510(k) Data Aggregation

    K Number
    K240053

    Validate with FDA (Live)

    Device Name
    Venue Go
    Date Cleared
    2024-06-10

    (154 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Venue Go is a general purpose diagnostic ultrasound system for use by qualified and trained healthcare professionals or practitioners that are legally authorized or licensed by law in the country, state or other local municipality in which he or she practices, for ultrasound imaging, measurement, display and analysis of the human body and fluid. The users may or may not be working under supervision or authority of a physician. Users may also include medical students working under the supervision or authority of a physician during their education / training. Venue Go is intended to be used in a hospital or medical clinic. Venue Go clinical applications include: abdominal (GYN and Urology), thoracic/pleural, ophthalmic, Fetal/OB, Small Organ (including breast, testes, thyroid), Vascular/Peripheral vascular, neonatal and adult cephalic, pediatric, musculoskeletal (conventional and superficial), cardiac (adults and pediatric), Transrectal, Transvaginal, Transesophageal, Intraoperative (vascular) and interventional guidance (includes tissue biopsy, fluid drainage, vascular and non-vascular access). Modes of operation include: B, M, PW Doppler, CW Doppler, Color Doppler, Color M Doppler, Power Doppler, Harmonic Imaging, Coded Pulse and Combined modes: B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD, B/CWD, B/Color/CWD.

    Device Description

    Venue Go is a general-purpose diagnostic ultrasound system intended for use by qualified and trained healthcare professionals to evaluate the body by ultrasound imaging and fluid flow analysis.

    The Venue Go is a compact, portable system with a small footprint. The system can be hand carried using an integrated handle, placed on a horizontal surface, attached to a mobile cart or wall mounted. It has a high resolution color LCD monitor, with a simple, multi-touch user interface that makes the system intuitive. The system can be powered through an electrical wall outlet for long term use or from an internal battery for a short time with full functionality and scanning.

    The Venue Go utilizes a variety of linear, convex, and phased array transducers which provide high imaging performance and support standard acquisition modes. Compatible biopsy kits can be used for needle-guidance procedures.

    The system is capable of displaying the patient's ECG trace synchronized to the scanned image. This allows the user to view an image from a specific time of the ECG signal which is used as an input for gating during scanning. The ECG signal can be input directly from the patient or as an output from an ECG monitoring device. ECG information is not intended for monitoring or diagnosis.

    A barcode reader and RFID scanner are available as additional input devices. A roller bag will also be available for the customer to use when transporting the system.
    Venue Go is capable of wireless communication through a built-in Wireless LAN device. The system meets DICOM requirements to support image storage and archiving (local PACS or products such as Q-Path) and allows for output to printing devices.

    AI/ML Overview

    The provided FDA 510(k) summary for the GE Venue Go ultrasound system does not include acceptance criteria for artificial intelligence (AI) performance or the study details to prove the device meets such criteria.

    The document primarily focuses on demonstrating substantial equivalence to a predicate device (Venue Go K220800) and other reference devices, based on:

    • Identical fundamental scientific technology.
    • Similar imaging capabilities, technological characteristics, safety, and effectiveness.
    • Minor updates to the Indications for Use statement for clarity regarding operator qualifications.
    • Addition of new transducers already cleared on other GE devices.
    • New features/functionality such as Auto Volume Flow (AVF), Bladder Volume Tool, expanded MSK diagrams, Venue Coach, electronic software delivery, expanded Auto Inferior Vena Cava (IVC), Auto B-Lines, Real-Time ejection fraction (RT-EF), cNerve for Vscan Air CL and SL probes (which are already cleared with these AI features on predicate Vscan Air K231301 and reference LOGIQ E10s/LOGIQ Fortis K231989), and AppAPI functionality.
    • Conformity to recognized performance standards for acoustic output, biocompatibility, cleaning/disinfection, thermal, electrical, electromagnetic, and mechanical safety.
    • Quality assurance measures like risk analysis, requirements/design reviews, and various levels of testing.

