Search Filters

Search Results

Found 48 results

510(k) Data Aggregation

    K Number
    K241366

    Validate with FDA (Live)

    Date Cleared
    2024-10-29

    (168 days)

    Product Code
    Regulation Number
    868.5870
    Age Range
    0 - 18
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The demand valve is used for the application of 100 vol% oxygen during manual ventilation with the resuscitator bag, as well as during direct non-invasive inhalation with the resuscitator mask in spontaneously breathing patients.

    Device Description

    The valve, non - rebreathing QualityFlow O2 demand valve is a medical device to deliver medical oxygen or a medical air - oxygen mixture to a patient on demand. Unlike continuous oxygen systems, the demand valve only supplies medical gas when the patient inhales, reducing oxygen consumption and ensuring the patient receives gas only when needed. The QualityFlow O2 Demand Valve from DEHAS consists of a membrane that is activated by negative pressure during inspiration and a valve mechanism that controls gas flow. During inspiration, the valve opens due to the movement of the membrane, while during expiration it is closed by a return spring. The function of the demand valve is based on providing gas only when a negative pressure is applied to the membrane, e.g. by the patient inhaling with a connected mask or suction of the resuscitator when connected (expansion after previous compression of the resuscitator). The valve therefore ensures that gas only flows when it is needed.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification for a medical device called "QualityFlow O2 Demand Valve". This type of submission aims to demonstrate that a new device is substantially equivalent to a legally marketed predicate device, and thus does not require clinical studies in many cases.

    Based on the provided text, the device in question (QualityFlow O2 Demand Valve) is a mechanical device, not an AI/software-driven one. Therefore, the questions related to AI/MRMC studies, training/test sets, ground truth establishment by experts, and adjudication methods are not applicable to this submission. The "acceptance criteria" discussed in this document refer to the device's ability to meet various engineering, safety, and performance standards, rather than diagnostic accuracy metrics common for AI/ML devices.

    Here's an analysis based on the information relevant to this type of medical device submission:

    1. Table of Acceptance Criteria (Relevant Standards) and Reported Device Performance:

    The document summarizes non-clinical tests performed according to international standards to demonstrate substantial equivalence and that the device performs as intended.

    Acceptance Criteria (Standard Used)Description of TestReported Device Performance
    ISO 5356-1 Third edition 2004-05-15: Anaesthetic and respiratory equipment - Conical connectors: Part 1: Cones and socketsSpecifies the dimension and dimensional requirements for cones and bushings intended for the connection of anesthesia and ventilator equipment (e.g., in ventilation systems, anesthetic gas transport systems and vaporizers).Result: pass
    IEC 62366-1 Edition 1.1 2020-06 CONSOLIDATED VERSION: Medical devices - Part 1: Application of usability engineering to medical devicesUsed to verify and validate the usability of the QualityFlow O2 demand valve.Result: pass
    ISO 14971 Third Edition 2019-12: Medical devices - Application of risk management to medical devicesProcess for identification and assessment of hazards and associated risks, control of these risks, and monitoring effectiveness of risk control measures.Result: pass
    ISO 15001:2010: Anesthetic and respiratory equipment - Compatibility with oxygenUsed to ensure the oxygen compatibility of the used materials.Result: pass
    ISO 10993-1 Fifth edition 2018-08: Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management processUsed for biological evaluation of medical devices and testing within a risk management process.Result: pass
    ISO 18562-1 First edition 2017-03: Biocompatibility evaluation of breathing gas pathways in healthcare applications - Part 1: Evaluation and testing within a risk management processEvaluated biocompatibility of breathing gas pathways, specifically for indirect patient contact via medical oxygen (gas pathway contact, limited exposure < 24 hours).Result: pass
    ISO 18562-3: Biocompatibility evaluation of breathing gas pathways in healthcare applications - Part 3: Tests for emissions of volatile organic compoundsAssessed gas emission of VOCs with a risk assessment of any identified chemical.Result: pass

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size for Test Set: The document does not specify a "sample size" in terms of patient data or images, as this is a non-AI/ML mechanical device. The tests performed are laboratory/bench tests on the device itself and its components, according to the listed international standards.
    • Data Provenance: Not applicable in the context of patient data. The provenance for the testing refers to the execution of tests in a laboratory setting by the manufacturer (DEHAS Medical Systems GmbH) or their designated testing facilities, adhering to international standards. The tests are non-clinical.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:

    • Not applicable. "Ground truth" in the context of device performance is established by the detailed specifications of the international standards themselves and the objective measurements derived from testing the physical device against those specifications, using calibrated equipment. There is no subjective human interpretation of patient data here that requires expert consensus for ground truth.

    4. Adjudication Method for the Test Set:

    • Not applicable. Adjudication methods (e.g., 2+1, 3+1) are used for resolving disagreements among human readers or expert reviewers, typically in diagnostic studies with subjective image interpretation. This is a mechanical device tested against objective engineering standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

    • No. MRMC studies are specific to evaluating the diagnostic performance of AI or imaging devices with human readers interpreting cases. This device is a mechanical demand valve, not an imaging or AI diagnostic device.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study Was Done:

    • Not applicable. This is not an algorithm. The "standalone" performance here refers to the device's inherent mechanical functionalities (e.g., flow rates, pressure triggers, material compatibility) as measured by laboratory testing according to the specified standards. All tests listed in the "Summary of non-clinical tests" (Table 3 on page 10-11) evaluate the device's technical performance.

    7. The Type of Ground Truth Used:

    • Ground Truth: The "ground truth" for this type of device is defined by the objective, quantifiable specifications and pass/fail criteria outlined in the referenced international standards (e.g., ISO 5356-1 for connector dimensions, ISO 15001 for oxygen compatibility, ISO 18562-3 for VOC emissions). These are engineering and safety standards, not medical diagnostic "ground truth" derived from patient outcomes or expert consensus.

    8. The Sample Size for the Training Set:

    • Not applicable. This is a mechanical device, not an AI/ML model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established:

    • Not applicable. As there is no training set for an AI/ML model, no ground truth was established for it.

    Summary of Study Proving Device Meets Acceptance Criteria:

    The study proving the device meets the acceptance criteria is a series of non-clinical bench tests and evaluations conducted according to recognized international standards. These tests assess various aspects of the device's design, materials, usability, and performance to ensure it is safe and effective and substantially equivalent to its predicate device. The results for all listed standards were "pass", indicating the device met the pre-defined criteria for each test. The manufacturer concluded that there are "no differences which raise any different questions of safety and effectiveness" compared to the predicate device, and that its performance is "better or identical" to the predicate. No animal or clinical testing was performed for this 510(k) submission, as it was deemed unnecessary to support substantial equivalence.

