Search Results
Found 4 results
510(k) Data Aggregation
(505 days)
Tamper Evident Caps (Sterile) are indicated for use as a sterile tamper evident cap for IV syringes.
The International Medical Industries, Inc. Tamper Evident Cap consists of a polystyrene outer sleeve and bottom cover and a polypropylene tip cap held in place by a breakaway ring integrated into the sleeve. Evidence of access to the Luer lock cap is achieved by physically breaking the connection and removing the polystyrene outer sleeve exposing the Luer lock cap. Removal of the Luer Lock Cap is restricted by the presence of the sleeve. The restrictive sleeve and cap are available in multiple colors (cosmetic purposes only) in individual blister packs and trays of 10. The Luer Lock Caps are provided sterile to the end user. The sterilization method used is Ethylene Oxide.
The provided text is a 510(k) summary for a medical device called "Tamper Evident Cap." It details the device's characteristics and its comparison to a predicate device to establish substantial equivalence for FDA clearance.
**However, the document does not describe:
- An AI-powered device.
- Acceptance criteria related to AI performance metrics (e.g., sensitivity, specificity, AUC).
- A study that proves an AI device meets acceptance criteria.**
Instead, the document focuses on the physical and material properties of a non-AI medical device (Tamper Evident Cap) and its compliance with ISO 80369-7:2016 ("Small-bore connectors for liquids and gases in healthcare applications – Part 7 Connectors for intravascular or hypodermic applications") and ISO 80369-20:2015 ("Common test methods"), as well as biocompatibility testing standards.
Therefore, I cannot extract the information required by your prompt regarding AI device acceptance criteria and study details from the provided text. The questions about sample sizes for test and training sets, data provenance, expert panels, adjudication methods, MRMC studies, standalone performance, and ground truth types are not applicable to the non-AI device described in this 510(k) summary.
The only relevant information that can be extracted from the document regarding "acceptance criteria" and "study" is for the physical device:
Acceptance Criteria and Study for the Tamper Evident Cap (Non-AI Device)
1. A table of acceptance criteria and the reported device performance:
| Acceptance Criteria (Stipulated in Design Verification Protocol) | Reported Device Performance |
|---|---|
| Meet ISO 80369-7:2016 requirements | TEC passes ISO 80369 testing and associated standards |
| Meet ISO 80369-20:2015 requirements | TEC passes ISO 80369 testing and associated standards |
| Biocompatibility: | |
| - ASTM Hemolysis Study Extract Method | Passed |
| - Cytotoxicity Study Using the ISO Elution Method | Passed |
| - Infrared Spectroscopy | Data supports substantial equivalence |
| - ISO Acute Systemic Toxicity Study | Passed |
| - ISO Guinea Pig Maximization Sensitization Test | Passed |
| - ISO Intracutaneous Irritation Study - Extract | Passed |
| - Microscopic Particle Count Test (USP 788) | Passed |
| - Particulate Analysis - Device - Light Obscuration Method (USP 788) | Passed |
| - USP Rabbit Pyrogen Study Material Mediated | Passed |
2. Sample sized used for the test set and the data provenance:
- Sample Size: Not explicitly stated in the document, but standard medical device testing involves a sufficient number of samples to ensure statistical significance for the specific tests (e.g., multiple samples for each biocompatibility test, mechanical tests as per ISO standards).
- Data Provenance: The tests were conducted as part of the Design Control process, presumably by the manufacturer (International Medical Industries, Inc.) or contracted labs. The document does not specify country of origin for the data or if it was retrospective/prospective, but implies prospective testing for regulatory submission.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
- Not applicable as this is a physical device being tested against established international standards (ISO, ASTM, USP). "Ground truth" in this context refers to the defined parameters and passing criteria of these standards, not expert interpretations of images or clinical outcomes.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not applicable. Testing is against objective, predefined criteria in established international standards.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This is a physical medical device, not an AI-powered diagnostic or assistive tool for human readers.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not applicable. This is a physical medical device, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
- The ground truth is based on the objective passing criteria defined by the relevant international standards (ISO 80369-7, ISO 80369-20, various ISO 10993 standards for biocompatibility, ASTM, USP) for mechanical performance and material safety.
8. The sample size for the training set:
- Not applicable. This is a physical device, not an AI model requiring a training set.
9. How the ground truth for the training set was established:
- Not applicable. This is a physical device.
Ask a specific question about this device
(73 days)
Tamper Evident Caps are indicated for use as a sterile tamper evident cap for IV syringes.
