Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K211840

    Validate with FDA (Live)

    Device Name
    Sight OLO
    Date Cleared
    2022-05-09

    (329 days)

    Product Code
    Regulation Number
    864.5220
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Sight OLO is a quantitative multi-parameter automated hematology analyzer intended for in vitro diagnostic use in screening capillary or venous whole blood samples collection tubes, or fingertip samples collected using the Sight OLO test kit micro-capillary tubes.

    When used with the Sight OLO cartridge, the Sight OLO utilizes computer vision algorithms to enumerate the following CBC parameters in whole blood: WBC, RBC, HCT, MCV, MCH, MCHC, RDW, PLT, NEUT%/#, LYMPH %/#, MONO %/#, EOS%/#, and BASO%/#.

    The Sight OLO is indicated for use by clinical laboratories to identify and classify one or more of the formed elements of blood in children 3 months and above, adolescents and adults.

    Device Description

    The Sight OLO device is a computer vision based platform for blood analysis. The platform combines computer-vision algorithms for image processing to identify and quantify blood components (e.g., red blood cells) and their characteristics (e.g., cell volume) in an automated fashion. Using dedicated staining, the proposed platform provides complete blood count analysis. The Sight OLO is a compact device, designed to be automated and simple to operate, to enable rapid testing and analysis. The Sight OLO consists of a scanning and analyzing device and a CBC test kit, including a disposable cartridge and sample preparation tools. The disposable cartridge containing the blood sample is loaded into the device through the loading slot. The device is operated through the touch screen interface.

    The Sight OLO provides complete blood count information with 5-part differentials for white blood cell types. Specifically, the CBC parameters measured by the Sight OLO are listed below and include: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, PLT, NEUT%/#, LYMPH %/#, MONO %/#, EOS%/# and BASO%/#. In addition, the Sight OLO signals specific WBC abnormal cases by flagging the sample.

    AI/ML Overview

    The original text provided is a 510(k) Premarket Notification from the FDA regarding the Sight OLO device. It details the device's technical specifications, indications for use, and performance data used to establish substantial equivalence to a predicate device.

    Here's an analysis of the acceptance criteria and study data based on the provided text:

    Acceptance Criteria and Reported Device Performance

    The core of the acceptance criteria in this submission appears to be demonstrating substantial equivalence to a predicate device (Sight OLO K190898) and a reference device (Sysmex XN-Series Hematology Analyzer, K112605) for various blood parameters and flagging capabilities. The performance is assessed primarily through method comparison studies and flagging studies.

    Table of Acceptance Criteria (Implied) and Reported Device Performance:

    The document explicitly states that "all measurands met the prespecified acceptance criteria for correlation, bias, slope, intercept (and the 95% two-sided confidence interval (CI) around the slope and intercept)" for the method comparison. For flagging, it states the "overall flagging capabilities of the Sight OLO device met the predefined acceptance criteria for both sensitivity and specificity." While the exact numerical criteria for each parameter's correlation, slope, intercept, and bias are not explicitly listed as "acceptance criteria" but rather the results that met them, the table below reflects the reported performance that demonstrates meeting these criteria.

    Metric (Implied Acceptance Criteria)Device ParameterReported Performance (Result that met acceptance)
    Method Comparison
    Correlation Coefficient (r) (High r expected)WBC0.997
    RBC0.991
    PLT0.984
    HGB0.990
    HCT0.983
    MCV0.941
    RDW0.941
    MCH0.976
    MCHC0.687
    NEUT%0.988
    NEUT#0.996
    LYMPH%0.991
    LYMPH#0.995
    MONO%0.926
    MONO#0.947
    EOS%0.978
    EOS#0.980
    BASO%0.658
    BASO#0.646
    Slope (Close to 1 expected, 95% CI covering 1)Most parameterse.g., WBC: 1.016 (1.008, 1.024)
    Intercept (Close to 0 expected, 95% CI covering 0)Most parameterse.g., WBC: 0.014 (-0.025, 0.067)
    Median Bias (Low bias expected)All parameterse.g., WBC: 0.11
    Relative Bias (%) (Low bias expected)All parameterse.g., WBC: 1.92%
    Flagging Capability
    Sensitivity (PPA) (High expected)Overall Flagging91.0%
    Specificity (NPA) (High expected)Overall Flagging92.6%
    Overall Agreement (High expected)Overall Flagging91.8%

    Note on MCHC, BASO% and BASO# Correlation: The correlation coefficients for MCHC, BASO%, and BASO# are notably lower than other parameters (0.687, 0.658, 0.646 respectively). However, the document states "all measurands met the prespecified acceptance criteria," implying these values were acceptable for the purpose of demonstrating substantial equivalence. This could be due to factors like the analytical measuring range of these parameters or inherent variability.


