Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K240948

    Validate with FDA (Live)

    Date Cleared
    2024-06-06

    (59 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TracStar LDP Large Distal Platform is indicated for the introduction of interventional devices into the peripheral. coronary, and neuro vasculature.

    The Zoom 88 Large Distal Platform and Zoom 88 Large Distal Platform Support are indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature.

    Device Description

    The Imperative Care Large Distal Platform™ (LDP) Catheters include the Zoom™ 88 Large Distal Platform™, Zoom™ 88 Large Distal Platform™ Support, and TracStar™ LDP Large Distal Platform. The LDP Catheters are 0.038" diameter or smaller guidewire compatible single lumen guide catheters that provide access to peripheral, coronary and neuro vasculature. The catheters are comprised of a hollow cylindrical tube bonded at the proximal end to a standard luer fitting. The wall of the tube is constructed using a combination of metal coils/braids and medical grade polymers. The distal section of each catheter has a hydrophilic coating to enhance tracking through tortuous vasculature. An angled distal soft tip facilitates smooth tracking past vessel branches. A radiopaque marker provides visual confirmation of the distal tip location under fluoroscopy. The LDP Catheters have an inner diameter (ID) of 0.088" (6F compatible), and a maximum outer diameter (OD) of 0.110". The LDP guide catheters are packaged with a rotating hemostasis valve (RHV) that is attached to the proximal luer to help maintain hemostasis.

    AI/ML Overview

    This document describes the acceptance criteria and the study conducted to demonstrate that the TracStar™ LDP Large Distal Platform, Zoom™ 88 Large Distal Platform, and Zoom™ 88 Large Distal Platform Support devices meet these criteria.

    1. Table of Acceptance Criteria and Reported Device Performance

    The device is a medical catheter, and the performance evaluation involved a series of bench and laboratory (in-vitro) tests to ensure its safety and effectiveness. The acceptance criteria are "Pass" for all tested attributes, meaning the device met the specified performance requirements.

    Test AttributeSpecificationReported Device Performance
    Delivery, Compatibility, and Retraction (Trackability)The catheter shall be able to be delivered, deployed, and retracted per the IFU within a simulated neurological model without incurring any damage to the catheter.Pass
    Tip FlexibilityThe catheter distal tip flexibility shall be comparable to the predicate.Pass
    Visual InspectionThe catheter shall meet visual inspection criteria. The printing on the strain relief must be legible.Pass
    Dimensional (Distal ID, Proximal ID, Distal OD, Proximal OD)All defined catheter dimensions are within the specified tolerances.Pass
    Catheter Bond StrengthThe catheter shall have sufficient bond strengths to remain intact throughout a procedure.Pass
    Freedom from Leakage - Positive PressureThe catheter must remain leak free under specified test conditions.Pass
    Freedom from Leakage - Negative PressureThe catheter must remain leak free under specified test conditions.Pass
    Burst PressureThe catheter must withstand pressure testing under dynamic flow conditions.Pass
    Static BurstThe catheter shall meet criteria for static burst pressure testing.Pass
    Catheter Torque StrengthWith the catheter tip constrained from movement, the proximal end was rotated until failure. The catheter shall not be damaged when rotated at least two (2) full rotations (720 degrees).Pass
    Kink ResistanceThere shall be no kinking of the catheter shaft around respective clinically relevant minimum bend radii in distal tip, medial and proximal locations.Pass
    FlexibilityThe catheters need to have acceptable flexure values for tracking in the vasculature.Pass
    Proximal Shaft StiffnessThe stiffness of the proximal shaft was evaluated to ensure comparable stiffness to the predicate.Pass
    Delivery ForceThe catheters shall not be too stiff or require excessive force to safely navigate and track to the target neurovasculature.Pass
    Compatibility with other Devices (External)The catheters shall be able to be delivered through the minimum introducer sheath or guide catheter size indicated in the product labeling.Pass
    Guidewire CompatibilityThe catheters shall be able to be delivered over the maximum size guidewire indicated in the product labeling.Pass
    Interventional Device Compatibility (internal)The catheters shall be able to accommodate other interventional devices (e.g., support catheter, diagnostic catheter) up to the maximum size indicated in the product labeling.Pass
    Luer CompatibilityDevices and accessories shall be compatible with standard syringe luer fittings per ISO 80369-7.Pass
    Accessory CompatibilityDevices shall be compatible with an RHV.Pass
    Coating - ParticulateThe amount of particulate matter generated during simulated use testing shall be determined and compared to competitive products and techniques.Pass
    Coating - Lubricity, Durability and IntegrityCoating must be lubricious with a specified average pull force. There were no coating anomalies or significant wear observed post simulated use.Pass

