510(k) Data Aggregation

    K Number
    K251024

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-04-30

    (28 days)

    Product Code
    Regulation Number
    864.5425
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TEG 6s Hemostasis System is intended for in vitro diagnostic use to provide semi-quantitative indications of the hemostasis state of a 3.2% citrated whole blood sample. The Citrated: K, KH, RT, FF Assay Cartridge, to be used with the TEG 6s analyzer, contains four independent assays (CK, CKH, CRT, and CFF), described below.

    The CK assay monitors the hemostasis process via the intrinsic pathway in 3.2% citrated whole blood specimens on the TEG 6s System. Clotting characteristics are described by the functional parameters Clotting Time (R), Speed of Clot Formation (K and Alpha angle) and Maximum Clot Strength (MA).

    The CKH assay monitors the effects of heparin in 3.2% citrated whole blood specimens on the TEG 6s System. CKH is used in conjunction with CK, and heparin influence is determined by comparing Clotting Times (R) between the two tests.

    The CRT assay monitors the hemostasis process via both the intrinsic and extrinsic pathways in 3.2% citrated whole blood specimens on the TEG 6s System. Clotting characteristics are described by the functional parameter Maximum Clot Strength (MA). The CRT MA parameter is equivalent to the CK MA parameter but the final MA value is reached more quickly using the CRT assay.

    The CFF assay monitors hemostasis of 3.2% citrated whole blood specimens in the TEG 6s System after blocking platelet contributions to clot strength. Clotting characteristics are described by the functional parameters Maximum Clot Strength (MA) and the Estimated Functional Fibrinogen Level (FLEV).

    Results from the TEG 6s analysis should not be the sole basis for a patient diagnosis, but should be evaluated together with the patient's medical history, the clinical picture and, if necessary, further hemostasis tests. The indication for TEG 6s System use is with adult patients where an evaluation of their blood hemostasis properties is desired. Hemostasis evaluations are commonly used to assess clinical conditions in cardiovascular surgery and cardiology procedures to assess hemorrhage or thrombosis conditions before, during and following the procedure. The TEG 6s Hemostasis System can be used in the laboratory or at the point-of-care.

    Device Description

    The TEG® 6s Hemostasis System (TEG® Hemostasis analyzer and TEG® 6s Assay Cartridges) is intended for in vitro diagnostic use to provide semi-quantitative indications of a blood sample's ability to form and maintain a clot. The TEG® 6s Hemostasis System records the kinetic changes in a sample of whole blood as the sample clots, retracts and/or lyses. The system output consists of a table of numerical values and graphs resulting from the hemostasis process over time. This information can be used by clinicians to aid in determining if a clotting dysfunction or coagulopathy is present.

    To perform a test, a disposable TEG® 6s Assay Cartridge is inserted into the TEG® 6s Hemostasis analyzer. The instrument reads the bar code on the cartridge and identifies the type of cartridge for operator confirmation. Blood (collected in a 3.2% sodium citrate tube) or Quality Control (QC) material is added to the entry port on the cartridge and drawn into the cartridge under the TEG® 6s Hemostasis analyzer control. The amount of the sample drawn into the cartridge is determined by the pre-set volume of the blood chambers in the cartridge. Once in the cartridge, the sample is metered into as many as 4 separate analysis channels, depending upon the assays being performed. Reconstitution of reagents dried within the cartridge is accomplished by moving the sample back and forth through reagent chambers, under the control of microfluidic valves and bellows (pumps) within the cartridge. After each sample has been mixed with reagent, it is delivered to a test cell where it is monitored for viscoelastic changes due to coagulation. Excess sample material is moved under microfluidic control into an enclosed waste chamber within the cartridge.

    The TEG® 6s technology is based on a disposable cartridge containing up to 4 independent measurement cells. Each cell consists of a short vertically-oriented injection molded tube (ring). Detection of clotting in the TEG® 6s Hemostasis System is performed optically. A piezoelectric actuator vibrates the measurement cell(s) through a motion profile composed of summed sinusoids at different frequencies. The movement of the measurement cells will induce motion in the sample meniscus, which will be detected by a photodiode. The resulting motion of the meniscus is monitored optically and analyzed by the instrument to calculate the resonant frequency and modulus of elasticity (stiffness) of the sample. By performing a Fast Fourier Transform (FFT) on meniscus motion data, the resonant frequencies can be determined. The analyzer monitors the harmonic motion of a hanging drop of blood in response to external vibration. As the sample transitions from a liquid state to a gel-like state during clotting, the modulus of elasticity (stiffness) and therefore resonant frequency increase. The TEG® 6s Hemostasis analyzer measures these variations in resonant frequency during clotting and lysis.

    Resonance is the tendency of a material or structure to oscillate with greater amplitude at some frequencies than others. The exact frequencies at which resonance occurs will depend on the stiffness and mass of the sample. Stiffness, in turn, is a function of a material's modulus of elasticity and the boundary conditions to which the material is exposed, such as the geometry and materials of a test cell. By holding these boundary conditions and sample mass constant from sample to sample, the TEG® 6s Hemostasis System allows direct comparison of elasticity between samples. The output measurements are displayed in a table and on a graphical tracing that reflects the hemostasis profile of the clot formation.