    Study Details:

    1. Sample sizes used for the test set and the data provenance:

      • Method Comparison Study:

        • Sample Size: A total of 700 residual clinical K2EDTA whole blood samples. The 'N' column in the method comparison results table shows slight variations (e.g., 662 for WBC, 674 for RBC), indicating samples might have been excluded for specific parameter analysis for various reasons (e.g., insufficient volume, issues during analysis).
        • Data Provenance:
          • Country of Origin: Three (3) US sites.
          • Retrospective or Prospective: The samples were "residual clinical" samples and re-run with the updated algorithm, suggesting they were previously collected, indicating a retrospective approach to the sample acquisition for the re-run study, though the initial collection for K190898 might have been prospective. The text says, "The samples previously collected in K190898 were re-run with the updated algorithm of the subject device."
          • Sample Characteristics: Included normal and pathological samples (e.g., acute inflammation, various anemias, leukemias etc.). Covered an age range of 3 months to 94 years old, with 32% pediatric samples (3M-21Y). 365 males (52%) and 335 females (48%).
      • Flagging Study:

        • Sample Size: Over 200 samples.
        • Data Provenance:
          • Country of Origin: 3 clinical study sites (location not explicitly stated but implied US given other elements of the submission).
          • Retrospective or Prospective: "The samples previously collected in K190898 were re-run with the updated algorithm of the subject device," indicating a retrospective re-analysis with the new algorithm.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Flagging Study (Ground Truth for Flagging):
        • Number of Experts: Two primary qualified morphology examiners (Reader A and Reader B), with a third qualified morphology examiner (arbitrator) used in case of disagreement. So, a total of 3 experts potentially involved for each discordant case.
        • Qualifications: Referred to as "qualified morphology examiners." Specific experience in years or board certification is not detailed, but their "qualification" is stated.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Flagging Study (for Ground Truth of Manual Microscopy): A 2+1 adjudication method was used. "Two qualified morphology examiners evaluated one of the three blood films... The third blood film was saved for reading by a third qualified morphology examiner (i.e., arbitrator) in the event that there was disagreement between Reader A and Reader B."
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No MRMC comparative effectiveness study involving human readers improving with AI assistance is described in this document. The studies presented are primarily a comparison of the device's performance to a reference device (Sysmex) and to manual microscopy (for flagging), not a human-AI teamed performance study. The device is an automated analyzer, not an AI-assisted human reading tool in the sense of an MRMC study.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Yes, the performance data presented for the Sight OLO device is for its standalone performance. The device is described as an "automated hematology analyzer" that "utilizes computer vision algorithms to enumerate" parameters. The listed performance metrics (correlation, bias, sensitivity, specificity) reflect the device's output independently.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • Method Comparison Study: The ground truth was established by a reference device, the Sysmex XN-Series Hematology Analyzer. This is a common method for validating new automated hematology analyzers, where an established, cleared device serves as the comparator.
      • Flagging Study: The ground truth was established by expert consensus (manual light microscopy combined with adjudication by "qualified morphology examiners"). This is often considered the gold standard for morphological assessment of blood cells.
    7. The sample size for the training set:

      • The document does not specify the sample size for the training set. It focuses on the performance data for the updated algorithm applied to previously collected samples. The exact details of the original training data for the computer vision algorithms are not provided within this summary for this specific 510(k).
    8. How the ground truth for the training set was established:

      • The document does not explicitly state how the ground truth for the training set was established. It describes the "minor modifications to analysis algorithms" made to increase actionable results and improve flagging specificity/reduce invalidation. It also mentions that "the device is a computer vision based platform for blood analysis." However, the methodology for establishing the ground truth for the training of these computer vision algorithms is typically proprietary and not detailed in this type of summary. It is generally assumed to involve expert-labeled data, similar to the "qualified morphology examiners" used for the flagging study's ground truth, but this is not confirmed in the provided text for the training set itself.
    Ask a Question

    Ask a specific question about this device

    K Number
    K190898

    Validate with FDA (Live)

    Device Name
    Sight OLO
    Date Cleared
    2019-11-01

    (210 days)

    Product Code
    Regulation Number
    864.5220
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Sight OLO is a quantitative multi-parameter automated hematology analyzer intended for in vitro diagnostic use in screening capillary or venous whole blood samples collected in KyEDTA blood collection tubes, or fingertip samples collected using the Sight OLO test kit micro-capillary tubes.

    When used with the Sight OLO cartridge, the Sight OLO enumerates the following CBC parameters in whole blood: WBC, RBC, HGB, HCT, MCV, MCH, RDW, PLT, NEUT%#, LYMPH %#, MONO %#, EOS%#, and BASO%/#.

    The Sight OLO is indicated for use in clinical laboratories to identify and classify one or more of the formed elements of blood in children 3 months and above, adolescents and adults.

    Device Description

    The Sight OLO device is a computer vision based platform for blood analysis. The platform combines computer-vision algorithms for image processing to identify and quantify blood components (e.g., red blood cells) and their characteristics (e.g., cell volume) in an automated fashion. Using dedicated staining, the proposed platform provides complete blood count analysis. The Sight OLO is a compact device, designed to be automated and simple to operate, to enable rapid testing and analysis. The Sight OLO consists of a scanning and analyzing device and a CBC test kit, including disposable cartridges and sample preparation tools. The disposable cartridge containing the blood sample is loaded into the device through the loading slot. The device is operated through the touch screen interface.

