Search Filters

Search Results

Found 29 results

510(k) Data Aggregation

    K Number
    K243917

    Validate with FDA (Live)

    Date Cleared
    2025-07-14

    (206 days)

    Product Code
    Regulation Number
    868.5330
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This device is intended for the delivery of high flow respiratory gases (within the limits of its stated technical specifications) to patients in a hospital by appropriately qualified healthcare professionals.

    This product is not intended to be used on patients with bypassed upper airways.

    This product is not intended to be life-sustaining or life-supporting.

    This product is not intended for apneic ventilation.

    Device Description

    The F&P Optiflow Air/Oxygen Flow Source is a respiratory high flow therapy device designed to generate a flow of air and/or oxygen. The device allows selectable flow rates up to 70L/min with selectable oxygen concentrations ranging from 21% to 100%.

    The subject device is a multi-patient use prescription only device, provided in a non-sterile state. It operates at flow ranges between 0 to 70 L/min and is intended to be used by appropriately qualified healthcare professionals in hospitals.

    AI/ML Overview

    The provided FDA 510(k) clearance letter and summary for the "F&P Optiflow Air/Oxygen Flow Source" is for a hardware device, a breathing gas mixer, not a software or AI-driven diagnostic device. Therefore, the majority of the requested information regarding acceptance criteria, study design, expert involvement, and ground truth for AI/software performance is not applicable based on the provided document.

    The document primarily focuses on demonstrating substantial equivalence to predicate devices through comparisons of technological characteristics and compliance with general medical device standards. There is no information regarding AI-specific performance metrics, clinical studies involving human readers, or detailed ground truth establishment for diagnostic capabilities.

    Below is a table summarizing the general performance data found, where applicable, according to the structure requested, along with an explanation for the absence of AI-specific information.


    Acceptance Criteria and Device Performance for F&P Optiflow Air/Oxygen Flow Source

    This device is a hardware breathing gas mixer, not an AI or software-driven diagnostic tool. As such, the typical acceptance criteria and study designs associated with AI performance (e.g., sensitivity, specificity, human reader improvement, adjudication, ground truth for AI training/testing) are not present in this 510(k) summary. The "performance data" section primarily refers to compliance with safety, electrical, and performance standards for medical electrical equipment and gas mixers.


    1. Table of Acceptance Criteria and Reported Device Performance:
    Acceptance Criteria (from recognized standards & predicate comparison)Reported Device Performance (as per 510(k) Summary)
    Flow Rate Accuracy: +/- 10%Accuracy of +/- 10% (Identical to Primary Predicate)
    Operating Temperature Range: (Comparable to predicates)64°F to 79°F (18 °C to 26 °C)
    % Oxygen Control: 21 – 100%21 – 100% (Identical to Predicates)
    % Oxygen Control Accuracy: (Based on ISO 11195:2018)+/- 5%
    Oxygen Supply Alarm: Triggers if pressure < 40psiIssues alarm if < 40psi (Equivalent to Predicate)
    FiO2 Alarm: Triggers if unable to achieve set FiO2Device alarms if unable to achieve set FiO2
    Power-fail Alarm: Audible alarm for t > 2 minutesAudible alarm for time, t > 2 minutes
    Disconnection Alarm: Activated in < 5 secondsActivated in < 5 seconds
    Obstruction Detected/Flow Alarm: Activated in < 5 secondsActivated in < 5 seconds
    General Alarm: For internal faultsActivated for internal faults
    Compliance with ANSI/AAMI ES 60601-1:2005/A2:2021Tested and confirmed compliant
    Compliance with IEC 60601-1-8:2006+AMD1:2012+AMD2:2020Tested and confirmed compliant
    Compliance with IEC 62304:2006 + A1:2015 (Software Lifecycle)Tested and confirmed compliant
    Compliance with IEC60601-1-6:2010+AMD1:2013+AMD2:2020 (Usability)Tested and confirmed compliant
    Compliance with ISO 17664-2:2021 (Processing of healthcare products)Tested and confirmed compliant
    Compliance with IEC 62366-1:2015 + A1:2020 (Usability Engineering)Tested and confirmed compliant
    Compliance with AIM 7351731 Rev 3:2021 (EMC for RFID Readers)Tested and confirmed compliant
    Compliance with ISO 80601-2-90:2021 (Respiratory high-flow equipment)Tested and confirmed compliant
    Compliance with ISO 11195:2018 (Gas mixers for medical use)Tested and confirmed compliant (specifically for % O2 accuracy)
    Compliance with ANSI/AAMI ST98:2022 (Cleaning validation)Tested and confirmed compliant

    The remaining requested information is not applicable to this 510(k) submission as it pertains to an AI/software-driven diagnostic device, which the "F&P Optiflow Air/Oxygen Flow Source" is not. The document describes a physical medical device (breathing gas mixer) with its performance verified through non-clinical bench testing and compliance with relevant safety and performance standards for such equipment.

    1. Sample size used for the test set and the data provenance: Not applicable. The document describes non-clinical performance testing against engineering specifications and relevant standards for a hardware device, not a test set of data for AI.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for diagnostic AI is not relevant for this hardware device.
    3. Adjudication method: Not applicable.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable. The "ground truth" for this device's performance would be the specified output values (e.g., flow rate, oxygen concentration) measured by calibrated equipment during engineering and validation testing, as required by the listed standards.
    7. The sample size for the training set: Not applicable. This is not an AI/ML device.
    8. How the ground truth for the training set was established: Not applicable. This is not an AI/ML device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K241465

    Validate with FDA (Live)

    Date Cleared
    2024-09-19

    (119 days)

    Product Code
    Regulation Number
    868.5330
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Midas Flowmeter is intended for use as a continuous flow conscious sedation system to deliver a mixture of nitrous oxide and oxygen gases to a patient. When used with the electronic automatic vacuum switch (eAVS), the Midas Flowmeter is used to control the scavenging of waste analgesic gas.

    Device Description

    The Midas Flowmeter is used within a healthcare environment during a nitrous oxide (N20)/oxygen (O2) conscious sedation system. The device is a digitally controlled, software driven, continuous flow device that precisely meters nitrous oxide and oxygen medical gases and delivers gas mixtures to a conscious, spontaneously breathing patient. The Midas Flowmeter uses electronic mechanisms to control the flowrate and mixture percentage of gases delivered to a patient. The optional eAVS is an accessory to the Midas Flowmeter. It is used within the nitrous oxide/oxygen conscious sedation system to allow removal of waste analgesic gases through a connected vacuum source. The eAVS connects the exhalation line of the patient's breathing circuit to vacuum tubing from the vacuum source and controls the vacuum flowrate (i.e., scavenging flowrate). The eAVS is specifically designed to work with the Midas Flowmeter and all functionality control of the eAVS has been integrated into the user interface of the Midas Flowmeter.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification from the FDA, which focuses on demonstrating substantial equivalence of a new medical device to existing predicate devices. It primarily discusses comparisons of technical characteristics and compliance with recognized standards.

    Crucially, this document does NOT contain explicit acceptance criteria tables or a detailed study report that proves the device meets specific performance metrics in a clinical or simulated environment with ground truth from experts, as would be expected for an AI/ML powered device.

    The document states:

    • "The characteristics of the Midas Flowmeter and optional eAVS do not require clinical investigation due to safety and efficacy being supported by non-clinical testing performed."
    • "The verification and validation testing of the Midas Flowmeter and optional eAVS was found to be acceptable and supports the claims of substantial equivalence."

    This indicates that the device's performance was evaluated through bench testing and compliance with recognized consensus standards, rather than a clinical study involving human readers or expert-established ground truth for a test set.

    Therefore, I cannot provide the requested information about acceptance criteria, study details, sample sizes for test/training sets, expert qualifications, adjudication methods, MRMC studies, or the type of ground truth used, because such a study (as described in the prompt) was not performed or detailed in this 510(k) submission.

