510(k) Data Aggregation

    K Number
    K200873

    Validate with FDA (Live)

    Device Name
    HALO
    Manufacturer
    Date Cleared
    2020-11-20

    (233 days)

    Product Code
    Regulation Number
    892.2080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    HALO is a notification only cloud-based image processing software artificial intelligence algorithms to analyze patient imaging data in parallel to the standard of care imaging interpretation. Its intended use is to identify suggestive imaging patterns of a pre-specified clinical condition and to directly notify an appropriate medical specialist.

    HALO's indication is to facilitate the evaluation of the brain vasculature on patients suspected of stroke by processing and analyzing contrast enhanced CT angiograms of the brain acquired in an acute setting. After completion of the data analysis. HALO sends a notification if a pattern suggestive for a suspected intracranial Large Vessel Occlusion (LVO) of the anterior circulation (ICA, M1 or M2) has been identified in an image.

    The intended users of HALO are defined as appropriate medical specialists that are involved in the diagnosis and care of stroke patients at emergency department where stroke patients are administered. They include physicians such as neurologists, and/or other emergency department physicians.

    HALO's output should not be used for primary diagnosis or clinical decisions; the final diagnosis is always decided upon by the medical specialist. HALO is indicated for CT scanners from GE Healthcare.

    Device Description

    HALO is a notification only, cloud-based clinical support tool which identifies image features and communicates the analysis results to a specialist in parallel to the standard of care workflow.

    HALO is designed to process CT angiograms of the brain and facilitate evaluation of these images using artificial intelligence to detect patterns suggestive of an intracranial large vessel occlusion (LVO) of the anterior circulation.

    A copy of the original CTA images is sent to HALO cloud servers for automatic image processing. After analyzing the images, HALO sends a notification regarding a suspected finding to a specialist, recommending review of these images. The specialist can review the results remotely in a compatible DICOM web viewer.

    AI/ML Overview

    Here's a summary of the acceptance criteria and study details for the HALO device, based on the provided FDA 510(k) summary:

    HALO Device Acceptance Criteria and Study Details

    1. Table of Acceptance Criteria and Reported Device Performance

    MetricAcceptance Criteria (Implicit)Reported Device Performance (HALO)
    SensitivitySufficiently high for LVO detection (comparable to predicate)91.1% (95% CI, 86.0%-94.8%)
    SpecificitySufficiently high for LVO detection (comparable to predicate)87.0% (95% CI, 81.2%-91.5%)
    AUCHigh (indicative of good discriminative power)0.97
    Notification TimeFast enough for acute stroke setting (comparable to predicate)Median: 4 minutes 31 seconds. Range: 3:47 to 7:12
    Substantial EquivalenceEquivalent to predicate device ContaCT in terms of indications for use, technological characteristics, and safety and effectiveness.Concluded to be substantially equivalent.

    2. Sample Size and Data Provenance for Test Set

    • Sample Size: 348 CTA scans were initially collected, with 364 patients included for further analysis after exclusion. It's unclear if the "348 CTA scans" and "364 patients" refer to the same dataset or if some patients had multiple scans or if there was an expansion of the dataset. Assuming 364 cases (patients with at least one scan) were used.
    • Data Provenance: Retrospective evaluation in a consecutive patient cohort. Data was collected from US comprehensive stroke centers.

    3. Number and Qualifications of Experts for Ground Truth

    • Number of Experts: 3 neuro radiologists.
    • Qualifications: "Neuro radiologists" implies specialized training and experience in interpreting neurological imaging, which is appropriate for stroke diagnosis. Specific years of experience are not mentioned.

    4. Adjudication Method for Test Set

    The adjudication method is not explicitly stated. It says "Ground truth was established by an expert panel consisting of 3 neuro radiologists," which suggests a consensus-based approach, but the specific rule (e.g., majority vote, unanimous agreement, review by a lead expert if disagreement) is not provided.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, an MRMC comparative effectiveness study involving human readers with and without AI assistance was not explicitly mentioned or conducted as detailed in the summary. The study focused on the standalone performance of the HALO algorithm.

    6. Standalone Performance (Algorithm Only)

    Yes, a standalone study was performed. The clinical study retrospectively evaluated the performance of the HALO clinical decision support algorithm for LVO detection using the collected CTA scans. The reported sensitivity, specificity, AUC, and notification time are all measures of the algorithm's standalone performance.

    7. Type of Ground Truth Used

    Expert Consensus. The ground truth for the test set was established by an expert panel consisting of 3 neuro radiologists.

    8. Sample Size for Training Set

    The sample size for the training set is not explicitly mentioned in the provided document. The document only covers the evaluation of the algorithm.

    9. How Ground Truth for Training Set was Established

    How the ground truth was established for the training set is not explicitly mentioned in the provided document. ("database of images" is stated for the core algorithm, but not how ground truth was applied to them).