510(k) Data Aggregation

    K Number
    K251564

    Validate with FDA (Live)

    Device Name
    microINR System
    Date Cleared
    2025-07-21

    (60 days)

    Product Code
    Regulation Number
    864.7750
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The microINR System measures prothrombin time (PT) expressed in International Normalized Ratio (INR), for monitoring oral anticoagulant therapy with warfarin.

    The microINR System consists of a meter and chips (test strips) and uses fresh capillary whole blood from a fingerstick. The microINR System is intended for patient self-testing use as well as for healthcare professionals at Point of Care settings.

    The microINR System is intended for use in patients 18 years old or older. Patients must be stable on warfarin medication for at least 6 weeks before starting to use the microINR System.

    For self-testing use: The system is intended for properly trained users under specific prescription of a physician.

    Caution: The microINR System is not intended for use in patients who are transitioning from heparin treatment to warfarin therapy. The microINR System is not intended to be used for screening purposes.

    Device Description

    The microINR System is comprised of a portable measuring device (microINR, microINR Link and microINR Expert meter) and test strips (microINR Chips) in which the capillary blood sample flows through capillary action.

    The microINR Chip contains a reagent in dried form which consists of thromboplastin, and contains two symmetrical regions, the measuring channel and a control channel. The microINR meters measure International Normalized Ratio (INR) based on a Prothrombin Time (PT) assay carried out in the microINR Chip based on microfluidic technology with machine vision detection.

    The microINR System has a multi-level On-board Quality Control. Multiple key functions and elements of the system are checked and if deviations are detected, error messages are displayed and test results are not reported.

    AI/ML Overview

    The microINR System measures prothrombin time (PT) expressed in International Normalized Ratio (INR) for monitoring oral anticoagulant therapy with warfarin.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided FDA 510(k) clearance letter:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state pre-defined "acceptance criteria" in terms of specific thresholds for slope, intercept, Pearson correlation, SD, or CV%. Instead, it presents the performance characteristics demonstrated by the studies. The comparison to predicate devices, and the conclusion of substantial equivalence, imply that these reported performance values were considered acceptable.

    MetricAcceptance Criteria (Implied)Reported Device Performance
    Accuracy (microINR System vs. Lab Reference)Strong correlation (r ≈ 1), small bias (slope ≈ 1, intercept ≈ 0)N: 1016, Slope: 1.00, Intercept: 0.08, Pearson (r): 0.97
    Accuracy (Self-testing patients vs. Lab Reference)Strong correlation (r ≈ 1), small bias (slope ≈ 1, intercept ≈ 0)N: 248, Slope: 0.95, Intercept: 0.13, Pearson (r): 0.97
    Accuracy (Self-testing patients vs. HCPs using microINR System)Strong correlation (r ≈ 1), small bias (slope ≈ 1, intercept ≈ 0)N: 461, Slope: 0.99, Intercept: 0.02, Pearson (r): 0.98
    Precision (INR < 2.0)Low variability (small SD, CV%)N test pairs: 131, Mean: 1.46, SD: 0.07, CV%: 4.9
    Precision (2.0 ≤ INR < 4.5)Low variability (small SD, CV%)N test pairs: 337, Mean: 2.79, SD: 0.14, CV%: 5.0
    Precision (INR ≥ 4.5)Low variability (small SD, CV%)N test pairs: 34, Mean: 5.30, SD: 0.26, CV%: 4.9
    Precision (Trained PSTs)Low variability (small SD, CV%)N: 111, Mean: 2.63, SD: 0.13, CV%: 4.9
    Precision (Untrained PSTs with high INR)Low variability (small SD, CV%)N: 13, Mean: 5.64, SD: 0.26, CV%: 4.6

    Note: The "Acceptance Criteria (Implied)" are derived from the typical expectations for such medical devices and the fact that these results led to FDA clearance, indicating they met regulatory standards.