    The document focuses on:

    • Comparison of Technical Characteristics: Showing the Midas Flowmeter and eAVS are similar in function, intended use, and technology to predicate devices (FlowStar Touch Digital Mixer Flowmeter and Nitronox Scavenger Plus).
    • Compliance with Standards: Listing various IEC, ISO, AAMI, and ASTM standards that the device was tested against to demonstrate safety and performance (e.g., IEC 60601-1 for basic safety, ISO 11195 for gas mixers).
    • Non-clinical Performance Data (Bench Testing): A bulleted list of "performance bench testing" categories like "Gas Flow Rate Adjustment," "Touchscreen Display," "Accuracies and Faults," and "Life Testing." However, it does not provide the specific acceptance criteria for these tests nor the detailed results in a quantitative manner as requested. It simply states that "The results of the testing demonstrated that the Midas Flowmeter and optional eAVS met all of the acceptance criteria for functional, operational, and performance characteristics..."

    In summary, none of the specific elements required to answer your prompt (e.g., table of acceptance criteria with reported performance, sample sizes for test sets, expert ground truth establishment, MRMC studies, standalone performance, training set details) are present in this 510(k) document because the device is a hardware system (a gas mixer) and was cleared based on non-clinical bench testing and adherence to recognized standards, not clinical performance studies with AI/ML components involving human expert review or data-driven ground truth.

    Ask a Question

    Ask a specific question about this device

    K Number
    K231895

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-12-04

    (159 days)

    Product Code
    Regulation Number
    868.5330
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Maxtec MaxBlend 2+p is intended to provide a continuous air/oxygen gas mixture and to continuously monitor the concentration of oxygen and pressure being delivered to infant, pediatric, and adult patients. It is a restricted medical device intended for use by qualified, trained personnel, under the direction of a physician, in professional healthcare settings, i.e., hospital, subacute, and nursing-care facilities where the delivery and monitoring of air/oxygen mixtures is required. This is not intended as a life-supporting device or life sustaining device.

    Device Description

    The Maxtec MaxBlend 2+p is an oxygen delivery device which incorporates an air/oxygen blender, battery powered oxygen and pressure monitor, and an adjustable flowmeter, all in a single assembly. The integral air/oxygen blender provides precise mixing of medical grade air and oxygen. The flowmeter provides control of the flow rate delivered. The oxygen monitor measures the oxygen concentration from the blender's gas flow, displays these measured concentrations, and provides user selectable high and low oxygen alarms. It also allows the user to monitor pressure simultaneously using adjustable high and low alarm limits.

    AI/ML Overview

    The document describes the Maxtec MaxBlend 2+p, a medical device combining an air/oxygen blender, oxygen monitor, pressure monitor, and flowmeter. The 510(k) submission seeks substantial equivalence to existing predicate devices (Maxtec MaxBlend 2 - K161718 and MaxO2ME+p - K221734), particularly highlighting the addition of a pressure monitoring feature.

    Here's an analysis of the acceptance criteria and the study information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for several performance aspects are implicitly derived from the comparative tables (Table 1 and 2) where the subject device's specifications are listed and compared to the predicate devices. The "Reported Device Performance" for the subject device is simply its stated characteristics, which are claimed to meet or be identical to the predicate device specifications, thereby meeting the acceptance criteria based on substantial equivalence.

    Feature / Performance AspectAcceptance Criteria (from predicate devices)Reported Device Performance (Maxtec MaxBlend 2+p)
    Air/Oxygen Mixer Features
    Gas Supply TypeAir / OxygenAir / Oxygen
    Pressure30 to 75 psi30 to 75 psi
    Mixed gas stability± 1% oxygen± 1% oxygen
    Flow range of BlendersLow flow model – 0-30 Lpm; High flow model – 0-100 LpmLow flow model – 0-30 Lpm; High flow model – 0-70 Lpm (Primary Predicate) / 0-100 Lpm (Secondary Predicate)
    Pressure supply differential alarmAir / oxygen pressure < 20 psi an alarm soundsAir / oxygen pressure < 20 psi an alarm sounds
    Pressure Drop< 6 psi @ 50 psi< 6 psi @ 50 psi
    Bleed flow3-13 Lpm at 50 psi depending upon model3-13 Lpm at 50 psi depending upon model
    Oxygen Monitor Features
    Oxygen measurement range0-100%0-100%
    Total Accuracy±3% Actual oxygen level over full operating temperature range±3% Actual oxygen level over full operating temperature range
    Response Time90% of final value in approx. 15 seconds at 23°C90% of final value in approx. 15 seconds at 23°C
    Warm-up TimeNone requiredNone required
    High alarm range16 – 100%16 – 100%
    Operating Temperature15°C – 40°C (59°F – 104°F)15°C – 40°C (59°F – 104°F)
    Storage Temperature-15°C – 50°C (5°F – 122°F)-15°C – 50°C (5°F – 122°F)
    Humidity0-95% (non-condensing)0-95% (non-condensing)
    Expected use-life of sensor1,500,000 O2% hours (approx. 2 years)1,500,000 O2% hours (approx. 2 years)
    Flowmeter Features
    Accuracy± 10% of indicated value when inlet pressure is 50 psi± 10% of indicated value when inlet pressure is 50 psi
    Flow meter ranges0 – 3 Lpm, 0 – 15 Lpm, 0 – 30 Lpm, 0 – 70 Lpm (some predicates mention 0-100Lpm as well)0 – 3 Lpm, 0 – 15 Lpm, 0 – 30 Lpm, 0 – 70 Lpm
    Pressure Monitor Features(Based on Secondary Predicate MaxO2 ME+p K221734)(Maxtec MaxBlend 2+p)
    TechnologyMicroprocessor controlled deviceMicroprocessor controlled device
    Pressure sensor typeSolid-state pressure transducerSolid-state pressure transducer
    Pressure Measurement Range-15 to +60 cmH2O-15 to +60 cmH2O
    Pressure Resolution1 cmH2O1 cmH2O
    Display resolution0.5 cm H2O0.5 cm H2O
    Total Accuracy± 1 cmH2O± 1 cmH2O
    Low alarm range1 - 30 cmH2O1 - 30 cmH2O
    High alarm range1 - 60 cmH2O1 - 60 cmH2O
    Alarm delay3 seconds (pressure only)3 seconds (pressure only)
    Zero calibrationYesYes
    Operating Temperature15°C - 40°C, 0-95% RH15°C - 40°C, 0-95% RH
    Storage Temperature-15°C - 50°C @ 95% RH-15°C - 50°C @ 95% RH
    Atmospheric Pressure800 – 1013 mBars800 – 1013 mBars
    Power requirements4x - AA alkaline batteries4x - AA alkaline batteries
    Battery Life5000 hours5000 hours

    2. Sample Size Used for the Test Set and Data Provenance

    The document explicitly states:

    • Non-clinical Testing: "We performed a number of tests to demonstrate that the proposed device performed as intended." It then lists several international standards that the device met. It does not specify a sample size for the test set or the data provenance (e.g., country of origin, retrospective/prospective). The testing appears to be primarily related to engineering and performance validation against standards rather than clinical data.
    • Clinical Testing: "No human clinical testing was performed."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Since "No human clinical testing was performed," and the primary testing involved adherence to international standards and non-clinical performance, there is no mention of experts establishing ground truth for a clinical test set. The "ground truth" for the non-clinical testing would be the defined requirements and specifications within the cited standards.

    4. Adjudication Method for the Test Set

    Not applicable, as no human clinical test set is mentioned, and the testing focuses on compliance with technical standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a medical gas mixer and monitor, not an AI-assisted diagnostic or interpretation tool that would involve human readers or image analysis.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The device is a hardware product with integrated monitoring capabilities. Its "standalone" performance is measured through its adherence to the technical specifications and international standards listed under "Non-clinical Testing." The document states: "In all cases the subject device met the standard requirements." This implicitly confirms standalone performance against established benchmarks.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the performance specifications (e.g., oxygen accuracy, pressure accuracy, flow rates), the ground truth is based on metrological standards and calibrated equipment used in accordance with the listed international standards (e.g., ISO 80601-2-55, ISO 11195). For biocompatibility, the ground truth is established by adherence to ISO 10993 and ISO 18562 standards.