    Specifically, the document states: "The subject of this premarket submission, Venue Go, did not require clinical studies to support substantial equivalence." This indicates that any evaluation of "AI performance" as typically understood (e.g., using a test set with ground truth, performance metrics, and acceptance criteria) was not part of this 510(k) submission.

    The AI-related features mentioned (Auto IVC, Auto B-Lines, RT-EF, cNerve) are noted as being "expanded" or already "cleared" on predicate/reference devices (K231301 Vscan Air, K231989 LOGIQ E10s/LOGIQ Fortis). This implies their performance was likely evaluated as part of those previous 510(k) submissions, and the current submission is leveraging that prior clearance for their inclusion in the Venue Go system when used with the particular probes.

    Therefore, since this 510(k) does not present new AI performance study data for the Venue Go, I cannot extract the requested information. The document focuses on demonstrating that the Venue Go, with its included features, is substantially equivalent to existing cleared devices, rather than proving novel AI algorithm performance.

    Ask a Question

    Ask a specific question about this device

    K Number
    K220800

    Validate with FDA (Live)

    Device Name
    Venue Go
    Date Cleared
    2022-06-21

    (95 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Venue Go is a general purpose diagnostic ultuse by qualified and trained healthcare professionals for ultrasound imaging, measurement, display and analysis of the human body and fluid. Venue Go is intended to be used in a hospital or medical clinic. Venue Go clinical applications include: abdominal (GYN and Urology), thoracic/pleural, ophthalmic, Fetal/OB, Small Organ (including breast, thyroid), Vascular/Peripheral vascular, neonatal and adult cephalic, pediatric, musculoskeletal (conventional and superficial), cardiac (adults and pediatic), Transrectal, Transvaginal, Transesophageal, Intraoperative (vascular) and interventional guidance (includes tissue biopsy, fluid drainage, vascular and non-vascular access). Modes of operation include: B, M, PW Doppler, CW Doppler, Color Doppler, Color M Doppler, Power Doppler, Harmonic Imaging, Coded Pulse and Combined modes: B/M, B/Color M, B/ PWD, B/Color/PWD, B/Power/PWD, B/CWD, B/Color/CWD.

    Device Description

    Venue Go is a general-purpose diagnostic ultrasound system intended for use by qualified healthcare professionals to evaluate the body by ultrasound imaging and fluid flow analysis. The Venue Go is a compact, portable system with a small footprint. The system can be hand carried using an integrated handle, placed on a horizontal surface, attached to a mobile cart or wall mounted. It has a high resolution color LCD monitor, with a simple, multi-touch user interface that makes the system intuitive. The system can be powered through an electrical wall outlet for long term use or from an internal battery for a short time. The Venue Go utilizes a varietv of linear, convex, and phased array transducers which provide high imaging performance and support standard acquisition modes. Compatible biopsy kits can be used for needle-guidance procedures. The system is capable of displaying the patient's ECG trace synchronized to the scanned image. This allows the user to view an image from a specific time of the ECG signal which is used as an input for gating during scanning. The ECG signal can be input directly from the patient or as an output from an ECG monitoring device. ECG information is not intended for monitoring or diagnosis. A barcode reader and RFID scanner are available as additional input devices. A roller bag will also be available for the customer to use when transporting the system. Venue Go is capable of wireless communication through a builtin Wireless LAN device. The system meets DICOM requirements to support image storage and archiving (local PACS or products such as Q-Path) and allows for output to printing devices. The user documentation is available electronically.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification Submission for the GE Venue Go ultrasound system. It includes a specific section detailing the AI feature cNerve. Based on this information, here's a description of the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance (cNerve AI Feature)

