria (Implied)Reported Proposed Device Performance (Vantera® Clinical Analyzer)Predicate Device Performance (NMR Profiler)
    LDL-P
    LoBComparable to predicate0 nmol/L0 nmol/L
    LoDComparable to predicate40.7 nmol/L41 nmol/L
    LoQComparable to predicate132 nmol/L157 nmol/L
    Measuring RangeComparable to predicate300-3500 nmol/L300-3500 nmol/L
    Linearity Regression (Y=mX+b)R² comparable to predicatey=1.02x+7.82, R²=0.9949y=0.99x-22.37, R²=0.9979
    Within-Run Precision (CV%)Comparable to predicateLevel 1: 5.8%, Level 2: 3.0%, Level 3: 2.7%Level 1: 5.0%, Level 2: 4.3%, Level 3: 3.7%
    Within-Lab Precision (CV%)Comparable to predicateLevel 1: 5.3%, Level 2: 4.0%, Level 3: 3.9%Level 1: 7.6%, Level 2: 4.5%, Level 3: 4.3%
    Method Comparison (Correlation R)Comparable to predicateR=0.978R=0.973
    Interference StudyNo significant interference for tested substancesSalicylic acid (≥ 1.3mmol/L) and Clopidogrel hydrogensulfate (≥ 95.7 µmol/L) determined to interfere.No interference found for 5 endogenous & 22 exogenous substances.
    Specimen Stability (Refrigerated)Comparable to predicate6 days5 days
    Triglycerides
    LoBComparable to predicate1.1 mg/dL1.4 mg/dL
    LoDComparable to predicate2.4 mg/dL2.6 mg/dL
    LoQComparable to predicate4 mg/dL2.6 mg/dL
    Measuring RangeComparable to predicate5-1100 mg/dL5-1100 mg/dL
    Linearity Regression (Y=mX+b)R² comparable to predicatey=1.008x-0.3979, R²=0.9999y=0.95x-12.21, R²=0.999
    Within-Run Precision (CV%)Comparable to predicateLevel 1: 2.3%, Level 2: 2.1%, Level 3: 1.2%Level 1: 2.6%, Level 2: 1.8%, Level 3: 1.3%
    Within-Lab Precision (CV%)Comparable to predicateLevel 1: 2.3%, Level 2: 2.4%, Level 3: 2.7%Level 1: 3.6%, Level 2: 2.6%, Level 3: 2.5%
    Method Comparison (Correlation R)Comparable to predicateR=0.998R=1.00
    Interference StudyNo significant interference for tested substancesNo interference found for 7 endogenous & 23 exogenous substances.Ibuprofen may interfere with TG measurement at and above 210µg/mL for 5 endogenous & 22 exogenous substances.
    Specimen Stability (Refrigerated)Comparable to predicate6 days10 days
    HDL-C
    LoBComparable to predicate2.7 mg/dL4.3 mg/dL
    LoDComparable to predicate3.5 mg/dL5.2 mg/dL
    LoQComparable to predicate4 mg/dL5.2 mg/dL
    Measuring RangeComparable to predicate7-140 mg/dL7-140 mg/dL
    Linearity Regression (Y=mX+b)R² comparable to predicatey=1.049x-0.3459, R²=0.9961y=1.004x-0.5956, R²=0.9998
    Within-Run Precision (CV%)Comparable to predicateLevel 1: 4.0%, Level 2: 2.8%, Level 3: 2.6%Level 1: 2.0%, Level 2: 1.9%, Level 3: 0.9%
    Within-Lab Precision (CV%)Comparable to predicateLevel 1: 2.8%, Level 2: 2.0%, Level 3: 1.8%Level 1: 3.3%, Level 2: 2.0%, Level 3: 1.8%
    Method Comparison (Correlation R)Comparable to predicateR=0.989R=0.999
    Interference StudyNo significant interference for tested substancesNo interference found for 7 endogenous & 23 exogenous substances.No interference found for 5 endogenous & 22 exogenous substances.
    Specimen Stability (Refrigerated)Comparable to predicate6 days10 days

    Study Proving Acceptance Criteria:

    The study conducted was an analytical validation comparing the performance of the Vantera® Clinical Analyzer with the NMR LipoProfile® test to its predicate device (NMR Profiler for the assay) across various analytical parameters. The overall conclusion is that the new device is "as safe and effective as its predicate device."

    2. Sample size used for the test set and the data provenance

    • Test Sets (Analytical Studies):
      • Analytical Sensitivity (LoB, LoD, LoQ): Five serum pools for low concentrations tested in replicates of 4 for 3 days. Non-lipoprotein specimens analyzed 60 consecutive times for 3 days for LoB.
      • Assay Precision (Within-run & Within-Laboratory): 20 replicates of three patient serum pools in the same run and in 20 different runs over 20 days. Reproducibility study used 5 levels of serum panels tested for 5 days, 6 runs per day, 2 replicates per run at 3 sites (n=60 per panel per site, total N=177-180 across all sites for each panel).
      • Linearity: Three serum pools prepared from patient specimens mixed and diluted to produce eleven (for LDL-P) or twelve (for TG and HDL-C) different samples, with four replicates of each pool analyzed.
      • Method Comparison:
        • LDL-P: n=1483 serum samples.
        • HDL-C: n=1518 serum samples.
        • Triglycerides: n=1520 serum samples.
      • Interfering Substances: 7 endogenous agents and 23 drugs were screened.
      • Reference Range: Serum samples (n=452) from apparently healthy men (n=158) and women (n=294).
    • Data Provenance: The document does not specify the country of origin. The studies were described as "analytical validations" and included testing using "patient specimens" and "serum pools." There is no explicit mention of the data being either retrospective or prospective, but the nature of the analytical studies suggests controlled laboratory environments rather than a large-scale clinical trial with patient follow-up. For the reference range, it states "serum samples... were analyzed from apparently healthy men and women," which implies a prospective collection for this specific purpose or a well-characterized existing cohort. The MESA (Multi-Ethnic Study of Atherosclerosis) is mentioned for the predicate device's reference range, suggesting a US context for that, but it's not explicitly stated for the proposed device's reference population.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document does not describe the use of human experts to establish "ground truth" for the test set in the context of interpretation or diagnosis. This device is an automated laboratory analyzer for quantifying analytes. The "ground truth" for its analytical performance studies (e.g., precision, linearity, method comparison) is established by comparing its measurements to a reference method or known concentrations, or by assessing consistency internally.

