Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K181414

    Validate with FDA (Live)

    Device Name
    uCT 530, uCT 550
    Date Cleared
    2018-08-14

    (76 days)

    Product Code
    Regulation Number
    892.1750
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The uCT Computed Tomography X-ray System uCT 530/550 is a computed tomography x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes and indicated for the whole body (including head, neck, vascular).

    Device Description

    The uCT 530/uCT 550 is a multi-slice X-ray computed tomography scanner which features a continuously rotating tube-detector system and functions according to the fan beam principle. The system provides the filter back-projection (FBP) algorithm to reconstruct images in DICOM format. which can be used by post-processing applications.

    The system consists of the Gantry, X-ray System, Data Management System, Patient Table, Console, Power Supply Cabinet, Image Processing Computer, and Software. The system software is a program used for patient management, data management, Xray scan control, image reconstruction, and image archive.

    A motorized patient table moves the patient through a circular opening in the Gantry. As the patient passes through the Gantry, a source of x rays rotates around the inside of the circular opening. Detectors on the exit side of the patient record the X rays exiting the section of the patient's body being irradiated as an X-ray "snapshot". Many different "snapshots" (angles) are collected during one complete rotation. The data are sent to a computer to reconstruct all of the individual "snapshots" into a crosssectional image (slice) of the internal organs and tissues for each complete rotation of the source of x rays.

    There are two features for denoising and reduce metal artifact, which are KARL iterative denoising reconstruction algorithm and MAC Metal artifact correction algorithm.

    AI/ML Overview

    The provided text does not contain acceptance criteria or a detailed study that proves the device meets specific acceptance criteria in the format requested.

    Instead, the document is a 510(k) summary for a Computed Tomography X-ray System (uCT 530, uCT 550) seeking FDA clearance based on substantial equivalence to a predicate device (uCT 760, uCT 780).

    Here's what can be extracted and inferred from the text regarding performance, though it doesn't meet all your requested points:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document provides a comparison table of technological characteristics between the proposed device (uCT 530, uCT 550) and a predicate device (uCT 760, uCT 780). While not explicitly "acceptance criteria" against numerical targets, it implies that comparable performance to the predicate is the acceptance criterion for substantial equivalence.

    ITEMAcceptance Criteria (Implicit: Similar to Predicate)Reported Device Performance (uCT 530, uCT 550)Notes/Differences from Predicate
    Intended UseThe uCT Computed Tomography X-ray System is a computed tomography x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes and indicated for the whole body (including head, neck, cardiac and vascular).The uCT Computed Tomography X-ray System uCT530/uCT550 is a computed tomography x-ray system intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes and indicated for the whole body (including head, neck, vascular).Note 1: Does not include cardiac, but states this does not affect safety and effectiveness.
    Product CodeJAKJAKSame
    Regulation No.21 CFR 892.175021 CFR 892.1750Same
    ClassIIIISame
    Scan RegimeContinuous RotationContinuous RotationSame
    Scan ModesScout, Axial Scan, Helical ScanScout, Axial Scan, Helical ScanSame
    Detector MaterialSolid-state GOSSolid-state GOSSame
    Z-plane coverage40mm22mmNote 2: Shorter coverage induces longer scanning time, but does not affect safety and effectiveness.
    Size of detector element in Z-plane0.5mm0.55mmNote 3: Bigger minimum element size induces lower spatial resolution, but does not affect safety and effectiveness.
    Number of element per row936864Note 4: Smaller number of elements induces less data sampling but provides sufficient data for reconstruction, not affecting safety and effectiveness.
    Number of detector row8040Note 5: Determined by Z-plane coverage and element size, not affecting safety and effectiveness.
    Maximum slices generated per rotation128 for uCT 760, 160 for uCT 78040 for uCT 530, 80 for uCT 550Note 6: Smaller slice number induces longer scanning time, but does not affect safety and effectiveness.
    Minimum slice thickness0.625mm for uCT 760, 0.5mm for uCT 7800.55mmNote 7: Determined by detector element size, not affecting safety and effectiveness.
    Maximum sampling rateUp to 4800 views per 360°Up to 4800 views per 360°Same
    Tube anode storage capacity7.5MHU5.3MHUNote 8: Lower capacity affects continuous scans' throughput, but not independent scans' effect.
    Maximum cooling rate1386 kHU/min815 kHU/minNote 9: Lower rate affects continuous scans' throughput, but not independent scans' effect.
    Focal spot size0.7x0.7mm, 1.0x1.0mm0.5x1.0mm, 1.0x1.0mmNote 10: Smaller size helps resolution, but image spatial resolution is "equivalent substantially."
    Power80kW for uCT 760, 100 kW for uCT 78050kWNote 11: Smaller power output induces lower ability of x-ray penetration for high BMI, but safety evaluated.
    mA Range6-667mA for uCT 760, 6-833mA for uCT 78010-420mANote 12: Smaller mA output induces lower ability of x-ray penetration for high BMI, but safety evaluated.
    kV Settings70, 80, 100, 120, 14070, 80, 100, 120, 140Same
    Aperture700mm700mmSame
    Rotation speedUp to 0.35 sec per 360° rotation for uCT 760, Up to 0.3 sec per 360° rotation for uCT 780Up to 0.5 sec per 360° rotationNote 13: Slower rotation speed induces longer scan time, but does not affect safety and effectiveness.
    Gantry Tilt± 30° with 0.5 increment± 30° with 0.5 incrementSame
    Scannable range1700 mm1700 mmSame
    Horizontal motion range2180 mm2180 mmSame
    Table Horizontal SpeedUp to 200mm/secUp to 200mm/secSame
    Vertical motion range480 mm-950 mm from the floor480 mm-950 mm from the floorSame
    Vertical speedUp to 40 mm/secUp to 40 mm/secSame
    Table Horizontal Position accuracy± 0.25mm± 0.25mmSame
    Table Maximum table load205kg205kgSame
    Image Spatial ResolutionHigh mode: >20 lp/cm @ MTF 0%, 16.5 ± 1.7 lp/cm @ MTF10%, 11.5 ± 1.2 lp/cm @ MTF50%High mode: >20 lp/cm @ MTF 0%, 16.5 ± 1.7 lp/cm @ MTF10%, 11.5 ± 1.2 lp/cm @ MTF50%Same
    Image Noise3.0 ± 0.5 HU at 120 kV, 5 mm slice thickness, CTDIvol 29.1mGy3.0 ± 0.5 HU at 120 kV, 5 mm slice thickness, CTDIvol 28.9 mGyNote 14: Noise level is "equivalent substantially" considering slightly different CTDIvol.
    CT Number Display Range-1024 ~+8191 HU-1024 ~+8191 HUSame
    Scan Field of ViewUp to 500 mm, 600mm with extend FOVUp to 500 mm, 600mm with extend FOVSame
    Reconstruction Field of View40mm-500mm, 40mm-600mm with extend FOV40mm-500mm, 40mm-600mm with extend FOVSame
    Maximum scannable length1700mm1700mmSame
    Image MatrixUp to 1024 x 1024Up to 1024 x 1024Same
    Reconstructed slice thicknessuCT 760: 0.625mm,1.25mm,2.5mm,5mm,10mm (axial), 0.625-10mm(helical); uCT 780: 0.5mm,0.625mm,1.25mm,2.5mm,5mm,10mm (axial), 0.5-10mm (helical)0.55mm,1.1mm,2.2mm,5.5mm,11mm (axial), 0.55-10mm(helical)Note 15: More slice thickness options for predicate, but similar capabilities overall, not affecting safety/effectiveness.
    Pitch0.1~2.00.1~2.0Same
    Maximum continuous exposure timeUp to 100secondsUp to 100secondsSame
    Electrical SafetyComply with ES60601-1Comply with ES60601-1Same
    EMCComply with IEC60601-1-2Comply with IEC60601-1-2Same
    BiocompatibilityPatient Contact Materials were tested and demonstrated no cytotoxicity (ISO 10993-5), no evidence for irritation and sensitization (ISO 10993-10).Patient Contact Materials were tested and demonstrated no cytotoxicity (ISO 10993-5), no evidence for irritation and sensitization (ISO 10993-10).Same
    Iterative noise reductionKARL 3DKARL 3DSame
    Adaptive FilterAdaptive FilterAdaptive FilterSame
    Metal artifact reductionMACMACSame