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the exact sample sizes for each test attribute. The study relies on bench and laboratory (in-vitro) testing, which typically involves testing a defined number of device units or components to statistical significance or to demonstrate compliance with a specification. The data provenance is prospective as the testing was conducted specifically for this submission to evaluate the subject devices. The country of origin of the data is not explicitly stated but is implied to be where Imperative Care, Inc. conducts its R&D and testing.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This information is not applicable as the study involves bench and laboratory (in-vitro) testing of physical device properties against established engineering and performance specifications, not human interpretation of medical data. Therefore, there is no "ground truth" established by medical experts in this context.

    4. Adjudication Method for the Test Set

    Not applicable, as this was not a human-reader study requiring adjudication of expert opinions. The adjudication for bench testing is typically based on whether the test results meet or exceed the predefined pass/fail criteria.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study is relevant for diagnostic imaging AI systems where human readers interpret medical images. The current submission pertains to the substantial equivalence of a medical catheter based on physical and mechanical performance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This question is not applicable to the device described. The device is a physical medical catheter, not a software algorithm. The "standalone" performance refers to the physical and mechanical performance of the catheter itself, which was evaluated through bench and laboratory testing without human "in-the-loop" performance in the context of interpretation or decision-making.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's evaluation is defined by established engineering and performance specifications based on standards such as ISO 10555-1, as well as internal design requirements. These specifications act as the objective criteria against which the device's performance is measured.

    8. The Sample Size for the Training Set

    This information is not applicable. The device is a physical medical catheter, not a machine learning or AI algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K231168

    Validate with FDA (Live)

    Date Cleared
    2023-05-22

    (27 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TracStar LDP Large Distal Platform is indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature.

    The Zoom 88 Large Distal Platform and Zoom 88 Large Distal Platform Support are indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature.

    Device Description

    The Imperative Care Large Distal Platform (LDP) Catheters include the Zoom™ 88 Large Distal Platform, Zoom™ 88 Large Distal Platform Support, and TracStar™ LDP Large Distal Platform. The LDP Catheters are 0.038" diameter or smaller guidewire compatible single lumen guide catheters that provide access to peripheral, coronary and neuro vasculature. The catheters are comprised of a hollow cylindrical tube bonded at the proximal end to a standard luer fitting. The wall of the tube is constructed using a combination of metal coils/braids and medical grade polymers. The distal section of each catheter has a hydrophilic coating to enhance tracking through tortuous vasculature. An angled distal soft tip facilitates smooth tracking past vessel branches. A radiopaque marker provides visual confirmation of the distal tip location under fluoroscopy. LDP Catheters have an inner diameter of 0.088" (6F compatible), and a maximum outer diameter of 0.110". The LDP guide catheters are packaged with a rotating hemostasis valve (RHV) that is attached to the proximal luer to help maintain hemostasis.

    AI/ML Overview

    This FDA 510(k) Summary describes a device (TracStar LDP Large Distal Platform, Zoom 88 Large Distal Platform, Zoom 88 Large Distal Platform Support) intended for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature. The submission is for a line extension and manufacturing process improvements, referencing a predicate device (K212224). Therefore, the study details provided are focused on demonstrating substantial equivalence to the predicate device, rather than a de novo clinical trial for a new product with absolute performance metrics against a disease state.

    Here's a breakdown of the requested information based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for this 510(k) submission are based on the performance of the device against a set of predefined specifications derived from an existing standard (ISO 10555-1 for catheter performance) and internal engineering requirements. The reported device performance is simply "Pass" for all tested attributes, indicating that the subject device met these specifications.