    4. Adjudication method for the test set

    Not applicable. There was no clinical study involving human interpretation or diagnosis that would require an adjudication method. The testing involved direct analytical measurements.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is an in-vitro diagnostic (IVD) device for quantifying analytes (LDL-P, HDL-C, Triglycerides), not an AI-assisted diagnostic imaging or interpretation device that would involve human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, the primary performance studies presented are "standalone" in the sense that they assess the device's analytical performance (algorithm + instrument) in quantifying the specified analytes without human-in-the-loop performance for diagnosis or interpretation. The device's output is explicit numerical analyte concentrations.

    7. The type of ground truth used

    The ground truth used for verifying the analytical performance of the device was:

    • Reference Methods/Known Concentrations: For analytical sensitivity (LoB, LoD, LoQ), linearity, and precision, the "ground truth" was established through precisely prepared serum pools and non-lipoprotein specimens with known or target concentrations, or through statistical determination methods (e.g., EP17-A).
    • Comparison to Predicate Device: For method comparison studies, the "ground truth" or reference was the measurements obtained from the legally marketed predicate device (NMR Profiler) using patient samples. The goal was to show high correlation and similar results between the two devices.
    • CLSI Guidelines: Standardized guidelines (e.g., EP5-A2, EP6-A, EP7-A2, EP9-A2, EP17-A) from the Clinical and Laboratory Standards Institute (CLSI) were referenced for establishing protocols for these analytical validations.

    8. The sample size for the training set

    The document does not explicitly mention a "training set" in the context of machine learning or AI models. This device is an automated NMR spectrometer that measures signals and converts them to concentrations based on specified deconvolution analysis models. The development of these deconvolution models would have involved a form of "training" or optimization, but the document does not detail the dataset size or methodology used for this prior model development. The document focuses on the analytical validation of the manufactured device.

    9. How the ground truth for the training set was established

    As there is no explicit mention of a "training set" in the conventional AI/ML sense, this question cannot be directly answered from the provided text. However, the assay description mentions:
    "The NMR signals from the various lipoprotein subclasses have unique and distinctive frequencies and lineshapes, each of which is accounted for in the deconvolution analysis model. Each subclass signal amplitude is proportional to the number of subclass particles emitting the signal, which enables subclass particle concentrations to be calculated from the subclass signal amplitudes derived from the spectral deconvolution analysis."
    This suggests that the deconvolution analysis model was developed using a "ground truth" based on the established biophysical properties of lipoprotein subclasses and their NMR spectral characteristics. This likely involved:

    • Carefully characterized lipoprotein samples with known subclass concentrations.
    • Expert knowledge of NMR spectroscopy and signal processing.
    • Calibration against established reference methods for lipoprotein analysis.

    The document does not provide details on the specific data sets or expert consensus used for the initial development and establishment of this deconvolution model.

    Ask a Question

    Ask a specific question about this device

    K Number
    K102993

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-05-19

    (223 days)

    Product Code
    Regulation Number
    862.1705
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ELITech Clinical Systems TRIGLYCERIDES SL is intended for the quantitative in vitro diagnostic determination of triglycerides in human serum and plasma on ELITech Vital Scientific Selectra/Flexor analyzers. It is not intended for use in Point of Care settings. Triglycerides measurements are used in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver obstruction, other diseases involving lipid metabolism, or various endocrine disorders.

    ELITech Clinical Systems CHOLESTEROL SL is intended for the quantitative in vitro diagnostic determination of cholesterol in human serum and plasma on ELITech Vital Scientific Selectra/Flexor analyzers. It is not intended for use in Point of Care settings. Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders.

    ELITech Clinical Systems ELICAL 2 is a multi-parametric calibrator for in vitro diagnostic use in the calibration of quantitative ELITech Clinical Systems methods on the ELITech Vital Scientific Selectra/Flexor Analyzers.

    ELITech Clinical Systems ELITROL I & ELITROL II are multiparametric control sera for use in quality control of ELITech Clinical Systems methods on ELITech Vital Scientific Selectra/Flexor Analyzers.

    Device Description

    The device for this submission is available as kit only. It consists of 1 reagent, "R". Reagent R contains: Pipes buffer, p-chlorophenol, ATP, Amino-4-antipyrine (4-AAP), Lipoprotein lipase (bacterial), Glycerol kinase (bacterial), Glycerol-3-phosphate oxidase (microorganisms), Peroxidase (horseradish), Potassium ferrocyanide, Magnesium (Mg2+) and Sodium azide.

    The device for this submission is available as kit only. It consists of 1 reagent, "R". Reagent R contains: Pipes buffer, 4-Aminoantipyrine (4-AAP), Cholesterol esterase (CHE bacterial), Cholesterol oxidase (CHO microorganisms), Peroxidase (POD horseradish), Sodium cholate, Phenol and Sodium azide.

    ELITech Clinical Systems ELICAL 2 is a lyophilized calibrator based on human serum containing constituents to ensure optimal calibration. ELICAL 2 is prepared exclusively from the blood of donors tested individually and found to be negative for HbsAg and to antibodies to HCV and HIV according to FDA-approved methods or methods in compliance with the European Directive 98/79/EC, Annex II, List A.

    ELITech Clinical Systems ELITROL I and ELITROL II are two level quality control products consisting of lyophilized human serum containing constituents at desired levels. Elitrol I and Elitrol II are prepared exclusively from the blood of donors tested individually and found to be negative for HbsAg and to antibodies to HCV and HIV according to FDA-approved methods or methods in compliance with the European Directive 98/79/EC, Annex II, List A.

    AI/ML Overview

    Here's a summary of the acceptance criteria and the studies for the ELITech Clinical Systems TRIGLYCERIDES SL reagent, ELITech Clinical Systems CHOLESTEROL SL reagent, ELICAL 2 calibrator, and ELITROL I/ELITROL II controls, based on the provided documents.

    ELITech Clinical Systems TRIGLYCERIDES SL Reagent

    1. Table of Acceptance Criteria and Reported Device Performance

    For TRIGLYCERIDES SL, the acceptance criteria are not explicitly stated as distinct numerical targets; instead, the performance is demonstrated through comparison with a predicate device. The implied acceptance is that the device performs comparably to the predicate or within clinically acceptable ranges for the specified parameters.