    Acceptance Criterion (cNerve)Reported Device Performance (cNerve)
    Overall cNerve Performance (Sequence Accuracy)
    At least 70% of the sequences are meaningfully detected.Performance not explicitly stated as a percentage for "meaningfully detected sequences" in the provided document, but implied by the "successfully detected" metric.
    At least 80% of the meaningfully detected sequences are successfully detected (meeting frame accuracy criteria).The document states performance requirements, but the actual achieved performance data (e.g., "cNerve achieved X% of meaningfully detected sequences" or "cNerve successfully detected Y% of these") against these specific criteria are not numerically reported in the "AI Summary of Testing" section. It only states what the performance requirements are, not what was measured against those requirements.
    Pixel Accuracy in Frames (Specific thresholds for appropriate intended use derived from preliminary survey)Specific numerical thresholds for pixel accuracy are not provided in the document.
    Frame Accuracy in Sequences (Specific thresholds for appropriate intended use derived from preliminary survey)Specific numerical thresholds for frame accuracy are not provided in the document.

    Important Note: The document outlines the acceptance criteria for cNerve but does not explicitly provide the measured numerical performance results of the device against these criteria. It states that the "cNerve performance requirements" are listed but does not follow up with a section detailing the actual tested performance numbers.

    2. Sample Size and Data Provenance for the Test Set

    • Sample Size (Individuals): A total of 44 individuals contributed to the verification dataset.
    • Sample Size (Sequences/Frames): The test dataset included 124 sequences and 3776 frames.
    • Relationship between Samples: Each individual contributed up to 2 sequences per view location (often both left and right laterals were scanned).
    • Demographic Distribution:
      • Gender: Male and Female
      • Age: 18-82 years
      • Ethnicity/Country: USA, Japan, Israel
    • Subgroups Tested: The algorithm performance was verified via frame accuracy on all demographic subgroups: Gender (M/F), Age (<60 / >=60), BMI (<=25 / >25). It was also tested for all supported nerve block locations and supported probe types.
    • Data Provenance: The document implies the data was collected from a mix of clinical settings (implied by "USA, Japan, Israel" for ethnicity/country) and clinical scenarios ("all supported nerve block locations and all supported probe types"). The study type (retrospective or prospective) is not explicitly stated, but the process of collecting and annotating an existing dataset suggests a retrospective approach for the dataset creation for validation.

    3. Number of Experts and Qualifications for Ground Truth

    • Number of Experts: A single clinical expert annotated the frames from scouting sequences.
    • Qualifications of Experts: The specific qualifications (e.g., years of experience, specific specialty like "radiologist") of the "single clinical expert" are not detailed in the provided text.

    4. Adjudication Method for the Test Set

    • The ground truth was established by a single clinical expert. Therefore, there was no adjudication method described (e.g., 2+1, 3+1 concensus), as only one expert was involved in marking the ground truth.

    5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study

    • The document does not indicate that an MRMC comparative effectiveness study was done to evaluate how much human readers improve with AI vs without AI assistance. The study focuses solely on the standalone performance measurement of the cNerve algorithm against defined ground truth.

    6. Standalone (Algorithm Only) Performance Study

    • Yes, a standalone performance study was done. The "AI Summary of Testing" section describes the evaluation of the cNerve algorithm's accuracy, focusing on its ability to detect and track nerve bundles independently. The acceptance criteria ("Sequence accuracy requirement - for testing overall cNerve performance") explicitly relate to the algorithm's performance.

    7. Type of Ground Truth Used

    • The ground truth used was expert annotation/consensus (from a single expert). The frames were "annotated by a single clinical expert, where the anatomical area of the nerves was marked in each frame."

    8. Sample Size for the Training Set

    • The sample size for the training set is not provided in the document. The document explicitly states: "The data used for verification is completely distinct from that used during training process and there is no overlap between the two." However, it does not disclose details about the training data itself.