    2. Sample Size Used for the Test Set and Data Provenance

    The document mentions "Clinical Evaluation for sample clinical images evaluation" and "sample clinical images for proposed devices are provided in Section 36 Clinical Evaluation. Electronic file for each image are provide in MISC Folder." However, it does not specify the sample size of the test set for image evaluation or the data provenance (e.g., country of origin, retrospective/prospective). It only refers to "sample clinical images."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    The document does not provide information on the number of experts used, their qualifications, or how ground truth was established for the "sample clinical images."

    4. Adjudication Method for the Test Set

    The document does not describe any adjudication method for a test set.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    The document does not mention a MRMC comparative effectiveness study or any effect size of AI assistance for human readers. This device is a CT scanner, not an AI-assisted interpretation tool in the context of this submission. The "KARL iterative denoising reconstruction algorithm and MAC Metal artifact correction algorithm" are features of the CT system itself, not necessarily AI for reader improvement.

    6. Standalone (Algorithm Only Without Human-in-the-Loop) Performance

    As mentioned above, the device is a CT scanner. Its performance verification focuses on technical specifications (e.g., Image Spatial Resolution, Image Noise). The "KARL iterative denoising reconstruction algorithm" and "MAC Metal artifact correction algorithm" are integral parts of the image reconstruction process, not standalone interpretive algorithms. Therefore, a separate "standalone" performance for an interpretive algorithm is not applicable in the context described.

    7. Type of Ground Truth Used

    For the "sample clinical images evaluation," the type of ground truth is not specified. It refers to "sample clinical images," implying they are real patient images, but how their "truth" was confirmed (e.g., pathology, subsequent expert consensus) is not detailed. For the technical performance parameters (e.g., image noise, spatial resolution), the "ground truth" is typically established through phantom studies and objective measurements according to industry standards.

    8. Sample Size for the Training Set

    The document does not mention a training set sample size. This submission is for a CT scanner and its image reconstruction algorithms, not a deep learning AI model that typically requires a large training set of labeled data for classification or detection tasks. While iterative reconstruction algorithms have parameters that might be "trained" or optimized, the document does not discuss this in the context of a "training set" of patient images.

    9. How the Ground Truth for the Training Set Was Established

    Since no training set and its ground truth are mentioned (as per point 8), this information is not provided.

    In summary: The provided document is a 510(k) summary focused on demonstrating substantial equivalence of a new CT system to a predicate device based on technical specifications and functional capabilities. It does not contain the detailed clinical study design, acceptance criteria (beyond substantial equivalence to predicate), and performance metrics typically associated with AI-driven diagnostic software, especially concerning human reader performance or defined ground truth for clinical datasets. The "Performance Verification" section lists standards compliance and "Clinical Evaluation for sample clinical images evaluation" but lacks the specifics requested.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1