    Performance CharacteristicAcceptance Criteria (Implied by Predicate)Reported Device Performance (ELITech Clinical Systems TRIGLYCERIDES SL)
    Measuring Range (Linearity)Comparable to predicate (3.1 to 1470 mg/dL)30 to 1000 mg/dL
    Precision (Within-run CV)Comparable to predicate (e.g., CV=0.82% to 2.83%)Level 48 mg/dL: CV=1.5%Level 142 mg/dL: CV=1.0%Level 273 mg/dL: CV=0.7%
    Precision (Total CV)Comparable to predicate (e.g., CV=1.37% to 2.96%)Level 48 mg/dL: CV=3.9%Level 142 mg/dL: CV=2.7%Level 273 mg/dL: CV=4.5%
    Method Comparison (Correlation with Predicate)High correlation (r²=0.9994) with predicatey=1.040x + 0.339 mg/dLr²= 0.998Range: 22 to 936 mg/dL
    Interference (Unconjugated Bilirubin)No significant influence up to 22.5 mg/dL (Total)No significant interference up to 15 mg/dL
    Interference (Conjugated Bilirubin)No significant influence up to 22.5 mg/dL (Direct)No significant interference up to 5.9 mg/dL
    Interference (Hemoglobin)No significant influence up to 500 mg/dLNo significant interference up to 250 mg/dL
    Interference (Uric Acid)Not explicitly stated for predicate in this tableNo significant interference up to 23.6 mg/dL
    Interference (Ascorbic Acid)Not explicitly stated for predicate in this tableNo significant interference up to 2.0 mg/dL (Concentrations above therapeutic levels interfere)
    Interference (Methyl-dopa)Not explicitly stated for predicate in this tableNo significant interference up to 1.0 mg/dL
    Calibration Frequency14 days14 days
    On-board Stability48 days (refrigerated)28 days (refrigerated)

    2. Sample Size and Data Provenance

    • Test Set Sample Size: The document does not explicitly state the sample size used for the test set for precision and method comparison studies.
    • Data Provenance: Not specified in the provided text. It is a submission by SEPPIM S.A.S. from FRANCE, suggesting data could be of French origin or from other regions. Studies are generally retrospective as they are performance evaluations of an existing reagent formulation.

    3. Number of Experts and Qualifications

    This information is not applicable. The studies are laboratory-based analytical performance evaluations of a diagnostic reagent, not interpretations by medical experts.

    4. Adjudication Method

    This information is not applicable for this type of analytical performance study.

    5. MRMC Comparative Effectiveness Study

    • This information is not applicable. This is an analytical performance study of a diagnostic reagent, not a study evaluating human reader performance with or without AI assistance.

    6. Standalone Performance Study

    • Yes, the performance data presented (measuring range, precision, method comparison, limitations) are for the standalone performance of the ELITech Clinical Systems TRIGLYCERIDES SL reagent on the ELITech Vital Scientific Selectra/Flexor analyzers.

    7. Type of Ground Truth Used

    • For precision and linearity, the "ground truth" refers to laboratory controls and reference materials with known concentrations, or samples run repeatedly to establish variability.
    • For method comparison, the "ground truth" is established by comparing the device's results to those obtained using the legally marketed predicate device (ABX PENTRA TRIGLYCERIDES CP). The predicate device's results serve as the reference for comparison.

    8. Sample Size for Training Set

    • This information is not applicable. This is a chemical reagent, not an AI/ML algorithm that requires a training set.

    9. How Ground Truth for Training Set was Established

    • This information is not applicable.

    ELITech Clinical Systems CHOLESTEROL SL Reagent

    1. Table of Acceptance Criteria and Reported Device Performance

    For CHOLESTEROL SL, similar to TRIGLYCERIDES SL, acceptance is implied by comparison to the predicate device and being within clinically acceptable ranges.

    Performance CharacteristicAcceptance Criteria (Implied by Predicate)Reported Device Performance (ELITech Clinical Systems CHOLESTEROL SL)
    Measuring Range (Linearity)Comparable to predicate (2.55 to 580 mg/dL)20 to 600 mg/dL
    Precision (Within-run CV)Comparable to predicate (e.g., CV=0.53% to 1.21%)Level 116 mg/dL: CV=2.4%Level 190 mg/dL: CV=1.9%Level 298 mg/dL: CV=1.7%
    Precision (Total CV)Comparable to predicate (e.g., CV=2.34% to 3.01%)Level 116 mg/dL: CV=2.6%Level 190 mg/dL: CV=2.7%Level 298 mg/dL: CV=2.7%
    Method Comparison (Correlation with Predicate)High correlation (r²=0.9943) with predicatey=1.006 x - 1.734 mg/dLr²= 0.999Range: 20 to 579 mg/dL
    Interference (Unconjugated Bilirubin)Not explicitly stated for predicate in this tableNo significant interference up to 6.0 mg/dL
    Interference (Conjugated Bilirubin)No significant influence up to 6.8 mg/dL (Direct)No significant interference up to 5.9 mg/dL
    Interference (Hemoglobin)No significant influence up to 336 mg/dLNo significant interference up to 250 mg/dL
    Interference (Turbidity)No significant influence up to 612.5 mg/dL triglyceridesNo significant interference up to 614 mg/dL triglyceride equivalent
    Interference (Ascorbic Acid)Not explicitly stated for predicate in this tableNo significant interference up to 2.0 mg/dL (Concentrations above therapeutic levels interfere)
    Interference (Methyl-dopa)Not explicitly stated for predicate in this tableNo significant interference up to 0.8 mg/dL (Concentrations above therapeutic levels interfere)
    Interference (Uric Acid)Not explicitly stated for predicate in this tableNo significant interference up to 23.4 mg/dL
    Calibration Frequency8 days28 days
    On-board Stability48 days (refrigerated)28 days (refrigerated)
    CRMLN CertificationCertifiedCertified

    2. Sample Size and Data Provenance

    • Test Set Sample Size: The document does not explicitly state the sample size used for the test set for precision and method comparison studies.
    • Data Provenance: Not specified in the provided text. It is a submission by SEPPIM S.A.S. from FRANCE, suggesting data could be of French origin or from other regions. Studies are generally retrospective as they are performance evaluations of an existing reagent formulation.

