510(k) Data Aggregation

    K Number
    K203723

    Validate with FDA (Live)

    Date Cleared
    2021-12-10

    (354 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Branchor Balloon Guide Catheter is indicated for use to facilitate the insertion and guidance of an intravascular catheter into a selected blood vessel in the neuro vasculature, and injection of contrast media.

    The balloon provides temporary vascular occlusion during these procedures.

    The Branchor Balloon Guide Catheter can also be used as a conduit for retrieval devices.

    Device Description

    This balloon guide catheter is a variable stiffness catheter that has a radio-opaque marker at the distal end of the balloon and a branched connector at the proximal end, and is equipped with a braid reinforced coaxial lumen. A balloon is attached to the distal end, and the dimensions of this balloon guide catheter and the recommended balloon injection volume are indicated on the product label. The balloon guide catheter is provided sterile, using ethylene oxide, and is intended for single use only by physicians who have been adequately trained in neurointerventional procedures.

    The outer surface of this balloon guide catheter is coated with a hydrophilic coating in order to enhance lubricity when the surface is wet. The shaft lumen is provided with PTFE coating, except the connector part, to facilitate the passage of the guidewire and other devices through the section.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information based on the provided text, focusing on what is present and noting what is not mentioned:

    It's important to note that this document is a 510(k) premarket notification, which demonstrates substantial equivalence to existing devices, usually through non-clinical bench testing. It is not typically a clinical study involving human patients or complex AI performance metrics like those for image analysis.


    Acceptance Criteria and Device Performance

    This submission is for a medical device (Balloon Guide Catheter), not an AI/ML imaging device. As such, the "reported device performance" refers to the results of non-clinical bench testing, which are primarily pass/fail against engineering and design specifications.

    Acceptance Criteria (Test)Reported Device Performance
    Air Leakage Into Hub Assembly During AspirationPass
    Balloon Diameter to Inflation PressurePass
    Balloon Inflation and Deflation TimesPass
    Balloon Maximum DiameterPass
    Burst PressurePass
    Coating Integrity/Particulate Evaluation(Characterization Only)
    Design Verification – Simulated UsePass
    Dimensional AnalysisPass
    Freedom from Leakage and Damage on InflationPass
    Kink Resistance TestPass
    Liquid Leakage Under PressurePass
    Lumen Collapse Under Maximum AspirationPass
    Luer Hub TestingPass
    Peak Tensile Strength - TipPass
    Peak Tensile Strength - Tube JunctionPass
    Radio-detectabilityPass
    Tip FlexibilityPass
    Torque StrengthPass
    Visual Inspection - Distal TipPass
    Visual Inspection - SurfacePass

    Biocompatibility Testing

    Acceptance Criteria (Test summary for "Conclusion")Reported Device Performance (Conclusion)
    Systemic Toxicity - Rabbit Pyrogen TestNon-pyrogenic
    Systemic Toxicity - Systemic InjectionNo acute systemic toxicity
    Cytotoxicity - L929 cells/MEM Elution TestNon-cytotoxic
    Sensitization - Guinea Pig Maximization TestNon-sensitizing
    Irritation - Intracutaneous Injection TestNon-irritant
    Hemocompatibility - HemolysisNon-hemolytic
    Hemocompatibility - Partial Thromboplastin Time Assay (PTT)Minimal Activator
    Hemocompatibility - Complement Activation Assay (SC5b-9)Non-Potential Activator
    Hemocompatibility – ThrombogenicityNo Thrombosis

    Study Details (Applicable to Non-Clinical Bench Testing)

    1. Sample Size Used for the Test Set and Data Provenance:

      • Test Set Sample Size: Not explicitly stated for each test, but standard engineering testing involves sufficient samples to achieve statistical confidence for each specific test (e.g., multiple units for burst pressure, tensile strength, etc.).
      • Data Provenance: The submission is from "ASAHI INTECC CO., LTD" in "Aichi 489-0071, Japan". This indicates the testing was conducted by or on behalf of the manufacturer, likely in Japan or a contracted laboratory. The nature of these tests (bench tests) means they are not "retrospective or prospective" in the sense of patient data.
    2. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

      • This is a non-clinical device performance study (bench testing), not a study assessing AI performance or diagnostic accuracy. Therefore, "experts" in the context of establishing ground truth for data interpretation (like radiologists for image analysis) are not applicable. The "ground truth" for these tests is defined by engineering specifications and standards (e.g., how much pressure a balloon should withstand).
    3. Adjudication Method for the Test Set:

      • Not applicable. Adjudication methods (like 2+1 voting) are used in clinical studies or studies where human interpretation of complex data (like medical images) is involved. Bench tests have objective pass/fail criteria based on measured physical properties.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No. This is not an AI/ML imaging device and thus no MRMC study was conducted.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • No. This is not an AI/ML device.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • For the non-clinical tests, the "ground truth" is based on engineering specifications, recognized industry standards (e.g., ISO), and physical measurements designed to ensure the device meets its intended performance characteristics and safety profile.
    7. The sample size for the training set:

      • Not applicable. This is not an AI/ML device, so there is no "training set."
    8. How the ground truth for the training set was established:

      • Not applicable. There is no training set for this type of device submission.