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Accuracy Study (microINR System vs. Laboratory Reference Device):

      • N: 1016 microINR System results.
      • Data Provenance: Capillary blood samples from control subjects and patients recruited at ten clinical sites. This suggests a prospective collection from multiple geographical locations, though specific countries are not mentioned.
    • Accuracy Study (Self-testing patients vs. Laboratory Reference Device):

      • N: 248 INR test results obtained by self-testing patients (compared to laboratory systems).
      • Data Provenance: Collected at seven clinical sites. This also implies prospective data collection.
    • Accuracy Study (Self-testing patients vs. Healthcare Professionals using microINR System):

      • N: 461 INR test results.
      • Data Provenance: Collected at seven clinical sites, comparing self-testers to healthcare professionals. This is likely prospective.
    • Precision Study (microINR System):

      • N: Duplicate measurements from 502 control subjects and patients.
      • Data Provenance: Performed at ten clinical sites, suggesting prospective collection.
    • Precision Study (Trained PSTs):

      • N: Duplicate measurements from 111 patients ("Paired results").
      • Data Provenance: Performed at four clinical sites. Implies prospective collection.
    • Precision Study (Untrained PSTs with high INR values):

      • N: Duplicate measurements from 13 patients ("Paired results").
      • Data Provenance: Performed at three clinical sites. Implies prospective collection.

    The general provenance is from multiple clinical sites, suggesting a diverse patient population, but specific countries are not detailed. All studies appear to be prospectively collected for the purpose of validating the device performance.

    3. Number of Experts Used to Establish the Ground Truth and Qualifications

    • The document mentions comparisons to "laboratory systems ACL TOP 500 or ACL TOP 750" as reference methods. These are established laboratory analyzers.
    • For the "Accuracy of the microINR System by self-testing patients" study, the comparison also includes "Healthcare professionals (HCP) using the microINR System." While HCPs are involved, specific qualifications beyond their professional role are not detailed for the purpose of establishing a "ground truth" other than operating the device.
    • The ground truth for the clinical method comparison studies seems to be based on the results from the ACL TOP 500 or ACL TOP 750 laboratory systems, which are considered the gold standard for PT/INR measurement, rather than human experts adjudicating individual cases.
    • The document does not explicitly describe "experts" in the sense of radiologists reviewing images; rather, it refers to standardized laboratory methods and healthcare professionals. Therefore, the specific "number of experts" or their "qualifications" for establishing ground truth, beyond the inherent accuracy of the reference laboratory devices, is not provided.

    4. Adjudication Method for the Test Set

    No explicit adjudication method (e.g., 2+1, 3+1) is mentioned. Given that the ground truth appears to be established by comparison to highly accurate and standardized laboratory reference instruments (ACL TOP 500/750), a human adjudication process for individual test results is typically not required or described in such device submissions. The comparison is quantitative against these reference devices.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study, detailing how much human readers improve with AI vs. without AI assistance, was performed or presented. This device is a standalone diagnostic system (a meter and test strips) and not an AI-assisted diagnostic tool that aids human interpretation of complex medical images or data. The "readers" in this context are the device users (patients or healthcare professionals), and the study compares their results directly to a laboratory reference, or compares self-testers to healthcare professionals using the same device, rather than measuring an improvement in human diagnostic accuracy with an AI assist.

    6. Standalone Performance

    Yes, standalone performance was done. The entire set of performance characteristics, particularly the "Accuracy of the microINR System" (comparing the microINR System against a laboratory reference device) and "Precision of the microINR System," represents the standalone performance of the device without explicit human-in-the-loop interpretation beyond operating the device as intended. The system itself measures and displays the INR.

    7. Type of Ground Truth Used

    The primary ground truth used is comparison to established laboratory reference methods, specifically ACL TOP 500 or ACL TOP 750 laboratory systems, which measure PT/INR from citrate venous plasma samples. This is a highly standardized and accepted method for determining INR.

    8. Sample Size for the Training Set

    The document does not provide details on a specific "training set" for the device, as it is not an AI/machine learning model in the typical sense that requires a training and a separate test set. The changes implemented are "only SW changes... for increasing the INR measuring range... and the removal of the limitation for LVAD patients," implying a modification to an existing algorithm rather than a completely new, data-trained AI model. The studies presented are for validation of these changes.

    9. How the Ground Truth for the Training Set Was Established

    As noted above, no specific "training set" is mentioned in the context of a machine learning model. The device's underlying principle relies on microfluidic technology and machine vision detection of a prothrombin time assay. The calibration of the chips is done against "International Reference Thromboplastin of the World Health Organization," which serves as a fundamental standard rather than a "ground truth" for a training set in data science terms.

    Ask a Question

    Ask a specific question about this device

    K Number
    K243543

    Validate with FDA (Live)

    Device Name
    microINR System
    Date Cleared
    2025-02-27

    (104 days)

    Product Code
    Regulation Number
    864.7750
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The microINR System measures prothrombin time (PT) expressed in International Normalized Ratio (INR), for monitoring oral anticoagulant therapy with warfarin.

