Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K200417

    Validate with FDA (Live)

    Date Cleared
    2020-05-02

    (72 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The RIST Radial Access Catheter is indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature.

    Device Description

    The RIST Radial Access Catheter is a flexible, single lumen catheter compatible with 0.035 inch and 0.038-inch guidewires. It is designed to deliver interventional devices into the peripheral, coronary, and neuro vasculature and is intended to provide access to the target site via transradial access. The RIST Radial Access Catheter consists of a radiopaque catheter shaft reinforced with stainless steel and has a luer connector at its proximal end for the attachment of accessories and the infusion of fluids. The distal tip of the catheter is shaped to have a smooth, rounded tip and is offered in two different distal segment shapes, namely Simmons 2 (SIM2) and Berenstein (BER). The catheter has a nominal outer diameter of 0.070 inches and a nominal inner diameter of 0.040 inches. It is available in two working lengths: 120 cm and 130 cm. The RIST Radial Access Catheter is compatible with the RIST Cath Radial Access Long Sheath (K191551) and off-theshelf accessories. The RIST Radial Access Catheter is supplied sterile, non-pyrogenic, and intended for single use only.

    AI/ML Overview

    The provided text is a 510(k) Summary for the RIST Radial Access Catheter. This document details the engineering and performance tests conducted to demonstrate the device's substantial equivalence to a predicate device, rather than the evaluation of an AI/ML medical device.

    Therefore, many of the typical acceptance criteria and study design elements requested in the prompt, such as:

    • A table of acceptance criteria and reported device performance related to AI/ML metrics (e.g., sensitivity, specificity, AUC)
    • Sample size for the test set and data provenance (e.g., country, retrospective/prospective) related to clinical data for AI/ML performance
    • Number of experts and their qualifications for establishing ground truth for AI/ML
    • Adjudication method for the test set for AI/ML
    • Multi-Reader Multi-Case (MRMC) comparative effectiveness study for human readers with and without AI
    • Standalone (algorithm-only) performance
    • Type of ground truth used (expert consensus, pathology, outcome data) for AI/ML
    • Sample size for the training set and how ground truth was established for training

    cannot be extracted from this document. The document describes a traditional medical device (catheter) and its physical and functional performance testing, not an AI or machine learning algorithm.

    The document focuses on establishing substantial equivalence based on:

    1. Indications for Use: Identical to the predicate device.
    2. Technological Characteristics: Comparison of materials, dimensions, and operational principles.
    3. Performance Data: A series of design verification and validation tests, and biocompatibility tests directly on the catheter itself.

    However, I can provide the acceptance criteria and reported performance for the physical device based on the available information:


    Acceptance Criteria and Device Performance (for a physical catheter; not an AI/ML device)