    8. The Sample Size for the Training Set

    Not applicable. This is a hardware medical device with integrated sensors and controls, not a machine learning or AI algorithm that requires a "training set."

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as there is no training set for an AI algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K230987

    Validate with FDA (Live)

    Date Cleared
    2023-09-14

    (161 days)

    Product Code
    Regulation Number
    868.5330
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Indicated for administering an adjustable mixture of Nitrous Oxide analgesic gas and Oxygen to a conscious, spontaneously breathing patient

    Device Description

    This is a device which can precisely dose a mixture of medical oxygen and medical nitrous oxide, depending on the adjusted mixture, for the sedation of patients in hospitals and at the dentist. The operator, a dentist is able to adjust the total flow and the concentration of nitrous oxide of this flow. Depending on these settings the level of sedation of the patient can be controlled by the operator. The FlowStar ® Analog Mixer device has been created with the following safety features. The total flow emitted can consist of a maximum of 70 % nitrous oxide, the mix up of gas types is prevented by mechanical encodings and the flow of pure nitrous oxide in the absence of a missing oxygen supply is prevented by a fail-safe device.

    Two rotary knobs are used to select the desired total flow and N20/O2 gas mixture. The total flow can consist of between 0 (100% O₂) and 70 (30% O₂) percent N₂O. The minimum output of 30 % O2 prevents from a hypoxic mixture for the patient. The FlowStar® Analog Mixer device has an integrated O2 Flush knob which delivers at least over 40 l/min of pure oxygen if the user presses the knob to flush the N20 out of the patient circuit after the treatment or in the situation the patient is over sedated.

    The FlowStar Analog Mixer device has an alarm system which generates an alarm in the case of an insufficient O2 supply. The device will shut off the sedation when the O2 supply reaches a critical level.

    The FlowStar ® Analog Mixer has an integrated Bag T. The mixed gas to it. The Bag T is a device with one additional input and two outputs. The additional input is a valve which opens when the adjusted total flow is too low for the patient. The gas mixture is then enriched with additional ambient air. The first output of the Bag T is for the connected breathing bag which is filled by the mixing device when the patient exhales and is emptied when the patient inhales. The second output leads the gas mixture to the connected scavenger system and finally to the patient

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification for a medical device called "FlowStar Analog Mixer Flowmeter." It focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study proving that the device meets specific performance acceptance criteria for an AI/ML-based algorithm.

    Therefore, the requested information regarding acceptance criteria, device performance from a test set, sample sizes, expert involvement, adjudication methods, MRMC studies, standalone performance, ground truth establishment, and training set details cannot be extracted from this document as it does not describe a clinical performance study of an AI/ML algorithm.

    The document discusses:

    • Device Description: A breathing gas mixer for administering Nitrous Oxide and Oxygen.
    • Indications for Use: Administering an adjustable mixture of Nitrous Oxide and Oxygen to a conscious, spontaneously breathing patient.
    • Safety Features: O2 fail-safe system, alarm for insufficient O2 supply, O2 flush knob, mechanical encodings to prevent gas mix-up.
    • Non-Clinical Testing: Compliance with standards like ISO 18562 (biocompatibility), ISO 14971 (risk management), ISO 15223-1 (labeling symbols), CGA V-5 (connectors), ISO 11195 (gas mixers), and IEC 62366-1 (usability testing). This mainly involves bench testing to ensure the physical and functional aspects of the device meet safety and performance standards.
    • Clinical Testing: The document explicitly states, "Clinical testing was not required to establish substantial equivalence."

    In summary, this document is a regulatory submission for a physical medical device (a gas mixer), not a clinical performance study for an AI/ML algorithm. As such, the specific metrics and study design elements requested in your prompt (e.g., acceptance criteria for an AI model's accuracy, expert reads, ground truth for data sets) are not relevant to this type of device and its regulatory clearance process as described here.

    Ask a Question

    Ask a specific question about this device

    K Number
    K222794

    Validate with FDA (Live)

    Date Cleared
    2023-07-28

    (315 days)

    Product Code
    Regulation Number
    868.5330
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Indicated for administering an adjustable mixture of Nitrous Oxide analgesic gas and Oxygen to a conscious, spontaneously breathing patient.

    Device Description

    The device will be available in two models: (Details below) FLOWSTAR TOUCH DIGITAL MIXER FLOWMETER 50 %, and FLOWSTAR TOUCH DIGITAL MIXER FLOWMETER 70 %.

    50% and 70% represent the maximum nitrous oxide concentration. The FlowStar Touch O2- N₂O Mixer for Analgesia is a device which can precisely dose a mixture of medical oxygen and medical nitrous oxide, depending on the adjusted mixture, for the sedation of patients in hospitals and at the dentist. The operator, a doctor or a dentist is able to adjust the total flow and the concentration of nitrous oxide of this flow. Depending on these settings the level of sedation of the patient can be controlled by the operator. The Baldus® Touch O2- N2O Mixer for Analgesia device has been created with the following safety features. The total flow emitted can consist of a maximum of 70 % nitrous oxide, the mix up of gas types is prevented by mechanical encodings and the flow of pure nitrous oxide in the absence of a missing oxygen supply is prevented by a fail-safe device. An integrated Touch pad is used to select the desired N₂O /O2 gas mixture. The total flow can consist of between 0 (100% O2) and 70 (30% O2) percent N₂O. The minimum output of 30 % O2 prevents from a hypoxic mixture for the patient. The FlowStar TouchO2- N2O Mixer for Analgesia device has an integrated O2 Flush function which delivers at least 30 l/min of pure oxygen to flush the N20 out of the patient circuit after the treatment. The FlowStar TouchO2- N₂O Mixer for Analgesia device has an alarm system which generates an alarm in the case of an insufficient O2 supply. The device will shut off the sedation when the O2 supply reaches a critical level. The gas mixture output of the FlowStar TouchO2- N2O Mixer for Analgesia device is connected to the Bag T. The Bag T is a device with two inputs and two outputs. One input is the gas mixture from the mixing device. The other input is a valve which opens when the adjusted total flow is too low for the patient. The gas mixture is then enriched with additional ambient air. The first output of the Bag T is for the connected breathing bag which is filled when the patient exhales and is emptied when the patient inhales. The second output leads the gas mixture to the scavenger system and finally to the patient. The main accessories for the proposed device are a Breathing Bag and a Double Mask Scavenger System.

    The device includes a microprocessor-controlled valve system that precisely meters Oxygen and Nitrous Oxide gases for analgesic purposes. The inputs of the touchpad influence the gas flow inside of the device. The firmware is considered to be of a moderate level of concern.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and supporting studies for the FlowStar Touch Digital Mixer Flowmeter:

    1. Table of Acceptance Criteria & Reported Device Performance:

    The document doesn't explicitly state "acceptance criteria" in a tabulated format with specific numerical targets. Instead, it focuses on demonstrating substantial equivalence to a predicate device (Accutron Digital Ultra K052335) through comparisons of characteristics and compliance with recognized standards. The "performance" is implicitly demonstrated by meeting the standards and showing similar or improved characteristics compared to the predicate.

