Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K102187

    Validate with FDA (Live)

    Date Cleared
    2010-12-10

    (129 days)

    Product Code
    Regulation Number
    880.2910
    Age Range
    All
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is an electronic clinical thermometer using an infrared sensor to detect body temperature from the auditory canal in the neonatal, pediatric and adult population used in the home setting.

    Device Description

    The Radiant Innovation Inc., Infrared Ear Thermometer, Model TH709L(E) is electronic thermometer using an infrared detector (thermopile detector) to detect body temperature from the auditory canal. Its operation is based on measuring the natural thermal radiation emanating from the tympanic membrane and the adjacent surfaces of the patient.

    To measure ear temperature, the ear thermometer is inserted into a patient's outer ear canal. A start button is pressed to start the measurement through the radiation exchanges. The electrical signal read out from the detector is fed to the circuit for amplification and calculation. The measured temperature then appears on a LCD display. The total operation takes a few seconds.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and supporting study for the Radiant Innovation Inc. Infrared Ear Thermometer TH709L(E), based on the provided document:

    This document is a 510(k) summary for a Special 510(k) submission, meaning the device TH709L(E) is being asserted as substantially equivalent to a previously cleared predicate device (TH8 series) due to minor modifications. The focus of such submissions is typically demonstrating that the modifications do not raise new questions of safety or effectiveness, often through compliance with recognized standards and comparison to the predicate. Therefore, the "study" described is primarily about demonstrating compliance with accuracy standards rather than a comparative effectiveness study or a complex standalone algorithm validation.

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (from ASTM E1965-98 and EN12470-5:2003, as implied by compliance)Reported Device Performance (TH709L(E))
    Accuracy: ±0.2°C (±0.4°F) for 35.5°C to 42°C (95.9°F to 107.6°F)±0.2°C (0.4°F) for 35.542°C (95.9107.6°F)
    Accuracy: ±0.3°C (±0.5°F) for other temperature ranges±0.3°C (0.5°F) for other temperature ranges

    Note: The document explicitly states "Compliance to applicable voluntary standards includes ASTM E1965-98 and EN12470-5;2003". These standards define the accuracy requirements for clinical infrared thermometers. The reported performance in the comparison table mirrors these standard accuracy criteria, implying the device met these thresholds.

    2. Sample size used for the test set and the data provenance

    The document does not explicitly state a specific "test set" sample size or data provenance in terms of human subjects for a clinical study. The evaluation appears to be based on bench testing ("bench testing contained in this submission supplied demonstrate that the modification of TH709L(E) do not raise any new questions of safety or effectiveness.") and compliance with voluntary standards (ASTM E1965-98 and EN12470-5;2003). These standards typically involve specified test procedures and equipment to verify accuracy under controlled conditions, rather than a clinical trial with a specific number of patients.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable as the evaluation focuses on compliance with technical standards through bench testing, not on interpreting medical images or data that would require expert adjudication. The "ground truth" for thermometer accuracy is established by highly accurate reference thermometers and controlled temperature environments as part of the standard testing protocols specified in ASTM E1965-98 and EN12470-5:2003.

    4. Adjudication method for the test set

    This information is not applicable for the same reasons as point 3.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable. The device is an infrared ear thermometer, not an AI-powered diagnostic tool that assists human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable in the context of an "algorithm only" performance as would be relevant for an AI device. The device performance itself (measuring temperature) is essentially a standalone function, and its accuracy is validated against established standards.

    7. The type of ground truth used

    The ground truth used for establishing the device's accuracy is derived from measurements taken by highly accurate reference instruments under controlled environmental conditions, as dictated by the specified voluntary standards (ASTM E1965-98 and EN12470-5:2003).

    8. The sample size for the training set

    This information is not applicable. This is not an AI/machine learning device that requires a training set.

    9. How the ground truth for the training set was established

    This information is not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1