    Ask a Question

    Ask a specific question about this device

    K Number
    K234053

    Validate with FDA (Live)

    Date Cleared
    2024-08-09

    (231 days)

    Product Code
    Regulation Number
    868.5870
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This product is intended for use as an in-line pressure relief device, designed to limit the system pressure within the limits of its stated technical specifications.
    This product is indicated for patients in a hospital for use by appropriately qualified healthcare professionals.

    Device Description

    The F&P Optiflow Flow Diverter is intended for use as an in-line pressure relief device, designed to limit the system pressure when used with the compatible F&P Optiflow Oxygen Kits and nasal interfaces.
    The F&P Flow Diverter is placed between the flow meter and the dryline during use. It is a multi-patient use prescription-only device, provided in a non-sterile state. It operates at flow rates between 5 to 70 L/min and is intended to be used by appropriately qualified healthcare professionals in hospitals.
    When an anesthesia mask is applied over the Optiflow Switch interface, the respiratory gas no longer has a pathway to flow through the nasal interface to the patient. This will trigger respiratory gas flow to be diverted into the atmosphere.

    AI/ML Overview

    The medical device in question is the "F&P Optiflow™ Flow Diverter", a nonrebreathing valve (Product Code: CBP).

    Here's an analysis of the acceptance criteria and study information provided:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document does not explicitly provide a table of acceptance (performance) criteria with numerical targets. Instead, it states that "Performance testing of the Flow Diverter was completed and confirms the subject device does not raise new questions of safety and effectiveness." It implies that the device met the safety and performance standards outlined in the listed ISO and IEC standards.

    However, based on the comparison table with the predicate device, we can infer some key performance characteristics. The document emphasizes equivalence to the predicate device, the "OJR215 Pressure Relief Manifold (K173770)".

    Characteristic / Acceptance Criteria (Inferred)F&P Optiflow™ Flow Diverter (Subject Device) PerformanceOJR215 Pressure Relief Manifold (Predicate Device) PerformanceComments (for equivalence)
    Indications for UseIn-line pressure relief, limits system pressure within stated specs, for hospital use by qualified professionals.Protects neonates, infants, children from excessive inspiratory pressure in downstream occlusion in continuous flow systems via nasal cannula. For flow rates 0.5-36 L/min. Upstream of patient, single use, prescription only.Equivalent (despite slight wording differences, the core function of pressure relief in a respiratory circuit is comparable).
    AvailabilityPrescription use (Part 21 CFR 801 Subpart D)Prescription use (Part 21 CFR 801 Subpart D)Identical
    Intended Use EnvironmentHospitalHospital/InstitutionalIdentical
    UsersQualified healthcare professionalsQualified healthcare professionalsIdentical
    Ambient Operating Temperature18 to 26 °C (64.4 to 78.8°F)18 to 26 °C (64.4 to 78.8°F)Identical
    Shelf LifeUp to 7 years3 yearsDifferent (Subject device has longer shelf life, which is generally a performance improvement from a user's perspective, but would have required stability testing).
    Storage Temperature-10°C to +50 °C (14 to 122 °F)-10°C to +50 °C (14 to 122 °F)Identical
    SterilityDevice not provided sterile.Device not provided sterile.Identical
    Operating Pressure - Maximum PressureMaximum system pressure of 60cmH2O at maximum rated flow rate of 70L/minMaximum relief pressure is 75 cm H2O at maximum rated flow of 36 L/min.Equivalent (Both devices are designed to provide pressure relief; the specific maximum pressures and flow rates differ but are within acceptable ranges for their respective claims. The subject device operates at a lower maximum pressure but at a higher maximum flow).
    Flow Range5-70 L/min0.5-36 L/minEquivalent (Both cover a range of flow rates, the subject device covers a higher range).
    Application/TherapyNasal High Flow TherapyNasal High Flow TherapyIdentical
    Principles of OperationValve seals, allows gas flow; if relief pressure exceeded, shaft moves, seal opens, pressure relieved, flow diverted to atmosphere.Valve seals, allows gas flow; if relief pressure exceeded, shaft moves, seal opens, pressure relieved, flow diverted to atmosphere.Equivalent
    Inlet ConnectionDISS 1240Inlet barbDifferent (Requires compatibility verification)
    Outlet ConnectionF&P female safety connection22mm female taper connection as per ISO 5356-1Different (Requires compatibility verification)
    Housing MaterialPolycarbonateABSEquivalent (Both are suitable medical-grade plastics).
    Plunger Seal FunctionRests on valve body, accurately relieves pressure by opening when relief pressure is reached.Seals manifold valve during normal use, accurately relieves pressure by opening when relief pressure is reached.Equivalent
    Spring Pressure Relief Valve MechanismShaft moved by valve seal that moves when set pressure is exceeded.Shaft moved by a spring pressure relief valve that moves when set pressure is exceeded.Equivalent
    Manifold Shroud / Tamper PreventionRear and front caps prevent user modification to preset pressure relief.Prevents access to pre-set operating pressures and tampering.Equivalent

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not explicitly state a sample size for a "test set" in the context of a clinical performance study. The testing described is non-clinical performance testing against standards. Therefore, concepts like sample size for clinical test data, country of origin, or retrospective/prospective data are not applicable here. The testing would have involved multiple units of the device under various conditions described by the cited standards.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    Not applicable. This is not a clinical study involving expert interpretation or "ground truth" establishment in a diagnostic context. The performance is assessed against engineering and safety standards.

    4. Adjudication Method for the Test Set:

    Not applicable. There's no human interpretation or diagnostic assessment that would require an adjudication method.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This device is a mechanical pressure relief valve, not an AI-powered diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable. This is not an algorithm-based device.

    7. The type of ground truth used:

    The "ground truth" for this device would be its performance against the objective criteria defined in the referenced industry standards (e.g., ISO 80601-2-74:2017 for respiratory humidifying equipment, IEC 60601-1 for basic safety and essential performance, ISO 80601-2-90:2021 for respiratory high-flow therapy equipment). This includes measurements such as maximum pressure relief, flow rates, temperature tolerances, and mechanical integrity, as well as adherence to design principles and safety features.

    8. The sample size for the training set:

    Not applicable. This is a medical device, not a machine learning model requiring a training set.

    9. How the ground truth for the training set was established:

    Not applicable.