    In a typical test, blood that has been delivered to the measurement cell will not clot for several minutes. During this time the sample has no inherent stiffness except that provided by surface tension, and since this remains constant the measured resonant frequencies will not change. Once clotting begins, however, the elastic modulus and thus the resonant frequencies increase rapidly. During fibrinolysis, the process is reversed, with elastic modulus and resonant frequencies decreasing. In tests where clotting does not occur, the resonant frequency of the sample will not change. During coagulation, however, a clot will bind to the ring contained in the cartridge and the resonant frequency will rise with increasing firmness of the clot. The TEG® 6s Hemostasis Analyzer collects meniscus motion data, tracks changing resonant frequencies and analyzes the frequency data to provide semi-quantitative parameters describing the clot.

    The TEG® 6s Hemostasis System monitors the interaction of platelets within the fibrin mesh of the clot during clot formation and lysis, all in a whole-blood setting. The TEG® 6s Hemostasis System uses thromboelastography to provide continuous measurement of clot elasticity.

    AI/ML Overview

    The provided document is a 510(k) clearance letter from the FDA for the TEG 6s Hemostasis System. This type of submission focuses on demonstrating substantial equivalence to a predicate device, rather than providing the detailed acceptance criteria and study data often found in novel device approvals or publications on algorithmic performance.

    Therefore, the document does not contain the level of detail requested for many of the points, especially those related to AI algorithm performance (e.g., sample sizes for test/training sets, expert qualifications, adjudication methods, MRMC studies, standalone performance, ground truth establishment for training). The device described is an in vitro diagnostic (IVD) system that measures blood coagulation properties, suggesting a direct measurement rather than an AI-driven interpretation of complex data like images.

    Below is an attempt to extract and infer information based on the provided text, while also clearly indicating what information is not available in this document.


    Acceptance Criteria and Device Performance Study

    The document primarily focuses on demonstrating substantial equivalence to a predicate device rather than providing specific acceptance criteria for a novel device or an AI algorithm. The core of the study described is a comparison of the modified device's performance against its predicate and within different sample matrices.

    1. Table of Acceptance Criteria and Reported Device Performance

    Parameter CategoryAcceptance Criteria (Inferred from "met clinically acceptable bias limits")Reported Device Performance
    Sample Matrix EquivalenceNo statistical difference in agreement between TEG 6s and TEG 5000 by blood source.Demonstrated.
    Predicate Bias (Comparison between new and predicate cartridge types)Bias met clinically acceptable bias limits across all parameters for TEG 6s Citrated: K, KH, RT, FF vs. TEG 6s Citrated: K, KH, RTH, FFH.Demonstrated.

    Note: Specific numerical acceptance criteria (e.g., ±X% bias) are not provided in this document. The document states that the bias "met clinically acceptable bias limits," but the limits themselves are not specified.

    2. Sample Sizes and Data Provenance for the Test Set

    • Sample Size for Test Set: Not specified in the document.
    • Data Provenance: The document mentions "an observational study" for the agreement analysis between TEG 6s and TEG 5000 by blood source. It does not specify the country of origin of the data or whether it was retrospective or prospective.

    3. Number of Experts and Qualifications for Ground Truth

    This information is not applicable/not provided for this type of IVD device clearance. The device measures physical properties of blood, and the "ground truth" for its performance would typically involve comparison to established laboratory methods or accepted clinical standards rather than expert consensus on subjective interpretations.

    4. Adjudication Method for the Test Set

    This information is not applicable/not provided. Adjudication methods are typically used in studies involving subjective interpretations (e.g., image reading) where multiple experts might disagree. This device measures quantifiable physiological parameters.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    Not performed/not applicable. An MRMC study is relevant to subjective interpretation tasks, often in medical imaging, where human readers interact with AI. This device directly measures blood coagulation parameters, so there is no "human reader" in the AI sense, nor is there an "AI assistance" component described.

    6. Standalone Performance Study (Algorithm Only)

    Not performed/not applicable. The TEG 6s Hemostasis System, as described, is a physical instrument that performs direct measurements. It is not an "algorithm only" device or an AI system in the typical sense that would have a standalone performance study independent of human interaction. The "algorithm" here refers to the internal processing of sensor data to derive hemostasis parameters. Its performance is intrinsically linked to the physical measurement system.

    7. Type of Ground Truth Used

    The ground truth used for this device would be the quantitative measurements of hemostatic parameters from established, validated laboratory methods or from the predicate device itself. The document mentions:

    • Agreement analysis with TEG 5000 (a prior generation or related device).
    • Comparison with predicate TEG 6s Citrated: K, KH, RT, FF (K243858).

    8. Sample Size for the Training Set

    The concept of a "training set" in the context of machine learning is not applicable to this device as described in the provided document. The device performs direct physical measurements rather than learning from a dataset via an AI algorithm. Therefore, no training set size is reported.