    3. Number of Experts and Qualifications

    This information is not applicable. The studies are laboratory-based analytical performance evaluations of a diagnostic reagent, not interpretations by medical experts.

    4. Adjudication Method

    This information is not applicable for this type of analytical performance study.

    5. MRMC Comparative Effectiveness Study

    • This information is not applicable. This is an analytical performance study of a diagnostic reagent, not a study evaluating human reader performance with or without AI assistance.

    6. Standalone Performance Study

    • Yes, the performance data presented (measuring range, precision, method comparison, limitations) are for the standalone performance of the ELITech Clinical Systems CHOLESTEROL SL reagent on the ELITech Vital Scientific Selectra/Flexor analyzers.

    7. Type of Ground Truth Used

    • For precision and linearity, the "ground truth" refers to laboratory controls and reference materials with known concentrations, or samples run repeatedly to establish variability.
    • For method comparison, the "ground truth" is established by comparing the device's results to those obtained using the legally marketed predicate device (ABX PENTRA CHOLESTEROL CP). The predicate device's results serve as the reference for comparison.

    8. Sample Size for Training Set

    • This information is not applicable. This is a chemical reagent, not an AI/ML algorithm that requires a training set.

    9. How Ground Truth for Training Set was Established

    • This information is not applicable.

    ELITech Clinical Systems ELICAL 2 Calibrator

    1. Table of Acceptance Criteria and Reported Device Performance

    For ELICAL 2, the primary acceptance criteria revolve around format, stability, and handling being comparable to the predicate, and traceability being established.

    Performance CharacteristicAcceptance Criteria (Implied by Predicate)Reported Device Performance (ELITech Clinical Systems ELICAL 2)
    FormatLyophilized human serum basedLyophilized human serum based
    LevelSingle levelSingle level
    Handling (Reconstitution)Exactly 3 mL water, gentle swirling, dissolve within 30 minExactly 3 mL water, gentle swirling, dissolve within 30 min
    Lyophilized StabilityStable at 2-8°C until expiration dateStable at 2-8°C until expiry date
    Reconstituted Stability (15-25°C)8 hours8 hours
    Reconstituted Stability (2-8°C)2 days2 days
    Reconstituted Stability (-25 to -15°C, frozen once)4 weeks4 weeks
    TraceabilityTraceability given in value sheets/instructionsTraceability given in value sheet

    2. Sample Size and Data Provenance

    • Test Set Sample Size: Not specified for stability studies.
    • Data Provenance: Not specified. From SEPPIM S.A.S. (FRANCE). Studies are for characterization of the calibrator's properties.

    3. Number of Experts and Qualifications

    Not applicable.

    4. Adjudication Method

    Not applicable.

    5. MRMC Comparative Effectiveness Study

    • Not applicable.

    6. Standalone Performance Study

    • Yes, the stability and handling characteristics are for the standalone performance of the ELICAL 2 calibrator.

    7. Type of Ground Truth Used

    • For stability, the "ground truth" involves testing the calibrator's analyte concentrations over time under specified storage conditions and comparing them against initial established values or specified ranges. For traceability, it refers to documentation linking the analyte values to international reference methods or materials.

    8. Sample Size for Training Set

    • Not applicable.

    9. How Ground Truth for Training Set was Established

    • Not applicable.

    ELITech Clinical Systems ELITROL I and ELITROL II Controls

    1. Table of Acceptance Criteria and Reported Device Performance

    For ELITROL I and ELITROL II, the acceptance criteria focus on comparable format, levels, handling, and stability to the predicate.

    Performance CharacteristicAcceptance Criteria (Implied by Predicate)Reported Device Performance (ELITech Clinical Systems ELITROL I / ELITROL II)
    FormatLyophilized human sera basedLyophilized human sera based
    LevelsTwo levelsTwo levels
    Handling (Reconstitution)Exactly 5 mL water, gentle swirling, dissolve within 30 minExactly 5 mL water, gentle swirling, dissolve within 30 min
    Lyophilized StabilityStable at 2-8°C until expiration dateStable at 2-8°C until expiry date
    Reconstituted Stability (15-25°C)12 hours12 hours
    Reconstituted Stability (2-8°C)5 days5 days
    Reconstituted Stability (-25 to -15°C, frozen once)4 weeks4 weeks

    2. Sample Size and Data Provenance

    • Test Set Sample Size: Not specified for stability studies.
    • Data Provenance: Not specified. From SEPPIM S.A.S. (FRANCE). Studies are for characterization of the control materials' properties.

    3. Number of Experts and Qualifications

    Not applicable.

    4. Adjudication Method

    Not applicable.

    5. MRMC Comparative Effectiveness Study

    • Not applicable.

    6. Standalone Performance Study

    • Yes, the stability and handling characteristics are for the standalone performance of the ELITROL I and ELITROL II controls.

    7. Type of Ground Truth Used

    • For stability, the "ground truth" involves testing the control's analyte concentrations over time under specified storage conditions and comparing them against initial established values or specified ranges. The control materials are "assayed," meaning their analyte concentrations are determined using reference methods or established analytical procedures.

    8. Sample Size for Training Set

    • Not applicable.

    9. How Ground Truth for Training Set was Established

    • Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K063804

    Validate with FDA (Live)

    Date Cleared
    2007-03-23

    (91 days)

    Product Code
    Regulation Number
    862.1705
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    System reagent for the quantitative determination of Triglyceride concentrations in human serum and plasma on OLYMPUS analyzers. Measurements of triglyceride are used in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver obstruction, other diseases involving lipid metabolism, or various enocrine disorders, and in the assessment of risk factors for atherosclerosis and coronary artery disease.

    Device Description

    This Olympus Triglyceride procedure is based on a series of coupled enzymatic reactions. The triglycerides in the sample are hydrolyzed by a combination of microbial lipases to give glycerol and fatty acids. The glycerol is phosphorylated by adenosine triphosphate (ATP) in the presence of glycerol kinase (GK) to produce glycerol-3-phosphate. The glycerol-3-phosphate is oxidized by molecular oxygen in the presence of GPO (glycerol phosphate oxidase) to produce hydrogen peroxide (H2O2) and dihydroxyacetone phosphate. The formed H2O2 reacts with 4-aminophenazone and N,N-bis(4-sulfobutyl)-3,5-dimethylaniline, disodium salt (MADB) in the presence of peroxidase (POD) to produce a chromophore, which is read at 660/800nm. The increase in absorbance at 660/800 nm is proportional to the triglyceride content of the sample.