    In summary:

    The document describes a 510(k) submission for a non-AI, mechanical medical device. The "acceptance criteria" are primarily established by compliance with recognized international and national standards for medical devices and respiratory equipment (ISO, IEC, CGA), as well as demonstrating substantial equivalence to a predicate device. The "study" proving this is a series of non-clinical performance and safety tests conducted in accordance with these standards. The concepts of clinical test sets, ground truth by experts, and training sets for algorithms are not relevant to this type of device submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K210992

    Validate with FDA (Live)

    Date Cleared
    2021-12-23

    (265 days)

    Product Code
    Regulation Number
    868.5870
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The exhalation valves are intended to be used in ventilator circuits consistent with the indicated use of the ventilator to which they are attached.

    The single limb model directs flow and allows for pressure monitoring and may be used within hospitals and for home care use.

    The dual limb model regulates flow and is for use in hospitals and for home care use.

    The exhalation valve are used with adults, pediatrics, and neonates.

    The exhalation valves are for single patient use and can be used for a maximum of 30 days.

    Device Description

    The Altech® Exhalation Valves (Single Limb) are available in 2 configurations intended for use in Single and Dual Limb ventilator circuits. The difference between the configurations is location of exhaust of the expired gas to the room. In the Single limb, the valve is integrated to the limb and expired gas exhausts through the opening below the valve whereas in the valve is attached to the end of the exhalation limb and the expired gas exhausts through device end of the valve.

    AI/ML Overview

    The Altech® Exhalation Valves (Single Limb and Dual Limb) underwent non-clinical performance testing to demonstrate substantial equivalence to the predicate device.

    1. Table of Acceptance Criteria and Reported Device Performance:

    Test MethodologyAcceptance CriteriaReported Device Performance
    ISO 5356-1 (Conical fittings)Meet performance criteriaFittings were tested according to ISO 5356-1 and meet the performance criteria
    ISO 5367 (Leakage)< 70 ml/minLeakage of <70 ml/min
    ISO 5367 (Resistance to Flow)Adult < 0.06 cmH2O/l/m, Pediatrics < 0.12 cmH2O/l/m, Neonate < 0.74 cmH2O/l/mAdult < 0.06 cmH2O/l/m, Pediatrics < 0.12 cmH2O/l/m, Neonate < 0.74 cmH2O/l/m
    Performance after 30 days continuous useLeakage and resistance to flow meet acceptance criteria per standardTesting of leakage, resistance to flow met the acceptance criteria per each standard
    Performance under various ventilator modesCompatibility of the device with a ventilatorTesting over different ventilator modes was performed to confirm the compatibility of the device with a ventilator
    ISO 10993-5 (Cytotoxicity)Non-cytotoxicFound non-cytotoxic
    ISO 10993-10 (Sensitization/Irritation)Non-sensitizer, Non-irritantNon-sensitizer, Non-irritant
    ISO 10993-11 (Pyrogenicity/Toxicity)Non-pyrogenic, Non-toxicNon-pyrogenic, Non-toxic
    ISO 18562-2 (Particulate Material)Met predefined acceptance criteriaMet the predefined acceptance criteria
    ISO 18562-3 (Volatile Organic Chemicals)Margin of Safety > 1Toxicological Risk Assessment resulted in Margin of Safety >1
    Leakage in main bodyNot explicitly stated as acceptance criteria, but compared to predicate (predicate had <0.5 ml/min)6 ml/min for Subject Device, 2 ml/min for Predicate Device

    2. Sample Size Used for the Test Set and Data Provenance:

    The document describes non-clinical performance data and comparative testing to a predicate device, rather than a clinical study with a "test set" of patient data. Therefore, information about patient sample size and data provenance (e.g., country of origin, retrospective/prospective) is not applicable in the context of this submission. The "test set" refers to the physical devices undergoing various laboratory tests. The devices were tested pre- and post-aging, after 30 days of use, and after storage conditions.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

    This information is not applicable. The "ground truth" for the non-clinical performance tests is established by published ISO standards and engineering specifications, not by expert consensus on clinical cases.

    4. Adjudication Method for the Test Set:

    This information is not applicable. Adjudication methods like "2+1" or "3+1" are typically used in clinical studies for resolving discrepancies in expert interpretations of patient data. The non-clinical performance tests rely on objective measurements and established standards.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No MRMC comparative effectiveness study was mentioned. The submission focuses on non-clinical performance and substantial equivalence to a predicate device, not on human-in-the-loop performance or comparison of human readers with and without AI assistance.

    6. Standalone (Algorithm Only) Performance:

    This information is not applicable. The device is an exhalation valve, a physical medical device, not a software algorithm or AI. The tests performed are on the physical device's function.

    7. Type of Ground Truth Used:

    The "ground truth" for these performance tests is based on:

    • International Standards (ISO): Specifically, ISO 5356-1, ISO 5367, ISO 10993-5, ISO 10993-10, ISO 10993-11, ISO 18562-2, and ISO 18562-3.
    • Engineering Specifications: Predefined acceptance criteria for various parameters like resistance to flow and material composition.
    • Toxicological Risk Assessment: For volatile organic chemicals.

    8. Sample Size for the Training Set:

    This information is not applicable. There is no "training set" as this is a physical medical device, not an AI or machine learning model.

    9. How the Ground Truth for the Training Set Was Established:

    This information is not applicable for the same reason as above.

    Ask a Question

    Ask a specific question about this device

    K Number
    K192285

    Validate with FDA (Live)

    Device Name
    CPR Face Shield
    Date Cleared
    2020-11-25

    (461 days)

    Product Code
    Regulation Number
    868.5870
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This device is used as a physical barrier for mouth to mouth resuscitation. The use of this device does not guarantee complete protection from disease transmission.

    Device Description

    Not Found

    AI/ML Overview

    I am sorry, but the provided text from the FDA 510(k) clearance letter for the "CPR Face Shield" (K192285) does not contain any information about a study proving the device meets acceptance criteria.

    The document is a standard FDA clearance letter, which confirms that the device is "substantially equivalent" to legally marketed predicate devices. It discusses regulatory matters such as registration, listing, manufacturing practices, and adverse event reporting.

    Therefore, I cannot fulfill your request to describe the acceptance criteria and the study that proves the device meets the acceptance criteria, as this information is not present in the provided text.

    The document details:

    • The device name: CPR Face Shield
    • Regulation Number: 21 CFR 868.5870 (Nonrebreathing Valve)
    • Regulatory Class: Class II
    • Product Code: CBP
    • Indications for Use: "This device is used as a physical barrier for mouth to mouth resuscitation. The use of this device does not guarantee complete protection from disease transmission."