    Below is a table summarizing the characteristics where performance could be inferred or compared:

    CharacteristicAcceptance Criteria (Inferred from predicate/standards)Reported Device Performance (FlowStar Touch Digital Mixer Flowmeter)
    Intended UseAdminister adjustable mixture of Nitrous Oxide and Oxygen to a conscious, spontaneously breathing patient.Indicated for administering an adjustable mixture of Nitrous Oxide analgesic gas and Oxygen to a conscious, spontaneously breathing patient. (Functionally identical)
    Max N2O Concentration70% (for 70% model)70% (FlowStar Touch Digital Mixer Flowmeter 70%)
    Min O2 Concentration30%30% (Prevents hypoxic mixture for patient)
    Oxygen FlushMinimum 20 LPM40-55 l/min (Improved/Higher)
    Oxygen Flow1.0 LPM - 9.9 LPM1.0 - 18 l/min (Wider Range/Improved)
    Oxygen Resuscitator FlowMinimum 100 LPM100 - 250 l/min (Wider Range/Improved)
    Nitrous Oxide Flow0 - 6.9 LPM0 - 12.6 l/min (Wider Range/Improved)
    Mixed Gas Flow1.0 - 9.9 LPM3 – 18 l/min (Wider Range/Improved)
    Power Source100-240 V AC, 50/60 Hz110-230 V 50-60 Hz (Essentially the same)
    ControlsTouch PanelTouch Panel (Same)
    Safety Standards ComplianceIEC 60601-1, IEC 60601-1-2IEC 60601-1, IEC 60601-1-2 (SAME), ISO 11195, CGA V-5, CGA C-9, ISO 5356-1
    Alarms (O2 Fail-Safe)O₂ ceases if O₂ supply interrupted/reduced; alarm sounds, N₂O flow ceases.If no oxygen supplied, visual/acoustic signals emitted. Unit shuts off sedation if O2 reaches critical level. (Equivalent)
    Alarms (N2O Supply)Alarm if N₂O supply cannot keep up; alarm sounds, flow switches to 100% O2.If no nitrous oxide supplied, visual/acoustic signals emitted (including low pressure or measurement failure). (Equivalent)
    Alarms (Mixed Gas Flow)Alarm if mixed gas line pressure obstructed.O₂ minimum flow 3 l/min (visual/acoustic signals if < 3l/min. Mixture never contains < 1L O2). Max flow reached (visual/acoustic signal). Flush flow low message if < 18l/min during O2 flush. (Equivalent/Similar)
    Dimensions/Weight7.5" x 7" x 8.75", 6.3 lbs5.3" x 11.8" x 4.75", 5.5 lbs (Similar Size and Weight)
    Gas Inlet ConnectorsDISS connectorDISS connector (SAME)
    External Interface ConnectorType DB-9 Serial Output for printerUSB Type A for memory sticks (More Convenient)
    Use EnvironmentHospital, Dental Facility, Healthcare FacilityHospital, Dental Facility, Healthcare Facility (SAME)
    BiocompatibilityImplicitly safe for breathing gas pathwaysISO 18562-2 and ISO 18562-3 compliance demonstrated for VOCs and particulate matter.
    CybersecurityImplicitly secure given predicateCompliance with FDA guidance "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices" (addressed via USB port).
    Other Bench TestingFunctionality as intended, durability, safety features.Alarms verified, silicon sealing, ambient air valve performance, pressure differential, marking durability, normal operating conditions, reverse gas flow, ambient temperature range, accessory compatibility, on demand flow, high altitude test, N2O room contamination, O2/N2O empty tests. (All successful)

    2. Sample Size Used for the Test Set and Data Provenance:

    The document primarily discusses non-clinical bench testing and compliance with recognized standards. It does not mention any specific "test set" in terms of patient data or clinical samples. The testing for safety and performance appears to have been conducted on the device itself (hardware and firmware).

    Therefore, information on:

    • Sample size for test set: Not applicable as no human- or patient-based test set is described.
    • Data provenance: Not applicable.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    As there is no mention of a human- or patient-based "test set" requiring ground truth establishment, this information is not applicable to the provided document. The ground truth for engineering characteristics would be defined by the specifications of the device and the requirements of the standards.

    4. Adjudication Method for the Test Set:

    Since no human- or patient-based "test set" is described, an adjudication method for such a dataset is not applicable.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

    No. The document explicitly states: "Clinical testing was not required to establish substantial equivalence." Therefore, an MRMC study comparing human readers with and without AI assistance was not performed, nor is it relevant given the device type (a gas mixer flowmeter, not an AI diagnostic tool).

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done:

    This question is not applicable as the device is a medical device (a gas mixer flowmeter), not an AI algorithm. Its performance is inherent in its hardware and software functionality, not as an algorithm independently diagnosing or interpreting data. The "algorithm" refers to the firmware controlling the valve system. Its performance was verified through bench testing.

    7. The Type of Ground Truth Used:

    For the non-clinical testing, the "ground truth" was established by:

    • Engineering specifications and design requirements: How the device should function according to its design.
    • Recognized industry standards: Such as IEC 60601-1, IEC 60601-1-2, ISO 11195, CGA V-5, CGA C-9, ISO 5356-1, ISO 18562-2, ISO 18562-3. Compliance with these standards serves as the benchmark for acceptable performance and safety.
    • Predicate device characteristics: The functionalities and safety features of the legally marketed predicate device served as the primary comparative "ground truth" to establish substantial equivalence.

    8. The Sample Size for the Training Set:

    This information is not applicable in the context of this device. The device does not employ machine learning or AI that requires a "training set" of data in the typical sense (e.g., for image recognition or predictive models). The "firmware" is mentioned as being of moderate concern and was likely developed through standard software engineering practices, not through a data-driven training process.

    9. How the Ground Truth for the Training Set Was Established:

    Since there is no "training set" in the context of this device's type (a gas mixer flowmeter), this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K221494

    Validate with FDA (Live)

    Date Cleared
    2023-01-23

    (245 days)

    Product Code
    Regulation Number
    868.5330
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The air-oxygen mixer Quality Mix is used to administer a continuous, precise mix of medical air and medical oxygen through its outlet ports - to infants, children and adults. The exact fractional inspiratory oxygen concentration (FiO2) corresponds to the selected FiO2 setting on the control knob (dial).

    It is a restricted medical device intended for use by qualified, trained personnel under the direction of a physician in professional healthcare settings.

    This product is not intended for use as a life supporting device.

    Warning: By using flows above 70 LPM, the risk of high flow such as ar trapping, barotrauma, and gastric insufflation (leading to possible aspiration) can occur.

    Device Description

    The air-oxygen blender Quality Mix is a compact air/oxygen mixing device, which provides a precise mixing of medical grade air and oxygen gases. The blender is a pneumatic powered device with flowmeters which can be connected to outlet ports to regulate the gas flow.

    The Quality Mix series medical air / oxygen blender includes following models: Quality Mix High Flow; Quality Mix High Flow XL and Quality Mix Low Flow; Quality Mix Low Flow XL.

    The Quality Mix series air / oxygen blender is designed, tested and manufactured in accordance with the standard ISO 11195:2018 Gas mixers for medical use - Stand - alone qas mixers.

    AI/ML Overview

    The Acceptance Criteria and the study proving the device meets them are detailed below, based on the provided text.

    1. Table of Acceptance Criteria and Reported Device Performance

    The device is the Quality Mix Blender, Oxymixer. The acceptance criteria are largely derived from compliance with the ISO 11195:2018 standard for gas mixers for medical use and are demonstrated through performance specified in comparison to the predicate device (Precision Blender K053232).