    9. How Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" or AI algorithm in the described context.

    Ask a Question

    Ask a specific question about this device

    K Number
    K243858

    Validate with FDA (Live)

    Date Cleared
    2025-01-15

    (30 days)

    Product Code
    Regulation Number
    864.5425
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TEG 6s Hemostasis System is intended for in vitro diagnostic use to provide semi-quantitative indications of the hemostasis state of a venous blood sample. The Citrated: K, KH, RT, FF Assay Cartridge, to be used with the TEG 6s analyzer, contains four independent assays (CK, CKH, CRT, and CFF), described below.

    The CK assay monitors the hemostasis process via the intrinsic pathway in 3.2% citrated whole blood specimens on the TEG 6s System. Clotting characteristics are described by the functional parameters Clotting Time (R), Speed of Clot Formation (K and Alpha angle) and Maximum Clot Strength (MA).

    The CKH assay monitors the effects of heparin in 3.2% citrated whole blood specimens on the TEG 6s System. CKH is used in conjunction with CK, and heparin influence is determined by comparing Clotting Times (R) between the two tests.

    The CRT assay monitors the hemostasis process via both the intrinsic and extrinsic pathways in 3.2% citrated whole blood specimens on the TEG 6s System. Clotting characteristics are described by the functional parameter Maximum Clot Strength (MA). The CRT MA parameter is equivalent to the CK MA parameter but the final MA value is reached more quickly using the CRT assay.

    The CFF assay monitors hemostasis of 3.2% citrated whole blood specimens in the TEG 6s System after blocking platelet contributions to clot strength. Clotting characteristics are described by the functional parameters Maximum Clot Strength (MA) and the Estimated Functional Fibrinogen Level (FLEV).

    Results from the TEG 6s analysis should not be the sole basis for a patient diagnosis, but should be evaluated together with the patient's medical history, the clinical picture and, if necessary, further hemostasis tests. The indication for TEG 6s System use is with adult patients where an evaluation of their blood hemostasis properties is desired. Hemostasis evaluations are commonly used to assess clinical conditions in cardiovascular surgery and cardiology procedures to assess hemorrhage or thrombosis conditions before, during and following the procedure. The TEG 6s Hemostasis System can be used in the laboratory or at the point-of-care.

    Device Description

    The TEG® 6s Hemostasis System (TEG® Hemostasis analyzer and TEG® 6s Assay Cartridges) is intended for in vitro diagnostic use to provide semi-quantitative indications of a blood sample's ability to form and maintain a clot. The TEG® 6s Hemostasis System records the kinetic changes in a sample of whole blood as the sample clots, retracts and/or lyses. The system output consists of a table of numerical values and graphs resulting from the hemostasis process over time. This information can be used by clinicians to aid in determining if a clotting dysfunction or coagulopathy is present.

    To perform a test, a disposable TEG® 6s Assay Cartridge is inserted into the TEG® 6s Hemostasis analyzer. The instrument reads the bar code on the cartridge and identifies the type of cartridge for operator confirmation. Blood (collected in a 3.2% sodium citrate tube) or Quality Control (QC) material is added to the entry port on the cartridge and drawn into the cartridge under the TEG® 6s Hemostasis analyzer control. The amount of the sample drawn into the cartridge is determined by the pre-set volume of the blood chambers in the cartridge. Once in the cartridge, the sample is metered into as many as 4 separate analysis channels, depending upon the assays being performed. Reconstitution of reagents dried within the cartridge is accomplished by moving the sample back and forth through reagent chambers, under the control of microfluidic valves and bellows (pumps) within the cartridge. After each sample has been mixed with reagent, it is delivered to a test cell where it is monitored for viscoelastic changes due to coagulation. Excess sample material is moved under microfluidic control into an enclosed waste chamber within the cartridge.

    The TEG® 6s technology is based on a disposable cartridge containing up to 4 independent measurement cells. Each cell consists of a short vertically-oriented injection molded tube (ring). Detection of clotting in the TEG® 6s Hemostasis System is performed optically. A piezoelectric actuator vibrates the measurement cell(s) through a motion profile composed of summed sinusoids at different frequencies. The movement of the measurement cells will induce motion in the sample meniscus, which will be detected by a photodiode. The resulting motion of the meniscus is monitored optically and analyzed by the instrument to calculate the resonant frequency and modulus of elasticity (stiffness) of the sample. By performing a Fast Fourier Transform (FFT) on meniscus motion data, the resonant frequencies can be determined. The analyzer monitors the harmonic motion of a hanging drop of blood in response to external vibration. As the sample transitions from a liquid state to a gel-like state during clotting, the modulus of elasticity (stiffness) and therefore resonant frequency increase. The TEG® 6s Hemostasis analyzer measures these variations in resonant frequency during clotting and lysis.

    Resonance is the tendency of a material or structure to oscillate with greater amplitude at some frequencies than others. The exact frequencies at which resonance occurs will depend on the stiffness and mass of the sample. Stiffness, in turn, is a function of a material's modulus of elasticity and the boundary conditions to which the material is exposed, such as the geometry and materials of a test cell. By holding these boundary conditions and sample mass constant from sample to sample, the TEG® 6s Hemostasis System allows direct comparison of elasticity between samples. The output measurements are displayed in a table and on a graphical tracing that reflects the hemostasis profile of the clot formation.

