510(k) Data Aggregation

    K Number
    K211242

    Validate with FDA (Live)

    Date Cleared
    2021-09-09

    (136 days)

    Product Code
    Regulation Number
    880.5860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    It is intended to be used for medical purposes to inject fluids into or withdraw fluids from the body.

    Device Description

    A sterile device consisting of a calibrated barrel with plunger and a fixed needle at the distal end intended to be used for injection/withdrawal of fluids/gas (e.g., medication) to/from a medical device or the body (i.e., capable of both). It is made of stainless steel needle, plastic and silicone materials, and is manually operated. It is intended for various medical applications and is not dedicated to medication administration.

    It is intended be used by health-care personnel (doctors, nurses, etc.). The needle is lubricated with silicone. Product is sterile and sterilization process is validated according to ISO 11135:2014.

    AI/ML Overview

    This document is a 510(k) Premarket Notification for Promisemed Sterile Hypodermic Syringes and Verifine Sterile Hypodermic Syringes. It is not a study proving the device meets acceptance criteria for an AI/ML powered device. Instead, it demonstrates substantial equivalence to a legally marketed predicate device (Sol-M TB Syringe K153537) for a standard medical device (syringes).

    Therefore, the requested information about acceptance criteria, study design for AI/ML performance, ground truth establishment, expert adjudication, MRMC studies, and training set details are not applicable to this document.

    The document primarily focuses on:

    • Device Description: Sterile hypodermic syringes with a fixed needle.
    • Intended Use: To inject or withdraw fluids from the body for medical purposes.
    • Substantial Equivalence: Comparison to a predicate device (Sol-M TB Syringe K153537) in terms of intended use, technological characteristics, and performance.
    • Performance Testing: Bench testing according to recognized ISO standards (ISO 7886-1, ISO 9626, ISO 7864) and USP <85> for bacterial endotoxin, along with sterility, package integrity, shelf-life, and biocompatibility testing per ISO 10993 series. These tests confirm the physical and biological safety of the device, not the performance of an AI algorithm.

    Summary of Device Performance (from the document), adapted to fit a "performance" concept for a non-AI device:

    Acceptance Criteria (Bench Test Standards)Reported Device Performance (Compliance)
    ISO 7886-1:2017 (Sterile hypodermic syringes for manual use)Complied
    ISO 9626:2016 (Stainless steel needle tubing)Complied
    ISO 7864:2016 (Sterile hypodermic needles for single use)Complied
    USP <85> (Bacterial Endotoxin Test - limit < 20 EU/Device)Complied (non-pyrogenic, endotoxin < 20 EU/Device)
    ISO 11135:2014 (Ethylene oxide sterilization - SAL 10^-6)Complied (Sterilization validated, EO/chlorohydrin residuals within limits)
    ASTM D4169 (Simulated transportation) / ASTM F88/F88-15 (Seal strength) / ASTM F1929-15 (Dye penetration) (Package integrity/sterility maintenance)Complied (Packaging acceptable for product protection and sterility maintenance)
    ASTM F1980-16 (Accelerated Aging - Shelf Life)5 years shelf life validated
    ISO 10993-5:2009 (Cytotoxicity)No cytotoxicity
    ISO 10993-10:2010 (Irritation and Skin Sensitization)No evidence of skin irritation or sensitization
    ISO 10993-11:2006 (Acute Systemic Toxicity)No systemic toxicity
    USP 42-NF37 <151> (Pyrogen Test)Non pyrogenic
    USP <788> (Particulate matter in injections)Complied
    ISO 10993-4:2017 (Interactions with blood)Complied (No evidence of hemolysis)
    ISO 10993-7:2008 (Ethylene oxide sterilization residuals)Within limit

    The remaining questions are not applicable as this is a traditional medical device submission, not an AI/ML device:

    1. Sample sized used for the test set and the data provenance: Not applicable. Performance is based on bench testing of device characteristics, not a clinical test set from patient data.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. No "ground truth" for diagnostic or prognostic purposes is established. Device performance is determined by objective physical and chemical tests against recognized standards.
    3. Adjudication method: Not applicable.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done: Not applicable.
    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable.
    6. The type of ground truth used: Not applicable. Performance is based on compliance with established engineering and biocompatibility standards.
    7. The sample size for the training set: Not applicable.
    8. How the ground truth for the training set was established: Not applicable.