    AI/ML Overview

    Here's a summary of the acceptance criteria and study information for the Olympus Triglyceride Test System, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally implied by the comparison to a predicate device, showing "similarities" and specific performance characteristics. The specific thresholds for acceptable performance are not explicitly stated as strict "acceptance criteria" but rather as "Performance Characteristics" which are compared to the predicate device.

    Performance CharacteristicAcceptance Criteria (Implied by Predicate)New Olympus Triglyceride Reported PerformancePredicate Reported Performance
    Precision AU400/400eComparable to predicateSample 1: 2.58% CVSample 1: 1.21% CV
    Sample 2: 2.54% CVSample 2: 1.67% CV
    Sample 3: 2.41% CVSample 3: 1.37% CV
    Precision AU600/640/640eComparable to predicateAU600:AU600:
    Sample 1: 1.65% CVSample 1: 1.83% CV
    Sample 2: 1.41% CVSample 2: 1.58% CV
    Sample 3: 1.46% CVSample 3: 2.80% CV
    Sample 4: 1.13% CV
    AU640/640e:AU640/640e:
    Sample 1: 1.00% CV
    Sample 2: 1.00% CV
    Precision AU2700/5400Comparable to predicateSample 1: 2.00% CVSample 1: 2.50% CV
    Sample 2: 1.72% CVSample 2: 2.00% CV
    Sample 3: 1.78% CVSample 3: 1.50% CV
    Sample 4: 1.20% CV
    Assay Range10 - 1000 mg/dL10 - 1000 mg/dL10 - 1000 mg/dL
    Method ComparisonSlope ≈ 1, Intercept ≈ 0, R ≈ 1Intercept: -0.871Intercept: 3.2
    Slope: 1.011Slope: 1.010
    R: 1.000R: 0.999
    Interfering SubstancesPerformance generally comparable to predicate, with limits on interferenceAU400/400e/600/640/640e/2700/5400:AU400/400e:
    Ascorbate≤ 2-10%≤ 5% up to 20 mg/dL≤ 2% up to 20mg/dL
    Bilirubin≤ 10%≤ 3% up to 40 mg/dL≤ 10% up to 20 mg/dL
    Hemolysis≤ 7-8%≤ 3% up to 500 mg/dL≤ 8% up to 500 mg/dL
    AU600/640/640e:
    Ascorbate≤ 1% up to 20mg/dL
    Bilirubin≤ 10% up to 32 mg/dL
    Hemolysis≤ 7% up to 500 mg/dL
    AU2700/5400:
    Ascorbate≤ 2% up to 20mg/dL
    Bilirubin≤ 10% up to 16 mg/dL
    Hemolysis≤ 8% up to 500 mg/dL

    Study Details:

    • 2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

      • The document mentions "samples" for precision and method comparison studies but does not specify the exact number of samples (sample size) used for the test sets. For precision, multiple "samples" (likely control or pooled patient samples) were tested across different Olympus analyzer models. For method comparison, it implies a set of samples were run on both the new and predicate devices, yielding the Intercept, Slope, and R values.
      • Data Provenance: Not specified. The document does not indicate the country of origin of the data or whether the studies were retrospective or prospective.
    • 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

      • This is a diagnostic reagent for quantitative measurement of triglyceride concentrations, not an imaging device requiring expert interpretation. Therefore, the concept of "experts establishing ground truth" in the interpretive sense does not directly apply. The "ground truth" for such devices is established through reference methods or highly characterized control materials. The document states traceability to College of American Pathology (CAP) Serum Lipid (RM016) #2, which serves as the reference for accuracy.
    • 4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

      • Not applicable as this is a quantitative chemical assay, not an interpretive device requiring adjudication.
    • 5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

      • Not applicable. This is a chemical diagnostic reagent device, not an AI or imaging device involving human readers.
    • 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

      • Yes, the performance data presented (precision, assay range, method comparison, interfering substances) represents the standalone performance of the Olympus Triglyceride Reagent when run on Olympus analyzers. There is no human-in-the-loop component for the direct measurement of triglycerides by this device.
    • 7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

      • The ground truth for accuracy and calibration is based on traceability to College of American Pathology (CAP) Serum Lipid (RM016) #2. This implies that the device's measurements are referenced against established standards or values obtained from a highly reliable and recognized reference material.
    • 8. The sample size for the training set

      • Not applicable. This is a chemical reagent, not a machine learning or AI device that typically requires a "training set." The development would involve chemical optimization and formulation, followed by validation studies.
    • 9. How the ground truth for the training set was established

      • Not applicable, as there is no "training set" in the context of an AI/ML algorithm for this type of device. The development and validation relied on established chemical principles and reference materials (like CAP standards).
    Ask a Question

    Ask a specific question about this device

    K Number
    K043546

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2005-01-18

    (26 days)

    Regulation Number
    862.1705
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Dimension® RxL Max™ with StreamLAB® Analytical Workeell and Sample Transfer Module is a discrete photometric chemistry analyzer for clinical use intended to duplicate analytical procedures by performing automatically various steps such as pipetting, preparing filtrates, heating, and measuring color intensity. This device is intended for use in conjunction with certain materials to measure a variety of analytes.

    Device Description

    The modified device includes the connection of a StreamLAB® Analytical Workcell and Sample Transfer Module. These are used to prepare specimens from the human body for testing on the Dimension® RxL Max™ system.

    AI/ML Overview

    This submission (K043546) is a Special 510(k) for a device modification; therefore, the provided document does not contain an acceptance criteria table or a detailed study report with performance metrics. The submission focuses on demonstrating substantial equivalence to a predicate device after a modification, rather than proving performance against specific acceptance criteria for a novel device.

    The modification involves connecting a StreamLAB® Analytical Workcell and Sample Transfer Module to the existing Dimension® RxL Max™ Clinical Chemistry System. The core functionality and assay performance are stated to be "the same" as the predicate device.