    9. How the Ground Truth for the Training Set was Established

    • The document does not describe how the ground truth for the training set was established. It only mentions that the test data was distinct from the training data.
    Ask a Question

    Ask a specific question about this device

    K Number
    K202233

    Validate with FDA (Live)

    Device Name
    Venue Go
    Date Cleared
    2020-11-18

    (103 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Venue Go is a general purpose diagnostic ultrasound system for use by qualified and trained healthcare professionals for ultrasound imaging, measurement, display and analysis of the human body and fluid. Venue Go is intended to be used in a hospital or medical clinic. Venue Go clinical applications include: abdominal (GYN and Urology), thoracic/pleural, ophthalmic, Fetal/OB, Small Organ (including breast, thyroid), VascularPeripheral vascular, neonatal and adult cephalic, pediatric, musculoskeletal (conventional and superficial), cardiac (adults and pediatric), Transrectal, Transvaginal, Transesophageal, Intraoperative (vascular) and interventional guidance (includes tissue biopsy, fluid drainage, vascular and non-vascular access). Modes of operation include: B, M, PW Doppler, Color Doppler, Color M Doppler, Power Doppler, Harmonic Imaging, Coded Pulse and Combined modes: B/M, B/Color M, B/ PWD, B/Color/PWD, B/Power/PWD, B/CWD, B/Color/CWD.

    Device Description

    Venue Go is a general-purpose diagnostic ultrasound system intended for use by qualified and trained healthcare professionals to evaluate the body by ultrasound imaging and fluid flow analysis. The Venue Go is a compact, portable system with a small footprint The system can be hand carried using an integrated handle, placed on a horizontal surface, attached to a mobile cart or wall mounted. It has a high resolution color LCD monitor, with a simple, multi-touch user interface that makes the system intuitive. The system can be powered through an electrical wall outlet for long term use or from an internal battery for a short time.

    The Venue Go utilizes a variety of linear, convex, and phased array transducers which provide high imaging capability, supporting all standard acquisition modes. Some transducers are compatible with OEM biopsy kits to support needle-guidance procedures.

    The system has the capability for displaying the patient's ECG trace synchronized to the scanned image. This allows the user to view an image from a specific time of the ECG signal. The ECG signal can be input directly from the patient leads or as an output from an ECG monitoring device. ECG information is not intended for monitoring or diagnosis.

    A barcode reader is available to be used as an input device. Venue Go is capable of wireless communication through a builtin Wireless LAN device. The system meets DICOM requirements to support users image storage and archiving needs (local PACS or products such as O-Path) and allows for output to printing devices. The user documentation is available electronically. An additional accessory that will also be available for the customer will be a roller bag.

    AI/ML Overview

    The provided text describes the GE Medical Systems Ultrasound and Primary Care Diagnostics' Venue Go device and its 510(k) premarket notification. However, it does not include information about acceptance criteria or a study proving the device meets those criteria, as typically understood in the context of diagnostic performance (e.g., sensitivity, specificity, accuracy).

    Instead, the document details the device's technical specifications, intended use, comparison to predicate devices, and compliance with non-clinical safety standards (acoustic output, biocompatibility, electrical safety, etc.). The "Summary of Non-Clinical Tests" section outlines various standards the device complies with, but these are safety and engineering standards, not performance criteria for a diagnostic algorithm.

    Specifically, the "Summary of Clinical Tests" section explicitly states: "The subject of this premarket submission, Venue Go, did not require clinical studies to support substantial equivalence." This means that no clinical performance study was conducted to establish diagnostic accuracy metrics for the device itself. The substantial equivalence argument is based on its similarity to existing, legally marketed ultrasound systems.

    Therefore, I cannot provide the requested information regarding acceptance criteria and a study proving their fulfillment, as it is not present in the provided text.

    If you have a document that describes the diagnostic performance claims and supporting studies for the Venue Go or a similar medical device, please provide that text, and I will be able to answer your questions.

    Ask a Question

    Ask a specific question about this device

    K Number
    K183362

    Validate with FDA (Live)

    Device Name
    Venue GO
    Date Cleared
    2019-01-28

    (55 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Venue Go is a general-purpose diagnostic ultrasound system intended for use by qualified healthcare professionals for ultrasound imaging, measurement and analysis of the human body and fluid. Venue Go clinical applications include the following: Abdominal (including Gynecology and Urology), Thoracic/Pleural, Ophthalmic, Fetal/Obstetrics, Small Organ (including breast, testes, thyroid), Peripheral vascular, Adult and neonatal cephalic, Musculoskeletal (Conventional and Superficial), Cardiac (Adults and Pediatric), Transesophageal, Transrectal, Transvaginal, Intraoperative and Imaging guidance for interventional procedures (e.g. Nerve block, biopsy, vascular access).