    In a typical test, blood that has been delivered to the measurement cell will not clot for several minutes. During this time the sample has no inherent stiffness except that provided by surface tension, and since this remains constant the measured resonant frequencies will not change.

    Once clotting begins, however, the elastic modulus and thus the resonant frequencies increase rapidly. During fibrinolysis, the process is reversed, with elastic modulus and resonant frequencies decreasing. In tests where clotting does not occur, the resonant frequency of the sample will not change. During coagulation, however, a clot will bind to the ring contained in the cartridge and the resonant frequency will rise with increasing firmness of the Clot. The TEC® 6s Hemostasis Analyzer collects meniscus motion data, tracks changing resonant frequencies and analyzes the frequency data to provide semi-quantitative parameters describing the clot.

    The TEG® 6s Hemostasis System monitors the interaction of platelets within the fibrin mesh of the clot during clot formation and lysis, all in a whole-blood setting. The TEG® 6s Hemostasis System uses thromboelastography to provide continuous measurement of clot elasticity.

    AI/ML Overview

    The provided document is a 510(k) Summary for the Haemonetics TEG® 6s Hemostasis System, specifically for the Citrated: K, KH, RT, FF Assay Cartridge. Its primary purpose is to demonstrate substantial equivalence to a predicate device (K150041). As such, it focuses on comparing the proposed device to the predicate rather than detailing a study that establishes novel acceptance criteria or proves performance against new, distinct benchmarks requiring extensive clinical trials with human-in-the-loop or standalone AI performance.

    The document states: "There is no change to the technology, design of the device, reported parameters, or mechanics of how the cartridge is run on the TEG® 6s analyzer. No additional product development of the TEG® 6s system was required." The only change mentioned is an expansion of the "Use Location" from "clinical laboratory" to "clinical laboratory or at the point-of-care."

    Therefore, the document does not contain the detailed information necessary to answer the prompt's request for:

    • A table of acceptance criteria and reported device performance for a new device or significant modification requiring such a study. Instead, it relies on the predicate's established performance.
    • Sample sizes, data provenance, expert ground truth establishment, adjudication methods, MRMC studies, standalone AI performance, or training set details as these are typically required for demonstrating efficacy or superiority of a new or substantially modified device, especially AI/ML-driven ones.

    The document is a submission for substantial equivalence for a minor modification (expanded use location) of an already cleared device, not a submission for a novel device or a device with a new AI/ML component that requires extensive performance validation against a defined ground truth.

    Therefore, based solely on the provided text, I cannot extract the information required by your prompt, as the study described is a demonstration of substantial equivalence via comparison to a predicate, not an independent performance study of a novel device against predefined acceptance criteria for its core functionality.

    The "study" described is a regulatory comparison. The "acceptance criteria" are effectively that the modified device performs equivalently to the predicate device, especially in the expanded use environment. No specific performance metrics or detailed study results are presented because the core technology and measured parameters are unchanged from the predicate.

    Ask a Question

    Ask a specific question about this device

    K Number
    K183160

    Validate with FDA (Live)

    Date Cleared
    2019-05-09

    (175 days)

    Product Code
    Regulation Number
    864.5425
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TEG® 6s Hemostasis System consists of the TEG 6s Hemostasis Analyzer and TEG 6s Citrated: K, RT, FF Assay Cartridge. The TEG 6s Hemostasis System is intended for in vitro diagnostic use to provide semi-quantitative indications of the hemostasis state of a venous blood sample. The TEG 6s Hemostasis System records the kinetic changes in a sample of 3.2% citrated whole blood as the sample clots.

    The Citrated: K, RT, FF Assay Cartridge contains three independent assays (CK, CRT and CFF) and the system output consists of a table of numerical values for parameters R, LY30, and MA.

    The CK assay monitors the hemostasis process via the intrinsic pathway in 3.2% citrated whole blood specimens on the TEG 6s Hemostasis System. Clotting characteristics are described by the functional parameters R (clotting time) and L Y 30 (fibrinolysis after 30 minutes of reaching maximum clot strength).

    The CRT assay monitors the hemostasis process via both the intrinsic pathways in 3.2% citrated whole blood specimens on the TEG 6s Hemostasis System. Clotting characteristics are described by the functional parameter MA (maximum clot strength).

    The CFF assay monitors hemostasis of 3.2% citrated whole blood specimens in the TEG 6s Hemostasis System after blocking platelet contributions to clot strength. Clotting characteristics are described by the functional parameter MA (maximum clot strength).

    Results from the TEG 6s analysis should not be the sole basis for a patient diagnosis, but should be evaluated together with the patient's medical history, the clinical picture and, if necessary, further hemostasis tests. The indication for TEG 65 Hemostasis System use is with adult patients (18 years and older) where an evaluation of their blood hemostasis properties is desired. Hemostasis evaluation with the TEG 6s Hemostasis System using the Citrated: K, RT, FF Assay Cartridge is used to assess clinical conditions in a trauma setting to assess hemorrhage or thrombosis conditions.