    Test AttributeAcceptance CriteriaReported Device Performance
    Delivery, Compatibility, and Retraction (Trackability)The catheter shall be able to be delivered, deployed, and retracted per the IFU within a simulated neurological model without incurring any damage to the catheter.Pass
    Flexibility and Kink ResistanceThere shall be no kinking of shaft (permanent deformation) after simulated use.Pass
    Compatibility with other Devices (external)The catheters shall be able to be delivered through the minimum introducer sheath or guide catheter size indicated in the product labeling.Pass
    Guidewire CompatibilityThe catheters shall be able to be delivered over the maximum size guidewire indicated in the product labeling.Pass
    Interventional Device Compatibility (internal)The catheters shall be able to accommodate other interventional devices (e.g., support catheter, diagnostic catheter) up to the maximum size indicated in the product labeling.Pass
    Luer CompatibilityDevices and accessories shall be compatible with standard syringe luer fittings per ISO 80369-7.Pass
    Accessory CompatibilityDevices shall be compatible with an RHV.Pass
    Catheter Bond StrengthThe catheter shall have sufficient bond strengths to remain intact throughout a procedure.Pass
    Dimensional (Distal ID, Proximal ID, Distal OD, Proximal OD)All defined catheter dimensions are within the specified tolerances.Pass
    Catheter Torque StrengthWith the catheter tip held in a static position, there shall be no separation of any portion of the catheter when rotated at the hub at least two (2) full rotations (720 degrees).Pass
    Coating - Particulate and IntegrityThe amount of particulate matter generated during simulated use testing shall be determined and compared to competitive products and techniques. There were no coating anomalies or significant wear observed post simulated use.Pass
    Coating – Lubricity and DurabilityCoating must be lubricious with a specified average pull force.Pass

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the exact sample sizes used for each bench and laboratory test. It states that "Bench and laboratory (in-vitro) testing was completed" and references "simulated neurological model" and "simulated use testing," indicating an in-vitro experimental setting.

    Data Provenance: The data is generated from in-vitro bench and laboratory testing conducted by the manufacturer, Imperative Care, Inc. The country of origin of this data is not explicitly stated, but it is implied to be internal testing. The data is retrospective in the sense that it evaluates the performance of the manufactured devices against pre-established specifications.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of information is not applicable to this 510(k) submission. "Ground truth" established by experts (e.g., radiologists) is typically relevant for studies validating diagnostic or AI-driven devices against clinical outcomes or expert interpretation, such as in an MRMC study. This submission focuses on the physical and mechanical performance of a medical device (catheter) through bench testing.

    4. Adjudication Method for the Test Set

    Since the evaluation is based on objective measurements and predefined specifications in bench testing (e.g., absence of kinking, passage through a simulated model), there is no adjudication method as typically understood in clinical or expert-reviewed studies. The "Pass/Fail" results are determined by whether the device meets the quantitative and qualitative requirements of the test specifications.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If So, What was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    No, an MRMC comparative effectiveness study was not done. This submission is for a percutaneous catheter, not an AI-driven or diagnostic device that involves human reader interpretation. Consequently, there is no discussion of human reader improvement with or without AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    No, a standalone algorithm performance study was not done. This device is a physical medical instrument (catheter), not a software algorithm.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" in this context refers to the pre-established physical and mechanical performance specifications laid out in the test attributes and derived from standards like ISO 10555-1 and internal engineering requirements. It is a benchmark of expected functional performance, rather than a clinical or pathological diagnosis.

    8. The Sample Size for the Training Set

    This question is not applicable as this is not a machine learning or AI-driven device. There is no concept of a "training set" for physical device performance testing.

    9. How the Ground Truth for the Training Set was Established

    This question is not applicable as this is not a machine learning or AI-driven device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K212224

    Validate with FDA (Live)

    Date Cleared
    2021-09-20

    (66 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TracStar LDP Large Distal Platform is indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature.

    The Zoom 88 Large Distal Platform is indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature.

    Device Description

    The Imperative Care Large Distal Platform (LDP) Catheters include the TracStar™ LDP Large Distal Platform and Zoom™ 88 Large Distal Platform. The LDP Catheters are 0.038" diameter or smaller guidewire compatible single lumen guide catheters that provide access to peripheral, coronary and neuro vasculature. The catheters are comprised of a hollow cylindrical tube bonded at the proximal end to a standard luer fitting. The wall of the tube is constructed using a combination of metal coils/braids and medical grade polymers. The distal section of each catheter has a hydrophilic coating to enhance tracking through tortuous vasculature. An angled distal soft tip facilitates smooth tracking past vessel branches. A radiopaque marker provides visual confirmation of the distal tip location under fluoroscopy. LDP Catheters have an inner diameter of 0.088" (6F compatible), and a maximum outer diameter of 0.110". The LDP guide catheters are packaged with a rotating hemostasis valve (RHV) that is attached to the proximal luer to help maintain hemostasis.