    The Sight OLO provides complete blood count information with 5-part differentials for white blood cell types. Specifically, the CBC parameters measured by the Sight OLO are listed below and include: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, PLT, NEUT%#, LYMPH %#, MONO %/#, EOS%# and BASO%#. In addition, the Sight OLO signals specific WBC abnormal cases by flagging the sample.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Study CategoryMeasurandAcceptance CriteriaReported Device Performance
    Method ComparisonWBC, RBC, PLT, HGB, HCT, MCV, RDW, MCH, MCHC, NEUT%, NEUT#, LYMPH%, LYMPH#, MONO%, MONO#, EOS%, EOS#, BASO%, BASO#Correlation, bias, slope, and intercept (and their 95% two-sided confidence intervals) met pre-specified criteria.All measurands met the pre-specified acceptance criteria for correlation, bias, slope, intercept (and the 95% two-sided confidence interval (CI) around the slope and intercept). (See Table 1 for specific values for each measurand).
    RepeatabilityAll measurandsAll measurands in all tested ranges met predefined acceptance criteria for mean, standard deviation (SD), and coefficient of variation (CV).All measurands in all tested ranges met the predefined acceptance criteria. (See Table 2 for specific values for each measurand across target ranges).
    ReproducibilityAll measurandsAll components of variation (within-run, between-run, between-day, between-instrument, between-site, total precision) met predefined acceptance criteria for SD and %CV.All components of variation that were calculated met the pre-defined acceptance criteria. (See Table 3 for specific values for each measurand across low, normal, and high levels).
    Detection Limits (LoB, LoD, LoQ)WBC, PLT, HGBLoB defined as 95th percentile of study variable. LoD determined mathematically. LoQ defined as lowest concentration (≥LoD) where total error (TE) accuracy goals were satisfied. Test results met these definitions.WBC: LoB = 0.04 x 10³/μL, LoD = 0.17 x 10³/μL, LoQ = 0.18 x 10³/μLPLT: LoB = 8.00 x 10³/μL, LoD = 11.00 x 10³/μL, LoQ = 13.40 x 10³/μLHGB: LoB = 0 g/dL, LoD = 3.9 g/dL, LoQ = 3.9 g/dL. (All met acceptance criteria).
    LinearityHGB, PLT, WBCThe final linearity range was determined as the intersection of the results from three instruments, and no lower than the LoD.HGB: (3.9-21.75) g/dLPLT: (18-1028.5) x 10³/μLWBC: (0.18-100.13) x 10³/μL. (All met acceptance criteria).
    Analytical Specificity/InterferenceRBC, WBC, HGB, HCT, PLTNo significant interference for D-Glucose, Bilirubin F, Bilirubin C, Chyle, Hemolytic Hemoglobin, Intralipids at specified concentrations. No interference from RBC fragments, high leukocytosis, and high thrombocytosis.Results demonstrated no significant interference from D-Glucose, Bilirubin F (except for WBC above 8.86 mg/dL), Bilirubin C, Chyle (except for PLT above 530 FTU), Hemolytic Hemoglobin, and Intralipids within tested concentrations. No interference from abnormal specimen conditions.
    Sample StabilityAll measurandsSupports a sample stability claim of 8 hours from blood collection at room temperature, with RDW specifically 4 hours.Data support a sample stability claim of 8 hours from blood collection at room temperature for all Sight OLO measurands, with RDW having a stability of 4 hours.
    Test Kit Shelf-LifeFunctional and analytical testsMet predefined acceptance criteria at defined time points. Initial shelf life of 6 months.Functional and analytical tests were conducted and results met the predefined acceptance criteria. Initial shelf life set at 6 months. Ongoing study with 3 lots (4, 8, 12, 18, 24, 26, 30, 36, 38 months).
    Transportation StabilityDevice and Test KitDevice maintains calibration and functional specifications. Test kit met all acceptance criteria for functionality and expected analytical results.The Sight OLO device maintains its calibration and functional specifications. The Sight OLO test kit met all acceptance criteria for functionality and expected analytical results.
    Flagging StudySensitivity and SpecificityMet predefined acceptance criteria.Sensitivity (PPA): 93.0%Specificity (NPA): 80.6%Overall Agreement: 86.5% (Met predefined acceptance criteria).
    Adult and Pediatric Reference IntervalsAll measurandsAdults: Reference intervals calculated for each measurand. Pediatrics: Supports validity of pre-established pediatric reference intervals.Adult reference intervals were calculated. Results supported the validity of pre-established pediatric reference intervals.
    Matrix Comparison (Capillary vs. Venous)All applicable measurandsComparable performance characteristics for capillary and venous whole blood specimens.The results show comparable performance characteristics for capillary and venous whole blood specimens.
    Matrix Comparison (Capillary Microtube vs. Sight OLO Test Kit Micro-capillaries)All applicable measurandsComparable performance between K2EDTA anticoagulated microtube capillary samples and 2-drop fingertip samples.The results of the regression and bias analysis showed comparable performance between the K2EDTA anticoagulated microtube capillary samples and the 2-drop fingertip samples.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Method Comparison Study: 679 residual clinical K₂EDTA whole blood samples.
      • Provenance: Collected from both adults (≥22 years old) and pediatric patients (3 months to 21 years old) at three (3) US sites. The majority were venous samples; a few pediatric samples were capillary whole blood. The study included normal and pathological samples, covering an age range of 3 months to 94 years old. 32% were pediatric samples. Consisted of 355 males (52%) and 324 females (48%). This was a retrospective study using residual clinical samples.
    • Repeatability Study: Minimum of 11 samples per site, covering the laboratory reference range and medical decision levels for HGB, PLT, WBC, and upper range for RBC, HGB, WBC, and PLT.
      • Provenance: Residual K₂EDTA whole blood samples. Performed at 3 US sites.
    • Reproducibility Study: 3 lots of commercial control material (low, normal, high concentrations). For a total of 240 measurements per level of control material (presumably 4 replicates x 2 runs/day x 5 days x 2 instruments/site x 3 sites, though the exact breakdown to reach 240/level is not fully detailed).
      • Provenance: Commercial control material. Conducted at 3 sites, including 2 US sites.
    • Detection Limits (LoB, LoD, LoQ):
      • LoB: Preserved RBC samples (WBC and PLT), diluent (HGB). Two Sight OLO instruments, two reagent lots, 60 measurements per instrument.
      • LoD and LoQ: Six low concentration samples (manipulated concentrated and diluted venous blood samples), each measured 10 times, for a total of 60 repeated measurements per test. This was repeated for a different test reagent lot (another 60 measurements).
      • Provenance: Manipulated venous blood samples for LoD/LoQ.
    • Linearity Studies: Ten (10) venous whole blood samples manipulated to create linearity panels. Seven concentrations for WBC and PLT, ten for RBC and HGB. Each concentration scanned in duplicate. Repeated on three OLO devices.
    • Analytical Specificity/Interference Study: Not explicitly stated as a sample size of patient samples, but implied various samples manipulated with interferents.
    • Sample Stability Study: 10 freshly collected venous whole blood samples (7 normal, 3 around medical decision levels).
    • Test Kit Shelf-Life: Three lots of test kits.
    • Transportation Studies: Not specified in terms of sample size, but indicates testing of the physical device and test kits.
    • Flagging Study: 208 samples.
      • Provenance: Collected at 3 clinical study sites.
    • Adult Reference Intervals Study: 240 (120 male and 120 female) apparently healthy adults.
      • Provenance: K₂EDTA venous whole blood samples collected from apparently healthy adults (≥22 years) at a single US site. This was a prospective study.
    • Pediatric Reference Intervals Study: 80 apparently healthy pediatric subjects (20 per subpopulation: baby (3-23 months), child (2-<12 years), adolescent (12-<18 years), transitional adolescent (18-<22 years)).
      • Provenance: Residual blood samples collected from apparently healthy pediatric subjects. This was a retrospective study.
    • Capillary vs. Venous Matrix Comparison: 67 normal and pathological paired capillary and venous whole blood specimens.
      • Provenance: Paired specimens collected from the same individuals.
    • Capillary Microtube vs. Sight OLO Test Kit Micro-capillaries Matrix Comparison: 40 paired K₂EDTA anticoagulated capillary microtube samples and 2-drop finger prick samples.
      • Provenance: Collected from healthy volunteer subjects at the company's internal laboratory.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    • Flagging Study:
      • Number of Experts: Two qualified morphology examiners (initially). A third qualified morphology examiner was used for adjudication.
      • Qualifications: "Qualified morphology examiners" (specific experience level not provided in the summary).

