Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K243756

    Validate with FDA (Live)

    Date Cleared
    2025-07-17

    (223 days)

    Product Code
    Regulation Number
    870.1330
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The 0.014" Willow Guidewire is intended for general intravascular use, including neurovascular and peripheral vasculatures. It can be used to selectively introduce and position catheters and other interventional devices within the peripheral and neuro vasculature. This device should be used only by physicians trained in percutaneous, intravascular techniques and procedures.

    Device Description

    Like the predicate device, the 0.014" Willow Guidewire is a single-use product with a shapeable tip available in straight and pre-shaped configurations, used to gain intravascular access to facilitate the positioning and exchange of interventional devices in small diameter, tortuous vasculature for neuro and peripheral diagnostic and interventional procedures. The wire can be torqued to facilitate navigation through the vasculature. The 0.014" Willow Guidewire comes in three stiffness profiles: Soft, Standard and Support.

    AI/ML Overview

    The provided FDA 510(k) clearance letter and summary for the 0.014" Willow Guidewire do not describe a study involving an AI/Machine Learning device.

    Instead, this document describes a traditional medical device (a guidewire) and its clearance process through substantial equivalence to a legally marketed predicate device. The "acceptance criteria" and "study that proves the device meets the acceptance criteria" in this context refer to engineering and biocompatibility bench testing, not clinical performance studies for an AI algorithm.

    Therefore, many of the questions related to AI/ML device performance (like sample size for test/training sets, expert ground truth, adjudication methods, MRMC studies, standalone performance, effect size of AI assistance) are not applicable to this document.

    However, I can extract the relevant information regarding the acceptance criteria and the engineering study (bench performance testing) for the 0.014" Willow Guidewire.


    Acceptance Criteria and Study for the 0.014" Willow Guidewire

    The 0.014" Willow Guidewire is a traditional medical device, not an AI/ML device. Therefore, the "acceptance criteria" and "study" refer to engineering specifications and bench performance testing, along with biocompatibility testing, to demonstrate substantial equivalence to a predicate device.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally implied by "Acceptance criteria were met" or "considered acceptable" for each test. The reported device performance is that these criteria were successfully achieved.

    TestTest Method SummaryReported Device Performance / Conclusions
    Bench Performance Testing
    Coating LubricityHydrophilic coating lubricity was assessed after multiple pull cycles through silicone pads.Acceptance criteria were met.
    Coating DurabilityCoating durability was assessed after repeating multiple pull cycles through silicone pads.Acceptance criteria were met.
    Coating IntegrityPer FDA guidance "Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling" (2019). Inspected pre- and post-simulated use tracking with comparison to predicate.Coating integrity was reported and considered acceptable.
    Corrosion ResistancePer FDA guidance "Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling" (2019).Acceptance criteria were met.
    Dimensional & Visual Insp.Per FDA guidance "Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling" (2019). Verified key dimensions.Acceptance criteria were met.
    Flex FatigueSubjected guidewire to multiple flexure cycles around cylindrical pins.Acceptance criteria were met.
    FractureSubjected guidewire to multiple wrappings around cylinder and visually inspected for signs of fracture.Acceptance criteria were met.
    Kink ResistancePer FDA guidance "Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling" (2019). Wrapped guidewire around test fixture with clinically relevant radii.Acceptance criteria were met.
    Particulate CharacterizationPer FDA guidance "Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling" (2019). Counted particulates after tracking through tortuous simulated use model, with comparison to predicate.Particulate counts were reported and considered acceptable.
    RadiopacityPer FDA guidance "Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling" (2019). Evaluated subject and predicate guidewires under fluoroscopy.All Willow Guidewires demonstrated acceptable radiopacity; comparable to predicate.
    Simulated UsePer FDA guidance "Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling" (2019). Tested with microcatheter, introducer, and torque device while navigating to target in a tortuous simulated use model.Acceptance criteria were met.
    Tensile StrengthPer FDA guidance "Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling" (2019). Measured force to break at each bond.Acceptance criteria were met.
    Tip FlexibilityPer FDA guidance "Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling" (2019). Measured force to deflect tip at 5 mm, 10 mm, and 20 mm gauge lengths, compared to predicate.Acceptance criteria were met.
    Tip ShapeabilityShaped the guidewire tip three times per labeling.Acceptance criteria were met.
    Tip Shape RetentionMeasured tip shape retention after tracking through a tortuous simulated use model.Acceptance criteria were met.
    TorqueabilityPer FDA guidance "Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling" (2019). Tracked through tortuous simulated use model; evaluated steerability and torque response.Acceptance criteria were met.
    Torque StrengthPer FDA guidance "Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling" (2019). Recorded rotations until failure.Acceptance criteria were met.
    Biocompatibility Testing
    Acute Systemic ToxicityPer ISO 10993-11.Pass: No evidence of acute systemic toxicity.
    Compliment ActivationPer ISO 10993-4.Pass: Non-activator.
    Cytotoxicity – MEM ElutionPer ISO 10993-5.Pass: Non-cytotoxic.
    HemolysisPer ISO 10993-4.Pass: Non-hemolytic.
    Intracutaneous ReactivityPer ISO 10993-10.Pass: Non-reactive.
    In Vivo ThrombogenicityPer ISO 10993-4.Pass: Non-Thrombogenic.
    PyrogenicityPer ISO 10993-11.Pass: Non-pyrogenic.
    SensitizationPer ISO 10993-10.Pass: Non-sensitizing.

    Information Not Applicable (NA) to this Device/Submission:

    The following questions are specifically targeted at AI/ML medical devices and are not relevant to the 0.014" Willow Guidewire, which is a traditional physical medical device.

    1. Sample size used for the test set and the data provenance: NA (This is a physical device, not an AI/ML algorithm requiring a data test set.)
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: NA (No "ground truth" in the AI/ML sense is established for this type of device.)
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: NA (No adjudication method described for physical device testing.)
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: NA (No MRMC study or AI assistance involved.)
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: NA (Not an algorithm.)
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): NA (Ground truth is not a concept applied in this context; performance is measured against engineering specifications and predicate characteristics.)
    7. The sample size for the training set: NA (No training set for a physical device.)
    8. How the ground truth for the training set was established: NA (No training set or ground truth in this context.)

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1