    For professional use only.

    Device Description

    The TEG® 6s Hemostasis System (TEG® Hemostasis analyzer and TEG® 6s Assay Cartridges) is intended for in vitro diagnostic use to provide semi-quantitations of a blood sample's ability to form and maintain a clot. The TEG® 6s Hemostasis System records the kinetic changes in a sample of whole blood as the sample clots, retracts and/or lyses. The system output consists of a table of numerical values resulting from the hemostasis process over time. This information can be used by clinicians to aid in determining if a dysfunction or coagulopathy is present.

    To perform a test, a disposable TEG® 6s Assay Cartridge is inserted into the TEG® 6s Hemostasis analyzer. The instrument reads the bar code on the cartridge and identifies the type of cartridge for operator confirmation. Blood (collected in a 3.2% sodium citrate tube) or Quality Control (QC) material is added to the entry port on the cartridge and drawn into the cartridge under the TEG® 6s Hemostasis analyzer control. The amount of the sample drawn into the cartridge is determined by the pre-set volume of the blood chambers in the cartridge. Once in the cartridge, the sample is metered into as many as 4 separate analysis channels, depending upon the assays being performed. Reconstitution of reagents dried within the cartridge is accomplished by moving the sample back and forth through reagent chambers, under the control of microfluidic valves and bellows (pumps) within the cartridge. After each sample has been mixed with reagent, it is delivered to a test cell where it is monitored for viscoelastic changes due to coagulation. Excess sample material is moved under microfluidic control into an enclosed waste chamber within the cartridge.

    The TEG® 6s technology is based on a disposable cartridge containing up to 4 independent measurement cells. Each cell consists of a short vertically-oriented injection molded tube (ring). Detection of clotting in the TEG® 6s Hemostasis System is performed optically. A piezoelectric actuator vibrates the measurement cell(s) through a motion profile composed of summed sinusoids at different frequencies. The movement of the measurement cells will induce motion in the sample meniscus, which will be detected by a photodiode. The resulting motion of the meniscus is monitored optically and analyzed by the instrument to calculate the resonant frequency and modulus of elasticity (stiffness) of the sample. By performing a Fast Fourier Transform (FFT) on meniscus motion data, the resonant frequencies can be determined. The analyzer monitors the harmonic motion of a hanging drop of blood in response to external vibration. As the sample transitions from a liquid state to a gel-like state during clotting, the modulus of elasticity (stiffness) and therefore resonant frequency increase. The TEG® 6s Hemostasis analyzer measures these variations in resonant frequency during clotting and lysis.

    AI/ML Overview

    This document describes the TEG® 6s Hemostasis System, an in vitro diagnostic device used to provide semi-quantitative indications of the hemostasis state of a venous blood sample.

    Here’s an analysis of the acceptance criteria and the study proving the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't explicitly state "acceptance criteria" for performance in a pass/fail format with exact numerical thresholds for all evaluated aspects. Instead, it details various performance characteristics and states that the data "supports a substantial equivalence determination." However, based on the performance testing presented, we can infer the implicit acceptance criteria by observing the reported results and the conclusions drawn.

    Acceptance Criteria (Inferred from Performance Testing)Reported Device Performance
    Electrical Safety & EMC Compliance: Meets relevant IEC and UL standards.The system complies with IEC 60601-2-10, IEC 60601-2-10, IEC 60601-2-101 and UL 61010-1 for safety, and IEC 60601-1-2, EN61326-1, EN61326-2-6, EN61000-3-2, EN61000-3-3 and EN55011 for EMC.
    Software Verification & Validation: Adheres to FDA guidance for "moderate" level of concern software.Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance. The software was considered "moderate" level of concern.
    Reference Ranges: Established according to CLSI C28-A3c.Reference ranges were established using citrated whole blood from normal donors. Non-parametric analysis determined the following ranges: - CK R (minutes): 4.6 - 9.1 (n=157) - CK LY30 (percent): 0.0 - 2.6 (n=132) - CRT MA (mm): 52 - 70 (n=152) - CFF MA (mm): 15 - 32 (n=151)
    Analytical Precision: Demonstrated across different coagulation levels (hypo, normal, hyper) and sample types (natural, contrived, patient-derived), using CLSI EP5-A2 guidance with acceptable Standard Deviations (SD) and Coefficients of Variation (%CV).Three precision studies were conducted. Results (SD and %CV) were reported across reagent lots, operators, instruments, and days for various parameters and sample types (Hypo, Normal, Hyper donors; Contrived samples with Dabigatran/Cytochalasin D/ReoPro, Kaolin/RiaStap; Patient-derived samples with Dabigatran/Warfarin; and low/high/no tPA for LY30). The values are presented in Tables 5, 6, 7, and 8, indicating the device maintains precision across these conditions. (Specific numeric thresholds for "acceptable" are not given, but the presentation implies the results are within acceptable industry limits for such a device).
    Interference: Identifies interfering factors and limits for CRT and CFF assays.For CK assay, no interfering factors were found among: Absence of a Discard Tube, Short Draw, Hemolysis, Hemodilution, Direct Oral Anticoagulants (FXa and direct thrombin inhibitors), Antiplatelet Drug (P2Y12 inhibitor). For CRT assay, Hemolysis and Hemodilution above 30% were found to be interfering factors. For CFF assay, Hemodilution above 40% was found to be an interfering factor.
    Method Comparison (Equivalency to Predicate Device): Slopes of linear regression lines close to 1.0 (with 95% CI containing 1.0), acceptable predictive bias relative to acceptance criteria, and high Pearson linear correlation (>0.9 for identical parameters, 0.86 for CRT MA vs CK MA).The linear regression slope estimates for all between-device comparisons were close to 1.0, with their respective 95% confidence intervals all containing 1.0 (range 0.99 to 1.06). The assessment of predictive bias and its 95% confidence interval relative to the bias acceptance criteria supports equivalency according to CLSI EP09-A3. Predicted biases at AMR limits were consistent with reference range limits. Pearson linear correlation estimates were above 0.9 for all identical parameters. The correlation for CRT MA (TEG® 6s) and CK MA (TEG® 5000) was 0.86. Conclusion: The method comparison data strongly supports the correlation between TEG® 6s and the TEG® 5000.