    The microlNR System consists of a meter and chips (test strips) and uses fresh capillary whole blood from a fingerstick. The microlNR System is intended for patient self-testing use as well as for healthcare professionals at Point of Care settings.

    The microlNR System is intended for use in patients 18 years old or older. Patients must be stable on warfarin medication for at least 6 weeks before starting to use the microINR System.

    For Self-testing use: The System is intended for properly trained users under specific prescription of a physician. Caution: The microINR System is not in patients who are transitioning from heparin treatment to VKA therapy. The microINR System is not intended to be used for screening purposes.

    Device Description

    The modified device microINR System containing the microINR Expert Meter is derived from the existing devices microINR Meter and microINR Link Meter. The change is for the incorporation of additional connectivity functions (Wi-Fi and Ethernet connectivity on top of the Bluetooth connectivity of the microINR Link), a touchscreen, a barcode scanner as well as other configurable settings and optional functions (such as the option of including operator and patient identification). All the actions related to the extra functionalities and configurable settings are set before or after performing the analytical test. The fundamental scientific technology of the modified device and its performance has not changed. No performance changes to the microINR System have been implemented since its initial clearance on K180780 and its home use clearance in K201185.

    AI/ML Overview

    The provided text is a 510(k) summary for the microINR System, specifically addressing changes made to the device (the incorporation of the microINR Expert Meter). The key takeaway from the document is that the fundamental scientific technology and performance of the device have NOT changed. The modifications primarily involve connectivity, user interface, and optional features.

    Therefore, the document does not present new performance data based on a new study for the microINR System's core functionality (measuring PT/INR). Instead, it relies on the performance previously cleared under K180780, K201185, and K231711. The testing described in section 7 ("Testing on the modified device was conducted as follows: Software verification and validation; testing for Radiofrequency, Electromagnetic Compatibility and Electrical Safety: Cybersecurity analysis and testing; and Usability testing.") relates to the new functionalities added to the expert meter, not the core PT/INR measurement.

    Given this, I cannot extract numerical "acceptance criteria" and "reported device performance" directly from this document for the PT/INR measurement. The document states:

    • "No performance changes to the microINR System have been implemented since its initial clearance on K180780 and its home use clearance in K201185."
    • "The fundamental scientific technology of the modified device and its performance has not changed."
    • "The performance characteristics of the microINR System remain the same."

    Therefore, for the purpose of answering your request based solely on the provided text, the answer will reflect that the core performance information is not detailed here, as the submission focuses on documenting that the new components do not negatively impact the already cleared performance.

    Here's the information based on the provided text, with explicit notes where information is not available or where the focus is on the new functionalities rather than the core PT/INR measurement:


    Acceptance Criteria and Device Performance for microINR System (Focus on New Expert Meter Functionalities)

    The provided 510(k) summary for the microINR System (K243543) primarily addresses changes related to the incorporation of the "microINR Expert Meter," which introduces additional connectivity, a touchscreen, barcode scanner, and other configurable settings. The document explicitly states that the fundamental scientific technology and the performance of the device related to its intended use (measuring Prothrombin Time/INR) have not changed from previous clearances (K180780, K201185, K231711).

    Therefore, the "study" described in the document primarily focuses on verifying and validating these new functionalities to ensure they do not introduce new questions of safety or effectiveness, rather than re-evaluating the core PT/INR measurement accuracy.

    1. Table of Acceptance Criteria and Reported Device Performance

    As the core performance of the PT/INR measurement is stated to be unchanged and reliant on previous clearances, a specific table of numerical acceptance criteria and reported new performance for PT/INR is not present in this document. The testing detailed pertains to the new functionalities of the Expert Meter. The acceptance criteria for these new functionalities are broadly stated as:

    Acceptance Criterion Type (for new functionalities)Reported Device Performance (for new functionalities)
    Software verification and validationAll testing results met the pre-determined acceptance criteria. Performed as intended.
    Radiofrequency, Electromagnetic Compatibility, Electrical SafetyAll testing results met the pre-determined acceptance criteria. Performed as intended.
    Cybersecurity analysis and testingAll testing results met the pre-determined acceptance criteria. Performed as intended.
    Usability testingAll testing results met the pre-determined acceptance criteria. Performed as intended.
    Risk analysis of implemented changesNo significant detectable risks compared to predicate device; all residual risks deemed acceptable.

