Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K102520

    Validate with FDA (Live)

    Date Cleared
    2011-01-06

    (126 days)

    Product Code
    Regulation Number
    880.5200
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    As indicated in 21 CFR Part 880.5200, The Nexiva™ intravascular catheter is inserted into a patient's vascular system for a short-term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravascularly. The needle-shielding feature and luer access port, aid in the prevention of needle-stick injuries. Blood is contained within the device during the catheter insertion process aiding in the prevention of blood exposure. This catheter may be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness of procedure.

    The 18-22 gauge Nexiva™ catheters are suitable for use with power injectors rated for a maximum of 300 psi when the luer access port(s) is removed and a direct connection is made.

    Device Description

    The BD Nexiva™ Closed IV Catheter System consists of an over-the needle, peripheral intravascular catheter made from Vialon™ polyurethane, integrated extension tubing with a Y adapter and clamp, BD Q-Syte™ luer access port, and a passive needleshielding mechanism.

    The design of the Nexiva™ IV catheter can be described as a closed system since it protects clinicians from blood exposure during the catheter insertion procedure. Since the needle is withdrawn through a septum that seals after the needle has been removed and both ports of the Y adapter are closed, blood is contained within the NexivaTM device during catheter insertion. The pressure exerted on the needle as it passes through the septum wipes blood from the needle, further reducing potential blood exposure. The slide clamp on the integrated extension tubing is provided to eliminate blood exposure when the vent plug is replaced with an infusion set connection of a BD Q-Syte™ luer access port.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the BD Nexiva™ Closed IV Catheter System, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The nonclinical tests compare the modified BD Nexiva™ Closed IV Catheter System to its predicate device (K032843). The reported device performance for all listed characteristics is "Pass," indicating that the device met the specified acceptance criteria.

    Associated ModificationProduct Performance Characteristics/Verification TestingAcceptance Criteria for Product VerificationReported Device Performance
    1Premature Decouple ForceDPPM < 280Pass
    2Catheter Pull ForceDPPM < 60Pass
    2Catheter Burst StrengthDPPM < 60Pass
    2Catheter Penetration ForceDPPM <1350Pass
    2Catheter DragDPPM <1350Pass
    2Tip AdhesionDPPM <1350Pass
    1System Drag -Cannula DragDPPM < 500Pass
    1System Drag -Ferrule Retraction ForceDPPM < 500Pass
    1Adapter Release ForceDPPM < 500Pass
    3Needle Cover RemovalDPPM USL < 18000Pass
    3(Needle Cover Removal)DPPM LSL < 200000Pass
    3(Needle Cover Removal)-20%/+25%, DPPM < 5000Pass
    4Flow Rate Label Claim-10%/+15%, DPPM < 5000Pass
    5Pinch Clamp Fluid SealDPPM < 10000Pass
    5Pinch Clamp Engagement Disengagement ForceDPPM < 2000Pass
    8Wing deflection forceDPPM USL < 2000Pass
    8(Wing deflection force)DPPM LSL < 500Pass
    6Bond Strength -Tubing Bond to AdapterDPPM < 0.1Pass
    4Extension Tube Burst PressureDPPM < 60Pass
    10Unit Package Seal Peel Force1.33 CpkPass
    10Seal Width1.33 CpkPass
    10Unit Package Material Thickness - Post Formed (Bottom Wed)1.33 CpkPass
    10Package Integrity (Water Leak)0 out of 2301Pass
    10Ship Testing (Drop/Vibration)0 out of 2301Pass
    7Y-Adapter Wishbone LoadingDPPM < 0.1Pass
    7ISO Liquid LeakageDPPM < 500Pass
    7ISO Air LeakageDPPM < 500Pass
    7ISO Unscrewing TorqueDPPM < 500Pass
    7ISO Resistance to OverridingDPPM < 500Pass
    7ISO Thread SeparationDPPM < 500Pass
    7Vent Plug Removal Force/TorqueDPPM USL < 6500Pass
    7(Vent Plug Removal Force/Torque)DPPM LSL <10000Pass
    7Q-Syte Removal TorqueDPPM USL < 6500Pass
    7(Q-Syte Removal Torque)DPPM LSL <10000Pass
    11ISO Thread Seperation -Both Y female luerDPPM < 500Pass
    11ISO Thread Seperation -Straight female luerDPPM < 500Pass
    11ISO Unscrewing Torque -Both Y female luerDPPM < 500Pass
    11ISO Unscrewing Torque -Straight female luerDPPM < 500Pass
    11ISO Resistance to Override -Both Y female luerDPPM < 500Pass
    11ISO Resistance to Override -Straight female luerDPPM < 500Pass
    11ISO Liquid LeakageDPPM < 500Pass
    11ISO Air LeakageDPPM < 500Pass
    12Vent Plug Torque RemovalDPPM USL < 10000, DPPM LSL< 20000Pass
    11Bond Strength - Tubing Bond to AdapterDPPM < 60Pass
    11Flow Rate Label Claim-10%/+15%, DPPM < 5000Pass
    14, 15Proximal Re-exposureDPPM < 100Pass
    17System Drag -Maximum Adhesion ForceDPPM < 1350Pass
    17System Drag - Average Drag ForceDPPM < 10,000Pass
    17System Drag - Offset Peak ForceDPPM < 2000Pass
    17Flow Rate Label Claim-20%/+25%, DPPM < 5000, -10%/+15%, DPPM < 5000Pass
    17Minimum FlowDPPM < 5000Pass
    17FlashbackDPPM < 1000Pass
    17Cannula PenetrationDPPM < 1350Pass
    17Catheter PenetrationDPPM < 1350Pass
    17, 19Premature DecoupleDPPM < 280Pass
    20Needle Hub PullDPPM < 500Pass

