510(k) Data Aggregation

    K Number
    K140878

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-08-20

    (135 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    12 - 120
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Straumann® dental implants are indicated for oral endosteal implantation in the upper and lower jaw and for the functional and esthetic oral rehabilitation of edentulous and partially dentate patients. Straumann dental implants can also be used for immediate or early implantation following extraction or loss of natural teeth. Implants can be placed with immediate function on single-tooth and/or multiple tooth applications when good primary stability is achieved and with appropriate occlusal loading, to restore chewing function. The prosthetic restorations used are single crowns, bridges and partial or full dentures, which are connected to the implants by the corresponding elements (abutments). In cases of fully edentulous patients, 4 or more implants must be used in immediately loaded cases.

    Device Description

    The subject devices represent a line extension of the previously cleared Bone Level implants of the Straumann Dental Implant System (K062129, K083550 and K121131). The subject devices have the same diameters (3.3, 4.1 and 4.8 mm), the same implant-to-abutment interfaces (NC, RC) the same lengths (8 to 14 mm), the same materials (CP Ti and Ti-13Zr) and the same surface finishes (SLA, SLActive) as the identified predicate devices. The subject devices differ in that the apical aspect of the implants incorporate a tapering thread form, the apical aspect of the implants incorporate three cutting flutes, an additional length of 16mm and the transfer piece is of a new design.

    AI/ML Overview

    The provided text is a 510(k) premarket notification for a medical device, specifically dental implants. It focuses on demonstrating substantial equivalence to existing predicate devices rather than providing acceptance criteria and performance data in the traditional sense of a study proving performance against defined metrics.

    Therefore, much of the requested information cannot be extracted directly from this document because it describes a different type of regulatory submission. This document aims to establish that the new device is as safe and effective as a legally marketed predicate device, not necessarily to meet specific, quantifiable performance targets with a detailed clinical study.

    However, I can extract information related to the device characteristics and the type of evidence provided for substantial equivalence.

    Here's what can be extracted and what cannot:

    1. A table of acceptance criteria and the reported device performance

    • Cannot be fully provided as requested. The document doesn't define explicit "acceptance criteria" against which the device performance is measured in a quantitative manner, like sensitivity, specificity, accuracy, or a specific failure rate.
    • Instead, for dental implants, "performance" is demonstrated through technological characteristics and non-clinical bench studies (specifically dynamic fatigue testing) to show "substantial equivalence" to predicate devices.
    • The comparison table provided (Page 6) shows technological characteristics of the subject device (Straumann® Bone Level Tapered Implants) versus predicate devices. This implicitly suggests that being similar or equivalent in these characteristics is the "acceptance criteria" for substantial equivalence.
    FeatureAcceptance Criteria (Predicate Characteristics)Reported Device Performance (Subject Device Characteristics)
    Implant-to-Abutment ConnectionNarrow CrossFit® (NC), Regular CrossFit® (RC)Narrow CrossFit® (NC), Regular CrossFit® (RC)
    Implant DiameterØ3.3mm, Ø4.1mm, Ø4.8mmØ3.3mm, Ø4.1mm, Ø4.8mm
    Implant Length8, 10, 12, 14mm8, 10, 12, 14, 16mm (Note: new 16mm length added)
    Coronal Thread FormConstant major and minor thread diameters (parallel wall), 0.8mm thread pitchConstant major and minor thread diameters (parallel wall), 0.8mm thread pitch
    Apical Thread FormConstant major and minor thread diameters (parallel wall), 0.8mm thread pitchAngled major and minor thread diameters (tapered wall), with differing major/minor diameters such that depth increases toward apical end, with cutting flutes. 0.8mm thread pitch. (Note: This is a design difference from the predicate, justified by equivalence to Nobel predicates with tapered forms and cutting flutes.)
    Surface FinishSLA, SLActiveSLA, SLActive
    Transfer PieceThread mount transfer pieceSnap fit mount Loxim™ transfer piece (Note: new design, but serves same intended function)
    MaterialCommercially pure grade 4 titanium & Titanium-13Zirconium alloy (Roxolid®)Commercially pure grade 4 titanium & Titanium-13Zirconium alloy (Roxolid®)
    Primary Package SLActiveVial of styrene-butadiene block copolymer (Styrolux), vial cap of LDPE, implant holder of Grilamid TR70 polyamide, filled with aqueous NaCl solution.Vial of styrene-butadiene block copolymer (Styrolux), vial cap of LDPE, implant holder of Grilamid TR70 polyamide, filled with aqueous NaCl solution.
    SterilizationGamma Irradiation, 25 kGy minimum, Validated per ISO 11137-1 and ISO 11137-2 to an SAL of 1 x 10^-6Gamma Irradiation, 25 kGy minimum, Validated per ISO 11137-1 and ISO 11137-2 to an SAL of 1 x 10^-6
    Dynamic Fatigue PerformanceConsistent with FDA guidance and ISO 14801 (implied successful performance by predicates)Data shows that the subject devices provide performance that is substantially equivalent to the identified predicate devices (demonstrated through bench studies per ISO 14801 and FDA guidance). This is the key "performance" reported to meet substantial equivalence. The specific numerical acceptance criteria for fatigue are not detailed here, only that the subject device meets the criteria that the predicates met.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Cannot be provided. The document refers to "bench studies" for dynamic fatigue testing (Page 7). This typically involves mechanical testing rigs, not human or animal subjects, so terms like "test set," "data provenance," "retrospective/prospective" in a clinical sense are not applicable. The sample size for materials testing would be part of the ISO 14801 standard and the specific test protocols, which are not detailed in this summary.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not applicable. This information is relevant for studies involving human interpretation or pathology, which is not the case for this technical submission based on bench testing.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. See point 3.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This device is a dental implant (a physical medical device), not an AI software.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Not applicable. See point 5.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • For the bench testing (dynamic fatigue), the "ground truth" or standard for acceptable performance would be defined by ISO 14801 standards and FDA guidance for dental implants. Success is failure to fracture or deform beyond acceptable limits under specified loading conditions.
    • For the overall substantial equivalence determination, the "ground truth" is typically the performance of the legally marketed predicate devices. The new device aims to be as safe and effective as these established devices.

    8. The sample size for the training set

    • Not applicable. This device is hardware, not an AI algorithm requiring a training set in the computational sense.

    9. How the ground truth for the training set was established

    • Not applicable. See point 8.