    Acceptance Criteria (from ISO 11195:2018 and performance specifications)Reported Device Performance (Quality Mix Blender)Result
    Essential Performance Requirements (Clause 4 of ISO 11195:2018)Fulfils Clause 4 essential performance requirementsPass
    Clause 9: Reverse gas flowPerforms as intendedPass
    Clause 11: Alarm systemsPerforms as intendedPass
    Clause 12: Accuracy of indicated oxygen concentration± 3%Pass
    Clause 13: Gas supply failurePerforms as intendedPass
    Primary Outlet Flow RangeHigh Flow: 15 – 120 LPM (at 50 psi)Pass
    Low Flow: 3 – 30 LPM (at 50 psi)Pass
    Auxiliary Outlet Flow Range (Bleed Closed)High Flow: 15 – 120 LPM (at 50 psi)Pass
    Low Flow: 3 – 30 LPM (at 50 psi)Pass
    Auxiliary Outlet Flow Range (Bleed Open)High Flow: 0 – 105 LPM (at 50 psi, with ±3% accuracy for low flows)Pass
    Low Flow: 0 – 27 LPM (at 50 psi, with ±3% accuracy for low flows)Pass
    Bleed FlowHigh Flow: ≤ 13 LPM at 50 psiPass
    Low Flow: ≤ 3 LPM at 50 psiPass
    Maximum Combined Flow (all outlets)High Flow: ≥ 120 LPMPass
    Low Flow: ≥ 30 LPMPass
    Bypass Flow (loss of air or oxygen supply)High Flow: > 85 LPMPass
    Low Flow: > 15 LPMPass
    Bypass Alarm Activation (pressure difference)Activates when pressure difference ≥ 13.05 psi (0.9 bar)Pass
    Alarm ResetResets when pressure difference ≤ 4.35 psi (0.3 bar)Pass
    Alarm Sound Level≥ to 80 dB at 1 ft (0.3 m)Pass
    Oxygen Concentration Adjustment Range21 – 100 %Pass
    Gas Supply Pressure46.41 – 94.27 psi (3.2 – 6.5 bar)Pass
    Air and oxygen pressure differentialShould be within max 0.7 barPass
    Mixed Gas Stability± 1% OxygenPass
    Operating Temperature Range41°F to 122°F (5°C to 50°C)Pass
    Pressure DropHigh Flow: ≤ 0.21 bar at 30 LPM and 60% FiO2 (inlet 3.2-6.5 bar)Pass
    Low Flow: ≤ 0.14 bar at 10 LPM and 60% FiO2 (inlet 3.2-6.5 bar)Pass
    Biocompatibility (ISO 18562-1, -2, -3)Pass for particulate matter and VOC emissionsPass
    Risk Management (ISO 14971)Applied for hazard identification, risk control, and monitoringPass
    Usability (IEC 62366-1)Verified and validatedPass
    Environmental Testing (Strength, Vibration, Shock)Passed EN 1789, EN 60068-2-6, EN 60068-2-27, EN 60068-2-64Pass
    Flow-metering devices (DIN EN ISO 15002)Flowmeter accuracy, leak test, environmental, and mechanical condition metPass
    Oxygen Compatibility (ISO 15001)Ensured oxygen compatibility of used materialsPass

    2. Sample Size Used for the Test Set and Data Provenance

    The provided text describes non-clinical performance and safety testing. There is no mention of a "test set" in the context of patient data or clinical samples for the device itself. The testing was performed on the device units in a laboratory setting. No specific sample size (i.e., number of devices tested) is explicitly stated for each test, but it is implied that testing was conducted on representative units of the Quality Mix Blender series. The provenance of this data is from the manufacturer's internal testing as well as cooperation with an authorized, certified competence center for medical oxygen blenders for usability testing (likely in Germany, as the manufacturer is based in Germany). This is a retrospective analysis of device performance against established standards.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This information is not applicable as the study described is non-clinical performance testing of a medical device against engineering standards, not a diagnostic or AI-assisted clinical study requiring expert ground truth in the traditional sense. The "ground truth" for these tests are the objective measurements and specifications outlined in the referenced international standards (e.g., ISO 11195:2018, IEC 62366-1, ISO 14971).

    4. Adjudication Method for the Test Set

    Not applicable. This was non-clinical device performance and safety testing against established engineering and performance standards, not a clinical study involving human interpretation or adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    No, an MRMC comparative effectiveness study was not done. This submission is for a breathing gas mixer, which is a hardware device (Quality Mix Blender), not an AI-powered diagnostic or assistive tool. Therefore, the concept of human readers improving with AI assistance is not relevant to this submission.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Not applicable. As stated above, this is a hardware device, not an algorithm or AI-powered system.

    7. The Type of Ground Truth Used

    The ground truth used for these evaluations is based on International Standards and established engineering specifications. These include:

    • ISO 11195:2018 (Gas mixers for medical use - Stand-alone gas mixers)
    • ISO 14971:2007 (Medical devices - Application of risk management to medical devices)
    • IEC 62366-1:2015 (Medical devices - Application of usability engineering to medical devices)
    • EN 1789:2007+A2:2014 (Medical vehicles and their equipment - Road ambulances)
    • EN 60068-2-6:2008 (Environmental testing - Part 2-6: Tests - Test Fc: Vibration (sinusoidal))
    • EN 60068-2-27:2009 (Environmental testing - Part 2-27: Tests - Test Ea and guidance: Shock)
    • EN 60068-2-64:2008 (Environmental testing – Part 2-64: Tests – Test Fh: Vibration, broadband random and guidance)
    • DIN EN ISO 15002:2020-05 (Flow-metering devices for connection to terminal units of medical gas pipeline systems)
    • ISO 15001:2010 (Anesthetic and respiratory equipment - Compatibility with oxygen)
    • ISO 18562-1, -2, -3 (Biocompatibility evaluation of breathing gas pathways in healthcare applications)

    These standards define the performance metrics, test methods, and acceptable ranges for the device's functionality, safety, and compatibility.

    8. The Sample Size for the Training Set

    Not applicable. There is no "training set" as this is a hardware medical device, not a machine learning or AI algorithm.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K202480

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-04-22

    (237 days)

    Product Code
    Regulation Number
    868.5330
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Nitronox Plus is intended to provide a mixture of N20/02, on demand, to a conscious, spontaneously breathing patient. The Nitronox Plus is designed for use with adult and pediatric patients and is not intended to be used with infants or neonates.

    Device Description

    The Nitronox Plus is a portable "on demand" analgesia system used to deliver a mixture of nitrous oxide and oxygen gases to a conscious, spontaneously breathing patient. The device is designed for use with adult and pediatric patients and is not intended to be used with infants or neonates. It is only for use in a healthcare setting by healthcare professionals. The Nitronox Plus is available in three models which differ only in adjustability and maximum ratio of nitrous oxide that can be delivered. The Nitronox Plus features a 0-70% N2O adjustable model, 0-50% N2O adjustable model, and 50% N2O/50% O2 fixed model.

    The Nitronox Plus is a pneumatic system. Oxygen and nitrous oxide are supplied to the device through flexible hoses. Inlet gas is regulated and equalized through the use of an oxygen piloted dual diaphragm regulator. The inlet pressure of both gases and the mixed gas pressure are displayed on the front panel of the device through the use of pressure gauges. The device is designed to function with a medical breathing circuit and face mask or mouthpiece accessory.

    Delivery of mixed gas starts with a patient's demand; upon inhalation through a face mask or mouthpiece, a demand valve opens and supplies mixed gas to the patient through the connected Medical Breathing Circuit. In adjustable models, the healthcare professional is able to set the desired concentration through the use of a control knob on the front panel. Mixed gas is supplied through the breathing circuit until inhalation ceases.

    The Nitronox Plus contains several safety features. Pneumatically powered alarms for both gas inlets alert the user when either gas supply begins to deplete. The device will also maintain delivery of the set concentration with depleting oxygen pressure. Once the oxygen pressure drops below a certain threshold, the device will allow ambient air to be delivered through an Emergency Air Intake valve. The device also features a key lock to control the nitrous oxide gas supply, allowing healthcare professionals to prevent unauthorized use of the device.

    AI/ML Overview

    The provided text describes the safety, operational, and performance requirements for the Nitronox Plus breathing gas mixer, but it does not provide a specific table of acceptance criteria or detailed results of a study proving the device meets those criteria, especially at the level of detail requested for AI/diagnostic studies. Instead, it broadly states that performance testing was conducted and demonstrates the device meets "all design input requirements."

    Therefore, many of the requested details cannot be extracted from this document, as it focuses on establishing substantial equivalence to a predicate device rather than presenting a detailed performance study with specific metrics, sample sizes, and ground truth methodologies relevant to AI or diagnostic devices.

    However, I can provide what can be inferred from the text based on its content about non-clinical performance testing.

    1. A table of acceptance criteria and the reported device performance

    The document does not provide a specific table of quantitative acceptance criteria or detailed reported device performance values. It generally states that "performance testing was conducted to verify performance characteristics, safety features, capabilities compared to the predicate device, and biocompatibility." The "reported device performance" is essentially that the device "meets all design input requirements, including functional, operational, and performance requirements."

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not explicitly provided in the document. The testing described is "non-clinical performance testing," which typically involves bench testing, engineering verification, and validation, rather than clinical studies with human subjects or data sets.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable and is not provided. The "ground truth" in this context would likely refer to engineering specifications and measurements, not expert consensus on diagnostic images.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable and is not provided. Adjudication methods are typically used in clinical trials or studies involving expert review, which is not the type of testing described here.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No such study was conducted or is mentioned in the document. This is a breathing gas mixer, not an AI or diagnostic imaging device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable as the device is a physical breathing gas mixer, not a software algorithm. Its "performance" is inherently standalone in that it functions according to its design specifications.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for this device would be established through engineering specifications, validated test methods, and measurement standards for gas mixtures, pressure regulation, alarm functionality, and other specified device characteristics. The document implies that these standards were used in the "performance testing."

