Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K180921

    Validate with FDA (Live)

    Date Cleared
    2018-06-28

    (80 days)

    Product Code
    Regulation Number
    870.5100
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Sapphire® II PRO balloon dilatation catheter (1.0-1.25mm configurations) is indicated for: balloon pre-dilatation of a stenotic portion of a coronary artery or bypass graft stenosis (≥70% stenosis) for the purposes of improving myocardial perfusion. The Sapphire® II PRO balloon dilatation catheter (1.5-4.0mm configurations) is indicated for: balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis in patients evidencing coronary ischemia for the purpose of improving myocardial perfusion. balloon dilatation of a coronary artery occlusion for the treatment of acute myocardial infarction. The Sapphire® II PRO balloon dilatation catheter is also indicated for: percutaneous transluminal angioplasty in the peripheral vasculature, including renal, femoral, popliteal, infra-popliteal, tibial, and peroneal arteries.

    Device Description

    The Sapphire II PRO balloon dilatation catheter is a rapid exchange balloon catheter with a working length of 140cm design for both coronary and peripheral indications. This catheter was previously cleared for coronary use (K173680, K163114) and this submission is intended to support the peripheral use of this balloon catheter. The proximal shaft is a PTFE coated stainless steel hypotube. Hydrophilic lubricious coatings are applied to the distal section. The semi-compliant balloons, available in diameters from 1.0-4.0mm and lengths from 5-30mm, can be inflated by injecting dilute contrast media solution through the trailing hub of the catheter. Two radiopaque platinum marker bands are located within the balloon segment (there is only one centrally located marker band for the Ø1.0-1.5mm configurations). The catheter is compatible with 4F or larger guiding sheaths or 5F or larger guiding catheters. The internal lumen of the catheter accepts a standard 0.014 inch guidewire. The proximal part of the guidewire enters the catheter tip and advances coaxially out the catheter proximal port, thereby allowing both coaxial guidance and rapid exchange of catheters with a single standard length guidewire. Two marked sections are located on the hypotube shaft to indicate catheter position relative to the tip of either a guiding sheath or guiding catheter. The design of this dilatation catheter does not incorporate a lumen for distal dye injections or distal pressure measurements.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the Sapphire II PRO Balloon Dilatation Catheter, focusing on the peripheral use indication:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly list acceptance criteria in a quantitative table format alongside performance data. Instead, it states that the device "met all acceptance criteria" and "were similar to the reference devices" for the performed tests.

    The performance tests conducted include:

    TestReported Device Performance
    SterilizationComplete (leveraged from reference device K173680, K163114)
    Shelf-LifeComplete (leveraged from reference device K173680, K163114)
    Dimensional VerificationMet acceptance criteria (partially leveraged)
    Balloon Rated Burst PressureMet acceptance criteria (partially leveraged)
    Shaft BurstMet acceptance criteria (partially leveraged)
    Balloon FatigueMet acceptance criteria (partially leveraged)
    Balloon ComplianceMet acceptance criteria (partially leveraged)
    Balloon Inflation and Deflation TimeMet acceptance criteria (partially leveraged)
    Catheter Bond StrengthMet acceptance criteria (partially leveraged)
    Tip Pull StrengthMet acceptance criteria (partially leveraged)
    Flexibility and KinkingMet acceptance criteria (partially leveraged)
    Torque StrengthMet acceptance criteria (partially leveraged)
    RadiopacityMet acceptance criteria (partially leveraged)
    Coating IntegrityMet acceptance criteria (partially leveraged)
    Visual Inspection (for peripheral use)Met acceptance criteria (additional testing performed)
    Balloon Preparation, Deployment, and Retraction (for peripheral use)Met acceptance criteria (additional testing performed)

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not provide specific sample sizes for any of the tests. It refers to testing being "leveraged from the reference device (K173680, K163114)" for many criteria, and "additional testing was performed" for others.

    There is no information on the country of origin of the data, and the studies are retrospective in the sense that existing data from previous clearances (K173680, K163114) were used, supplemented by new (implicitly prospective) tests for the peripheral use indication.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This document describes a medical device clearance for a balloon catheter, not an AI/software device. Therefore, the concept of "experts" establishing ground truth in the context of clinical imagery or diagnostic data is not applicable here. The ground truth for device performance is established through engineering and performance testing against predetermined specifications.

    4. Adjudication Method for the Test Set

    Not applicable for this type of medical device performance testing. Adjudication methods like "2+1" or "3+1" are relevant for expert review of clinical cases, often in AI development contexts.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Its Effect Size

    No, an MRMC comparative effectiveness study was not done. This is a physical medical device, not an AI system being evaluated for diagnostic or interpretive tasks by human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This is not an algorithm or AI system.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance is based on engineering specifications and established performance standards for balloon dilatation catheters. These standards are likely derived from industry best practices, regulatory guidance, and clinical requirements for the safe and effective operation of such devices. For example, balloon burst pressure needs to meet a certain PSI, and shaft flexibility needs to fall within a defined range.

    8. The Sample Size for the Training Set

    Not applicable. This document pertains to a physical medical device, not a machine learning model, so there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1