    None of these sections imply or describe performance studies with acceptance criteria as one would expect for an AI/software as a medical device (SaMD) or other complex diagnostic/therapeutic device. For a simple physical barrier like a CPR face shield, the "acceptance criteria" for FDA clearance primarily revolve around substantial equivalence to existing devices, safety considerations inherent to its material and design, and its stated intended use. A formal "study" with the characteristics you outlined (test sets, experts, MRMC, etc.) would not typically be required or relevant for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K173770

    Validate with FDA (Live)

    Date Cleared
    2019-04-05

    (480 days)

    Product Code
    Regulation Number
    868.5870
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OJR215 Pressure Relief Manifold is designed to protect neonates, infants and children from excessive inspiratory pressure in the event of a downstream occlusion occurring in continuous flow breathing gas delivery systems via nasal cannula.

    The device is intended for use with flow rates greater than 0.5 L/min up to, and including. 36 L/min. The OJR215 is fitted upstream of the patient. The OJR215 is single use only and is prescription only.

    Device Description

    Not Found

    AI/ML Overview

    I am sorry, but I cannot answer the question with the provided context. The document is an FDA 510(k) clearance letter and Indications for Use statement for a medical device called the "OJR215 Pressure Relief Manifold." This document explains the device's intended use and regulatory classification but does not include information about acceptance criteria, device performance studies, sample sizes, expert qualifications, or ground truth establishment. These details would typically be found in the 510(k) submission itself or related study reports, which are not part of this document.

    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CPR Mask with Oxygen Port is designed to assist in providing immediate life support (mouth to mask ventilation) to health emergency victims requiring oxygen support (inhalation) or cardiopulmonary resuscitation (CPR) rescue techniques. The CPR Mask with Oxygen Port is intended for prescription use.

    The CPR Mask without Oxygen Port is designed to assist in providing immediate life support (mouth to mask ventilation) to health emergency victims requiring cardiopulmonary resuscitation (CPR) rescue techniques. The CPR Mask without Oxygen Port is intended for over-the-counter use.

    The Infant CPR Mask without Oxygen Port is designed to provide immediate life support (mouth to mask ventilation) to health emergency victims requiring cardiopulmonary resuscitation (CPR) rescue techniques. The Infant CPR mask is indicated for use on infants with body weight up to 10 kg (22 lbs). The Infant CPR Mask is intended for over-the-counter use.

    The Non-Rebreathing Valve is a single-use, non-sterile device intended to be used with a ventilation mask to provide mouth-to-mask ventilation to health emergency victims requiring cardiopulmonary resuscitation (CPR) rescue techniques. It is intended for over-the-counter use.

    Device Description

    Compower Emergency CPR masks with non-rebreathing valves are designed to assist in providing immediate life support to health emergency victims requiring oxygen support and cardiopulmonary resuscitation rescue techniques. The devices subject to this 510(k) include the following: CPR Mask with Oxygen Port, CPR Mask without Oxygen Port, Infant CPR Mask without Oxygen Port, Non-Rebreathing Valve (15mmOD), Non-Rebreathing Valve (18.5mm).

    CPR Mask with Oxygen Port: comprised of an air cushion, foldable mask cover with oxygen port, a non-rebreathing valve and an elastic head strap. Constructed of transparent polyvinyl chloride. Provided non-sterile and is intended for single use only.

    CPR Mask without Oxygen Port: comprised of an air cushion, foldable mask cover, a non-rebreathing valve and an elastic head strap. Constructed of transparent polyvinyl chloride. Provided non-sterile and is intended for single use only.

    Infant CPR Mask without Oxygen Port: comprised of an air cushion, mask cover and non-rebreathing valve. Constructed of transparent polyvinyl chloride. Provided non-sterile and is intended for single use only.

    Non-Rebreathing Valve: a non-sterile, single-use device designed with a rigid styrene-butadiene copolymer housing, a silicone valve plate and polypropylene filter. Compatible for use with 22mm connectors used in standard ventilation masks. Designed with 15mmOD and 18.5mm ports.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document presents performance data primarily in the context of comparison to predicate devices, rather than explicit acceptance criteria with pass/fail values. The "acceptance criteria" are implied by meeting or falling within ranges comparable to the predicate devices and adherence to relevant standards.

    Comparison of Compower CPR Mask (with/without Oxygen Port) to KYOLING CPR Mask (Predicate)

    MetricAcceptance Criteria (Implied by Predicate & Standard)Reported Compower CPR Mask PerformanceReported KYOLING CPR Mask Performance (Predicate)
    Inspiratory Resistance Specification≤0.5 kPa at 50 L/min (from AS 4259)0.40 - 0.49 kPa at 50 L/min0.30 - 0.31 kPa at 50 L/min
    Expiratory Resistance Specification≤0.5 kPa at 50 L/min (from AS 4259)0.20 - 0.30 kPa at 50 L/min0.31 - 0.32 kPa at 50 L/min
    Biocompatibility - CytotoxicityPass (ISO 10993-5)PassSame (Implied by Predicate Comparison)
    Biocompatibility - SensitizationPass (ISO 10993-10)PassSame (Implied by Predicate Comparison)
    Biocompatibility - IrritationPass (ISO 10993-10)PassN/A (Not explicitly listed for predicate)
    Standards UsedConformance to ISO 5356-1 and AS 4259-1995Conformance demonstratedConformance demonstrated
    Visual InspectionImplied to meet specificationsPassedN/A
    Dimensional MeasurementImplied to meet specificationsPassedN/A
    TransparencyImplied to meet specificationsPassedN/A
    Function after Contamination/ImmersionImplied to meet specificationsPassedN/A
    Performance After Storage and AgingImplied to meet specificationsPassedN/A
    Mean Delivered Oxygen ConcentrationImplied to meet specificationsPassedN/A

    Comparison of Compower Infant CPR Mask to EMS Child/Infant Mask (Primary Predicate) & Laerdal Paediatric Pocket Mask (Reference)