    8. The sample size for the training set

    This information is not applicable as the device is not an AI model requiring a training set.

    9. How the ground truth for the training set was established

    This information is not applicable as the device is not an AI model.


    Summary of available information regarding the acceptance criteria and study:

    The document states that "performance testing was conducted to verify performance characteristics, safety features, capabilities compared to the predicate device, and biocompatibility; validate reprocessing activities; and ensure usability of the device to meet patient needs when used as intended."

    The study proving the device meets its acceptance criteria is referred to as "non-clinical performance testing."

    • Acceptance Criteria (Implied): The device "meets all design input requirements, including functional, operational, and performance requirements." These requirements would likely include specifications for N2O/O2 mixture accuracy, pressure regulation (45-60 psi input, low pressure mixed gas output), alarm activation thresholds (e.g., N2O/O2 input below 35 psi), demand valve responsiveness, and the proper functioning of safety features like the O2 piloted system, DISS fittings, and the Emergency Air Intake Valve.
    • Reported Device Performance: "The results of performance testing demonstrate that the Nitronox Plus meets all design input requirements..." and "Performance testing supports the conclusion that the subject device is as safe and effective as the predicate device, and therefore the two devices are substantially equivalent."

    This is a substantial equivalence (510(k)) submission, which focuses on demonstrating that a new device is "as safe and effective" as a legally marketed predicate device, rather than providing detailed clinical efficacy trial data, which would be typical for a PMA submission or for AI/diagnostic devices. The "study" here is a set of "non-clinical performance tests" that show the device functions according to its engineering and safety specifications, aligning with the predicate device's performance profile (where applicable, and noting differences and new features).

    Ask a Question

    Ask a specific question about this device

    K Number
    K173205

    Validate with FDA (Live)

    Date Cleared
    2018-08-29

    (331 days)

    Product Code
    Regulation Number
    868.5330
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Equinox® Advantage Nitrous Oxide/Oxygen Analgesic Gas Mixing and Delivery System is indicated for administering an adjustable mixture of Nitrous Oxide analgesic gas and Oxygen, on demand, to a conscious, spontaneously breathing patient.

    Device Description

    Equinox® Advantage Nitrous Oxide/Oxygen Analgesic Gas Mixing and Delivery System is a portable pneumatically powered device intended to administer an adjustable mixture of Nitrous Oxide analgesic gas and Oxygen mixture to provide pain relief, on demand, to a conscious, spontaneously breathing, patient.

    The device provides two input connection with nitrous oxide and oxygen cylinders through pressure regulators. The device has a control knob for turning ON or OFF the device. When it is turned ON, the output of N2O/O2 gas mixture will only be administered by an inspiratory effort by the patient.

    A selection knob is used to select desired N20/02 gas mixture. The adjustable range is from 25% O2 (NJO 75%) to 100% O2 (N3O 0%), with the minimum of 25% oxygen output to eliminate the possibility of delivering a hypoxic mixture.

    The Oxygen Enrichment Function is designed to provide 30 L/min of 100% oxygen in order to flush any residual gas from the patient circuit following the completion of the patient treatment.

    The built-in pneumatically powered alarm system will generate both visual and audible alarms should either nitrous oxide or oxygen input fall below 40 PSI, and the device will be automatically shut off should oxygen input falls below 35 PSI.

    The main accessories for the proposed device are a disposable Patient Circuit and a Mouth Piece or a universal Face Mask.

    AI/ML Overview

    The provided text describes the regulatory clearance of a medical device, the Equinox® Advantage Nitrous Oxide/Oxygen Analgesic Gas Mixing and Delivery System, and its substantial equivalence to a predicate device, the Equinox® Relieve (K113687). However, the document does not contain the typical comprehensive study report details that would allow for a complete answer to all parts of your request. This document is a 510(k) Summary, which focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than presenting a de novo clinical trial.

    Here's an analysis based on the information provided:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of "acceptance criteria" in the format typically seen for algorithm performance (e.g., specific sensitivity, specificity, or AUC thresholds). Instead, it relies on demonstrating equivalence to a predicate device and adherence to relevant standards. The "performance" reported is primarily comparative.

    Characteristic / TestAcceptance Criteria (Implied by Predicate Equivalence & Standards)Reported Device Performance (Equinox® Advantage)
    Intended UseAdminister adjustable N2O/O2, on demand, to a conscious, spontaneously breathing patient (comparable to predicate's 50/50% N2O/O2, on demand, to a conscious, spontaneously breathing patient)Administers adjustable mixture of N2O/O2, on demand, to a conscious, spontaneously breathing patient. (Meets)
    Environments of UsePre-hospital and in-hospital use (equivalent to predicate)Pre-hospital and in-hospital environment. (Meets)
    Patient PopulationConscious, spontaneous breathing patients requiring pain relief (equivalent to predicate)Conscious, spontaneous breathing patients requiring pain relief. (Meets)
    Contra-IndicationsSame as predicateSame as predicate. (Meets)
    Gas input and regulating50-70 PSI Oxygen and Nitrous Oxide regulated by internal regulators (equivalent to predicate)50-70 PSI Oxygen and Nitrous Oxide regulated by internal regulators. (Meets)
    Gas mixingOxygen and Nitrous Oxide equalized and mixed (predicate was 1:1 ratio, proposed is variable ratios)Oxygen and Nitrous Oxide equalized and mixed at variable ratios. (Meets, with modification)
    Gas deliveryProportional to patient's demand (equivalent to predicate)Proportional to patient's demand. (Meets)
    Monitoring and alarmsAudible/Visual alarms activated if N2O or O2 input below 40 PSI (equivalent to predicate)Audible/Visual alarms activated if N2O or O2 input below 40 PSI. (Meets)
    Safety features (O2 supply failure)Protection circuit shuts off device in event of O2 supply failure (equivalent to predicate)Protection circuit shuts off the device in the event of O2 supply failure. (Meets)
    Safety features (N2O supply failure)Predicate: Shuts off device. Proposed: Continues with higher O2 output / 100% O2 if no N2O input, as per ISO 11195:1995 Sec.12.4.Continues to operate with higher oxygen output; delivers 100% oxygen if no N2O input. (Meets ISO 11195:1995 Sec.12.4)
    Input pressure50 to 70 PSI (equivalent to predicate)50 to 70 PSI. (Meets)
    Output mixture concentrationPredicate: Preset 50/50% (V/V) O2 / N2O. Proposed: Adjustable 25%-100% Oxygen.Adjustable Oxygen: 25%-100%. (Meets, with new feature)
    Oxygen EnrichmentNot available on predicate. Proposed: 30 L/min oxygen.30 L/min oxygen. (New feature, meets ISO 80601-2-13:2011 Sec. 201.101.8)
    Output failure shutoffPredicate: Shut off if O2 much lower than N2O. Proposed: Removed function; minimum 25% oxygen setting prevents hypoxic mixture."Output failure shutoff" function removed; minimum 25% oxygen setting prevents delivering a hypoxic mixture. (Meets ISO 11195:1995 Sec.11.1)
    Inspiratory ResistanceEquivalent to predicateEquivalent to predicate. (Meets)
    Expiratory ResistanceEquivalent to predicateEquivalent to predicate. (Meets)
    Volume-time waveformsEquivalent to predicateEquivalent to predicate. (Meets)
    Pressure-time waveformsEquivalent to predicateEquivalent to predicate. (Meets)
    Flow-time waveformsEquivalent to predicateEquivalent to predicate. (Meets)
    "Low Oxygen supply" alarm characteristicsEquivalent to predicateEquivalent to predicate. (Meets)
    "Low Nitrous Oxide supply" alarm characteristicsEquivalent to predicateEquivalent to predicate. (Meets)
    "Oxygen Failure Shutoff" characteristicEquivalent to predicateEquivalent to predicate. (Meets)
    Oxygen concentration (at 50% setting)Equivalent to predicate (when predicate is 50%)Equivalent to predicate. (Meets)
    Nitrous Oxide concentration (at 50% setting)Equivalent to predicate (when predicate is 50%)Equivalent to predicate. (Meets)
    Electrical Safety & Essential PerformanceCompliance with ISO 11195 and other applicable standardsBench testing performed as per ISO 11195 and other applicable standards. (Meets)

    2. Sample size used for the test set and the data provenance

    The document describes "comparative performance testing" and "bench testing." These are laboratory-based tests comparing the proposed device against the predicate and against standards. It does not involve human subjects or patient data in the sense of a clinical trial. Therefore, specific "sample sizes" of a test set derived from patient data are not applicable here.

