Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K190606

    Validate with FDA (Live)

    Date Cleared
    2019-05-28

    (81 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. These gloves were tested for use with chemotherapy drugs and Fentanyl Citrate as per ASTM D6978-05 (Reapproved 2013) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

    Device Description

    Powder Free Nitrile Patient Examination Glove, Black Colored, Non-Sterile, Low Dermatitis Potential. Tested for Use with Chemotherapy Drugs and Fentanyl Citrate.

    AI/ML Overview

    This document describes the acceptance criteria and the study results for Kossan International Sdn. Bhd.'s Powder Free Nitrile Patient Examination Gloves.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the gloves are based on their resistance to permeation by chemotherapy drugs and Fentanyl Citrate, as determined by ASTM D6978-05 (Reapproved 2013). The reported device performance is the "Minimum Breakthrough Detection Time in Minutes" for each substance.

    Chemotherapy Drug and ConcentrationAcceptance Criteria (Minimum Breakthrough Detection Time in Minutes)Reported Device Performance (Minimum Breakthrough Detection Time in Minutes)
    Carmustine (BCNU) (3.3 mg/ml)(Not explicitly defined, but a longer time is better)25.2
    Cisplatin (1.0 mg/ml)(Not explicitly defined, but a longer time is better)>240
    Cyclophosphamide (Cytoxan) (20.0 mg/ml)(Not explicitly defined, but a longer time is better)>240
    Cytarabine (100 mg/ml)(Not explicitly defined, but a longer time is better)>240
    Dacarbazine (DTIC) (10.0 mg/ml)(Not explicitly defined, but a longer time is better)>240
    Doxorubicin Hydrochloride (2.0 mg/ml)(Not explicitly defined, but a longer time is better)>240
    Etoposide (20.0 mg/ml)(Not explicitly defined, but a longer time is better)>240
    Fluorouracil (50.0 mg/ml)(Not explicitly defined, but a longer time is better)>240
    Ifosfamide (50.0 mg/ml)(Not explicitly defined, but a longer time is better)>240
    Methotrexate (25.0 mg/ml)(Not explicitly defined, but a longer time is better)>240
    Mitomycin C (0.5 mg/ml)(Not explicitly defined, but a longer time is better)>240
    Mitoxantrone (2.0 mg/ml)(Not explicitly defined, but a longer time is better)>240
    Paclitaxel (Taxol) (6.0 mg/ml)(Not explicitly defined, but a longer time is better)>240
    Thiotepa (10.0 mg/ml)(Not explicitly defined, but a longer time is better)35.9
    Vincristine Sulfate (1.0 mg/ml)(Not explicitly defined, but a longer time is better)>240
    Fentanyl Citrate Injection (100 mcg/2ml)(Not explicitly defined, but a longer time is better)>240
    Simulated Gastric Acid Fluid(Not explicitly defined, but a longer time is better)>240

    Note: While specific numerical "acceptance criteria" are not explicitly stated as pass/fail thresholds in the document for each drug, the implication of conducting the test and reporting breakthrough times is that these values met the manufacturer's internal standards and/or the expectations for substantial equivalence based on the ASTM D6978-05 standard. The goal is to maximize the breakthrough time for safety.

    2. Sample size used for the test set and the data provenance

    The document does not specify the exact sample size (number of gloves or trials) used for the permeation tests. It only states that the gloves "were tested for use with chemotherapy drugs and Fentanyl Citrate as per ASTM D6978-05 (Reapproved 2013) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs." Following an ASTM standard typically implies a specific number of replicates, but this detail is not provided.

    The data provenance is from prospective testing conducted by the manufacturer according to the specified ASTM standard. The country of origin of the data is not explicitly stated, but the manufacturer, Kossan International Sdn. Bhd., is based in Malaysia.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This is not applicable to this type of device and study. The "ground truth" for glove permeation is established by objective laboratory measurements using analytical techniques to detect the breakthrough of chemical substances, not by expert interpretation.

    4. Adjudication method for the test set

    This is not applicable. Adjudication methods are typically used in studies involving human interpretation or clinical endpoints. For a technical performance test like chemical permeation, the results are objective measurements.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. This document pertains to the performance of examination gloves against chemical permeation, not a diagnostic device involving human readers or AI.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This is not applicable. This document pertains to the performance of examination gloves against chemical permeation, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth used is objective laboratory measurement (detection of chemical permeation through the glove material) based on the methodology outlined in ASTM D6978-05.

    8. The sample size for the training set

    This is not applicable. The device is a physical product (gloves) and does not involve a "training set" in the context of machine learning or AI algorithms.

    9. How the ground truth for the training set was established

    This is not applicable for the same reason as above (physical product, no training set).

