510(k) Data Aggregation

    K Number
    K152625

    Validate with FDA (Live)

    Date Cleared
    2016-03-09

    (176 days)

    Product Code
    Regulation Number
    870.1210
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MMS Guide Catheter is intended for the introduction of interventional devices and infusion of diagnostic or therapeutic agents into the peripheral vasculature.

    Device Description

    The MMS Guide Catheter is a small diameter tubular device. It is inserted into the vasculature using standard access techniques and advanced over a guidewire to the physician desired location. The MMS Guide Catheter can be used as a pathway for other diagnostic or therapeutic devices to enter the vasculature. The distal end has a standard luer fitting.

    AI/ML Overview

    This appears to be a 510(k) summary for the MMS Guide Catheter, making a case for substantial equivalence to a predicate device (Marksman Catheter, K111490). The document describes non-clinical testing to demonstrate performance, not diagnostic performance of an AI/ML device. Therefore, a direct answer to the request regarding "acceptance criteria and reported device performance" in the context of an AI/ML device is not applicable for this document.

    However, I can extract the general acceptance criteria and performance reported for the MMS Guide Catheter as presented in the document, treating "acceptance criteria" as the planned tests and "reported device performance" as the outcome of those tests.

    Here's an analysis based on the provided text, adapted to the closest interpretations:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Test)Reported Device Performance
    Biocompatibility:
    CytotoxicityFound to be biocompatible
    SensitizationFound to be biocompatible
    Acute Systemic ToxicityFound to be biocompatible
    Hemolysis ExtractFound to be biocompatible
    Hemolysis Direct ContactFound to be biocompatible
    PyrogenFound to be biocompatible
    Complement ActivationFound to be biocompatible
    Canine ThrombogenicityFound to be biocompatible
    Animal Testing (Simulated Clinical Performance):Performed to specification
    Cadaver Experience (Simulated Clinical Use):Performed to specification
    Non-Clinical Testing:
    Dimensional verificationMet all acceptance criteria
    Surface ReviewMet all acceptance criteria
    LeakMet all acceptance criteria
    TrackingMet all acceptance criteria
    FlowMet all acceptance criteria
    Flex/KinkMet all acceptance criteria
    TorqueMet all acceptance criteria
    TensileMet all acceptance criteria
    PackagingMet all acceptance criteria
    Overall Performance Data:All necessary verification and validation testing has been performed to assure substantial equivalence to the predicate devices.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    This document does not specify sample sizes for mechanical or biocompatibility tests. It briefly mentions "Animal testing" and "Cadaver testing" but provides no details on the number of animals or cadavers used, let alone their provenance or whether the studies were retrospective or prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. This device is a guide catheter, not a diagnostic AI/ML system requiring expert ground truth for its performance evaluation (e.g., image interpretation). The "ground truth" for this device's performance would be engineering specifications and biological reactions, evaluated through laboratory and preclinical tests.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This is not a study involving human interpretation or adjudication. Performance is assessed against predetermined engineering and biological specifications.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI/ML device, nor does it involve human readers in a diagnostic context. This is a medical device clearance submission for a guide catheter.

    6. If a standalone (i.e. algorithm only, without human-in-the-loop performance) was done

    Not applicable. This is a physical medical device (catheter), not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the MMS Guide Catheter, the "ground truth" for demonstrating its performance relies on:

    • Engineering Specifications: Defined physical and mechanical properties (e.g., dimensions, tracking, flex/kink, torque, tensile strength).
    • Biocompatibility Standards (ISO 10993): Established tests for biological safety.
    • Simulated Clinical Performance: Performance in animal models and cadaveric vasculature against expected functionality.
    • Predicate Device Comparison: Establishing substantial equivalence by comparing technological characteristics and intended use to an already cleared device.

    8. The sample size for the training set

    Not applicable. This is a physical medical device, not an AI/ML system that requires a "training set."

    9. How the ground truth for the training set was established

    Not applicable. See point 8.