The Tamper Evident Cap (TEC) is a sterile, single use device which will be used to cover male luer ports on medical devices and provide evidence of access. TEC is offered in multiple colors and all the components used in the device are comprised of polystyrene and polypropylene resin. The TEC device is marketed as a stand-alone device and packaged in bulk.
This document is a 510(k) summary for the Prep-Lock™ Tamper Evident Cap for IV Syringes (K193192). It focuses on demonstrating substantial equivalence to a predicate device (K861276), primarily due to a change in packaging configuration.
Here's an analysis of the provided text in relation to your request:
1. A table of acceptance criteria and the reported device performance:
The document lists various non-clinical bench tests performed to assess the new packaging configuration of the Tamper Evident Cap. It states that "All pre-determined acceptance criteria were met." However, it does not explicitly define the specific acceptance criteria for each test in a quantitative manner, nor does it provide detailed numerical results for the reported device performance. It only states that the device "conforms to the same functional and sterility specifications as the predicate device."
| Test Name | Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|---|
| Sterility Assurance 10-6 SAL | Sterilization per ANSI/AAMI/ISO 11737-2:2009/(R)2014 achieved 10-6 SAL. | Meets established sterility assurance level. |
| Sterilization Product Adoption | Conformance to AAMI TIR 28 for ethylene oxide sterilization. | Conforms to AAMI TIR 28. |
| Shelf Life | Conformance to ASTM F1980-16 for accelerated aging. | Maintains 3-year shelf life. |
| EO/ECH Residuals | Conformance to ISO 10993-1 and relevant guidance for residuals. | Meets requirements for EO/ECH residuals. |
| Device Function (HE75, ISO 80369-7, IMI Internal Tests) | Design per AAMI ANSI HE75, connectors per ISO 80369-7, and satisfactory internal tests for tamper evident break force, simulated use, and weld strength. | Meets functional specifications. |
| Packaging Integrity (ASTM F 1886, ASTM F 88) | Integrity of seals per ASTM F 1886, minimum peel strength per ASTM F 88. | Packaging maintains integrity. |
| Transportation (ISTA 3B, ASTM F1886, F1929) | Packaging withstands ISTA 3B test, visual seal inspection per ASTM F1886, and dye penetration per F1929. | Withstands transportation challenges. |
| Product Validation (Human factors) | Usability assessment of packaging use. | Packaging deemed usable. |
| Biocompatibility Testing | Conformance to ISO 10993-1 and FDA guidance for indirect blood contact, limited duration. | Device is biocompatible. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document does not specify the sample sizes used for any of the described non-clinical bench tests. It also does not explicitly state the country of origin of the data or whether the tests were retrospective or prospective. Given the nature of bench testing for a device modification (packaging), these would typically be conducted prospectively by the manufacturer (International Medical Industries, Inc., Pompano Beach, Florida, USA).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This is not applicable to the provided document. The study described is a non-clinical bench testing for a physical medical device (tamper evident cap) and its packaging, not an AI/diagnostic device that would require expert-driven ground truth establishment.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable, as this is non-clinical bench testing. Human adjudication methods like 2+1 are typically used for clinical studies involving interpretation of medical images or patient outcomes.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is not an AI-powered diagnostic tool, and therefore, no MRMC comparative effectiveness study was conducted.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This device is a physical medical device, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
For the non-clinical bench tests, the "ground truth" is implied by the standards and internal specifications that the device and its packaging must meet (e.g., ANSI/AAMI/ISO 11737-2 for sterility, ASTM F1980-16 for shelf life, ISO 10993-1 for biocompatibility). The tests directly measure whether the physical properties and performance characteristics of the device and its packaging conform to these established benchmarks.
8. The sample size for the training set
Not applicable. This document describes the testing of a physical medical device, not an AI or machine learning model that requires a training set.
9. How the ground truth for the training set was established
Not applicable, as no training set was used.
Ask a specific question about this device
(116 days)
The Tamper Evident Cap with Male Luer Lock is indicated for use as a cap for female Luer ports on medical devices such as manifolds, stopcocks or set.
The Tamper Evident Cap with Male Luer Lock (MTEC) is a sterile, single use device which will be used to cover female Luer ports on medical devices and provide evidence of access. MTEC is provided in two packaging configurations; a 10-up tray and single unit blisters. Both configurations are offered in multiple colors and all the components used in the manufacture of MTEC are comprised of polystyrene and polypropylene resin.
