510(k) Data Aggregation

    K Number
    K140022

    Validate with FDA (Live)

    Date Cleared
    2014-10-24

    (294 days)

    Product Code
    Regulation Number
    878.4780
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Cork Medical Products Nisus Negative Pressure Wound Therapy System is indicated for use in patients who would benefit from negative pressure wound therapy particularly as the device may promote wound healing by the removal of excess exudates, infectious material, and tissue debris.

    Device Description

    Cork Medical Products has developed a negative pressure wound therapy system. The Cork Medical Products Nisus Negative Pressure Wound Therapy System is indicated for use in patients who would benefit from negative pressure wound therapy particularly as the device may promote wound healing by the removal of excess exudates, infectious material, and tissue debris.

    The components included within the Cork NPWT System are:

    • Nisus Negative Pressure Wound Therapy Pump
    • Nisus NPWT Canister 250-mL
    • Nisus Pump Battery Charger
    • Cork Medical Products NPWT Wound Dressing Kit – Medium
    • Cork Medical Products NPWT Wound Dressing Kit - Large
    AI/ML Overview

    This document is a 510(k) premarket notification for the "Cork Medical Products Nisus Negative Pressure Wound Therapy System." It details the device, its intended use, and a comparison to a predicate device to establish substantial equivalence.

    Here's an analysis of the provided text in relation to your questions:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state formal "acceptance criteria" for each performance test in a tabular format with corresponding reported performance metrics for the Cork Medical Products Nisus NPWT System. Instead, it states that "All biocompatibility test results were negative and passed the pre-defined test acceptance criteria," and that nonclinical tests were conducted "to ensure the Cork Medical Products Nisus Negative Pressure Wound Therapy System is substantially equivalent to the Genadyne A4 Wound Dressing Vacuum System Kit and A4-XLR8 Foam Dressing." This implies the acceptance criteria for the nonclinical tests were equivalence to the predicate device's performance characteristics.

    The comparison table on pages 5 and 6 provides a direct comparison of technical data between the new device and the predicate device. This essentially serves as the 'performance' of both devices that the FDA used to determine substantial equivalence.

    Here's a summarized table based on the provided "Predicate Product Comparison Table" and "Nonclinical Tests" section, inferring acceptance criteria as being substantially equivalent to the predicate:

    Acceptance Criteria (Inferred: Substantially equivalent to Predicate Device)Cork Medical Products Nisus NPWT System PerformancePredicate Device (Genadyne A4 Wound Dressing Vacuum System Kit) PerformanceStudy Type
    PUMP - TECHNICAL DATANonclinical Tests
    Suction Capacity4 liters / minute5 liters / minuteHead to head performance testing
    Maximum Vacuum Pressure220-mmHg230-mmHgHead to head performance testing
    Power Requirements18 VDC, 2A24 VDC, 1AHead to head performance testing
    Battery TypeLi-ionNi-MHComparison of specifications
    Dimensions151 x 108 x 71-mm (~6 x 4.3 x 2.8-inches)200 x 180 x 80-mm (~7.9 x 7.1 x 3.1-inches)Comparison of specifications
    Weight0.575-kg (~1.27-lb)1.36-kg (~3-lb)Comparison of specifications
    ReusableYesYesComparison of specifications
    SterileNon-sterileNon-sterileComparison of specifications
    ComplianceIEC 60601-1:2005, 3rd Edition (AAMI ES 60601-1, CAN/CSA C22.2 No. 60601-1-08, EN 60601-1)IEC 60601-1-2:2007IEC 60601-1-6:2010 / IEC 62366:2010IEC 60601-1-11:2010UL 60601-1CAN/CSA C22.2 No. 601-1-M90Comparison of specifications
    Storage & Shipping Conditions-25°C (-13°F) without relative humidity control to 44°C (111°F) up to 93% relative humidity (non-condensing)-18°C to +43°C (0°F to 110°F)Relative Humidity 10% to 95%700 – 1060 mbar Atmospheric pressureComparison of specifications
    Environmental ConditionsOperating Temperature: 18°C to 34°C (65°F to 94°F)Operating Relative Humidity: 10% - 95%Operating Pressure: 700-hPA – 1060-hPA (10.15-atm – 15.37-atm) atmospheric pressure18°C to 34°C (65°F to 94°F) Relative Humidity 10% to 95%700 – 1060 mbar Atmospheric pressureComparison of specifications
    ACCESSORIES
    Canisters250-mL disposable canisterCanister includes hydrophobic membrane filter and liquid solidifier800-mL disposable canister with a built-in hydrophobic shut off filter for overflow protectionComparison of specifications
    Wound Dressing KitWound Foam: Reticulated polyether based polyurethane foam (A30M)Wound Drape: Transparent polyurethane film with adhesive backingPort Pad Assembly: Silicone port pad, Port Pad Skirt (Transparent polyurethane film with adhesive backing), Drainage tubing, Luer connector, Pinch clampK082676:Non-adherent gauzeAnti-microbial gauzeTransparent film dressingSilicone flat drainK092992: Polyurethane foam (A30M)Comparison of specifications
    SterilizationIndividual kit components individually sterilized by Ethylene Oxide or Gamma Irradiation(cleared on K132004) - implied similar sterilizationComparison of specifications
    PERFORMANCE TESTSNonclinical Tests
    Continuous Mode Low Pressure (40-mmHg) TestPassed (implied by "substantially equivalent")(Performance of predicate device)Head to head performance testing
    Continuous Mode Typical Pressure (125-mmHg) TestPassed (implied by "substantially equivalent")(Performance of predicate device)Head to head performance testing
    Continuous Mode High Pressure (200-mmHg) TestPassed (implied by "substantially equivalent")(Performance of predicate device)Head to head performance testing
    Intermittent Mode Test (high pressure: 125-mmHg, low pressure: 40-mmHg)Passed (implied by "substantially equivalent")(Performance of predicate device)Head to head performance testing
    Leakage Alarm TestPassed (implied by "substantially equivalent")(Performance of predicate device)Head to head performance testing
    Canister Full Alarm TestPassed (implied by "substantially equivalent")(Performance of predicate device)Head to head performance testing
    Blockage Alarm TestPassed (implied by "substantially equivalent")(Performance of predicate device)Head to head performance testing
    Operating Parameter TestPassed (implied by "substantially equivalent")(Performance of predicate device)Head to head performance testing
    Storage & Shipping Parameter TestPassed (implied by "substantially equivalent")(Performance of predicate device)Head to head performance testing
    Canister Real-Time AgingPassed (implied by "substantially equivalent")(Performance of predicate device)Head to head performance testing
    Wound Dressing Kit AgingPassed (implied by "substantially equivalent")(Performance of predicate device)Head to head performance testing
    BIOCOMPATIBILITY
    Cytotoxicity TestNegative/Passed(Implied standard biocompatibility)ISO 10993 testing standards
    Intracuteneous Reactivity TestNegative/Passed(Implied standard biocompatibility)ISO 10993 testing standards
    Sensitization TestNegative/Passed(Implied standard biocompatibility)ISO 10993 testing standards

    2. Sample size used for the test set and the data provenance

    • Sample size for nonclinical tests: Not explicitly stated for each test beyond "head to head performance testing," "Canister Real-Time Aging," and "Wound Dressing Kit Aging." For the biocompatibility tests, the sample size refers to the "wound foam and wound drape components" and "individual components included in the Cork Medical Products Wound Dressing Kit." Specific numbers are not provided.
    • Data provenance: The tests were conducted internally by Cork Medical Products ("Cork Medical conducted head to head performance testing") and a Biocompatibility Risk Assessment was completed by "Nelson Laboratories." The data origin is thus from Cork Medical's testing and a third-party lab. No information about country of origin of the data or whether it was retrospective or prospective is mentioned, which is typical for premarket notifications focus on non-clinical engineering and bench testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This device is a negative pressure wound therapy system, not an AI/imaging device requiring expert interpretation for ground truth. The "ground truth" for its performance is derived from physical measurements (pressure, flow, alarm functionality) and standardized biocompatibility testing.

    4. Adjudication method for the test set

    Not applicable. As noted above, this device does not involve expert adjudication for establishing ground truth, as it's a mechanical/electrical device.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is not an AI-assisted diagnostic or imaging tool.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This device is not an algorithm, but a physical medical device.

    7. The type of ground truth used

    • Nonclinical Performance Tests: The "ground truth" for the nonclinical performance tests (e.g., pressure, alarms) was established through engineering and bench testing protocols using "simulated wound exudate" and a "wound test bed fixture." The implicit ground truth standard for these tests was the demonstrated performance of the legally marketed predicate device.
    • Biocompatibility Tests: The ground truth was based on international standards, specifically ISO 10993 testing standards, with predefined acceptance criteria for cytotoxicity, intracutaneous reactivity, and sensitization.

    8. The sample size for the training set

    Not applicable. This device does not use machine learning or AI, and therefore does not have "training sets" in that context. Its performance is based on engineering design and testing, not learned data.

    9. How the ground truth for the training set was established

    Not applicable (as above).