    2. Sample Sizes Used for the Test Set and Data Provenance

    The document does not specify quantified sample sizes for the testing of the new functionalities. It generally refers to "Software verification and validation; testing for Radiofrequency, Electromagnetic Compatibility and Electrical Safety: Cybersecurity analysis and testing; and Usability testing."

    Data provenance (e.g., country of origin, retrospective/prospective) for these specific verification and validation activities is not detailed.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    Given that the document focuses on technical verification (software, electrical, cybersecurity) and usability of new features (connectivity, touchscreen, barcode scanner), and not the diagnostic accuracy of PT/INR measurement, the concept of "experts establishing ground truth" in the clinical sense (e.g., radiologists for imaging) is not directly applicable here. The ground truth for these tests would be technical specifications and user expectations. The document does not specify the number or qualifications of personnel involved in these specific verification and validation activities.

    4. Adjudication Method for the Test Set

    Not applicable in the context of the described technical verification and validation of new functionalities. Adjudication methods like 2+1 or 3+1 are typical for clinical outcome or diagnostic accuracy studies involving human interpretation, which is not the focus of the testing described for this device's modifications.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done

    No MRMC study was done or described in this document. This type of study is relevant for AI-assisted diagnostic tools where human reader performance is being evaluated, which is not the nature of the microINR System or the modifications herein.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    The microINR System is a point-of-care device that requires human interaction (fingerstick, applying blood, reading results). Therefore, a purely "standalone" (algorithm only) performance is not a direct concept for this type of device. The document states that the "fundamental scientific technology... has not changed," implying that the core measurement algorithm's performance remains as previously cleared. The current submission focuses on added functionalities (connectivity, display) to an already cleared system.

    7. The Type of Ground Truth Used

    For the new functionalities tested: The ground truth would be based on engineering specifications, software requirements, cybersecurity standards, and usability design principles. The document does not specify the exact methods or "ground truth" for each of these technical tests, only that they met pre-determined acceptance criteria.

    For the core PT/INR measurement: This information would reside in the previous 510(k) submissions (K180780, K201185, K231711). Typically, ground truth for INR devices is established against a reference laboratory method (e.g., plasma-based PT/INR using a validated coagulometer).

    8. The Sample Size for the Training Set

    The document does not describe a "training set" in the context of machine learning model development. This device is an in-vitro diagnostic (IVD) device, not an AI/ML diagnostic tool that requires a training set for model development in the same way. The core measurement algorithm's development (if it had a "training" phase) would have occurred prior to the initial clearances.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable as this is not an AI/ML device that uses a "training set" in the described context.

    Ask a Question

    Ask a specific question about this device

    K Number
    K231711

    Validate with FDA (Live)

    Device Name
    microINR System
    Date Cleared
    2023-07-11

    (29 days)

    Product Code
    Regulation Number
    864.7750
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The microINR System measures prothrombin time (PT) expressed in International Normalized Ratio (INR), for monitoring oral anticoagulant therapy with warfarin.

    The microlNR System consists of a meter and chips (test strips) and uses fresh capillary whole blood from a fingerstick. The microlNR System is intended for patient self-testing use as well as for healthcare professionals at Point of Care settings.

    The microlNR System is intended for use in patients 18 years old or older. Patients must be stable on warfarin medication for at least 6 weeks before starting to use the microINR System.

    For Self-testing use: The System is intended for properly trained users under specific prescription of a physician. Caution: The microINR System is not in patients who are transitioning from heparin treatment to VKA therapy. The microINR System is not intended to be used for screening purposes.

    Device Description

    The modified device microINR System containing the microINR Link Meter is derived from the existing device microINR Meter. The change is for the addition of Bluetooth connectivity to the meter. Bluetooth functionality stays inactive during the testing/measuring process. All the actions related to Bluetooth functionality are set before or after performing the analytical test. The fundamental scientific technology of the modified device and its performance has not changed. No performance changes to the microINR System have been implemented since its clearance on K201185.

    AI/ML Overview

    This document (K231711) describes a Special 510(k) submission for the microINR System. This type of submission is used when changes are made to a previously cleared device, and it aims to demonstrate that the modified device remains substantially equivalent to the original cleared device. In this specific case, the change introduced is the addition of Bluetooth connectivity to the meter.

    Because this is a Special 510(k), the focus of the testing is to demonstrate that the new functionality (Bluetooth) does not adversely affect the device's original performance and does not raise new questions of safety or effectiveness. Therefore, the acceptance criteria and study described are not for the primary device performance (e.g., accuracy of INR measurement), but rather for the impact of the newly added feature.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not provide a specific table detailing acceptance criteria and reported device performance for the primary function of the microINR System (PT/INR measurement accuracy). This is because the original performance was established in the predicate K201185, and this submission focuses on the new Bluetooth functionality.