Here's a breakdown of the acceptance criteria and study information based on the provided text, while noting limitations in detail for device performance and several other categories:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| ISO 80369-7:2016 (Small-bore connectors for liquids and gases in healthcare applications -- Part 7: Connectors for intravascular or hypodermic applications) | The MTEC device meets the bench testing requirements of ISO 80369-7:2016. Specifically, it meets requirements for: - Separation Force - Unscrewing torque - Ease of assembly - Stress cracking - Liquid leakage test - Resistance to overriding |
| Shelf Life Evaluation | Design verification of the shelf life was conducted and implied to meet criteria. The document states "All test results meet the acceptance criteria." |
| Sterilization Product Adoption (AAMI TIR 28:2009/(R)2013) | Implied to meet criteria: "All test results meet the acceptance criteria." |
| Packaging (ISO 11607-1:2006) | Implied to meet criteria: "All test results meet the acceptance criteria." |
| Sterility (ISO 11135:2014) | Implied to meet criteria: "All test results meet the acceptance criteria." |
| Residuals (ISO 10993-7:2008) | Implied to meet criteria: "All test results meet the acceptance criteria." |
| Biocompatibility (ISO 10993-1, relevant FDA Guidance) | Biocompatibility testing was conducted according to the cited guidance and standards for an External Communicating Device, Blood Path, Indirect Contact, Limited Duration, and implied to meet criteria: "All test results meet the acceptance criteria." |
| Tamper Evident Design Feature | The MTEC device's tamper evident design feature was tested and successfully passed all internal specifications. |
2. Sample Size Used for the Test Set and Data Provenance
The document does not specify the sample size for the test set used for any of the bench tests (e.g., ISO 80369-7:2016, tamper evident feature testing).
The data provenance is from non-clinical bench testing performed by International Medical Industries, Inc. (the manufacturer). The country of origin for the data is not explicitly stated but implies testing was done internally by the manufacturer. The data is prospective in the sense that it was generated specifically for this submission.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This information is not applicable as the studies are non-clinical bench tests comparing device performance to established international standards and internal specifications, not studies involving human interpretation or ground truth establishment by experts in a clinical context.
4. Adjudication Method for the Test Set
This information is not applicable as the studies are non-clinical bench tests against objective criteria, not studies requiring expert adjudication of results.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No, an MRMC comparative effectiveness study was not done. This submission is for a medical device (tamper evident cap) and relies on non-clinical bench testing for substantial equivalence, not clinical effectiveness studies with human readers.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
This question is not applicable as the device is a physical medical component, not an algorithm or AI system.
7. The Type of Ground Truth Used
The "ground truth" for the non-clinical bench tests is established by:
- International Standards: e.g., ISO 80369-7:2016, ISO 11607-1:2006, ISO 11135:2014, ISO 10993-7:2008.
- Internal Specifications: For the tamper-evident design feature.
- FDA Recognized Guidance Documents: For biocompatibility testing (e.g., ISO 10993-1 and specific FDA guidance for intravascular administration sets).
These are objective, quantitative criteria rather than expert consensus, pathology, or outcomes data.
8. The Sample Size for the Training Set
This question is not applicable as the device is a physical medical component, not an AI/ML algorithm that requires a training set.
9. How the Ground Truth for the Training Set Was Established
This question is not applicable as the device is a physical medical component, not an AI/ML algorithm.
Ask a specific question about this device
(214 days)
Tamper Evident Cap For use with ENFit® Syringes is indicated for use to prevent fluid loss and contamination of syringe contents until ready for use with evidence of access.
Tamper Evident Cap For use with ENFit® Syringes is a sterile, single use disposable device for use as a closure cap with evidence of access for Enteral Syringes with ENFit™ Connectors. Removal of the cap is achieved by removing a restrictive sleeve covering the ENFit™ Lock Cap. A damaged or missing restrictive sleeve indicates potential tampering. A non-sterile version of the device is also offered.
The provided text does not contain information about the acceptance criteria and the study that proves the device meets the acceptance criteria in the format requested. The document is a 510(k) summary for a medical device (Tamper Evident Cap For use with ENFit® Syringes) and focuses on demonstrating substantial equivalence to a predicate device.
Specifically, the document lists various performance bench tests that were conducted (Liquid Leakage, Stress Cracking, Resistance to Separation from Axial Load, etc., as per AAMI/ISO 80369-3, and additional internal specifications like Visual Indication of Tamper Evidence). However, it does not provide specific acceptance criteria values or the reported performance data in a tabular format as requested. It only states that the device "meets the bench testing requirements" and "passed all biocompatibility tests."
Therefore, I cannot populate the requested table or provide details on sample size, data provenance, expert qualifications, adjudication methods, MRMC studies, standalone performance, or training set information.
Ask a specific question about this device
Page 1 of 1