Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K180476

    Validate with FDA (Live)

    Date Cleared
    2018-05-15

    (82 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim (Green) is a specialty medical glove which is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between examiner and patient. In addition, these gloves are worn to protect the wearer against exposure to chemotherapy drugs. Tested for use chemotherapy drugs. Tested drugs are as follows:

    Average Breakthrough Detection Time (minutes)
    Cisplatin 1.0 mg/ml≥ 240
    Cyclophosphamide (Cytoxan) 20 mg/ml≥ 240
    Dacarbazine (DTIC) 10.0 mg/ml≥ 240
    Doxorubicin Hydrochloride 2.0 mg/ml≥ 240
    Etoposide (Toposar) 20.0 mg/ml≥ 240
    Fluorouracil 50.0 mg/ml≥ 240
    Paclitaxel (Taxol) 6.0 mg/ml≥ 240

    Please note that the following drugs have extremely low permeation time for
    *Carmustine (BCNU) 3.3mg.ml - 23.4 (mins)
    *Thiotepa 10.0 mg/ml - 16.2 (mins)

    Device Description

    The proposed device is the Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim (Green). The proposed device is for single use and is provided non-sterile. The device meets the requirements for dimensions, physical properties, and thickness in ASTM D6319 - 10(2015) Standard Specification for Nitrile Examination Gloves for Medical Application. The device accomplishes its intended function by creating a physical barrier between the examiner and patient and creates a barrier for the chemotherapy drugs listed in the Indications for Use.

    AI/ML Overview

    This document is a 510(k) Premarket Notification for Powder Free Nitrile Examination Gloves tested for use with chemotherapy drugs. It focuses on demonstrating substantial equivalence to a legally marketed predicate device.

    Here's the information extracted and organized according to your request:

    1. Table of Acceptance Criteria and Reported Device Performance

    CharacteristicAcceptance Criteria (Standard)Reported Device Performance (New Device - K180476)Comparison
    Chemotherapy Drugs Permeation (Minimum Breakthrough Detection Time)ASTM D6978-05
    Cisplatin 1.0 mg/mlN/A (Comparative to predicate)>240 minutesSimilar to predicate
    Cyclophosphamide (Cytoxan) 20 mg/mlN/A (Comparative to predicate)>240 minutesSimilar to predicate
    Dacarbazine (DTIC) 10.0 mg/mlN/A (Comparative to predicate)>240 minutesSimilar to predicate
    Doxorubicin Hydrochloride 2.0 mg/mlN/A (Comparative to predicate)>240 minutesSimilar to predicate
    Etoposide (Toposar) 20.0 mg/mlN/A (Comparative to predicate)>240 minutesSimilar to predicate
    Fluorouracil 50.0 mg/mlN/A (Comparative to predicate)>240 minutesSimilar to predicate
    Paclitaxel (Taxol) 6.0 mg/mlN/A (Comparative to predicate)>240 minutesSimilar to predicate
    *Carmustine (BCNU) 3.3 mg/mlN/A (Comparative to predicate, with warning)23.4 minutesSimilar to predicate, with warning
    *Thiotepa 10.0 mg/mlN/A (Comparative to predicate, with warning)16.2 minutesSimilar to predicate, with warning
    DimensionsASTM D6319 - 10(2015)Length - Min 240mm; Thickness palm and finger - Min 0.05mmSame
    Physical PropertiesASTM D6319 - 10(2015)MeetsSame
    Thickness (Finger & Palm)ASTM D6319 - 10(2015)MeetsSame
    Powder ContentASTM D6124 - 06(2011) (≤ 2 mg/glove)MeetsSame
    Biocompatibility: Primary Skin IrritationISO 10993-10:2010Passes (non-irritating)Same
    Biocompatibility: Dermal SensitizationISO 10993-10:2010Passes (non-sensitization)Same
    Biocompatibility: Cytotoxicity and acute systemic toxicityISO 10993-5:2009 & ISO 10993-5:2017Passes (Certificate of Analysis for color imparting compound indicates non-cytotoxic, not systemically toxic)Different (Predicate N/A for this test)
    Watertight (1000ml)21 CFR 800.20, ASTM D5151PassesSame
    Indication for UseN/A (Comparative to predicate)Specialty medical glove for medical purpose, worn on examiner's hand/finger to prevent contamination between examiner and patient, and to protect against exposure to listed chemotherapy drugs.Similar
    MaterialASTM D6319 - 10(2015)NitrileSame
    ColorN/AGreenDifferent (Predicate: Blue or White)
    SizeMedical Glove Guidance Manual - LabelingExtra Small, Small, Medium, Large, Extra LargeSame
    Single UseMedical Glove Guidance Manual - LabelingSingle UseSame

    The acceptance criterion for the chemotherapy drug permeation test appears to be established by comparison to the predicate devices. For most drugs, the criterion is a minimum breakthrough detection time of ">240 minutes". For Carmustine and Thiotepa, the low permeation times observed for both the new device and the predicate lead to a warning statement, indicating that these drugs have "extremely low permeation times" and are not held to the ">240 minutes" criterion.

    2. Sample size used for the test set and the data provenance:

    • The document does not explicitly state the sample size used for the test set for properties like chemotherapy drug permeation, physical properties, or biocompatibility. It only reports "meets" or specific breakthrough times, implying that tests were conducted according to relevant standards (e.g., ASTM D6978-05 for chemotherapy drug permeation and ASTM D6319-10(2015) for physical properties and dimensions). These standards typically define sampling plans.
    • Data Provenance: The manufacturer is Comfort Rubber Gloves Industries Sdn. Bhd., located in Matang, Perak, Malaysia. The document does not specify if the testing was performed in Malaysia or another country. The data is non-clinical performance data, likely from prospective testing conducted to support the 510(k) submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This information is not applicable as the device is a medical glove, and its performance is evaluated through objective, standardized laboratory tests against established physical, chemical, and biological criteria, not through expert interpretation of images or patient data to establish a ground truth.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • This is not applicable. The evaluation of medical gloves against performance standards (e.g., ASTM, ISO) involves quantitative measurements and predefined pass/fail criteria, not subjective interpretation requiring adjudication by experts.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • This is not applicable. The device is a physical medical glove, not an AI-powered diagnostic or assistive technology. Therefore, no MRMC study or AI-related effectiveness assessment was conducted or reported.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • This is not applicable. The device is a physical medical glove and does not involve any algorithm or AI component.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The "ground truth" for this device is based on objective, standardized test methods and criteria defined by international and national standards (e.g., ASTM D6319, ASTM D6124, ASTM D6978, ISO 10993, 21 CFR 800.20). These standards establish the acceptable performance characteristics for medical gloves. For biocompatibility, it's based on "passes" criteria of ISO standards. For chemotherapy permeation, it's the objectively measured breakthrough detection time.

    8. The sample size for the training set:

    • This is not applicable. The device is a physical medical glove and does not utilize a "training set" in the context of machine learning or AI. Its design and manufacturing are based on engineering principles and material science, not data-driven model training.

    9. How the ground truth for the training set was established:

    • This is not applicable for the reasons stated in point 8.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1