    However, based on the provided text, here's what can be extracted or inferred about the device and the nature of this submission:

    1. A table of acceptance criteria and the reported device performance

    This information is not provided in the given text. As a Special 510(k) for a device modification, the manufacturer primarily asserts that the modified device has "the same operating principles, design, manufacturing materials, method of manufacture, assay performance characteristics and intended use as the predicate device." Therefore, explicit acceptance criteria and corresponding reported device performance for the modified components are not detailed, as the focus is on maintaining equivalence to the existing, cleared predicate.

    2. Sample sized used for the test set and the data provenance

    This information is not provided in the given text.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not provided in the given text.

    4. Adjudication method for the test set

    This information is not provided in the given text.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    An MRMC study is not applicable to this device. This is a clinical chemistry analyzer, not a device requiring human reader interpretation of images or other subjective data.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is a standalone clinical chemistry analyzer. The device itself performs the analysis; human interpretation is involved in reviewing the results it produces. However, the document does not describe "standalone performance" in the context of comparing an algorithm to human performance, as it's an automated analytical system.

    7. The type of ground truth used

    This information is not explicitly stated. For clinical chemistry analyzers, ground truth would typically be established through reference methods (e.g., gold standard laboratory tests) for accuracy and precision studies, but such studies are not described for this specific modification. The submission implies that the "assay performance characteristics" are the same as the predicate, which would have had its performance validated against appropriate ground truth.

    8. The sample size for the training set

    This information is not provided in the given text. Given that this is a modification to an existing device, it's unlikely that a new "training set" for an algorithm, as understood in AI/ML, would be applicable or explicitly mentioned in such a submission.

    9. How the ground truth for the training set was established

    This information is not provided in the given text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K034000

    Validate with FDA (Live)

    Date Cleared
    2004-03-10

    (77 days)

    Product Code
    Regulation Number
    862.1705
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Vitalab Triglycerides Reagent, the Vitalab Calibrator and the Vitalab Selectra Analyzer are intended for use as a system for the quantitative determination of triglycerides in serum and plasma. Triglycerides results may be used for the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver diseases involving lipid metabolism, various endocrine disorders, or for assessing of the risk of developing cardiovascular diseases.

    Device Description

    The Vitalab Triglycerides Reagent, the Vitalab Calibrator and the Vitalab Selectra Analyzer are used as a system for the quantitative analysis of triglycerides in serum and plasma. The Vitalab Triglycerides Reagent determines triglycerides using the lipase/GPO enzymatic assay procedure coupled to a Trinder indicator reaction. The resulting increase in absorbance at 505 nm is proportional to the triglycerides concentration of the sample.

    AI/ML Overview

    This response is based on the provided text.

    1. Table of Acceptance Criteria and Reported Device Performance:

    Performance MetricAcceptance Criteria (Implied)Reported Device Performance
    Linearity/RecoveryLinear from 5 to at least 900 mg/dLLinear from 5 to at least 900 mg/dL, with a span from 0 mg/dL to 930 mg/dL. Regression statistics forced through the origin: sy.x = 3.7 mg/dL, (Vitalab Recoveries) = 1.164 x (Dilution Factor), n = 44.
    Precision (Within Run)Implicitly comparable to predicate or acceptable for intended useSerum 1 (mean 69 mg/dL): 1SD = 0.6, %CV = 0.9%Serum 2 (mean 291 mg/dL): 1SD = 1.6, %CV = 0.6%Serum 3 (mean 624 mg/dL): 1SD = 3.5, %CV = 0.6%
    Precision (Total)Implicitly comparable to predicate or acceptable for intended useSerum 1 (mean 69 mg/dL): 1SD = 1.1, %CV = 1.6%Serum 2 (mean 291 mg/dL): 1SD = 4.4, %CV = 1.5%Serum 3 (mean 624 mg/dL): 1SD = 11.3, %CV = 1.8%
    Method Comparison (Serum)Low intercept, slope near 1, low sy.x (comparable to predicate)Intercept: -4.0 mg/dL (-6.9 to 1.16 mg/dL 95% CI)Slope: 1.071 (1.059 to 1.083 95% CI)sy.x: 3.6 mg/dL
    Method Comparison (Plasma)Low intercept, slope near 1, low sy.x (comparable to predicate)Intercept: -0.2 mg/dL (-2.4 to 2.1 mg/dL 95% CI)Slope: 1.068 (1.057 to 1.079 95% CI)sy.x: 3.6 mg/dL
    Minimum Detection LimitClearly defined and adequately low for clinical use1 mg/dL (calculated as mean + 2 standard deviations of 30 replicate measurements of normal saline, where both mean and SD were 0 mg/dL. Rounded up due to assay's round-off error).
    Onboard Reagent StabilityTotal imprecision of triglycerides recoveries less than 2 mg/dL over 14 daysDocumented by assay of serum controls over 14 days. Total imprecision of triglycerides recoveries was less than 2 mg/dL.
    Calibration StabilityTotal imprecision of triglycerides recoveries less than 2 mg/dL over 7 daysDocumented by assay of serum controls over 7 days. Total imprecision of triglycerides recoveries was less than 2 mg/dL.
    Reconstituted Calibrator StabilityObserved change in triglycerides concentration less than 5% and statistically insignificant over 3 daysDocumented by assaying calibrators of increasing ages. The observed change in triglycerides concentration over three days was less than 5% and statistically insignificant.

    (Note: The "Acceptance Criteria (Implied)" column contains interpretations of what would generally be considered acceptable performance for such a device in comparison to a predicate, as explicit criteria are not always stated but inferred from the study design and conclusion of substantial equivalence.)