    However, it does state general acceptance regarding the performed testing:

    Acceptance Criteria CategoryReported Device Performance (Satisfied)
    RadiofrequencyAll testing results met pre-determined acceptance criteria.
    Electromagnetic CompatibilityAll testing results met pre-determined acceptance criteria.
    Electrical SafetyAll testing results met pre-determined acceptance criteria.
    Cybersecurity AnalysisAll testing results met pre-determined acceptance criteria.
    Usability TestingAll testing results met pre-determined acceptance criteria.
    Software Verification and ValidationAll testing results met pre-determined acceptance criteria.
    Risk AnalysisConcluded no significant detectable risks compared to the predicate; all residual risks deemed acceptable. Bluetooth connectivity features were correctly implemented.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample size used for the test set in the context of clinical performance or the data provenance (country of origin, retrospective/prospective). This is because the core analytical performance related to measuring INR was established with the predicate device (K201185). The testing for this Special 510(k) focused on the new Bluetooth functionality, which typically involves engineering or software testing, not patient-based clinical trials for accuracy.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. The testing described for the Bluetooth feature (Radiofrequency, EMC, Electrical Safety, Cybersecurity, Usability, Software V&V) does not typically involve expert clinical ground truth establishment in the same way that a diagnostic accuracy study would.

    4. Adjudication Method for the Test Set

    Not applicable. As noted above, the testing is for the impact of the new Bluetooth feature, not for diagnostic accuracy, and therefore does not rely on clinical adjudication of results.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    Not applicable. The microINR System is a prothrombin time test device, not an AI-assisted diagnostic imaging device, and therefore MRMC studies or AI assistance effect sizes are not relevant to this submission.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    The device is a system (meter and test strips) for measuring PT/INR. While the new Bluetooth functionality itself operates without direct human intervention once initiated, the overall device includes human interaction. The "standalone" performance here refers to the device functions for Radiofrequency, EMC, Electrical Safety, Cybersecurity, Usability, and Software verification and validation, all of which were tested without requiring a human operator to compensate for potential deficiencies in the Bluetooth module itself. The document states "Bluetooth functionality stays inactive during the testing/measuring process. All the actions related to Bluetooth functionality are set before or after performing the analytical test." This indicates the Bluetooth module's performance was assessed without interfering with the core measurement, and implicitly, its standalone function (transferring data) was verified.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    For the testing related to the Bluetooth functionality:

    • Radiofrequency, Electromagnetic Compatibility, Electrical Safety: Ground truth would be defined by relevant international and national standards and regulations (e.g., ISO, IEC standards for medical devices and radio equipment).
    • Cybersecurity Analysis: Ground truth is established by recognized cybersecurity standards, best practices, and vulnerability assessment methodologies.
    • Usability Testing: Ground truth is established by user feedback and observation against established usability heuristics and design principles for medical devices, potentially defined by standards like IEC 62366.
    • Software Verification and Validation: Ground truth is established by the software requirements specification, design documents, and coding standards, ensuring the software performs as intended according to its formal definition.

    For the primary device function (PT/INR measurement accuracy) which was established with the predicate, the ground truth would typically be a laboratory reference method for PT/INR.

    8. The Sample Size for the Training Set

    Not applicable. This device is not an AI/machine learning device that requires a "training set" in the conventional sense. The testing performed here (Radiofrequency, EMC, Electrical Safety, Cybersecurity, Usability, Software V&V) involves engineering and software validation, not the development of a predictive model.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" for this type of device and submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K201185

    Validate with FDA (Live)

    Device Name
    microINR System
    Date Cleared
    2020-12-18

    (231 days)

    Product Code
    Regulation Number
    864.7750
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The microlNR System measures prothrombin time (PT) expressed in International Normalized Ratio (INR), for monitoring oral anticoagulant therapy with warfarin.

    The microINR System consists of a microINR Meter and microINR Chip and uses fresh capillary whole blood from a fingerstick.

    The microINR System is intended for patient self-testing use as well as for healthcare professionals at Point of Care settings.

    The microlNR System is intended for use in patients 18 years old or older. Patients must be stable on warfarin medication for at least 6 weeks before starting to use the microINR System.

