Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K190898

    Validate with FDA (Live)

    Device Name
    Sight OLO
    Date Cleared
    2019-11-01

    (210 days)

    Product Code
    Regulation Number
    864.5220
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Sight OLO is a quantitative multi-parameter automated hematology analyzer intended for in vitro diagnostic use in screening capillary or venous whole blood samples collected in KyEDTA blood collection tubes, or fingertip samples collected using the Sight OLO test kit micro-capillary tubes.

    When used with the Sight OLO cartridge, the Sight OLO enumerates the following CBC parameters in whole blood: WBC, RBC, HGB, HCT, MCV, MCH, RDW, PLT, NEUT%#, LYMPH %#, MONO %#, EOS%#, and BASO%/#.

    The Sight OLO is indicated for use in clinical laboratories to identify and classify one or more of the formed elements of blood in children 3 months and above, adolescents and adults.

    Device Description

    The Sight OLO device is a computer vision based platform for blood analysis. The platform combines computer-vision algorithms for image processing to identify and quantify blood components (e.g., red blood cells) and their characteristics (e.g., cell volume) in an automated fashion. Using dedicated staining, the proposed platform provides complete blood count analysis. The Sight OLO is a compact device, designed to be automated and simple to operate, to enable rapid testing and analysis. The Sight OLO consists of a scanning and analyzing device and a CBC test kit, including disposable cartridges and sample preparation tools. The disposable cartridge containing the blood sample is loaded into the device through the loading slot. The device is operated through the touch screen interface.

