510(k) Data Aggregation

    K Number
    K153225

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-06-21

    (228 days)

    Product Code
    Regulation Number
    870.2300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Masimo Root Monitoring System and Accessories are indicated for use by healthcare professionals for the monitoring of multiple physiological parameters in healthcare environments.

    The Masimo Root Monitoring System and Accessories can transmit data for supplemental remote viewing and alarming (e.g., at a central station).

    The optional Masimo Radical-7 Pulse CO-Oximeter and Accessories are indicated for the continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (SpHb), and/or respiratory rate (RRa). The Masimo Radical-7 Pulse CO-Oximeter and accessories are indicated for use with adult, pediatric, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused in hospital-type facilities, mobile, and home environments. In addition, the Masimo Radical-7 Pulse CO-Oximeter and accessories are indicated to provide the continuous non-invasive monitoring data obtained from the Masimo Radical-7 Pulse CO-Oximeter and accessories of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate to multi-parameter devices for the display of those devices.

    The optional Masimo Radius-7 Wearable Pulse CO-Oximeter and Accessories are indicated for the continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (SpHb), and/or respiratory rate (RRa). The Masimo Radius-7 Wearable Pulse CO-Oximeter and accessories are indicated for use with adult and pediatric patients during both no motion conditions, and for patients who are well or poorly perfused in hospitals and hospital-type facilities.

    The optional ISA product family consists of three types of sidestream gas analyzers (ISA CO2, ISA AX+ and ISA OR+), intended to be connected to other medical backboard devices for monitoring of breath rate and the following breathing gases:

    ISA CO2: CO2

    ISA AX+: CO2, N2O, Halothane, Isoflurane, Enflurane, Sevoflurane and Desflurane

    ISA OR+: CO2, O2, N2O, Halothane, Isoflurane, Enflurane, Sevoflurane and Desflurane

    ISA CO2, ISA AX+ and ISA OR+ are intended to be connected to a patient breathing circuit for monitoring of inspired/expired gases during anesthesia, recovery and respiratory care. The intended environment is the operating suite, intensive care unit and patient room. ISA CO2 is also intended to be used in road ambulances. The intended patient population is adult, pediatric and infant patients.

    The optional SEDLine Sedation Monitor is indicated for use in the operating room (OR), intensive care unit (ICU), and clinical research laboratory. It is intended to monitor the state of the brain by real-time data acquisition and processing of EEG signals. The system includes the Patient State Index (PSI), a proprietary computed EEG variable that is related to the effect of anesthetic agents.

    Device Description

    The Root Monitoring System (Root) is a multifunctional device that monitors vital signs of patients from neonates to adults. Parameters monitored by Root include non-invasive functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (SpHb), respiratory rate (RRa), inspired/expired gases during anesthesia, recovery and respiratory care, state of the brain by real-time data acquisition and processing of EEG signals, and Patient State Index (PSI) which is an EEG variable that is related to the effect of anesthetic agents.

    Root is intended to be used with the previously FDA cleared measurement technologies for the modules of:

    • . Masimo Radical-7 Pulse CO-Oximeter (Radical-7 module), with cleared technologies of SpO2, pulse rate, SpCO, SpMet, SpHb and RRa monitoring per K110028.
    • Masimo Radius-7 Pulse Oximeter (Radius-7 module), with cleared technologies of ● SpO2, pulse rate, SpCO, SpMet, SpHb and RRa monitoring per K110028.
    • . ISA-Infrared Sidestream Gas Analyzer (ISA module), with cleared technologies of breathing gases and respiratory rate monitoring per K103604.
    • Sedline Sedation Monitor with Frontal PSI and SEDTrace EEG Electrode Set ● (Sedline module), with cleared technologies of EEG and PSI monitoring per K051874.

    Root is intended to be used as an alternative user interface to facilitate access control and monitoring device functions and to connect system networks such as the Patient SafetyNet (K071047).

    Root displays patient monitoring information from the connected modules. Visual alarms are shown on the Root display and audible alarms are generated through the Root internal speaker. When the module is disconnected from Root, the monitoring information from the module is no longer displayed on Root.

    Data from connected modules, including patient monitoring data, can be communicated to network systems. Root also functions as a pass-through means for communicating information between connected devices and network systems.

    The predicate device, Masimo Root Monitoring System (Root) was cleared in K142394, is the same as the subject device, Masimo Root Monitoring System (Root). Both the predicate and subject devices include the option to connect the Masimo Radius-7 Pulse Oximeter (Radius-7) module. The main difference is that the Radius-7 in the subject device now provides the same measurement functionality as the Radical-7 module.

    AI/ML Overview

    The Masimo Root Monitoring System and Accessories is a multifunctional device designed for monitoring multiple physiological parameters in healthcare environments.

    Here's an analysis of its acceptance criteria and supporting studies based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document describes performance in terms of compliance with various standards and successful completion of verification and validation activities. It doesn't list specific quantitative acceptance criteria for each physiological parameter (e.g., accuracy ranges for SpO2, pulse rate, etc.) for the Root system itself, as these are attributed to the individual modules (Radical-7, Radius-7, ISA, Sedline). The Root system primarily acts as an interface and data aggregator.