    Acceptance Criteria (Test Performed)Reported Device Performance (Results)
    Design Verification Testing
    Tensile StrengthAll units met the Tensile Strength acceptance criteria.
    Torque StrengthAll units met the Torque Strength acceptance criteria.
    Particulates (per USP <788>)The acceptance criteria for Particulate testing was met.
    Dimensional Verification - IDThe acceptance criteria for ID Dimensional Verification was met.
    Dimensional Verification - ODThe acceptance criteria for OD Dimensional Verification was met.
    Dimensional Verification - Working LengthThe acceptance criteria for Working Length Dimensional Verification was met.
    Dimensional Verification - Tip LengthThe acceptance criteria for Tip Length Dimensional Verification were met.
    Kink ResistanceAll units met the Kink Resistance acceptance criteria.
    Visual Inspection (Tip Taper/Transitions)All units met the Tip Taper/Transition Visual Inspection acceptance criteria.
    Shape Retention (Catheter Tip Shape)All units met the Shape Retention Acceptance criteria.
    Burst Test (per ISO 10555-1)All units met the Catheter Burst acceptance criteria.
    Liquid Leak Test (per ISO 10555-1)All units met the Liquid Leak acceptance criteria.
    Air Leak Test (per ISO 80369-7)All units met the Air Leak Test acceptance criteria.
    Chemical Compatibility (with saline, dextrose, heparin, radiocontrast)All units met the Chemical Compatibility acceptance criteria.
    Corrosion (per ISO 10555-1 Annex A)All units met the Corrosion acceptance criteria.
    Packaging - Visual InspectionAll units met the Packaging Visual Inspection.
    Packaging - Pouch Leak (per ASTM F-1929-15)All units met the Pouch Leak test.
    Packaging - Pouch Peel (per ASTM F-88/F88M-15)All units met the Pouch Peel test.
    Packaging - Seal WidthAll seals met the acceptance criteria for Seal Width.
    Comparative Testing
    Friction Force (in simulated-use anatomical model)The RIST Radial Access Catheter and predicate device had comparable friction forces.
    Design Validation Testing
    In vitro Simulated Use Study - Bench (performance by physicians in simulated procedure, compared to competitive device)All acceptance criteria were met. The performance was found to be substantially equivalent.
    Radiopacity (physician identification of distal tip, shape, curves from fluoroscopic images)All acceptance criteria were met. The performance was found to be substantially equivalent.
    In vitro Simulated Use Study - Usability (evaluators representative of intended user population, per IFU)All acceptance criteria were met. The catheter was able to be used as per the Instruction for Use.
    Leveraged Testing (from previous submissions)
    Hub Compatibility (Catheter luers per ISO 594)All units met the Hub Compatibility acceptance criteria.
    Labeling Legibility - LabelThe acceptance criteria for Labeling Legibility was met.
    BarcodeThe acceptance criteria for Barcode was met.
    Labeling Legibility - IFUThe acceptance criteria for IFU Legibility was met.
    Sterilization (per ANSI/AAMI/ISO 11135:2014 and AAMI TIR 28:2016)Product was sterile (achieved SAL of at least 10⁻⁶).
    Biocompatibility Testing (per EN ISO 10993-1)
    Cytotoxicity (ISO MEM Elution)The test article is considered non-cytotoxic.
    Sensitization (ISO Guinea Pig Maximization Sensitization)The test article did not elicit a sensitization response; classified as non-sensitizer.
    Irritation (ISO Intracutaneous Irritation)No significant dermal reactions observed; non-irritant.
    Systemic Toxicity (ISO Acute Systemic Injection)No abnormal clinical signs indicative of toxicity; non-cytotoxic (conclusion oddly repeats cytotoxicity).
    Complement Activation, SC5b-9 AssaySC5b-9 assay results not statistically significant when compared to reference material and comparison.
    Hemocompatibility (ASTM Hemolysis - Direct Contact and Extract Method)Test article returned a blank corrected percent hemolysis above the negative control of 0.0% (interpreted as non-hemolytic based on conclusion).
    Pyrogenicity (Materials Mediated Rabbit Pyrogen)Test article extracts did not cause a pyrogenic response; non-pyrogenic.

    Study Details (for a physical catheter, not AI/ML):

    1. Sample sizes used for the test set and the data provenance: The document generally states "All units" or "Samples" without specifying the exact number of units tested for each criterion. The data provenance is internal RIST Neurovascular testing, leveraging some data from previously cleared devices (K152202, K161262, K152876). The studies are evidently prospective, laboratory/benchtop tests, not clinical studies on human subjects, hence "country of origin of the data" in a clinical sense is not applicable.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: For the In vitro Simulated Use Study - Bench and Radiopacity tests, "physicians" were used. For the In vitro Simulated Use Study - Usability, "evaluators representative of the intended user population" were used. No specific number or qualifications (e.g., years of experience, specialty) of these experts are provided in the summary.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable and not mentioned, as these are primarily objective engineering performance tests rather than subjective human assessments requiring adjudication. For the tests involving physicians/evaluators, specific adjudication methods are not detailed.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done: No, an MRMC study was not done, as this is a physical medical device (catheter), not an AI/ML system being evaluated for diagnostic accuracy improvement.
    5. If a standalone performance was done: Yes, the described performance tests (e.g., tensile strength, torque strength, dimensional verification, kink resistance, burst, leak tests, biocompatibility, sterilization) were performed directly on the RIST Radial Access Catheter in a standalone manner to demonstrate its inherent properties and safety.
    6. The type of ground truth used: For most engineering tests, the ground truth is established by the specified ISO/ASTM standards and internal acceptance criteria (e.g., a specific pressure for burst test, a specific force for tensile strength). For the simulated use and radiopacity tests, the 'ground truth' is the observable, functional performance and perception by the evaluating physicians/users, often compared to the predicate device. Biocompatibility used standard biological test methods.
    7. The sample size for the training set: Not applicable. This is a physical device submission, not an AI/ML algorithm that requires a training set.
    8. How the ground truth for the training set was established: Not applicable, as there is no training set for a physical device.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1