    MetricAcceptance Criteria (Implied by Predicate/Reference & Standard)Reported Compower Infant CPR Mask PerformanceReported EMS Child/Infant Mask Performance (Predicate)Reported Laerdal Paediatric Pocket Mask Performance (Reference)
    Inspiratory Resistance Specification<0.5 kPa at 50 L/min (from AS 4259)0.46-0.48 kPa at 50 L/minNot Known0.32 kPa at 50 L/min
    Expiratory Resistance Specification<0.5 kPa at 50 L/min (from AS 4259)0.27-0.29 kPa at 50 L/minNot Known0.42 kPa at 50 L/min
    Dead Space≤120 ml (AS 4259-1995)44.4 mlNot Known90.5 ml
    Biocompatibility - CytotoxicityPass (ISO 10993-5)PassNot KnownNot Known
    Biocompatibility - SensitizationPass (ISO 10993-10)PassNot KnownNot Known
    Biocompatibility - IrritationPass (ISO 10993-10)PassNot KnownNot Known
    Standards UsedConformance to ISO 5356-1 and AS 4259-1995Conformance demonstratedConformance to ASTM 920-93 (other not known)Conformance to ISO 5356-1 and AS 4259-1995

    Comparison of Compower Non-Rebreathing Valve to Seal Rite™ Non-Rebreathing Valve (Predicate)

    MetricAcceptance Criteria (Implied by Predicate & Standard)Reported Compower Non-Rebreathing Valve PerformanceReported Seal Rite™ Non-Rebreathing Valve Performance (Predicate)
    Inspiratory Resistance Specification≤0.5 kPa at 50 L/min (from AS 4259)15mmOD: 0.46-0.49 kPa; 18.5mm: 0.46-0.48 kPa0.113 kPa at 50 L/min
    Expiratory Resistance Specification≤0.5 kPa at 50 L/min (from AS 4259)15mmOD: 0.26-0.30 kPa; 18.5mm: 0.26-0.30 kPa0.0321 kPa at 50 L/min
    Biocompatibility - CytotoxicityPass (ISO 10993-5)PassPass
    Biocompatibility - SensitizationPass (ISO 10993-10)PassPass
    Biocompatibility - IrritationPass (ISO 10993-10)PassPass
    Standards UsedConformance to AS 4259-1995 and ISO 5356-1Conformance demonstratedConformance demonstrated

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample size used for the test set for performance testing, nor does it detail the country of origin for the data or whether it was retrospective or prospective. The non-clinical testing was conducted to confirm conformance to specified requirements, implying internal laboratory testing.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    N/A. This type of information (experts, ground truth for test set, qualifications) is typically relevant for studies involving subjective interpretation (e.g., medical imaging AI). For medical devices like CPR masks and valves, performance is assessed through objective, quantifiable non-clinical tests based on engineering standards.

    4. Adjudication Method for the Test Set

    N/A. Adjudication methods are typically used in clinical studies or studies requiring expert consensus for ground truth, which is not applicable to the non-clinical testing described for these devices.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    N/A. This is a non-AI medical device. MRMC studies are not applicable.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    N/A. This is a non-AI medical device. Standalone algorithm performance is not applicable.

    7. The type of ground truth used

    For the non-clinical performance and biocompatibility testing, the "ground truth" was established by objective measurements against established international and national standards (ISO 5356-1, AS 4259-1995, ISO 10993 series) and product specifications. This means the device's performance was compared to predefined, quantifiable limits and characteristics outlined in these standards.

    8. The sample size for the training set

    N/A. There is no "training set" as this is not an AI/ML device. The device design and materials are based on established engineering principles and comparison to existing predicate devices.

    9. How the ground truth for the training set was established

    N/A. There is no "training set" or corresponding ground truth for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K152521

    Validate with FDA (Live)

    Date Cleared
    2016-06-03

    (274 days)

    Product Code
    Regulation Number
    868.5870
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Seal Rite Non-Rebreathing Valve is a single-use, non-sterile device intended to be used with a ventilation mask (without oxygen port) to provide mouth-to-mask ventilation to health emergency victims requiring cardiopulmonary resuscitation ("CPR") rescue techniques. It is intended for over-the-counter use.

    Device Description

    The Seal Rite™ Non-Rebreathing Valve ("Seal Rite Valve") is a resuscitation valve designed for resuscitation using expired air for ventilation. It has fittings on the grooved patient end that will adapt to standard masks (22 mm ID) used for resuscitation. The Seal Rite Valve allows the rescuer to blow air into a mask through its silicone one-way valve and directs the patient's exhaled air away from the rescuer via a rigid one-way valve and multiple exhaust ports. The Seal Rite Valve helps protect the rescuer from potential patient contamination. The Seal Rite Valve does not have an oxygen port.

    AI/ML Overview

    The provided text is a 510(k) summary for the Seal Rite Non-Rebreathing Valve (K152521). It describes the device, its intended use, and the testing conducted to establish its substantial equivalence to a predicate device.

    Here's an analysis of the acceptance criteria and study information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The device was tested to ensure conformance with the FDA Recognized AS 4259-1995 standard ("Ancillary devices for expired air resuscitation") and passed all testing requirements. The table below summarizes the key performance parameters tested and the reported results. The explicit acceptance criteria (e.g., maximum allowable resistance values per the standard) are not fully detailed in the provided text, but the text states that the device "passed all testing requirements" and "performs as intended and meets the requirements of the AS 4259-1995 standard." I've extracted specific quantitative performance values where available and indicated "Pass" for qualitative tests.

    Acceptance Criteria (Implied by standard AS 4259-1995)Reported Device Performance (Seal Rite™ Non-Rebreathing Valve)Predicate Device Performance (Respironics Rescue Valve)
    Biocompatibility:
    Meet Cytotoxicity requirements (ISO 10993-5)PassNot explicitly stated for predicate in this document
    Meet Irritation requirements (ISO 10993-10)PassNot explicitly stated for predicate in this document
    Meet Sensitization requirements (ISO 10993-12)PassNot explicitly stated for predicate in this document
    Bench Testing (Functional Performance/Conformity to AS 4259-1995):
    Ventilation Performance requirementsPassNot explicitly stated for predicate in this document
    Expiratory Resistance for the Patient requirementsPassNot explicitly stated for predicate in this document
    Expiratory Resistance for the Rescuer requirementsPassNot explicitly stated for predicate in this document
    Inspiratory Resistance for a Spontaneously Breathing Patient requirementsPassNot explicitly stated for predicate in this document
    Function After Contamination with Stomach Contents requirementsPassNot explicitly stated for predicate in this document
    Function After Immersion in Water requirementsPassNot explicitly stated for predicate in this document
    Measurement of Dead Space requirementsPassNot explicitly stated for predicate in this document
    Resistance to Disengagement of Parts requirementsPassNot explicitly stated for predicate in this document
    High and Low Storage Conditions requirementsPassNot explicitly stated for predicate in this document
    High Temperature Operation requirementsPassNot explicitly stated for predicate in this document
    Low Temperature Operation requirementsPassNot explicitly stated for predicate in this document
    Attempted Displacement of Valve by Finger, Function After Dropping, Function After Potentially Deforming Weight requirementsPassNot explicitly stated for predicate in this document
    Conformance of Connectors requirementsPassNot explicitly stated for predicate in this document
    ISTA Procedure 2A Shipping Testing requirementsPassNot explicitly stated for predicate in this document
    Expiratory Resistance at 50 LPM0.327 cm H2O (0.0321 kPa)0.25 cm H2O at 50 LPM (per K142402 510(k) Summary)
    Inspiratory Resistance at 50 LPM1.15 cm H2O (0.113 kPa)1.85 cm H2O at 50 LPM (per K142402 510(k) Summary)
    Inlet ConnectorStandard 22 mm IDStandard 22 mm ID
    Outlet ConnectorStandard 22 mm OD/15 mm IDStandard 22 mm OD/15 mm ID

