Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K233081

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-11-15

    (416 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    NUVENTUS NV.C™ Dental Implants are indicated for the functional and esthetic oral rehabilitation of the upper or lower jaw of edentulous or partially edentulous patients. NUVENTUS NV.C™ dental implants may be used for immediate, early or delayed implantation following the extraction or loss of natural teeth. The implants can be placed with immediate loading for single-tooth or multiple teeth restorations when good primary stability is achieved and with appropriate occlusal loading to restore chewing function. NUVENTUS NV.CTM Dental Implants are compatible for use with the following prosthetic interfaces.

    Implant System Prosthetic CompatibilityPlatform Size/Designation
    Nobel Biocare Internal Conical Connection (CC)NP CC
    RP CC
    Device Description

    The purpose of this submission is to obtain marketing clearance for NUVENTUS NV.C™ Dental Implant System endosseous dental implants and cover screws. The dental implants are intended to interface with Internal Conical Connection (CC) prosthetic components from Nobel Biocare.

    A summary of the subject device implant and the associated compatible OEM prosthetic connection is provided in the table Summary of Subject Device Implant Designs.

    Subject Device ImplantDescriptionPlatformDesignationLength (mm)*OEM Prosthetic Compatibility(K071370, K161435, K161416)
    Implant, NV.C, Platform NP, Ø3.5 mmNP8.51011.51315Nobel Biocare Internal Conical, NP Platform
    Implant, NV.C, Platform NP, Ø4.3 mmNP8.51011.51315Nobel Biocare Internal Conical, NP Platform
    Implant, NV.C, Platform RP, Ø5.0 mmRP8.51011.51315Nobel Biocare Internal Conical, RP Platform

    The subject device dental implants have a conical abutment seating surface on the interior of the implants and internal threads so that prosthetic components may be fastened to the implant. The implant lines have two (2) abutment interface connections with internal geometric features to allow for rotational resistance of the mating abutment. All subject device implants are manufactured from Ti-6Al-4V alloy conforming to ASTM F136.

    The external surface of all subject device implants is threaded, and the implant body tapers at the apical end. which includes two (2) cutting flutes. At the coronal end. the Ø4.3mm and Ø5.0mm (body diameter) subject device implants have two (2) fluted features on the body of the implant spaced 180° apart. Each fluted feature has horizontal grooves spaced vertically within the flute surface. The number of grooves within each flute ranges from 3 to 5 and the actual number is a function of the implant length. The Ø3.5mm (body diameter) subject device implants do not have fluted features. The endosseous surface of all subject implants is textured by blasting with resorbable media

    The subject device implants are compatible with prosthetic components that interface with Nobel Biocare Internal Connection implants. The subject device cover screws are manufactured from Ti-6A1-4V alloy conforming to ASTM F136 and are anodized to identify the prosthetic platform (NP and RP).

    The compatible Nobel Biocare Internal Conical Connection prosthetic components (NP and RP platforms) include cover screws, healing abutments, temporary abutments esthetic abutments, straight multi-unit abutment, and angled multi-unit abutments.

    All subject device implants and cover screws are individually packaged and are provided sterile.

    AI/ML Overview

    The FDA document provided pertains to a 510(k) premarket notification for a dental implant system (NUVENTUS NV.C™ Dental Implant System). This document focuses on demonstrating substantial equivalence to predicate devices through engineering analysis, non-clinical performance data (e.g., mechanical testing, biocompatibility), and a review of clinical literature. It does not present a study proving the device meets specific acceptance criteria based on AI/ML performance metrics, expert consensus, or clinical outcomes from a new study specific to an AI device.

    Therefore, I cannot extract the requested information regarding acceptance criteria, AI/ML study design (sample size, data provenance, expert ground truth, adjudication, MRMC study, standalone performance), or training set details because this information is not present in the provided text.

    The document discusses performance data related to the dental implant itself (e.g., sterilization, biocompatibility, mechanical testing, surface analysis), and clinical literature data to support the substantial equivalence of the implant's design features with established predicate devices. The "Clinical Testing Literature Table" outlines existing studies on predicate implants, reviewing their clinical outcomes (like bone level changes) to support the subject device's design principles. This is not the same as a study testing an AI device's performance against defined acceptance criteria.

    In summary: The provided text is a 510(k) submission for a dental implant system, not an AI/ML-based medical device. Thus, the requested details specific to AI/ML device performance and validation studies are not available in this document.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1