    2. Sample Size Used for the Test Set and Data Provenance

    • Reference Ranges:

      • CK R: 157 samples
      • CK LY30: 132 samples
      • CRT MA: 152 samples
      • CFF MA: 151 samples
      • Provenance: Citrated whole blood from normal donors, representative of normal population distributions (age, gender, race), with no known coagulopathies and not taking any drugs that would potentially affect patient hemostasis. (Implies prospective collection for range establishment).
    • Analytical Precision (First Study):

      • Sample Size: Blood from 4 donors (2 Hypo, 1 Normal, 1 Hyper). Run in duplicate for 5 non-consecutive days, across 3 reagent lots, by 2 operators, using 12 analyzers. Each data point in Table 5 represents n=120 replicates.
      • Provenance: Donors with natural coagulation levels (Hypo, Normal, Hyper). (Implies prospective collection from specific donor profiles).
    • Analytical Precision (Second Study - Contrived/Patient-Derived):

      • Sample Size:
        • 3 contrived samples (Hypo, Hyper types)
        • 4 patient-derived dabigatran samples
        • 2 patient-derived warfarin samples
        • Each run in duplicate for 5 non-consecutive days, by 3 operators, using up to 12 analyzers. Each data point in Tables 6 and 7 represents N=12 replicates per parameter/sample/patient ID.
      • Provenance: Contrived samples (blood from normal donors spiked with drugs/reagents). Patient-derived samples (blood from clinical patients being treated with therapeutic levels of anticoagulants). (Combination of prospective collection for spiking and retrospective/prospective from clinical patients).
    • Analytical Precision (Third Study - CK LY30):

      • Sample Size: 3 sample types (no tPA, low tPA, high tPA). 12 replicates per sample type, by 3 operators, using 3 reagent lots for 5 days. Each data point in Table 8 represents N=58 to N=60 replicates.
      • Provenance: Normal donors (for no tPA) and normal donor blood spiked to create low/high tPA samples. (Implies prospective collection for spiking).
    • Method Comparison:

      • Sample Size:
        • CFF MA: 450 samples
        • CK R: 405 samples
        • CK LY30: 86 samples
        • CRT MA vs. CK MA: 336 samples
      • Provenance: Conducted at 12 US clinical sites, enrolling adult patients (18+ years) who met full or limited trauma team criteria of the American College of Surgeons or similar institutional guidelines. (This indicates prospective, multi-center, real-world clinical data collection).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    For this device, the "ground truth" is typically established by the inherent physical properties of the blood samples (clotting characteristics) and comparison to an established predicate device (TEG® 5000), not by human expert interpretation of results in the traditional sense of image analysis or diagnostic decision-making.

    • Reference Ranges: Established using samples from "normal donors" with "no known coagulopathies." The determination of "normal" is likely based on standard medical criteria, possibly verified by physicians or lab personnel, but not explicitly stated as "expert consensus" in the document.
    • Precision Studies: Ground truth is the measured value of the parameters (R, MA, LY30) from the samples themselves, along with their known characteristics (e.g., "hypo," "hyper," spiked with certain drugs). This is measured intrinsically by the device itself, not by external human experts.
    • Method Comparison: The predicate device, TEG® 5000, serves as the "ground truth" or reference for comparison. The study assesses how closely the TEG® 6s results correlate with the TEG® 5000, rather than comparing to a diagnostic expert's opinion. Clinical experts (physicians) would ultimately interpret both TEG® 5000 and TEG® 6s results to make patient diagnoses, but they are not directly establishing the "ground truth" for the device's functional performance in this context.

    Therefore, the document does not mention human experts establishing ground truth for the test sets in the way AI/ML studies often describe.