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample sizes for the non-clinical tests (biocompatibility and bench testing). It mentions that the testing was conducted according to recognized standards (FDA Blue Book Memo G95-1, ISO 10993 series, AS 4259-1995, ISTA Procedure 2A). Typically, these standards specify the number of samples required for each test. The data provenance is implied to be from laboratory testing as mandated by these standards, not from human subjects or real-world use. It is retrospective in the sense that the testing was performed to support the 510(k) submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable as the evaluation relies on a non-clinical, benchtop and laboratory testing against established engineering and biocompatibility standards, not on expert interpretation of medical data.

    4. Adjudication Method for the Test Set

    This information is not applicable as the evaluation relies on a non-clinical, benchtop and laboratory testing against established engineering and biocompatibility standards. There is no human adjudication process described.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    This information is not applicable. The Seal Rite Non-Rebreathing Valve is a mechanical medical device, not an AI or imaging diagnostic tool. Therefore, an MRMC study with human readers (or AI assistance) would not be relevant.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This information is not applicable as the device is a mechanical non-rebreathing valve, not an algorithm or AI system. The performance evaluated is the standalone performance of the physical device according to engineering and biocompatibility standards.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance is based on established engineering standards and biocompatibility standards. Specifically:

    • AS 4259-1995: "Ancillary devices for expired air resuscitation"
    • FDA Blue Book Memo G95-1: Guidance for biocompatibility testing.
    • ISO 10993-5: Biocompatibility testing (Cytotoxicity)
    • ISO 10993-10: Biocompatibility testing (Irritation)
    • ISO 10993-12: Biocompatibility testing (Sensitization)
    • ISTA Procedure 2A: Shipping testing.

    These standards define the acceptable performance parameters and test methods for such devices.

    8. The Sample Size for the Training Set

    This information is not applicable. The device is a mechanical non-rebreathing valve, not a machine learning model. Therefore, there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K142764

    Validate with FDA (Live)

    Date Cleared
    2015-07-09

    (287 days)

    Product Code
    Regulation Number
    868.5870
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CPR Mask with Oxygen Port is designed to assist in providing immediate life support (mouth to health emergency victims requiring oxygen support (inhalation) or cardiopulmonary resuscitation (CPR) rescue techniques. The CPR Mask with Oxygen Port is for prescription use.

    The CPR Mask without Oxygen Port is designed to assist in providing immediate life support (mouth to health emergency victims requiring cardiopulmonary resuscitation (CPR) resuscitator without oxygen port is for overthe-counter use.

    Device Description

    The KYOLING CPR mask is used during cardiopulmonary resuscitation procedures, which are used in emergency situations to supply oxygen and produce blood flow in the heart and lungs.

    The CPR mask is made of PVC and one-way valve made of medical grade K-resin, the mask is used for mouth-to-mask breathing, it provides a physical barrier between the rescuer and victim, eliminating direct contact of the rescuer's lips with the unknown subject; and also it promotes an airtight seal to the face allowing ventilation through both the mouth and nose simultaneously. The KYOLING CPR Mask includes transparent dome, universal breathing tube, one-way filtered valve, head strap, and with Oxygen Port or without Oxygen Port.

    AI/ML Overview

    The provided document describes the acceptance criteria and a study to prove that the KYOLING CPR mask with and without an oxygen port meets these criteria.

    Here's a breakdown of the requested information:

    1. A table of acceptance criteria and the reported device performance

    ParameterAcceptance CriteriaReported Device Performance (KYOLING CPR Mask with Oxygen Port)Reported Device Performance (KYOLING CPR Mask without Oxygen Port)
    BiocompatibilityMeeting ISO 10993 standards for cytotoxicity, sensitization, and irritationAll tests conducted in accordance with FDA Blue Book Memorandum #G95-1 and ISO 10993-1, with tests for cytotoxicity, sensitization, and irritation completed. Cushion and valve considered surface contacting for < 24 hours.All tests conducted in accordance with FDA Blue Book Memorandum #G95-1 and ISO 10993-1, with tests for cytotoxicity, sensitization, and irritation completed. Cushion and valve considered surface contacting for < 24 hours.
    Inspiratory Resistance< 5 cmH₂O (at 50 L/min)3.19~3.24 cmH₂O3.19~3.24 cmH₂O
    Expiratory Resistance< 5 cmH₂O (at 50 L/min)3.31~3.37 cmH₂O3.31~3.37 cmH₂O

    2. Sample size used for the test set and the data provenance

    The document does not specify a distinct "test set" sample size for the resistance testing or biocompatibility. The resistance testing likely refers to a sample of manufactured devices. The data provenance is not explicitly stated as "country of origin," but the manufacturer is Hangzhou Jinlin Medical Appliances Co., Ltd. in China, implying the testing was conducted there or arranged by them. The studies appear to be retrospective in the sense that they are conducted on the finished device to demonstrate compliance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. The ground truth for this device's performance is established by objective measurements against international standards (ISO 10993 and AS-4259-1995), not expert consensus.

    4. Adjudication method for the test set

    Not applicable. As noted above, performance is determined by objective measurements against established standards, not by human adjudication of outcomes.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a CPR mask, not an AI-powered diagnostic or interpretive tool that would involve multi-reader studies.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a physical medical device, not an algorithm.

    7. The type of ground truth used

    The ground truth used for this device's performance evaluation is based on:

    • International Standards: ISO 10993 for biocompatibility and AS-4259-1995 for inspiratory and expiratory resistance. These standards define the acceptable range for the device's physical and biological properties.