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document specifies a sample size of 2301 units for "Package Integrity (Water Leak)" and "Ship Testing (Drop/Vibration)" tests. For other tests, specific sample sizes are not explicitly stated, but the acceptance criteria are generally expressed in "DPPM" (Defective Parts Per Million) or "Cpk" (Process Capability Index), suggesting statistical process control and quality assurance testing.
    • Data Provenance: The tests are described as "Nonclinical Tests" and "design verification activities," conducted by Becton Dickinson Infusion Therapy Systems, Inc. The data is retrospective in the sense that it was generated during device development and verification to support the 510(k) submission. There is no information on the country of origin of the data, but it can be assumed to be related to the manufacturer's testing facilities, potentially in the US (headquarters in Sandy, UT).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This document primarily describes physical and mechanical performance testing of a medical device, not diagnostic or clinical performance involving expert interpretation of data (like radiology images). Therefore, the concept of "experts" establishing ground truth in this context doesn't directly apply as it would for a software device. The ground truth for these tests is based on established engineering standards (e.g., ANSI/AAMI/ISO 10993-1 2003 (E) for biocompatibility, and presumably other relevant engineering and ISO standards for mechanical performance), internal product specifications, and quality control methodologies.

    4. Adjudication Method for the Test Set

    Not applicable. As described above, these are physical and mechanical tests with quantitative acceptance criteria (e.g., DPPM, Cpk, force measurements). There is no "adjudication" in the sense of reconciling differing expert opinions or classifications. The results are objective measurements compared against predefined thresholds.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. This document does not describe an MRMC comparative effectiveness study. The submission relates to a physical medical device (intravascular catheter) and its modifications, not an AI or software device that would typically involve human readers interpreting cases.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    No. This concept is not applicable as the submission is for a physical medical device, not an algorithm or AI. The tests are focused on the device's physical and mechanical properties.

    7. The Type of Ground Truth Used

    The ground truth used for these tests is based on:

    • Engineering Standards: Adherence to established industry and international standards (e.g., ANSI/AAMI/ISO 10993-1 2003 (E), ISO Liquid Leakage, etc.).
    • Product Specifications: Internal design requirements and performance targets defined by the manufacturer.
    • Predicate Device Performance: The modified device is benchmarked against its predicate device (K032843) to demonstrate "substantial equivalence."

    8. The Sample Size for the Training Set

    Not applicable. This is not a machine learning or AI device that requires a training set. The "training" for such physical devices involves design, prototyping, and iterative testing processes, not data-driven algorithmic training.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" in the context of this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1