    4. Adjudication Method for the Test Set

    Not applicable in the conventional sense. The studies described are performance evaluations of an in vitro diagnostic device, involving quantitative measurements. There is no mention of an adjudication process (e.g., 2+1, 3+1 consensus) because the "ground truth" for the performance tests (precision, method comparison) is based on instrumental measurements and comparison to a predicate, not subjective human interpretations that would require adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    No MRMC comparative effectiveness study was done. This device is an in vitro diagnostic tool that provides numerical parameters describing blood hemostasis. It is not an AI-assisted diagnostic imaging device or a tool designed to directly improve human reader performance for tasks like lesion detection in images. The output is a graphical tracing and a table of numerical values for R, LY30, and MA, which clinicians then interpret.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    The performance studies described are standalone (algorithm/device only) in nature. The TEG® 6s Hemostasis System automatically measures and outputs parameters without human-in-the-loop diagnostic assistance within the measurement process itself. The interpretation of these results for patient diagnosis is then performed by a professional. The studies evaluate the device's ability to accurately and precisely perform these measurements.

    7. The Type of Ground Truth Used

    The types of "ground truth" used are:

    • Defined Reference Ranges: Established from "normal donors" (i.e., empirically derived from a healthy population).
    • Known Sample Characteristics: For precision and interference studies, samples were classified as Hypo, Normal, Hyper, or specifically manipulated (spiked with drugs/reagents like Dabigatran, Warfarin, Kaolin, RiaStap, tPA) to represent known coagulation states.
    • Predicate Device Measurements: For method comparison, the measured values from the legally marketed Thrombelastograph® Coagulation Analyzer (TEG®) – 5000 served as the reference for determining substantial equivalence.

    8. The Sample Size for the Training Set

    The document does not specify a separate "training set" in the context of machine learning. The TEG® 6s Hemostasis System is described as a measurement device based on physical principles (resonant frequency, modulus of elasticity) rather than a machine learning algorithm that is "trained" on data. Therefore, the concept of a separate training set, as typically defined in AI/ML, is not applicable or discussed here. The reference ranges and performance characteristics are established via analytical studies on various sample types.

    9. How the Ground Truth for the Training Set Was Established

    As explained in point 8, there is no explicit "training set" in the AI/ML sense. The device's operational characteristics and parameter calculations are based on established biophysical principles of coagulation, not on machine learning from a labeled training dataset.

    Ask a Question

    Ask a specific question about this device

    K Number
    K160502

    Validate with FDA (Live)

    Date Cleared
    2017-04-19

    (421 days)

    Product Code
    Regulation Number
    864.5425
    Age Range
    2 - 12
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TEG 6s System is intended for in vitro diagnostic use to provide semi-quantitative indications of the hemostasis state of a blood sample. The TEG 6s System records the kinetic changes in a venous sample of 3.2% citrated whole blood as the sample clots, and retracts in real time. The system output consists of a table of numerical values for parameters R, K, Angle, MA, and FLEV. The TEG 6s System provides specific blood modifiers, in the form of reagents dried-in-place within TEG 6s cartridges.

    Results from the TEG 6s analysis should not be the sole basis for a patient diagnosis, but should be evaluated together with the patient's medical history, the clinical picture and, if necessary, further hemostasis tests.

    The indication for use for the TEG 6s System is with adult patients where an evaluation of their blood hemostasis properties is desired. Hemostasis evaluations are commonly used to assess clinical conditions in cardiovascular surgery and cardiology procedures to assess hemorrhage or thrombosis conditions before, during and following the procedure.

    Device Description

    The TEG6s analyzer is a four-channel in vitro diagnostic instrument with an integrated computer module, a display touch screen for operator interaction, and a slot for inserting one TEG 6s cartridge. The TEG 6s analyzer is for use in laboratories and near patient use. It consists of an assembly of controllers, sensors, and displays, all managed and sequenced by a central processor. The embedded programming in the processor provides the necessary information for the automation of hemostasis testing. The program sequences the instrument hardware through the appropriate cycles to perform the test. Using a compressor, pressure and vacuum sensors, and a series of valves, actuators and controls, blood samples are introduced into the microfluidics of a disposable cartridge.

    To perform a test, a disposable cartridge is inserted into the instrument. The instrument reads the bar code on the cartridge, identifies the type of cartridge for operator confirmation. Then, the operator adds a blood sample to the entry port on the cartridge and uses the touch screen to issue the command to the instrument to proceed with the test. The sample is then drawn into the cartridge under instrument control. The amount of the sample drawn into the cartridge is automatically determined by the volume of the reagent chambers in the cartridge.

    The TEG 6s analyzer firmware provides features for and manages all functions of the instrument, including user interface (via the touch display screen) and external communications for service and installation via Service Maintenance Software (SMS). SMS is run on a separate computer connected to one or more TEG 6s analyzer(s) via the Ethernet port and a router. Its purpose is to allow additional control of analyzer functions by authorized remote users, such as administrative and service personnel.