    The Sight OLO provides complete blood count information with 5-part differentials for white blood cell types. Specifically, the CBC parameters measured by the Sight OLO are listed below and include: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, PLT, NEUT%#, LYMPH %#, MONO %/#, EOS%# and BASO%#. In addition, the Sight OLO signals specific WBC abnormal cases by flagging the sample.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Study CategoryMeasurandAcceptance CriteriaReported Device Performance
    Method ComparisonWBC, RBC, PLT, HGB, HCT, MCV, RDW, MCH, MCHC, NEUT%, NEUT#, LYMPH%, LYMPH#, MONO%, MONO#, EOS%, EOS#, BASO%, BASO#Correlation, bias, slope, and intercept (and their 95% two-sided confidence intervals) met pre-specified criteria.All measurands met the pre-specified acceptance criteria for correlation, bias, slope, intercept (and the 95% two-sided confidence interval (CI) around the slope and intercept). (See Table 1 for specific values for each measurand).
    RepeatabilityAll measurandsAll measurands in all tested ranges met predefined acceptance criteria for mean, standard deviation (SD), and coefficient of variation (CV).All measurands in all tested ranges met the predefined acceptance criteria. (See Table 2 for specific values for each measurand across target ranges).
    ReproducibilityAll measurandsAll components of variation (within-run, between-run, between-day, between-instrument, between-site, total precision) met predefined acceptance criteria for SD and %CV.All components of variation that were calculated met the pre-defined acceptance criteria. (See Table 3 for specific values for each measurand across low, normal, and high levels).
    Detection Limits (LoB, LoD, LoQ)WBC, PLT, HGBLoB defined as 95th percentile of study variable. LoD determined mathematically. LoQ defined as lowest concentration (≥LoD) where total error (TE) accuracy goals were satisfied. Test results met these definitions.WBC: LoB = 0.04 x 10³/μL, LoD = 0.17 x 10³/μL, LoQ = 0.18 x 10³/μLPLT: LoB = 8.00 x 10³/μL, LoD = 11.00 x 10³/μL, LoQ = 13.40 x 10³/μLHGB: LoB = 0 g/dL, LoD = 3.9 g/dL, LoQ = 3.9 g/dL. (All met acceptance criteria).
    LinearityHGB, PLT, WBCThe final linearity range was determined as the intersection of the results from three instruments, and no lower than the LoD.HGB: (3.9-21.75) g/dLPLT: (18-1028.5) x 10³/μLWBC: (0.18-100.13) x 10³/μL. (All met acceptance criteria).
    Analytical Specificity/InterferenceRBC, WBC, HGB, HCT, PLTNo significant interference for D-Glucose, Bilirubin F, Bilirubin C, Chyle, Hemolytic Hemoglobin, Intralipids at specified concentrations. No interference from RBC fragments, high leukocytosis, and high thrombocytosis.Results demonstrated no significant interference from D-Glucose, Bilirubin F (except for WBC above 8.86 mg/dL), Bilirubin C, Chyle (except for PLT above 530 FTU), Hemolytic Hemoglobin, and Intralipids within tested concentrations. No interference from abnormal specimen conditions.
    Sample StabilityAll measurandsSupports a sample stability claim of 8 hours from blood collection at room temperature, with RDW specifically 4 hours.Data support a sample stability claim of 8 hours from blood collection at room temperature for all Sight OLO measurands, with RDW having a stability of 4 hours.
    Test Kit Shelf-LifeFunctional and analytical testsMet predefined acceptance criteria at defined time points. Initial shelf life of 6 months.Functional and analytical tests were conducted and results met the predefined acceptance criteria. Initial shelf life set at 6 months. Ongoing study with 3 lots (4, 8, 12, 18, 24, 26, 30, 36, 38 months).
    Transportation StabilityDevice and Test KitDevice maintains calibration and functional specifications. Test kit met all acceptance criteria for functionality and expected analytical results.The Sight OLO device maintains its calibration and functional specifications. The Sight OLO test kit met all acceptance criteria for functionality and expected analytical results.
    Flagging StudySensitivity and SpecificityMet predefined acceptance criteria.Sensitivity (PPA): 93.0%Specificity (NPA): 80.6%Overall Agreement: 86.5% (Met predefined acceptance criteria).
    Adult and Pediatric Reference IntervalsAll measurandsAdults: Reference intervals calculated for each measurand. Pediatrics: Supports validity of pre-established pediatric reference intervals.Adult reference intervals were calculated. Results supported the validity of pre-established pediatric reference intervals.
    Matrix Comparison (Capillary vs. Venous)All applicable measurandsComparable performance characteristics for capillary and venous whole blood specimens.The results show comparable performance characteristics for capillary and venous whole blood specimens.
    Matrix Comparison (Capillary Microtube vs. Sight OLO Test Kit Micro-capillaries)All applicable measurandsComparable performance between K2EDTA anticoagulated microtube capillary samples and 2-drop fingertip samples.The results of the regression and bias analysis showed comparable performance between the K2EDTA anticoagulated microtube capillary samples and the 2-drop fingertip samples.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Method Comparison Study: 679 residual clinical K₂EDTA whole blood samples.
      • Provenance: Collected from both adults (≥22 years old) and pediatric patients (3 months to 21 years old) at three (3) US sites. The majority were venous samples; a few pediatric samples were capillary whole blood. The study included normal and pathological samples, covering an age range of 3 months to 94 years old. 32% were pediatric samples. Consisted of 355 males (52%) and 324 females (48%). This was a retrospective study using residual clinical samples.
    • Repeatability Study: Minimum of 11 samples per site, covering the laboratory reference range and medical decision levels for HGB, PLT, WBC, and upper range for RBC, HGB, WBC, and PLT.
      • Provenance: Residual K₂EDTA whole blood samples. Performed at 3 US sites.
    • Reproducibility Study: 3 lots of commercial control material (low, normal, high concentrations). For a total of 240 measurements per level of control material (presumably 4 replicates x 2 runs/day x 5 days x 2 instruments/site x 3 sites, though the exact breakdown to reach 240/level is not fully detailed).
      • Provenance: Commercial control material. Conducted at 3 sites, including 2 US sites.
    • Detection Limits (LoB, LoD, LoQ):
      • LoB: Preserved RBC samples (WBC and PLT), diluent (HGB). Two Sight OLO instruments, two reagent lots, 60 measurements per instrument.
      • LoD and LoQ: Six low concentration samples (manipulated concentrated and diluted venous blood samples), each measured 10 times, for a total of 60 repeated measurements per test. This was repeated for a different test reagent lot (another 60 measurements).
      • Provenance: Manipulated venous blood samples for LoD/LoQ.
    • Linearity Studies: Ten (10) venous whole blood samples manipulated to create linearity panels. Seven concentrations for WBC and PLT, ten for RBC and HGB. Each concentration scanned in duplicate. Repeated on three OLO devices.
    • Analytical Specificity/Interference Study: Not explicitly stated as a sample size of patient samples, but implied various samples manipulated with interferents.
    • Sample Stability Study: 10 freshly collected venous whole blood samples (7 normal, 3 around medical decision levels).
    • Test Kit Shelf-Life: Three lots of test kits.
    • Transportation Studies: Not specified in terms of sample size, but indicates testing of the physical device and test kits.
    • Flagging Study: 208 samples.
      • Provenance: Collected at 3 clinical study sites.
    • Adult Reference Intervals Study: 240 (120 male and 120 female) apparently healthy adults.
      • Provenance: K₂EDTA venous whole blood samples collected from apparently healthy adults (≥22 years) at a single US site. This was a prospective study.
    • Pediatric Reference Intervals Study: 80 apparently healthy pediatric subjects (20 per subpopulation: baby (3-23 months), child (2-<12 years), adolescent (12-<18 years), transitional adolescent (18-<22 years)).
      • Provenance: Residual blood samples collected from apparently healthy pediatric subjects. This was a retrospective study.
    • Capillary vs. Venous Matrix Comparison: 67 normal and pathological paired capillary and venous whole blood specimens.
      • Provenance: Paired specimens collected from the same individuals.
    • Capillary Microtube vs. Sight OLO Test Kit Micro-capillaries Matrix Comparison: 40 paired K₂EDTA anticoagulated capillary microtube samples and 2-drop finger prick samples.
      • Provenance: Collected from healthy volunteer subjects at the company's internal laboratory.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    • Flagging Study:
      • Number of Experts: Two qualified morphology examiners (initially). A third qualified morphology examiner was used for adjudication.
      • Qualifications: "Qualified morphology examiners" (specific experience level not provided in the summary).