    4. Adjudication Method for the Test Set:

    • Flagging Study: A 2+1 adjudication method was used. Two qualified morphology examiners evaluated one of the three blood films for each sample. In case of disagreement between the first two readers (who performed a 200-cell count each, totaling 400 cells), a third qualified morphology examiner reviewed the third blood film.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No MRMC comparative effectiveness study was done. The studies compared the Sight OLO device's performance against a predicate device (Sysmex XN-Series) or manual microscopy (for flagging), but not a direct comparison of human readers with and without AI assistance. The study focuses on the performance of the device itself.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

    • Yes, the performance studies described (Method Comparison, Repeatability, Reproducibility, Detection Limits, Linearity, Analytical Specificity, Stability) are all standalone performance evaluations of the Sight OLO device. It tests the accuracy and precision of the automated hematology analyzer without human intervention in the CBC parameter enumeration process.
    • The Flagging Study involves a comparison to manual light microscopy, which is a human-read method, to establish agreement in flagging abnormal samples. While this involves human input for comparison, the Sight OLO's flagging itself is an automated algorithm-only output being evaluated.

    7. Type of Ground Truth Used:

    • Method Comparison: Predicate device (Sysmex XN-Series) results.
    • Repeatability/Reproducibility/Detection Limits/Linearity/Analytical Specificity/Stability: Internal reference methods, commercial control materials, or manipulated samples (e.g., concentrated/diluted blood samples), and comparison to expected analytical results.
    • Flagging Study: Manual light microscopy by qualified morphology examiners.
    • Adult and Pediatric Reference Intervals: Literature reference limits (for verification in pediatric study), and statistical calculation from healthy population samples.
    • Matrix Comparison: Paired samples compared to each other on the OLO device.

    8. Sample Size for the Training Set:

    • This information is not provided in the given 510(k) summary. The document focuses on performance testing (validation) datasets.

    9. How the Ground Truth for the Training Set Was Established:

    • This information is not provided in the given 510(k) summary, as it does not detail the development or training of the device's algorithms.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1