    The TEG Manager 2.0.0 is an optional accessory to Haemonetics TEG 6s Hemostasis System. TEG Manager 2.0.0 is an application that provides remote viewing of current and historical patient tracing and test results created by the TEG 6s analyzers, and administration of all connected TEG 6s devices. TEG Manager interfaces with the TEG analyzers to obtain clinical data and retrieves patient information from external Hospital Information System (HIS). Users cannot manipulate the data that is stored and displayed within TEG Manager or input any additional clinical data in the software.

    AI/ML Overview

    The provided document is a 510(k) summary for the Haemonetics TEG® 6s Hemostasis System and TEG Manager 2.0.0. The primary purpose of this submission is to demonstrate substantial equivalence to previously cleared predicate devices, specifically the TEG 6s Hemostasis System (K140893 and K150041), with the addition of TEG Manager 2.0.0 as an optional accessory.

    Here's the breakdown of the requested information based on the document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of specific numerical acceptance criteria (e.g., sensitivity, specificity, accuracy targets) for the TEG Manager 2.0.0. Instead, the acceptance criterion for the software is described in a more general sense:

    Acceptance Criteria (General)Reported Device Performance
    TEG Manager 2.0.0 Software Functionality: The software must be thoroughly verified and validated at different levels (unit, system/integration, verification, validation, beta testing)."The TEG Manager 2.0.0 software was thoroughly verified and validated via test standards. The overall approach to testing was to test the software at different levels, including unit, system/integration, verification, validation and beta testing."
    Test Case Results: Acceptance criteria for verification and validation testing were identified in individual test cases as expected results. Test cases were comprised of scenarios with multiple steps along with their respective expected results."The acceptance criteria for verification and validation testing were identified in the individual test cases as the expected results. Test cases were comprised of scenarios with multiple steps along with their respective expected results. When all the steps were tested and the actual results matched the expected results, the test case result was passed." "Test data demonstrated that the device met all acceptance requirements, and that the subject device is as safe, as effective, and performs as well as or better than the predicate device." "The addition of TEG Manager 2.0.0 as an optional accessory to the TEG 6s Analyzer does not render the device non-substantially equivalent because non-clinical testing has demonstrated that the subject device is as safe and effective as the predicate and the results of verification and validation have not raised different questions of safety and effectiveness than the predicate."
    Substantial Equivalence: The device, with the optional accessory, must be as safe and effective and perform as well as or better than the predicate device, without raising new questions of safety or effectiveness."The data and information provided in the submission support a substantial equivalence determination for the TEG 6s Analyzer and the proposed TEG 6s Analyzer with TEG Manager as an optional accessory." The document implies that the TEG Manager, being primarily a remote viewer of existing data from the predicate device, does not introduce new analytical or interpretive data and thus maintains the safety and effectiveness of the system.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a "test set" in the context of clinical data for a direct performance evaluation of the TEG Manager 2.0.0 as an AI/ML device. The testing described is primarily software verification and validation (V&V). Therefore, there are no details on clinical sample size or data provenance (country of origin, retrospective/prospective) for a performance study. The focus is on ensuring the software functions as intended and accurately displays data from the TEG 6s Analyzer.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    Since the "test set" refers to software V&V rather than clinical performance data requiring expert adjudication, there is no mention of experts establishing ground truth or their qualifications. Software testing typically relies on predefined requirements and expected outputs rather than expert consensus on clinical findings.

    4. Adjudication Method for the Test Set

    No adjudication method (e.g., 2+1, 3+1) is mentioned because the V&V of the TEG Manager 2.0.0 does not involve clinical interpretation or a "ground truth" that would require expert consensus. The software's function is to accurately display data generated by the TEG 6s Analyzer.

    5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study is described. The TEG Manager 2.0.0 is an auxiliary viewing and management tool for data already generated by the TEG 6s Analyzer. It does not involve human readers interpreting images or data with and without AI assistance to assess an effect size.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    The TEG Manager 2.0.0 is described as an "optional accessory" that provides "remote viewing of current and historical patient tracing and test results created by the TEG 6s analyzers, and administration of all connected TEG 6s devices." It "does not provide any additional analytical or interpretive data outside of the TEG analyzer capabilities." Therefore, its performance is intrinsically tied to the TEG 6s Analyzer, and it does not operate as a standalone diagnostic algorithm in the way an AI/ML device typically would. Its "performance" would be related to its ability to accurately reflect the analyzer's output and manage devices, which is covered by the software V&V.

    7. Type of Ground Truth Used

    For the TEG Manager 2.0.0, the "ground truth" for the software validation would be the expected output values and display capabilities based on the data directly produced by the TEG 6s Analyzer and the specified software requirements. It's not clinical ground truth in the sense of pathology, outcomes data, or expert consensus on a diagnosis.

    8. Sample Size for the Training Set

    The document does not describe any training set as would be relevant for an AI/ML model. The TEG Manager 2.0.0 is a software application for data display and management, not a learning algorithm that undergoes training.

    9. How the Ground Truth for the Training Set Was Established

    As there is no training set for an AI/ML model, there is no discussion of how ground truth was established for such a set.