    For Self-testing use: The System is intended for properly trained users under specific prescription of a physician. Caution: The microINR System is not in patients who are transitioning from heparin treatment to VKA therapy. The microINR System is not intended to be used for screening purposes.

    Device Description

    The microINR® System is comprised of a portable measuring device (microINR® Meter) and test strips (microINR® Chips) in which the capillary blood sample flows through capillary action.

    The microINR® Chip contains a reagent in dried form which consists of thromboplastin, and contains two symmetrical regions, the measuring channel and a control channel. The microINR® Meter measures International Normalized Ratio (INR) based on a Prothrombin Time (PT) assay carried out in the microINR® Chip based on microfluidic technology with machine vision detection.

    The microINR® System has a multi-level On-board Quality Control. Multiple key functions and elements of the system are checked and if deviations are detected, error messages are displayed and test results are not reported.

    AI/ML Overview

    Here's a summary of the acceptance criteria and study detailed in the provided document for the microINR System:

    1. Table of Acceptance Criteria and Reported Device Performance

    The FDA 510(k) summary does not explicitly list "acceptance criteria" for the self-testing use case in a table format with pass/fail thresholds. Instead, it describes clinical studies and their outcomes designed to demonstrate that the device performs equivalently for patient self-testers as for healthcare professionals, particularly in terms of accuracy and precision. The success of the study itself serves as the demonstration of meeting the required performance for the new intended use.

    Here's a breakdown of the key performance metrics reported, framed against the implicit acceptance criterion of equivalence to healthcare professional (HCP) results and a laboratory reference:

    Performance CharacteristicAcceptance Criteria (Implicit)Reported Device Performance (Patient Self-Testers vs. HCPs/Laboratory)
    Accuracy (PST vs. HCP)INR test results from PST should be equivalent to HCP results.N = 225, Slope = 1.00, Intercept = 0.00, Correlation Coefficient (r) = 0.932. Shows equivalence.
    Accuracy (PST vs. Lab)INR test results from PST should be accurate compared to reference laboratory system.N = 112, Slope = 0.95, Intercept = 0.12, Correlation Coefficient (r) = 0.922. Shows accuracy.
    Precision (PST)Demonstrate acceptable precision for self-testing.N = 110 (paired results), Mean = 2.61, SD = 0.13, CV (%) = 5.0.

    2. Sample Size Used for the Test Set and Data Provenance

    • Accuracy (PST vs. HCP): N = 225 patients. The study was conducted at "four clinical sites" and involved patients self-testing in a "home setting for up to 2 weeks" with "2 scheduled visits to their study site." This indicates a prospective study design. The country of origin of the data is not specified.
    • Accuracy (PST vs. Laboratory System): N = 112 patients. Data collected at the second visit by PSTs was compared to a laboratory system. As part of the same study as above, this is also prospective, and the country of origin is not specified.
    • Precision (PST): N = 110 patients (paired results). Data collected by patient self-testers at the second visit to the clinical sites. This is also prospective, and the country of origin is not specified.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    For the ground truth comparison:

    • Healthcare Professionals (HCPs): The study compares patient self-tester results to "healthcare professionals using the microINR® System." While the exact number of HCPs is not stated, it implies multiple professionals across the four clinical sites. Their specific qualifications (e.g., radiologist with 10 years of experience) are not provided, beyond being "healthcare professionals."
    • Laboratory System: The "ACL TOP 500" is cited as the reference laboratory system. This is an automated coagulation analyzer, implying the ground truth was established by a validated laboratory method rather than individual expert consensus.

    4. Adjudication Method for the Test Set

    The document does not describe an explicit adjudication method (like 2+1 or 3+1) for resolving discrepancies in the test set. Instead, it's a direct comparison of the microINR results from patient self-testers against the microINR results from healthcare professionals and against a reference laboratory system.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    This is not applicable. The microINR System is a diagnostic medical device for measuring prothrombin time (PT) expressed as INR, not an AI-assisted diagnostic tool for interpretation by human readers. Therefore, an MRMC comparative effectiveness study where human readers improve with AI assistance is not relevant to this device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, in a way. The device itself, the microINR System, is the "standalone algorithm" that measures INR. The studies evaluate its performance (accuracy, precision) in the hands of two different user groups (patient self-testers and healthcare professionals) and against a laboratory reference, effectively assessing its standalone performance for each user group.