    • Sample size for test set: Not applicable (bench testing, not patient data).
    • Data provenance: Bench testing results, comparing the proposed device to the predicate device and relevant international standards (e.g., ISO 11195, ISO 80601-2-13). The country of origin for the test data is not specified but is presumably where the manufacturing and R&D for O-Two Medical Technologies Inc. takes place (Canada, as per the address). The testing is prospective for the purpose of demonstrating equivalence for the new device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not provided. For bench testing, ground truth is established by objective measurements against known physical and chemical standards and the specifications of the predicate device. Experts in this context would likely be engineers and testers.

    4. Adjudication method for the test set

    Not applicable. The testing described is objective, physical measurement and comparison, not expert adjudication of a diagnostic outcome.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a medical gas mixing and delivery system, not an AI-powered diagnostic or assistive tool for human readers. No MRMC study was conducted or mentioned.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This refers to an algorithm's performance without human intervention. This concept is not applicable to a physical gas mixing and delivery system. Its "performance" is inherent in its mechanical and pneumatic operation.

    7. The type of ground truth used

    For the aspects involving comparison to the predicate device, the "ground truth" for equivalence was the performance and specifications of the legally marketed predicate device (Equinox® Relieve/K113687). For aspects related to safety and functionality specifications, the ground truth was defined by international standards such as ISO 11195:1995 and ISO 80601-2-13:2011.

    8. The sample size for the training set

    Not applicable. This is not an AI/machine learning device that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K161718

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-10-27

    (127 days)

    Product Code
    Regulation Number
    868.5330
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MaxBlend 2 and MaxBlend Lite are designed to provide a continuous air/oxygen gas mixture and to continuously monitor the concentration of oxygen being delivered to infant, pediatric, and adult patients. It is a restricted medical device intended for use by qualified, trained personnel, under the direction of a physician, in professional healthcare settings, i.e.. hospital, sub-acute, and nursing-care facilities where the delivery and monitoring of air/oxygen mixtures is required. This is not intended as a life supporting device.

    Device Description

    The MaxBlend family of blenders, MaxBlend 2 and MaxBlend Lite, are oxygen delivery devices which incorporate an air/oxygen blender, battery powered oxygen monitor, and an adjustable flowmeter, all in a single assembly. The air/oxygen blender provides precise mixing of medical grade air and oxygen. The flowmeter provides control of the flow rate delivered. The oxygen monitor measures the oxygen concentration from the blender's gas flow, displays these measured concentrations, and provides user selectable high and low oxygen alarms.

    The MaxBlend 2 incorporates the air/oxygen blender within its enclosure. The MaxBlend Lite may be provided with or without the blender component pre-assembled, allowing the user to install the oxygen monitor/flowmeter module, the MaxBlend Lite component, on an existing compatible blender. The addition of the MaxBlend Lite module is intended to improve the safety of an existing air/oxygen blender that is in the user's possession. Both devices use the exact same oxygen monitor sensor and electronics, MaxO2ME (K153659). Both devices use the exact same components/assemblies in the flowmeter and sensor bleed manifold which form the gas pathway to the patient. The only substantive difference is the form of the enclosure for the monitor electronics.

    AI/ML Overview

    This document is a 510(k) Premarket Notification from Maxtec, LLC, regarding their MaxBlend 2 and MaxBlend Lite devices. It aims to demonstrate substantial equivalence to a predicate device, the Bird Sentry Blender (K973646), and references other cleared devices (K153659 – Maxtec MaxO2ME, K883038 - CareFusion / Bird - Blender, K925982 - BioMed Devices - Blender).

    Here's an analysis based on your request, highlighting that this document outlines non-clinical testing for a medical device that mixes and monitors breathing gases, not a typically AI-powered diagnostic imaging device. Therefore, many of the requested categories (like MRMC studies, roles of experts for ground truth, sample sizes for training/test sets for AI models, etc.) are not applicable or are addressed differently in the context of hardware device performance testing.

    Acceptance Criteria and Device Performance (Non-Clinical/Hardware)

    The acceptance criteria are primarily based on meeting the performance specifications of the predicate device and established standards for medical devices of this type. The "reported device performance" is generally stated as "passed" or "meets the acceptance criteria" for various tests.

    1. Table of Acceptance Criteria and Reported Device Performance

    AttributeAcceptance Criteria (Predicate / Standards Equivalent)Reported Device Performance (MaxBlend 2 and MaxBlend Lite)
    Indications for UseContinuous air/oxygen gas mixture, monitor O2 concentration for infant, pediatric, adult patients in institutional settings; not life supporting.Continuous air/oxygen gas mixture, continuously monitor O2 concentration for infant, pediatric, adult patients in professional healthcare settings (hospital, sub-acute, nursing-care facilities); not life supporting.
    Environments of UseInstitutional (healthcare settings), Not for MRI.Professional healthcare settings (hospital, sub-acute, nursing-care facilities), Not for MRI.
    Patient PopulationInfant, pediatric, and adult patients.Infant, pediatric, and adult patients.
    Weight2 kg2.4 kg
    Power source of oxygen monitor2 x AA Alkaline batteries4 x AA Alkaline batteries
    Gas Supply TypeAir / Oxygen, 30 to 75 psiAir / Oxygen, 30 to 75 psi
    % Oxygen Control21 - 100%, Accuracy ± 3%21 - 100%, Accuracy ± 3%
    Mixed Gas Stability± 1% oxygen± 1% oxygen
    Flow Range of BlendersLow flow: 0-30 Lpm; High flow: 0-100 Lpm.Low flow: 0-30 Lpm; High flow: 0-100 Lpm.
    Pressure Supply Differential AlarmAir / oxygen pressure < 20 psi alarms sounds.Air / oxygen pressure < 20 psi alarms sounds.
    Pressure Drop≤ 6 psi @ 50 psi≤ 6 psi @ 50 psi
    Bleed Flow3-13 Lpm at 50 psi depending on model.3-13 Lpm at 50 psi depending on model.
    Oxygen Measurement Range0-100%0-100%
    Total Accuracy (Oxygen)+3% Actual oxygen level over full operating temperature range.+3% Actual oxygen level over full operating temperature range.
    Response Time (Oxygen Monitor)< 20 seconds90% of final value in approx. 15 seconds at 23°C (meets/exceeds criterion).
    Warm-up TimeNone required.None required.
    High Alarm Range (Oxygen)18 - 99%16 - 100% (broader range, considered equivalent as it allows broader control and is set by clinician).
    Operating Temperature15°C - 40°C (59°F – 104°F)15°C - 40°C (59°F – 104°F)
    Storage Temperature-15°C - 50°C (5°F – 122°F)-15°C - 50°C (5°F – 122°F)
    Humidity0-100% (non-condensing)0-95% (non-condensing) (slight difference, but likely acceptable for intended environment).
    Expected Use-Life of Sensor750,000 O2% hours1,500,000 O2% hours (approx. 2 years) (superior to predicate, identical to reference device).
    Flowmeter AccuracyTypically ± 10% of indicated value.± 10% of indicated value when inlet pressure is 50 psi.
    Flow Meter Ranges0-30 Lpm for Low flow; 0-100 Lpm for High flow (user discretion).0-3 Lpm, 0-30 Lpm, 0-70 Lpm, 0-100 Lpm (broader range tested, at user's discretion, accuracy is the same).
    Standards ComplianceTested to older versions of equivalent standards (e.g., electrical safety).ANSI/AAMI/ES 60601-1, IEC 60601-1-2, ISO 80601-2-55, IEC 60601-1-8, ISO11195, ISO 15002:2008, ISO15001. (Indicates compliance with current standards).
    Biocompatibility (Materials in Gas Pathway)Materials externally communicating, tissue (permanent), no humidified gases.Externally communicating, tissue (permanent), no humidified gases. VOC and PM2.5 testing performed and passed. Materials identical to cleared Maxtec products and blenders.
    Cleaning Durability[Implied standard durability]Durability of marking tested with standard cleaning wipes/disinfectants. (Passed)
    Packaging Validation[Implied standard validation]Testing per ISTA 2A. (Passed)
    Usability[Implied safe and effective use]Usability testing performed with healthcare professionals. (Passed)

    Overall Conclusion found in the document: "In all cases the proposed device passed or meets the acceptance criteria. One can find the proposed device substantially equivalent to the predicate Bird Sentry Blender (K973646)."