    8. The sample size for the training set

    Not applicable. This device is a physical medical device, not a machine learning model. There is no concept of a "training set" in this context.

    9. How the ground truth for the training set was established

    Not applicable. As there is no training set, there is no ground truth for a training set to be established.

    Ask a Question

    Ask a specific question about this device

    K Number
    K142402

    Validate with FDA (Live)

    Date Cleared
    2015-03-23

    (208 days)

    Product Code
    Regulation Number
    868.5870
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BigEasy™ Non-Rebreathing Valve is intended to assist in providing immediate life support (mouth to mask ventilation) to health emergency victims requiring oxygen support or cardiopulmonary resuscitation (CPR) rescue techniques.

    Device Description

    The BigEasy™ Non-Rebreathing Valve is a resuscitation valve designed for resuscitation using expired air for ventilation. It has fittings on the patient end that will adapt to standard masks (22mm ID) or endotracheal tubes (15mm OD) used for resuscitation. It provides a path for a rescuer to blow expired air into a mask or endotracheal tube through a silicone valve in the device and has a rigid one-way valve that directs exhaled air from the patient away from the rescuer. It also incorporates an oxygen inlet port on the rescuer's side of the valve for the adding of supplemental oxygen without requiring a separate adapter.

    AI/ML Overview

    The provided text describes the acceptance criteria and the study that proves the device meets those criteria for the BigEasy™ Non-Rebreathing Valve.

    Here's the breakdown of the information requested:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria CategoryAcceptance CriteriaReported Device Performance
    BiocompatibilityMeeting ISO 10993-1 requirements for Cytotoxicity, Irritation, and Sensitization.The BigEasy™ Non-Rebreathing Valve passed all required biocompatibility tests (Cytotoxicity, Irritation, Sensitization) based on ISO10993-1:2009/2010 and the 2013 FDA Guidance Document.
    Packaging and Assembly / ShippingMeeting packaging and assembly / shipping requirements.The BigEasy™ Non-Rebreathing Valve met the packaging and assembly / shipping testing acceptance criteria.
    EnvironmentalMeeting environmental requirements for storage and operation. (Specific ranges provided: Operating Temperature: -18°C to 50°C, Storage Temperature: -40°C to 60°C, Storage Relative Humidity: 10% to 95%)The BigEasy™ Non-Rebreathing Valve met the environmental requirements acceptance criteria. (Specific values: Operating Temperature: -18°C to 50°C, Storage Temperature: -40°C to 60°C, Storage Relative Humidity: 10% to 95% were also listed as specifications).
    Volatile Organic Compounds (VOCs), Ozone, CO, CO2 and Fine Particle DischargeMeeting TO-15 standards for VOCs and requirements for the discharge of other gases and particles below the standards' thresholds.The BigEasy™ Non-Rebreathing Valve met the requirements for VOCs, Ozone, CO, CO2, and Fine Particles.
    Materials SpecificationsMeeting materials specifications requirements.The BigEasy™ Non-Rebreathing Valve met the materials specifications acceptance criteria.
    Labeling VerificationMeeting labeling verification requirements.The BigEasy™ Non-Rebreathing Valve met the labeling verification acceptance criteria.
    Inlet and Outlet FittingsMeeting inlet and outlet fitting requirements (Standard 22mm ID for inlet, Standard 22mm OD/15mm ID for outlet).The BigEasy™ Non-Rebreathing Valve met the fittings acceptance criteria.
    Resistance to Inhalation and ExhalationMeeting resistance requirements after and during environmental exposure, vomitus contamination, water submersion, and mechanical displacement.The BigEasy™ Non-Rebreathing Valve met the resistance criteria. (Specific values: Inspiratory Resistance: 1.9 cm H2O at 50 LPM, Expiratory Resistance: 0.25 cm H2O at 50 LPM were also listed as specifications).
    Drop TestWithstanding a drop on a concrete floor from 1 meter.The BigEasy™ Non-Rebreathing Valve met the drop test requirements.
    Mean Concentration of Oxygen at 15 LPM and Circuit Backpressure at 30 LPMDetermining the mean oxygen concentration and backpressure during oxygen delivery.The BigEasy™ Non-Rebreathing Valve met the requirements for oxygen delivery.
    Assembly and ApplicationValidation that the device can be assembled and applied according to the Instructions For Use (IFU).Participants were able to perform the intended actions while following the IFU.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not explicitly state the sample sizes for each specific test or the data provenance (country of origin, retrospective/prospective). The studies are described as "Non-clinical test results" and "Bench Tests," which typically implies laboratory testing rather than human subject data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not provided. Given that the tests are primarily "Bench Tests" and "Biocompatibility Testing," they typically involve objective measurements against established standards, rather than expert interpretation for ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not provided. The tests appear to be objective measurements against predefined standards, thus not requiring an adjudication method by human experts.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC comparative effectiveness study was done. This device is a non-rebreathing valve, which is a physical medical device, not an AI-powered diagnostic tool. The "performance testing" focuses on physical and material properties, not diagnostic accuracy with or without AI.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    A standalone performance evaluation was done, but it was for the physical device itself, not an algorithm. The device was tested independently against various physical, material, and operational standards.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for the performance tests was based on established industry standards and regulations, such as:

    • AS 4259-1995 (Ancillary devices for expired air resuscitation)
    • ISO 13544-2:2002 (Respiratory Therapy Equipment – Part 2: Tubing and Connectors)
    • ISO 5356-1:2004 (Anaesthetic and Respiratory Equipment – Conical Connectors - Part 1: Cones and Sockets)
    • BS EN ISO 10651-4:2009 (Lung Ventilators – Part 4: Particular requirements for operator-powered non-rebreathing valves)
    • ISTA-2A:2011 (Packaged-Products weighing 150 lbs (68 kg) or Less)
    • ISO 10993-1:2009 COR 1 2010 (Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing within a Risk Management Process)
    • FDA Guidance Document (Draft) (Reviewer Guidance on Face Masks and Shield for CPR)
    • BS EN ISO 15223-1:2012 (Medical Devices – Symbols to be used with Medical Devices Labels, Labelling and Information to be supplied– Part 1: General Requirements)
    • EPA-453/R-98-008B (Method TO-15, Determination Of Volatile Organic Compounds (VOCs) In Air)

    8. The sample size for the training set

    Not applicable. This is a physical medical device, not an AI algorithm requiring a 'training set.'

