    K222469

    Validate with FDA (Live)

    Device Name
    TIB Abutments
    Date Cleared
    2023-04-17

    (244 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TIB Abutments are premanufactured prosthetic components directly connected to endosseous dental implants and are intended for use as an aid in prosthetic rehabilitation. The TIB abutments consist of two major parts. Specifically, the titanium base and mesostructure components make up a two-piece abutment. The system integrates multiple components of the digital dentistry workflow: Scan files from desktop scanners, CAM software, ceramic material, milling machine and associated tooling and accessories.

    The intended use for the TiB narrow abutments used with the Ø3.0 mm External hex and Ø3.0 mm Deep Conical implants is limited to replacement of maxillary lateral incisors and mandibular lateral and central incisors.

    Device Description

    The Narrow TIB Abutment bases are standard premanufactured titanium allov abutments for supporting a dental restoration and mesostructure. The dental laboratory is to fabricate the mesostructure restoration by CAD/CAM technique out of zirconia. The TIB abutment base then serves as the interface between the endosseous implant and the Zirconia restoration. The TIB Abutment Base is designed to support the restoration on an endosseous implant in order to restore chewing function for the patient.

    The DC Ø3.0mm Narrow TiB abutments are only to be placed straight in the patient. Occlusal loading forces are only to be applied through the central longitudinal axis of the implant body.

    The mesostructured restoration is a CAD/CAM designed prosthesis milled out of Zirconia, which is designed to fit the abutment base in order to restore chewing for the patient. Each restoration is custom designed using 3Shape Abutment Designer Software in order to meet the requirements of each patient on a case-by-case basis. Limitations have been put in place in 3Shape Abutment Designer in order to prevent malfunctioning of the restoration.

    The subject Narrow TIB Abutments are compatible with the Southern Implants' Deep Conical and External Hex implants and screws. The subject TIB abutment bases are manufactured from Titanium alloy conforming to ASTM F136 and are color coded by gold anodizing. The anodization process is the same as used for previously cleared anodized titanium alloy devices in K193084. The mesostructured restoration is to be manufactured from Zirconia - Sage Max NexxZr which has been previously cleared for use in K130991.

    The digital workflow includes the following products (not subject devices to this submission):

    • Ceramic material: Sage Max NexxZr Zirconia Restorative material (K130991)
    • . Cement: Ivoclar Vivadent Multilink Hybrid Abutment Cement (K130436)
    • Intra-oral scanner ●
    • . Lab scanner: 3Shape E3 Desktop Scanner (3Shape A/S)
    • Abutment design software: 3Shape Abutment Designer Software (K151455) ●
    • Milling machine: Roland DXW51D ●
    AI/ML Overview

    This is a summary of the provided text regarding the TIB Abutments.

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state "acceptance criteria" in a tabular format with corresponding reported performance for the subject device. Instead, it demonstrates substantial equivalence by comparing the technological characteristics of the subject device (TIB Abutments) with primary and reference predicate devices. The implicit acceptance criterion is that the subject device's performance, as demonstrated by the listed tests and comparisons, is equivalent to the predicate devices.

    Here's a table comparing the subject device's characteristics to the primary predicate, which serves as the basis for demonstrating equivalence:

    CharacteristicSubject Device (Narrow TIB Abutments)Primary Predicate Device (K193084)Reported Performance/Equivalence Justification
    Indications for UsePremanufactured prosthetic components directly connected to endosseous dental implants for prosthetic rehabilitation. Two-piece abutment (titanium base & mesostructure). Integrates digital dentistry workflow. Specific limitation for Ø3.0mm External hex & Deep Conical: replacement of maxillary lateral incisors and mandibular lateral and central incisors.Premanufactured prosthetic components directly connected to endosseous dental implants for prosthetic rehabilitation. Two-piece abutment (titanium base & mesostructure). Integrates digital dentistry workflow."Similar indications for use," "only with restrictions imposed on the Ø3.0 Deep Conical and Ø3.0 External Hex abutments." Thus, the subject device's Indications for Use are a subset or modified version of the predicate, deemed equivalent.
    Product CodeNHA, PNPNHA, PNPIdentical.
    Abutment Design2 Piece – Premanufactured titanium abutment, mounted onto the implant and fixed with a screw. SageMaxx Zirconia (K130991) hybrid/crown restoration milled and bonded to the titanium abutment.2 Piece – Premanufactured titanium abutment, mounted onto the implant and fixed with a screw. SageMaxx Zirconia (K130991) hybrid/crown restoration milled and bonded to the titanium abutment."Equivalent design," "differing only in dimension." "All the predicate device design features... are present."
    Collar Height1.5 or 3mm (External HEx) / 1.5 or 3mm (Deep Conical)0.6, 1.5 or 3mm (for predicate versions)Subject device offers a subset of predicate's collar heights. Considered equivalent.
    Maximum Abutment Top-Cap Angulation20° (External Hex) / 0° (for Ø3mm implant) and 20° (Deep Conical)20°"Identical maximum angulation for the ceramic restoration" (with the noted exception for Ø3.0 Deep Conical narrow TIB abutments).
    Implant ConnectionExternal Hex / Deep ConicalExternal Hex / Deep Conical"Reference devices K173706 and K163060... contain the identical connection types and sizes to those of the subject devices."
    Post HeightMinimum 4.5mmMinimum 4.5mmIdentical.
    Abutment Restorative Platform Diameter3.85mm (External Hex) / 3.85mm (Deep Conical)4.3mm (External Hex predicate) / 4.5mm (Deep Conical predicate)Subject device dimensions are different but considered equivalent in overall design and function within the context of substantial equivalence.
    Abutment MaterialTitanium Grade 5 Alloy (ASTM F136)Titanium Grade 5 Alloy (ASTM F136)"Identical in formulation, processing, component interactions, and storage conditions" to the predicate.
    Abutment SurfaceMachined and anodizedMachined and anodizedIdentical. Anodization process is the same as used for previously cleared devices in K193084.
    Abutment Screw MaterialTitanium Grade 5 Alloy (ASTM F136)Titanium Grade 5 Alloy (ASTM F136)Identical.
    Restoration MaterialZirconia - Sage Maxx NexxZr (K130991)Zirconia - Sage Maxx NexxZr (K130991)"Substantially equivalent... with reference to the abutments restoration material."
    CAD/CAM Design Workflow3Shape E3 Desktop Scanner (3Shape A/S), 3Shape Abutment Designer Software3Shape E3 Desktop Scanner (3Shape A/S), 3Shape Abutment Designer Software"Same scanners and software to design the restoration."
    CAD/CAM Manufacturing WorkflowWorkNC CAM software, Roland DWX51D milling unitWorkNC CAM software, Roland DWX51D milling unit"Follow the same design and manufacturing CAD/CAM workflows."
    Mechanical Fatigue TestingDynamic Fatigue Testing per ISO 14801Dynamic Fatigue Testing per ISO 14801"Performed on worst-case subject device constructs" and references K193084 and K220841 for "fatigue strength of the worst-case combinations." Demonstrates equivalent mechanical performance.
    SterilityProvided non-sterileProvided non-sterileIdentical.
    UsageSingle-patient single-useSingle-patient single-useIdentical.
    BiocompatibilityReferenced from K193084(Predicate devices are biocompatible)"Supported by the fact that materials are identical in formulation, processing, component interactions, and storage conditions to the predicate device in K193084."
    MR Safety TestingAs per FDA Guidance Document "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment" (referenced from K193084)(Predicate devices meet MR safety requirements)Referenced from K193084, implying equivalence.
    Software Validation TestingPer FDA Guidance Document for Off-The-Shelf Software Use in Medical Devices (referenced from K193084)(Predicate devices meet software validation requirements)Referenced from K193084, implying equivalence.
    Scanning and Milling ValidationReferenced from K193084(Predicate devices meet scanning and milling validation)Referenced from K193084, implying equivalence.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The document primarily relies on non-clinical performance data for mechanical fatigue testing (ISO 14801).

    • Sample size: The specific sample size for the dynamic fatigue testing is not provided in the summary. It states "Dynamic testing was performed on worst-case subject device constructs." This typically implies multiple samples were tested to statistical confidence.
    • Data provenance: The document does not explicitly state the country of origin of the data for the performance tests, nor whether it was retrospective or prospective. Given that this is a 510(k) submission for a South African company, it is likely the testing was conducted at a certified lab, possibly in South Africa or internationally, to ISO standards. The type of testing (mechanical fatigue) is inherently prospective in its execution (i.e., new tests are run on new devices).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This device is a physical dental implant component, not a diagnostic AI system or image analysis tool. Therefore, the concept of "ground truth" established by experts in the context of diagnostic accuracy or clinical interpretation is not applicable here. The "ground truth" for this device's performance is established by engineering tests and material science standards, such as ISO 14801 for mechanical fatigue.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This is not a diagnostic device involving human interpretation that would require an adjudication method.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a physical dental implant component, not a diagnostic or AI-assisted device. No MRMC study was conducted or is relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an algorithm or software-only device in the context of a diagnostic or interpretive tool. The "digital dentistry workflow" involves software (3Shape Abutment Designer Software) but it's for design and manufacturing, not standalone diagnostic performance. Software validation was referenced from K193084.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the TIB Abutments is based on:

    • Compliance with material specifications (e.g., Titanium Grade 5 Alloy (ASTM F136), Zirconia - Sage Maxx NexxZr).
    • Performance against international engineering standards (e.g., ISO 14801 for dynamic fatigue testing).
    • Biocompatibility established through material equivalence to a predicate device with established biocompatibility.

    8. The sample size for the training set

    Not applicable. This is a physical medical device, not an AI/machine learning model that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set mentioned for an AI/machine learning model.