    4. Adjudication Method for the Test Set:

    • Flagging Study: A 2+1 adjudication method was used. Two qualified morphology examiners evaluated one of the three blood films for each sample. In case of disagreement between the first two readers (who performed a 200-cell count each, totaling 400 cells), a third qualified morphology examiner reviewed the third blood film.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No MRMC comparative effectiveness study was done. The studies compared the Sight OLO device's performance against a predicate device (Sysmex XN-Series) or manual microscopy (for flagging), but not a direct comparison of human readers with and without AI assistance. The study focuses on the performance of the device itself.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

    • Yes, the performance studies described (Method Comparison, Repeatability, Reproducibility, Detection Limits, Linearity, Analytical Specificity, Stability) are all standalone performance evaluations of the Sight OLO device. It tests the accuracy and precision of the automated hematology analyzer without human intervention in the CBC parameter enumeration process.
    • The Flagging Study involves a comparison to manual light microscopy, which is a human-read method, to establish agreement in flagging abnormal samples. While this involves human input for comparison, the Sight OLO's flagging itself is an automated algorithm-only output being evaluated.

    7. Type of Ground Truth Used:

    • Method Comparison: Predicate device (Sysmex XN-Series) results.
    • Repeatability/Reproducibility/Detection Limits/Linearity/Analytical Specificity/Stability: Internal reference methods, commercial control materials, or manipulated samples (e.g., concentrated/diluted blood samples), and comparison to expected analytical results.
    • Flagging Study: Manual light microscopy by qualified morphology examiners.
    • Adult and Pediatric Reference Intervals: Literature reference limits (for verification in pediatric study), and statistical calculation from healthy population samples.
    • Matrix Comparison: Paired samples compared to each other on the OLO device.

    8. Sample Size for the Training Set:

    • This information is not provided in the given 510(k) summary. The document focuses on performance testing (validation) datasets.

    9. How the Ground Truth for the Training Set Was Established:

    • This information is not provided in the given 510(k) summary, as it does not detail the development or training of the device's algorithms.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1