    9. How the ground truth for the training set was established

    Not applicable. There is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K132143

    Validate with FDA (Live)

    Device Name
    EXHALATION VALVE
    Date Cleared
    2014-06-04

    (328 days)

    Product Code
    Regulation Number
    868.5870
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The exhalation valves are used to control Inspiratory pressure and expel the expired air from a patient being ventilated via a single limb breathing system. The exhalation valve and single limb breathing systems are used with adults/pediatrics and prescribed by a physician. A pediatric is defined as 10 kg in weight. The product is not for use in neonates. The device can be used within hospitals and for home care use. It is a single patient use device and can be used for a maximum of 30 days.

    Device Description

    Double limb breathing systems comprising of an inspiratory and expiratory limb is commonly used for respiratory care. To reduce waste and clutter single limb breathing systems can be used with an exhalation valve to dispel exhaled air from the patient. Some ventilator designs or specifications do not require proximal pressure monitoring; airway pressure is monitored inside the ventilator. There are three variants of the exhalation valve in this submission.

    Exhalation valve with proximal pressure port (1924501 &1924504): A single breathing tube connects the patient to the ventilator allowing the patient to receive respiratory care. The exhalation valve is connected at the patient end onto the breathing tube. A pressure monitor line and exhalation valve control line directly connects the valve to the ventilator. Via exhalation valve control line, the ventilator applies a pressure which controls the balloon valve. During the Inspiratory phase the balloon is inflated to close off the exhalation port in the valve body. During the expiratory phase the pressure is released in the valve chamber and the balloon deflates allowing air to be expelled to the surrounding environment via the exhalation port. This prevents the patient rebreathing exhaled gases yet allowing for a single limb breathing system to be used instead of a double limb system. The pressure line monitors the pressure in the valve body. The patient end of the exhalation valve has a swivel connector end.

    Exhalation valve without proximal proximal prossure port (1923500): A single breathing tube connects the patient to the ventilator allowing the patient to receive respiratory care. The exhalation valve is connected at the patient end onto the breathing tube. An exhalation valve control line directly connects the valve to the ventilator. Via the exhalation valve control line, the ventilator applies a pressure which controls the balloon valve. During the Inspiratory phase the balloon is inflated to close off the exhalation port in the valve body. During the expiratory phase the pressure is released in the valve chamber and the balloon deflates allowing air to be expelled to the surrounding environment via the exhalation port. This prevents the patient rebreathing exhaled gases vet allowing for a single limb breathing system to be used instead of a double limb system. The patient end of the exhalation valve has a swivel connector end.

    Different ventilators require valves which work with different pressure ratios. The MKIII exhalation valve (1923500 and 1924501) has a 1:1.5 pressure ratio which caters for the majority of ventilators. However, there are markets which have applications for ratio of 1:1, and 1:2. The MK3b exhalation valve (1924504) has a 1:2 pressure ratio. The pressure ratio is the pressure difference between the control line pressure and the pressure that will be generated on the patient side of the valve.

    AI/ML Overview

    This K132143 submission describes the Intersurgical Exhalation Valve (MK3 and MK3b models) and demonstrates its substantial equivalence to a predicate device (Intersurgical 1922500 exhalation valve) through non-clinical testing.

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission does not explicitly list "acceptance criteria" in a numerical format that would typically be seen for a device like software. Instead, it presents a comparison of key performance characteristics between the new devices and the predicate, implying that performance comparable to the predicate is the acceptance criteria for substantial equivalence. The "Acceptance Criteria" below are inferred from the predicate's performance or internal requirements.

    Characteristic ComparedAcceptance Criteria (Inferred from Predicate/Internal req.)MK3 Exhalation Valve (1923500 non-ported & 1924501 ported) PerformanceMK3b Exhalation Valve (1924504 ported) Performance
    Intended UseAdult/pediatric patients using a single limb breathing tubeAdult/pediatric patients using a single limb breathing tubeAdult/pediatric patients using a single limb breathing tube
    Duration of UseSingle patient use, up to 24 hours (Predicate)Single patient use, up to 30 daysSingle patient use, up to 30 days
    Where usedHospital and homeHospital and homeHospital and home
    Flow Resistance at 10L/min≤ 0.6 mbar0.6 mbar0.6 mbar
    Flow Resistance at 30L/min≤ 1.0 mbar0.9 mbar0.9 mbar
    Pressure Ratio1:2 (Predicate)1:1.51:2
    Leakage balloon (ml/min)< 0.5 ml/min< 0.5 ml/min0.0 ml/min
    Leakage main body (ml/min)< 3.5 ml/min (and < 5ml/min internal req.)< 0.5 ml/min1.3 ml/min
    Conditioned/Storage testingComplete and PassComplete and PassComplete and Pass
    TapersPassPassPass
    Aging (5 year)N/A (no predicate data)PassPass
    30-day testingN/A (24-hour predicate)PassPass
    Trigger featureN/A (predicate different design)Trigger by 30msSame as MK3
    BiocompatibilityISO 10993ISO 10993ISO 10993
    SterilityNon SterileNon SterileNon Sterile
    Standards MetISO 5356 (Connectors)ISO 5356 (Connectors)ISO 5356 (Connectors)

    2. Sample size used for the test set and the data provenance

    The submission describes non-clinical laboratory testing. There is no mention of a "test set" in the context of clinical data or patient samples. The tests were performed on the device prototypes themselves. The data provenance is laboratory testing conducted by Intersurgical Incorporated. There is no mention of country of origin for specific test data, but the company is based in Liverpool, NY, USA. The testing is prospective for the new devices, demonstrating their performance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This is a non-clinical device performance study based on engineering specifications and physical measurements, not a diagnostic study requiring expert interpretation of results or establishment of "ground truth" as it would pertain to medical imaging or diagnostics.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as this is a non-clinical device performance study with objective measurements, not a study requiring adjudication of expert opinions.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a medical device (exhalation valve) and not an AI or diagnostic imaging device that would typically undergo MRMC studies.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a mechanical medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For this non-clinical study, the "ground truth" is defined by established engineering principles, international standards (e.g., ISO 5356-1, ISO 10993-1), and the performance characteristics of the legally marketed predicate device. For example, resistance to flow is measured against a standard, and biocompatibility is confirmed against ISO 10993. The "performance" of the predicate device serves as a benchmark for substantial equivalence.

    8. The sample size for the training set

    Not applicable. There is no concept of a "training set" for a physical medical device submission of this nature. The devices themselves are the "samples" produced and tested.

    9. How the ground truth for the training set was established

    Not applicable. No training set as per item 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 5