    7. The Type of Ground Truth Used

    The ground truth used for the clinical studies was a combination of:

    • Clinical Reference (HCP results): INR results obtained by trained healthcare professionals using the microINR System. This serves as a "human expert" benchmark for the device's performance when operated by trained professionals.
    • Laboratory Reference Method: INR results obtained from a standard "laboratory system (ACL TOP 500)." This is an established independent gold standard for PT/INR measurement.

    8. The Sample Size for the Training Set

    The document does not explicitly state a "training set" sample size for the new intended use (patient self-testing). The microINR System was previously cleared, meaning its core algorithm for INR measurement was already established. The studies for this 510(k) are focused on validating its performance in the hands of self-testers, rather than "training" a new algorithm. The document mentions that "following training, the patients self-tested." This implies the patients themselves received training on how to use the device, not that their data was used to train the device's internal algorithm.

    9. How the Ground Truth for the Training Set Was Established

    Given that this 510(k) primarily addresses a new user population for an already cleared device, there isn't a "training set" in the sense of machine learning algorithm development. The core measurement principle and validation were established under the previous clearance (K180780), which would have involved its own validation studies against established reference methods. For the current submission, the focus is on demonstrating that trained patients can achieve equivalent results, with ground truth established as described in point 7.

    Ask a Question

    Ask a specific question about this device

    K Number
    K180780

    Validate with FDA (Live)

    Device Name
    microINR System
    Date Cleared
    2019-01-25

    (305 days)

    Product Code
    Regulation Number
    864.7750
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    microINR System:

    The microlNR System (consisting of the microINR Chip) is intended for multiple-patient use by professional healthcare providers for the determination of International Normalized Ratio (INR) to monitor Oral Anticoagulation Therapy (OAT) warfarin. The microINR System uses fresh capillary whole blood. The microNR System is intended for in vitro diagnostic use at the point-of-care.

    The microlNR System is intended for use in patients 18 years of age and older. Patients must be stabilized (≥ 6 weeks) on warfarin.

    Caution: The microINR System is not intended for use in patients who are transitioning from heparin treatment to warfarin therapy. The microINR System is not intended to be used for screening purposes.

    The microINR chips are only intended to be used with the microINR meter.

    microINR Control Level A / microINR Control Level B:

    The microlNR Control is intended for quality control performed on the microINR Meter when used with the disposable microINR Chips.

    iLine Microsystems has available two microNR Control Levels (A and B) to perform quality control checks in the therapeutic range on the microINR System.

    The microINR Control is intended for professional use only.

    Device Description

    The microINR® System is comprised of a portable measuring device (microINR® Meter) and test strips (microINR® Chips) in which the capillary blood sample flows through capillary action.

    The microINR® Chip contains a reagent in dried form which consists of thromboplastin, and contains two symmetrical regions, the measuring channel and a control channel. The microINR® Meter measures International Normalized Ratio (INR) based on a Prothrombin Time (PT) assay carried out in the microINR® Chip based on microfluidic technology with machine vision detection.

    The microINR® System has a multi-level On-board Quality Control. Multiple key functions and elements of the system are checked and if deviations are detected, error messages are displayed and test results are not reported.

    In addition to the microINR® On-Board Controls, the microINR® Control has been developed as an optional liquid quality control solution. The microINR® Control must be used only by healthcare professionals in order to meet the regulatory compliance requirements applicable to the facility where they are to be used.

    The microINR® Control contains lyophilized human citrated plasma from healthy donors (not heparinized plasma or plasma samples under oral anticoagulant therapy) modified by means of a dedicated process to simulate an abnormal coagulation sample and a calcium chloride solution. Before use, the lyophilized plasma must be reconstituted with the calcium chloride solution.

    AI/ML Overview

    Here's an analysis of the provided text to extract information about the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't explicitly state "acceptance criteria" for accuracy in a tabulated format that is distinct from the reported performance. The "Performance Characteristics" section outlines various analytical performance aspects, and the "Method comparison with Laboratory Reference device" section provides data related to accuracy. We can infer the performance metrics from this section.