    Ask a Question

    Ask a specific question about this device

    K Number
    K143131

    Validate with FDA (Live)

    Date Cleared
    2015-02-12

    (104 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

    Device Description

    Powder Free Nitrile Patient Examination Glove, Black Colored, Non-Sterile is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

    The gloves are powder free, ambidextrous with beaded-cuff, black colored, single-use disposable devices that come in six sizes (XS, S, M, L, XL, XXL), and supplied in Non-Sterile state.

    The gloves are made of Nitrile Butadiene Rubber, designed and manufactured in accordance with ASTM D6319-10, Standard Specification for Nitrile Examination Gloves for Medical Application.

    Its physical and performance characteristics meet all requirements of ASTM D6319-10.

    AI/ML Overview

    The provided document describes the acceptance criteria and performance of the Kossan International Sdn. Bhd. Powder Free Nitrile Patient Examination Glove, Black Colored, Non-Sterile (K143131). This is a medical device, and the "study" referred to is the non-clinical testing performed to demonstrate substantial equivalence to a predicate device.

    Here's the information extracted and organized as requested:


    1. Table of Acceptance Criteria and Reported Device Performance

    CharacteristicAcceptance Criteria (Standards Requirements)Reported Device Performance (Results Summary)
    DimensionsASTM D6319-10 (Length $\ge$ 230mm, Palm Thickness $\ge$ 0.05mm, Finger Thickness $\ge$ 0.05mm, Specific Width ranges for XS, S, M, L, XL, XXL)Meets Standard Requirements (Explicit values not provided in summary, but stated agreement with standard)
    Physical PropertiesASTM D6319-10 (Before Aging: Tensile Strength $\ge$ 14 MPA, Elongation $\ge$ 500%; After Aging: Tensile Strength $\ge$ 14 MPA, Elongation $\ge$ 400%)Meets Standard Requirements (Explicit values not provided in summary, but stated agreement with standard)
    Freedom from PinholesASTM D5151-11, ASTM D6319-10 (Pass quality level at G1 AQL 1.5)Tested in accordance with ASTM D5151 test method. Pass quality level at G1 AQL 1.5. Meets Standard Requirements.
    Powder Free ResidueASTM D6124-11, ASTM D6319-10 (Values $\le$ 2 mg of residual powder per glove)Result generated values $\le$ 2 mg of residual powder per glove. Meets Standard Requirements.
    BiocompatibilityDermal Sensitization (as ISO 10993-10:2010), Primary Skin Irritation Test (as ISO 10993-10:2010)Dermal Sensitization: Magnusson & Kligman Scale is '0'. Device is not a sensitizer. Meets Standard Requirements.Primary Skin Irritation: Primary Irritation Index for Erythema and Edema is '0'. Device is not an irritant. Meets Standard Requirements.

    2. Sample size used for the test set and the data provenance

    The document does not specify the exact sample sizes (number of gloves or animals/humans) used for each non-clinical test (dimensions, physical properties, pinholes, powder residue, biocompatibility). It only references adherence to ASTM and ISO standards, which would dictate appropriate sample sizes for each test.

    • Data Provenance: The tests were conducted by Kossan International Sdn. Bhd. (Malaysia), making the data provenance likely from Malaysia. The data is non-clinical/laboratory test data, not patient data, so "retrospective or prospective" is not an applicable distinction in this context.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable. The "ground truth" for this device's performance is established by objective measurements against recognized industry standards (ASTM, ISO) in a laboratory setting, not by expert consensus on clinical findings. Therefore, no experts were used to establish a clinical ground truth for the test set in the traditional sense of medical image analysis or diagnostic studies.


    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies where multiple human readers assess data (e.g., images), and discrepancies need to be resolved. For non-clinical, objective laboratory tests of physical properties and biocompatibility, the results are typically determined by measurements and observations according to established protocols, not by expert adjudication.


    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC comparative effectiveness study was done. This document pertains to the regulatory clearance of a physical medical device (examination gloves) based on non-clinical performance and biocompatibility data. It does not involve AI or human readers for diagnostic assessment.


    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable. This document is about a physical medical device, not an algorithm or AI system.


    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" is based on objective laboratory measurements and tests against predefined industry standards (ASTM D6319-10, ASTM D5151-11, ASTM D6124-11, ISO 10993-10:2010 for biocompatibility). For example:

    • Physical dimensions: measured values compared to specified ranges.
    • Physical properties (tensile strength, elongation): measured values compared to minimum requirements.
    • Freedom from pinholes: water leak test results compared to AQL.
    • Powder free residue: measured weight of residual powder compared to maximum limit.
    • Biocompatibility: animal study results (Magnusson & Kligman Scale, Primary Irritation Index) compared to "non-sensitizer" and "non-irritant" conclusions.

    8. The sample size for the training set

    This information is not applicable. This is not an AI/machine learning device. No training set was used.


    9. How the ground truth for the training set was established

    This information is not applicable. This is not an AI/machine learning device. No training set was used.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1