    Regarding the AI/Diagnostic-Specific Questions (Not Directly Applicable to this Hardware Device but Addressed for Completeness):

    2. Sample Size Used for the Test Set and Data Provenance:

    • Not Applicable in the AI/ML sense. This submission relates to physical device performance and safety testing. The "test set" would refer to devices manufactured and subjected to various non-clinical engineering and performance tests (e.g., electrical safety, EMC, accuracy, durability, environmental conditions, etc.). The sample size for these tests is not explicitly stated in terms of a "data set" but rather implies a sufficient number of representative devices for engineering validation.
    • Data Provenance: The testing was conducted by or on behalf of Maxtec, LLC, in the USA (Salt Lake City, Utah). It's "prospective" in the sense that the new devices were built and tested to meet pre-defined criteria, not a retrospective analysis of patient data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts:

    • Not Applicable in the AI/ML sense. Ground truth for device performance is established by engineering specifications, international and national standards (e.g., ISO, IEC, ANSI/AAMI), and comparison to a legally marketed predicate device.
    • Qualifications of Experts: The experts are likely engineers, quality assurance personnel, and regulatory affairs professionals involved in device design, manufacturing, and testing, potentially with input from clinical users for usability studies. Their qualifications would be in engineering, quality management systems (QMS), and regulatory compliance for medical devices.

    4. Adjudication Method for the Test Set:

    • Not Applicable in the AI/ML sense. Device performance testing involves objective measurements against a standard or specification. There isn't an "adjudication" process like a clinical expert consensus on an image. Test results either pass or fail the predetermined criteria.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

    • No, not done. This is a hardware device focused on gas mixing and oxygen monitoring, not a diagnostic imaging AI algorithm that assists human readers. Therefore, an MRMC study is not relevant.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    • Not applicable as an "algorithm only" study. The core components (oxygen monitor and blender) are "standalone" in function (they perform their tasks independently) and their performance was evaluated by integrating them into the MaxBlend family. The document states "The major components... have already been cleared for the identical indications for use, population, and environment of use. Integrating them into a single assembly for user convenience does not alter their performance."

    7. The Type of Ground Truth Used:

    • The "ground truth" for this device's performance is based on:
      • Engineering specifications and design requirements.
      • Performance characteristics of the legally marketed predicate device (Bird Sentry Blender - K973646).
      • Compliance with recognized industry standards (e.g., ISO, IEC, ANSI/AAMI).
      • Comparison to previously cleared reference devices (K153659 – Maxtec MaxO2ME, K883038 - CareFusion / Bird - Blender, K925982 - BioMed Devices - Blender) which contribute components to the MaxBlend family.
      • Objective physical measurements and test results (e.g., accuracy of oxygen control, response time, flow rates, pressure, etc.).

    8. The Sample Size for the Training Set:

    • Not Applicable in the AI/ML sense. There is no "training set" for an AI model. This is a hardware device.

    9. How the Ground Truth for the Training Set Was Established:

    • Not Applicable in the AI/ML sense. No training set exists.
    Ask a Question

    Ask a specific question about this device

    K Number
    K142291

    Validate with FDA (Live)

    Date Cleared
    2015-05-19

    (274 days)

    Product Code
    Regulation Number
    868.5330
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GMX Series Medical Air/Oxygen Blender is designed to dispense a continuous blend of Medical Air and Oxygen via outlet ports to infant, pediatric, and adult patients.

    Device Description

    The GMX Series Medical Air/Oxygen blender is a restricted medical device intended for use by qualified and trained personnel under the direction of a physician in institutional environments where delivery of air/oxygen mixtures is required. The GMX Series Medical Air/Oxygen blender includes two models: GMX30U-AIR/O2, GMX120U-AIR/O2. The GMX Series Medical Air/Oxygen blender is designed, tested, and will be manufactured in accordance with the following standard: ISO 11195 Gas mixers for medical use - Stand-alone gas mixers 1995.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document compares the GMX Series Medical Air/Oxygen Blender (the device) to the Precision Blender (the predicate device). The acceptance criteria are implicitly those of the predicate device, demonstrating that the new device meets or is substantially equivalent to these performance metrics.

    ItemAcceptance Criteria (Predicate Device K053232)Reported Device Performance (GMX Series Air/Oxygen Blender)
    Oxygen % Range21-100%21-100%
    Accuracy± 3%± 3%
    Energy SourceAir SupplyAir Supply
    Supply Pressure30-75 Psi30-72.5 psi
    Max Flow (GMX120U-AIR/O2)≥120 LPM≥120 LPM
    Pressure Drop (GMX120U-AIR/O2)≤6psi at 50Psi inlet pressure and 40lpm flow≤6psi at 50Psi inlet pressure and 40lpm flow
    Alarm/Bypass Reset≤6psi≤6psi
    Alarm Intensity0.3 meter ≥ 80db1 foot ≥ 80db
    Operating Temperature15°C-40°C10°C-40°C
    ConnectorNIST or DISSNIST or DISS
    Max Flow (GMX30U-AIR/O2)≥30 LPM≥30 LPM
    Pressure Drop (GMX30U-AIR/O2)≤6psi at 50Psi inlet pressure and 10lpm flow≤6psi at 50Psi inlet pressure and 10lpm flow

    2. Sample Size Used for the Test Set and Data Provenance

    The document describes non-clinical tests based on compliance with voluntary standards (ISO, USP, ASTM, IEC). It does not specify a "test set" in terms of patient data or a specific number of items tested. Instead, it refers to performance testing and final acceptance testing. The data provenance is not explicitly stated as country of origin but implies laboratory testing to meet international and US standards. The study is prospective in the sense that the device was manufactured and then tested against these standards.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable. The "ground truth" for this device is based on established engineering and medical device performance standards. It doesn't involve clinical evaluations by medical experts to establish a "ground truth" for diagnostic accuracy, but rather verification that the device physically performs as specified and meets safety requirements.

    4. Adjudication Method for the Test Set

    This information is not applicable. The compliance is determined by meeting the specified values in the standards, not by an adjudication process as would be used in clinical studies.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    An MRMC comparative effectiveness study was not done. The device is a medical air/oxygen blender, not an AI-powered diagnostic or assistive tool for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is a physical medical device (an air/oxygen blender), not an algorithm. Therefore, "standalone" performance in the context of an algorithm is not applicable. Its performance is inherent to its mechanical and pneumatic design and is tested directly.

    7. The Type of Ground Truth Used

    The ground truth used is engineering and performance standards, specifically:

    • ISO 11195 Gas mixers for medical use – Stand-alone gas mixers 1995
    • USP <788>:2012—Particulate Matter in Injections
    • ASTM D 5466-01(2007): Standard Test Method for Determination of Volatile Organic Chemicals in Atmospheres (Canister Sampling Methodology)
    • IEC 60601-1 Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance.
    • The performance specifications of the legally marketed predicate device (Precision Blender K053232).

    8. The Sample Size for the Training Set

    This information is not applicable. The device is not an AI algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 3