Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K113300

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-12-09

    (30 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For percutaneous transluminal angioplasty (PTA) of lesions in peripheral arteries including iliac, renal, popliteal, infrapopliteal, femoral and iliofemoral as well as obstructive lesions of native or synthetic arteriovenous dialysis fistulae. The ATB Advance® PTA Dilatation Catheter is also intended for post dilatation of balloon-expandable peripheral vascular stents.

    Device Description

    The ATB Advance® PTA Dilatation Catheter is an over-the-wire catheter indicated for percutaneous transluminal angioplasty (PTA) of lesions in peripheral arteries including iliac, renal, popliteal, infrapopliteal, femoral and iliofemoral as well as obstructive lesions of native or synthetic arteriovenous dialysis fistulae. The ATB Advance® PTA Dilatation Catheter is also intended for post dilatation of balloon-expandable peripheral vascular stents. The 6 Fr balloon catheter will be compatible with a 0.035 inch wire guide and an 8 Fr sheath. It will be supplied sterile, intended for one-time use.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the ATB Advance® PTA Dilatation Catheter:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Balloon Minimum Burst Strength: Burst at or above the minimum rated burst pressure.Testing shows the balloons will burst at or above the minimum rated burst pressure, with all failure modes being linear tears. Predetermined acceptance criteria met.
    Balloon Compliance: Meet labeled diameter within tolerance at the nominal pressure under simulated body temperature conditions.Testing shows that, under simulated body temperature conditions, each balloon will meet its labeled diameter within tolerance at the nominal pressure. Predetermined acceptance criteria met.
    Balloon Profile: Compatible with an 8 Fr sheath (< 0.113 inch profile).Measurement of the diameter of catheter shaft, bonds, and folded balloon shows that the device is compatible with an 8 Fr sheath (< 0.113 inch profile). Predetermined acceptance criteria met.
    Balloon Fatigue: Free from leakage and damage on inflation after 10 cycles of inflation/deflation (inflating to rated burst pressure, holding for 30 seconds and deflating).Testing shows that balloons are free from leakage and damage on inflation, withstanding 10 cycles of inflation/deflation (inflating to rated burst pressure, holding for 30 seconds and deflating). Predetermined acceptance criteria met.
    Sheath Compatibility: Compatible with an 8 Fr sheath.Qualitative and quantitative evaluations show that the balloons are compatible with an 8 Fr sheath. Predetermined acceptance criteria met.
    Balloon Bond Strength: Tensile force during proper clinical use should not fracture or rupture the balloon catheter bond (in conformance with ISO 10555-1: 1995).Testing shows the tensile force during proper clinical use should not fracture or rupture the balloon catheter bond. In conformance with the applicable sections of ISO 10555-1: 1995, predetermined acceptance criteria were met.

    2. Sample Size Used for the Test Set and Data Provenance

    The provided text describes bench testing only. There is no mention of a "test set" in the context of patient data. The testing involves physical properties of the device (balloon, catheter).

    • Sample Size: Not explicitly stated for each test (e.g., "how many balloons were tested for burst strength"). It only states "Testing shows..." which implies multiple units were tested to demonstrate conformity.
    • Data Provenance: N/A, as this is laboratory bench testing, not clinical data.
    • Retrospective/Prospective: N/A

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    N/A. This 510(k) summary focuses on device design and performance through engineering and laboratory testing, not on clinical evaluations requiring expert interpretation of patient data or ground truth establishment.


    4. Adjudication Method for the Test Set

    N/A. Adjudication methods are typically used for clinical studies with subjective interpretations. The tests described are objective, measurable physical properties.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. This document does not mention any MRMC study, human readers, AI, or comparative effectiveness with or without AI assistance. This device is a medical instrument (balloon catheter), not an AI-powered diagnostic tool.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    No. This concept is irrelevant to a physical medical device like a balloon catheter.


    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    N/A. For physical device testing, the "ground truth" is typically defined by engineering specifications, industry standards (like ISO 10555-1: 1995), and predefined acceptance thresholds derived from those specifications. For example, the ground truth for "Balloon Minimum Burst Strength" is the "minimum rated burst pressure."


    8. The Sample Size for the Training Set

    No "training set" is applicable here. The device is a physical product, not a machine learning model.


    9. How the Ground Truth for the Training Set Was Established

    No "training set" and therefore no ground truth establishment for it.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1