    AI/ML Overview

    This document describes the premarket notification for the Imperative Care TracStar LDP Large Distal Platform and Zoom 88 Large Distal Platform, which are guide catheters. The submission is based on substantial equivalence to predicate devices, meaning extensive new studies were not required. The provided text details the acceptance criteria and study findings for this medical device.

    1. Table of Acceptance Criteria and Reported Device Performance

    Test AttributeAcceptance CriteriaReported Performance
    Delivery, Compatibility, and Retraction (Trackability)The catheter shall be able to be delivered, deployed, and retracted per the IFU within a simulated neurological model without incurring any damage to the catheter.Pass
    Flexibility and Kink ResistanceThere shall be no kinking of shaft (permanent deformation) after simulated use.Pass
    Compatibility with other Devices (external)The catheters shall be able to be delivered through the minimum introducer sheath or guide catheter size indicated in the product labeling.Pass
    Guidewire CompatibilityThe catheters shall be able to be delivered over the maximum size guidewire indicated in the product labeling.Pass
    Interventional Device Compatibility (internal)The catheters shall be able to accommodate other interventional devices (e.g., support catheter, diagnostic catheter) up to the maximum size indicated in the product labeling.Pass
    Luer CompatibilityDevices and accessories shall be compatible with standard syringe luer fittings per ISO 80369-7.Pass
    Accessory CompatibilityDevices shall be compatible with an RHV.Pass
    Catheter Bond StrengthThe catheter shall have sufficient bond strengths to remain intact throughout a procedure.Pass
    Freedom from Leakage – positive pressureNo liquid leakage from the hub or catheter shaft at 46psi for 30 seconds.Pass
    Freedom from leakage - negative pressureNo air leakage into a 20cc syringe when vacuum pulled for 15 seconds.Pass
    Dynamic Burst PressureCatheter does not burst under pressures that could be seen when performing contrast injections with a standard 10cc syringe.Pass
    Static Burst PressureCatheter does not burst under pressures that could be seen when performing contrast injections with a standard 10cc syringe.Pass
    Catheter Torque StrengthWith the catheter tip held in a static position, there shall be no separation of any portion of the catheter when rotated at the hub at least two (2) full rotations (720 degrees).Pass
    Kink ResistanceThere shall be no kinking of the catheter shaft (permanent deformation) after wrapping around anatomically relevant bend radii.Pass
    PushabilityThe proximal shaft of the catheters shall have sufficient stiffness that the user can easily push the catheter to the target anatomy without buckling.Pass
    Access ForceCatheters shall not require excessive force to safely navigate and track to the target anatomy.Pass

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the specific sample sizes (number of units tested) for each individual bench and laboratory test. It generally refers to "bench and laboratory (in-vitro) testing." The data provenance is in-vitro, conducted in a laboratory setting. There is no information regarding country of origin of the data or if it was retrospective or prospective, as these terms typically apply to clinical studies with human participants.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided. For performance testing of a medical device like a catheter, "ground truth" is typically established by physical measurements, engineering specifications, and adherence to recognized standards (like ISO 10555-1), rather than expert consensus on interpretive data.

    4. Adjudication Method for the Test Set

    This information is not applicable and not provided. Adjudication methods (e.g., 2+1, 3+1) are typically used in clinical studies involving interpretation of data (e.g., imaging) by multiple readers to resolve discrepancies. The performance tests described here are objective, pass/fail engineering tests.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The submission is for a medical device (catheter) and relies on bench and laboratory testing for substantial equivalence, not comparative clinical effectiveness with human readers.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    This question is not applicable to the device described. The TracStar LDP Large Distal Platform and Zoom 88 Large Distal Platform are physical guide catheters, not algorithms or AI software. Therefore, there is no "algorithm only" or "human-in-the-loop" performance to report.

    7. The Type of Ground Truth Used

    For the bench and laboratory tests, the "ground truth" is based on pre-defined engineering specifications and performance standards, particularly ISO 10555-1 for catheters. The results are objective measurements against these established criteria, leading to a "Pass" or "Fail" outcome.

    8. The Sample Size for the Training Set

    This information is not applicable. The device is a physical catheter, not an AI/ML algorithm that requires a training set. The substantial equivalence pathway relies on demonstrating that the new device performs similarly to a predicate device through bench testing, not on learning from a dataset.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable, as there is no training set for a physical medical device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1