    Performance CharacteristicAcceptance Criteria (Inferred from context and predicate comparison)Reported Device Performance (microINR® System)
    Accuracy (Method Comparison with Laboratory Reference Device):A strong correlation to the reference method (e.g., Pearson (r) close to 1) and acceptable slope/intercept.N: 260 Slope: 1.04 Intercept: 0.02 Pearson (r): 0.978
    Repeatability (INR < 2.0)Low SD and CV% (indicating good precision)n test pairs: 91SD: 0.06 CV%: 5.25
    Repeatability (2.0 ≤ INR < 3.5)Low SD and CV% (indicating good precision)n test pairs: 129SD: 0.12 CV%: 4.57
    Repeatability (3.5 ≤ INR < 4.5)Low SD and CV% (indicating good precision)n test pairs: 32SD: 0.21 CV%: 5.64
    Repeatability (INR ≥ 4.5)Low SD and CV% (indicating good precision)n test pairs: 17SD: 0.26 CV%: 5.46
    Device Reproducibility (High Abnormal INR level plasma)Low SD and CV% for within-run, between-run, between-day, between-device, within-device and overall reproducibility.N: 90 Mean: 3.45 Device Reproducibility SD: 0.19 Device Reproducibility CV%: 5.5
    Device Reproducibility (Low Abnormal INR level plasma)Low SD and CV% for within-run, between-run, between-day, between-device, within-device and overall reproducibility.N: 90 Mean: 2.25 Device Reproducibility SD: 0.12 Device Reproducibility CV%: 5.5
    Hematocrit RangeNo significant effect on test results within the specified range.25 - 55% does not significantly affect test results.
    Bilirubin InterferenceNo significant effect up to the specified concentration.Up to 40 mg/dL has no significant effect on test results.
    Triglyceride InterferenceNo significant effect up to the specified concentration.Up to 3270 mg/dL has no significant effect on test results.
    Reference RangeEstablished normal range95% of INR results ranged between 0.8-1.2 in healthy subjects.

    2. Sample size used for the test set and the data provenance:

    • Accuracy Study (Method Comparison):
      • Sample Size: 260 capillary samples from patients (assumed to be from patients on warfarin therapy, based on the study's purpose).
      • Data Provenance: The document does not explicitly state the country of origin. It indicates that the study was performed at "three clinical sites" for the repeatability study which is likely part of the overall clinical evaluation. It is a prospective study as fresh capillary samples were tested.
    • Repeatability Study:
      • Sample Size: 269 paired tests (meaning 269 subjects, with two fingersticks each)
      • Data Provenance: Performed at "three clinical sites". Subjects included "normal subjects and patients on warfarin therapy." Likely prospective.
    • Reproducibility Study:
      • Sample Size: 90 (N=90 for each plasma level, though the number of individual devices/runs contributes to this count).
      • Data Provenance: Performed using "two levels of plasma samples". No country of origin mentioned. Likely prospective laboratory study.
    • Reference Range Study:
      • Sample Size: 134 healthy subjects.
      • Data Provenance: Conducted across "three sites." Likely prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    The document does not mention the use of experts to establish ground truth for the test set in the context of diagnostic interpretation.

    For the Accuracy Study (Method Comparison), the "ground truth" (reference method) was established by the Instrumentation Laboratory ACL TOP® 500 laboratory analyzer using HemosIL® RecombiPlasTin 2G recombinant human tissue thromboplastin reagent. This is a recognized laboratory method, not human expert consensus.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    No adjudication method for human interpretation is mentioned, as the primary comparison is against a reference laboratory instrument, not human readers.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    No MRMC study was performed or described. This device is an in-vitro diagnostic system that directly measures INR, not an AI-assisted diagnostic imaging or interpretation tool for human readers.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    Yes, the studies described are of the device in a standalone capacity. The microINR® System itself performs the measurement and displays the INR result. The "Performance Characteristics" and "Comparison Studies" sections detail the analytical performance of the device without human intervention impacting the measurement itself, other than proper sample application.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The ground truth used for the accuracy study was the measurement from a validated laboratory reference device:

    • Instrumentation Laboratory ACL TOP® 500 laboratory analyzer using HemosIL® RecombiPlasTin 2G recombinant human tissue thromboplastin reagent.

    For other analytical studies like precision and reproducibility, the ground truth is often inherent in the repeatability/reproducibility metrics themselves (i.e., how consistently the device measures its own result) or against certified control materials.

    8. The sample size for the training set:

    The document does not explicitly describe a separate "training set" in the context of algorithm development. In IVD devices like this, the device is typically developed and validated using a structured R&D process, and these performance studies represent the final validation. If any internal algorithm development or calibration involved data, it's not disclosed as a distinct "training set" in the submission summary.

    9. How the ground truth for the training set was established:

    As no explicit training set is detailed, information on how its ground truth was established is not available in the provided text. The calibration of the microINR® Chips is done against a reference lot of human recombinant thromboplastin traceable to the World Health Organization International Reference Preparation. This could be considered part of the "ground truth" establishment for the device's inherent measurement capability.