    Feature/StandardAcceptance CriteriaReported Device Performance
    Electrical SafetyCompliance with IEC60601-1Testing per IEC60601-1 performed; compliant.
    Electromagnetic Compatibility (EMC)Compliance with IEC-60601-1-2Testing per IEC-60601-1-2 performed; compliant.
    Alarm SystemCompliance with IEC-60601-1-8Testing per IEC-60601-1-8 performed; compliant.
    BiocompatibilityCompliance with ISO-10993Testing per ISO-10993 performed; compliant.
    UsabilityCompliance with FDA Human Factors and Usability Draft GuidanceUsability testing performed; compliant.
    Wireless FunctionalityCompliance with FDA Wireless GuidanceWireless testing performed; compliant.
    Software VerificationCompliance with FDA Software GuidanceSoftware verification performed; compliant.
    Mechanical and Environmental CharacteristicsCompliance with Mil-Std-810Mechanical and environmental testing performed; compliant.
    Functionality (Root as an interface)Alternative user interface to integrate modules, provide access/control/monitoring of measurement technologies (of connected modules), transmit data for remote viewing/alarming, communicate with network systems.Functions as intended, displays data from modules, handles alarms, communicates with networks.
    Modified Radius-7 Module PerformanceTo provide all measurements in the Radical-7 module (SpO2, pulse rate, SpCO, SpMet, SpHb, RRa).Thoroughly tested through verification and validation; all requirements and performance specifications satisfied.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document is for a 510(k) premarket notification, which often relies on non-clinical testing and substantial equivalence to a predicate device rather than extensive clinical trials for the primary monitor system itself.

    The document indicates that for the Root Monitoring System and the modified Radius-7 Module, "thoroughly tested through verification and validation" was performed. This suggests a series of engineering and performance tests on the device hardware and software.

    • Sample Size: Not explicitly stated in numerical terms within the provided text. The testing appears to be primarily laboratory-based and engineering verification rather than a patient-based test set.
    • Data Provenance: The nature of the tests (electrical, mechanical, software, etc.) implies that the data was generated internally by Masimo Corporation during product development and validation phases. No information about country of origin of data or whether it was retrospective or prospective is given, as it is non-clinical.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts:

    This information is typically relevant for clinical studies or studies involving human judgment (e.g., image interpretation). Since the testing described is non-clinical (electrical safety, software verification, mechanical, etc.), the concept of "experts establishing ground truth for a test set" in the context of clinical outcomes or diagnostic accuracy doesn't directly apply.

    Instead, the "ground truth" for these engineering and performance tests would be established by:

    • Engineering specifications and design documents.
    • Applicable voluntary standards (e.g., IEC, ISO, Mil-Std).
    • FDA guidances (e.g., Human Factors, Wireless, Software).

    The "experts" would be the engineering and quality assurance teams responsible for designing, conducting, and evaluating these tests against established requirements and standards. Their qualifications would involve expertise in relevant engineering disciplines, regulatory compliance, and quality assurance.

    4. Adjudication Method for the Test Set:

    Adjudication methods (like 2+1, 3+1) are typically used in clinical trials or studies where human interpretation or endpoint determination is subjective and requires consensus. As the described testing is non-clinical, an adjudication method for a test set in this sense is not directly applicable.

    The verification and validation processes would involve:

    • Test protocols defining clear pass/fail criteria.
    • Independent review of test results.
    • Conformity assessment against standards.

    Any discrepancies or failures would likely be resolved through engineering review, root cause analysis, and retesting, rather than an adjudication panel.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No, an MRMC comparative effectiveness study was not done. The document explicitly states: "No clinical testing was done." The Masimo Root system functions as an interface and aggregator for other FDA-cleared modules; its primary mode of submission is through substantial equivalence based on non-clinical tests and its role not altering the intended use of the connected modules.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

    The concept of "standalone" performance, often used for AI algorithms, isn't directly applicable here in the same way. The Masimo Root Monitoring System is a hardware device with software that integrates and displays data from other cleared modules. Its "performance" is in correctly displaying data, processing alarms, and interacting with devices and networks.

    • The non-clinical performance testing described (electrical, EMC, software, etc.) essentially assesses the system's "standalone" functional performance relative to its design specifications and applicable standards, without human intervention in the data generation process, but with a human-in-the-loop for monitoring.
    • However, if "standalone" refers to an algorithm making a diagnostic decision without human input, then no such study was done because this device is a monitoring system and not a diagnostic AI algorithm.

    7. Type of Ground Truth Used:

    For the non-clinical tests specified:

    • Engineering specifications and design requirements: For software verification, functional testing, mechanical characteristics, and interface performance.
    • Compliance with harmonized standards: For electrical safety (IEC60601-1), EMC (IEC-60601-1-2), alarm systems (IEC-60601-1-8), biocompatibility (ISO-10993).
    • Compliance with FDA Guidances: For Usability (Human Factors) and Wireless performance.
    • Predicate Device Equivalence: The foundation of the 510(k) submission is demonstrating substantial equivalence to a legally marketed predicate device (K142394 Masimo Root Monitoring System). The performance of the predicate device serves as an implicit "ground truth" for overall device function and safety. The modifications to the Radius-7 module were then tested to ensure they achieved the same performance as the Radical-7 module's existing cleared functionalities.

    8. Sample Size for the Training Set:

    This question is related to machine learning models. The Masimo Root Monitoring System is described as a medical device for monitoring and data integration, not as a device utilizing a machine learning algorithm that requires a "training set." Therefore, this information is not applicable and not provided in the document.

    9. How the Ground Truth for the Training Set Was Established:

    As there is no indication of a machine learning algorithm or a "training set" in the context of this device, this question is not applicable. The device's foundational components (Radical-7, Radius-7, ISA, Sedline modules) are previously cleared technologies with their own established ground truths for their respective physiological measurements, which the Root system then displays and manages.