    2. Sample Size Used for the Test Set and Data Provenance:

    • Linearity/Recovery: n = 44 (This sample size likely refers to dilution factors or individual measurements across the range). Provenance: Not explicitly stated, but likely laboratory-prepared standards or spiked samples.
    • Precision:
      • Serum 1: 60 replicates
      • Serum 2: 60 replicates
      • Serum 3: 60 replicates
        Provenance: Commercially available control serum.
    • Method Comparison:
      • Serum samples: 59 specimens
      • Heparinized plasma samples: 60 specimens
        Provenance: Collected from adult patients. Country of origin not specified, but the submission is to the FDA, implying studies likely conducted in the US or internationally to US standards. Retrospective or prospective design is not explicitly stated, but the description "were collected from adult patients and were assayed for triglycerides" often implies a prospective or at least recently collected set of samples for the purpose of the study.
    • Minimum Detection Limit: 30 replicate measurements of normal saline. Provenance: Laboratory materials.
    • Onboard Reagent Stability & Calibration Stability: Not explicitly stated beyond "serum controls."
    • Reconstituted Calibrator Stability: Not explicitly stated beyond "calibrators of increasing ages."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    • This device is a quantitative clinical chemistry assay. The "ground truth" for such assays is typically established by reference methods or comparison to a legally marketed predicate device, rather than expert consensus on individual cases.
    • In this case, the "ground truth" for the method comparison study was the results from "another commercially available method," which served as the reference standard for comparison. The predicate device, the Beckman Trighycerides Reagent Kit and Synchron Multi-Calibrator, also serves as a benchmark for substantial equivalence.
    • Therefore, traditional "experts" like radiologists establishing ground truth for images are not applicable here. The "expertise" lies in the validated performance of the reference method and the established accuracy of commercially available control sera.

    4. Adjudication Method for the Test Set:

    • Not applicable. This is a quantitative chemical assay, not a subjective interpretation task that would typically involve adjudication by multiple experts. The comparison is statistical (Deming regression) between the new device and a reference method.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance:

    • Not applicable. This is a standalone in-vitro diagnostic (IVD) device for quantitative measurement of triglycerides, not an AI-assisted diagnostic tool that involves human readers interpreting results. Therefore, no MRMC study with human readers was conducted, and the concept of "improvement with AI vs without AI" does not apply.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    • Yes, this entire submission describes standalone performance studies for the Vitalab Triglycerides Reagent and Calibrator on the Vitalab Selectra Analyzer. The performance metrics (linearity, precision, method comparison, detection limit, stability) are all intrinsic to the device system itself, operating without direct human interpretive input influencing the quantitative result. The device outputs a numerical value for triglyceride concentration.

    7. The Type of Ground Truth Used:

    • Reference Method Comparison: For the method comparison studies, the "ground truth" was established by analyzing patient samples with "another commercially available method" (presumably a validated and accepted diagnostic method for triglycerides).
    • Known Concentrations/Standards: For linearity and detection limit, the ground truth was based on known concentrations of standards or normal saline (effectively 0 mg/dL).
    • Certified Control Sera: For precision and stability studies, commercially available control serum with established target values was used.

    8. The Sample Size for the Training Set:

    • This submission describes performance validation studies for an IVD reagent and calibrator. It does not mention any "training set" in the context of machine learning or AI algorithms. The assay procedure is based on a well-established lipase/GPO enzymatic reaction coupled to a Trinder indicator, not a statistical model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established:

    • Not applicable, as there is no mention of a training set for an AI/machine learning algorithm.
    Ask a Question

    Ask a specific question about this device

    K Number
    K020488

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2002-04-15

    (61 days)

    Product Code
    Regulation Number
    862.1705
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CARESIDE Triglyceride cartridge is intended for in vitro diagnostic use in conjunction with the CARESIDE Analyzer to quantitatively measure the concentration of triglycerides in anti-coagulated whole blood, plasma or serum. This product is indicated for use in the diagnosis and treatment of patients with primary or secondary hyperlipidemias. Hyperlipidemias may result from liver obstruction, diseases involving lipid metabolism, or various endocrine disorders. Triglyceride results are used together by the CARESIDE Analyzer with total cholesterol and HDL-cholesterol results to calculate LDL-cholesterol levels.

    Device Description

    CARESIDE Trighceride cartridges are used with the CARESIDE Analyzer to measure triglyceride concentration in anti-coagulated whole blood, plasma or serum specimens. The CARESIDE Triglyceride cartridge, a single use disposable in vitro diagnostic test cartridge, aids in specimen separation and delivers a measured volume of plasma or serum to a dry film to initiate the measurement of triglyceride concentration. The patented film cartridge contains all reagents necessary to measure triglyceride concentration. Each CARESIDE Triglyceride cartridge consists of a triglyceride-specific multi-layer reagent film mounted in a plastic base with a hinged lid. The user introduces the anticoagulated whole blood, serum, or plasma specimen into the cartridge Sample Well, closes the lid and inserts the cartridge into the CARESIDE Analyzer. Once loaded, the CARESIDE Analyzer scans the cartridge barcode, brings the cartridge and the contained specimen to 37℃, and spins the cartridge to move the sample from the sample deposition well into the cartridge channels and chambers. As the cartridge continues to spin, the blood cells are separated from the plasma/serum and the cells accumulate in the separation well. Approximately 8.5 microliters of plasma (or serum, as applicable) remain in the metering passage. Any excess sample flows into an overflow well. The plasma (or serum, as applicable) is automatically dispensed onto the multi-layer The triglyceride-containing specimen is distributed uniformly by the reagent film. spreading layer. The sample then passes through a reflection layer and into the reaction layer. Finally, the reaction mixture is pulled through the reaction layer by a suction layer where the NTB chromogen is converted into a purple formazan dye. As the cartridge spins, a photodiode measures film reflectance of light emitted from a wavelength-specific light emitting diode (LED) at a lixed time. The instrument uses the reflectance measurements and the lot-specific standard curve to calculate triglyceride concentration.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    CARESIDE Triglyceride Premarket Notification (K020488)

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implicitly defined by comparing the CARESIDE Triglyceride device's performance characteristics against its predicate device (Vitros TRIG DT Slides) and general analytical standards. The document doesn't explicitly state "acceptance criteria" as a separate section, but rather presents performance characteristics of both devices side-by-side. The key areas of comparison are:

    CharacteristicCARESIDE Triglyceride Performance (Reported)Predicate Device (Vitros TRIG DT Slides) PerformanceImplied Acceptance Criteria (relative to predicate)
    Intended UseAid in diagnosis/treatment of hyperlipidemiasSameMust be substantially equivalent
    IndicationsIn vitro diagnostic useIn vitro diagnostic useMust be substantially equivalent
    Measurement TypeQuantitativeQuantitativeMust be quantitative
    Method PrincipleDry film based lipase hydrolysis, reflectanceSameMust be substantially equivalent
    Specimen DilutionNot requiredNot requiredNot required
    MaterialsLipoprotein lipase and coupling enzymes/co-factorsLipoprotein lipase and coupling enzymes/co-factorsSubstantially equivalent (some same, some different)
    DetectorReflectance (570 nm)Reflectance (555 nm)Similar technology
    Test TimeApprox. 4 min warm-up + 6 min test time15 min warm-up + 5 min test timeComparable or improved efficiency
    Sample TypeSerum, plasma, whole bloodSerum, plasmaBroader (allowing whole blood) is considered an advantage and meets criteria
    Specimen Volume8.5 μl test volume (90 ± 10 μl applied)10 μlComparable small volume required
    CalibrationBar-coded on each cartridge, lot-specificRun Vitros DT II calibrators per new lot/as neededReliable and convenient calibration method
    Quality Control2 levelsSameStandard QC practices
    Reporting Unitsmg/dL or mmol/LSameStandard clinical units
    Reaction Temperature37 °CSameStandard biological reaction temperature
    Direct Blood SpecimenYes, whole bloodNoImproved functionality, meets criteria
    Reportable Range25 to 500 mg/dL15 to 400 mg/dLComparable or broader clinical range (improved upper limit, slightly higher lower)
    Accurate PipettingNot requiredRequiredImproved ease of use, meets criteria
    Reagent Pre-warmingNot requiredRequiredImproved ease of use, meets criteria
    Detection Limit25 mg/dL15 mg/dLClinical relevance within range
    Accuracy (Recovery)Mean recovery 99%Not providedHigh accuracy demonstrated
    Precision (Total CV)3.4% at 146 mg/dL2.1% at 189 mg/dLAcceptable clinical precision
    Method ComparisonCARESIDE = 0.98 (BM/Hitachi 902) + 2.92 mg/dLr= 0.99 (implied against a reference method)Strong correlation with a recognized reference method
    LinearitySlope and correlation coefficient within acceptable limits by mixing and dilutionNot providedDemonstrated linearity across reportable range
    InterferenceNo significant interference observed for tested interferents (Ascorbic acid, Bilirubin, Hemoglobin)None statedDemonstrated robustness against common interferents
    Specimen Types (Anticoagulants)No significant difference for sodium heparinized whole blood, sodium heparin plasma, EDTA plasma. Serum slightly higher.No significant difference for serum, heparin plasma, or EDTA plasma. Whole blood unsuitable.Demonstrated compatibility with relevant specimen types and anticoagulants, and an advantage with whole blood.

    Study Proving Device Meets Acceptance Criteria:

    The document describes a comparative performance characteristics study, though the details are somewhat summarized rather than a full protocol.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: Not explicitly stated for each characteristic. For "Precision," a single sample concentration (146 mg/dL) is mentioned as tested. For "Method Comparison," the slope and intercept are provided, indicating a regression analysis was performed on a set of samples, but the number of samples is not stated. Similarly, for "Interference" and "Specimen Types & Anticoagulants," various conditions were tested, but specific sample numbers (n) are not given.
    • Data Provenance: Not explicitly stated. Given that it's a premarket notification for a US market, it's highly likely the data was generated in a lab environment (prospective testing) but the country of origin of the samples themselves is not specified.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    • Not Applicable. This is an in vitro diagnostic device for measuring a biochemical analyte (triglycerides). The "ground truth" for the test set would be established by a well-characterized reference measurement method (e.g., an assay on a high-precision clinical chemistry analyzer like the BM/Hitachi 902, or a gas chromatography-mass spectrometry method if available at the time for direct comparison). It does not involve human expert interpretation of images or clinical cases.

    4. Adjudication Method for the Test Set:

    • Not Applicable. As this is a quantitative chemical assay, there is no human adjudication process involved in establishing the "ground truth" or comparing results. The comparisons are statistical and analytical.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

    • No. An MRMC study is relevant for diagnostic imaging or interpretation tasks where human readers make subjective assessments (e.g., radiologists reading scans). This device is an automated in vitro diagnostic assay, so MRMC studies are not applicable.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) was Done:

    • Yes. The entire performance characteristic section (Section IV.C. "Comparative Performance Characteristics") is effectively a standalone performance evaluation of the CARESIDE Triglyceride system (cartridge + Analyzer). The device is designed to be an automated measurement device, and the reported accuracy, precision, linearity, and method comparison data reflect its standalone performance. The document highlights features like "accurate pipetting not required" and "reagent pre-warming not required," which further emphasize its automated, standalone nature.

    7. The Type of Ground Truth Used:

    • For Method Comparison, the ground truth was established by another well-accepted clinical chemistry analyzer, specifically the "BM/Hitachi 902." The comparison yielded the equation: CARESIDE = 0.98 (BM/Hitachi 902) + 2.92 mg/dL. This indicates the BM/Hitachi 902 served as the reference or comparative "ground truth" for evaluating the CARESIDE device's quantitative accuracy.
    • For Accuracy (Recovery), the ground truth would be the known concentration of triglycerides in spiked samples or certified reference materials, where the device's measured value is compared to the expected true value.
    • For Precision, the ground truth is the statistical property of repeatability and reproducibility, typically assessed by running samples multiple times to determine the variability (CV%).
    • For Linearity, the ground truth is derived from preparing samples with known serially diluted or mixed concentrations and ensuring the device's readings are proportional to these known values.

    8. The Sample Size for the Training Set:

    • Not applicable / Not explicitly stated for this type of device. Clinical chemistry devices like this typically undergo extensive analytical validation (accuracy, precision, linearity, interference) using characterized samples. While there's an internal "lot-specific standard curve" mentioned for calibration, this isn't a "training set" in the machine learning sense. The device's underlying chemistry and optical detection principles are well-established. The development process would involve optimizing reagents and calibration, but the specific term "training set" with a defined sample size as used in AI/ML is not relevant here.

    9. How the Ground Truth for the Training Set Was Established:

    • Not applicable for a "training set" in the AI/ML sense. For the calibration curves that the instrument uses, "ground truth" is established through carefully characterized calibrator materials with known triglyceride concentrations, measured by highly accurate reference methods. The device then generates a standard curve based on these calibrators to convert raw reflectance data into triglyceride concentrations.
    Ask a Question

    Ask a specific question about this device

    K Number
    K010871

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2001-07-09

    (108 days)

    Product Code
    Regulation Number
    862.1705
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    Page 1 of 2