    Device Description

    Venue Go is a general-purpose diagnostic ultrasound system intended for use by qualified healthcare professionals to evaluate the body by ultrasound imaging and fluid flow analysis. The Venue Go is a compact, portable system with a small footprint. The system can be hand carried using an integrated handle, placed on a horizontal surface, attached to a mobile cart or wall mounted. It has a high resolution color LCD monitor, with a simple, multi-touch user interface that makes the system intuitive. The system can be powered through an electrical wall outlet for long term use or from an internal battery for a short time.

    The Venue Go utilizes a variety of linear, convex, and phased array transducers which provide high imaging performance and support standard acquisition modes. Some biopsy kits are available for needle-guidance procedures.

    The system has the capability for displaying the patient's ECG trace synchronized to the scanned image. This allows the user to view an image from a specific time of the ECG signal. The ECG signal can be input directly from the patient or as an output from an ECG monitoring device. ECG information is not intended for monitoring or diagnosis.

    A barcode reader is available to be used as an input device. Venue Go is capable of wireless communication through a builtin Wireless LAN device. The system meets DICOM requirements to support users image storage and archiving needs (local PACS or products such as O-Path) and allows for output to printing devices. The user documentation is available electronically. An additional accessory that will also be available for the customer will be a roller bag.

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification Submission for the GE Venue Go Ultrasound System. This type of submission is used to demonstrate that a device is substantially equivalent to a legally marketed predicate device. As such, it primarily focuses on comparing the new device against existing ones rather than detailing a study that proves the new device meets specific, quantitative acceptance criteria for clinical performance.

    The document states: "The subject of this premarket submission, Venue Go, did not require clinical studies to support substantial equivalence." This means that traditional clinical performance studies with specific quantitative acceptance criteria or comparative effectiveness studies (like MRMC studies) demonstrating improvement with AI assistance were not performed or submitted for this specific premarket notification.

    Instead, the submission relies on demonstrating substantial equivalence to predicate devices (Venue, LOGIQ e, LOGIQ P9) through:

    • Identical or similar intended use and clinical applications.
    • Similar imaging capabilities and technological characteristics.
    • Compliance with applicable medical device safety standards (e.g., electrical, electromagnetic, mechanical safety, biocompatibility, acoustic output).
    • Quality assurance measures applied during development (risk analysis, requirements reviews, design reviews, various levels of testing: unit, integration, performance, safety).

    Therefore, I cannot provide a table of acceptance criteria and reported device performance from a clinical study for this specific device, nor can I answer questions about sample sizes, expert ground truth establishment, adjudication methods, or MRMC studies, as these types of clinical performance evaluations were explicitly not required or conducted for this 510(k) submission.

    The document indicates that new indications or features are being added to the device (e.g., Intraoperative indication, new transducers and imaging modes). For these new features, substantial equivalence is established by showing that they were previously cleared on other predicate devices. For instance:

    • The Intraoperative indication was cleared on LOGIQ e (K151028).
    • L4-12t-RS, L8-18i-RS, and 6S-RS transducers were cleared in LOGIQ e (K151028).
    • The Neonatal Cephalic application for L8-18i-RS was cleared in LOGIQ P9 (K181783).
    • New applications for L4-12t-RS and 6S-RS, and new imaging modes like "Color M Doppler" and "Combined Modes" were cleared in LOGIQ P9 (K181783) or existing predicates.

    In summary, the "acceptance criteria" for a 510(k) device like this are primarily demonstrated through substantially equivalent performance to existing, legally marketed predicate devices by showing similar features, technology, and compliance with recognized safety standards in non-clinical tests, rather than through new clinical performance studies with quantitative metrics.